ultimate healthcare Tamora Plus II Digital User Guide

October 30, 2023
Ultimate Healthcare

ultimate healthcare Tamora Plus II Digital

ultimate healthcare Tamora Plus II Digital PRODUCT

Product Information: Tamora Plus II Digital

The Tamora Plus II Digital is a medical air mattress system designed for patients who are at risk of developing pressure ulcers. It is equipped with various features to provide comfort and prevent pressure sores.

ultimate healthcare Tamora Plus II Digital 03Comfort Control

The Soft and Firm buttons allow adjustment of pressures within a safe pre-set range, providing customized comfort to the patients.

.ultimate healthcare Tamora Plus II Digital 04When pressing the Firm button, the output pressure will increase to provide a higher pressure output and thus increased support.

  • When pressing the Soft button, the output pressure will be decreased to provider a lower pressure output and thus increased comfort.

Auto-Detection of Pressure Settings

The Tamora Plus II Digital automatically detects the patients’ weight to set internal pressures appropriately. It has three modes available such as Alternating, Static, and Constant Low Pressure. The mattress will revert back to Alternating mode if the patient moves. If the Constant Low Pressure mode is required, pressing the Soft button illuminates the mode, and the time is displayed on the LCD screen.

CPR

In case of a cardiac arrest or Cardio Pulmonary Resuscitation (CPR), the CPR valve located at the head end on the left-hand side of the mattress can be activated. To activate CPR, gently pull and rotate the dial of the CPR valve to click into the OPEN position while disconnecting the air pipe connector from the pump. The mattress deflates in 10-15 seconds. To reset CPR, ensure that the dial of the CPR attachment is rotated until it clicks into the CLOSED position.

In the event of a patient suffering a cardiac arrest and CPR (Cardio Pulmonary Resuscitation) needing to be administered:

To activate the CPR:

  1. CPR valve is located at the head end on the left hand side of the mattress.
  2. Gently pull and rotate the dial of the CPR valve to ‘click’ into the OPEN position. At the same time, disconnect the air pipe connector from the Pump to speed up the air release. The mattress deflates in 10-15 seconds.ultimate healthcare Tamora Plus II Digital 12

Transport Mode

The transport mode helps maintain the cells in their present state for up to 24 hours.

  • Simply disconnect air pipe connector and place the connector cover over it. The air pressures in the mattress will now equalise, but maintain a degree of comfort.
  •  It is important to restore the mattress as quickly as possible by reconnecting the supply tubes to the power unit.ultimate healthcare Tamora Plus II Digital 13

Alarms

The Tamora Plus II Digital has three types of alarms: Low Pressure, Power Failure, and Service alarms. The Alarm Mute button temporarily resets the audible alarms. The annual Service Indicator remains illuminated until serviced.

Alarm Mute
The Alarm Mute button temporarily resets the audible Low Pressure/Power Failure/Service alarms. Should the situation not be resolved and the fault condition continues the alarm will resume.

Low Pressure Alarm

  • Upon detection of low pressure, an audible alarm will be heard and the Low Pressure indicator will illuminate.
  •  The audible alarm may be cancelled by pressing the Alarm Mute button.

Power Failure Alarm

  •  If mains power is removed from power unit or the power cable is unplugged without turning the power unit OFF, an audible alarm will sound and the Power Failure indicator will illuminate.
  • The audible alarm may be cancelled by pressing the Alarm Mute button.

Service Alarm

  •  Should your system develop a fault condition whilst in use, an audible alarm will be heard and the Service indicator will illuminate.
  •  The audible alarm may be cancelled by pressing the Alarm Mute button.

Annual Service Indicator

  •  A dedicated Annual Service indicator on the control panel will illuminate and stay ON to alert nursing staff of the need for the system to be professionally serviced.
  • The Annual Service indicator will illuminate after 365 days of use.

Product Usage Instructions

  1. Refer to the product User Manual for full instructions for use.
  2. Adjust the Soft and Firm buttons to provide customized comfort to the patients.
  3. The Tamora Plus II Digital has three modes available: Alternating, Static, and Constant Low Pressure. Press the Soft button to illuminate Constant Low Pressure mode.
  4. In case of a cardiac arrest or CPR, activate the CPR valve located at the head end on the left-hand side of the mattress by gently pulling and rotating the dial to click into the OPEN position while disconnecting the air pipe connector from the pump. The mattress deflates in 10-15 seconds. To reset CPR, ensure that the dial of the CPR attachment is rotated until it clicks into the CLOSED position.
  5. Use the transport mode to maintain the cells in their present state for up to 24 hours.
  6. The Tamora Plus II Digital has three types of alarms: Low Pressure, Power Failure, and Service alarms. The Alarm Mute button temporarily resets the audible alarms. The annual Service Indicator remains illuminated until serviced.

ultimate healthcare Tamora Plus II Digital 01
Control Panel Lock Out
To unlock panel:

  • Press & hold for 3 seconds to unlock control panel.

To lock panel:

  • To lock out the control panel press the Lock Out button and all functions will be locked.
  • If the control panel is not used for a period of 30 seconds it will automatically lock out and inhibit the use of the functions.

ultimate healthcare Tamora Plus II Digital 05 Auto Firm Mode

  •  Pressing the Auto Firm button quickly inflates all cells to a single pressure setting.
  • The system will automatically return to Alternating mode after 20 minutes.

Auto-Detection of Pressure Settings

  • The Tamora Plus II Digital features Auto-Detection of cell pressures to provide appropriate care for each patient.
  • Utilized through ‘Auto’ activation or ‘Manual’ activation.

Auto-Detection

  • Every 2 hours system will automatically check for changes in patient weight/position, during this time the Auto-Detection indicator will flash.
  • Once the system has finished its detection it will set the internal cell pressures accordingly and the indicator light will remain ON.

ultimate healthcare Tamora Plus II Digital 06Function Mode Button

  • Pressing the Mode button allows you to cycle between the different modes available such as Alternating, Static and Constant Low Pressure.

Alternating Mode

  • Alternating mode is the default mode for the system. Within this mode the mattress will inflate and deflate in an alternating 1-in-2 cell cycle.
  •  The alternating cycle will continue at the selected cycle time until another mode is selected.

Static Mode

  •  Pressing the Mode button until the Static indicator illuminates puts the system into Static mode.
  •  After 30 minutes the system will automatically revert back to Alternating mode.

Constant Low Pressure (CLP) Mode

  • Pressing the Mode button until the CLP indicator illuminates puts the system into Constant Low Pressure mode.

ultimate healthcare Tamora Plus II Digital 10Changing Cycle Times

  •  Choice of 10/15/20/25/30 minutes.
  • To select press Cycle button until required cycle time is displayed in LCD screen.

ultimate healthcare Tamora Plus II Digital 11Manual Auto-Detection

  •  Simultaneously pressing the Soft and Firm buttons on the control panel will manually activate Auto-Detection.
  •  Once activated the indicator light will flash and the system will start detecting the patients weight to set internal pressures appropriately.

Troubleshooting

  1. No lights on power unit
    •  Check the power unit is connected to the mains power supply and that the mains switch is turned ON.
    • Check power unit for any blown fuses.
  2.  Patient is bottoming out (without alarm being triggered)
    • Pressure setting might be inadequate for the patient, adjust comfort level to Firm and wait for a few minutes for a better comfort level.
  3.  Low Pressure indicator is flashing and sounding
    • Check to ensure that the CPR valve is in the closed position.
    • Check whether power was suddenly shut down.
    •  Check that the connection between air tube and power unit is tightly secured.
    • Check that all tubing connections along the mattress are secured.
    •  If all of above steps have been checked. Press Alarm Mute button for system to be verified again.
  4.  The power unit is operating but mattress is not alternating
    • Ensure that the mattress inflation process is complete.
    • Check that Alternate indicator on control panel is illuminated. If not, press Mode button to switch to Alternating mode.
  5. Service indicator remains illuminated
    • Indicates that system is scheduled for an annual service. Please contact Uitimate Healthcare or your local dealer to arrange.
  6. Power Failure alarm indicator is flashing and sounding
    •  Check the power unit is connected to the mains power supply and that the mains switch is turned ON
  7. Power unit is noisy
    •  Ensure that the power unit is resting against a solid surface.

Ultimate Healthcare
Calmore Industrial Estate, Nutwood Way, Totton, Hampshire, SO40 3WW
tel: 0333 321 8996, fax: 0845 051 0035, email: sales@ultimatehealthcare.co.uk
www.ultimatehealthcare.co.uk