Vitalograph copd-6 Fast and Accurate COPD Screener User Guide

October 30, 2023
Vitalograph

Vitalograph logo

copd-6™

COPD Screener

Quick Start Guide

Prepare Device

  1. ****Switch On.
    Vitalograph copd-6 Fast and Accurate COPD Screener - Fig
2

  2. Set test subject’s age.
    Press or to change age.
    Press to set.

  3. Set test subject’s height.
    Press or to change height.
    Press to set.

  4. Set test subject’s gender.
    Press or to change gender.
    Press to set.

  5. Set test subject’s population group.
    Press or to change population group.
    Press to set.

No. Population Group Ref
1 European (default) ERS 1993
2 Asian ERS 1993
3 African ERS 1993
4 Chinese ERS  1993
5 Japanese ERS 1993
6 Polynesian ERS 1993
7 Caucasian NHANES III
8 African-American NHANES III
9 Mexican-American NHANES III
or
20 and above Weight (weight not normally required) kg

Perform Test

  1. Insert disposable Safe Tway mouthpiece.

  2. Demonstrate to test subject.
    • Fully breathe in, hold …
    • Insert and lightly bite the mouthpiece
    • Then blow as hard as possible for 6 s (wait for beep)
    • WELL DONE! Rest test subject for a short while and wait for blow icon to reappear

  3. Repeat blow 2 more times.

Vitalograph copd-6 Fast and Accurate COPD Screener - Fig
10

Test Result

  1. Press to show result

Negative

  • Spirometry not required
  • COPD Classification green

Positive

  • Refer for spirometry
  • COPD Classification I to IV

View Parameters

  1. Display shows FEV1 and % of predicted.

  2. Press to show FEV6 and % of predicted.

  3. Press to show FEV Ratio and % of predicted.

  4. Press to show Lung Age.