asma-1 Child
Guide for Healthcare Professionals
Testing Subject
-
Insert mouthpiece and switch on
-
Blow 3 times
-
View values for parameters FEV.₅,FEV.₇₅, FEV₁ and PEF
-
Best blow
PEF +/- 5%; FEV₁ +/- 3%
Review History
-
Switch On
-
Hold for 3 seconds
-
or Review History
-
Return
Delete History
- Switch on
- + Hold for 10 seconds
Cleaning and Disinfecting:
The outer surfaces should be cleaned every week, more often if necessary. The
use of an ordinary alcohol wipe is recommended, with special attention to the
mouthpiece area.
EN ISO 23747:2009
Specifies accuracy: ± 10% PEF
Resistance to flow: < 0.35 kPa/L/s
Vitalograph Childrens
Respiratory Monitor
Accuracy: ± 5% PEF, ± 3% FEV₁, .₅,.₇₅
Resistance to flow: < 0.15kPa/L/s