MSR63 BudgetLine Data Loggers Instruction Manual

June 4, 2024
MSR

MSR63 BudgetLine Data Loggers

Introduction

MSR63-BudgetLine-Data-Loggers-1

MSR BudgetLine: reusable low-cost data loggers for transport and storage

Reusable PDF data loggers for cost-effective monitoring of temperature and humidity values during transport and storage.

Features

  • high measuring accuracy, DIN12830 / 21 CFR Part 1- compliant
  • secured START and STOP buttons
  • clear status and alarm LEDs, battery indicator
  • 6-point calibration certificate included
  • fast data transfer and evaluation on PC
  • auto generated PDF report | Excel | CSV | email
  • low battery indication
  • battery life approx. 1 year, replaceable battery

Requirements

To use the MSR BudgetLine data loggers, you will require a PC with an active USB port. PC requirements: Windows 10.

Connect Data Logger

MSR BudgetLine data loggers are available with integrated USB 2.0 A-Type plug. Hence, no additional hardware or cable is required to connect the logger with a PC. Note:The logger will be STOPPED automatically after 10 seconds when connected with a PC. You can also STOP the logger manually by pressing the STOP button continuously for 5 seconds until the Red LED is on. Open Data Logger USB drive «PDF_Logger» that contains soft copy of the Quick Start Guide Calibration Certificate «CXXXXXXXXXX.pdf» and «UI4Report» – to get a report of recorded readings in PDF / Excel / CSV formats.MSR63-BudgetLine-Data-
Loggers-2

Configuring Data Logger

Connect the data logger to the PC, open the data logger folder «PDF_LOGGER», double-click on UI4Report and click on Configuration window at the top on top.MSR63-BudgetLine-Data-Loggers-3

Enter User Description, Logger Interval, Start After (delay start time), Alarm Limit, Alarm Delay Time, Alarm Type, Temperature Unit, Time Zone, Device Password (optional), User Email (optional) and click on Configure. By configuring the data logger the last recorded readings saved in its internal memory will be erased automatically. Click Yes to continue.MSR63-BudgetLine-
Data-Loggers-4

New data is loaded in the device. Please validate data in configuration screen before removing the device. Click OK. Verify the set parameters and click Close or Print to get a printout of the confirmation window.

Just after disconnecting the data logger, the green and the red LED will start flickering and then the green LED will be ON for a second indicating the configuration of the logger has completed successfully. Check if configuration is completed successfully: MSR BudgetLine loggers with display will continuously indicate current reading only (START | STOP | RUN | OKAY will not be shown on the display). Press «START» button on the data logger once; twice blinking of the green LED will indicate that the configuration is completed successfully.
Note: FailSafe Logging will be automatically started after the configuration of the data logger. This helps the receiver to get recorded readings from the logger even if the sender forgets to press the START button while activating the data logger.

START Data Loggers MSR63, MSR64

MSR63-BudgetLine-Data-Loggers-5

Check the status of the logger:
Press the START button once and check the LED indication

LED Indication                                Data Logger Status

  • green LED flashes twice       waiting to START (press START button for 5 seconds to START)
  • green LED flashes once      STARTED (in delay start mode)
  • green LED flashes once recording – OK (no alarm condition)
  • red LED flashes once   recording – alarm condition
  • red LED flashes twice STOPPED
  • yellow LED flashes twice   low battery

START the data logger:
Keep the START button pressed continuously for 5 seconds until the green LED flashes, followed by a flashing of the red and green LED»s. Once the red and green LED»s stop flashing you may check the status by pressing the START button again for once which will flash the green LED for once confirming the MSR data logger has been started successfully. Note: Do not use the data logger if no LED is ON or if the red LED flashes twice after pressing the START button. Replace the battery if the yellow LED flashes twice. Avoid using the logger without battery replacement.

START Data Loggers MSR83, MSR84, MSR86

MSR63-BudgetLine-Data-Loggers-6

Check the status of the data logger: During recording [RUN] & Stop [STOP] mode, press the START button once to view MIN | MAX | AVERAGE values.

Display Indication:

LCD Indication                                Data Logger Status

  • Only Current Reading           not started
  • START waiting to START recording (in delay start mode)
  • RUN started recording
  • STOP recording is stopped
  • OK no alarm condition
  • NOT OK LO | HI alarm condition
  • Lo VOLT low battery

START the data logger:
Initially, the current reading will be shown on the display before starting the data logger. Keep the START button pressed continuously for 5 seconds to start the logger. The green LED and the red LED will start flickering confirming the logger has been started successfully. The display will indicate START until the delay start time is over. After completion of the delay start time, the display will indicate RUN and the current reading confirming logger has been started successfully. Note: Do not use the data logger if START is not shown on the display or if the LED is not ON after pressing the START button for 5 seconds. «Lo VOLT» on the display indicates low battery. Avoid using logger without battery replacement.MSR63-BudgetLine-Data-
Loggers-7

Precautions for Sender

  • Store the data logger in a cool and dry place.
  • Don’texpose the data logger to direct sunlight and avoid condensation forming inside the unit.
  • A little pre-cooling is advisable in case of lower temperature application.
  • Don’t use the data logger if START is not shown on the display or if the LED is not ON after pressing the START button for 5 seconds.MSR63-BudgetLine-Data-Loggers-8
  • «Lo VOLT» on display / yellow LED flashes indicates low battery. Avoid using the data logger without battery replacement.
  • Record the complete details (serial number of data logger, start data & time) prior to starting the logger.
  • Place the data logger as close to the application as possible.
  • Never allow the data logger to remain loose, especially during transit.
  • Frequent rattling may reset the logger.
  • Clearly mark the box containing the loggers so that it is easily identified at the end of the measurement period

Precautions for Receiver

  • Identify the box containing the data logger.
  • Note down the time when the logger is removed from the box.
  • Check LCD / LED indications.MSR63-BudgetLine-Data-Loggers-10
  • Inform the sender immediately if NO LCD/LED indications are found at the time of receipt/removed from the box and do not try to connect the data logger with a PC.
  • Press the STOP button for 5 seconds (until the display indicates STOP) to stop the logger.
  • The data logger will be automatically stopped after 10 seconds when connected with the USB port of a PC.
  • Before connecting it with the USB port of a PC, allow the logger to come to room temperature.
  • Retrieve the recorded data from the logger as your first priority.
  • After stopping, when the logger is connected for the first time to the USB port of the PC, both LEDs (green and red) will start flickering, indicating the report generation of the recorded data is in process. Do not disconnect the logger until both LEDs (green and red) are ON continuously, which indicates the report of the recorded data is ready.
  • Once the report of the recorded data is generated, whenever the logger is connected to the USB port of the computer, both LEDs (green and red) are ON continuously, which indicates the data logger is in stop mode and the report of the recorded data is already generated.
  • Plug the logger on the USB port, open the folder «PDF_Logger» and double click on «UI4Report». This will open «UI4Report 1.0». Now select your required «Report Format» – PDF | Excel | CSV and click on «Process Report». Your selected report format will be automatically saved in the default location «C:\Documents\UI4Report\data». You may also choose your location to save the report or send to a third party by adding their email ID.
  • Don’t remove the USB cap or connect until the logger comes to room temperature.
  • Don’t retry connecting the logger if the report generation process is not started/completed after connecting the logger with the PC in the first attempt.
  • Don’t open the casing of the data logger or remove the battery. Doing this may delete the recorded readings from the logger permanently.

STOP Data Logger

To stop recording, press the «STOP» button continuously for 5 seconds until the red LED is ON. Note: Logger will be automatically STOPPED after 10 seconds when connected with the USB port of a PC.

Report Generation

MSR BudgetLine data loggers are available with integrated USB 2.0 A-Type plug. Hence, no additional hardware or cable is required to connect the logger with the PC. After stopping, when the logger is connected for the first time to the USB port of the computer, both LEDs (green & red) will start flashing to show the report generation is in process. Do not disconnect the logger until both LEDs (green and red) are ON continuously, which indicates the report is ready. Once the report is generated, whenever the logger is connected to the USB port of the computer, both LEDs (green & red) are ON continuously, which indicates the data logger is in stop mode and the report is already generated. Open data logger USB drive «PDF_Logger» that contains soft copy of the Quick Start Guide and the Calibration Certificate «CXXXXXXXXXX.pdf» and «UI4Report» – to get a report of the recorded readings in PDF / Excel / CSV formats.MSR63
-BudgetLine-Data-Loggers-11

Process Report – PDF / EXCEL / CSV Format

Double-click on UI4Report shown in the drive «PDF_Logger» to open the «UI4 Report 1.0» window.MSR63-BudgetLine-Data-Loggers-12

Report Format: Select your required report format (PDF or Excel or CSV).

Report Action:

  • Default Location: This will open a report of recorded readings in your selected format and save it in the default location C:\Users\{User}\Documents\UI4Report\data
  • Choose Your Location: This will open a report of recorded readings in your selected format and save it in your specified location.
  • Send Report by Email: This will send an email of the selected report format of recorded readings to the email ID specified.
  • Click on Process Report to perform the selected action.

Report Analysis

Selected report format of recorded data will be automatically opened and saved at your selected location just after clicking «Process Report».MSR63
-BudgetLine-Data-Loggers-13

Contact Us

We are pleased that you are using our data loggers. If you have any questions, please contact the sales office where you purchased your MSR BudgetLine data logger. More than 90 MSR sales partners in over 50 countries worldwide are available to provide you with competent advice.

MSR Electronics GmbH

  • Mettlenstrasse 6
  • CH-8472 Seuzach, Switzerland

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

MSR User Manuals

Related Manuals