Supereyes 10 Series Television Microscope Instruction Manual
- June 4, 2024
- SUPEREYES
Table of Contents
10 Series Television Microscope
Instruction Manual
Please read this operation instruction before you use a microscope
CATALOGUE
Notes ……………………………. Error! Bookmark not defined.
一、Brief ………………………. Error! Bookmark not defined.
二、Usage …………………… Error! Bookmark not defined.
三、Structure ……………………………………………………………… 4
四、Main Specifications …………………………………………….. 4
五、install and operation error! Bookmark not defined.
六、care and maintanceError! Bookmark not defined.
七、form a complete set of instrument errors! Bookmark not defined.
Attention
-
be sure not to disassemble
Random disassembling can easily lead to the instrument breaking Please do not randomly disassemble any parts except this operation instruction mentioned can disassemble of parts. If find the instrument out-of-order, Please ask a professional person to repair it or direct to contact the supplier. -
Inspect the voltage
1)Microscope base marked rated input voltage values. Pay attention to match the local service voltage. If the microscope is used under un-nominal voltage will be serious damage.
2)Power supply socket should earth fine. -
prevent burn and fire
1)during illumination, the bulb will be very hot.Be careful that do not to touch your skin.to avoid burn.
2)change the bulb should after cooling
3)to avoid fire, inflammable do not close to the bulb. -
change the bulb
1)to avoid electric shock and damage to the instrument, before changing the bulb, should pull out the attaching plug.
2)only use specified halogen and fluorescence lamp tubes.
3)should wear gloves or use a protective jacket mantle the bulb, Your hand does not touch the glass of the bulb. After changing the bulb, use the clean cloth to dip in straight alcohol, then wipe off the fingerprint and splodge that showed on the bulb.
Otherwise, bulbs can be easily broken and reduce the lightness. -
carry and place
1)should turn off the power switch before carrying the microscope
2)working environment requirement: room tempreture:0℃~40℃ maximum relative humidity: 85%
3)when using the microscope should avoid sunlight perpendicular incidence
4)avoid the microscope put on a dust environment, If disused, should coat on a dust cover.
5)the microscope should be placed in no vibrative place.
Brief
10 series of television microscope is a zoom stereo microscopes, that adopt unique achromatism and large depth of field technician, clear image, strong stereoscopic impression and long working distance, wide visual field, and so on. Matching high definition of colors CCD and television. Designed with 0.5X,2X Series of additional objectives and 0.35X,0.5X,1X Series of camera eyepieces. According to users’ different situation requirements, specially designed many kinds of illuminator devices. can according to the observed sample choose an up & down double halogen bulb; up halogen bulb, or down fluorescence bulb. ring fluorescence lamp; fiber optic cold light source, LED coaxial illumination, etc. Also can optional X.Y Moving stage, image analysis device to analyze the picture. Pillar stand base and boom stand base can option.
Usage
10 series of microscopes widely use for micro-electronics, mold, power metallurgy, precision machinery, jewelry, horologe, decorations, fingerprint identification, ticket recognize the false, print, fiber optic, maintaining the mobile, archaeology, living beings, anatomy, teaching, etc. Especially for observing the surface microstructure of large objectives. (many boom stand bases can option)
Structure (check below view):
1. stand base | 8. rotating sleeve |
---|---|
2. focus arm | 9. the main objective |
3. hand wheel | 10. camera eyepiece |
4. focusing hand wheel | 11. CCD camera(can be optional) |
5. pillar | 12. color television or monitor(can be optional) |
6. pillar cover | 13. the signal line(optional) |
7. additional objective(can be optional) | 14. CCDpower adapter(can be |
optional)
Main specifications
- main optical parameter
Television microscope magnification、working distance、dia of viewing field of object space; Effective for television image magnification. parts is additional objective, main objective, camera eyepiece and CCD camera, and television(display). The front 3 parts form an optical magnifying glass, back 2 parts form digital magnification. The largest Optical magnification is 30X, the Total magnification is1776X(1/3” CCD,14” monitor)largest working distance is 320mm,dia of viewing field of object space:90.7mm, As XDC-0745(ZOOM RATIO1:6.5) is 0.7X-1.5X for example:
a.total magnification=optical magnification×digital magnification optical magnification=zoom main objective ( 0.7X ~ 4.5X )×Camera eyepiece magnification× additional objective(ref sheet 1) digital magnification:relative with CCD camera specification and television(monitor)specification.(ref sheet 2)
b.wroking distance: only relative with an objective and additional objective magnification. (ref sheet1)
c.dia of viewing field object space=CCD camera diagonal size of target surface/optical magnification.
Optical magnification
Sheet 1:
Additional
objective| Camera eyepiece| Working
distance (mm)
---|---|---
O. 35X| O. 5X| 1X
0. 5X| 0. 12-0. 19X| 0. 18-1. 13X| 0. 35-2. 25X| 180
No
additional
eyepiece| 0. 25-1. 58X| 0. 35—2. 25X| 0. 70-4. 50X| 95
2X| 0.49″-3. 15X| 0. 70-4. 5X| 1.40–9. 00X| 32
Digital magnification
Sheet 2:
CCD camera
specification| television (display) specification
---|---
14″| 15″| 17″| 21″| 25″| 29″
1 / 3″| 59. 2X| 63. 5X| 72. OX| 88. 5X| 105. 8X| 122. 8X
1 / 2″| 44.5X| 47.6X| 54. OX| 66.7X| 79.4X| 92.1X
2 / 3″| 32.3X| 34.6X| 39.3X| 48.5X| 57.7X| 67. OX
CCD camera diagonal size of the target surface
CCD camera specification | 1 | / 3″ | 1 | / 2″ | 2/ 3″ |
---|---|---|---|---|---|
CCD camera diagonal size of the target surface | 6mm | 8mm | 11 mm |
hand wheel focus range:65mm Drawtube size: φ50mm, up to φ25mm pillar center
distance 140mm
lift range(φ25mm pillar): 178mm(XDS-10A lift range:265mm)
illumination: illumination: input voltage:110V/60HZ or 220V/50H output:
oblique incidence illuminator 12V10W with halogen lamp.
Transillumination:5W energy saving fluorescent lamp (according to the user)