SUGOLD HH.CP-100 CO2 Incubator Instruction Manual

October 30, 2023
SUGOLD

SUGOLD HH.CP-100 CO2 Incubator

SUGOLD HH.CP-100 CO2 Incubator product

Instruction Manual for CO2 Incubator

Thank you for purchasing our CO2 Incubator. Please read this Instruction Manual carefully before use and keep it for easy reference.

 Summary

Product Features:
  • Large color display screen: The liquid crystal screen displays humidity and temperature. The dynamic icons visually indicate the running condition of the incubator. High precision: The temperature is accurately controlled by microcomputer, which ensures a highly precise constant temperature and humidity. The touch-button facilitates the temperature regulation.

  • Safe and reliable: Over-voltage protection, under-voltage protection, and double temperature limits (Alarms are given in case of a superhigh temperature (over 69.9) or an ultralow temperature (under 0). The upper-limit protection can be provided when the temperature reaches 70.)

  • Environment-friendly and energy-saving: The insulating layer is integrally injected with hard PU, thus ensuring a long durability and a high effectiveness in energy saving.

  • Self-diagnosis: The fault can be diagnosed automatically, and displayed visually. The supply voltage can be detected automatically, and alarms can be given.

  • Automatic printing and recording: The data printer or curve recorder prints or records the historical data or curve of temperature automatically for easy reference and filing. (The configuration is optional. Please mark it when placing an order.)

  • RS485 computer communication interfaces (The configuration is optional. Please mark it when placing an order.)

  • Large color display screen:
    The liquid crystal screen displays humidity and temperature. The dynamic icons visually indicate the running condition of the incubator.

  • High precision:
    The temperature is accurately controlled by microcomputer, which ensures a highly precise constant temperature and humidity. The touch-button facilitates the temperature regulation.

  • Safe and reliable:
    Over-voltage protection, under-voltage protection, and double temperature limits (Alarms are given in case of a superhigh temperature (over 69.9°C) or an ultralow temperature (under 0°C). The upper-limit protection can be provided when the temperature reaches 70°C.)

  • Environment-friendly and energy-saving:
    The insulating layer is integrally injected with hard PU, thus ensuring a long durability and a high effectiveness in energy saving.

  • Self-diagnosis:
    The fault can be diagnosed automatically, and displayed visually. The supply voltage can be detected automatically, and alarms can be given.

  • Automatic printing and recording
    The data printer or curve recorder prints or records the historical data or curve of temperature automatically for casy reference and filing.
    (The configuration is optional. Please mark it when placing an order.)

  • RS485 computer communication interfaces
    (The configuration is optional. Please mark it when placing an order.)

Application Scope and Main Purposes:

The CO2 Incubator is a kind of high-precision and thermostatic device that can control heating and refrigeration designed for the required purposes.
The CO2 Incubator (referred to as the incubator, the same below) is a kind of high-precision laboratory device providing constant temperature, constant humidity and CO2, thus making itself ideal for colleges and universities, scientific research institutions, medical field, chemical, food, animal husbandry, aquaculture, environmental sciences, biogenetic engineering and so on. It can find wide application in testing effectiveness and aging of medical field, chemical raw materials, wood and building materials, as well as in culture of mould, cells, tissues, micro-organisms, antibiont and so on.

Type and Specification:
The incubator has a variety of specifications.

Composition and Meaning of Model:SUGOLD HH.CP-100 CO2 Incubator
01

The model name is HH.CP – XXX, where the nominal volume is measured in liters and the product code means CO2 Incubator.

Service Environment:
The CO2 incubator should be used under the following conditions:

  • Ambient temperature: 5-30°C
  • Ambient humidity: 80%RH
  • Range of atmospheric pressure: 86kPa-106kPa.

Operating Conditions:

The CO2 incubator should be used according to the instructions provided in this manual.

  1. It should be operated on an AC (220 ± 22) V (50 ± 1) Hz power source.
  2.  Operating mode: Continuous operation.

Structural Features

  • Body: The exterior housings are made of superior cold-roll steel sheets and the inner housing is made of brushed stainless steel. The insulating layer is made of hard PU foam.

  • Door: The door is designed to provide easy access and is equipped with a safety lock.
    The door is mounted with hollow glass in its middle so as to facilitate observing the experiment effect. The insulating layer is made of hard PU foam. The door gasket is sealed together with the housing with magnetic adhesive.

  • Refrigerating system:
    It is made up by the compressor, condenser, capillary tube, evaporator and so on.

  • Control system:
    It is controlled by microcomputer.

  • Supporting net:
    The round steel is made into the strip shelf which is then sprayed with plastics. The height is adjustable.

Technical Features

The main technical parameters are shown in the chart below:

Product Model| Nominal volume| Temperature control range (Note1)| Allowance of temperature fluctuation| Allowance of temperature uniformity| Lighting power| Compressor power| Heating power| Refrigerant
---|---|---|---|---|---|---|---|---
HH.CP-100| 100L| RT+5-65°C| ±0.1°C| ±0.5°C| 6VA| 98VA| 300VA| R134a

The incubator is a kind of high-precision and thermostatic device that can control heating and refrigeration designed for the required purposes (2nd in the summary).
The main technical parameters are shown as follows( chart 1 and 2):

Dimension of incubator Body

Product Model Boundary Dimension Dimension of Inner Body
HH.CP-100 500mm×510mm×1120mm 430mm×410mm×570mm

Major Performance Parameters

Model Item HH.CP-100
Nominal volume 100L
Temperature control range RT+5℃~65℃ (Note1)
Allowance of temperature fluctuation

**** ±0.3℃

Allowance of temperature uniformity| **** ±0.8℃
Lighting power| 6VA

**** Compressor power

| **** 98VA
Heating power| 300VA
Refrigerant| R134a

Note: The above parameters are measured in the 30℃±2℃ thermostatic incubator.

Product Usage Instructions

  1. Place the CO2 incubator in an appropriate location according to the service environment requirements in the manual.
  2. Connect the power supply and turn on the power switch.
  3. Press the touch-button to set the desired temperature. The temperature is accurately controlled by microcomputer.
  4. The liquid crystal screen displays humidity and temperature, and the dynamic icons visually indicate the running condition of the incubator.
  5. The CO2 incubator has over-voltage protection, under-voltage protection, and double temperature limits. Alarms are given in case of a superhigh temperature (over 69.9) or an ultralow temperature (under 0). The upper-limit protection can be provided when the temperature reaches 70.
  6. The fault can be diagnosed automatically, and displayed visually. The supply voltage can be detected automatically, and alarms can be given.
  7. The data printer or curve recorder prints or records the historical data or curve of temperature automatically for easy reference and filing. (The configuration is optional. Please mark it when placing an order.)
  8. RS485 computer communication interfaces (The configuration is optional. Please mark it when placing an order.)

 Directions for Use and Precautions

The indication panel of the incubator :
SUGOLD HH.CP-100 CO2 Incubator 02 Key and identifier description:
The definition of Identifier

  • TI Reservation: when entering the reservation timing state, TI will flash and [ (2) the display window] will display the appointment countdown;

  • T2 Heating: when heating has output, T2 lights up;

  • T3Compressor: when the compressor starts, T3 lights up; when the compressor is waiting for start-up, T3 flashes;

  • T4 Humidification: when humidifying has output, T4 lights up;

  • T5 Daytime: when entering day mode, T5 lights up

  • T7 Timing: when entering the operation timing state, T7 will flash, and [ (2) display window) will display the timed countdown;

  • T8 Water shortage: when there is a signal of water shortage, T8 lights up; when there is an alarm of water shortage, T8 flashes;

  • T9 Open door: when the door is opened, T9 lights up;
    TO Alarm: when there are alarms of temperature and humidity, T10 lights up; when entering the low temperature or high temperature protection state, T10 flashes

  • T11 Lock the screen: when the screen is locked, TIl lights up;

  • T12 Adding water: when the pump has output, T12 lights up;

  • T13 Defrosting: when defrosting has output, T13 lights up;

  • T14 Solenoid valve: when the solenoid valve has output, T14 lights up

Display window
【(1)Display window】 shows: values of period or segment ; 【(2)Display window】 shows: The values of timing or time setting;
【(3)Display window】 shows: the measured temperature; 【(4)Display window】 shows: the setting temperature;
【(5)Display window】 shows: the measured humidity; 【(6)Display window】 shows :the setting temperature;
【(7)Display window】 shows: the setting illumination or the output power of heating .

Definition of Keys

key:| In normal display state, click this key to switch on and off the disinfection and sterilization function.
---|---
Key:| In normal display state, click this key to switch on and off lights.
Key:| In normal display state, click this key to enter the modification interface of the setting values , and long press this key for 3 seconds to enter the modification interface of the parameter table.
Key:| In the setting state, click this key to make the setting values shift, flash and modify.
In normal display state, if it runs in 【day / night 】  mode, it can switch between day and night running state; if it runs in program mode, it can switch
| to display the number of segments or the cycle.
---|---
Key:| In the setting state, click this key to increase the setting values.
Key:| In the setting state, click this key to decrease the setting values.
In normal display state, long press this key to turn on the defrosting function manually.
Key:| In normal display state, click or long press this key to start or stop the operation of the controller.

Instructions of operation:

  1. When the controller is powered on,,【(3)Display window】 shows “TH”,【(5)Display window】 shows“V01”, the buzzer gives a short beep. After about 2 seconds, it will enter the normal display state.

  2. The modification of setting values In normal display state, click 【setting】 key, the two identifiers, “TIME” and “SV” flash at the same time. At this time, the value of the number flashed by the cursor can be modified by 【shift】, 【increase】, 【decrease】 keys, and click 【setting】 key again to switch to the next group of values. After modification, long press the 【setting】 key for 1 second to exit the setting state, or in the fixed mode, continuously click the 【setting】 key and modify a group of data and exit. The buzzer will give a short beep to enter the normal display interface, and the parameters will be saved automatically. When the controller is running in the program mode, click the【 setting】 key, and the value of segments starts to flash. At this time, you can view the setting values of time, temperature, humidity and illuminance under the corresponding number of segments by modifying the value of segments with the 【increase】, 【decrease】 keys. And then click the 【setting】 key again to modify the setting parameters one by one in the present segment. When the cursor returns to the value of segments and flashes, modify the value of segments again to modify the next segment. When the controller is running in the 【DAY /NIGHT】 mode, click the 【setting】 key, and the “DAY” identifier will flash. Through the 【increase】, 【decrease】 keys, it can be switched, and the “NIGHT” identifier flashes, to view the setting parameters under the day and night. Click the 【setting】 key to modify the setting parameters in the day or night mode.

  3. To set the number of segments or cycle In the non-fixed value mode (see parameter table-1-u1 for details), when the controller stops running, long press the 【setting】 key for 3 seconds, the 【(1) display window】 displays the “LC” prompt, and the 【(2) display window】 displays the password. Modify the password to 3 through the 【increase】, 【decrease】 keys to enter the cycle and segment number settings. In the program mode, if the “PRO” prompt flashes, you can set the total operation sections; when you click the 【setting】 key, and the “CYC” prompt flashes, you can set the total operation cycles (when the number of cycles is set to 0, the controller always runs). After setting, long press the 【setting】 key for 3 seconds to exit the setting and the parameters will be saved automatically. In the day and night mode, there is no total operation sections, only the cycle can be set.

  4. .Start and stop
    Long press the 【R / S】 key U7 (see user parameter TABLE-1 for details) seconds to start the the controller, and the remaining running time will be displayed in 【(2) display window】. When the operation time is over, the controller will stop working, and the buzzer will beep u9 (see user parameter TABLE-1 for details) seconds.【(2) display window】 will display “End”, or long press 【R / S】 key U7 seconds to stop the controller working and 【(2) display window】 will display “oFF”. When the controller is running in program mode and the total number of segments is greater than 1, if the constant temperature and humidity condition (see parameter table-1-u5, U6 for details) is not on, it is necessary to set each segment of time before starting the controller. When the time is all set to 0, the start will fail.

  5. Reservation function
    When the value of the reservation function AP (see the user parameter table-7) is set to 1, click the 【setting】 key again to set the reservation time, and the unit is minutes. Otherwise, it will directly exit the normal display interface. After the reservation time setting, it is needed to click the 【R / S】 key to start the machine. In the reservation timing, you can enter the parameter table again to modify the reservation time, or click the 【R / S】 key to stop the controller, and the setting time of running the machine is automatically cleared, which is only valid once.

  6. The memory function for power-down

By modifying the parameter of power down memory (see “U2” parameter: internal parameter TABLE-1), you can select whether there is memory function for power- down.

Functions of Intrinsic Parameters
Press  ****to enter into the parameter setup menu after the instrument is powered on.

Internal Operation Parameters of the Control Panel

Press **| key :LC=000| |
Prompt| Name| Setting Range| Instruction
| |
LCK=9| the parameter table of users
U1| Operation mode| (0~2)0| 0:Fixed value operation mode;
1: Day and night mode, 0 ~ 99 cycles;
2: Program mode, which is programmable 1 ~ 30 segments, 0 ~ 99 cycles.
U2| Operation mode of| (0~2)0| 0: No operation;
1:     Operate from the first segment (daytime); 2:     Operate from power- down time.
| power-down| |
---|---|---|---
U3**| Timing correction| (-999~999) 0| Correct the total timing error, The correction value =  【 opera (seconds)】 * 10 ÷ actual time| ting time (seconds) – actual (minutes).| time
U4| Time unit| (1~2) 1| 1:minute 0~9999; 2:hour 0~9999。
U5|

Timing deviation in the condition of constant temperature

| (0~10.0℃) 0| If the deviation between the measured temperature and the setting      value is less than U5, start the timing. Note: 0 means that there is no need to judge the temperature when timing.
U6| Timing deviation in the condition of constant humidity| (0~50.0%) 0| If the deviation between the measured humidity and the setting value is less than U6, start timing. Note: 0 means that there is no need to judge the humidity when timing.
U7| Key Effective time| (0~10s)0| After long press U7(the time), | function key is valid
UA| illumination time| (0~9999min)  5| After the lighting is turned on, the illumination time is automatically turned off.
Note: 0 means that the lighting must be turned off manually.
Ub| sterilization time| (0~9999min) 30| After the sterilization is turned on, the sterilization time is automatically turned off.
Note: 0 means that sterilization must be turned off manually .
Uc| Address| (1~16) 1| Address of the machine.
LCK=103 **parameter table of temperature
TH| the upper deviation is over temperature and an alarm appears;| (0~20.0℃) 4.0| If “the measured value > the setting value + TH “, the upper deviation alarms and the temperature and humidity output will be turned off.
The relay has output when there is a temperature alarm; the buzzer beeps; the alarm indicator is on, and the temperature unit flashes rapidly. Click any key to cancel the buzzer.
TL| the lower deviation is over temperature and an alarm appears;| (-50.0~℃) 0| If “the measured value < the setting value + TL “, the lower deviation alarms.
The relay has output when there is a temperature alarm; the buzzer beeps, and the temperature unit flashes slowly. Click any key to cancel the buzzer.
Note: when “TL = 0”, this function is invalid.
Tb| Deviation correction| (-99.9~99.9℃)
0| Correct measurement errors of the sensor in low temperature;
Tb = the actual temperature – the measured value.
TA| Slope correctio0n| (-999~999) 0| Correct measurement errors of the sensor in high temperature;
TA = 1000 * (the actual temperature – the measured value) / the measured value.
TP| Heating ratio| (0.1~50.0)      8.0| Adjust the function of the proportion of time
TI| Heating integral| (1~2000s)     500| Adjust the function of the integral
TD| Heating differential| (0~2000s)     200| Adjust the function of the differential
TT| Heating cycle| (1~60s)     5| The cycle of controlling Heating
Tc| Heating off| (-5.0~5.0℃)0| Restrict the heating overshoot; when “the measured temperature
| | |
the setting temperature + Tc ”, heating stops output.
---|---|---|---
To| heating power| (0~100%) 100| The percentage of the maximum power of the heating output.
| | | When| the| compressor| is| in| manual| start-stop| mode| and| the
Tu| Turn on refrigeration| (-10.0~10.0℃)
0.6| compressor  is  under  disconnected  controlling  mode,  if  “the measured temperature
the setting temperature + Tu “, start the
| | | compressor.
Tn| Turn off cooling| (-10.0~ uP )
0.6| When the compressor is in manual start-stop mode and the compressor is under disconnected controlling mode, if “the measured temperature  ≤ the setting temperature
+ Tu Tn “, shut down the compressor .
TE| High-temperature control, heating off| 0| Under controlling of high temperature, the temperature point where heating stops is effective only when the setting temperature is higher than the ambient temperature.
| |
LCK=203| parameter table of humidity| |
HH| The upper deviation is
over humidity and an| (0~50.0%)
20.0| If “the measured humidity > the setting value +
HH** “, the upper deviation alarms; the humidity alarm relay is disconnected and the humidity output is turned off.

The humidity alarm relay has output when there is an alarm; the alarm indicator is on, and the humidity unit flashes rapidly.

| alarm appears|
HL| The lower deviation is over humidity and an| (-50.0~0%)
0| If “the measured humidity < the setting value+ HL “, the lower deviation alarms

The humidity alarm relay has output when there is an alarm; the alarm indicator is on, and the humidity unit flashes rapidly.

| alarm appears| | Note: this function is invalid when “HL = 0”
Hb| Deviation correction| (-99.9~99.9%)
0| Correct measurement errors of the sensor in low humidity;
Hb = the actual humidity – the measured value.
HA| Slope correction| (-999~999) 0| Correct measurement errors of the sensor in high humidity;
HA = 1000 *** ( the actual humidity the measured value ) / the measured value
HP| Humidifying ratio| (0.0~90.0)| 10.0| Adjust the function of the proportion of time
HI| Humidifying integral| (1~999s)| 200| Adjust the function of the integral
Hd| Humidifying
differential| (0~999s)| 30| Adjust the function of the differential
HT| Humidifying cycle| (0~60s)| 5| The cycle of controlling Humidifying
Hc| Low-temperature control, Humidifying off| (-50.0~50.0%)
0.0| Under  controlling  of  low  humidity,  the  humidity  point  where humidifying stops.
Ho| Humidification power| (0~100%) 100| The percentage of the maximum power of the humidifying output.
Hu| Turn on dehumidification| ( Hn ~20.0%)3.0| When the compressor is in manual start-stop mode compressor  is  under  disconnected  controlling mode, measured  humidity the  setting  humidity  +  Hu “,| and if start| the “the the
| | | compressor.
---|---|---|---
Hn| Turn off dehumidification| (-20.0~ Hu )3.0| When the compressor is in manual start-stop mode and the compressor is under disconnected controlling mode, if “the measured  humidity the  setting  humidity  + Hn “,  stop  the compressor.
HE**| The point where there is no humidification| (0.0~10.0)2.0| When the compressor is working in intermittent mode, if the setting value meets the conditions, the value can be modified and stop the humidification in advance.

HF

| The point where there is no humidification in low humidity| (-10.0~10.0%)
-5.0| When controlling low humidity, if “the measured humidity  ≥ the setting humidity + HF “, the controller will prohibit humidification.

 Instructions for the intelligent supplement device of CO2

  1. Monitor and adjust CO2 levels at any time. The resolution is 0.1%. ( When the new CO2 device changes from a low-temperature use to a high one, the display value of the CO2 device may rise for the reason that there is a small amount of glass glue in the incubator but it will automatically become normal for some time. To solve the rising problem, open the door of the incubator for 5 seconds, then close the door.
  2. Instrument operation:
  3. Introduction for setting symbols: AL1 represents the upper limit value(this device is set to zero); AL2 represents the settings value( the modified value of CO2 concentration ); AH1 is the upper limit value that the relay outputs difference value; AH2 is the setting value that the relay outputs difference value (AH1, AH2 have been set up when leaving factory, needing no change).
  4. Control example: a certain system control technics requires: The control concentration should be 5%.
  *  Under the condition that instrument PV displays the measured value, when pressing SET button, the instrument displays: AL1, and then pressing the SET button, the instrument displays (AL1) a parameter, by pressing increase or decrease button to change it to 0%.
  *  After the AL1 parameter has set, when pressing the SET button, the instrument displays AL2. Then press the SET button, and the instrument displays (AL2) a parameter, by pressing increase or decrease button to change it to 5%.
  *  After the AL2 parameter has set, when pressing the SET button, the instrument displays AH1. Then press the SET button, and the instrument displays (AH1) a parameter, by pressing increase or decrease button to change it to 0.01%.
  * Continue to press the SET button, and the instrument displays AH2. Then press SET button, and the instrument displays (AH2) a parameter, by pressing increase or decrease button to change it to 0.01%. After setting, press the reset button to return to the instrument to display the measured values.

Fault Analysis and Removal

The simple faults and removal methods are shown in Chart 4:

Fault Symptom| Analysis of Causes| Troubleshooting| Remarks
---|---|---|---
The power light isn’t working.|

  1.  There is no power supply in the socket.
  2.  The fuse protector blows out.

|

  1. Check the socket
  2. . Replace the fuse protector

| Behind the box
The unit does not heat (or temperature does not rise.).|

  1. The heating light is off, indicating the control board failure.
  2. The heating light is on, indicating the blowing out of heating tube.

|

  1.  Replace the control board.
  2. Replace the heating tube.

|
The unit can not refrigerate (or cool).|

  1.  The refrigeration light is off, indicating the control board failure.
  2.  The refrigeration light is on, indicating the failure in the refrigerating system.
  3. The time delay for compressor startup or re-operation is less than three minutes. Please wait

| 1.  Replace the control board
2.  Repair the refrigerating system
3.  Please wait for a moment.|
The heating or refrigeration light is on, but there is no sign of temperature rise or drop.| Control board failure| Replace the control board|

Located      in     the instrument

The temperature display stays at“ —” always| There is a short circuit in the temperature probe or an open circuit in the leading wire.| Replace the probe Pt100|
The buzzer sounds, and the power light flickers.| The voltage is not stable.| Check the power supply circuit, improve the mains supply, or add the regulated power supply.|

Notes: In case the failures are not included in the above table or the above failures can not be removed, please contact the technical service department of our company or our maintenance points.

 Maintenance

  1.  The incubator must be kept in a well-ventilated and dry place away from sunshine. The distance between the unit and the wall shall not be less than 10cm.
  2.  The incubator shall be equipped with a three-pin plug which shall be reliably grounded, in order to ensure the normal operation and personal safety.
  3. It is not allowed to clean the inner and exterior surfaces of the incubator with acid, alkali and other corrosive substances for the sake of beauty. It can be cleaned with neutral detergent and then wiped with dry cloth.
  4. When replacing the fuse protector, the one with the same blowing current and the same dimension as marked in the fuse protector shall be selected.
  5. During use, the incubator should be checked often to ensure that there is water in the humidifier tank.
  6. When humidifying is not used, the water in the humidifier should be removed and dried.
  7.  When the incubator is not used, unplug it, and keep it dry and clean within.
  8.  The dirt on the condenser (at the back of the incubator) shall be cleaned every six months, so as to ensure a favorable heat dissipation during the operation.

Transport and Storage

  1. Transport
  2.  Rough transport is not allowed.
  3.  During the transportation, the incubator shall be inclined at an angle less than 45 degrees. It is not allowed to place it flatwise or upside-down.
  4.  During the transportation of unpacked incubator, pay attention not to impact the refrigerating pipeline at the back of the incubator.
  5.  Storage

In case the incubator is not used for a long time before or after unpacking, it shall be stored in a place with the following conditions.

  1.  Ambient humidity: 0~40℃
  2.  Relative humidity: ≤80%RH;
  3. Range of atmospheric pressure: 86kPa~106kPa。

Out of box audit

  1. After unpacking the incubator, you shall take out the Instruction Manual to check whether all the attachments are complete and whether the product specification conforms to the Manual according to the “List of Attachments”.
  2. Please inspect and operate the incubator according to the Instruction Book.

 List of Attachments

Model     Name HH.cp-100
1. Instruction Manual 1
2.Quality tracking card 1
3. Supporting net 3
4. Hook 14

5. Drip tray| 1

Special announcement
Accordine to related regulations. the product. from the date of purchase. will be guaranteed

  1. According to related regulations, the product, from the date of purchase, will be guaranteed for one year, except the man-made faults Our company reserves the nght to change the design of the product. The design is subieet ic
  2. Our company reserves the right to change the design of the product. The design is subject to change without prior notice.
  3. The Instruction Manual covers the contents of technical instruction.

Wiring diagram

SUGOLD HH.CP-100 CO2 Incubator 11