mPower Electronics MP342 4-Bay CaliCase Automatic Calibration Station User Guide

June 11, 2024
mPower Electronics

mPower Electronics MP342 4-Bay CaliCase Automatic Calibration Station User Guide
mPower Electronics MP342 4-Bay CaliCase Automatic Calibration
Station

Product Overview

Product Overview

1 Unit LED x 4 8 Exhaust Port
2 Status LED 9 Gas Inlet2
3 CAL Button 10 Pressure Gauge
4 BUMP Button 11 Thumb Screw x 2
5 USB Data Line Port 12 Demand-flow Regulator
6 Air Inlet 13 Gas Inlet1
7 Charging Port 14 UDisk Port

Caution: Use in non-hazardous-gas locations only!

SETUP AND INSTALLATION

  1. Connect the case to a PC by USB cable [5]. Use mPower Suite software to configure the gas concentrations & other parameters on the CaliCase. Be sure to set the gas types to match the sensor types and the expiration date past the current date.
  2. Insert the UDisk thumb drive into a PC USB port and configure the gas concentrations & expiration dates on the UDisk using mPower Suite. Be sure to Write both the CaliCase Info and the MUNI Info to the UDisk.
  3. With the Calicase powered off, insert the UDisk into port [14] and then turn the power on. IMPORTANT! The CaliCase will not operate properly without the UDisk!
  4. Thread the gas cylinder into the demand-flow regulator (DFR) [12] C10 inlet fitting [13]. Once tightly connected, the gauge [10] will indicate the cylinder gas pressure. If needed, a second gas can be connected to port [9] using a separate DFR.

CALIBRATION OR BUMP

  1. Turn on 1-4 MUNI instrument(s) and place them facedown into their cradle(s).
  2. If the power is off, press and hold the CAL button [3] until the Status LED [2] turns green.
  3. CALIBRATION: Press the CAL button [3] to initiate. The Unit LED(s) [1] should blink green for about 60-100 seconds and then stay GREEN if passed or RED if failed.
  4. BUMP: Press the BUMP button [4] to initiate. The buzzer will be tested and the Unit LED will blink green for about 30 seconds and then stay GREEN if passed or RED if failed.
  5. A Calibration or Bump Report will be saved on the UDisk.
  6. To abort a Calibration or Bump, press and hold the BUMP button [4] until the Unit LED(s) turn off.
  7. To power off, press CAL until the Status LED turns off.

LEDS

LED Color Description
Status LED Green Power on
Green blinking Low battery
Orange Charging
**** Red blinking 1)  USB Disk error2)  Datetime error3)  No pump detected
None Power off
Unit LEDs Green blinking fast Cal/Bump testing
Green blinking slow Unit firmware upgrading
Orange Match fail
Green 1)  Cal/Bump test pass2)  Unit firmware upgrade
Red 1)  Cal/Bump test fail2)  Unit firmware upgrade fail