mPower Electronics MP110-v1.1 UNI Lite Single Gas Detectors User Guide
- June 1, 2024
- mPower Electronics
Table of Contents
- MP110-v1.1 UNI Lite Single Gas Detectors
- Read Before Operating
- Turning the Unit On
- Event Log
- Enter Configuration Mode
- Configuration Mode Menu
- Zero (Fresh Air) Calibration
- Span Calibration
- Maintenance and Service
- Proper Product Disposal at The End Of Life
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
UNI LITE
MP110 Series
Quick Start Guide mPower Electronics Inc.
2910 Scott Blvd. Santa Clara, CA 95054
www.mpowerinc.com
info@mpowerinc.com
PN:M027-4002-000 v1.1
MP110-v1.1 UNI Lite Single Gas Detectors
Warning
- Never operate the monitor when the cover is removed.
- Remove the monitor cover and battery only in area known as non-hazardous.
- Use only mPower’s lithium battery part number M5000038-000 (EVE 14335) (3.6 V, 1650 mAh, 2/3 AA size).
- This instrument has not been tested in an explosive gas/air atmosphere having oxygen concentration greater than 21%.
- Substitution of components will impair suitability for intrinsic safety and void warranty.
- It is recommended to bump test with a known concentration gas to confirm the instrument is functioning properly before use.
- Before use, ensure that the colorless ESD layer on the display is not damaged or peeling. (The blue protective film may be removed.)
Read Before Operating
The User’s Guide must be carefully read by all individuals who have or will
have the responsibility ofusing, maintaining, or servicing this product. The
product will perform as designed only if it is used, maintained, and serviced
in accordance with the manufacturer’s instructions.
User Interface
The UNI’s user interface consists of the LCD display, LEDs, an alarm buzzer,
one push-button key, an alligator clip, and a chemical sensor.
Turning the Unit On
Press and hold the Operation Key ( ) for 3 seconds, until the LCD displays
“On” and the unit enters a self-test sequence. The alarm buzzer, vibrator, and
LEDs are tested, and the firmware version, (e.g. “VER”,”1.0.0.0″) and the
programmed alarm limits, STEL, TWA, etc. are shown. After the start-up is
completed, the instrument enters the detection mode and displays the real-time
gas concentration.
Turning the Unit Off
In normal display mode, press and hold the Key for a 3-second count down until
the unit powers off.
Normal Mode
In normal mode, the gas concentration is displayed on the LCD continuously and
the unit alarms if a pre-set limit is exceeded. Press and hold the Key until
it beeps to test the function of the buzzer, LEDs, and vibration alarm. Or
short press the Key to cycle through values of STEL, TWA, MIN (O2 only), and
PEAK, and view EVT LOG.
Event Log
From the EVT LOG screen, long-press the button until it beeps to show the latest alarm event A1 and then short-press the button to cycle through the last 10 alarm events. Up to 50 events can be viewed using mPower Suite.
Enter Configuration Mode
With the unit off, double-click the Operation Key and the unit shows “PROG”. Short-press the Key to prompt password entry with the first digit flashing. Short-press the Key to increase the number, and longpress until the beep to move the cursor to the next digit. The default password is 0000. After all four digits are entered, long-press to move to “OK” and short-press to accept and enter Configuration Mode.
Configuration Mode Menu
In Configuration Mode the user can calibrate the unit and change various operating parameters*:
- AIR: Zero calibration,
- SPAN: Span calibration
- SET HIGH\LOW\STEL\TWA alarm limits
- SET CAL Change calibration span value
- SET UNIT: Change concentration unit
- EXIT: Exit Config Mode
Configuration Mode navigation: In general, longpress the Key to enter the
menu item and short-press to scroll to the menu next item, increase a number,
confirm, or to move to the item withinthe menu. Adjust numerical digits as
for password.
*Additional parameters beyond those accessible in Configuration Mode may be set using mPower Suite software, which requires an MP312 4-bay CaliCase.
Exit Configuration Mode
Scroll to “EXIT?” and long-press to exit and return to Normal Mode.
Alarm Limits
Alarms are triggered when readings are above the Low, High, STEL or TWA Alarm
limit. To adjust an alarm limit, enter Configuration Mode and scroll to: SET
HIGH\LOW\STEL\TWA?.
- Long-press to display the alarm value with the first digit flashing
- Short-press to increase the value and cycle 0-9.
- Long-press to move the cursor to the next digit.
- When finished, long-press to scroll to OK and short-press to save and exit.
Default Alarm and Span Settings (ppm)
Sensor | Low | High | STEL | TWA | Span |
---|---|---|---|---|---|
CO | 35 | 200 | 100 | 35 | 100 |
H2S | 10 | 20 | 15 | 10 | 25 |
Zero (Fresh Air) Calibration
Zero calibration sets the base line for the sensor and is done in fresh air or
other clean air source.
Enter Configuration Mode and “AIR?” displays as the first menu item. Long-
press to start the15-second zero calibration count-down, after which a “pass”
or “fail” result will be displayed.
To abort, long-press during the 15-second count, and “ABRT” displays to
confirm.
Span Calibration
Span calibration uses a known concentration gas to determine the response of the sensor to the gas.
- Ensure that the CAL value is set to the same concentration as the Span gas cylinder.
- Attach the Calibration Adapter over the inlet port on the front of unit by pressing it into place.Use a fixed-flow regulator with a flowrate preferably of 0.3 LPM. Up to 0.5 LPM can be used but not higher.
- Enter Config Mode and scroll to “SPAN?”
- Start the gas flow and long-press to start the calibration countdown. The default count time is 45 seconds but may vary with sensor type.
- When finished, a “pass” or “fail” result will be displayed. Turn off the gas supply, remove the calibration adapter, and exit to Normal Mode.
- To abort at any time during the count-down, long-press and “ABRT” is displayed.
Maintenance and Service
Battery: MP110 has a built-in lithium battery. If the battery is drained,
replace it with a new one. The alarm signal is 1 beep and flash per minute
until a new battery is installed.Until the battery is drained, the screen
will display “bAT LoW” and the instrument readings will no longer be
displayed. After the battery dies, it will continue to sound and light alarm
for 1 minute. If the battery has not been drained, the user can long-press the
control key for a manual shutdown.
Sensor: If used in high-dust environments, purge the sensor inlet opening
with compressed air to prevent dust build-up from reducing the sensitivity of
the detector. Replace the sensor as needed when it fails calibration or gives
noisy readings.
Cautions
When disassembling the case and replacing the battery, be careful not to
damage the internal circuit of the detector and pay attention to the positive
and negative poles of the battery.
It is recommended that the detector be calibrated every three to six months or
according to company regulations.
Proper Product Disposal at The End Of Life
**** The Waste Electrical and Electronic Equipment (WEEE) directive
(2002/96/EC) is intended to promote recycling of electrical and electronic
equipment and their components at the end of life.
This symbol (crossed-out wheeled bin) indicates separate collection of waste
electrical and electronic equipment in the EU countries. This product may
contain one or more nickel-metal hydride (NiMH), lithium-ion, or alkaline
batteries. Specific battery information is given in this user guide. Batteries
must be recycled or disposed of properly. At the end of its life, this product
must undergo separate collection and ecycling from general or household waste.
Please use the return and collection system available in your country for the
disposal of this product.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>