TECHNO VIETNAM KEW 1061 Digital Multimeters Instruction Manual

June 11, 2024
TECHNO VIETNAM

TECHNO VIETNAM KEW 1061 Digital Multimeters

Product Information

The KEW 1061/1062 Digital Multimeter is a measuring tool that uses modulation for A/D conversion, allowing for fast and accurate measurements. It has a 5-digit LCD display with a maximum reading of 50000 and a bar graph indicator. The multimeter has switching detection modes for effective value (root mean square value) detection (RMS) and mean value detection (MEAN) during AC voltage or AC current measurement (KEW1062 only). Additionally, it has a low- pass filter that can be switched on/off during AC voltage or AC current measurement (KEW1062 only). The multimeter complies with CE standards and features a current-input terminal shutter for preventing wrong input. It uses high-performance UL-standard fuses.

Product Usage Instructions

Safety Instructions
Always observe the following instructions to avoid the risk of electrical shock or other dangers that may lead to serious injury or loss of life:

  • Refer to the WARNING instruction manual to avoid the risk of serious injury or loss of life.
  • Refer to the instruction manual to avoid the risk of injury or damage to the product.
  • Handle with care to avoid the risk of injury or death of personnel or damage to the instrument.
  • Do not use the product if it is exposed, removed, or wet.

Measurement Category

Ensure that you use the KEW 1061/1062 Digital Multimeter in the correct measurement category based on the following:

  • CAT III: For measurements on installations connected to the distribution network.
  • CAT II: For measurements on appliances, portable tools, and similar equipment.

User Calibration Function
The KEW 1061/1062 Digital Multimeter has a user calibration function that allows you to correct measurement errors caused by changes in the environment or aging of the instrument. Follow the instructions in the instruction manual to perform user calibration.

**Battery and Fuse Replacement

**

When the battery or fuse needs to be replaced, follow the instructions below:

Battery Replacement

  1. Remove the battery cover from the back of the instrument.
  2. Remove the old battery and insert a new one.
  3. Replace the battery cover.

Fuse Replacement

  1. Turn off the power to the instrument.
  2. Remove the fuse cover from the back of the instrument.
  3. Remove the old fuse and insert a new one.
  4. Replace the fuse cover.

Calibration and Maintenance

Regular calibration and maintenance of the KEW 1061/1062 Digital Multimeter is essential for accurate and safe operation. Follow the instructions in the instruction manual to perform calibration and maintenance.

Disposing the Product
When disposing of the KEW 1061/1062 Digital Multimeter, follow all local regulations and guidelines for electronic waste disposal.

INSTRUCTION MANUAL
DIGITAL MULTIMETER
KEW 10611062

Thank you for purchasing our Model KEW1061, KEW1062 Digital Multimeter. This user’s manual describes the specifications and handling precaution for this Digital Multimeter. Before using this Digital Multimeter, thoroughly read this manual to get a clear understanding on proper use. Always observe the following instructions. Failure to do so may impair the protection provided by the instrument and probes, and may result in electrical shock or other dangers that may lead to serious injury or the loss of life. KYORITSU is in no way liable for any damage resulting from the user’s mishandling of the product. For safe use of this product, the following safety symbols are used on the product: n About This Manual · Every effort has been made to ensure accuracy in the preparation of this manual.
However, should any errors or omissions come to the attention of the user, please contact KYORITSU. · The contents of this manual are subject to change without prior notice because of improvement in performance or function. · All rights reserved. No part of this manual may be reproduced in any from without KYORITSU’S written permission.
60

Regarding Safe Use of This Product For safe use of this product, the following safety symbols are used on the product and manual:
This indicates that the operator must refer to an explanation in the WARNING instruction manual in order to avoid the risk of serious injury or the loss
of life.

CAUTION
This indicates that the operator must refer to an explanation in the instruction manual in order to avoid the risk of injury or damage to the product.

Note
This indicates information that is essential for handling the instrument or should be noted in order to familiarize yourself with the instrument’s operating procedures and/or functions.

Danger! Handle with Care This symbol indicates that the operator must refer to an explanation in the instruction manual in order to avoid risk of injury or death of personnel or damage to the instrument.

Double Insulation This symbol indicates double insulation or reinforced insulation.

Direct Current This symbol indicates DC voltage/current.

Alternating Current This symbol indicates AC voltage/current.

DC/AC This symbol indicates AC and DC.

Fuse This symbol indicates a fuse.

Battery This symbol indicates a battery.

Ground This symbol indicates ground (earth).

61

WARNING
n Always observe the following instructions. Failure to do so may result in electrical shock or other dangers that may lead to serious injury or the loss of life.
Test leads / Test leads with crocodile clip (optional accessory) · Use the probes supplied by KYORITSU with this instrument. · Do not use test leads/test leads with crocodile clip that have deteriorated or are defective.
Check test leads/test leads with crocodile clip continuity. · Disconnect test leads/test leads with crocodile clip from the circuit under test before opening
the casing to replace the batteries or for any other reason. · Disconnect test leads/test leads with crocodile clip from the circuit under test before
attaching/detaching the test leads/test leads with crocodile clip to/from the instrument. · Disconnect test leads/test leads with crocodile clip from the instrument before opening the
casing to replace the batteries or for any other reason. · A cap is provided on the tip of a test lead.
Use a test lead with the cap on for safety (safety standards: IEC 61010-031). · Do not use the crocodile clip of test leads being loosen or removed conditions. · Stop using the test lead if the outer jacket is damaged and the inner metal or color jacket is
exposed.
Casing · Do not use the instrument if there is any damage to the casing or when the casing is
removed.
Fuses · Use fuses of the specified rating when the fuse is replaced.
Operating Environment · Do not operate the instrument in an atmosphere where any flammable or explosive gas is
present. · Avoid using the instrument if it has been exposed to rain or moisture or if your hands are
wet.
Disassembly · No person, except personnel from KYORITSU, is authorized to disassemble this instrument.

Overview

Fast and more accurate measurement The KEW1061 and KEW1062 digital multimeters use modulation for A/D conversion, which enable fast and more accurate measurement.
l Display 5-digit (LCD) Maximum Reading: 50000 Bar graph indicator
l Supports a variety of measurement function Measurement function DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature, Capacitor, Duty cycle ratio, Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower- Other functions Data Hold (D·H), Auto Hold (A·H), Peak Hold (P·H), Range Hold (R·H), Maximum value (MAX), Minimum value (MIN), Average value (AVG), Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory, LCD backlight. LowPower-: Measures resistance under lowmeasurement current. *: For model KEW1062 only
l Switching detection modes Effective value (root mean square value) detection (RMS) and mean value detection (MEAN) can be switched during AC voltage or AC current measurement (KEW1062 only).
l Low-pass filter The low-pass filter can be switched on/off during AC voltage or AC current measurement (KEW1062 only).
l Communication: optional communication package is required · Measurement data can be transferred to a PC using an optional USB communication set. The data can be read by certain applications to make trend graphs or can be converted into Excel files. · The data can also be output from an optional printer via an optional printer communication set.
l Safety design Complied standards: CE standards Uses a current-input terminal shutter for preventing wrong input. Uses high-performance UL-standard fuses.
65

Measurement Category

WARNING

n Measurement Category (CAT.) The restrictions on the maximum voltage level for which the KEW1061, KEW1062 can be used, depend on the measurement categories specified by the safety standards.

Do not apply any input level higher than maximum allowable input.

AC/DC 1000V

CAT III

AC/DC 600V

CAT

Measurement category

Description

O None, Other Other circuits that are not directly connected to MAINS.

CAT II For measurements performed on circuits directly connected to the low voltage installation.

CAT III For measurements performed in the building installation.

CAT For measurements performed all the source of the low-voltage installation.

Remarks
Appliances, portable equipment, ect. Distribution board, circuit breaker, ect. Overhead wire, cable systems, ect.

Incoming wire Socket

Interior wiring

:Device which is not directly connected to the mains power supply

Note
Radiation immunity affects the accuracy of the KEW1061, KEW1062 under the conditions specified in IEC61326-1. Use of this instrument is limited to domestic, commercial, and light industry applications. If equipment generating strong electromagnetic interference is located nearly, the instrument may malfunction.
66

3. Specifications

3.1 General Specifications

Measurement function:

DC Voltage, AC Voltage, DC Current, AC Current, Resistance,

Frequency, Temperature, Capacitor, Duty cycle ratio,

Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower-*

Other functions: Data Hold (D·H), Auto Hold (A·H), Peak Hold* (P·H),

Range Hold (R·H), Maximum value (MAX), Minimum value (MIN),

Average value (AVG), Zero Adjustment (Capacitor, Resistance),

Relative values, Save to Memory, LCD backlight.

LowPower-: Measures resistance under lowmeasurement current.

*: For model KEW1062 only

Measuring method: modulation

Display:

5-digit (LCD)/7-segment

Maximum Reading: 50000

Polarity Indicator: “­” Appears automatically when the polarity is

negative

Overrange Indicator: ” OL ”

Low-battery Indicator: ”

” Appears when the batteries become

low

Measurement cycle: 6 times per second

(except frequency measurement: one times per second, Resistance

measurement : four times per second, capacitor measurement (50mF):

max. 0.03 time per second)

Bar graph display 15 times per second

Operating temperature and humidity ranges:

-20 to 55ºC, 80%RH or less (no condensation)

70%RH or less at 40 to 55ºC.

Température de fonctionnement:

-20 à 55°C

Humidité de fonctionnement:

80% HR ou moins (sans condensation), 70% HR ou moins à 40 ~ 55°C

Storage temperature and humidity ranges:

-40 to 70ºC, 70%RH or less (no condensation)

Temperature coefficient:

(Accuracy at 23±5ºC) × 0.05/ºC or less

At -20 to 18ºC and 28 to 55ºC

When continuously DCV and DCA measuring, add 1digit / ºC

(except for 50mV, 5A, 10A range, add 3digits / ºC).

Power supply:

AA-size (R6) 1.5V batteries: 4

67

Battery life:

Approximately 100 hours

(Operating hours of alkaline batteries when in DC voltage-mode.)

Note: The battery life varies depending on the operating conditions.

Insulation resistance : 1000V DC, 100M or more

Withstand voltage: 6.88kVrms AC for five seconds

(across input terminals and casing)

External dimensions: Approximately 192(L) × 90(W) × 49(D) mm

Weight:

Approximately 560g (including batteries)

Complied standards: Safety standards

IEC61010-1, IEC61010-2-033, IEC61010-031

CAT III (Max. input voltage: AC/DC1000V)

CAT IV (Max. input voltage: AC/DC600V)

Pollution degree 2, indoor use,

2000m max. above sea level

Utilisation interne 2000m max. au-dessus du niveau de la mer.

EMC standards

IEC61326-1 Class B

Effect of radiation immunity:

In the radio-frequency electromagnetic field of 3 V/m, accuracy is

within five times the rated accuracy.

Environmental standard:

EU RoHS Directive compliant

68

Standard accessories: Batteries : 4 Test leads: 1set (M-7220A) Fuse (included): 440mA/1000V (M-8926), 10A/1000V (M-8927) Instruction manual: 1

Optional accessories: Carrying case

M-9154

(for the main unit with test leads

and communication cable)

Test leads (1set)

M-7220A

Test leads with crocodile clip (1set) M-7234

Fuse

440mA/1000V

M-8926

10A/1000V

M-8927

Temperature probes

M-8405, 8406, 8407, 8408

USB Communication set

M-8241 (Software, USB adapter and

cable)

Printer Adapter and Cable M-8243

Printer

M-8246

AC adapter (for printer, Europe) M-8248A

Thermal paper for printer (10 rolls) M-8247

69

3.2 Accuracy

Test conditions: Temperature and humidity: 23±5ºC at 80%RH or less Accuracy: ±(% of reading + digits)

Note: Each response time is a value to rated accuracy within selected range.

DC Voltage Measurement V

Range
50mV 500mV 2400mV
5V 50V 500V 1000V

Resolution
0.001mV 0.01mV 0.1mV 0.0001V 0.001V 0.01V 0.1V

Accuracy KEW1061, KEW1062
0.05+10 0.02+2 0.025+5
0.03+2

Input Resistance Approx. 100M
10M

Maximum Input Voltage
1000V DC
1000V rms AC

NMRR: 80dB or more 50/60Hz ±0.1% (70dB or more 50/60Hz ±0.1% when 50mV Range)
CMRR: 100dB or more 50/60Hz (Rs=1k) Response time: 0.3 sec max.

AC Voltage Measurement [RMS] KEW1061

V AC Coupling, Rms-value detection, Crest factor*: <3

Range
500mV 5V 50V
500V 1000V *

Resolution 10 to 20Hz

0.01mV 0.0001V 0.001V
0.01V 0.1V

1.5+30 1 2

Accuracy 20Hz to 1k to
1kHz 10kHz
0.7+30 1
2 3+30 *2

10k to 20kHz
2+50 *2

: Crest factor <1.5 at 1000V range Accuracy 1: At 5 to 100% of range, *2: At 10 to 100% of range CMRR: 80dB or more DC to 60Hz (Rs=1k) Response time: 1 sec max.

Input Impedance

Maximum Input Voltage

11M <50pF
10M <50pF

1000V rms AC
1000V DC

70

AC Voltage Measurement [RMS]

V

KEW1062

Range

Resolution

AC Coupling, Rms-value detection, Crest factor*: <3

Accuracy
10 to 20Hz to 1k to 10k to 20Hz 1kHz 10kHz 20kHz

20k to 50kHz

50k to 100kHz

Input Impedance

Maximum Input Voltage

50mV 0.001mV
500mV 0.01mV 5V 0.0001V 50V 0.001V
500V 0.01V
1000V* 0.1V

2+80 2
1+30
1
*2

0.4+40 5+40 5.5+40 2 2 *2

0.4+30 *1

1+40 *1

*2

3+30 *2

15+40 *2

2+70 5+200

*2

*2

11M <50pF
10M <50pF

1000V rmsAC
1000V DC

: Crest factor <1.5 at 1000V range Accuracy 1: At 5 to 100% of range, *2: At 10 to 100% of range CMRR: 80dB or more DC to 60Hz (Rs=1k) Response time: 1 sec max.

AC Voltage Measurement [MEAN]

V

KEW1062

Range

Resolution

AC Coupling, MEAN value detection, RMS value calibration (sine wave)

10 to 20 Hz

Accuracy 20 to 500 Hz

Input 500Hz to 1 kHz Impedance

Maximum Input Voltage

50mV 0.001mV
500mV 0.01mV 5V 0.0001V 50V 0.001V
500V 0.01V 1000V 0.1V

4+80 2
2+30
1
*2

1.5+30 2
1+30
1
*2

5+30 2
3+30
1
*2

11M <50pF

1000V rmsAC

10M <50pF

1000V DC

Accuracy 1: At 5 to 100% of range, 2: At 10 to 100% of range CMRR: 80dB or more DC to 60Hz (Rs=1k) Response time: 1 sec max.

71

DCV+ACV + KEW1061

Maximum Reading 50000, Crest factor*: <3

Range

Accuracy Resolution DC, 10 to DC, 20Hz DC, 1k to DC, 10k
20Hz to 1kHz 10kHz to 20kHz

5V
50V 500V 1000V*

0.0001V
0.001V 0.01V 0.1V

1.5+10 1
2

1+10 1
2

2+10 *2

Input Impedance 11M <50pF
10M <50pF

Maximum Input Voltage
1000V rms AC
1000V DC

DCV+ACV + KEW1062

Maximum Reading 50000, Crest factor*: <3

Range
5V 50V 500V 1000V*

Resolution
0.0001V 0.001V 0.01V 0.1V

DC, 10 to 20Hz

DC, 20Hz to
1kHz

Accuracy DC, DC, 1k to 10k to 10kHz 20kHz

DC, 20k to 50kHz

1.5+10 *1

0.5+10 *1

1+10 2+10

*1

*2

*2

*2

DC, 50k to 100kHz

Input Impedance

Maximum Input Voltage

11M 1000V

5+20 <50pF rms AC

*2

10M <50pF

1000V DC

: Crest factor <1.5 at 1000V range Accuracy 1: At 5 to 100% of range, *2: At 10 to 100% of range
CMRR: 80dB or more DC to 60Hz (Rs=1k) Response time: Approx. 2 sec

DC Current Measurement A

Range
500A 5000A
50mA 500mA*3
5A 10A

Accuracy Resolution
KEW1061, KEW1062

0.01A

0.1A 0.001mA

0.2+5

0.01mA

0.0001A

0.6+10

0.001A

0.6+5

Voltage Drop <0.11mV/A
<4mV/mA <0.1V/A

Maximum Input Current
440mA Protected by a 440mA/1000V fuse.
10A Protected by a 10A/1000V fuse.

*3: Maximum measurement current : 440mA at 500mA range Response time: 0.3 sec max.
Note : After measuring over 500mA DC (especially 10A DC), “Zero error” occurs for a while. In that case, please wait for a while at zero input until the value stabilizes before measuring again.

72

AC Current Measurement [RMS]

A

KEW1061

Rms-value detection, Crest factor: <3

Range Resolution

500A 5000A
50mA 500mA*3
5A 10A

0.01A 0.1A 0.001mA 0.01mA 0.0001A 0.001A

Accuracy Voltage Drop
10 to 20Hz 20Hz to 1kHz

Maximum Input Current

1.5+20

1+20

<0.11 mV/A 440mA Protected

by a 440mA/

<4mV/mA

1000V fuse.

<0.1V/A

10A Protected by a 10A/1000V fuse.

AC Current Measurement [RMS]

A

KEW1062

Rms-value detection, Crest factor: <3

Range

Resolution

Accuracy

10 to 20Hz 20Hz to 1kHz 1k to 5kHz

Voltage Drop

Maximum Input Current

500A 5000A
50mA 500mA *3

0.01A 0.1A
0.001mA 0.01mA

1+20

0.75 +20

1+30

<0.11 mV/A 440mA Protected
by a 440mA/ <4 mV/mA 1000V fuse.

5A

0.0001A 1.5+20

10A

0.001A

1+20

2+30

<0.1 V/A

10A Protected by a 10A/1000V fuse.

Model KEW1061/1062 Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range *3: Maximum measurement current : 440mA at 500mA range Response time: 1 sec max.

73

AC Current Measurement [MEAN]

A

KEW1062

MEAN value detection, RMS value calibration (sine wave)

Range
500A 5000A
50mA 500mA*3
5A
10A

Accuracy Resolution 10 to 20Hz 20 to 500Hz

0.01A 0.1A 0.001mA 0.01mA 0.0001A
0.001A

2+20 3+20

1.5 +20 2+20

500Hz to 1kHz 2+30
4+30

Voltage Maximum Input

Drop

Current

<0.11 mV/A 440mA Protected
by a 440mA/ <4 mV/mA 1000V fuse.

10A Protected <0.1 V/A by a 10A/
1000V fuse.

Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range *3: Maximum measurement current : 440mA at 500mA range Response time: 1 sec max.

DCA+ACA KEW1061
Range
500A 5000A
50mA 500mA*3
5A 10A

  • Resolution
    0.01A 0.1A 0.001mA 0.01mA 0.0001A 0.001A

Maximum Reading 50000, Crest factor: <3

Accuracy

DC, 10 to DC, 20Hz to Voltage Drop

20Hz

1kHz

Maximum Input Current

2+10

1.5+10

<0.11 mV/A 440mA Protected by a
<4mV/mA 440mA/1000V fuse.

<0.1V/A

10A Protected by a 10A/1000V fuse.

Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range *3: Maximum measurement current : 440mA at 500mA range Response time: 2 sec max.

74

DCA+ACA + KEW1062

Maximum Reading 50000, Crest factor: <3

Range

Accuracy Resolution DC, 10 to DC, 20Hz DC, 1k to Voltage Drop
20Hz to 1kHz 5kHz

Maximum Input Current

500A 5000A
50mA 500mA*3

0.01A 0.1A 0.001mA 0.01mA

1.5+10

1+10

1.5+10

<0.11 mV/A 440mA Protected

by a 440mA/

<4 mV/mA

1000V fuse.

5A 10A

0.0001A 0.001A

2+10

1.5+10

3+10

<0.1 V/A

10A Protected by a 10A/1000V fuse.

Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range *3: Maximum measurement current : 440mA at 500mA range Response time: Approx. 2 sec

Resistance Measurement

Range
500 5k 50k
500k 5M
50M

Accuracy

Maximum

Resolution

Measuring

KEW1061 KEW1062 Current

0.01

<1mA

0.0001k 0.001k

0.1+2 *1

0.05+2 <0.25mA

*1

<25A

0.01k

<2.5A

0.0001M

0.5+2

<1.5A

0.001M

1+2

<0.13A

*1: Accuracy is specified after zero adjustment (resistance). Response time: 1 sec max. at 500 to 500k
5 sec max. at 5M to 50M

Open-loop Voltage
<2.5V

Input Protective Voltage
1000V rms

LowPower- LP-

Range Resolution

5k 50k 500k 5M

0.001k 0.01k 0.1k 0.001M

Accuracy KEW1062 only
0.2+3 1+3

Maximum Measuring
Current
<10A <1.0A <0.6A <0.05A

Maximum Reading 5000

Open-loop Input Protective

Voltage

Voltage

<0.7V

1000V rms

LowPower-: Measures resistance under lowmeasurement current.

75

Continuity Check

Range 500

Resolution 0.1

Range of Operation
The buzzer turns on for resistances lower
than 100±50.

Measuring Current
Approx. 0.5mA

Maximum Reading 5000

Open-loop Input Protective

Voltage

Voltage

<5V

1000V rms

Diode Test Range 2.4V

Resolution 0.0001V

Accuracy

Measuring Current (Vf=0.6V)

Open-loop Voltage

Input Protective Voltage

1+2

Approx. 0.5mA

<5V

1000V rms

Temperature Measurement TEMP

Range

Resolution

Accuracy

-200 to 1372°C -328 to 2501.6°F

0.1°C 0.1°F

1%+1.5°C 1%+2.7°F

Use optional Temperature Probe: Thermocouple Type K

Capacitor Measurement

Range 5nF 50nF
500nF 5F 50F
500F 5mF 50mF

Resolution 0.001nF 0.01nF 0.1nF 0.001F 0.01F 0.1F 0.001mF 0.01mF

Accuracy
1+5 *1
2+5 3+5

*1: Accuracy is specified after zero adjustment (capacitor).

Input Protective Voltage
1000V rms
Maximum Reading 5000 Input Protective Voltage
1000V rms

76

Frequency Measurement Hz
Range (AUTO) 2.000 to 9.999Hz
9.00 to 99.99Hz 90.0 to 999.9Hz 0.900 to 9.999kHz 9.00 to 99.99kHz

AC Coupling, Maximum Reading 9999

Resolution 0.001Hz 0.01Hz 0.1Hz 0.001kHz 0.01kHz

Accuracy
0.02+1 1
2

Accuracy 1: At 10 to 100% of input voltage or current range 2: At 40 to 100% of input voltage or current range

Duty cycle ratio % Range
10 to 90%

Resolution 1%

Accuracy ±1%*1

Accuracy *1: At 10.00Hz to 500.0Hz, square wave At 40 to 100% of input voltage or current range

Peak Hold P·H Model KEW1062 only
Range DCV, DCA

Accuracy ±100 digit

Maximum Reading 5000
Response Time Maximum >250S

77

4. Operation
4.1 Precautions Before Measurement
n Examining Items Contained in the Package After opening the package, be sure to examine the product as instructed below before use. Should the delivered product be the wrong model, lack any item, or show any flaw in its appearance, contact the vendor from which you purchased the product.
n Precautions of Operation and Storage
CAUTION
· Insert the batteries in the instrument by referring to “6.1 Battery Replacement.” · A Blank cover is provided on the upper part of back casing.
Don’t remove the Blank cover except when the USB adapter or Printer adapter is connected. · Do not use the instrument near noise-emitting equipment or where there may be a sudden change of temperature. Otherwise, the instrument may give an unstable reading or errors. Removal of Dirt Do not wipe the instrument using any solvent (chemicals) such as benzine or paint thinner, as this may damage or discolor the front panel. Use a dry cloth to clean the instrument. Storage Conditions · Do not leave the instrument exposed to direct sunlight or in a hot and humid location such as the inside of a vehicle, for any prolonged length of time. · If the instrument will not be used for a prolonged period, remove the batteries.
78

Caps of Testing leads Protective fingerguard

4.2 Components
n Panel Description

n Test leads
Display (LCD)
SHIFT key RANGE key HOLD (SAVE) key LIGHT (READ) key SELECT (PRINT) key REL /% key MIN/MAX (LOG) key MEMORY key
Function switch
Input terminal block

Red

Black

With: 1000V10A CAT. /600V 10A CAT. Without: 1000V10A CAT. /600V 10A CAT.

n Test leads with crocodile clip(optional accessory)

Protective fingerguard

Red

Protective fingerguard

Black

Cable

Crocodile clip

79

Cap

Protective fingerguard

Protective fingerguard (Barrier) It is a part providing protection against electrical shock and ensuring the minimum required air
and creepage distances.
Cap Uncapped condition for CAT II environment Capped condition for CAT III/ IV environments The Cap shuld be firmly attached to the probes.

  1. Function switch Turns off the power or select the measurement mode (function).

OFF

Turns off the power.

Resistance measurement

V AC voltage (V) measurement

Capacitor measurement

mV AC voltage (mV) measurement V DC voltage (V) measurement mV DC voltage (mV) measurement
Continuity check, Diode Test

TEMP
A mA A

Temperature measurement DC/AC current measurement

  1. SELECT key Pressing this key in each measurement modes (function) described above selects other measurement modes (function).

V,
V µ A/mA/A

mV 1

HzV

: Frequency measurement (Voltage value is displayed.)

2

Hz%

: Frequency measurement (Duty cycle ratio)

3

dBV

: dBV measurement (Voltage value is displayed.)

4

dBm

: dBm measurement

(DC+AC) measurement, (DC, AC) Dual display

LP- (Model KEW1062 only)

Diode Test

Pressing this key in each measurement modes (function) described above selects other measurement modes (function).

1

: AC voltage measurement

2

: (DC+AC) measurement

3

·

: (DC, AC) Dual display

4

Hz%

: Frequency measurement (Duty cycle ratio)

80

  1. RANGE key Allows the operator to select the measuring range. Fixed ranges : The display shows the ” R·H ” symbol. The range increases every time this key is pressed. AUTO range : The display shows the ” AUTO ” symbol. To return to the auto-ranging mode, hold down RANGE key for more than one second.

  2. HOLD key Selects between the DATA HOLD, AUTO HOLD and PEAK HOLD functions. To cancel functions, press this key once again. DATA HOLD: Holds the display readings. The display shows the ” D·H ” symbol. AUTO HOLD: Holds the measured value when the Test leads are handled. The display shows the ” A·H ” symbol. PEAK HOLD: Holds the peak value. The display shows the ” P·H ” symbol. (Model KEW1062 only)

  3. LIGHT key LIGHT key: Use to turn on the LCD backlight. Press this key once to turn on the LCD backlight for approximately one minute. The LCD backlight is lit for approximately one minute. (To postpone turned on time, press this key once again.) To cancel the function, hold down this key for more than one second.

  4. REL / % key The instrument can calculate relative values or differences, and percentage values from the reference measurement values. 1 : Relative Calculation The display shows the ” ” symbol. The sub-display shows the reference voltage value.
    2 : Percentage Calculation The display shows the ” “, ” % ” symbol. The sub- display shows the reference voltage value.

  5. MIN/MAX key Displays the minimum value (MIN), maximum value (MAX) and average value (AVG) during measurement.
    Pressing this key starts recording and at the same time the display shows MIN/MAX/AVG to release AUTO POWER OFF.
    81

  6. MEMORY key Data can be stored in internal memory using this key. Used when outputting to printer with the optional adapter and cable.

  7. SHIFT key While this key is pressed, “Shift” appears on the display. Pressing the following keys with the SHIFT key held down enables the following settings.

SHIFT+

LIGHT key RANGE key REL key HOLD key

Set-up function Change to [RMS] mode Change to [MEAN] mode Turn filter on/off

(KEW1062 only) (KEW1062 only) (KEW1062 only)

n Display (LCD) Description

82

Symbol and Unit
+ ­
R·H AUTO
AUTO OFF RMS LP- Filter Shift nF, F, mF mV, V A, mA, A M, k, °C / °F kHz, Hz dB, dBm % (Main-display) % (Sub- display) mV, V (Sub- display) s (Sub- display) (Sub- display)
OL

Description
Appears when in DC-mode measurement
Appears when in AC-mode measurement Appears when in DC+AC-mode measurement Appears when the polarity is negative
Appears when in diode test
Appears when in continuity check
Relative calculation indicator
Fixed ranges indicator AUTO range indicator DATA HOLD indicator AUTO HOLD indicator PEAK HOLD indicator
Lit when in MIN/MAX/AVG-mode
Lit when in MIN/MAX/AVG-mode
Lit when in MIN/MAX/AVG-mode
Lit when in Memory function Auto power off indicator Appears in RMS mode Appears in Low Power- measurement Appears while filter is on Appears while the SHIFT key is held down Unit for capacitance measurement Unit for voltage measurement Unit for current measurement Unit for resistance measurement Unit for temperature measurement Unit for frequency measurement Decibel calculation indicator Unit for percentage calculation Unit for duty cycle ratio calculation Unit for voltage measurement (dBV, HzV) Unit for recording time when in MIN/MAX/AVG-mode Recording time indicator when in MIN/MAX/AVG-mode Number of saved data indicator Reference value indicator when relative calculation Duty cycle ratio indicator Voltage value (dBV, HzV) indicator Reference resistance value (dBm) indicator
Overrange Indicator
Appears when the batteries become low
Bar graph indicator, Range indicator

83

4.3 Measuring Instructions
WARNING
To avoid damage to instrument or equipment · Before starting measurement, make sure that the position of function switch and the
input terminals for connecting the test leads are appropriate for the desired mode of measurement. · Temporarily remove the test leads from the device under test before operating the function switch. · Verify proper operation on a known source before use or taking action as a result of the indication of the instrument. · Vérifiez le fonctionnemnet adéquat sur une source connue avant de procéder tout en se basant sur un affichage trompeur de l’instrument.” ou une annotation équivalente sera reprise dans le manuel d’utilisation. · Stop using the test lead if the outer jacket is damaged and the inner metal or color jacket is exposed. · Arrêtez d’utiliser le cordon de test si la gaine extérieure est endommagée et si la gaine métallique ou interne est exposée.

Test leads here include a test leads with crocodile clip (optional accessory). 4.3.1 AC Voltage Measurement ( V, mV)

  1. Turn the function switch to the ” V ” or ” mV ” position. 2) Plug the test leads into the input terminals. 3) Connect the test leads to the circuit under test and then read the value
    when it stabilizes.

Test leads

Black

Red

4.3.2 DC Voltage Measurement ( V, mV) 1) Turn the function switch to the ” V “or ” mV” position. 2) Plug the test leads into the input terminals. 3) Connect the test leads to the circuit under test and then read the value
when it stabilizes.

Test leads

Black

Red

Note
If ” mV “range is selected and the test leads are left open-circuited, the instrument may give a certain reading. This dose not affect your measurement.

84

4.3.3 DC+AC Voltage Measurement ( + )

  1. Turn the function switch to the ” V ” position. 2) Press the SELECT key to select DC+AC voltage measurement.
    (The display shows the ” + ” symbol.) 3) Plug the test leads into the input terminals. 4) Connect the test leads to the circuit under test and then read the value
    when it stabilizes.

Test leads

Black

Red

4.3.4 DC, AC Voltage Dual Display ( · ) 1) Turn the function switch to the “DCV” position. 2) Press the SELECT key twice to change the mode to DC/AC voltage dual
display. DC voltage measurement appears on the main display and AC voltage on the sub-display. 3) Plug the test leads into the input terminals. 4) Connect the test leads to the circuit under test and then read the value when it stabilizes.

Test leads

Black

Red

4.3.5 Resistance Measurement ()

CAUTION

To avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to the instrument.

  1. Turn the function switch to the ” ” position. 2) Plug the test leads into the input terminals. 3) Connect the test leads to the circuit under test and then read the value
    when it stabilizes.

Test leads

Black

Red

Note
Zero adjustment Zero adjustment is recommended for correct measurement. After executing 1), 2) above, short the two test leads. Press the REL key for adjust. (The display shows the “0.0” value.) The value (zero adjustment) stores till turn off.

85

4.3.6 Low Power- (LP-)

This function is for measuring the resistance of parts on a printed board under low measurement

current.

In Low Power- measurement, up to 5,000 counts are displayed and the measurement range

is from 5 k to 5 M.

  1. Turn the function switch to the “” position. Press the SELECT key to set LP-mode. “LP-” appears on the display.

  2. Plug the test leads into the input terminals.

  3. Connect the test leads to the resistance and then read the value when it

Test leads

stabilizes.

Black

Red

4.3.7 Continuity Check ( )
CAUTION

To avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to the instrument.

  1. Turn the function switch to the ” ” position. 2) Plug the test leads into the input terminals. 3) Connect the test leads to the circuit under test. If the circuit is continuous
    (no more than approximately 100), the buzzer sounds.

Test leads

Black

Red

86

4.3.8 Diode Test ( )
CAUTION
To avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to the instrument.

  1. Turn the function switch to the ”

· ” position.

Press the SELECT key to select Diode test.

(The display shows the

symbol.)

  1. Plug the test leads into the input terminals.

  2. Connect the test leads to the diode and then read the value when it

stabilizes.

Test leads

Black

Red

Connect the black test lead to the cathode and the red test lead to the anode. Silicon diodes should give a reading of approximately 0.5V and light-emitting diodes a reading between approximately 1.5V and 2.0V. Connect the black test lead to the anode and the red test lead to the cathode. Normally, the display shows the ” OL ” symbol, indicating that the diode under test is normal. The diode is defective if the display gives a certain voltage level.

Black test lead

Red test lead

Red test lead

Black test lead

Figure 1 Forward-bias Diode Test

Figure 2 Reverse-bias Diode Test

87

4.3.9 Temperature Measurement (TEMP)
CAUTION
To avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to the instrument.

Note
Optional Temperature probe is required for temperature measurement. Temperature Probe: Thermocouple Type K
Model: 8405, 8406, 8407, 8408 Check the measurable range of respective probes.

  1. Turn the function switch to the ” TEMP ” position. 2) Plug the measuring probe into the input terminals. 3) Contact the measuring probe to the under test and then read the value
    when it stabilizes.

Measuring

probe

Black

Red

Note
The default temperature read-out of the Digital Multimeters is in Celsius (°C). To change it to Fahrenheit (°F), it is necessary to proceed as follows:
Changing the temperature unit setting to Fahrenheit Displaying ” ºC ” only is configured at factory before shipment. Carry out the following setting procedure to display ” ºF “. While pressing the SELECT, RANGE and HOLD keys simultaneously, turn the function switch to the ” TEMP ” position. Then, upon pressing the SELECT key, the temperature unit switches from ºC to ºF. Once the temperature is displayed in ºF, press the SELECT key to alternately switch the temperature units between ºF and ºC.
The conversion from Celsius to Fahrenheit is performed using the equation Fahrenheit temperature = 1.8×Celsius temperature + 32

88

4.3.10 Current Measurement (µA/mA/A)
WARNING

To avoid damage to instrument or equipment · Before starting measurement, make sure that the position of function switch and the
input terminals for connecting the test leads are appropriate for the desired mode of measurement. · The maximum input current (limited by fuses) of the “µA” and “mA” ranges is 440 mA. Be sure not to exceed the limit in the 500 mA range.
Be careful not to burn yourself · When measuring more than 6A under exceeding 40 conditions, the continuous
measuring time shall be within 3 minutes, then keep disconnected for more than 10 minutes. · En mesurant plus de 6A dans des conditions en dessous de 40°C, le temps de mesure ininterrompue se situera dans les 3 minutes; déconnectez l’instrument ensuite pendant plus de 10 minutes.” ou une annotation équivalente sera reprise dans le manuel d’utilisation.

  1. Turn the function switch to the “µA ” , “mA ” or “A “position. (If the magnitude of the current being measured is not known, select the “A “position. Make sure the current being measured is no more than 440mA before the “µA ” or “mA ” position is selected.)
  2. Please select between DC and AC. When selecting AC, press the SELECT key.
  3. Plug the black test lead into the ” COM ” input terminal and the red test lead into the “A “input terminal. If the current is in the order of mA or less, plug the red test lead into the “µA · mA” input terminal.
  4. Connect the test leads to the circuit under test and then read the value when it stabilizes.

Test leads Red

Black

Test leads

Red

Black

89

4.3.11 DC+AC Current Measurement ( + )
WARNING

To avoid damage to instrument or equipment · Before starting measurement, make sure that the position of function switch and the
input terminals for connecting the test leads are appropriate for the desired mode of measurement. · The maximum input current (limited by fuses) of the “µA” and “mA” ranges is 440 mA. Be sure not to exceed the limit in the 500 mA range.
Be careful not to burn yourself · When measuring more than 6A under exceeding 40 conditions, the continuous
measuring time shall be within 3 minutes, then keep disconnected for more than 10 minutes. · En mesurant plus de 6A dans des conditions en dessous de 40°C, le temps de mesure ininterrompue se situera dans les 3 minutes; déconnectez l’instrument ensuite pendant plus de 10 minutes.” ou une annotation équivalente sera reprise dans le manuel d’utilisation.

  1. Turn the function switch to the “µA” , “mA” or “A” position. (If the magnitude of the current being measured is not known, select the “A” position. Make sure the current being measured is no more than 440mA before the “µA” or “mA” position is selected.)
  2. Press the SELECT key twice to select the DC +AC measurement. (The display shows the + symbol.) Plug the black test lead into the ” COM ” input terminal and the red test lead into the “A” input terminal. If the current is in the order of mA or less, plug the red test lead into the “µA · mA” input terminal.
  3. Connect the test leads to the circuit under test and then read the value when it stabilizes.

Test leads Red

Black

Test leads

Red

Black

90

4.3.12 DC, AC Current Dual Display ( · ) 1) Turn the function switch to the “µA,” “mA” or “A” position.
(If the magnitude of the current being measured is not known, select the “A” position. Make sure the current being measured is no more than 440 mA before the “µA” or “mA” position is selected.) 2) Press the SELECT key three times to select DC/AC dual display. DC current measurement appears on the main display and AC current on the sub-display. 3) Plug the black test lead into the “COM” input terminal and the red test lead into the “A” input terminal. If the current is in the order of mA or less, plug the red test lead into the “A/mA” input terminal. 4) Connect the test leads to the circuit under test and then read the value when it stabilizes.
4.3.13 Capacitor Measurement ( )
CAUTION

Test leads Red

Black

Test leads

Red

Black

To avoid damage to instrument · Turn off the power to the circuit under test before starting measurement in order to prevent
any excessive voltage from being applied to the instrument. · Before starting measurement, be sure to discharge the capacitor under check.

  1. Turn the function switch to the ” “position. 2) Plug the test leads into the input terminals. 3) Open the test lead and press the REL key in 5nF range to adjust the
    capacitance to zero. (The display shows ” 0.000 “.)
  2. Connect the test leads to the circuit under test and then read the value when it stabilizes.

Test leads

Black

Red

Note
The value (zero adjustment) remains displayed until power-off.

91

4.3.14 Frequency Measurement (Hz), Duty cycle ratio (Hz)
CAUTION
To avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to the instrument.

  1. Turn the function switch to the voltage( V, mV) or the current (µ A, mA, A ) position.
  2. Press the SELECT key to select the range of frequency. (The display shows the unit of frequency.)
  3. Plug the test leads into the input terminals. Plug the red test lead into the suitable input terminal (” A” or “µA·mA”) for current value when current measurement.
  4. Connect the test leads to the under test and then read the value when it stabilizes. The display shows frequency value in main-display and duty cycle ratio value in sub-display.

Test leads

Black

Red

Test leads

Red

Black

Test leads

Red

Black

92

4.3.15 Function to change RMS detection to/from MEAN detection mode (KEW1062 only) The instrument has a function to change RMS detection to/from MEAN detection modes.

1) Select the appropriate AC measurement mode (ACV, ACmV, ACµA, ACmA, ACA) by using the function switch and the SELECT key. 2) Press the SHIFT key to display “Shift” on the display. 3) Press the REL key while holding down the SHIFT key to change to MEAN detection mode. “RMS” disappears on the display. 1) Select the appropriate AC measurement mode (ACV, ACmV, ACµA, ACmA, ACA) by using the function switch and the SELECT key. 2) Press the SHIFT key to display “Shift” on the display. 3) Press the RANGE key while holding down the SHIFT key to change to RMS detection mode. “RMS” appears on the display.

4.3.16 Function to turn the filter on/off (KEW1062 only) The instrument has a function to turn the filter on/off during AC measurement.

  1. Select the appropriate AC measurement mode (ACV, ACmV, ACµA, ACmA, ACA) by using
    the function switch and the SELECT key. 2) Press the SHIFT key to display “Shift” on the display. 3) Then press the HOLD key to turn the low-pass filter on.
    While the filter on, “Filter” appears on the display. Refer to filter characteristics in the diagram below.

Amplitude ratio [dB]

0 -10 -20 -30 -40 -50 -60 -70
10

Third-order low-pass filter

100 Frequency [Hz] 1,000

10,000

  1. Repeat step 2) and 3) to turn the filter off. (“Filter” disappears from the display.)

93

4.3.17 AUTO HOLD Function The instrument can automatically retain the measured value when the test leads are handled as described below. 1) Press the HOLD key to select Auto hold function.
(The display shows the” A·H ” symbol.) 2) Connect the test leads to the circuit under test. 3) When the reading stabilized, the buzzer sounds. 4) Remove the test leads from the circuit under test. 5) The display shows the measured value that is retained.
You can repeat steps 2) to 4)as many times as you like as long as the display shows the ” A· H ” symbol.
Note
· In DC/AC voltage measurement, the Auto hold function is only available for ranges greater than the 5V range.
· This function is not available for Temperature, Capacitor and Frequency measurement. · The Auto hold function can not be applied to unstable signals.
94

4.3.18 PEAK HOLD Function This instrument can always detect, update and display the peak value (instantaneous) in DCV and DCA measurement. The peak value of the wave can be seen.

  1. Turn the function switch to DCV or DCA position. 2) Connect the test leads to the circuit under test. 3) Press the HOLD key to select Peak Hold. (The display shows the ” P·H ” symbol.) 4) The display shows the peak value. 5) When resetting the peak value in HOLD, press the MIN/MAX key.
    Then new peak value can be in HOLD.
    Note
    Even though the input signals (DCV, DCA) have negative polarity, the peak value can be measured when the peak is in the positive direction. Relative values from the reference values can be shown during the peak value measuring.
  2. Press the REL /% key to relative calculation in PEAK HOLD mode. The display shows the ” ” symbol and the relative peak value.
  3. Press the REL /% key once again to percentage calculation. The display shows the ” % ” symbol and the percentage peak value.
    SEE ALSO Next section ” Relative and percentage calculation”
    When resetting the peak value, press the MIN/MAX key. Then new peak value can be in HOLD.
    To cancel the percentage calculation, Press the REL /% key again. ” % ” symbol disappears, then retrieves the PEAK HOLD mode.
    4.3.19 Relative and percentage calculation The instrument can calculate relative values or difference, and percentage values from the reference measurement values. (The range will be fixed.)
    <Relative (REL) calculation> Subtracts the reference value from the measured value to display the relative value or difference.
  4. Take a measurement to set the reference value. 2) Press the REL /% key.
    (The display shows the” ” symbol and the sub-display shows the reference value.) 3) Take another measurement.
    95

<Percentage (%) calculation> Calculates and display the percentage value according to the following equation: % value =(measured value ­ reference value)/reference value

  1. Take a measurement to set the reference value. 2) Press the REL /% key.
    (The display shows the” ” symbol and the sub-display shows the reference value.) 3) Take another measurement.
    Press the REL /% key again. (The display shows the” % ” symbol.)
    4.3.20 Decibel calculation (dBm, dBV) The instrument can perform logarithmic calculations on a AC voltage.

dBm : 20log dBV : 20log

Measured voltage value Reference resistance value × 10-3 (1mW/Reference resistance ()=0dBm)
Measured voltage value 1(V)

  1. Turn the function switch to the V or mV position. 2) Press the SELECT key to select dBm dBV .
    (The display shows the ” dBm “, ” dB ” symbol.) 3) Connect the test leads to the circuit under test and then read the value when it stabilizes. 4) When calculating relative value, press the REL /% key.
    Note
    The instrument can switch (select) reference resistance value when measuring dBm. The reference resistance value is switched as follows every time the RANGE key is pressed. (Shown in sub-display.)
    Reference resistance value: 4,8,16,32,50,75,93,110,125,135,150 200,250,300,500,600,800,900,1000,1200 Default value: 600
    The default settings can be changed. Refer to the Set-up function.

96

4.3.21 MIN/MAX/AVG Function The minimum value (MIN), maximum value (MAX) and average value (AVG) during measurement are shown. (The range is fixed.) The average value is shown by dividing the integrated record data by the number of recording times. Pressing this key starts recording and at the same time the display shows “MIN”, “MAX” and “AVG” to release AUTO POWER OFF.

The timer is activated to show the elapsed time from the start and simultaneously the renewed time for MIN/MAX is also recorded. The elapsed time is displayed as follows: 0 sec. to 99 min. and 59 sec.: steps of 1 sec. 100 min. or more: steps of 1 min. Press the HOLD key to stop recording. (The display shows the ” D·H ” symbol.) For confirming the recording time, press the MIN/MAX key. Subsequent pressing of this key repeats to display the present minimum value (MIN), maximum value (MAX) and average value (AVG). Press the HOLD key once again to restart recording. To cancel the confirming mode, hold down the MAX / MIN key for one second. (“MAX” “MIN” “AVG” symbol disappears.) Note · No influence is exerted on the recorded data even if the test leads are disconnected while the recording is stopped. · If overload is recorded, the MIN or MAX display changes to ” OL ” display, resulting in incorrect average data. · For widely varying signal measurement, set the appropriate range in which the MAX or MIN does not change to ” OL ” display. 97

4.4 Memory Function

The instrument can save a data using with the following two types of modes. SAVE-mode: Saves a data for one measurement by manual operation. LOGGING-mode: Automatically saves a data from the start of logging.

Memory capacity SAVE-mode: 100 data LOGGING-mode: Logging data of one time

Model KEW1061 1,000 data Model KEW1062 10,000 data

Number of saved data
Number of saved data is 4-digit numbers. When LOGGING-mode, ” L ” is attached
to the top of 4-digit numbers. The instrument allocates the smallest number, between 0000 to 9999, that has not yet been used. Use the (RANGE) key or (REL /%) key switches the number of saved data.

To save a Data (SAVE-mode)

  1. Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the SAVE (HOLD) key.
    (The display shows the number of saved data.) 3) Press the SAVE (HOLD) key to save the data.
    Another press of the SAVE (HOLD) key saves a data for the second time measurement or later. 4) To cancel the function, hold down the MEMORY key for one second. (” MEM ” symbol disappears.)

Note
HOLD data can be saved. Hold the display and save it according to the above steps.

The number of saved data

98

To save a Data (LOGGING-mode) In logging measurement mode, the time needs to be set. Note that changing the batteries resets the time to 00:00. Set the time by referring to the Set-up function.

  1. Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the LOG (MIN/MAX) key.
    (The display shows the logging interval (period).) Set the value with the (RANGE) key or (REL /%) key. The default setting is one second. (The default settings can be changed. Refer to the Set-up function.) The display shows ” FULL ” when the logging data is already saved. When saving the new data, perform delete of data. 3) Press the LOG (MIN/MAX) key to start logging. (The ” MEM “symbol is flashing.) Every time the MIN/MAX key is pressed, the sub- display changes. (Sub display : saved number saved time (min : sec)(hour : min)saved number) 4) To cancel the function, hold down the MEMORY key for one second. When memory capacity becomes full, the function is automatically canceled. (” MEM ” symbol disappears.)
    Note
    LOGGING-mode operation during HOLD-mode disables HOLD-mode.
    To load a Data (SAVE-mode) 1) Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the READ (LIGHT) key. 3) Press the SAVE (HOLD) key to select the number of saved data.
    Select the number with the (RANGE) key or (REL /%) key. 4) To cancel the function, hold down the MEMORY key for one second.
    (” MEM ” symbol disappears.)
    To load a Data (LOGGING-mode) 1) Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the READ (LIGHT) key. 3) Press the LOG (MIN/MAX) key to select the number of saved data.
    Select the number with the (RANGE) key or (REL /%) key. Every time the MIN/MAX key is pressed, the sub-display changes. (Sub display : saved number saved time (min : sec)(hour : min)saved number) 4) To cancel the function, hold down the MEMORY key for one second. (” MEM ” symbol disappears.)
    99
Delete method (SAVE-mode) · To delete all data 1) Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Hold down the SAVE (HOLD) key for one second. (The display shows the ” CLr ? “symbol.) 3) Press the SAVE (HOLD) key. All data is deleted. · To overwrite selected data 1) Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the SAVE (HOLD) key. (The sub-display shows the number of saved data.) 3) Use the (RANGE) key or (REL /%) key to select the number of saved data. 4) Press the SAVE (HOLD) key to save (over write) the data. 5) To cancel the function, hold down the MEMORY key for one second. (“MEM” symbol disappears.) Delete method (LOGGING-mode) · To delete all data 1) Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Hold down the LOG (MIN/MAX) key for one second. (The display shows the ” CLr ? “symbol.) 3) Press the LOG (MIN/MAX) key. All data is deleted. LOG MIN/MAX 100

4.5 AUTO POWER OFF Function

The display shows the ” AUTO OFF “indication. · The instrument automatically turns off twenty minutes after the last key operation. The instrument will beep for approximately 30 seconds to alert the operator before the AUTO POWER OFF function takes effect. · Pressing any key or switch while the instrument is beeping postpones the power-off time. · Turning the function switch once after the power to the instrument is automatically turned off switches the instrument on again. 1) Turn the function switch to the OFF. 2) With pressing the HOLD key, turn the function switch to the desired position of any measurement mode (function). The ” AUTO OFF ” indication turns off when the function is canceled. Note Additional functions simply set when POWER ON can be used. 1) Turn the function switch to the OFF. 2) Turn the function switch to the desired position of any measurement mode (function). The AUTO POWER OFF function is enabled again. The display shows the ” AUTO OFF ” indication. 101

4.6 Set-up Function
The following settings can be made using the Set-up function: · time setting · default setting of dBm measurement reference resistance · default setting of detection method during AC measurement · default setting of LOGGING interval · default setting of number display/time display during LOGGING mode · sound on/off setting (beep of buzzer) · reset to factory preset mode 1) Press the SHIFT key shows “Shift” on the display. 2) Press the LIGHT key while holding down the SHIFT key changes the mode to Set-up mode
(from Set-up to time display).
3) Press the LIGHT key changes the setting items accordingly. 4) Change values by using the (RANGE) key or (REL) key. 5) Press the HOLD key to save/finish each setting.
“SEt” appears and the display returns to the setting items. 6) Hold down the LIGHT key for more than one second to return from Set-up mode to
measurement mode.
Note
To cancel any setting, hold down the LIGHT key for more than one second, or turn off by using the function key.

Set a default value of the reference resistance during dBm measurement mode. 1) Display “db-r” by using the LIGHT key. The reference value appears on the display. HOLD
  1. Select the reference resistance by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting.
    “SEt” appears and then “db-r.”
    Setting values of reference resistance 4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250, 300, 500, 600, 800, 900, 1000, 1200(default value is 600)
(KEW1062 only) Set a default setting of detection methods during AC measurement. RMS or MEAN: The default setting is RMS. 1) Display “Ac” by using the LIGHT key.

RMS MEAN

HOLD

  1. Select the detection method by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting. 4) “SEt” appears and then “Ac.”

    Set a default value of the saving interval during LOGGING mode. 1) Display “LG. int” by using the LIGHT key. The default setting is 1 sec. HOLD
  2. Select the saving interval by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting. “SEt” appears and then “LG. int.”
    Settings of saving interval 1, 2, 5, 10, 30, 60, 600, 1800 sec
    103

<Default setting of number display/time display during LOGGING mode> Set the sub-display during LOGGING mode. (The numbers of saved data or time (minute : second)) Default is the numbers of saved data. 1) Pressing the LIGHT key shows “LG. Unt” on the display.

No. CLOCK

HOLD

  1. Select the desired setting by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting.
    “SEt” appears and then “LG. Unt.”
    <Sound on/off setting> Set the sound on/off (beep of buzzer) Even if the user sets the sound off, it goes off at the following points. · checking continuity · alarm for over-input · alarm for auto power-off

  2. Pressing the LIGHT key shows “bEEP” on the display. Set on/off on the sub- display. Default is ON.

: ON : OFF

HOLD

  1. Select on/off by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting.
    “SEt” appears and then “bEEP.”

    Reset all the settings to factory preset mode except for time. 1) Pressing the LIGHT key shows “dEF.” on the display. HOLD
  2. Press the HOLD key to reset the settings. “donE” appears and then “dEF.”
    104

CAUTION
To avoid damage to instrument When the measurement function is completed, turn the function switch back to the OFF position to turn off.

4.7 Additional functions simply set when POWER ON

With pressing the following keys, turn the function switch to the desired position of any measurement mode (POWER ON-state). This enables the following functions corresponding to the press keys.

Keys MIN/MAX RANGE SELECT HOLD HOLD + REL /% SELECT + RANGE

Functions to be set Averaging Function (calculates the average of 8 times) 5000 (3.5-digit display) LCD check (Lit only while pressing the SELECT key) Cancels the Auto power off function Reset all calibration values to those before shipment. Calibration function

4.8 Averaging Function
The measured value may greatly fluctuate, the instrument can calculate the average (8 times / approx. 2 seconds). This function is available for Voltage, Current and Resistance-mode measurement. The function (averaging) operates until the power is turned off.
Note
Additional functions simply set when POWER ON can be used.

105

4.9 5000 display mode
This function switches 3.5-digit display (5000) and 5-digit display (50000). The function is not available for Capacitor, Temperature, DC+AC, Continuity and Frequencymode measurement. The function (5000 display) operates until the power is turned off.
Note
Additional functions simply set when POWER ON can be used.
4.10 LCD Check
The instrument can lit all segments and mark for LCD check. (Lit only while pressing the SELECT key.)
106

5. User Calibration Function
It is recommended that the instrument be calibrated periodically. The instrument can be calibrated.
CAUTION
To avoid electrical shock · Only authorized engineers are allowed to calibrate the instrument
using dedicated facilities. · Connect the calibrator to the instrument with the calibrator’s test leads. · Before carrying out calibration, read the instruction manual of the calibrator. · Temporarily remove the test leads from the instrument before switching measurement mode
(function).

Calibrator: With accuracy higher than of this instrument Ambient Environment: Temperature: 23±3ºC Humidity: 55%RH or less Leave the instrument for 30 minutes under above conditions before carrying out calibration. After reference valve of Calibrator stabilizes, Press the key to confirm for calibration valve. 107

<Table 1> Carry out calibration of ranges in accordance with Table 1. Calibration for 2 points (Input 1 and Input 2) is required other than DC range. After Input 1, carry out calibration of Input 2 repeating steps 6) and 7). For AC voltage and AC current ranges (marked with ), calibration is carried out at 50Hz or 60Hz frequency. 1) Turn the function switch from the OFF position to the mV position while pressing the
SELECT and RANGE keys at the same time. The display shows the ” CAL ” symbol then the ” PASS ” symbol. 2) Press the SELECT key. (The display shows the ” – ” symbol.) 3) Press the HOLD key twice. (The display shows the ” – – – ” symbol.) 4) Press the RANGE key. (The display shows the ” mV ” symbol.) 5) Connect the instrument to the calibrator with the test leads. 6) Set the calibrator to Input 1 value as an input to the instrument. 7) Press the HOLD key. 8) Be sure to confirm that the function switch and input terminal are set to the desired range. Carry out calibration of other ranges by repeating steps 6) and 7). 9) To quit calibration, turn the function switch back to the OFF position.
Note
The model KEW1062 needs to be calibrated at mean value detection (MEAN) mode and filter on of AC current.
108

Table 1. Input Signal for Calibration

Range

Input 1

Input 2

Unit

DC 50mV

0.000

50.000

mV

DC 500mV

500.00

mV

DC 2400mV

2000.0

mV

DC 5V

5.0000

V

DC 50V

50.000

V

DC 500V

500.00

V

DC 1000V

1000.0

V

AC 50mV*1

5.000

50.000

mV

AC 500mV

50.00

500.00

mV

AC 5V *2 AC 50V

50Hz/60Hz

0.5000 5.000

5.0000 50.000

V V

AC 500V

50.00

500.00

V

AC 1000V

100.0

1000.0

V

500

0.00

500.00

5k 50k 500k 5M 50M Continuity Check ( )

0.0000

5.0000

k

0.000

50.000

k

0.00

500.00

k

0.0000

5.0000

M

0.000

30.000

M

0.0

500.0

DC 500 µA

0.00

500.00

µA

DC 5000 µA

0.0

5000.0

µA

DC 50mA

0.000

50.000

mA

DC 500mA

0.00

400.00

mA

DC 5A

0.0000

5.0000

A

DC 10A

0.000

10.000

A

AC 500 µA

50.00

500.00

µA

AC 5000 µA

500.0

5000.0

µA

AC 50mA

5.000

50.000

mA

AC 500mA

50Hz/60Hz

50.00

400.00

mA

AC 5A

0.5000

5.0000

A

AC 10A

1.000

10.000

A

1: For model KEW1062.
2: For model KEW1062. Calibration points added in 5 V AC range (setting, input value). [RMS], Filter OFF, Input1 (0.50000), Input2 (5.0000) [MEAN], Filter OFF, Input1 (0.50000), Input2 (5.0000) [RMS], Filter ON, Input1 (0.5000) , Input2 (5.0000)

109

<Table 2> After completing the calibration for ranges in Table 1, carry out calibration for ” Frequency Characteristic”. The calibration for is frequency characteristic required for AC voltage and AC current ranges (marked with ). Calibration is carried out at the designated frequency in table 2.

  1. Set the calibrator to Input value as an input to the instrument. 2) Press the MEMORY key. 3) After 20 seconds, buzzer sounds and the instrument confirm calibration.
    (Do not next key operation till buzzer sounds.)

Table 2. Input Signal for Calibration

Range

Input

Unit

AC 50mV*1

50.000

mV

AC 500mV

500.00

mV

AC 5V

5kHz

5.0000

V

AC 50V

50.000

V

AC 500V

500.00

V

AC 1000V 600Hz

1000.0

V

AC 500A AC 5000A

1kHz

500.00 5000.0

A A

*1: For model KEW1062.

110

l Calibration of Capacitor Range

Before start calibration of the Capacitor range,

turn the function switch back to the OFF position.

  1. Turn the function switch from the OFF position to the pressing the SELECT and RANGE keys at the same time.

(Capacitor) position while

The display shows the ” CAL ” symbol then the ” PASS ” symbol.

  1. Press the SELECT key. (The display shows the ” – ” symbol.) 3) Press the HOLD key twice. (The display shows the ” – – – ” symbol.) 4) Press the RANGE key. (The display shows the ” nF ” symbol.)

  2. Connect the instrument to the calibrator with the test leads.

  3. Set the calibrator to Input 1 value as an input to the instrument. 7) Press the HOLD key to confirm. 8) Set the calibrator to Input 2 value as an input to the instrument. 9) Press the HOLD key to confirm. 10) Carry out calibration of other ranges by repeating steps 6) to 9).

  4. To quit calibration, turn the function switch back to the OFF position.

Range

Input 1

Input 2

Unit

5nF

0.500

5.000

nF

50nF

5.00

50.00

nF

500nF

50.0

500.0

nF

5F

0.500

5.000

F

50F

5.00

50.00

F

500F

50.0

500.0

F

5mF

0.500

5.000

mF

50mF

5.00

40.00

mF

111

6. Battery and Fuse Replacement
WARNING
Be careful not to burn yourself. · Fuse becomes a high temperature after current measurement, it is dangerous by touching it
directly. When fuse or batteries are replaced after current measurement, please be sure to leave the main unit for 10 minutes for cooling. · La température du fusible augmente sensiblement après une mesure de courant; il est dangereux de le toucher directement. Après le remplacement du fusible et des piles après une mesure de courant, laissez refroidir l’instrument pendant 10 minutes” ou une annotation équivalente sera reprise dans le manuel d’utilisation.

6.1 Battery Replacement

If the batteries fall below the normal operating voltage, the ”

“symbol turns on.

Follow the steps below to replace the batteries with new ones.

(AA-size (R6) 1.5V batteries)

Set the time by referring to the Set-up function after changing the batteries.

If you remove the batteries, the time resets to 00:00 and the time for LOGGING

measurement will not be correct.

WARNING

Be sure to disconnect the instrument from the circuit under test and test leads before replacing the batteries.

CAUTION

· Turn the function switch to OFF (turn off the power). · Do not mix batteries of different types or new batteries with used ones. · Make sure the polarities of the new batteries are exactly as shown on the battery holder.

To replace the batteries: 1) Remove the screw on the back of the
casing. 2) Remove the back cover. 3) Take the batteries out of the housing. 4) Replace the batteries with new ones. 5) Close the casing and fasten it with the
screw.

Battery housing Screw hole

Blank cover (For USB adapter, Printer adapter)

Back cover removed

112

6.2 Fuse Replacement
If a current greater than the rated value flows when the instrument is in the currentmeasurement range, a protection fuse may blow. If this happens, replace that fuse. The instrument contains the following types of fuses.
WARNING
· Turn the function switch to OFF(turn off the power). · Be sure to disconnect the instrument from the circuit under test and test leads before
replacing the fuses. · Do not operate the instrument with the casing left open. · In order to avoid damage to the instrument or any possible accident, use fuses of the
specified rating. Fuse rating: F1 M-8926 (440mA/1000V, SIBA GmbH & Co. KG, 50 210 06.0.44)
F2 M-8927 (10A/1000V, SIBA GmbH & Co. KG, 50 199 06.10) High breaking capacity type
To replace the fuse: 1) Remove the screw on the back of the casing. 2) Remove the back cover. 3) Remove the blown fuse from the fuse holder. 4) Install a new fuse in the holder.
(Make sure the fuse rating.) 5) Close the casing and fasten it with the screw.

Back cover removed

F1: (440mA/1000V)
F2: (10A/1000V) Screw hole

113

7. Calibration and Maintenance
Calibration Cycle It is recommended that the instrument be calibrated once every year. (SEE ALSO: User Calibration Function)
Contacts of Services Please contact the sales representative from which you purchased the instrument.
8. Disposing the Product
Waste Electrical and Electronic Equipment (WEEE), Directive 2002/96/EC This Product complies with the WEEE Directive (2002/96/EC) marking requirement. The affixed product label (see below) indicates that you must not discard this electrical/ electronic product in domestic household waste.
Product Category With reference to the equipment types in the WEEE directive Annex 1, this product is classified as a “Monitoring and Control instrumentation” product.
114

DISTRIBUTOR

Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.

3-22

92-1986J

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals