TECHNO VIETNAM KEW 1061 Digital Multimeters Instruction Manual
- June 11, 2024
- TECHNO VIETNAM
Table of Contents
TECHNO VIETNAM KEW 1061 Digital Multimeters
Product Information
The KEW 1061/1062 Digital Multimeter is a measuring tool that uses modulation for A/D conversion, allowing for fast and accurate measurements. It has a 5-digit LCD display with a maximum reading of 50000 and a bar graph indicator. The multimeter has switching detection modes for effective value (root mean square value) detection (RMS) and mean value detection (MEAN) during AC voltage or AC current measurement (KEW1062 only). Additionally, it has a low- pass filter that can be switched on/off during AC voltage or AC current measurement (KEW1062 only). The multimeter complies with CE standards and features a current-input terminal shutter for preventing wrong input. It uses high-performance UL-standard fuses.
Product Usage Instructions
Safety Instructions
Always observe the following instructions to avoid the risk of electrical
shock or other dangers that may lead to serious injury or loss of life:
- Refer to the WARNING instruction manual to avoid the risk of serious injury or loss of life.
- Refer to the instruction manual to avoid the risk of injury or damage to the product.
- Handle with care to avoid the risk of injury or death of personnel or damage to the instrument.
- Do not use the product if it is exposed, removed, or wet.
Measurement Category
Ensure that you use the KEW 1061/1062 Digital Multimeter in the correct measurement category based on the following:
- CAT III: For measurements on installations connected to the distribution network.
- CAT II: For measurements on appliances, portable tools, and similar equipment.
User Calibration Function
The KEW 1061/1062 Digital Multimeter has a user calibration function that
allows you to correct measurement errors caused by changes in the environment
or aging of the instrument. Follow the instructions in the instruction manual
to perform user calibration.
**Battery and Fuse Replacement
**
When the battery or fuse needs to be replaced, follow the instructions below:
Battery Replacement
- Remove the battery cover from the back of the instrument.
- Remove the old battery and insert a new one.
- Replace the battery cover.
Fuse Replacement
- Turn off the power to the instrument.
- Remove the fuse cover from the back of the instrument.
- Remove the old fuse and insert a new one.
- Replace the fuse cover.
Calibration and Maintenance
Regular calibration and maintenance of the KEW 1061/1062 Digital Multimeter is essential for accurate and safe operation. Follow the instructions in the instruction manual to perform calibration and maintenance.
Disposing the Product
When disposing of the KEW 1061/1062 Digital Multimeter, follow all local
regulations and guidelines for electronic waste disposal.
INSTRUCTION MANUAL
DIGITAL MULTIMETER
KEW 10611062
Thank you for purchasing our Model KEW1061, KEW1062 Digital Multimeter. This
user’s manual describes the specifications and handling precaution for this
Digital Multimeter. Before using this Digital Multimeter, thoroughly read this
manual to get a clear understanding on proper use. Always observe the
following instructions. Failure to do so may impair the protection provided by
the instrument and probes, and may result in electrical shock or other dangers
that may lead to serious injury or the loss of life. KYORITSU is in no way
liable for any damage resulting from the user’s mishandling of the product.
For safe use of this product, the following safety symbols are used on the
product: n About This Manual · Every effort has been made to ensure accuracy
in the preparation of this manual.
However, should any errors or omissions come to the attention of the user,
please contact KYORITSU. · The contents of this manual are subject to change
without prior notice because of improvement in performance or function. · All
rights reserved. No part of this manual may be reproduced in any from without
KYORITSU’S written permission.
60
Regarding Safe Use of This Product For safe use of this product, the following
safety symbols are used on the product and manual:
This indicates that the operator must refer to an explanation in the WARNING
instruction manual in order to avoid the risk of serious injury or the loss
of life.
CAUTION
This indicates that the operator must refer to an explanation in the
instruction manual in order to avoid the risk of injury or damage to the
product.
Note
This indicates information that is essential for handling the instrument or
should be noted in order to familiarize yourself with the instrument’s
operating procedures and/or functions.
Danger! Handle with Care This symbol indicates that the operator must refer to an explanation in the instruction manual in order to avoid risk of injury or death of personnel or damage to the instrument.
Double Insulation This symbol indicates double insulation or reinforced insulation.
Direct Current This symbol indicates DC voltage/current.
Alternating Current This symbol indicates AC voltage/current.
DC/AC This symbol indicates AC and DC.
Fuse This symbol indicates a fuse.
Battery This symbol indicates a battery.
Ground This symbol indicates ground (earth).
61
WARNING
n Always observe the following instructions. Failure to do so may result in
electrical shock or other dangers that may lead to serious injury or the loss
of life.
Test leads / Test leads with crocodile clip (optional accessory) · Use the
probes supplied by KYORITSU with this instrument. · Do not use test leads/test
leads with crocodile clip that have deteriorated or are defective.
Check test leads/test leads with crocodile clip continuity. · Disconnect test
leads/test leads with crocodile clip from the circuit under test before
opening
the casing to replace the batteries or for any other reason. · Disconnect test
leads/test leads with crocodile clip from the circuit under test before
attaching/detaching the test leads/test leads with crocodile clip to/from the
instrument. · Disconnect test leads/test leads with crocodile clip from the
instrument before opening the
casing to replace the batteries or for any other reason. · A cap is provided
on the tip of a test lead.
Use a test lead with the cap on for safety (safety standards: IEC 61010-031).
· Do not use the crocodile clip of test leads being loosen or removed
conditions. · Stop using the test lead if the outer jacket is damaged and the
inner metal or color jacket is
exposed.
Casing · Do not use the instrument if there is any damage to the casing or
when the casing is
removed.
Fuses · Use fuses of the specified rating when the fuse is replaced.
Operating Environment · Do not operate the instrument in an atmosphere where
any flammable or explosive gas is
present. · Avoid using the instrument if it has been exposed to rain or
moisture or if your hands are
wet.
Disassembly · No person, except personnel from KYORITSU, is authorized to
disassemble this instrument.
Overview
Fast and more accurate measurement The KEW1061 and KEW1062 digital multimeters
use modulation for A/D conversion, which enable fast and more accurate
measurement.
l Display 5-digit (LCD) Maximum Reading: 50000 Bar graph indicator
l Supports a variety of measurement function Measurement function DC Voltage,
AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature,
Capacitor, Duty cycle ratio, Decibel (dBV, dBm), Continuity Check, Diode Test,
LowPower- Other functions Data Hold (D·H), Auto Hold (A·H), Peak Hold (P·H),
Range Hold (R·H), Maximum value (MAX), Minimum value (MIN), Average value
(AVG), Zero Adjustment (Capacitor, Resistance), Relative values, Save to
Memory, LCD backlight. LowPower-: Measures resistance under lowmeasurement
current. *: For model KEW1062 only
l Switching detection modes Effective value (root mean square value) detection
(RMS) and mean value detection (MEAN) can be switched during AC voltage or AC
current measurement (KEW1062 only).
l Low-pass filter The low-pass filter can be switched on/off during AC voltage
or AC current measurement (KEW1062 only).
l Communication: optional communication package is required · Measurement data
can be transferred to a PC using an optional USB communication set. The data
can be read by certain applications to make trend graphs or can be converted
into Excel files. · The data can also be output from an optional printer via
an optional printer communication set.
l Safety design Complied standards: CE standards Uses a current-input terminal
shutter for preventing wrong input. Uses high-performance UL-standard fuses.
65
Measurement Category
WARNING
n Measurement Category (CAT.) The restrictions on the maximum voltage level for which the KEW1061, KEW1062 can be used, depend on the measurement categories specified by the safety standards.
Do not apply any input level higher than maximum allowable input.
AC/DC 1000V
CAT III
AC/DC 600V
CAT
Measurement category
Description
O None, Other Other circuits that are not directly connected to MAINS.
CAT II For measurements performed on circuits directly connected to the low voltage installation.
CAT III For measurements performed in the building installation.
CAT For measurements performed all the source of the low-voltage installation.
Remarks
Appliances, portable equipment, ect. Distribution board, circuit breaker, ect.
Overhead wire, cable systems, ect.
Incoming wire Socket
Interior wiring
:Device which is not directly connected to the mains power supply
Note
Radiation immunity affects the accuracy of the KEW1061, KEW1062 under the
conditions specified in IEC61326-1. Use of this instrument is limited to
domestic, commercial, and light industry applications. If equipment generating
strong electromagnetic interference is located nearly, the instrument may
malfunction.
66
3. Specifications
3.1 General Specifications
Measurement function:
DC Voltage, AC Voltage, DC Current, AC Current, Resistance,
Frequency, Temperature, Capacitor, Duty cycle ratio,
Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower-*
Other functions: Data Hold (D·H), Auto Hold (A·H), Peak Hold* (P·H),
Range Hold (R·H), Maximum value (MAX), Minimum value (MIN),
Average value (AVG), Zero Adjustment (Capacitor, Resistance),
Relative values, Save to Memory, LCD backlight.
LowPower-: Measures resistance under lowmeasurement current.
*: For model KEW1062 only
Measuring method: modulation
Display:
5-digit (LCD)/7-segment
Maximum Reading: 50000
Polarity Indicator: “” Appears automatically when the polarity is
negative
Overrange Indicator: ” OL ”
Low-battery Indicator: ”
” Appears when the batteries become
low
Measurement cycle: 6 times per second
(except frequency measurement: one times per second, Resistance
measurement : four times per second, capacitor measurement (50mF):
max. 0.03 time per second)
Bar graph display 15 times per second
Operating temperature and humidity ranges:
-20 to 55ºC, 80%RH or less (no condensation)
70%RH or less at 40 to 55ºC.
Température de fonctionnement:
-20 à 55°C
Humidité de fonctionnement:
80% HR ou moins (sans condensation), 70% HR ou moins à 40 ~ 55°C
Storage temperature and humidity ranges:
-40 to 70ºC, 70%RH or less (no condensation)
Temperature coefficient:
(Accuracy at 23±5ºC) × 0.05/ºC or less
At -20 to 18ºC and 28 to 55ºC
When continuously DCV and DCA measuring, add 1digit / ºC
(except for 50mV, 5A, 10A range, add 3digits / ºC).
Power supply:
AA-size (R6) 1.5V batteries: 4
67
Battery life:
Approximately 100 hours
(Operating hours of alkaline batteries when in DC voltage-mode.)
Note: The battery life varies depending on the operating conditions.
Insulation resistance : 1000V DC, 100M or more
Withstand voltage: 6.88kVrms AC for five seconds
(across input terminals and casing)
External dimensions: Approximately 192(L) × 90(W) × 49(D) mm
Weight:
Approximately 560g (including batteries)
Complied standards: Safety standards
IEC61010-1, IEC61010-2-033, IEC61010-031
CAT III (Max. input voltage: AC/DC1000V)
CAT IV (Max. input voltage: AC/DC600V)
Pollution degree 2, indoor use,
2000m max. above sea level
Utilisation interne 2000m max. au-dessus du niveau de la mer.
EMC standards
IEC61326-1 Class B
Effect of radiation immunity:
In the radio-frequency electromagnetic field of 3 V/m, accuracy is
within five times the rated accuracy.
Environmental standard:
EU RoHS Directive compliant
68
Standard accessories: Batteries : 4 Test leads: 1set (M-7220A) Fuse (included): 440mA/1000V (M-8926), 10A/1000V (M-8927) Instruction manual: 1
Optional accessories: Carrying case
M-9154
(for the main unit with test leads
and communication cable)
Test leads (1set)
M-7220A
Test leads with crocodile clip (1set) M-7234
Fuse
440mA/1000V
M-8926
10A/1000V
M-8927
Temperature probes
M-8405, 8406, 8407, 8408
USB Communication set
M-8241 (Software, USB adapter and
cable)
Printer Adapter and Cable M-8243
Printer
M-8246
AC adapter (for printer, Europe) M-8248A
Thermal paper for printer (10 rolls) M-8247
69
3.2 Accuracy
Test conditions: Temperature and humidity: 23±5ºC at 80%RH or less Accuracy: ±(% of reading + digits)
Note: Each response time is a value to rated accuracy within selected range.
DC Voltage Measurement V
Range
50mV 500mV 2400mV
5V 50V 500V 1000V
Resolution
0.001mV 0.01mV 0.1mV 0.0001V 0.001V 0.01V 0.1V
Accuracy KEW1061, KEW1062
0.05+10 0.02+2 0.025+5
0.03+2
Input Resistance Approx. 100M
10M
Maximum Input Voltage
1000V DC
1000V rms AC
NMRR: 80dB or more 50/60Hz ±0.1% (70dB or more 50/60Hz ±0.1% when 50mV Range)
CMRR: 100dB or more 50/60Hz (Rs=1k) Response time: 0.3 sec max.
AC Voltage Measurement [RMS] KEW1061
V AC Coupling, Rms-value detection, Crest factor*: <3
Range
500mV 5V 50V
500V 1000V *
Resolution 10 to 20Hz
0.01mV 0.0001V 0.001V
0.01V 0.1V
1.5+30 1 2
Accuracy 20Hz to 1k to
1kHz 10kHz
0.7+30 1
2 3+30 *2
10k to 20kHz
2+50 *2
–
: Crest factor <1.5 at 1000V range Accuracy 1: At 5 to 100% of range, *2: At 10 to 100% of range CMRR: 80dB or more DC to 60Hz (Rs=1k) Response time: 1 sec max.
Input Impedance
Maximum Input Voltage
11M <50pF
10M <50pF
1000V rms AC
1000V DC
70
AC Voltage Measurement [RMS]
V
KEW1062
Range
Resolution
AC Coupling, Rms-value detection, Crest factor*: <3
Accuracy
10 to 20Hz to 1k to 10k to 20Hz 1kHz 10kHz 20kHz
20k to 50kHz
50k to 100kHz
Input Impedance
Maximum Input Voltage
50mV 0.001mV
500mV 0.01mV 5V 0.0001V 50V 0.001V
500V 0.01V
1000V* 0.1V
2+80 2
1+30 1
*2
0.4+40 5+40 5.5+40 2 2 *2
0.4+30 *1
1+40 *1
*2
3+30 *2
15+40 *2
2+70 5+200
*2
*2
–
11M <50pF
10M <50pF
1000V rmsAC
1000V DC
: Crest factor <1.5 at 1000V range Accuracy 1: At 5 to 100% of range, *2: At 10 to 100% of range CMRR: 80dB or more DC to 60Hz (Rs=1k) Response time: 1 sec max.
AC Voltage Measurement [MEAN]
V
KEW1062
Range
Resolution
AC Coupling, MEAN value detection, RMS value calibration (sine wave)
10 to 20 Hz
Accuracy 20 to 500 Hz
Input 500Hz to 1 kHz Impedance
Maximum Input Voltage
50mV 0.001mV
500mV 0.01mV 5V 0.0001V 50V 0.001V
500V 0.01V 1000V 0.1V
4+80 2
2+30 1
*2
1.5+30 2
1+30 1
*2
5+30 2
3+30 1
*2
11M <50pF
1000V rmsAC
10M <50pF
1000V DC
Accuracy 1: At 5 to 100% of range, 2: At 10 to 100% of range CMRR: 80dB or more DC to 60Hz (Rs=1k) Response time: 1 sec max.
71
DCV+ACV + KEW1061
Maximum Reading 50000, Crest factor*: <3
Range
Accuracy Resolution DC, 10 to DC, 20Hz DC, 1k to DC, 10k
20Hz to 1kHz 10kHz to 20kHz
5V
50V 500V 1000V*
0.0001V
0.001V 0.01V 0.1V
1.5+10 1
2
1+10 1
2
2+10 *2
–
Input Impedance 11M <50pF
10M <50pF
Maximum Input Voltage
1000V rms AC
1000V DC
DCV+ACV + KEW1062
Maximum Reading 50000, Crest factor*: <3
Range
5V 50V 500V 1000V*
Resolution
0.0001V 0.001V 0.01V 0.1V
DC, 10 to 20Hz
DC, 20Hz to
1kHz
Accuracy DC, DC, 1k to 10k to 10kHz 20kHz
DC, 20k to 50kHz
1.5+10 *1
0.5+10 *1
1+10 2+10
*1
*2
*2
*2
–
DC, 50k to 100kHz
Input Impedance
Maximum Input Voltage
11M 1000V
5+20 <50pF rms AC
*2
10M <50pF
1000V DC
: Crest factor <1.5 at 1000V range Accuracy 1: At 5 to 100% of range, *2: At 10 to 100% of range
CMRR: 80dB or more DC to 60Hz (Rs=1k) Response time: Approx. 2 sec
DC Current Measurement A
Range
500A 5000A
50mA 500mA*3
5A 10A
Accuracy Resolution
KEW1061, KEW1062
0.01A
0.1A 0.001mA
0.2+5
0.01mA
0.0001A
0.6+10
0.001A
0.6+5
Voltage Drop <0.11mV/A
<4mV/mA <0.1V/A
Maximum Input Current
440mA Protected by a 440mA/1000V fuse.
10A Protected by a 10A/1000V fuse.
*3: Maximum measurement current : 440mA at 500mA range Response time: 0.3 sec max.
Note : After measuring over 500mA DC (especially 10A DC), “Zero error” occurs
for a while. In that case, please wait for a while at zero input until the
value stabilizes before measuring again.
72
AC Current Measurement [RMS]
A
KEW1061
Rms-value detection, Crest factor: <3
Range Resolution
500A 5000A
50mA 500mA*3
5A 10A
0.01A 0.1A 0.001mA 0.01mA 0.0001A 0.001A
Accuracy Voltage Drop
10 to 20Hz 20Hz to 1kHz
Maximum Input Current
1.5+20
1+20
<0.11 mV/A 440mA Protected
by a 440mA/
<4mV/mA
1000V fuse.
<0.1V/A
10A Protected by a 10A/1000V fuse.
AC Current Measurement [RMS]
A
KEW1062
Rms-value detection, Crest factor: <3
Range
Resolution
Accuracy
10 to 20Hz 20Hz to 1kHz 1k to 5kHz
Voltage Drop
Maximum Input Current
500A 5000A
50mA 500mA *3
0.01A 0.1A
0.001mA 0.01mA
1+20
0.75 +20
1+30
<0.11 mV/A 440mA Protected
by a 440mA/ <4 mV/mA 1000V fuse.
5A
0.0001A 1.5+20
10A
0.001A
1+20
2+30
<0.1 V/A
10A Protected by a 10A/1000V fuse.
Model KEW1061/1062 Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range *3: Maximum measurement current : 440mA at 500mA range Response time: 1 sec max.
73
AC Current Measurement [MEAN]
A
KEW1062
MEAN value detection, RMS value calibration (sine wave)
Range
500A 5000A
50mA 500mA*3
5A
10A
Accuracy Resolution 10 to 20Hz 20 to 500Hz
0.01A 0.1A 0.001mA 0.01mA 0.0001A
0.001A
2+20 3+20
1.5 +20 2+20
500Hz to 1kHz 2+30
4+30
Voltage Maximum Input
Drop
Current
<0.11 mV/A 440mA Protected
by a 440mA/ <4 mV/mA 1000V fuse.
10A Protected <0.1 V/A by a 10A/
1000V fuse.
Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range *3: Maximum measurement current : 440mA at 500mA range Response time: 1 sec max.
DCA+ACA KEW1061
Range
500A 5000A
50mA 500mA*3
5A 10A
- Resolution
0.01A 0.1A 0.001mA 0.01mA 0.0001A 0.001A
Maximum Reading 50000, Crest factor: <3
Accuracy
DC, 10 to DC, 20Hz to Voltage Drop
20Hz
1kHz
Maximum Input Current
2+10
1.5+10
<0.11 mV/A 440mA Protected by a
<4mV/mA 440mA/1000V fuse.
<0.1V/A
10A Protected by a 10A/1000V fuse.
Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range *3: Maximum measurement current : 440mA at 500mA range Response time: 2 sec max.
74
DCA+ACA + KEW1062
Maximum Reading 50000, Crest factor: <3
Range
Accuracy Resolution DC, 10 to DC, 20Hz DC, 1k to Voltage Drop
20Hz to 1kHz 5kHz
Maximum Input Current
500A 5000A
50mA 500mA*3
0.01A 0.1A 0.001mA 0.01mA
1.5+10
1+10
1.5+10
<0.11 mV/A 440mA Protected
by a 440mA/
<4 mV/mA
1000V fuse.
5A 10A
0.0001A 0.001A
2+10
1.5+10
3+10
<0.1 V/A
10A Protected by a 10A/1000V fuse.
Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range *3: Maximum measurement current : 440mA at 500mA range Response time: Approx. 2 sec
Resistance Measurement
Range
500 5k 50k
500k 5M
50M
Accuracy
Maximum
Resolution
Measuring
KEW1061 KEW1062 Current
0.01
<1mA
0.0001k 0.001k
0.1+2 *1
0.05+2 <0.25mA
*1
<25A
0.01k
<2.5A
0.0001M
0.5+2
<1.5A
0.001M
1+2
<0.13A
*1: Accuracy is specified after zero adjustment (resistance). Response time: 1 sec max. at 500 to 500k
5 sec max. at 5M to 50M
Open-loop Voltage
<2.5V
Input Protective Voltage
1000V rms
LowPower- LP-
Range Resolution
5k 50k 500k 5M
0.001k 0.01k 0.1k 0.001M
Accuracy KEW1062 only
0.2+3 1+3
Maximum Measuring
Current
<10A <1.0A <0.6A <0.05A
Maximum Reading 5000
Open-loop Input Protective
Voltage
Voltage
<0.7V
1000V rms
LowPower-: Measures resistance under lowmeasurement current.
75
Continuity Check
Range 500
Resolution 0.1
Range of Operation
The buzzer turns on for resistances lower
than 100±50.
Measuring Current
Approx. 0.5mA
Maximum Reading 5000
Open-loop Input Protective
Voltage
Voltage
<5V
1000V rms
Diode Test Range 2.4V
Resolution 0.0001V
Accuracy
Measuring Current (Vf=0.6V)
Open-loop Voltage
Input Protective Voltage
1+2
Approx. 0.5mA
<5V
1000V rms
Temperature Measurement TEMP
Range
Resolution
Accuracy
-200 to 1372°C -328 to 2501.6°F
0.1°C 0.1°F
1%+1.5°C 1%+2.7°F
Use optional Temperature Probe: Thermocouple Type K
Capacitor Measurement
Range 5nF 50nF
500nF 5F 50F
500F 5mF 50mF
Resolution 0.001nF 0.01nF 0.1nF 0.001F 0.01F 0.1F 0.001mF 0.01mF
Accuracy
1+5 *1
2+5 3+5
*1: Accuracy is specified after zero adjustment (capacitor).
Input Protective Voltage
1000V rms
Maximum Reading 5000 Input Protective Voltage
1000V rms
76
Frequency Measurement Hz
Range (AUTO) 2.000 to 9.999Hz
9.00 to 99.99Hz 90.0 to 999.9Hz 0.900 to 9.999kHz 9.00 to 99.99kHz
AC Coupling, Maximum Reading 9999
Resolution 0.001Hz 0.01Hz 0.1Hz 0.001kHz 0.01kHz
Accuracy
0.02+1 1
2
Accuracy 1: At 10 to 100% of input voltage or current range 2: At 40 to 100% of input voltage or current range
Duty cycle ratio % Range
10 to 90%
Resolution 1%
Accuracy ±1%*1
Accuracy *1: At 10.00Hz to 500.0Hz, square wave At 40 to 100% of input voltage or current range
Peak Hold P·H Model KEW1062 only
Range DCV, DCA
Accuracy ±100 digit
Maximum Reading 5000
Response Time Maximum >250S
77
4. Operation
4.1 Precautions Before Measurement
n Examining Items Contained in the Package After opening the package, be sure
to examine the product as instructed below before use. Should the delivered
product be the wrong model, lack any item, or show any flaw in its appearance,
contact the vendor from which you purchased the product.
n Precautions of Operation and Storage
CAUTION
· Insert the batteries in the instrument by referring to “6.1 Battery
Replacement.” · A Blank cover is provided on the upper part of back casing.
Don’t remove the Blank cover except when the USB adapter or Printer adapter is
connected. · Do not use the instrument near noise-emitting equipment or where
there may be a sudden change of temperature. Otherwise, the instrument may
give an unstable reading or errors. Removal of Dirt Do not wipe the instrument
using any solvent (chemicals) such as benzine or paint thinner, as this may
damage or discolor the front panel. Use a dry cloth to clean the instrument.
Storage Conditions · Do not leave the instrument exposed to direct sunlight or
in a hot and humid location such as the inside of a vehicle, for any prolonged
length of time. · If the instrument will not be used for a prolonged period,
remove the batteries.
78
Caps of Testing leads Protective fingerguard
4.2 Components
n Panel Description
n Test leads
Display (LCD)
SHIFT key RANGE key HOLD (SAVE) key LIGHT (READ) key SELECT (PRINT) key REL /%
key MIN/MAX (LOG) key MEMORY key
Function switch
Input terminal block
Red
Black
With: 1000V10A CAT. /600V 10A CAT. Without: 1000V10A CAT. /600V 10A CAT.
n Test leads with crocodile clip(optional accessory)
Protective fingerguard
Red
Protective fingerguard
Black
Cable
Crocodile clip
79
Cap
Protective fingerguard
Protective fingerguard (Barrier) It is a part providing protection against
electrical shock and ensuring the minimum required air
and creepage distances.
Cap Uncapped condition for CAT II environment Capped condition for CAT III/ IV
environments The Cap shuld be firmly attached to the probes.
- Function switch Turns off the power or select the measurement mode (function).
OFF
Turns off the power.
Resistance measurement
V AC voltage (V) measurement
Capacitor measurement
mV AC voltage (mV) measurement V DC voltage (V) measurement mV DC voltage (mV)
measurement
Continuity check, Diode Test
TEMP
A mA A
Temperature measurement DC/AC current measurement
- SELECT key Pressing this key in each measurement modes (function) described above selects other measurement modes (function).
V,
V µ A/mA/A
mV 1
HzV
: Frequency measurement (Voltage value is displayed.)
2
Hz%
: Frequency measurement (Duty cycle ratio)
3
dBV
: dBV measurement (Voltage value is displayed.)
4
dBm
: dBm measurement
(DC+AC) measurement, (DC, AC) Dual display
LP- (Model KEW1062 only)
Diode Test
Pressing this key in each measurement modes (function) described above selects other measurement modes (function).
1
: AC voltage measurement
2
: (DC+AC) measurement
3
·
: (DC, AC) Dual display
4
Hz%
: Frequency measurement (Duty cycle ratio)
80
-
RANGE key Allows the operator to select the measuring range. Fixed ranges : The display shows the ” R·H ” symbol. The range increases every time this key is pressed. AUTO range : The display shows the ” AUTO ” symbol. To return to the auto-ranging mode, hold down RANGE key for more than one second.
-
HOLD key Selects between the DATA HOLD, AUTO HOLD and PEAK HOLD functions. To cancel functions, press this key once again. DATA HOLD: Holds the display readings. The display shows the ” D·H ” symbol. AUTO HOLD: Holds the measured value when the Test leads are handled. The display shows the ” A·H ” symbol. PEAK HOLD: Holds the peak value. The display shows the ” P·H ” symbol. (Model KEW1062 only)
-
LIGHT key LIGHT key: Use to turn on the LCD backlight. Press this key once to turn on the LCD backlight for approximately one minute. The LCD backlight is lit for approximately one minute. (To postpone turned on time, press this key once again.) To cancel the function, hold down this key for more than one second.
-
REL / % key The instrument can calculate relative values or differences, and percentage values from the reference measurement values. 1 : Relative Calculation The display shows the ” ” symbol. The sub-display shows the reference voltage value.
2 : Percentage Calculation The display shows the ” “, ” % ” symbol. The sub- display shows the reference voltage value. -
MIN/MAX key Displays the minimum value (MIN), maximum value (MAX) and average value (AVG) during measurement.
Pressing this key starts recording and at the same time the display shows MIN/MAX/AVG to release AUTO POWER OFF.
81 -
MEMORY key Data can be stored in internal memory using this key. Used when outputting to printer with the optional adapter and cable.
-
SHIFT key While this key is pressed, “Shift” appears on the display. Pressing the following keys with the SHIFT key held down enables the following settings.
SHIFT+
LIGHT key RANGE key REL key HOLD key
Set-up function Change to [RMS] mode Change to [MEAN] mode Turn filter on/off
(KEW1062 only) (KEW1062 only) (KEW1062 only)
n Display (LCD) Description
82
Symbol and Unit
+
R·H AUTO
AUTO OFF RMS LP- Filter Shift nF, F, mF mV, V A, mA, A M, k, °C / °F kHz, Hz
dB, dBm % (Main-display) % (Sub- display) mV, V (Sub- display) s (Sub-
display) (Sub- display)
OL
Description
Appears when in DC-mode measurement
Appears when in AC-mode measurement Appears when in DC+AC-mode measurement
Appears when the polarity is negative
Appears when in diode test
Appears when in continuity check
Relative calculation indicator
Fixed ranges indicator AUTO range indicator DATA HOLD indicator AUTO HOLD
indicator PEAK HOLD indicator
Lit when in MIN/MAX/AVG-mode
Lit when in MIN/MAX/AVG-mode
Lit when in MIN/MAX/AVG-mode
Lit when in Memory function Auto power off indicator Appears in RMS mode
Appears in Low Power- measurement Appears while filter is on Appears while the
SHIFT key is held down Unit for capacitance measurement Unit for voltage
measurement Unit for current measurement Unit for resistance measurement Unit
for temperature measurement Unit for frequency measurement Decibel calculation
indicator Unit for percentage calculation Unit for duty cycle ratio
calculation Unit for voltage measurement (dBV, HzV) Unit for recording time
when in MIN/MAX/AVG-mode Recording time indicator when in MIN/MAX/AVG-mode
Number of saved data indicator Reference value indicator when relative
calculation Duty cycle ratio indicator Voltage value (dBV, HzV) indicator
Reference resistance value (dBm) indicator
Overrange Indicator
Appears when the batteries become low
Bar graph indicator, Range indicator
83
4.3 Measuring Instructions
WARNING
To avoid damage to instrument or equipment · Before starting measurement, make
sure that the position of function switch and the
input terminals for connecting the test leads are appropriate for the desired
mode of measurement. · Temporarily remove the test leads from the device under
test before operating the function switch. · Verify proper operation on a
known source before use or taking action as a result of the indication of the
instrument. · Vérifiez le fonctionnemnet adéquat sur une source connue avant
de procéder tout en se basant sur un affichage trompeur de l’instrument.” ou
une annotation équivalente sera reprise dans le manuel d’utilisation. · Stop
using the test lead if the outer jacket is damaged and the inner metal or
color jacket is exposed. · Arrêtez d’utiliser le cordon de test si la gaine
extérieure est endommagée et si la gaine métallique ou interne est exposée.
Test leads here include a test leads with crocodile clip (optional accessory). 4.3.1 AC Voltage Measurement ( V, mV)
- Turn the function switch to the ” V ” or ” mV ” position. 2) Plug the test
leads into the input terminals. 3) Connect the test leads to the circuit under
test and then read the value
when it stabilizes.
Test leads
Black
Red
4.3.2 DC Voltage Measurement ( V, mV) 1) Turn the function switch to the ” V
“or ” mV” position. 2) Plug the test leads into the input terminals. 3)
Connect the test leads to the circuit under test and then read the value
when it stabilizes.
Test leads
Black
Red
Note
If ” mV “range is selected and the test leads are left open-circuited, the
instrument may give a certain reading. This dose not affect your measurement.
84
4.3.3 DC+AC Voltage Measurement ( + )
- Turn the function switch to the ” V ” position. 2) Press the SELECT key to
select DC+AC voltage measurement.
(The display shows the ” + ” symbol.) 3) Plug the test leads into the input terminals. 4) Connect the test leads to the circuit under test and then read the value
when it stabilizes.
Test leads
Black
Red
4.3.4 DC, AC Voltage Dual Display ( · ) 1) Turn the function switch to the
“DCV” position. 2) Press the SELECT key twice to change the mode to DC/AC
voltage dual
display. DC voltage measurement appears on the main display and AC voltage on
the sub-display. 3) Plug the test leads into the input terminals. 4) Connect
the test leads to the circuit under test and then read the value when it
stabilizes.
Test leads
Black
Red
4.3.5 Resistance Measurement ()
CAUTION
To avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to the instrument.
- Turn the function switch to the ” ” position. 2) Plug the test leads into
the input terminals. 3) Connect the test leads to the circuit under test and
then read the value
when it stabilizes.
Test leads
Black
Red
Note
Zero adjustment Zero adjustment is recommended for correct measurement. After
executing 1), 2) above, short the two test leads. Press the REL key for
adjust. (The display shows the “0.0” value.) The value (zero adjustment)
stores till turn off.
85
4.3.6 Low Power- (LP-)
This function is for measuring the resistance of parts on a printed board under low measurement
current.
In Low Power- measurement, up to 5,000 counts are displayed and the measurement range
is from 5 k to 5 M.
-
Turn the function switch to the “” position. Press the SELECT key to set LP-mode. “LP-” appears on the display.
-
Plug the test leads into the input terminals.
-
Connect the test leads to the resistance and then read the value when it
Test leads
stabilizes.
Black
Red
4.3.7 Continuity Check ( )
CAUTION
To avoid damage to instrument Turn off the power to the circuit under test before starting measurement in order to prevent any excessive voltage from being applied to the instrument.
- Turn the function switch to the ” ” position. 2) Plug the test leads into
the input terminals. 3) Connect the test leads to the circuit under test. If
the circuit is continuous
(no more than approximately 100), the buzzer sounds.
Test leads
Black
Red
86
4.3.8 Diode Test ( )
CAUTION
To avoid damage to instrument Turn off the power to the circuit under test
before starting measurement in order to prevent any excessive voltage from
being applied to the instrument.
- Turn the function switch to the ”
· ” position.
Press the SELECT key to select Diode test.
(The display shows the
symbol.)
-
Plug the test leads into the input terminals.
-
Connect the test leads to the diode and then read the value when it
stabilizes.
Test leads
Black
Red
Black test lead
Red test lead
Red test lead
Black test lead
Figure 1 Forward-bias Diode Test
Figure 2 Reverse-bias Diode Test
87
4.3.9 Temperature Measurement (TEMP)
CAUTION
To avoid damage to instrument Turn off the power to the circuit under test
before starting measurement in order to prevent any excessive voltage from
being applied to the instrument.
Note
Optional Temperature probe is required for temperature measurement.
Temperature Probe: Thermocouple Type K
Model: 8405, 8406, 8407, 8408 Check the measurable range of respective probes.
- Turn the function switch to the ” TEMP ” position. 2) Plug the measuring
probe into the input terminals. 3) Contact the measuring probe to the under
test and then read the value
when it stabilizes.
Measuring
probe
Black
Red
Note
The default temperature read-out of the Digital Multimeters is in Celsius
(°C). To change it to Fahrenheit (°F), it is necessary to proceed as follows:
Changing the temperature unit setting to Fahrenheit Displaying ” ºC ” only is
configured at factory before shipment. Carry out the following setting
procedure to display ” ºF “. While pressing the SELECT, RANGE and HOLD keys
simultaneously, turn the function switch to the ” TEMP ” position. Then, upon
pressing the SELECT key, the temperature unit switches from ºC to ºF. Once the
temperature is displayed in ºF, press the SELECT key to alternately switch the
temperature units between ºF and ºC.
The conversion from Celsius to Fahrenheit is performed using the equation
Fahrenheit temperature = 1.8×Celsius temperature + 32
88
4.3.10 Current Measurement (µA/mA/A)
WARNING
To avoid damage to instrument or equipment · Before starting measurement, make
sure that the position of function switch and the
input terminals for connecting the test leads are appropriate for the desired
mode of measurement. · The maximum input current (limited by fuses) of the
“µA” and “mA” ranges is 440 mA. Be sure not to exceed the limit in the 500 mA
range.
Be careful not to burn yourself · When measuring more than 6A under exceeding
40 conditions, the continuous
measuring time shall be within 3 minutes, then keep disconnected for more than
10 minutes. · En mesurant plus de 6A dans des conditions en dessous de 40°C,
le temps de mesure ininterrompue se situera dans les 3 minutes; déconnectez
l’instrument ensuite pendant plus de 10 minutes.” ou une annotation
équivalente sera reprise dans le manuel d’utilisation.
- Turn the function switch to the “µA ” , “mA ” or “A “position. (If the magnitude of the current being measured is not known, select the “A “position. Make sure the current being measured is no more than 440mA before the “µA ” or “mA ” position is selected.)
- Please select between DC and AC. When selecting AC, press the SELECT key.
- Plug the black test lead into the ” COM ” input terminal and the red test lead into the “A “input terminal. If the current is in the order of mA or less, plug the red test lead into the “µA · mA” input terminal.
- Connect the test leads to the circuit under test and then read the value when it stabilizes.
Test leads Red
Black
Test leads
Red
Black
89
4.3.11 DC+AC Current Measurement ( + )
WARNING
To avoid damage to instrument or equipment · Before starting measurement, make
sure that the position of function switch and the
input terminals for connecting the test leads are appropriate for the desired
mode of measurement. · The maximum input current (limited by fuses) of the
“µA” and “mA” ranges is 440 mA. Be sure not to exceed the limit in the 500 mA
range.
Be careful not to burn yourself · When measuring more than 6A under exceeding
40 conditions, the continuous
measuring time shall be within 3 minutes, then keep disconnected for more than
10 minutes. · En mesurant plus de 6A dans des conditions en dessous de 40°C,
le temps de mesure ininterrompue se situera dans les 3 minutes; déconnectez
l’instrument ensuite pendant plus de 10 minutes.” ou une annotation
équivalente sera reprise dans le manuel d’utilisation.
- Turn the function switch to the “µA” , “mA” or “A” position. (If the magnitude of the current being measured is not known, select the “A” position. Make sure the current being measured is no more than 440mA before the “µA” or “mA” position is selected.)
- Press the SELECT key twice to select the DC +AC measurement. (The display shows the + symbol.) Plug the black test lead into the ” COM ” input terminal and the red test lead into the “A” input terminal. If the current is in the order of mA or less, plug the red test lead into the “µA · mA” input terminal.
- Connect the test leads to the circuit under test and then read the value when it stabilizes.
Test leads Red
Black
Test leads
Red
Black
90
4.3.12 DC, AC Current Dual Display ( · ) 1) Turn the function switch to the
“µA,” “mA” or “A” position.
(If the magnitude of the current being measured is not known, select the “A”
position. Make sure the current being measured is no more than 440 mA before
the “µA” or “mA” position is selected.) 2) Press the SELECT key three times to
select DC/AC dual display. DC current measurement appears on the main display
and AC current on the sub-display. 3) Plug the black test lead into the “COM”
input terminal and the red test lead into the “A” input terminal. If the
current is in the order of mA or less, plug the red test lead into the “A/mA”
input terminal. 4) Connect the test leads to the circuit under test and then
read the value when it stabilizes.
4.3.13 Capacitor Measurement ( )
CAUTION
Test leads Red
Black
Test leads
Red
Black
To avoid damage to instrument · Turn off the power to the circuit under test
before starting measurement in order to prevent
any excessive voltage from being applied to the instrument. · Before starting
measurement, be sure to discharge the capacitor under check.
- Turn the function switch to the ” “position. 2) Plug the test leads into
the input terminals. 3) Open the test lead and press the REL key in 5nF range
to adjust the
capacitance to zero. (The display shows ” 0.000 “.) - Connect the test leads to the circuit under test and then read the value when it stabilizes.
Test leads
Black
Red
Note
The value (zero adjustment) remains displayed until power-off.
91
4.3.14 Frequency Measurement (Hz), Duty cycle ratio (Hz)
CAUTION
To avoid damage to instrument Turn off the power to the circuit under test
before starting measurement in order to prevent any excessive voltage from
being applied to the instrument.
- Turn the function switch to the voltage( V, mV) or the current (µ A, mA, A ) position.
- Press the SELECT key to select the range of frequency. (The display shows the unit of frequency.)
- Plug the test leads into the input terminals. Plug the red test lead into the suitable input terminal (” A” or “µA·mA”) for current value when current measurement.
- Connect the test leads to the under test and then read the value when it stabilizes. The display shows frequency value in main-display and duty cycle ratio value in sub-display.
Test leads
Black
Red
Test leads
Red
Black
Test leads
Red
Black
92
4.3.15 Function to change RMS detection to/from MEAN detection mode (KEW1062 only) The instrument has a function to change RMS detection to/from MEAN detection modes.
4.3.16 Function to turn the filter on/off (KEW1062 only) The instrument has a function to turn the filter on/off during AC measurement.
- Select the appropriate AC measurement mode (ACV, ACmV, ACµA, ACmA, ACA) by
using
the function switch and the SELECT key. 2) Press the SHIFT key to display “Shift” on the display. 3) Then press the HOLD key to turn the low-pass filter on.
While the filter on, “Filter” appears on the display. Refer to filter characteristics in the diagram below.
Amplitude ratio [dB]
0 -10 -20 -30 -40 -50 -60 -70
10
Third-order low-pass filter
100 Frequency [Hz] 1,000
10,000
- Repeat step 2) and 3) to turn the filter off. (“Filter” disappears from the display.)
93
4.3.17 AUTO HOLD Function The instrument can automatically retain the measured
value when the test leads are handled as described below. 1) Press the HOLD
key to select Auto hold function.
(The display shows the” A·H ” symbol.) 2) Connect the test leads to the
circuit under test. 3) When the reading stabilized, the buzzer sounds. 4)
Remove the test leads from the circuit under test. 5) The display shows the
measured value that is retained.
You can repeat steps 2) to 4)as many times as you like as long as the display
shows the ” A· H ” symbol.
Note
· In DC/AC voltage measurement, the Auto hold function is only available for
ranges greater than the 5V range.
· This function is not available for Temperature, Capacitor and Frequency
measurement. · The Auto hold function can not be applied to unstable signals.
94
4.3.18 PEAK HOLD Function This instrument can always detect, update and display the peak value (instantaneous) in DCV and DCA measurement. The peak value of the wave can be seen.
- Turn the function switch to DCV or DCA position. 2) Connect the test leads
to the circuit under test. 3) Press the HOLD key to select Peak Hold. (The
display shows the ” P·H ” symbol.) 4) The display shows the peak value. 5)
When resetting the peak value in HOLD, press the MIN/MAX key.
Then new peak value can be in HOLD.
Note
Even though the input signals (DCV, DCA) have negative polarity, the peak value can be measured when the peak is in the positive direction. Relative values from the reference values can be shown during the peak value measuring. - Press the REL /% key to relative calculation in PEAK HOLD mode. The display shows the ” ” symbol and the relative peak value.
- Press the REL /% key once again to percentage calculation. The display
shows the ” % ” symbol and the percentage peak value.
SEE ALSO Next section ” Relative and percentage calculation”
When resetting the peak value, press the MIN/MAX key. Then new peak value can be in HOLD.
To cancel the percentage calculation, Press the REL /% key again. ” % ” symbol disappears, then retrieves the PEAK HOLD mode.
4.3.19 Relative and percentage calculation The instrument can calculate relative values or difference, and percentage values from the reference measurement values. (The range will be fixed.)
<Relative (REL) calculation> Subtracts the reference value from the measured value to display the relative value or difference. - Take a measurement to set the reference value. 2) Press the REL /% key.
(The display shows the” ” symbol and the sub-display shows the reference value.) 3) Take another measurement.
95
<Percentage (%) calculation> Calculates and display the percentage value according to the following equation: % value =(measured value reference value)/reference value
- Take a measurement to set the reference value. 2) Press the REL /% key.
(The display shows the” ” symbol and the sub-display shows the reference value.) 3) Take another measurement.
Press the REL /% key again. (The display shows the” % ” symbol.)
4.3.20 Decibel calculation (dBm, dBV) The instrument can perform logarithmic calculations on a AC voltage.
dBm : 20log dBV : 20log
Measured voltage value Reference resistance value × 10-3 (1mW/Reference
resistance ()=0dBm)
Measured voltage value 1(V)
- Turn the function switch to the V or mV position. 2) Press the SELECT key
to select dBm dBV .
(The display shows the ” dBm “, ” dB ” symbol.) 3) Connect the test leads to the circuit under test and then read the value when it stabilizes. 4) When calculating relative value, press the REL /% key.
Note
The instrument can switch (select) reference resistance value when measuring dBm. The reference resistance value is switched as follows every time the RANGE key is pressed. (Shown in sub-display.)
Reference resistance value: 4,8,16,32,50,75,93,110,125,135,150 200,250,300,500,600,800,900,1000,1200 Default value: 600
The default settings can be changed. Refer to the Set-up function.
96
4.3.21 MIN/MAX/AVG Function The minimum value (MIN), maximum value (MAX) and average value (AVG) during measurement are shown. (The range is fixed.) The average value is shown by dividing the integrated record data by the number of recording times. Pressing this key starts recording and at the same time the display shows “MIN”, “MAX” and “AVG” to release AUTO POWER OFF.
4.4 Memory Function
Memory capacity SAVE-mode: 100 data LOGGING-mode: Logging data of one time
Model KEW1061 1,000 data Model KEW1062 10,000 data
Number of saved data
Number of saved data is 4-digit numbers. When LOGGING-mode, ” L ” is attached
to the top of 4-digit numbers. The instrument allocates the smallest number,
between 0000 to 9999, that has not yet been used. Use the (RANGE) key or (REL
/%) key switches the number of saved data.
To save a Data (SAVE-mode)
- Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the
SAVE (HOLD) key.
(The display shows the number of saved data.) 3) Press the SAVE (HOLD) key to save the data.
Another press of the SAVE (HOLD) key saves a data for the second time measurement or later. 4) To cancel the function, hold down the MEMORY key for one second. (” MEM ” symbol disappears.)
Note
HOLD data can be saved. Hold the display and save it according to the above
steps.
The number of saved data
98
To save a Data (LOGGING-mode) In logging measurement mode, the time needs to be set. Note that changing the batteries resets the time to 00:00. Set the time by referring to the Set-up function.
- Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the
LOG (MIN/MAX) key.
(The display shows the logging interval (period).) Set the value with the (RANGE) key or (REL /%) key. The default setting is one second. (The default settings can be changed. Refer to the Set-up function.) The display shows ” FULL ” when the logging data is already saved. When saving the new data, perform delete of data. 3) Press the LOG (MIN/MAX) key to start logging. (The ” MEM “symbol is flashing.) Every time the MIN/MAX key is pressed, the sub- display changes. (Sub display : saved number saved time (min : sec)(hour : min)saved number) 4) To cancel the function, hold down the MEMORY key for one second. When memory capacity becomes full, the function is automatically canceled. (” MEM ” symbol disappears.)
Note
LOGGING-mode operation during HOLD-mode disables HOLD-mode.
To load a Data (SAVE-mode) 1) Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the READ (LIGHT) key. 3) Press the SAVE (HOLD) key to select the number of saved data.
Select the number with the (RANGE) key or (REL /%) key. 4) To cancel the function, hold down the MEMORY key for one second.
(” MEM ” symbol disappears.)
To load a Data (LOGGING-mode) 1) Press the MEMORY key. (The display shows the ” MEM “symbol.) 2) Press the READ (LIGHT) key. 3) Press the LOG (MIN/MAX) key to select the number of saved data.
Select the number with the (RANGE) key or (REL /%) key. Every time the MIN/MAX key is pressed, the sub-display changes. (Sub display : saved number saved time (min : sec)(hour : min)saved number) 4) To cancel the function, hold down the MEMORY key for one second. (” MEM ” symbol disappears.)
99
4.5 AUTO POWER OFF Function
4.6 Set-up Function
The following settings can be made using the Set-up function: · time setting ·
default setting of dBm measurement reference resistance · default setting of
detection method during AC measurement · default setting of LOGGING interval ·
default setting of number display/time display during LOGGING mode · sound
on/off setting (beep of buzzer) · reset to factory preset mode 1) Press the
SHIFT key shows “Shift” on the display. 2) Press the LIGHT key while holding
down the SHIFT key changes the mode to Set-up mode
(from Set-up to time display).
3) Press the LIGHT key changes the setting items accordingly. 4) Change values
by using the (RANGE) key or (REL) key. 5) Press the HOLD key to save/finish
each setting.
“SEt” appears and the display returns to the setting items. 6) Hold down the
LIGHT key for more than one second to return from Set-up mode to
measurement mode.
Note
To cancel any setting, hold down the LIGHT key for more than one second, or
turn off by using the function key.
- Select the reference resistance by using the (RANGE) key or (REL) key. 3)
Press the HOLD key to save the setting.
“SEt” appears and then “db-r.”
Setting values of reference resistance 4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250, 300, 500, 600, 800, 900, 1000, 1200(default value is 600)
RMS MEAN
HOLD
-
Select the detection method by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting. 4) “SEt” appears and then “Ac.”
Set a default value of the saving interval during LOGGING mode. 1) Display “LG. int” by using the LIGHT key. The default setting is 1 sec. HOLD -
Select the saving interval by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting. “SEt” appears and then “LG. int.”
Settings of saving interval 1, 2, 5, 10, 30, 60, 600, 1800 sec
103
<Default setting of number display/time display during LOGGING mode> Set the sub-display during LOGGING mode. (The numbers of saved data or time (minute : second)) Default is the numbers of saved data. 1) Pressing the LIGHT key shows “LG. Unt” on the display.
No. CLOCK
HOLD
-
Select the desired setting by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting.
“SEt” appears and then “LG. Unt.”
<Sound on/off setting> Set the sound on/off (beep of buzzer) Even if the user sets the sound off, it goes off at the following points. · checking continuity · alarm for over-input · alarm for auto power-off -
Pressing the LIGHT key shows “bEEP” on the display. Set on/off on the sub- display. Default is ON.
: ON : OFF
HOLD
-
Select on/off by using the (RANGE) key or (REL) key. 3) Press the HOLD key to save the setting.
“SEt” appears and then “bEEP.”Reset all the settings to factory preset mode except for time. 1) Pressing the LIGHT key shows “dEF.” on the display. HOLD -
Press the HOLD key to reset the settings. “donE” appears and then “dEF.”
104
CAUTION
To avoid damage to instrument When the measurement function is completed, turn
the function switch back to the OFF position to turn off.
4.7 Additional functions simply set when POWER ON
With pressing the following keys, turn the function switch to the desired position of any measurement mode (POWER ON-state). This enables the following functions corresponding to the press keys.
Keys MIN/MAX RANGE SELECT HOLD HOLD + REL /% SELECT + RANGE
Functions to be set Averaging Function (calculates the average of 8 times) 5000 (3.5-digit display) LCD check (Lit only while pressing the SELECT key) Cancels the Auto power off function Reset all calibration values to those before shipment. Calibration function
4.8 Averaging Function
The measured value may greatly fluctuate, the instrument can calculate the
average (8 times / approx. 2 seconds). This function is available for Voltage,
Current and Resistance-mode measurement. The function (averaging) operates
until the power is turned off.
Note
Additional functions simply set when POWER ON can be used.
105
4.9 5000 display mode
This function switches 3.5-digit display (5000) and 5-digit display (50000).
The function is not available for Capacitor, Temperature, DC+AC, Continuity
and Frequencymode measurement. The function (5000 display) operates until the
power is turned off.
Note
Additional functions simply set when POWER ON can be used.
4.10 LCD Check
The instrument can lit all segments and mark for LCD check. (Lit only while
pressing the SELECT key.)
106
5. User Calibration Function
It is recommended that the instrument be calibrated periodically. The
instrument can be calibrated.
CAUTION
To avoid electrical shock · Only authorized engineers are allowed to calibrate
the instrument
using dedicated facilities. · Connect the calibrator to the instrument with
the calibrator’s test leads. · Before carrying out calibration, read the
instruction manual of the calibrator. · Temporarily remove the test leads from
the instrument before switching measurement mode
(function).
<Table 1> Carry out calibration of ranges in accordance with Table 1.
Calibration for 2 points (Input 1 and Input 2) is required other than DC
range. After Input 1, carry out calibration of Input 2 repeating steps 6) and
7). For AC voltage and AC current ranges (marked with ), calibration is
carried out at 50Hz or 60Hz frequency. 1) Turn the function switch from the
OFF position to the mV position while pressing the
SELECT and RANGE keys at the same time. The display shows the ” CAL ” symbol
then the ” PASS ” symbol. 2) Press the SELECT key. (The display shows the ” –
” symbol.) 3) Press the HOLD key twice. (The display shows the ” – – – ”
symbol.) 4) Press the RANGE key. (The display shows the ” mV ” symbol.) 5)
Connect the instrument to the calibrator with the test leads. 6) Set the
calibrator to Input 1 value as an input to the instrument. 7) Press the HOLD
key. 8) Be sure to confirm that the function switch and input terminal are set
to the desired range. Carry out calibration of other ranges by repeating steps
6) and 7). 9) To quit calibration, turn the function switch back to the OFF
position.
Note
The model KEW1062 needs to be calibrated at mean value detection (MEAN) mode
and filter on of AC current.
108
Table 1. Input Signal for Calibration
Range
Input 1
Input 2
Unit
DC 50mV
0.000
50.000
mV
DC 500mV
500.00
–
mV
DC 2400mV
2000.0
–
mV
DC 5V
5.0000
–
V
DC 50V
50.000
–
V
DC 500V
500.00
–
V
DC 1000V
1000.0
–
V
AC 50mV*1
5.000
50.000
mV
AC 500mV
50.00
500.00
mV
AC 5V *2 AC 50V
50Hz/60Hz
0.5000 5.000
5.0000 50.000
V V
AC 500V
50.00
500.00
V
AC 1000V
100.0
1000.0
V
500
0.00
500.00
5k 50k 500k 5M 50M Continuity Check ( )
0.0000
5.0000
k
0.000
50.000
k
0.00
500.00
k
0.0000
5.0000
M
0.000
30.000
M
0.0
500.0
DC 500 µA
0.00
500.00
µA
DC 5000 µA
0.0
5000.0
µA
DC 50mA
0.000
50.000
mA
DC 500mA
0.00
400.00
mA
DC 5A
0.0000
5.0000
A
DC 10A
0.000
10.000
A
AC 500 µA
50.00
500.00
µA
AC 5000 µA
500.0
5000.0
µA
AC 50mA
5.000
50.000
mA
AC 500mA
50Hz/60Hz
50.00
400.00
mA
AC 5A
0.5000
5.0000
A
AC 10A
1.000
10.000
A
1: For model KEW1062.
2: For model KEW1062. Calibration points added in 5 V AC range (setting, input value). [RMS], Filter OFF, Input1 (0.50000), Input2 (5.0000) [MEAN], Filter OFF, Input1 (0.50000), Input2 (5.0000) [RMS], Filter ON, Input1 (0.5000) , Input2 (5.0000)
109
<Table 2> After completing the calibration for ranges in Table 1, carry out calibration for ” Frequency Characteristic”. The calibration for is frequency characteristic required for AC voltage and AC current ranges (marked with ). Calibration is carried out at the designated frequency in table 2.
- Set the calibrator to Input value as an input to the instrument. 2) Press
the MEMORY key. 3) After 20 seconds, buzzer sounds and the instrument confirm
calibration.
(Do not next key operation till buzzer sounds.)
Table 2. Input Signal for Calibration
Range
Input
Unit
AC 50mV*1
50.000
mV
AC 500mV
500.00
mV
AC 5V
5kHz
5.0000
V
AC 50V
50.000
V
AC 500V
500.00
V
AC 1000V 600Hz
1000.0
V
AC 500A AC 5000A
1kHz
500.00 5000.0
A A
*1: For model KEW1062.
110
l Calibration of Capacitor Range
Before start calibration of the Capacitor range,
turn the function switch back to the OFF position.
- Turn the function switch from the OFF position to the pressing the SELECT and RANGE keys at the same time.
(Capacitor) position while
The display shows the ” CAL ” symbol then the ” PASS ” symbol.
-
Press the SELECT key. (The display shows the ” – ” symbol.) 3) Press the HOLD key twice. (The display shows the ” – – – ” symbol.) 4) Press the RANGE key. (The display shows the ” nF ” symbol.)
-
Connect the instrument to the calibrator with the test leads.
-
Set the calibrator to Input 1 value as an input to the instrument. 7) Press the HOLD key to confirm. 8) Set the calibrator to Input 2 value as an input to the instrument. 9) Press the HOLD key to confirm. 10) Carry out calibration of other ranges by repeating steps 6) to 9).
-
To quit calibration, turn the function switch back to the OFF position.
Range
Input 1
Input 2
Unit
5nF
0.500
5.000
nF
50nF
5.00
50.00
nF
500nF
50.0
500.0
nF
5F
0.500
5.000
F
50F
5.00
50.00
F
500F
50.0
500.0
F
5mF
0.500
5.000
mF
50mF
5.00
40.00
mF
111
6. Battery and Fuse Replacement
WARNING
Be careful not to burn yourself. · Fuse becomes a high temperature after
current measurement, it is dangerous by touching it
directly. When fuse or batteries are replaced after current measurement,
please be sure to leave the main unit for 10 minutes for cooling. · La
température du fusible augmente sensiblement après une mesure de courant; il
est dangereux de le toucher directement. Après le remplacement du fusible et
des piles après une mesure de courant, laissez refroidir l’instrument pendant
10 minutes” ou une annotation équivalente sera reprise dans le manuel
d’utilisation.
6.1 Battery Replacement
If the batteries fall below the normal operating voltage, the ”
“symbol turns on.
Follow the steps below to replace the batteries with new ones.
(AA-size (R6) 1.5V batteries)
Set the time by referring to the Set-up function after changing the batteries.
If you remove the batteries, the time resets to 00:00 and the time for LOGGING
measurement will not be correct.
WARNING
Be sure to disconnect the instrument from the circuit under test and test leads before replacing the batteries.
CAUTION
· Turn the function switch to OFF (turn off the power). · Do not mix batteries of different types or new batteries with used ones. · Make sure the polarities of the new batteries are exactly as shown on the battery holder.
To replace the batteries: 1) Remove the screw on the back of the
casing. 2) Remove the back cover. 3) Take the batteries out of the housing. 4)
Replace the batteries with new ones. 5) Close the casing and fasten it with
the
screw.
Battery housing Screw hole
Blank cover (For USB adapter, Printer adapter)
112
6.2 Fuse Replacement
If a current greater than the rated value flows when the instrument is in the
currentmeasurement range, a protection fuse may blow. If this happens, replace
that fuse. The instrument contains the following types of fuses.
WARNING
· Turn the function switch to OFF(turn off the power). · Be sure to disconnect
the instrument from the circuit under test and test leads before
replacing the fuses. · Do not operate the instrument with the casing left
open. · In order to avoid damage to the instrument or any possible accident,
use fuses of the
specified rating. Fuse rating: F1 M-8926 (440mA/1000V, SIBA GmbH & Co. KG, 50
210 06.0.44)
F2 M-8927 (10A/1000V, SIBA GmbH & Co. KG, 50 199 06.10) High breaking capacity
type
To replace the fuse: 1) Remove the screw on the back of the casing. 2) Remove
the back cover. 3) Remove the blown fuse from the fuse holder. 4) Install a
new fuse in the holder.
(Make sure the fuse rating.) 5) Close the casing and fasten it with the screw.
F1: (440mA/1000V)
F2: (10A/1000V) Screw hole
113
7. Calibration and Maintenance
Calibration Cycle It is recommended that the instrument be calibrated once
every year. (SEE ALSO: User Calibration Function)
Contacts of Services Please contact the sales representative from which you
purchased the instrument.
8. Disposing the Product
Waste Electrical and Electronic Equipment (WEEE), Directive 2002/96/EC This
Product complies with the WEEE Directive (2002/96/EC) marking requirement. The
affixed product label (see below) indicates that you must not discard this
electrical/ electronic product in domestic household waste.
Product Category With reference to the equipment types in the WEEE directive
Annex 1, this product is classified as a “Monitoring and Control
instrumentation” product.
114
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.
3-22
92-1986J
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>