cetac ASX-560-280 Home Position Error Instructions

June 9, 2024
cetac

cetac ASX-560-280 Home Position

ASX-560-280 Home Position Error

Introduction

Refer: New install CETAC autosampler ASX 560/280 on icap 7000 series or ICAPPRO and Qtegra software.

Object: After autosampler installation it may occur running a labbook, that Qtegra stops when the probe returns to home position from random rack position, the autosampler led in dashboard turns red, showing false communication issue.

When you are installing a new autosampler CETAC ASX-560/230, ensure that it moves correctly. Create a labbook with a simple calibration, including all standard positions.

Running the labbook, when the autosampler probe returns from a random standard rack position (typical is from position 5 or 6) or sample rack position, it may happen that Qtegra stops unexpectedly, autosampler led turn red, and seems that it has lost communication with PC.

Check Qtegra log file, if you find the below error sequence


The error confirm that is not a communication issues but is due of a bad alignment.

To solve this issue, install Teledyne CETAC Autosampler Alignment or Teledyne CETAC Dashboard tool and perform full alignment.

Follow instructions on document “ASX-280 & ASX-560 Alignment Guide.pdf” that you can find inside Cetac CD or attached to this KB article.

If during alignment procedure, Alignment utility doesn’t find autosampler port, it may be in use by Qtegra software, in Qtegra change the configuration to one that doesn’t use the autosampler, (manual configuration or empty configuration).

IF there are any problems to load alignment files, add keyboard language and set to English.

To confirm problem resolution run again a labbook with calibration.
If problem persist report the issues to 2nd Level Tech support Specialist
matteo.bignardi@thermofisher.com and Bremen Support support.TEA.bremen@thermofisher.com