RICHTER OPTICA HS-3M Student Compound Microscope Instructions
- June 10, 2024
- RICHTER OPTICA
Table of Contents
- HS-3M Student Compound Microscope
- Product Information
- Product Usage Instructions
- DESCRIPTION
- UNPACKING
- BEFORE USE
- SETTING UP THE INSTRUMENT
- MAINTENANCE
- ASSEMBLING THE MICROSCOPE
- TROUBLESHOOTING
- CARE AND MAINTENANCE
- LED Light Replacement
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HS-3M Student Compound Microscope
Product Information
The Richter Optica microscope comes in three models, the HS-3M, HS-3B, and HS- 3T. It has a carrying handle for easy movement, and it also has a monocular head and a binocular head for viewing. The microscope has a revolving nosepiece, objectives, and an eyepiece. It also has a specimen fine/coarse holder and a focus knob. Additionally, it has a condenser and an aperture adjustment lever. The microscope comes with a lamp housing and an indicator light. The stage has X/Y axis travel knobs and a mechanical stage. The microscope also has a fine/coarse focus knob and a lock screw.
Product Usage Instructions
When unpacking the microscope, do not discard the styrofoam container or packing materials until you are sure the shipment is complete and undamaged. Do not touch the lens surfaces with dirt, fingerprints, or oil as it can damage the lens surfaces and reduce image quality. When moving the microscope, use both hands and only lay it on flat, even surfaces. Do not hold the stage, focusing knobs, or head when moving the microscope. Always hold the carrying handle when moving the microscope. Before replacing the bulb or fuse, ensure that the power switch is turned off, and wait until both the bulb and bulb holder have cooled down. To assemble the microscope, insert the power connector end of the power cord into the power socket located on the back of the microscope. Then plug the other end of the power cord into a wall outlet (110~240V). If you are using the cordless option, replace the four AA rechargeable batteries by locating the battery box on the underside of the microscope and using a compatible screwdriver to remove the screw found on the battery box cover. It is recommended to wipe the lenses only with lens paper and to cover the microscope with the dust cover provided after each use. Avoid placing the instrument in locations exposed to direct sunlight, dust, vibration, high humidity, and where it is difficult to unplug the power supply cord. Do not disassemble the microscope other than to change the bulb. Disassembling the microscope will affect performance and void the warranty. When the microscope is not in use, wind the power cord as shown in the photo provided.
DESCRIPTION
Thank you for your purchase of a Richter Optica microscope. The information in this manual is provided to answer most questions that can arise when operating your microscope and to help you avoid unnecessary maintenance expenses in the future. Please carefully read instructions before operating microscope. Nomenclature used to describe components and controls are identified on opposite page.
UNPACKING
Do not discard styrofoam container or packing materials until you are sure shipment is complete and undamaged (retain styrofoam shipping container to store your microscope when it is not in use). Remove all tape and packing material used to protect microscope during shipment. Make certain lens surfaces do not come in contact with dirt, fingerprints or oil. Damage of lens surfaces occur when they come in contact with such contaminants, and image quality is reduced.
BEFORE USE
When moving the microscope, use both hands and only lay on flat, even surfaces. Damage will occur by holding the stage, focusing knobs or head when moving the microscope. Hold carrying handle when moving the microscope. For safety, make sure the power switch is always turned to the off position “O” before replacing the bulb or fuse and wait until both the bulb and bulb holder have cooled down. Replacement bulb: 2W LED replacement module. Part #HS3-001
SETTING UP THE INSTRUMENT
Avoid placing the instrument in locations exposed to direct sunlight, dust, vibration, high humidity and where it is difficult to unplug the power supply cord.
MAINTENANCE
- Wipe lenses only with lens paper.
- Never disassemble the microscope other than to change the bulb (instructions on page 6). Disassembling the microscope will affect performance and void warranty.
- Cover microscope with dust cover provided after each use.
ASSEMBLING THE MICROSCOPE
POWER OPTION
Corded option: Ensuring the light adjustment knob is “O” (OFF), insert the
power connector end of the power cord into the power socket located on the
back of the microscope. Then plug the other end of the power cord into a wall
outlet (110~240V).
POWER OPTION
Cordless option: Your microscope can be operated with four AA rechargeable
batteries. To replace batteries,ensure the light adjustment knob is “O” (OFF).
Locate the battery box on the underside of
the microscope. Using a compatible screw driver, remove the screw found on the
battery box cover and replace the rechargeable batteries.
POWER CORD STORAGE
When the microscope is not in use, wind the power cord as shown here in the
photo.
TROUBLESHOOTING
Optical
PROBLEM | POSSIBLE CAUSE |
---|---|
Dust or dirt in field of view. | Dust or dirt on objective, illuminator or |
eyepiece.
Double Image
| Interpupillary distance is not set correctly.
Diopter adjustment not set on eyepieces.
Two different eyepieces are being used.
Sample blurs when magnification changes.
| Diopter adjustment is not set properly on eyepieces.
Specimen is not properly in focus.
Focusing is loose or too tight.| Adjust tension adjustment knob on the side of
the microscope next to the focusing knob.
Electrical
PROBLEM | POSSIBLE CAUSE |
---|
Lamp does not turn on and light up.
| Power cord not plugged in, or batteries are not charged.
Light is burned out.
Inadequate brightness.| Rheostat knob needs adjustment.
Lamp blows out immediately.| Incorrect bulb being used.
Lamp flickers.
| Connector wires are not secured.
Lamp near end of life.
Lamp not securely plugged into socket.
CARE AND MAINTENANCE
Do Not Disassemble
- Disassembly may significantly affect the performance of the instrument, may result in electric shock or injury, and will void the warranty.
- Never attempt to dismantle any parts other than described in this manual. If you notice any malfunction, contact your Richter Optica representative.
Cleaning the Microscope
- Do not use organic solvents such as ether, alcohol or paint thinner on painted surfaces or plastic components. Doing so could result in discoloration of surfaces.
- When cleaning lenses do not use any solvents other than absolute alcohol, as they may damage lens bonding cement.
- Do not use petroleum benzene when cleaning components such as lenses.
- Absolute alcohol and petroleum benzene are highly flammable. Keep away from open flames and when turning power switch on and off.
- For stubborn dirt, dampen a piece of gauze with diluted neutral detergent and wipe gently.
When Not In Use
- When not in use, cover the instrument with a dust cover and store in a place low in humidity where mold is not likely to form.
- Proper handling of the microscope will ensure years of trouble free service.
- If repair becomes necessary, please contact Richter Optica at info@richter-optica.com.
LED Light Replacement
Before performing any of these steps, the microscope MUST be turned off and unplugged.
- Unplug the microscope and gently lay the microscope down on a sturdy surface ensuring the carrying handle side is down.
- Using a screw driver, remove the 4 foot screws located on the underside of the microscope.
- Gently remove the black plastic cover.
- Locate the LED bulb.
- Disconnect the wires from the LED bulb by using your thumb and forefinger to pinch the sides of the plastic wire housing that holds the black and red wires.
- Using a screw driver, remove the small screws on either side of the LED bulb.
- Place the new LED bulb (part# HS3-001) on the LED bulb housing and attach using the small screws previously removed in step 6.
- Insert the plastic wire housing back into the underside of the LED assembly making sure it is fully inserted and secure.
- Replace the black plastic cover and secure using the 4 foot screws removed in step 2.
- Return the microscope to its upright position. Plug the microscope back in, power on, and adjust the rheostat control as needed.
Replacement 2W LED Assembly Part #HS3-001
info@richter-optica.com www.richter-
optica.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>