Amico Nova Motorized Exam Light Instruction Manual

June 10, 2024
Amico

Nova Motorized Exam Light
Instruction ManualAmico Nova Motorized Exam Light Operating and Maintenance Instructions
Nova Motorized Exam Light

Introduction
Thank you for your purchase of the Nova motorized Exam light.
This installation manual gives you instructions on how to install the Nova motorized exam on the Ceiling.
For instructions on operation, maintenance and further description of the product please refer to the Operation and Maintenance Manual included.
Please read the following instructions very carefully and follow the safety instructions and requirements.

Symbols Used in This Manual

The warning symbols indicate all instructions that are important for safety. Failure to comply with them can lead to injury, damage to the luminaire or the equipment. In combination with the following signal words the warning symbols means:

DANGER Can lead to death or serious injury
WARNING Can lead to injury
**** Comply with operating instructions
**** Protective conductor
N Neutral conductor/return conductor for the electrical current
L1 Live conductor
Storage humidity
Storage temperature
**** Disposal
Manufacturer
Read manual
RH Relative humidity
P Live conductor

Trademarks: All trade names mentioned in these instructions are the exclusive property of the relevant manufacturers.

Intented Use

The Nova motorized exam Light is intended to provide illumination for birthing room, exam room, and intensive care unit. It is to be installed to the ceiling to support treatments, which pose no risk to the patient in the event of any interruption caused by light failure. It is meant for continuous operation.

User Profiles

Medical Staff
All persons with medical training who work in the field for which they were trained.
Cleaning Staff
All persons familiar with national and workplace hygiene requirements.
Electricians
All persons with training in electronics and electrical technology who are familiar with the relevant standards and regulations.
Certified Installers
Persons qualified by training, experience and knowledge of the regulations to install and remove the equipment.

Safety Instructions

Please note that certain duties must be carried out by people with the appropriate corresponding qualifications.
The instructions are part of the product and must be kept so they are accessible for all subsequent users:

  • These instructions are for the Amico Lights Nova motorized exam lights. This includes ceiling mounted equipment
  • Please carefully read these operating instructions before using the equipment. It will help protect you and others from potential injuries that may occur
  • All work on the light (including repairs) must be done by qualified electricians only. The light must be installed by a certified installer
  • The light must not be modified or manipulated. Use in any other manner than with the original parts may result in different specifications and life-threatening dangers
  • Only persons with medical training are permitted to operate the equipment
  • Only trained cleaning personnel are permitted to clean the equipment
  • The light must not be used in an explosive environment. The power supply to the light is a potential ignition source
  • The light must be used in dry rooms only
  • The light must not be left on unattended
  • Keep all flammable equipment and objects away from the fixture during use
  • Do not use a damaged light. Defective cables can also be a potential hazard. Do not position cables near heat sources or sharp edges
  • Eye damage. Never look directly at the light cone
  • Replace damaged lenses before using the light again
  • The light must be connected to the mains with a protective earth to prevent electric shock
  • Never apply additional loads to the light fixture
  • Do not cover the light with a cloth or similar material while operating
  • If problems are encountered that have not been addressed in these operating instructions, please contact your supplier in the interest of your own safety
  • Do not place the fixture such that it is difficult to remove the power cord

Please Keep These Operating Instructions in a Secured Place Near the Device. This Ensures Easy Access to Safety Instructions and Important Information Should the Need Arise.

  • The inspections and maintenance work, as described on page 13, should only be done by trained service technicians. Anyone else is not authorized to perform inspections and maintenance work
  • The adjustment work should only be carried out by a hospital technician or someone with similar qualifications. Anyone else is not authorized to perform adjustment work
  • This equipment has been built accordingly and is operationally safe. Nonetheless, it may still present some danger, especially if not used according to the operating instructions as specified in this manual
  • Modifications are never to be made for any reason. This is to ensure safety for all persons involved. Any modifications made to the system void the warranty
  • The contents of the operating instructions are subjected to change without notice

Specific Fitting Instruction Nova Motorized Exam

  • Mounting materials are included in delivery
  • The ceiling mount may be fitted to ceiling only with a concrete strength of B25 (C20/25) or greater
  • Contact with reinforcing elements of the solid slab ceiling must be avoided. In case of doubt, a structural engineer should control the mounting on the respective surface
  • The drill holes should be made by a skilled person in accordance with the drill hole tolerances recommended by the manufacturer of the fastening anchors
  • The screws must be carefully tightened using a torque wrench in accordance with the recommendations of the manufacturer of the fastening anchors

DANGER
Indication of hazards that can lead to death or serious injury if measures are disregarded.
WARNING
Indication of hazards that can lead to injury if measures are disregarded.
CAUTION
Indication of hazards that can lead to damage to property if measures are disregarded.
EXPLOSION HAZARD
The above is an explosion hazard warning symbol. Non-compliance with this symbol may result in damages or injuries due to explosion.
ELECTRIC SHOCK
The above is an electric shock warning symbol. Non-compliance with this symbol may result in damages or injuries due to electric shock.

Scope of Delivery

Ceiling Mount
Included:Amico Nova Motorized Exam Light - Scope of
Delivery

Installing Nova Motorized Exam Light

DANGER
MOUNTING BY QUALIFIED PERSONNEL.

  • Mounting must be done by qualified personnel only. Lack of appropriate knowledge could be life-threatening

DANGER
LIFE-THREATENING DANGER FROM FALLING LUMINAIRE.

  • The ceiling must be made of solid concrete to guarantee a secure hold

DANGER
ELECTRIC SHOCKS ARE LIFE-THREATENING.

  • The ceiling must be made of solid concrete to guarantee a secure hold

DANGER
RISK OF INJURY FROM FALLING PARTS.

  • Mounting must be performed by two people

DANGER
USE PROTECTIVE EQUIPMENT IN LINE WITH TOOL MANUFACTURER’S INSTRUCTIONS.
DANGER
Danger of injury from loose parts.
NOTE: INSTALL THE MOUNTING PLATE TO THE CEILING AS PER THE STRUCTURAL ENGINEER.

  1. Loosen the screw and open conduit door on the Nova box.Amico Nova Motorized Exam Light - Nova box 1

  2. Use a hammer and flathead screwdriver to open the knockout.
    Wiring needs to be done after opening the knockout.
    Amico Nova Motorized Exam Light - Nova box 2(Refer to Wiring on page 12).
    Push conduit into the opened knockout and reassemble knockout door with the Nova box.

  3. Make sure to adjust vertical position of mounting bars within the bracket so that the parts rest on adjacent T-channel grid. Secure the mounting bars to the grid as required.Amico Nova Motorized Exam Light - Nova box 3

  4. For drywall or plaster ceiling, be sure to adjust vertical position of mounting bars within the bracket so that the parts rest on adjacent ceiling joists. Secure the mounting bars to joists as required.Amico Nova Motorized Exam Light - Nova box 4

  5. Recommended dimension of the hole cutout will be given. Recommended Diameter: 11.5 inches.Amico Nova Motorized Exam Light - Nova box 5

  6. Insert the bezel into the cutout, make sure the bottom edge of the bezel is flush on the bottom edge of the cut hole. Secure the bezel to the light fixture using flathead screw.Amico Nova Motorized Exam Light - Nova box 6

  7. Section view of the installed fixture. Make sure the bezel is secured and flush to the ceiling.Amico Nova Motorized Exam Light - Nova box 7

Electrical Wiring

Nova Motorized Exam Light Electrical Wiring – One Keypad
Single Light Fixture, One Keypad
Amico Nova Motorized Exam Light - Keypad 1Double Light Fixtures, One KeypadAmico Nova Motorized Exam Light - Electrical
Wiring

Single Light Fixture, Two KeypadsAmico Nova Motorized Exam Light -
Electrical Wiring 1Double Light Fixtures, Two Keypads
Amico Nova Motorized Exam Light - Electrical Wiring 2

Operation

Nova Motorized Exam Light Control Keypad
Switch on the Nova light from control keypad.
Keypad Options:Amico Nova Motorized Exam Light - Keypad
Options

Commissioning and Testing Checklist

  • Ensure the power is working on the keypad by ensuring the keypad is LED backlit.
  • Press the ON/OFF button on the keypad to ensure power is connected to the fixture.
  • Ensure the keypad controls are working properly by moving the fixture in all directions (left, right, up, down).
  • Note: UP direction should point towards the head of the patient bed and the DOWN direction should point towards the foot of the patient bed.
  • Check that the Dimming feature is working correctly by pressing and holding the Dimming levels on the keypad.
  • Ensure both fixtures (#1 & 2) are working correctly from the same keypad, and repeat steps above (if applicable).

Keypad Options:

Amico Nova Motorized Exam Light - Keypad Options 1 Ensure Lens (bezel) is securely fastened to the fixture.
Insert the bezel into the cutout, make sure the bottom edge of the bezel is flush on the bottom edge of the cut hole. Secure the bezel to the light fixture using flathead screw.
Amico Nova Motorized Exam Light - securely fastened

Cleaning, Disinfection, and Sterilization

NOTICE:
Damage due to improper cleaning authorized to perform inspections and maintenance work.

  • Do not use alcohol, solvent or cleaning agents containing chlorine or scouring agents
  • Damage to lens by concentrated disinfectant or solvent
  • Follow the manufacturer’s recommended mixing ratio
  • Use of incorrect cloths may scratch lens

NOTICE:
A dirty lens will reduce the illuminating power.

  • Clean lens regularly
  • Spray and wipe disinfection is permitted
  • Clean the lens with a leather cloth soaked in cleaner
  • Spray disinfectant in horizontal position

NOTICE :
The external components of the light can be disinfected with the following diluted products:

  • Lysoformin
  • Dismozon
  • Hexaquart Plus
  • Sagrotan – Fast Disinfecting Cleaner
  • Tuffie 5

NOTE: THE LOCAL GOVERNMENT REGULATIONS FOR DISINFECTION MUST ALSO BE OBSERVED IN ADDITION TO THESE INSTRUCTIONS.

Maintenance

DANGER
ELECTRIC SHOCKS ARE LIFE-THREATENING.

  • Remove the connector from the power supply and put the switch in the “OFF” position
  • The connector cable must be checked for damage at least once per year

NOTE:

  • Maintenance and repairs must only be carried out by qualified electricians
  • Nova motorized exam light is not serviceable or repairable by anyone other than the manufacturer
  • Locking mechanism should be provided by contractor as means of isolation

Notes on the Systems
Selection of Fastening Material
We recommend the use of (not included in the delivery) counter-plates for fastening to drywall.
Cleaning and Disinfecting
DANGER
ELECTRIC SHOCKS ARE LIFE-THREATENING.

  • The devices can conduct electricity and must be handled with care during cleaning and disinfection
  • If there is a power plug, remove it from the socket
  • Do not use spraying methods for cleaning and/or disinfection
  • Do not spray liquid into the openings of electrical power sockets, gas sockets or devices, or allow liquid to enter these

Recommended Cleaning Agents
Use mild soap solutions or standard detergents as cleaning agents. Wipe down the surfaces of the equipment with a dampened cloth. If necessary, use small quantities of mild soap solution as a cleaning agent. Then wipe the outer surfaces with a soft, clean cloth until they are completely dry.
Disinfection Procedure
Disinfection by wiping is intended to be the standard disinfection procedure for the support arm system. Hygiene guidelines and corresponding safety measures must be determined by the operator for the disinfection procedures that are applied.

  • After the contamination by any potentially infectious material (such as blood, secretions or excrement), the areas must be disinfected immediately in a targeted way
  • Comply with application concentrations
  • Wipe the area to be disinfected rather than spraying it
  • Do not use areas that have been wiped until the disinfectant is dry

Testing
DANGER
ELECTRIC SHOCKS ARE LIFE-THREATENING.

  • During all testing work, disconnect the device from the power supply/remove the plug from the power socket and secure the device so that it cannot be switched on again

Every six (6) months check for:

  • Paint damage
  • Cracks in plastic parts
  • Deformation of the support system
  • Loosening of parts

Every two (2) years:

  • The securing element of the cantilever arm must be checked and greased by a qualified professional

Troubleshooting

Malfunction Possible Cause Remedy User Profiles
Light does not come on Contact fault Switch on again All
Illuminant defective Contact manufacturer’s AfterSales
Service Only by manufacturer’s

After-Sales Service
No mains power| Check power supply and all
connections| Qualified electrician

Disassembly

DANGER
ELECTRIC SHOCKS ARE LIFE-THREATENING.

  • Remove connector from the power supply and put the switch in the “OFF” position before disassembling

Disposal
**** Do not dispose of this product with normal refuse. Dispose of the product at a recycling center or return it to a dealer with a disposal service depending on local regulations. Cut off the cable flush with the housing.

Technical Data of Nova Motorized Exam Light

| Units| Nova
---|---|---
Color Temperature| K| 4000 K
Central Illuminance @2.5 m| LUX| 24408 Lux
Ultraviolet radiation| W/m2| 0.0025 W/m2
CRI| Ra| 92.5
Voltage| Vac| 120 V
Current| A| 0.4 A
Frequency| Hz| 50/60 Hz
Total Irradiance| W/m2| 80.46 W/m2
R9| R9| 70.2

NOTE: ESSENTIAL PERFORMANCE PARAMETERS:

  • Delivery of minimum and adequate illumination on the diagnostic field
  • Limitation of energy in the diagnostic field

Electrical:

Nominal inputvoltage
Frequency range
Input current
Device Protection| 120 VAC
50/60 Hz
0.4 A Max
Class 1 Approval ETL/cETL
---|---

Environmental Conditions for Transport, Storage, and Operation:
Storage and transport
After 15 weeks the conditions for operation apply.
-10°C to +34°C Ambient 20% to 90% humidity 70 kPa to 94 kPa
Operation:
This product should be acclimatized to its operating conditions for 24 hours prior to its first use.
+10°C to +34°C 70 kPa to 94 kPa 20% to 90% humidity
Weight:
Nova LED Motorized Exam Light………….. 9.8 kg (21.5 lbs)
Operating Mode:
Continuous operation
Classification:

Nova LED Motorized Exam Light
Classification under 93/42 EEC – Annex IX
Safety testing according to:| Class I
Class I (Medical device class)
CSA C22.2 #60601 – 1 IEC/EN 60601 – 1 2nd. UL 60601-1
---|---

Warranty Policy

Medical Lighting Solutions
(Clarity Series, Gamma Series, Magna Series, Mira Series, Nova Series, Vega Series)
Amico Lights Corporation warrants its products against defective materials and workmanship for twelve (12) months from the date of shipment. Within this period, Amico Lights Corporation will repair or replace any part on site or at the factory which is proven to be defective at Amico Lights Corporation’s cost.
Furthermore, Amico Lights Corporation will warrant its material to be free from defect for an additional period of four (4) years (five [5] years from the date of shipment). Within this period, Amico Lights Corporation will replace any part at no charge, which is proven to be defective. Shipping and installation costs after the first twelve (12) months will be borne by the customer.
The spring arms for the above mentioned are warranted for a period of twelve (12) months. The warranty for the spring arms shall include parts and labor and shall commence from the date of shipment.
The warranty applies to normal usage and does not apply to any product that has been subject to alteration, abuse, negligence, or use (including voltage and/or current) other than that for which the product was designed.
Amico Lights Corporation shall not be liable for incidental or consequential damages resulting from the use of the equipment. All claims for warranty must first be approved by Amico Lights Corporation. A valid Return Goods Authorization (RGA) number must be obtained from Amico Lights Corporation prior to commencement of any service work. Warranty work which has not been pre-authorized by Amico Lights Corporation will not be reimbursed.

Amico Lighting Solutions
122-B East Beaver Creek Road
Richmond Hill, ON L4B 1G6, Canada
Toll Free Phone: 1.877.462.6426
Toll Free Fax: 1.866.440.4986
Tel: 905.764.0800
Fax: 905.764.0862
Email: info@amico.com 
www.amico.com
ALT-INSTAL-INSTR-NOVA-MOTORIZED-EXAM-LIGHT 09.19.2022