Product Information
The Toilet seat lift Liftolet-Tilty (models Liftolet-Tilty, Liftolet-Tilty Duo) is a high-quality and secure product designed to assist individuals with mobility issues. It features an easy-to-operate design and modern aesthetics. The product comes with a variety of accessories and options to meet different needs and preferences. The product is CE-marked and complies with the latest directives. It is subject to ongoing quality control.
Product Usage Instructions
Before using the Toilet seat lift, carefully read through the user manual provided. The user manual is an integral part of the product, and it should be kept safely accessible to care and maintenance staff.
Safety Instructions
It is important to follow the safety instructions provided in the user manual to prevent accidents. The user manual provides a guide to the most important handling, but it does not replace the experience and knowledge that operators may have gained from training courses or supervisory staff. Ensure all users of the product understand its handling and the meaning of the symbols attached to the device.
Assembly and Adjustment
After reading the user manual, follow the instructions provided for initial operation, assembly, and adjustment. For the basic model, assembly and adjustment instructions are provided starting from page 13. For accessories and options, refer to page 18. It is recommended that assembly and adjustment be carried out by trained personnel.
Cleaning and Disinfection
The Toilet seat lift should be cleaned and disinfected regularly according to the instructions provided on page 22 of the user manual. Failure to do so may result in contamination or infection.
Maintenance and Overhaul
Regular maintenance and overhaul are important to ensure the product remains in good condition and continues to function correctly. Instructions for maintenance and overhaul are provided on page 24 of the user manual.
Operating Life and Re-use
The operating life of the product and its ability to be reused depend on proper handling, maintenance, and overhaul. Instructions for determining the operating life and reusability of the product are provided on page 26 of the user manual.
Faults, Repairs, Spare Parts
If faults occur or repairs are needed, refer to page 27 of the user manual for instructions. It is important to use genuine spare parts to ensure continued functionality and safety of the product.
Storage
When not in use, store the Toilet seat lift safely in a dry and secure location. Instructions for storage are provided on page 30 of the user manual.
Dear Customer!
We would like to thank you for your trust extended to us and for purchasing
our product.
The Toilet seat lift Liftolet-Tilty (models Liftolet-Tilty, Liftolet-Tilty
Duo) distinguishes itself by its quality, security, easy operation and modern
design.
For optimal use of this aid, we ask you and possible other users to carefully
read this operating manual.
Please read through these instructions carefully before first use and handle equipment appropriately!
instructions
The Toilet seat lift Liftolet-Tilty is CE-marked and fulfills the requirements
of the latest directives. It was developed in line with most recent findings.
Production is subject to an on-going quality control.
Please read this user manual carefully before operating the Toilet seat lift.
The user manual is an integral part of the Rehab aid. In case of loss, request
a new copy or download directly from our down-load area on our homepage www
.ato-form.com.
Please keep the user manual in a safe place, accessible to care and
maintenance staff.
The user manual does not replace the experience of supervisory staff with the
experience and knowledge of operators, which they gained from (for example)
training courses.
In the event of specific or difficult applications, the user manual represents
a guide to the most important handling.
Ensure all users of this aid understand its handling and the meaning of the
symbols attached to the device.
Accidents can be prevented when technical instructions according to the EWG
guidelines are followed.
Protective coverings, labels and warning notices may not be removed.
The maximum load-bearing capacity of the Liftolet-Tilty (standard version) is
85 kg.
The maximum load-bearing capacity of the Liftolet-Tilty Duo is 155 kg.
Regular monitoring at your specialist dealer (at least once a year) lengthens
the life time of the Toilet seat lift and gives you security.
Please also see section: maintenance and overhaul.
-
Before mounting or using the Toilet
seat lift Liftolet-Tilty carefully read all instructions of this user manual, as these have been drawn up for your safety
and as a guideline for the correct use and maintenance of this product. This operating manual is an indispensable component of the delivery and should be stored carefully for further use. -
The operating manual is intended for retailers and users of the Toilet seat lift Liftolet-Tilty, (or their careers). It contains the required information to secure the correct use of the product, it cannot, however replace the necessary experience of the user or assistant in handling the device.
-
The reseller has the duty to deliver the device ready for operation and to explain precisely the functions and technical data of the device, as well as provide the necessary support and briefing needed for a correct and efficient use of the device.
-
The given instructions of the operating manual, refers to the point in time of the device being placed on the market. The manufacturer reserves the right to introduce any modifications which prove necessary to the device and associated operating manual, without being bound to change previous devices and associated operating manuals.
-
The Toilet seat lift Liftolet-Tilty must only be operated under supervision of a qualified person, or respectively a family member of the user.
-
The Liftolet-Tilty may only be operated by persons who have familiarized themselves with the standards and rules of the present operating manual.
-
Any potential assistant must be familiar with the clinical picture of the user.
-
The Liftolet-Tilty may only be used by persons of a suitable stature and physical aptitude.
-
Never place a patient in an aid, which
is not coordinated with his or hers body height and weight. -
Adjust settings only in the unloaded condition.
-
Do not place any hot beverages or objects onto the aid you have purchased, as this would seriously damage the aid.
-
Do not load any heavy objects onto the aid, as this would impair the structure, the centre of gravity and consequently impair the balance.
-
Always ensure there are no obstacles in operating range of the rehab aid, such as carpets, stairs or objects on the floor, which could lead to the Toilet seat lift Liftolet-Tilty overturning.
-
Check all adjustable parts are locked, before every use of the aid.
-
Please note the individual and overall safety relevant information when using several rehabilitation aids simultaneously.
-
All aspects of spatial influences (uneven, wet and smooth floors, steps or similar) must be taken into consideration by the career or user.
-
Check regularly if all screws and handles etc. are firmly tightened. Worn and defect parts are to be repaired or replaced immediately. This is especially important to ensure a safe use of the rehab aid! Carefully check the clamping jigs and the like are securely fixed.
-
Do not use the aid, if parts of it are damaged or missing. Missing parts are only to be replaced with original spare parts.
-
Do not carry out any interferences or measurements if you are unsure of their effects. For your own safety contact the nearest service center or contact the dealer directly.
-
In case of accidents or malfunctions, contact the dealer or manufacturer from whom you purchased the device.
-
The Liftolet-Tilty is connected via a power plug to a standard household wall socket. The common safety regulations according to the German VDE (Verb and Detacher Elektrotechniker) standards for electricity grids in damp rooms are to be observed (electrical safety areas in damp rooms, socket type, protective conductor).
-
Regularly inspect the connector or
cable connections and the socket for damage. In the case of damage contact immediately contact the manufacturer or an electrician. -
A connection with an extension cable or multiple sockets is not permitted.
-
Battery operated model: when in use never allow the battery to discharge completely (observe charging indicator)! When the load limit has been reached an acoustic warning signal sounds.
-
Battery operated model :the battery must be charged for at least 24 hours before the first use of the Liftolet.
-
The batteries are only to be charged in dry conditions.
-
The manufacturer is not liable for damage to persons, objects or animals resulting from false usage of the equipment by the user or third parties, through incorrect mounting of parts and accessories, by neglect or poor maintenance, neglect, through unauthorized alterations and modifications, usage of non-original spare parts or accessories, through unpredictable and exceptional events resulting from non-compliance of the regulations in the presented operating manual.
-
Clean the Toilet seat lift Liftolet-Tilty in accordance with the instructions given in the chapter „cleaning”.
-
Please ensure before use the product is cleaned according to the cleaning instructions and if indicated also disinfected.
-
Any serious incidents occurring in connection with the product must be reported to the manufacturer
(ATO FORM GmbH) and the competent authority of the country in which the user is resident.
Warranty and product liability
ATO FORM grants the required by law warranty of your country against defects
in workmanship and materials.
The warranty would be adversely affected if the customer’s responsibility of
servicing and/or daily maintenance is not carried out according to the
guidelines and intervals prescribed by the supplier and/ or stated in the
manual.
For further information, we refer to the ATO FORM homepage/download www.ato-
form.com.
The warranty can only be sustained if the ATO FORM product is in use in the same country where it was purchased and if the product can be identified by the serial number. The warranty does not cover accidental damage, including damage caused by misuse or neglect.
The manufacturer‘s obligations transfer then to the person carrying out the modifications.
The warranty does not extend to non-durable parts, which are subject to
normal wear and tear and need periodic replacement.
The warranty is null and void if non-original ATO FORM parts/accessories are used, or if the product is repaired or altered by anyone other than an authorized ATO FORM representative or by trained personnel officially recognized by ATO FORM for repair and maintenance of ATO FORM products.
ATO FORM reserves the right to inspect the product being claimed for and the
relevant documentation before agreeing to the warranty claim and to decide
upon whether to replace or repair the defective product. It is the customer’s
responsibility to return the item being claimed for under warranty to the
address of purchase.
The warranty is given by ATO FORM or, subsequently, an ATO FORM dealer.
CE sign and position
The CE label with details of the manufacturer, article number, serial number and date of manufacture is located on the base at the rear right.
ato form
www.ato-form.com
Indications and prophylaxis
- Amyotrophic lateral sclerosis (ALS) Arthritis
- Osteoarthritis
- Multiple sclerosis
- Muscular dystrophy
- Paresis
- Parkinson‘s disease/Morbus Parkinson
- Rheumatism
- Gout patients
- Polio and other muscle disorders
contraindications
- Inability to walk or stand
- Inability to sit
- Inability to put weight on the feet for a short time despite support
The list of contraindications does not grant guarantee completeness.
The patient‘s general condition is an important part of the risk assessment.
In individual cases, the decision must be made by the attending physician, who
must weigh the risks versus the expected benefits.
Purpose of use
The device must be prescribed by a medical specialist, who has tested
beforehand and determined if the Liftolet-Tilty is suitable for the individual
therapeutic goal of the patient. Furthermore, the medical specialist has
determined the aid to be suitable according to measurements, weight and
willingness of the patient to use the aid.
The aid is delivered via an authorized technician in compliance with the
requirements of the country where it is being used and in strict accordance
with the information and instructions in this manual for adjustments of the
aid according to the measurements of the user. Only compliance of the
regulations and instructions can lead to positive ergo-nomically effects.
We wish you every success when using the Toilet seat lift Liftolet-Tilty.
Product overview
Basic model
- Lifting columns
- Carrier plate for the WC seat
- WC-seat with lid
- Upward folding support handles
- Remote control
- Rechargeable battery unit
Image model Liftolet-Tilty Duo with maximum capacity 155 kg.
Accessories / options
For a complete overview of our extensive range of accessories for the
Liftolet-Tilty see item 11.2 Types and accessories .
Initial operation
The reseller is to unpack the unit, ensure that the components and accessories
listed in the scope of delivery are all present and undamaged, and give the
buyer the initial instructions on use of the aid. Above all the reseller
should read through the complete present user manual carefully, in order to
then explain it to the customer in detail.
This way, with help of the reseller, the user as well as the attendant must
become familiar with the correct operation of the aid.
Once the demonstration has been completed, the reseller must ensure the user
and/or the assistant have fully understood the basic functions of the
equipment. This is the only way to prevent the user from dangerous situations.
The above mentioned requirements must be observed for first use and also for
reuse.
The seller must give instruction for use, as well as explain the basic
functions of the device according to MPG (medical products law) to qualified
staff.
Assembly and adjustment
Basic model
Preparation:
- Unpack the product and remove the packing materials.
- The device is completely assembled.
- Ensure that the product is complete and undamaged.
One-man assembly:
- Liftolet can be mounted by one person.
Space requirement:
- If the Liftolet is used by a wheel chair user, then the entry side is according to the space available.
Preparing for self-assembly:
- Unpack the product and remove the packing materials.
- Ensure that all supplied components are present and undamaged.
- Have the required tools, components and mounting elements ready to hand before beginning the assembly.
- The floor area on which the Liftolet is to be installed should be level. If the floor surface is uneven, measures should be taken on-site to produce a surface that is even or suitably leveled out.
- Ensure that the floor area on which the Liftolet is to be installed is dry and clean.
Mounting :
-
After the conditions are given for mounting, place the Liftolet-Tilty over the WC and position correctly.
Attention:
When positioning the Liftolet-Tilty, do not grasp it by the support handles but only by the lifting columns. -
Slight unevenness in the floor can be compensated for with the adjusting screws on the feet.
-
Loosen the locking screw
-
Level out the height with the adjusting screws so that the Liftolet tilt stands securely on the floor with all four feet.
-
Secure the height of the adjusting screws with the locking screws by tightening them firmly.
-
Liftolet-Tilty should stand horizontally. The use of a suitable aid (e.g. spirit level) is recommended for adjustment.
Attention:
Only minor unevenness can be compensated for with the adjusting screws. Make sure that you do not turn them out too far. The adjustment range is 10 mm. -
Please note that the structural rules are to be observed. It must be ensured that the floor condition can withstand the load and no cables in the ground are dam-aged.
Attention:
Make sure that Liftolet-Tilty does not fall over when it is raised.
Attention:
Check at regular intervals that Liftolet-Tilty stands securely on the floor with all four adjusting screws. This is absolutely necessary for safe handling of this aid.
Liftolet Tilty can be equipped with a rechargeable battery unit, which can be permanently supplied with charging current via the control unit.
- Steady yellow Charging (battery not ready)
- No LED light Fully charged (battery ready)
- Yellow flashing Error during charging
LED operating display control unit
-
Orange
Over battery and operation System is working properly -
No LED
Over battery but not in operation, not connected to mains power
Buzzer function
The buzzer sounds a warning when a but-ton on the hand control is pressed and
the battery capacity is low.
The buzzer can also be activated via the control unit to signal other states.
This must be specified in the software of the control unit.
The battery is permanently connected and is supplied with charging current via the control unit.
Attention
Make sure that the charging cable is outside the area of use.
Attention
Before first use, the batteries must be charged for 24 hours to ensure proper
function and extended life.
Non-liability
The person carrying out the works is solei responsible for proper and
professional mounting of the Lift owlet. The manufacturer accepts no liability
for the carried out works or any damages resulting from this.
Commissioning
The following steps are required for restarting the system when the system is
started up for the first time and in the event of a fault.
Initial assembly / initialization of the drives:
- Press and hold the DOWN button until the actuators move.
- Now press the button again until the actuators have reached the lower end position.
- Then press the UP button until the lifting columns have moved up approx. 5 cm.
- Then press the DOWN button again until the lifting columns have reached the lower end position again and the control unit switches off both drives.
- The sensors of the drives are now calibrated.
- The system is now initialized and can be put into operation.
Attention,
To avoid an unintended function, it is pointed out that several keys must not
be pressed simultaneously.
Attention
If the system cannot be put into operation after initialization or after a
fatal error reset, contact the manufacturer.
Attention
Batteries are excluded from the warranty!
Battery maintenance
The batteries must be replaced after four years at the latest, depending on the intensity of use, possibly even earlier. Frequent and rapid discharging reduces the life of the batteries. To achieve an optimal life span, the batteries should be connected to the mains as often as possible. The batteries must be recharged every three months at the latest, otherwise they will be damaged due to self-discharge.
The battery is only to be charged in dry conditions.
When in use never allow the battery to discharge completely (observe charging indicator)! Ensure the battery is recharged (overnight) when in use on a daily basis. The battery cannot be over-charged as the battery charger switches automatically as soon as the battery is fully charged.
How it works:
There are two buttons on the hand control that move the toilet seat up and
down together with the support folding handles.
Sit down:
To sit down, the toilet seat is first raised until you can sit down on the
toilet seat sur-face from a standing position. By pressing a button on the
hand control, the seat and the support handles then move down simul-timeously
to the normal toilet seat height. The user slowly takes up his normal sitting
position without any effort.
Stand up:
It is just as comfortable to stand up as it is to sit down. At the touch of a
button, the seat and support arms are lifted upwards until the user is
straightened up again.
Support arms:
The support arms can be folded upwards.
Attention:
To prevent unintended operation for example by children, the support arms
should be folded upwardly when not in use.
Attention danger of clamping!
Attention:
Please also observe the safety instructions for the power supply.
Attention:
In doing so, ensure the feet always stay firmly on the ground.
Accessories / options
For a complete overview of our extensive range of accessories for the Liftolet-Tilty see item 11.2 Types and accessories on page 19.
Types and accessories
Basic model
- Article number description
- LT-1000-T toilet seat lift Liftolet-Tilty
- LT-1000-TD toilet seat lift Liftolet-Tilty Duo
Accessories / options
Liftolet-Tilty can optionally be equipped with a battery.
The most up-to-date overview of our extensive range of accessories for Liftolet-Tilty and their description as well as much more useful information can be found on our homepage www.ato-form.com.
For special problems, or if it seems the accessories you require are not
demonstrated nor described, please contact us.
We may already have the right solution for you.
Technical details
Liftolet-Tilty / Liftolet-Tilty Duo| | LT-1000-T|
LT-1000-TD
---|---|---|---
Overall width| cm| | 54.6
Overall depth| cm| | 64.2
Seat height front| cm| | 45.6
Seat height back| cm| | 48.7
Seat inclination (support plate)| Degrees| | 4°
Inclination of the support plate extended| Degrees| | 25°
Lifting height support plate front| cm| | 53.4
Lifting height support plate back| cm| | 73.3
Width between the support arms| cm| | 45.4
Space between the support arms| cm| | 40.5
Lifting speed| mm/s| approx. 10
Electrical supply| V| | 230
Electric drive with link motor| V| | 24
Protection class
(splash water protection)| | | IPX6
Product weight| kg| 25| 28
Maximum capacity| kg| 85| 155
Technical changes reserved!
Construction
All outside parts are made out of high quality corrosion-resistant materials:
stainless steel, aluminum anodized and plastics. The surface of the equipment
parts which by normal use come in contact with the skin, consist of skin
friendly plastics (WC-seat and support arms) or are coated with pow-der
coating. The metal parts and all screws are out of Stainless steel.
Operating conditions
- Temperature: +5 °C to +40 °C
- Relative humidity: 20% to 90% – noncondensing
- Air pressure: 800 to 1060 hPA (Operation in an estimated height ≤ 2000 m)
Cleaning and disinfection
- In order to ensure lasting trouble-free operation of the Liftolet in almost all operating conditions influenced by the existing conditions on site, please note the following:
- Clean the device regularly with a mild soapy solution. We recommend cleaning the Liftolet after each use.
- When using cleaning agents, always follow the manufacturer’s instructions.
- After cleaning ensure that a light lubricating film is present on the sliding surface of the lifting columns.
- The use of a sprayable lubricant to renew the lubricating film after each cleaning of the lifting columns is recommended.
- Drive the device to the lowest position of the lifting columns once per week.
- To do this, hold down the control button for the DOWN direction until both lifting columns have reached the lower end position and shut off.
- By regularly initializing the drive sensors, a possible position difference can be prevented.
Disinfection:
- For disinfecting, use household disinfectants and consider the following indications:
- The assistant must wear suitable protective clothing (for example gloves) in order to avoid infections.
- When using disinfecting agents always follow the instructions of the manufacturer.
Warning
Do not use strong solvents, which could damage the surface of the device.
The use of a high-pressure cleaner will void the warranty.
Maintenance and overhaul
For the protection of the warranty and safety, it is especially important that
the Toilet seat lift is serviced regularly once a year from the date of
purchase.
It is recommended to make service arrangements.
The manufacture will gladly provide the maintenance check list to your
specialist dealer on request.
Please contact your specialist dealer to carry out corrective maintenance
works for your Liftolet-Tilty.
The device should be checked by the manufacturer at the latest after 3 years.
Maintenance Checklist
Frame
- Metal parts; no signs of damage or wear, especially for adjustable parts
- Welding seams are intact
- Fasteners, bolts, nuts and springs are secure
- Replacement of parts if there is any sign of wear
- No signs of wear of damage of the safety components
Date | Service performed | Liftolet-Tilty | Sign |
---|---|---|---|
Operating life and Re-use
The Liftolet-Tilty is made from high quality materials and is subject to a
continuous quality control in our house.
Because of this and the solid construction, the expected operating life of our
product is up to 5 years, depending on the usage intensity and number of re-
uses, under normal and compliance with its intended use (according to the
instructions given in our user manual).
The expected operating life does not apply to wear parts.
The product Liftolet-Tilty was designed so it could be reused after the application of various measures.
Following measures should be followed:
- Thorough cleaning and disinfecting using appropriate cleaning agents.
- Replacement of wear parts
- Maintenance, following the maintenance schedule of the manufacturer or authorized specialist dealer.
- Preparing accompanying documents and release for re-use.
Faults, repairs, spare parts
If your Toilet seat lift Liftolet-Tilty cannot be operated?
Please ensure…..
- The plug is plugged in the socket and is supplied with power.
- All plug connections are plugged in correctly and secured.
Try and shut down the fault.
Failing that, contact your specialist dealer / reseller.
Only original accessories and original spare parts may be used. Non-compliance with these instructions and repairs carried out or modifications made to the Liftolet-Tilty or its parts by unauthorized persons, voids any warranties.
In these cases, any product liability granted by the manufacturer (ATO FORM GmbH, Zur Lauterhecke 34, 63877 Sailauf, Germany) is voided in advance.
For all questions about spare parts and accessories contact your specialist dealer or the ATO FORM-Service +49 6093 944-0.
Spare parts and accessories must not be removed or replaced while the patient is using the Toilet seat lift!
Storage
Keep this operating manual in a safe place, where you or care and maintenance
staff have access to it.
The current version of these instructions for use can be found on our homepage in the download area.
You can find more information about the product on our homepage at www.ato- form.com or under this QR code.
Disposal of battery operated devices
The battery pack must not be disposed of together with normal household waste
at the end of its service life. Please observe the legal requirements in your
area. Your municipal government and locally based waste disposal companies
will be happy to provide you with information in this regard.
In environmental protection matters:
All packaging parts are re-useable. The cardboard packaging is to be recycled
as waste paper. We therefore use paper as filler material. Possible used
plastic film of the packaging are made out of PE and are therefore low in
pollutant and can either be recycled or burnt.
Complete Range:
- Head and body protection
- Walking aids
- Gait trainer
- Standing aids
- Seating aids
- Hoists
- Customized solutions
Thank you for your interest in our products.
For questions please contact ATO FORM GmbH or your specialist dealer.
Your ATO FORM Team
ATO FORM GmbH
Zur Lauterhecke 34, 63877 Sailauf, Germany
Tel. +49 6093 944-0,
Fax +49 6093 944-444
E-Mail: info@ato-form.com
www.ato-form.com
Your specialist dealer
References
- Kopfschutz, Lifter, Gehtrainer, Kinderreha - Rehahilfsmittel von ATO FORM
- FORM | The Digital Assistant for the Frontline | Mobile Forms App
- Kopfschutz, Lifter, Gehtrainer, Kinderreha - Rehahilfsmittel von ATO FORM
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>