Drucker Diagnostics Dash Series Apex 6 Centrifuge User Manual
- June 9, 2024
- Drucker Diagnostics
Table of Contents
- Drucker Diagnostics Dash Series Apex 6 Centrifuge
- Instructions
- MODEL DESCRIPTION
- FEATURES
- CAUTION AND WARNING STATEMENTS
- INITIAL SETUP
- OPERATION
- REVIEW CYCLE TIME AND SPEED SETTINGS
- CHANGING CYCLE TIME AND SPEED SETTINGS
- BALANCING LOADS
- CARE AND PREVENTATIVE MAINTENANCE
- CLEANING AND DISINFECTION
- TROUBLESHOOTING
- GENERAL SPECIFICATIONS
- CALCULATING THE G-FORCE
- REPLACEMENT PARTS
- 6,811,531, #D718,463, & #D734,489. Other Patents Pending
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Drucker Diagnostics Dash Series Apex 6 Centrifuge
Instructions
SYMBOLS
MODEL DESCRIPTION
DASH Apex is engineered for STAT sample processing. When used with the Drucker DASH Approach to centrifugation, the DASH Apex cuts your turnaround time (TAT) by up to 20 minutes. This general-purpose laboratory centrifuge may also be used to spin approved containers with biologics, chemicals (non-flammable, non-explosive, non-volatile, and non-highly reactive), and environmental samples.
FEATURES
- ATURES
- Simple 2-Button interface
- Three (3) easily selectable pre-set cycles are conveniently labeled for your lab’s most common applications. Use the default cycles or settings can be customized. An LED light indicates the current selected setting.
- If desired, the control panel can be locked on one preset cycle, ideal for standardization to a single spin.
- Lid lighting indicates the centrifuge’s status (ready, running, done), keeping your TAT down (patent pending).
- A traditional audible alert indicates the completion of the cycle.
- Cool–Flow air flow design prevents overheating of samples by maintaining room temperature.
- Carbon fibers are used to reinforce the tube holders and provide high strength and durability.
- A clear lid permits safe observation of samples and optical calibration of speed.
- The lid safety system only allows entry into the centrifuge after the rotor has completely stopped.
- The lid safety system prevents the centrifuge from operating unless the lid is closed and latched.
- The high-power brushless DC motor provides years of operation with no routine maintenance.
INTENDED USE
General purpose laboratory centrifuge, intended for the density-based separation of fluids through centripetal acceleration.
WARRANTY
Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years.
CAUTION AND WARNING STATEMENTS
- This device is intended to be operated by properly trained personnel who have carefully read the operating manual and are familiar with the function of the device. [Refer to the clinical laboratory method specified by the specimen receptacle manufacturer or established by the medical technology for the product’s applications.]
- For the safety of both the operator and service personnel, care should be taken when using this centrifuge if handling substances that are known to be toxic, radioactive or contaminated with pathogenic microorganisms. Use appropriate personal protection equipment (PPE). When Risk Group II materials are used, (as identified in the World Health Organization “Laboratory Bio-Safety Manual”), a Bio- Seal should be employed. In the event that materials of a higher risk group are being used, more than one level of protection must be provided. The use of flammable or explosive materials as well as those materials which have a vigorous chemical reaction is prohibited.
- Unplug the centrifuge before cleaning or performing maintenance.
- Inspect centrifuge for cracks or physical damage to cabinet, lid, rotor, or tube holders. Damage may result in unsafe operation. Discontinue use until repairs have been performed.
- This equipment generates, uses, and can radiate radio frequency energy, and if not installed and used in accordance with this operator manual, may cause interference to radio communications.
- Operation of this equipment in a residential area may cause interference, in which case the user will be required to correct the interference at his own expense.
- Due to the lack of the possibility of human exposure, all Drucker centrifuges and accessories sold by Drucker Diagnostics, Inc. are compliant without any special labeling required by the California Safe Drinking Water and Toxic Enforcement Act (Proposition 65).
INITIAL SETUP
- Unpack and verify that all the following are included:
- Centrifuge
- Power cord
- Tube holders: 6 for Apex 6, 12 for Apex 12, or 6 buckets for Apex 24
- Quick Start Guide
- Setup the centrifuge on flat and level surface. A bench top clearance height of 21” (54 cm) is required to open the lid.
- The centrifuge should have 6” (15 cm) of clear space around the centrifuge. Proper ventilation is necessary to prevent the overheating of samples as well as premature failure of the centrifuge. Choose an area which allows unencumbered air flow, and where the temperature remains between 16°C and 32°C.
- No hazardous material shall be permitted in the clearance envelope during operation.
- The operator time within the envelope shall be limited to the time necessary for loading, unloading, and centrifuge operation only.
- Plug the line cord into the centrifuge.
- Plug the line cord into an electrical outlet.
- Turn on the power switch on the back of the centrifuge
BE SURE THE ELECTRICAL OUTLET IS ALWAYS ACCESSIBLE AS THE LINE CORD IS THE MEANS OF EMERGENCY DISCONNECTION!
OPERATION
-
Place the tubes into the tube holders. Be sure to follow the rules for balanced loads as listed on page 8.
-
The front panel LED is illuminated for the currently selected cycle. The selected cycle determines the run time and speed. To select another cycle, press the UNLOCK button in succession until the desired cycle is selected.
Note: cycle selection is only available with the lid open. -
Close the lid and turn the lid knob clockwise to its complete stop position.
-
Pushing the START button on the control panel starts the spin cycle.
-
When the cycle is completed, the rotor will slow to a complete stop and the lid light will flash.
-
The locking mechanism will disengage for 60 seconds allowing entry into the rotor chamber. To unlock after more than 60 seconds have elapsed, press the UNLOCK button. The lid will unlock for another 15 seconds.
-
Turn the lid knob counterclockwise and open the lid. The lid light will turn off.
-
You may now safely remove the samples.
The LED indicator light is on for the cycle currently selected. Depending on your unit, you may have the following cycles:
- BarricorTM: For BD Vacutainer® BarricorTM tubes
- Coag: For Coagulation or Platelet Poor Plasma(PPP)
- PST/SSTTM: For BD PSTTM & SSTTM tubes with gel or blood tubes without gel
- Chemistry: For BD PSTTM II & SSTTM II tubes with gel or blood tubes without gel
- 3 Min: STAT centrifugation at 4,000 xg
- 5 Min: STAT centrifugation at 4,000 xg (6 and 12) or 3,000 xg (24)
- 6 Min: STAT centrifugation at 3,000 xg
- Custom: This setting can be customized to your lab’s validated cycles.
SETTINGS
Factory Settings | DASH Apex 6 | DASH Apex 12 | DASH Apex 24 |
---|---|---|---|
Cycle | RPM | Time | G-Force |
3 min | 5,300 | 3 | 4,000 |
5 min | 5,300 | 5 | 4,000 |
Custom | 3,800 | 7 | 2,000 |
Factory Settings | DASH Apex 6 | DASH Apex 12 | DASH Apex 24 |
---|---|---|---|
Cycle | RPM | Time | G-Force |
BarricorTM | 5,300 | 3 | 4,000 |
Coag | 5,300 | 5 | 4,000 |
PST/SSTTM | 3,800 | 7 | 2,000 |
Factory Settings | DASH Apex 6 | DASH Apex 12 | DASH Apex 24 |
---|---|---|---|
Cycle | RPM | Time | G-Force |
BarricorTM | 5,300 | 3 | 4,000 |
Coag | 5,300 | 5 | 4,000 |
Chemistry
(PSTII/SSTIITM)
| 4,600| 5| 3,000| 4,400| 5| 3,000| 4,300| 5| 3,000
Custom cycles, if desired:
Cycle/ Tube Type/ RPM/ G-Force (RCF)
- 1
- 2
- 3
NOTE: Timer starts when speed reaches 90% of set speed. Deceleration time is not included in cycle time.
REVIEW CYCLE TIME AND SPEED SETTINGS
Your settings may not be standard. To review current settings, follow this procedure:
- The lid must be opened to review the selected cycle time and speed.
- Press and hold the START button until you hear a beep.
- Release the START button, the centrifuge will beep and the LED light will flash once for each minute of run time in the current cycle.
- Pressing the START button again will cause the unit to beep and the LED light to flash once for each 100 rpm in the current cycle.
- The centrifuge will automatically revert to normal mode at the end.
CHANGING CYCLE TIME AND SPEED SETTINGS
- The lid must be opened to change the selected cycle time and speed.
- Select the cycle you wish to change.
- Press and hold the START and UNLOCK buttons together until the LED light flashes.
- Press the START button for each minute of run time.
- Move to speed setting mode by pressing the UNLOCK button.
- Press the START button once for each 100 rpm.
- Press the UNLOCK button to exit setting mode.
BALANCING LOADS
Your centrifuge must contain a balanced load to work properly. Spinning balanced loads will extend the life of the centrifuge and produce betterresults. Use the following rules when loading the rotor. If an odd number of samples is to be spun, fill a tube with water to match the weight of the unpaired sample and place it across from this sample.
Opposing tube holders must be equally loaded or empty or loaded with equally weighted samples. When loading only 3 tubes, they must be of equal weight.
Buckets can be placed around the rotor in any of the rotor loading configurations shown. Each bucket must be loaded symmetrically with tubes as above.
CARE AND PREVENTATIVE MAINTENANCE
With proper care and maintenance, your centrifuge will provide years of laboratory service. For proper care, the following steps should be taken:
- Always Spin Balanced Loads: Make certain that you are always spinning a balanced load, as shown in the previous section. These centrifuges have a unique counter balanced motor mounting design which produces excellent vibration dampening. However, out–of–balance loads may break glass test tubes and may produce unsatisfactory separation results. Proper load balancing will improve sample separation and extend the life of the centrifuge.
- Motor and Electrical Maintenance: The highest quality electrical components have been selected for the DASH Apex centrifuges and should not need maintenance or servicing for the life of the centrifuge.
- Tube Holder Replacement: It is recommended that the tube holders be replaced after 24 months of use. Inspect tube holders regularly for cracks. If cracks are discovered, replace immediately.
- Remove Accessories Before Moving: All tube holders, samples, and caps must be removed from the rotor chamber before transporting orstoring the centrifuge to prevent damage and injury.
CLEANING AND DISINFECTION
To prolong the life of the centrifuge, cleaning and disinfection isrecommended every six months or whenever there is a spillage or tube breakage. Contaminants must be removed immediately or corrosion and premature degradation of components can occur. Before using any cleaning or decontamination methods other than those recommended by the manufacturer, users should verify with the manufacturer that the proposed method will not damage the equipment.
-
Unplug the centrifuge before cleaning.
-
Use appropriate personal protective equipment (PPE).
-
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in water or other cleaning solutions as this will cause damage and void the warranty.
-
ONLY isopropyl alcohol or a 10% (5500 PPM) bleach solution should be used to disinfect the centrifuge and its accessories.
-
All surfaces must be dried immediately after cleaning and disinfecting.
TBQ GERMICIDAL PRODUCTS ARE NOT RECOMMENDED AS THEY MAY CAUSE DAMAGE TO THE CENTRIFUGE. REFRAIN FROM USING TO PREVENT VOIDING THE WARRANTY. -
Fully/partially halogenated hydrocarbons, ketones, esters, ethers, benzyls, ethyl benzenes, and all other chemicals not prescribed by the manufacturer shall not be used as they may cause damage to the rotor chamber, rotor, tube holders, accessories and centrifuge exterior and void the warranty.
TROUBLESHOOTING
NOTE: The latch must be turned completely clockwise to its stop position for the centrifuge to operate.
The centrifuge does not run| ****
o Verify that the centrifuge is powered. One of the LED lights should be on.
o Make sure the lid latch is turned completely clockwise to its stop position.
o If the centrifuge still does not run, contact Customer Service.
---|---
The rotor does not spin freely| ****
o Make sure nothing has fallen into the rotor chamber, following the procedure above.
o If nothing obstructs the rotor, the rotor may be damaged. Contact Customer Service for further assistance.
The centrifuge makes a rattling noise when running| o Stop the centrifuge. Open the lid.
o Wearing PPE, remove tubes and tube holders/buckets and look for fallen objects or debris. Carefully reach inside the rotor chamber with a tool to remove them.
o Inspect the rotor, tube holders or buckets for damage.
o If the tube holders or buckets have any damage, even slight, safely dispose of them and replace them.
o If the rotor appears damaged, contact Customer Service for further assistance.
Excessive noise or vibration when the centrifuge is running| ****
o Verify that all four centrifuge feet are properly seated on a flat surface.
o Ensure that the load is balanced according to instructions in the “Balancing Loads” section of this manual.
o Make sure that nothing has fallen into the rotor chamber.
The centrifuge continuously| stops| and| beeps| ****
The load is not balanced. Press the UNLOCK button, open the lid, and balance the load as recommended elsewhere in this manual.
The cycle time and speed are not set to the desired value| ****
Check the setting by following the instructions in the section on Changing Cycle Settings. If the preset is not the desired length, follow the procedure on the same page to change the run preset time.
The centrifuge does not unlock after a run is completed| o Wait until the rotor has come to a complete stop. If the lid knob still cannot be rotated, press the UNLOCK button and try again.
o If no LED light is on, the unit is not powered and the lid will not unlock by conventional means. Remove the latch label and use a pen to manually disengage the locking mechanism. Pull the mechanism towards the control panel and then unlatch and open the lid.
o If the unit is damaged, contact your authorized dealer or Drucker Diagnostics.
---|---
The lid does not open| o Ensure that the lid knob is turned fully
counterclockwise.
o If the knob cannot be turned counterclockwise, turn it fully clockwise, press UNLOCK, and turn counterclockwise.
o If the lid remains locked after this and will not unlock, the electronics may have been damaged. Contact customer service for assistance.
Clicking noise during braking gets loud| o Make sure that the screw in
the center of the rotor is tight.
Lid does not stay up| o Tighten the center screw on the lid hinge.
GENERAL SPECIFICATIONS
The rotor and accessories are rated for the maximum rotation frequency shown in the table below.
Apex 6 | Tube Capacity | 6 tubes – 3 to 10 mL |
---|---|---|
Radius with included accessories | 5 in (12.7 cm) | |
Dimensions (Height x Width x Depth) | 8 in x 11 in x 13 in (20 cm x 28 cm x 34 |
cm)
Weight| 12 lbs. (5.4 kg)
Noise Level| 61 dB
Environmental Range| 16 – 32 0C
Voltage| 100 – 240 VAC
Frequency| 50/60 Hz
Power Requirement| 225 Watts
Centrifuge Motor| ½ H.P. Brushless DC
Maximum Speed| 5,300 RPM
Cycle Time| 1 to 30 minutes (+/- 2%)
Apex 12| Tube Capacity| 12 tubes – 3 to 10 mL
---|---|---
Radius with included accessories| 5.25 in (13.3 cm)
Dimensions (Height x Width x Depth)| 9 in x 12.5 in x 14.75 in (23 cm x 32 cm
x 37 cm)
Weight| 34 lbs. (15 kg)
Noise Level| 61 dB
Environmental Range| 16 – 32 0C
Voltage| 100 – 240 VAC
Frequency| 50/60 Hz
Power Requirement| 415 Watts
Centrifuge Motor| ½ H.P. Brushless DC
Maximum Speed| 5,200 RPM
Cycle Time| 1 to 30 minutes (+/- 2%)
Apex 24| Tube Capacity| 24 tubes – 3 to 10 mL
---|---|---
Radius with included accessories| 6 in (15.3 cm)
Dimensions (Height x Width x Depth)| 9 in x 14.5 in x 17 in (23 cm x 37 cm x
43 cm)
Weight| 39 lbs. (17 kg)
Noise Level| 64 dB
Environmental Range| 16 – 32 0C
Voltage| 95 – 253 VAC
Frequency| 50/60 Hz
Power Requirement| 220 Watts
Centrifuge Motor| ½ H.P. Brushless DC
Maximum Speed| 4,300 RPM
Cycle Time| 1 to 30 minutes (+/- 2%)
Use only with approved accessories from the original manufacturer. A complete list of accessories is available at www.DruckerDiagnostics.com.
CALCULATING THE G-FORCE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force. For included accessories, a conversion table is shown below.
For other tube holders, the G-Force can be calculated if you know the RPM and the radius:
- In Centimeters: RCF or G-force = 0.00001118 x Rotor Radius (cm) x (RPM)2
| APEX 6| APEX 12| APEX 24
---|---|---|---
| Radius| Radius| Radius
| 5.0 in| 5.25 in| 6.0 in
| 12.7 cm| 13.3 cm| 15.3 com
RPM| G-Force| G-Force| G-Force
1000| 140| 150| 170
1100| 170| 180| 210
1200| 200| 210| 250
1300| 240| 250| 290
1400| 280| 290| 330
1500| 320| 340| 380
1600| 360| 380| 450
1700| 400| 450| 500
1800| 450| 500| 550
1900| 500| 550| 600
2000| 600| 600| 700
2100| 650| 650| 750
2200| 700| 700| 800
2300| 750| 800| 900
2400| 800| 850| 1000
2500| 900| 900| 1050
2600| 950| 1000| 1150
2700| 1050| 1100| 1250
2800| 1100| 1200| 1350
2900| 1200| 1250| 1400
3000| 1300| 1350| 1500
3100| 1350| 1400| 1650
3200| 1450| 1500| 1750 - In Inches: RCF or G-force = 0.0000284 x Rotor Radius (in) x (RPM)2
| APEX 6| APEX 12| APEX 24
---|---|---|---
| Radius| Radius| Radius
| 5.0 in| 5.25 in| 6.0 in
| 12.7 cm| 13.3 cm| 15.3 com
RPM| G-Force| G-Force| G-Force
3300| 1550| 1600| 1850
3400| 1650| 1700| 2000
3500| 1750| 1800| 2100
3600| 1850| 1950| 2200
3700| 1950| 2050| 2300
3800| 2050| 2150| 2450
3900| 2150| 2300| 2600
4000| 2300| 2400| 2850
4100| 2400| 2500| 2900
4200| 2500| 2600| 3000
4300| 2600| 2750| 3150
4400| 2750| 2900| N/A
4500| 2900| 3000| N/A
4600| 3000| 3150| N/A
4700| 3150| 3300| N/A
4800| 3300| 3450| N/A
4900| 3400| 3600| N/A
5000| 3550| 3750| N/A
5100| 3700| 3900| N/A
5200| 3850| 4000| N/A
5300| 4000| N/A| N/A
REPLACEMENT PARTS
DASH APEX 6
Part No. Description
-
7724037K Foot, rubber
-
02-002-1-0028 Lid Tray Assembly
-
02-001-0-0011 Rotor Assembly, Dash Apex 6
-
02-005-1-0010 Motor Assembly
-
02-006-0-0011 PC Board
-
7760006 Power cord
-
03-1-0005-0192 Internal Power Supply
-
02-002-1-0027 Lid Assembly
-
7724071K Hinge, friction
-
02-002-1-0056 Seal, lid gasket
-
03-0-0003-0313 Open/Close Label
-
03-1-0007-0046K 75/100mm Tube Holder, Black
-
02-002-1-0026 Lid LED Assembly, Green
-
00-100-100-009 6 Series Soft Button Replacement Kit
DASH APEX 12 -
7724177K Foot, rubber
-
02-006-1-0044 Locking Lid Tray Assembly, 11”
-
02-001-0-0009 12-Place Rotor, Horizontal
-
02-005-1-0012 Motor Assembly
-
02-006-0-0011 PC Board
-
7760006 Power cord
-
03-1-0005-0193 Power Supply
-
02-002-1-0041 Lid Assembly
-
7724071K Hinge, friction
-
02-002-1-0057 Seal, lid gasket
-
03-0-0003-0313 Open/Close Label
-
03-1-0007-0046K 75/100mm Tube Holder, Black
-
00-100-100-010 12/24 Series Soft Button & Spacer Replacement Kit
-
00-100-100-005 Replacement Grommet & Bushing Kit
DASH APEX 24 -
7728052K Foot, rubber
-
02-006-1-0044 Locking Lid Tray Assembly, 11”
-
02-001-0-0008 24-Place Rotor, Horizontal
-
02-005-1-0012 Motor Assembly
-
02-006-0-0011 PC Board
-
7760006 Power cord
-
03-1-0005-0193 Power Supply
-
02-002-1-0037 Lid Assembly
-
7724071K Hinge, friction
-
02-002-1-0058 Seal, lid gasket
-
03-0-0003-0313 Open/Close Label
-
02-004-0-0012 4 Place Carrier, Carbon Fiber
-
00-100-100-010 12/24 Series-Soft Button & Spacer Replacement Kit
-
00-100-100-005 Replacement Grommet & Bushing Kit
Product Family: DASH Apex Series (Apex 6, Apex 12, Apex 24) Complies with UL61010-1/CSA C22.2 No. 61010-1 and IEC61010-2-020 Protected by U.S. Patents
6,811,531, #D718,463, & #D734,489. Other Patents Pending
To access previous manuals, please click the relevant link below: DASH Apex 6, 12, 24 Operator’s Manual, Rev G
INSTRUCTIONS FOR DISPOSAL OF WEEE BY USERS IN THE EUROPEAN UNION This product must not be disposed of with other waste. Instead, it is the user’s responsibility to dispose of their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment. The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment. For more information about where you can drop off your waste equipment for recycling, please contact your local city office, waste disposal service, or where you purchased the product.
200 SHADY LANE, SUITE 170 – PHILIPSBURG, PA 16866, USA +1-866-265-1486 (U.S. ONLY) – +1-814-692-7661
CUSTOMERSERVICE@DRUCKERDIAGNOSTICS.COM DRUCKERDIAGNOSTICS.COM
References
- Clinical Centrifuges & Hematology Instruments | Drucker Diagnostics
- Clinical Centrifuges & Hematology Instruments | Drucker Diagnostics
- Clinical Centrifuges & Hematology Instruments | Drucker Diagnostics
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>