Drucker Diagnostics DASH Coag 2-3-5 Minute STAT Coag Centrifuge User Manual
- June 9, 2024
- Drucker Diagnostics
Table of Contents
- Symbols
- MODEL DESCRIPTION
- FEATURES
- INTENDED USE
- CAUTION AND WARNING STATEMENTS
- INITIAL SETUP
- OPERATION
- CHANGING CYCLE TIME AND SPEED SETTINGS
- BALANCING LOADS
- CARE AND PREVENTATIVE MAINTENANCE
- CLEANING AND DISINFECTION
- TROUBLESHOOTING
- GENERAL SPECIFICATIONS
- REPLACEMENT PARTS
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
DASH Cog 2-3-5 Minute STAT Cog Centrifuge
User Manual
DASH COAG
Operator’s Manual
Symbols
Symbol | Definition | Use |
---|---|---|
**** | Caution | Caution to safety hazard. |
Potential risk of personal injury or damage to the instrument if improperly
handled. Consult the manual before proceeding.
| Manufacturer| Manufacturer of record.
| Electrical and electronic products recycling symbol| Recycle only as
electronic waste. Do not dispose in normal waste.
RoHS Compliant
| RoHS Compliant| Compliance with RoHS environmental standards.
| CE Mark| Denotes conformity to specific European directives and regulations.
| MET Listing| Denotes conformity to specific safety standards and
regulations.
FDA LISTED
| FDA Listed| Denotes that the product has been properly listed with the FDA.
ISO13485 certified
| ISO Certification| Denotes conformity to quality standards and quality management systems.
MODEL DESCRIPTION
The DASH Cog centrifuge is engineered to reduce turnaround time (TAT) and
simplify cog processing in the STAT laboratory. Produce error-free samples in
as little as 3 minutes with the DASH Cog’s simple set-and-lock controls,
preset cycle settings, and an LED lid lighting indicator system.
This general-purpose laboratory centrifuge may also be used to spin approved
containers with biologics, chemicals (non-flammable, non-explosive, non-
volatile, and non-highly reactive), and environmental samples.
FEATURES
- Simple 2-Button interface
- Three (3) easily selectable pre-set cycles are conveniently labeled for your lab’s most common applications. Use the default cycles or customize them as needed. An LED light indicates the current selected setting.
- If desired, the control panel can be locked on one preset cycle; ideal for standardization to a single spin.
- Lid lighting indicates the centrifuge’s status (ready, running, done), keeping your TAT down (patent pending).
- A traditional audible alert indicates the completion of the cycle.
- Cool–Flow air flow design prevents overheating of samples by maintaining room temperature.
- Carbon fibers are used to reinforce the tube holders and provide high strength and durability.
- A clear lid permits safe observation of samples and optical calibration of speed.
- The lid safety system only allows entry into the centrifuge after the rotor has completely stopped.
- The lid safety system prevents the centrifuge from operating unless the lid is closed and latched.
- The high-power brushless DC motor provides years of operation with no routine maintenance.
INTENDED USE
General purpose laboratory centrifuge, intended for the density-based separation of fluids through centripetal acceleration.
WARRANTY
Drucker Diagnostics warrants that this centrifuge is free from defects in
workmanship and parts for 2 years.
CAUTION AND WARNING STATEMENTS
This device is intended to be operated by properly trained personnel who have
carefully read the operating manual and are familiar with the function of the
device. Refer to the clinical laboratory method specified by the specimen
receptacle manufacturer or established by the medical technology for the
products applications.
For the safety of both the operator and service personnel, care should be
taken when using this centrifuge if handling substances that are known to be
toxic, radioactive or contaminated with pathogenic microorganisms. Use
appropriate personal protection equipment (PPE). When Risk Group II materials
are used, (as identified in the World Health Organization “Laboratory Bio-
Safety Manual”), a Bio- Seal should be employed. In the event that materials
of a higher risk group are being used, more than one level of protection must
be provided. The use of flammable or explosive materials as well as those
materials which have a vigorous chemical reaction is prohibited.
Unplug the centrifuge before cleaning or performing maintenance.
Inspect centrifuge for cracks or physical damage to cabinet, lid, rotor, or
tube holders. Damage may result in unsafe operation. Discontinue use until
repairs have been performed.
This equipment generates, uses, and can radiate radio frequency energy, and if
not installed and used in accordance with this operator manual, may cause
interference to radio communications.
Operation of this equipment in a residential area may cause interference, in
which case the user will be required to correct the interference at his own
expense.
Due to the lack of the possibility of human exposure, all Drucker centrifuges
and accessories sold by Drucker Diagnostics, Inc. are compliant without any
special labeling required by the California Safe Drinking Water and Toxic
Enforcement Act (Proposition 65).
INITIAL SETUP
-
Unpack and verify that all the following are included:
• Centrifuge
• Power cord
• Quick Start Insert Guide -
Setup the centrifuge on flat and level surface. A bench top clearance height of 21” (54 cm) is required to open the lid.
-
The centrifuge should have 6” (15 cm) of clear space around the centrifuge. Proper ventilation is necessary to prevent the overheating of samples as well as premature failure of the centrifuge. Choose an area which allows unencumbered air flow, and where the temperature remains between 16°C and 32°C.
-
No hazardous material shall be permitted in the clearance envelope during operation.
-
The operator time within the envelope shall be limited to the time necessary for loading, unloading, and centrifuge operation only.
-
Plug the line cord into an approved electrical outlet.
• BE SURE THE ELECTRICAL OUTLET IS ALWAYS ACCESSIBLE AS THE LINE CORD IS THE MEANS OF EMERGENCY DISCONNECTION!
OPERATION
- Place the tubes into the tube holders. Be sure to follow the rules for balanced loads as listed on page 7.
- Close the lid and turn the lid knob clockwise to its complete stop position.
- The LED light is on for the cycle currently selected. The selected cycle determines the run time and speed. To change the selected cycle, press the UNLOCK button in rapid succession until the desired cycle is selected. Two seconds after selection, the button reverts to its UNLOCK function.
- Pushing the START button on the control panel starts the spin cycle.
- When the cycle is completed, the rotor will slow to a complete stop and the lid light will flash.
- The locking mechanism will disengage for 60 seconds allowing entry into the rotor chamber. To unlock after more than 60 seconds have elapsed, press the UNLOCK button. The lid will unlock for another 15 seconds.
- Turn the lid knob counterclockwise and open the lid. The lid light will turn off.
- You may now safely remove the samples.
QUICK START
The LED indicator light is on for the cycle currently selected:
2 min | This setting is factory preset for 2 minutes at 6,600 RPM/4,400 xg |
---|---|
3 min | This setting is factory preset for 3 minutes at 6,600 RPM/4,400 xg |
5 min | This setting is factory preset for 5 minutes at 5,200 RPM/2,700 xg |
| Start| Begins running the cycle indicated by the cycle indicator LED
light. The lid must be closed.
---|---|---
| Unlock| Allows access into the rotor chamber by disengaging the
locking mechanism. Entry is only possible when the rotor is stopped.
| Stop| Pressing the UNLOCK button during operation will terminate the
run and unlock the lid after the rotor has come to a stop.
| Cycle Selection| The LED light is on for the cycle currently selected.
To change the selected cycle, press the UNLOCK button in succession until the
desired cycle is selected. Two seconds after selection, the button reverts to
its UNLOCK function.
---|---|---
| Lock Cycle Selection| Select desired cycle. Press and hold the UNLOCK
button for 5 seconds. Two beeps will confirm that cycle selection is locked.
| Unlock Cycle Selection| To re-enable cycle selection, press and hold
the UNLOCK button for 5 seconds. Three beeps will confirm that cycle selection
is now unlocked.
SETTINGS
| Setting 1| Setting 2| Setting 3
---|---|---|---
RPM| 6,600| 6,600| 5,200
Time| 2 min| 3 min| 5 min
G-Force| 4,400 xg| 4,400 xg| 2,700 xg
**Brake Setting***| 5| 5| 5
*Braking cannot be adjusted on the Dash Cog.
REVIEW CYCLE TIME AND SPEED SETTINGS
Factory programmed cycles are shown on the rear of the centrifuge, on the
Factory Set Cycles label. To confirm your centrifuge’s current settings,
follow this procedure:
o Open the lid. The lid must remain open until cycle confirmation is complete.
o Press and hold the START button until you hear a beep.
o Release the START button. The centrifuge will beep and the LED light will
flash once for each minute of run time in the current cycle. 10 beeps /
flashes equal 10 minutes of run time. Run time starts when the rotor reaches
90% of desired speed and stops when the rotor starts decelerating.
o Pressing the START button again will cause the LED light to flash once for
each 100 revolutions per minute (RPM) in the current cycle. 10 beeps / flashes
equal 10 x 100 or 1,000 RPM.
o The centrifuge will automatically revert to normal mode at the end.
CHANGING CYCLE TIME AND SPEED SETTINGS
o Open the lid. The lid must be open throughout setting adjustment.
o Select the cycle you wish to change.
o Press and hold the START and UNLOCK buttons together until the LED light
flashes.
o Press the START button for each minute of run time.
o Move to speed setting mode by pressing the UNLOCK button.
o Press the START button once for each 100 rpm.
o Press the UNLOCK button to exit setting mode.
BALANCING LOADS
Your centrifuge must contain a balanced load to work properly. Spinning
balanced loads will extend the life of the centrifuge and produce better
results. Use the following rules when loading the rotor. If an odd number of
samples is to be spun, fill a tube with water to match the weight of the
unpaired sample and place it across from this sample.
Opposing tube holders must be equally loaded or empty or loaded with equally
weighted samples.
When loading only 3 tubes, they must be of equal weight.
CARE AND PREVENTATIVE MAINTENANCE
With proper care and maintenance, your centrifuge will provide years of
laboratory service. For proper care, the following steps should be taken:
o Always Spin Balanced Loads: Make certain that you are always spinning a
balanced load, as shown in the previous section. These centrifuges have a
unique counter balanced motor mounting design which produces excellent
vibration dampening. However, out–of–balance loads may break glass test tubes
and may produce unsatisfactory separation results. Proper load balancing will
improve sample separation and extend the life of the centrifuge.
o Motor and Electrical Maintenance: The highest quality electrical
components have been selected for the DASH Apex centrifuges and should not
need maintenance or servicing for the life of the centrifuge.
o Remove Accessories Before Moving: All samples must be removed from the
rotor chamber before transporting or storing the centrifuge to prevent damage
and injury.
CLEANING AND DISINFECTION
To prolong the life of the centrifuge, cleaning and disinfection is recommended every six months or whenever there is a spillage or tube breakage. Contaminants must be removed immediately or corrosion and premature degradation of components can occur. Before using any cleaning or decontamination methods other than those recommended by the manufacturer, users should verify with the manufacturer that the proposed method will not damage the equipment.
-
Unplug the centrifuge before cleaning.
-
Use appropriate personal protective equipment (PPE).
-
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in water or other cleaning solutions as this will cause damage and void the warranty.
-
ONLY isopropyl alcohol or a 10% (5500 PPM) bleach solution should be used to disinfect the centrifuge and its accessories.
-
All surfaces must be dried immediately after cleaning and disinfecting.
TBQ GERMICIDAL PRODUCTS ARE NOT RECOMMENDED AS THEY MAY CAUSE DAMAGE TO THE CENTRIFUGE. REFRAIN FROM USING TO PREVENT VOIDING THE WARRANTY. -
Fully/partially halogenated hydrocarbons, ketones, esters, ethers, benzyls, ethyl benzenes, and all other chemicals not prescribed by the manufacturer shall not be used as they may cause damage to the rotor chamber, rotor, tube holders, accessories and centrifuge exterior and void the warranty.
TROUBLESHOOTING
NOTE: The latch must be turned completely clockwise to its stop position for the centrifuge to operate.
The centrifuge does not run| o Verify that the centrifuge is powered on.
One of the LED lights should be on.
o Make sure the lid latch is turned completely clockwise to its stop position.
o If the centrifuge still does not run, contact Customer Service.
---|---
The rotor does not spin freely| o Make sure nothing has fallen into the
rotor chamber, following the procedure above.
o If nothing obstructs the rotor, the rotor may be damaged. Contact Customer
Service for further assistance.
The centrifuge makes a rattling noise when running| ** o Stop the
centrifuge. Open the lid.
o Wearing PPE, remove tubes and look for fallen objects or debris. Carefully
reach inside the rotor chamber with a tool to remove them.
o Inspect the rotor for damage.
o If the rotor appears damaged, contact Customer Service for further
assistance.
Excessive noise or vibration when the centrifuge is running| ** o
Verify that all four centrifuge feet are properly seated on a flat surface.
o Ensure that the load is balanced according to instructions in the “Balancing
Loads” section of this manual.
o Make sure that nothing has fallen into the rotor chamber.
The centrifuge stops and beeps continuously| ** The load is not
balanced. Press the UNLOCK button, open the lid, and balance the load as
recommended elsewhere in this manual.
The centrifuge does not unlock after a run is completed**| o Wait until
the rotor has come to a complete stop. If the lid knob still cannot be
rotated, press the UNLOCK button and try again.
o If no LED light is on, the unit is not powered and the lid will not unlock
by conventional means. Remove the latch label and use a pen to manually
disengage the locking mechanism. Pull the mechanism towards the control panel
and then unlatch and open the lid.
o If the unit is damaged, contact your authorized dealer or Drucker
Diagnostics.
The cycle time and speed are not set to the desired value| o Check the
setting by following the instructions in the section on Reviewing Cycle
Settings. If the preset is not the desired length, follow the procedure on the
same page to change the run preset time.
---|---
The lid does not open| o Ensure that the lid knob is turned fully
counterclockwise.
o If the knob cannot be turned counterclockwise, turn it fully clockwise,
press UNLOCK, and turn counterclockwise.
o If the lid remains locked after this and will not unlock, the electronics
may have been damaged. Contact customer service for assistance.
Clicking noise during braking gets loud| o Make sure that the screw in
the center of the rotor is tight.
Lid does not stay up| o Tighten the center screw on the lid hinge.
GENERAL SPECIFICATIONS
The rotor and accessories are rated for the maximum rotation frequency shown in the table below.
Tube Capacity | 12 tubes, up to 75 mm long (4 mL) * |
---|---|
Dimensions (W x D x H) | 8 in x 11 in x 13 in (20 cm x 28 cm x 34 cm) |
Weight | 12 lbs (5.4 kg) |
Sound Level | 61 dB A |
Environmental Range | 16 – 32°C |
Voltage | 100 – 240VAC |
Frequency | 50/60 Hz |
Power Requirement | 225 Watts |
Centrifuge Motor | ½ H.P. Brushless DC |
Max G-Force | 4,400 xu |
Max Speed | 6,600 RPM |
Min Cycle Time | 1 minute |
Max Cycle Time | 30 minutes |
Use only with approved accessories from the original manufacturer. A complete
list of accessories is available at
druckerdiagnostics.com.
*Maximum sample density is 1.15 grams/mL (water density = 1.0 grams/mL).
Any use other than those specified by the Manufacturer is explicitly
prohibited.
CALCULATING THE G-FORCE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force, which
can be calculated if you know the RPM and the radius. Use the formula below or
go to www.druckerdiagnostics.com/g-force-
calculator/.
In Centimeters:
RCF or G-force = 0.00001118 x
Rotor Radius (cm) x (RPM)²| In Inches:
RCF or G-force = 0.0000284 x
Rotor Radius (in) x (RPM)²
---|---
Radius 3.5 in (9 cm)
REPLACEMENT PARTS
Part No. | Description |
---|---|
7724037 | Foot, rubber |
02-002-1-002802- 002-1-0073 | Lid Tray Assembly |
02-001-0-0017 | Cog Rotor Assembly |
02-005-1-0014 | Motor Assembly |
02-006-0-0011 | PC Board |
7760006 | Power cord |
03-1-0005-0192 | Internal Power Supply |
02-002-1-0027 | Lid Assembly |
7724071 | Hinge, friction |
02-002-1-0056 | Seal, lid gasket |
03-0-0003-0313 | Open/Close Label |
03-1-0007-0059 | Pediatric Tube Adapter, Stat Rotor, Printed |
03-0-0003-0579 | Dash Cog Front Panel Label |
03-1-0001-0090 | Button Cover No Emboss |
03-1-0001-0089 | Button Cover Embossed |
03-1-0002-0099 | Button Protector |
03-1-0002-0093 | Button Spacer |
Complies with UL61010-1/CSA C22.2 No. 61010-1 and IEC61010-2-020
Protected by U.S. Patents #6,811,531, #D718,463, & #D734,489. Other Patents
Pending
INSTRUCTIONS FOR DISPOSAL OF WEEE BY USERS IN THE EUROPEAN UNION
This product must not be disposed of with other waste. Instead, it is the
user’s responsibility to dispose of their waste equipment by handing it over
to a designated collection point for the recycling of waste electrical and
electronic equipment. The separate collection and recycling of your waste
equipment at the time of disposal will help to conserve natural resources and
ensure that it is recycled in a manner that protects human health and the
environment. For more information about where you can drop off your waste
equipment for recycling, please contact your local city office, waste disposal
service, or where you purchased the product.
200 SHADY LANE, SUITE 170 – PHILIPSBURG, PA 16866, USA
+1-866-265-1486 (U.S. ONLY) –
+1-814-692-7661
CUSTOMERSERVICE@DRUCKERDIAGNOSTICS.COM
DRUCKERDIAGNOSTICS.COM
ISO 13485 certified
This operator’s manual is part number
03-0-0002-0195 Rev. D
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>