Meraw Cedar Arm Blood Pressure Monitor User Manual
- June 9, 2024
- Meraw
Table of Contents
- Indications for Use
- Contraindications
- Safety Information
- Monitor Components
- LCD Display
- Power Supply
- Installing Batteries
- Setting Date and Time
- Tie The Cuff
- Sit Correctly
- Start The Measurement
- Measurement Tips
- Download App
- Add My Device
- View Data
- Recall the records
- Sync the records
- Delete the records
- Systolic pressure and diastolic pressure
- Standard blood pressure classification
- Irregular Heartbeat Detector
- Blood pressure fluctuate throughout the day
- Different blood pressure at home compared to the hospital
- Measuring on the right arm
- Trouble Shooting
- Specifications
- Product Maintenances
- FCC Statement
- EMC Guidance
- Technical description
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Arm Blood Pressure Monitor
User Manual
Indications for Use
This Blood Pressure Monitor TMB-2084 is a digital monitor intended for use in measuring blood pressure and heartbeat rate with arm circumference ranging from 22 cm to 42 cm (about 8¾˝-16½˝). It is intended for indoor, adult use only.
Contraindications
- The device is not suitable for use on the women who are or may be pregnant.
- The device is not suitable for use on patients with implanted electrical devices, such as cardiac pacemakers, defibrillators.
Safety Information
The signs below might be in the user manual, labeling or other component. They are the requirement of standard and using.
| Refer to instruction manual/booklet To signify that the instruction
manual/booklet must be read.| | Symbol for “Type BF applied part”
---|---|---|---
| Caution Indicates that caution is necessary when operating the device or
control close to where the symbol is placed, or that the current situation
needs operator awareness or operator action in order to avoid undesirable.
consequences.|
| The symbol indicates that the product should not be discarded as unsorted
waste but must be sent to separate collection facilities for recovery and
recycling.
****| Symbol for “SERIAL NUMBER”|
| Symbol for “Recycle”
****| Symbol for “DIRECT CURRENT”| | For indoor use only
| Date and Country of manufacture| | Symbol for “Class II Equipment”
| Symbol for “MANUFACTURER”
| MR Unsafe
To identify an item which poses unacceptable risks to the patient, medical
staff or other persons within the MR environment.
Monitor Components
- CUFF
- AIR HOSE
- AIR CONNECTOR PLUG
- BLUETOOTH BUTTON
- START/STOP BUTTON
- LCD DISPLAY
- Micro USB
- RECORD BUTTON
Component list of pressure measuring system:
- Cuff
- Air pipe
- PCBA
- Pump
- Valve
List
LCD Display
SYMBOL | DESCRIPTION | EXPLANATION |
---|
SYS
| Systolic blood pressure| The high pressure measured.
DIA
| Diastolic blood pressure| The low pressure measured.
MMHG
| mmHg| Measurement Unit of the blood pressure.
AVG
| Average value| Displays average of last 3 readings.
PULSE/MIN
| Pulse display| Pulse in beats per minute.
| Heartbeat| Heartbeat detection during measurement.
| Irregular heartbeat| Irregular heartbeat detected during measurement.
| Data pending to transmit| Appears in next measurement, when the data
transmission fails.
| Bluetooth transfer icon| The bluetooth transfer icon blinks when the
bluetooth is working.
| Battery indicator| Indicate the current battery.
| Excessive body motion detector| Appears when talking, moving, or shaking of
the arm with the cuff on is detected during the measurement.
| Cuff wrap detection| appears when the cuff is secured well.
| Memory display| Indicate it is in the memory mode and which group of memory
it is.
| Current time| Time and date (year/month/day; hour: minute)
| Blood pressure level| Indicates the blood pressure level, See page 24 for
more information.
Power Supply
- Battery powered mode: 6V DC 4× AAA batteries
- AC adapter powered mode: 5V 1A
CAUTION
In order to get the best effect and protect your monitor, please use the right
batteries and special power adapter which complies with local safety standard.
Installing Batteries
-
Slide off the battery cover.
-
Install or replace 4× AAA size batteries as indicated in the battery compartment. The display will light up and show the icon . Press “START/ STOP” button to turn off, otherwise it will power off automatically after 10 seconds.
Any time the battery is low, it will display the icon“ bAt Lo” & . It will power off automatically after 5 seconds. -
Close the battery cover.
Replace the batteries if:
• The low battery symbol appears on the display.
• When any button is pressed and nothing is displayed on the screen.
CAUTION
- Do not use new and used batteries together.
- Do not use different types of batteries together.
- Do not dispose the batteries in fire. Batteries may explode or leak.
- Remove batteries if the device is not likely to be used for some time.
- Worn batteries are harmful to the environment. Do not dispose with daily garbage. Remove the old batteries from the device following your local recycling guidelines.
Setting Date and Time
It is important to set the clock before using your blood pressure monitor, so
that a time stamp can be assigned to each record that is stored in the memory.
(year: 2022—2050 Time format: 12H/24H)
Auto Setting
Once connected to smartphone App via Bluetooth, date and time will be syned to
BPM automatically.
Manual Setting
- When the monitor is off, press and hold the “START/ STOP ” button to display the date format. Press the “Bluetooth” or “Record” button to switch the date format between [month/day] and [day/month].
- Press the “START/ STOP ” button to confirm the date format, then the year will flash. Press the “Bluetooth” or “Record” button to change the year.
- When you get the right year, press the “START / STOP” button to confirm the year. The screen will then show a blinkingnumber representing the [MONTH].
- Repeat steps 2 and 3 to set the [MONTH] and [DAY].
- Repeat steps 2 and 3 to set the time format.
- Repeat steps 2 and 3 to set the [HOUR] and [MINUTE].
- After the minute is set, the LCD will display “do NE” and then it will turn off.
Tie The Cuff
- Insert the air plug into the air jack
- Wrap the arm cuff firmly in place around your left upper arm.
NOTES
The bottom edge of the arm cuff, should be 0.5 inch (1 to 2 cm) above the
elbow. Air tube is on the inside of your arm and aligned with your middle
finger.
3. Secure closed with the fabric fastener.
NOTES
- When you take a measurement on the right arm, the air tube will be at the side of your elbow. Be careful not to rest your arm on the air tube.
- The blood pressure can differ between the right arm and the left arm, and the measured blood pressure values can be different. We recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your measurements.
Sit Correctly
To take a measurement, you need to be relaxed and comfortably seated, under
comfortable room temperature.
Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30
minutes before taking a measurement.
- Sit in a chair with your legs uncrossed and feet flat on the floor.
- Sit with your back and arm supported.
- The arm cuff should be placed on your arm at the same level as your heart, with the arm resting comfortably on a table.
Start The Measurement
Notes:
-To stop the measurement, press the START/STOP button once to defalte the arm cuff.
-Remain still and quiet while taking a measurement.
Press the START/STOP button.
-The arm cuff will start to inflate automatically.
-The whole measurement takes around 35 seconds.
Measurement Tips
To help ensure an accurate reading, follow these directions:
- Stress raises blood pressure. Avoid taking measurements during stressful times.
- Measurements should be taken in a quiet place.
- It is important to take measurements at the same times each day. Taking measurements in the morning and in the evening is recommended.
- Remember to have a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time.
CAUTION
- Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for at least 30 minutes before taking a measurement.
- Rest for at least 5 minutes before taking a measurement.
Download App
Merah Health app is available both on Google play and App Store, search and
download.
Make sure your phone has enough storage and meet lowest system version
requirement before downloading
https://merawlabs.com/pages/meraw-app
Scan above QR code to download App and get App instruction.
Android system: Android 9.0 or above
iOS system: iOS 10.0 or above
Add My Device
Step 1. Add device
After opening Merah App for the first time, it will guide you to add new
device.
Step 2. Choose device
Choose Merah Cedar and continue.
Step 3. Start Connection
Follow the instruction and start connection.
Step 4. Connection successfully
The connection process will complete automatically.
View Data
All synced data will be listed by month, and show brief information including:
-Blood pressure level
-SYS
-DIA
-measured time
Some factors may influence the accuracy of measurement, click a data and add remark on it.
Recall the records
- When the monitor is off, press “RECORD” button, the display will show the average value of the latest three records first.
- When the memory record is less than three groups, the display will show the latest record(memory record 01)
2. Press “Record” or “Bluetooth” button to get the record you want.
Each press “Record” bluetooth will increase the memory record by one in a
cycling manner (AVG-01-02…).
Each press “Bluetooth” bluetooth will increase the memory record by one in a
cycling manner (AVG-…-02-01).
Sync the records
- Turn on Bluetooth and open App on your smartphone before you sync the records.
- Each time you complete a measurement, data will be synced to App automatically, will blink during uploading and turn solid on after uploading successful.
- If you data is not synced to App, will show in data page, indicating the record is not uploaded to App.
- You could press ‘Bluetooth’ button when the blood pressure monitor is off to sync data that are not uploaded.
Delete the records
If you did not get the correct measurement, you can delete results by following steps below.
A: To delete a single measurement:
- Enter the memory recall mode as described in section [Recall the Records]. Press “Record” or “Bluetooth” button to get the measurement you would like to erase.
- Press and hold “Record” button for 3 seconds, and the display will show a blinking “dell yes” along with the memory number of the reading.
- Use the “Record” or “Bluetooth” button to toggle between “dell yes” and “deli no”, Press “START/STOP” to confirm the selection.
If “dell yews” is selected, the unit will delete the record and display “dell do NE” along with the memory number of the reading. After then it will return to last memory record and All the records are pushed forward one digit (e.g., 03 becomes 02, and so on)
If “dell no” is selected, it will stop the deletion.
B: To delete all measurements:
- Enter the memory recall mode as described in section [Recall the Records].
- Press and hold “BLUETOOTH” and “RECORD” button for 5 seconds, and the display will show a blinking “dEL ALL” along with the user ID.
- Use the “BLUETOOTH” or “RECORD” button to toggle between “dell ALL” and “dell no”, Press “START/STOP” to confirm the selection.
If “dEL ALL” is selected, the unit will display “dell ALL do NE” + User ID and delete all the record of the current user.
Several seconds later, it will display “—”. If “dell no” is selected, it will stop the deletion.
Systolic pressure and diastolic pressure
When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
Standard blood pressure classification
The blood pressure classification published by AAC/AHA in 2017 is as follows:
CAUTION
Only a physician can tell your normal BP range.Please contact a physician if
your measuring result falls out of the range. Please note that only a
physician can tell whether your blood pressure value has reached a dangerous
point.
Level Blood Pressure (mmHg) | Normal | Elevated | Grade 1 | Grade 2 | Grade 3 |
---|---|---|---|---|---|
SYS | <120 | 120-129 | 130-139 | 140-159 | ≥160 |
DIA | <80 | <80 | 85-89 | 90-99 | ≥100 |
Irregular Heartbeat Detector
An irregular heartbeat is detected when a heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure. During each measurement, the monitor records all the pulse intervals and calculate the average; if there are two or more pulse intervals, the difference between each interval and the average is more than the average value of ±25%, or there are four or more pulse intervals, the difference between each interval and the average is more than the average value of ±15%, the irregular heartbeat symbol appears on the display when the measurement results are appeared.
CAUTION
The appearance of the IHB icon indicates that a pulse irregularity consistent
with an irregular heart-beat was detected during measurement. Usually this is
NOT a cause for concern. However, if the symbol appears often, we recommend
you seek medical advice. Please note that the device does not replace a
cardiac examination, but serves to detect pulse irregularities at an early
stage.
Blood pressure fluctuate throughout the day
- Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.
- If the person takes medicine, the pressure will vary more.
- Wait at least 3 minutes for another measurement.
Different blood pressure at home compared to the hospital
The blood pressure is different even throughout the day due to weather,
emotion, exercise etc. Also, there is the “white coat” effect, which means
blood pressure usually increases in clinical settings.
What you need to pay attention to when you measure your blood pressure at
home:
If the cuff is tied properly.
If the cuff is too tight or too loose.
If the cuff is tied on the upper arm.
If you feel anxious.
Taking 2-3 deep breaths before beginning will be better for measuring.
Advice: Relax yourself for 4-5 minutes until you calm down.
Measuring on the right arm
It is ok for both arms, but there will be some different results for different
people.
We suggest you measure the same arm every time.
Trouble Shooting
If any abnormality arises during use,please check the following points:
PROBLEM | SYMPTOM | CHECK THIS | REMEDY |
---|---|---|---|
No power | Display can not light up. | Batteries are exhausted. | Replace with |
new batteries.
Batteries are inserted incorrectly.| Insert the batteries correctly.
Adapter is inserted incorrectly.| Insert the AC adapter correctly.
High power| bAt H shows| Adapter voltage is higher than 7.5V| Replace with the
authorized adapter
Low batteries| bAt Lo & shows| Batteries are low.| Please replace batteries.
if voice on, it read for you.
Error message
| E 1 shows| The cuff is not wrapped or wrapped incorrectly, or the cuff air
plug is loose.| Refasten the cuff and insert air tube plug correctly then
measure again.
E 2 or shows| Hand shaking or talking or moving is detected during measuring.|
Relax for 5 minutes. and then keep still, measure again.
E 3 shows| Pulse is not detected during measuring.| Loosen the clothing on
the arm and then measure again.
E 4 shows| The measurement failed.| Relax for 5 minutes and measure again.
Exec shows| Hardware error (X can be some digital symbol, such as 1, 2, etc.)|
Turn off monitor and measure again. If Exec still appears on the display
,please contact the retailer or our customer service
Err USB shows| The output voltage of the adapter is too high or too low|
Replace with the authorized adapter
Warning message| “out” shows| Out of measure- mint range| Relax for a moment.
Refasten the cuff and then measure again.
If the problem persists, contact your physician.
Specifications
Power supply| Battery powered mode: 6V DC 4× AAA batteries AC adapter powered
mode: 5V 1A
(Please only use the recommended AC adapter).
---|---
Display mode| Digital LCD V.A. 61 mm × 74 mm
Measurement mode| Oscillography testing mode
Measurement range| Rated cuff pressure:
0 mmHg ~ 299 mmHg Measurement pressure:
SYS: 60 mmHg ~ 230 mmHg
DIA: 40 mmHg ~ 130 mmHg Pulse value: (40-199) beat/minute
Accuracy| Pressure:
5°C – 40°C within ±3 mmHg Pulse value: ±5%
Normal working condition| A temperature range of: +5°C to +40°C
A relative humidity range of 15% to 90%, non-condensing
An atmospheric pressure range of: 70 kPa to 106 kPa
Storage & transportation condition| Temperature: -20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing
Measurement perimeter of the upper arm| About 22 cm to 42 cm
Weight| Approx.252 g (Excluding the batteries and cuff)
External dimensions| Approx.100.3 mm × 132.3 mm × 47.1 mm
Attachment| 4× AAA batteries, user manual, AC adapter, USB cable
Mode of operation| Continuous operation
Degree of protection| Type BF applied part
Protection against ingress of water| IP21 It means the device could protected
against solid foreign objects of 12.5 mm and greater, and protect against
vertically falling water drops.
Device Classification| Battery Powered Mode:
Internally Powered ME Equipment
AC Adapter Powered Mode: Class II ME Equipment
Software Version| A01
Product Maintenances
CAUTION
-
This device is intended for indoor, home use.
-
This device is not intended for public use.
-
This device is portable, but it is not intended for use during patient transport.
-
This device is not suitable for continuous monitoring during medical emergencies or operations.
-
This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than the arm, or for any purpose other than obtaining a blood pressure measurement.
-
This device is for adults. Do not use this device on neonates or infants.
Do not use it on children unless otherwise instructed by a medical professional. -
Do not use on the women in pregnant, including pre-eclamptic, patients.
-
The device is not suitable for use on patients with implanted, electrical devices, such as cardiac pacemakers, defibrillators.
-
The effectiveness of this device has not been established for use:
-on users with common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation,
-on users with peripheral arterial disease,
-on users undergoing intravascular therapy, or with arteriovenous (AV) shunt.
Consult a medical professional before use. -
Do not use this device for diagnosis or treatment of any health problem or disease. Contact your physician if you have or suspect any medical problem. Do not change your medications without the advice of your physician or health care professional.
-
If you are taking medication, consult your physician to determine the proper time to measure your blood pressure.
-
This device may be used only for the intended use described in this manual, the manufacturer shall have no liability for any incidental, consequential, or special damages caused by misuse or abuse.
CAUTION
-
Report any unexpected operation or events to the manufacturer.
Do not apply the cuff on an arm that has an intravenous drip or a blood transfusion attached. -
Warning:Do not kink,fold,stretch,compress,or otherwise deform the tube during measuring, as the cuff pressure might continuously increase, which could prevent blood flow and result injury.
-
Warning: Taking blood pressure measurements too frequently could disrupt blood circulation and cause injuries.
-
Warning:Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site frequently for irritation.
Warning:Do not place the cuff on the arm of a person whose arteries or veins are undergoing medical treatment,i.e.intra-vascular access or intra-vascular therapy or an arteriovenous (A-V)shunt, which could disrupt blood circulation and cause injuries. -
Do not place the cuff on the arm on the same side of a masticator. my (especially when lymph nodes have been removed).it is recommended to take measurements on the unaffected side.
-
Do not wrap the cuff on the same arm to which another monitoring device is applied. One or both devices could temporarily stop functioning if you try to use them at the same time.
-
Please check that the operation of the device do not result in prolonged impairment of patient blood circulation.
-
Warning: On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, loosen and remove the cuff immediately. Prolonged high pressure applied to the arm (cuff pressure >300 mmHg or constant pressure >15 mmHg for more than 3 minutes) might lead to bruising and discolored skin.
-
Warning:Do not use this device with high-frequency (HF)surgical equipment at the same time.
CAUTION
- Warning: This device is not AP/APG equipment. Do not use the device where flammable anesthetic are present, or in environments mixture with air of with oxygen or nitrous oxide.
- The device contains sensitive electronic components.To avoid measurement errors, avoid taking blood pressure measurements near a strong electromagnetic field radiated interference signal or electrical fast transient/burst signal.
- Wireless communication equipment, such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies may cause interference that may affect the accuracy of measurements. A minimum distance of 1 foot (30 cm) should be kept from such devices during a measurement.
- You can use this device to take your own measurement, no third-party operator is required.
- Please use the device under the environment which is provided in the user manual. Otherwise, the performance and lifetime of the device will be impacted and reduced.
- The device may require up to 30 minutes to warm up / cool down from the minimum / maximum storage temperature before it is ready for use.
- Warning: Excessive cuff tube lengths could cause strangulation if you don’t manage them properly.
- Warning: Do not touch output of the batteries/adapter and the user simultaneously.
- Adapter is specified as a part of ME EQUIPMENT.
- Warning: The power cord is considered the disconnect device for isolating this equipment from supply mains. Do not position the equipment so that it is dicult to reach or disconnect.
- The blood pressure monitor, its adapter, and the cuff are suitable for use within the patient environment.
- Warning: Do not use this device if you are allergic to polyester, nylon, or plastic.
CAUTION
- Warning: Only use accessories approved by manufacturer. Using unapproved accessories might cause damage to the unit and injure users.
- Warning: If you experience discomfort during a measurement, such as pain in the arm or other complaints, press the Power button immediately to release the air from the cuff.
- No calibration is required within two years of reliable service.
- Do not attempt to repair the unit yourself if it malfunctions. Only have repairs carried out by authorized service centers.
- At the request of authorized service personnel, circuit diagrams, component part lists, descriptions, and calibration procedures will be made available by the manufacturer or distributor.
- It is recommended that the performance should be checked after repair, maintenance, and every two years of use, by retesting the requirements in limits of the error of the cuff pressure indication and air leakage (testing at least at 50 mmHg and 200 mmHg).
- Warning: Do not use the device while under maintenance, or being serviced.
- Store your device, cuff and adapter in a clean and dry place, protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on it.
- Make sure the rubber tube of the cuff is not squeezed, stretched, or kinked during storage.
- Warning: Keep the device, cuff, and batteries away from children as they may pose a risk of choking or strangulation if used improperly.
- Clean both device and cuff with a soft, dry cloth. If necessary use a dampened cloth and natural detergent. Do not use alcohol, benzene, or other harsh chemicals.
- Do not wash the cuff in a washing machine or dishwasher!
- The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The typical service life is 10000 times.
- Dispose of accessories, detachable parts, and the device according to the local guidelines.
FCC Statement
FCC ID: OUT9TMB2080-B
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference
that may cause undesired operation.
Caution: The user is cautioned that changes or modifications not expressly
approved by the party responsible for compliance could void the user’s
authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits
for a class b digital device, pursuant to part 15 of the fcc rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions. May cause harmful interference to radio
communications. However there is no quarantine that interference will not
occur in a particular installation.
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on,the
user is encouraged to try to correct the interference by one or more of the
following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
EMC Guidance
The ME EOUIPMENT or ME SYSTEM is suitable for home healthcare environments.
Warning: Don’t be near the active He surgical equipment and the RF shielded
room of an me system for magnetic resonance imaging, where the intensity of EM
disturbances is high.
Warning: Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use
is necessary. This equipment and the other equipment should be observed to
verify that they are operating normally.
Warning: Use of accessories, transducers and cables other than those specified
or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Warning: Portable RE communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12
inches)to any part of the equipment.
TMB-2084 including cables specified by the manufacturer.
Otherwise. degradation of the performance of this equipment could result.
Technical description
- All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life.
- Guidance and manufacturer’s declaration-electromagnetic emissions and Immunity.
Table 1
Guidance and manufacturer’s declaration electromagnetic emissions
Emissions test| Compliance
RF emissions CISPR 11| Group 1
RF emissions CISPR 11| Class [ B ]
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations / flicker emissions IEC 61000-3-3| Comply
Table 2
Guidance and manufacturer’s declaration electromagnetic Immunity
Immunity Test| IEC 60601-1-2 Test level| Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2| ±8 kV contact ±2 kV, ±4 kV, ±8
kV, ±15 kV air| ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC 61000-4-4| +2 kV for power supply lines +1
kV signal input/output 100 kHz repetition frequency| +2 kV for power supply
lines Not Applicable 100 kHz repetition frequency
Surge IEC61000-4-5| 10.5 kV,1 kV differential mode ±0.5kV,±1kV,±2 kV common
mode| ±0.5 kV, ±1 kV differential mode Not Applicable
Voltage dips, short interruptions and voltage variations on power supply
input lines IEC 61000-4-11| 0% UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0% UT; 1 cycle and 70% UT; 25/30 cycles; Single phase: at 0°.
0% UT; 250 / 300 cycle| 0% UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0% UT; 1 cycle and 70% UT; 25/30 cycles; Single phase: at 0°.
0% UT; 250 / 300 cycle
Power frequency magnetic field IEC 61000-4-8| 30 A/m
50 Hz / 60 Hz| 30 A/m
50 Hz / 60 Hz
Conduced RF IEC61000-4-6| 3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80% AM at 1 kHz| 3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80% AM at 1 kHz
Radiated RF IEC61000-4-3| 10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz| 10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz
NOTE UT is the a.c.mains voltage prior to application of the test level.
Table 3
Guidance and manufacturer’s declaration – electromagnetic Immunity
Radiated RF IEC61000 -4-3
(Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communicati-
ons equipment)| Test Frequency
(MHz)| and
(MHz)| Service| Modulation| Maximum
Power (W)| Distance (m)| IEC 60601-1-2 Test Level (V/m)
---|---|---|---|---|---|---|---
385| 380-390| TETRA 400| Pulse modulation 18 Hz| 1.8| 0.3| 27
450| 430-470| GMRS 460, FRS 460| FM ± 5k Hz deviation 1 kHz sine| 2| 0.3| 28
710| 704-787| LTE Band
13, 17| Pulse modulation 217 Hz| 0.2| 0.3| 9
745
780
810| 800-960| GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5| Pulse modulation 18 Hz| 2| 0.3| 28
870
930
1720| 17001990| GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band 1, 3, 4,25;
UMTS| Pulse modulation 217 Hz| 2| 0.3| 28
1845
1970
2450| 24002570| Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE Band 7| Pulse modulation 217 Hz| 2| 0.3| 28
5240| 51005800| Bluetooth, WLAN,
802.11 b/g/n, RFID 2450, LTE
Band 7| Pulse modulation 217 Hz| 2| 0.3| 9
5500
5785
Guangdong Transit Medical Electronics Co., Ltd.
Zone A, No.105, Dongle Road, Torch Development District,
528437 Zhongzheng, Guangdong, China
Distributor & Address
Merah Labs Inc.
4500 Great America Parkway, Suite 1023, Santa Clara, CA 95054
Global Customer Service
support@merawlabs.com
www.merawlabs.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>