RKI INSTRUMENTS 65-2391RK CO2 Transmitter User Manual
- June 9, 2024
- RKI INSTRUMENTS
Table of Contents
INSTRUMENTS 65-2391RK CO2 Transmitter
User Manual
65-2391RK CO2 Transmitter
WARNING
Read and understand this instruction manual before operating transmitter.
Improper use of the transmitter could result in bodily harm or death.
Periodic calibration and maintenance of the transmitter is essential for
proper operation and correct readings. Please calibrate and maintain this
transmitter regularly! Frequency of calibration depends upon the type of use
you have and the sensor types. Typical calibration frequencies for most
applications are between 6 and 12 months, but can be required more often or
less often based on your usage.
Product Warranty
RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from
defects in materials, workmanship, and performance for a period of one year
from date of shipment from RKI Instruments, Inc. Any parts found defective
within that period will be repaired or replaced, at our option, free of
charge. This warranty does not apply to those items which by their nature are
subject to deterioration or consumption in normal service, and which must be
cleaned, repaired, or replaced on a routine basis. Examples of such items are:
a) Absorbent cartridges
b) Pump diaphragms and valves
c) Fuses
d) Batteries
e) Filter elements
Warranty is voided by abuse including mechanical damage, alteration, rough
handling, or repair procedures not in accordance with the operator’s manual.
This warranty indicates the full extent of our liability, and we are not
responsible for removal or replacement costs, local repair costs,
transportation costs, or contingent expenses incurred without our prior
approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND
REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR
LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO,
THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO
EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR
CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS
PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized
distributors, dealers, and representatives as appointed by RKI Instruments,
Inc.
We do not assume indemnification for any accident or damage caused by the
operation of this gas monitor, and our warranty is limited to the replacement
of parts or our complete goods.
Overview
This manual describes the 65-2391RK CO2 transmitter. This manual also describes how to install, start up, configure, maintain, and calibrate the transmitter when it is used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the CO2 transmitter.
Specifications
Table 1 lists specifications for the CO2 transmitter.
Table 1:Specifications
Target/Calibration Gas | Carbon Dioxide (CO2) |
---|---|
Area Classification | Explosionproof for Class I, Groups B, C, and D |
Sampling Method | Diffusion |
Signal Output | 4 to 20 mA |
Detection Range | 65-2391 RK-02: 0 – 5,000 ppm 65-2391 RK-03: 0 – 5% Volume |
65-2391 RK-05: 0 – 50% Volume 65-2391RK-10: 0 – 100% Volume
Response Time| 90% in 45 seconds
Accuracy| ± 5% of reading or ± 2% of full scale (whichever is greater)
WARNING: When using the 65-2391RK, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the 65-2391RK
and to minimize the risk of personal injury. Be sure to maintain and
periodically calibrate the 65-2391RK as described in this manual.
Description
This section describes the components of the CO2 transmitter. The transmitter
is a 4 – 20 mA type detector head. It consists of the infrared CO2 detector,
amplifier, and junction box. Infrared CO2 Detector
The infrared CO2 detector is made up of a miniature infrared CO2 detector
housed and encapsulated in a pipe nipple. The pipe nipple has 3/4 inch NPT
threads on each end and a 1 1/4 inch hex that allows removal or installation
of the detector with a wrench. A porous flame arrestor that is coated with a
hydrophobic film that repels liquids is on one end of the detector and allows
sample gas to enter the detector. Four color coded leads, red, white, green,
and black, extend from the other end of the detector. The leads allow you to
connect the detector to the amplifier.
To distinguish the different ranges of CO2 detectors from one another, a short
length of shrink tubing is applied to the wiring where it comes out of the
nipple. The following table indicates the color of the shrink tubing and the
color of the wire to which it is applied.
Table 2: CO2 Detector Color Designations
Detector | Color Designations |
---|---|
0 – 5,000 ppm | black shrink tubing on white wire |
0 – 5% volume | green shrink tubing on white wire |
0 – 50% volume | green shrink tubing on red wire |
0 – 100% volume | red shrink tubing on green wire |
Amplifier
The amplifier converts the electrical output from the detector to a 4 to 20 mA
signal that corresponds to the detection range and transmits the signal to a
gas monitoring controller. The amplifier includes the interconnect terminal
strip, span potentiometer, zero potentiometer, and test points (see Figure 1).
Interconnect Terminal Strip
The interconnect terminal strip is a seven-point terminal strip. Use the
interconnect terminal strip to connect the IR CO2 detector to the amplifier
and the amplifier to a controller.
NOTE: The IR CO2 detector is factory-wired to the amplifier. See the
Installation section of this manual for all wiring procedures related to the
transmitter.
Span Potentiometer
The span potentiometer is near the bottom of the amplifier (see Figure 1). Use
the span potentiometer to adjust the transmitter’s response output during the
calibration procedure. Turn the adjustment screw clockwise to increase the gas
response and counterclockwise to decrease the gas response.
Zero Potentiometer
The zero potentiometer is to the right of the span potentiometer (see Figure
1). Use the zero potentiometer to adjust the transmitter’s target gas-free
output during the start-up and calibration procedures. Turn the adjustment
screw clockwise to increase the zero output and counterclockwise to decrease
the zero output.
CAUTION: The amplifier includes two additional potentiometers. They are
factory-set. Do not adjust them.
Test Points
The test points (labeled TP+ and TP-) are to the left of the interconnect
terminal strip (see Figure 1). The test points produce a 100 to 500 mV output
that corresponds to the transmitter’s 4 to 20 mA output. Use the test points
and a voltmeter to measure the transmitter’s output during the start-up and
calibration procedures.
Junction Box
Use the junction box to install the CO2 transmitter at a mounting site that is
remote from the controller. The junction box also protects the amplifier and
wiring connections made to the amplifier. Use the two 3/4 in. conduit hubs to
mount the detector to the junction box (bottom hub) and connect wiring from
the amplifier to the controller (top hub).
NOTE: The detector and amplifier are factory-mounted to the junction box.
Use the junction box’s two mounting holes to mount the CO2 transmitter to a
vertical surface at the monitoring site. Use the cover on the front of the
junction box to access the interior of the junction box.
Installation
This section describes procedures to mount the CO2 transmitter in the
monitoring environment and wire the transmitter to a controller.
Mounting the CO2 Transmitter
-
Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
• Select a site where the transmitter is not likely to be bumped or disturbed.
• Select a site where there is sufficient room to perform start-up, maintenance, and calibration procedures.
• Mount the transmitter at least six feet from the floor to minimize the possibility of someone breathing on the unit. The exhaled CO2 may cause an alarm. -
If the detector is mounted to the junction box, skip to step 5. If not, continue with step 2.
-
Remove the junction box cover.
-
Guide the four wires that extend from the top of the detector through the bottom conduit hub of the junction box.
-
Screw the detector into the bottom conduit hub of the junction box.
-
At the monitoring site, use #10 screws through the junction box’s two mounting holes to secure the junction box to a vertical surface.
CAUTION: Mount the CO2 transmitter with the detector facing down (see
Figure 2.)
Wiring the CO2 Transmitter to a Controller
WARNING: Always verify that the power to the controller is off before you
make wiring connections.
-
Turn off power to the controller.
-
Place the controller’s power switch in the OFF position.
-
Remove the junction box cover.
-
Verify that the detector leads are wired to the amplifier’s interconnect terminal strip.
If necessary, connect the detector leads to the interconnect terminal strip as shown in Figure 3. -
Guide a three-conductor, shielded cable or three wires in conduit through the top conduit hub of the junction box.
-
Connect the three wires to the interconnect terminal strip as follows (see Figure 3).
• Connect the positive wire to the 24VDC terminal.
• Connect the feedback wire to the 4-20 (FB) terminal.
• Connect the negative wire to the GND (DC -) terminal.
CAUTION: If shielded cable is used, leave the cable shield’s drain wire insulated and disconnected at the transmitter. You will connect the opposite end of the cable’s drain wire at the controller to chassis ground. -
Secure the junction box cover to the junction box.
-
Route the cable or wires leading from the CO2 transmitter through one of the conduit hubs at the controller housing.
CAUTION: Do not route power and transmitter wiring through the same conduit hub. The power cable may disrupt the transmission of the transmitter signal to the controller. -
Connect the wires to the applicable transmitter terminal strip at the controller as shown in Figure 3 below.
-
If shielded cable is used, connect the cable’s drain wire to an available chassis ground at the controller.
Start Up
This section describes procedures to start up the CO2 transmitter and place
the transmitter into normal operation.
Introducing Incoming Power
- Complete the installation procedures described earlier in this manual.
- Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual.
- Turn on or plug in power to the controller, then place the controller’s power switch in the ON position.
- Verify that the controller is on and operating properly. Refer to the controller operator’s manual.
NOTE: When first powered up, the transmitter will enter about a one
minute period when the 4-20 mA output is stabilizing and may be above the
controller alarm points or well below zero momentarily. RKI controllers have a
one minute warmup period when the controller does not display any gas reading
or give any alarm indication. The CO2 transmitter’s 4-20 mA signal should be
stable by the time the controller’s warmup period is over.
CAUTION: Allow the detector to warm up for 5 minutes before you continue
with the next section, “Setting the Zero Signal”.
Setting the Zero Signal
Since there is a background of CO2 in air of typically 300 – 600 ppm (0.03 –
0.06 %volume), it is necessary to use a calibration kit with a 100% nitrogen
cylinder to set the zero signal of a CO2 transmitter with a low range
detector. Fresh air can be used to zero the transmitter if a 0-50 %volume or
0-100 %volume detector is being used.
NOTE: The following procedure assumes the use of a calibration kit which
includes a 100% nitrogen calibration gas cylinder, a 0.5 LPM fixed flow
regulator with an on/off knob, a calibration cup for the detector, and a short
piece of sample tubing to connect the regulator to the calibration cup.
-
Place the controller into its calibration program or disable external alarms to avoid accidental alarms during the zero setting procedure.
-
Screw the calibration cup onto the bottom of the detector.
-
Use the sample tubing that comes with the calibration kit to connect the regulator to the calibration cup.
-
Set a voltmeter to measure in the millivolt (mV) range.
-
Remove the junction box cover, then plug the voltmeter leads into the test points on the amplifier.
Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point labeled TP-. -
Screw the regulator into the 100% nitrogen calibration cylinder.
-
Turn the regulator knob counterclockwise to open the regulator.
-
Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use the zero potentiometer on the amplifier to adjust the reading to 100 mV
(±2 mV). -
Turn the regulator knob clockwise to close the regulator.
-
Unscrew the regulator from the 100% nitrogen calibration cylinder.
-
Unscrew the calibration cup from the detector.
-
Reinstall the junction box cover.
-
Return the controller to normal operation or enable external alarms.
Maintenance
This section describes maintenance procedures. It includes preventive
maintenance, troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum
performance of the CO2 transmitter. It includes daily, monthly, and biannual
procedures.
Daily
Verify a display reading at the controller of the background concentration of
CO2. Typical background concentrations of CO2 vary from about 300 to 600 ppm
(0.03 to 0.06 %volume) depending on location. The 0-5,000 ppm and 0-5 %volume
detectors will display a reading in a fresh air environment because they have
low ranges. The 0-50 %volume and 0-100% volume detectors will not display a
reading in fresh air because their ranges are too large. Investigate
significant changes in the display reading.
Monthly
This procedure describes a test to verify that the CO2 transmitter responds
properly to the target gas.
WARNING: The controller is not an active gas monitoring device during the
response test procedure.
NOTE: Performing a response test on the CO2 transmitter may cause alarms.
Be sure to put the controller into its calibration program or disable external
alarms before performing this test.
NOTE: The following procedure assumes the use of a calibration kit which
includes a calibration gas cylinder, a 0.5 LPM fixed flow regulator with an
on/off knob, a calibration cup for the detector, and a short piece of sample
tubing to connect the regulator to the calibration cup.
Preparing for the response test
-
Place the controller into its calibration program or disable external alarms.
-
Verify that the controller display reading for the channel you are testing is consistent with typical background levels of CO2.
If the display reading is not consistent with typical background levels of CO2, set the zero reading of the transmitter as described in “Start Up” on page 6 of this manual, then continue this procedure. -
Screw the calibration cup onto the bottom of the detector.
-
Use the calibration kit sample tubing to connect the regulator to the calibration cup.
-
Set a voltmeter to measure in the millivolt (mV) range.
-
Remove the junction box cover, then plug the voltmeter leads into the test points on the amplifier.
Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point labeled TP-. -
Use the following formula to determine the correct test points output for the test sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100 For example, with a test sample of 2.5% CO2 and a fullscale setting of 5% CO2, the correct output is 300 mV.
300 (mV) = (2.5/5) X 400 +100
Performing the response test
-
Screw the regulator into the calibration cylinder.
-
Turn the regulator knob counterclockwise to open the regulator.
-
Allow the gas to flow for one minute.
-
Verify that the reading is within ± 20% of the response reading you determined earlier.
NOTE: If the reading is not within ± 20% of the correct response reading, calibrate the transmitter as described in “Calibration” on page 13 of this manual. -
Turn the regulator knob clockwise to close the regulator.
-
Unscrew the regulator from the calibration cylinder.
-
Unscrew the calibration cup from the detector.
-
Remove the voltmeter leads from the amplifier test points.
-
Reinstall the junction box cover.
-
When the controller display reading falls below the alarm setpoints, return the controller to normal operation.
-
Store the components of the calibration kit in a safe place.
Biannually
Calibrate the CO2 transmitter as described in “Calibration” on page 13 of this
manual.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the CO2 transmitter.
NOTE: This troubleshooting guide describes transmitter problems only. See
the controller operator’s manual for problems you may encounter with the
controller.
Table 3:Troubleshooting the CO2 Transmitter
Condition | Symptom(s) | Probable Causes | Recommended Action |
---|---|---|---|
Fail Condition | • Controller indicates a fail condition. | • The transmitter |
wiring is disconnected or misconnected.
• The transmitter’s reading in air is low
enough to cause a fail condition.
• The transmitter is malfunctioning.| 1. Verify that the transmitter wiring
is
correct and secure.
2. Calibrate the transmitter.
3. If the fail condition continues, replace
the detector.
4. If the fail condition continues, contact
RKI for further instruction.
Slow or No Response/
Difficult or Unable to
Calibrate| • Transmitter respondsslowly or does not
respond to response test.
• Unable to accurately set the zero or
response reading during calibration.
• Transmitter requires frequent calibration.
Note: Under “normal” circumstances, the transmitter requires calibration
once every six months. Some applications
may require a more frequent calibration
schedule.| • The calibration cylinder is low, out-dated, or defective.
• The transmitter is malfunctioning.| 1. Verify that the calibration cylinder
contains an adequate supply of a fresh test sample.
2. Verify that the regulator used for
calibration is a 0.5 LPM regulator.
3. If the calibration/response difficulties
continue, replace the detector.
4. If the calibration/response difficulties
continue, contact RKI for further instruction.c
Replacing Components of the CO2 Transmitter
This section includes procedures to replace the IR CO2 detector and amplifier.
Replacing the IR CO2 Detector
-
Turn off power to the controller.
-
Place the controller’s power switch in the OFF position.
-
Remove the junction box cover.
-
Disconnect the detector leads from the interconnect terminal strip. Note the position of the color-coded leads as you remove them.
-
Unscrew the detector from the junction box.
-
Guide the detector leads of the replacement detector through the bottom conduit hub of the junction box, then screw the mounting threads of the detector into the conduit hub.
-
Connect the detector leads to the interconnect terminal strip as shown in Table 4 below and Figure 3 on page 6 of this manual.
Table 4: Reconnecting the CO2 Detector to the Amplifier** Detector Lead| Amplifier Interconnect Terminal Strip**
---|---
Red| RED
White| WHT
Green| GRN
Black| BLK -
Turn on power to the controller.
-
Place the controller’s power switch in the ON position.
NOTE: When first powered up, the transmitter will enter about a one minute period when the 4-20 mA output is stabilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The CO2 transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue with the next step. -
Calibrate the replacement detector as described in “Calibration” on page 13 of this manual.
Replacing the Amplifier
-
Turn off power to the controller
-
Place the controller’s power switch in the OFF position.
-
Remove the junction box cover.
-
Disconnect the detector leads from the interconnect terminal strip.
-
Disconnect the wiring that connects the CO2 transmitter to the controller from the amplifier’s interconnect terminal strip.
-
Unscrew and remove the two screws that secure the amplifier to the junction box. The screws are at the top left and bottom right of the amplifier.
-
Remove the amplifier.
-
Place the new amplifier in the same position as the amplifier you removed in the previous step.
-
Use the two screws you removed in step 6 to secure the amplifier to the junction box.
-
Reconnect the wiring that connects the controller to the CO2 transmitter at the amplifier’s interconnect terminal strip as shown in Table 5 below and Figure 3 on page 6 of this manual.
Table 5: Reconnecting the CO2 Amplifier to the Controller** Amplifier Interconnect Terminal Strip| Controller Transmitter Terminal Strip (typical)**
---|---
GND| – (DC -)
20-Apr| 4 – 20 mA (FB or S)
24V| + 24V -
Reconnect the detector leads to the amplifier’s interconnect terminal strip as shown in Table 6 below and Figure 3 on page 6 of this manual.
Table 6: Reconnecting the CO2 Detector to the Amplifier** Detector Lead| Amplifier Interconnect Terminal Strip**
---|---
Red| RED
White| WHT
Green| GRN
Black| BLK -
Turn on power to the controller.
-
Turn on the controller and place it into normal operation.
NOTE: When first powered up, the transmitter will enter about a one minute period when the 4-20 mA output is stabilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The CO2 transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next step. -
Calibrate the CO2 transmitter as described in “Calibration” on page 13 of this manual.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all
applications, a calibration frequency of every 6 months is adequate for most
infrared CO2 transmitter applications. Unless experience in a particular
application dictates otherwise, RKI Instruments, Inc. recommends a calibration
frequency of every 6 months for the infrared CO2 transmitter.
If an application is not very demanding, for example detection in a clean,
temperature controlled environment where calibration adjustments are minimal
at calibration, then a
calibration frequency of every 9 to 12 months is adequate.
If an application is very demanding, for example if the environment is not
well controlled, then more frequent calibration than every 6 months may be
necessary.
Calibration
This section describes how to calibrate the CO2 transmitter. It includes
procedures t o prepare for calibration, set the zero reading, set the response
reading, and return to normal operation.
WARNING: The controller is not an active gas monitoring device during the
calibration procedure.
NOTE: The following procedure assumes the use of a calibration kit which
includes a CO2 calibration gas cylinder, a 100% nitrogen cylinder, a 0.5 LPM
fixed flow regulator with an on/off knob, a calibration cup for the detector,
and a short piece of sample tubing to connect the regulator to the calibration
cup.
Preparing for Calibration
-
Screw the calibration cup onto the bottom of the CO2 detector.
-
Use the sample tubing that comes with the calibration kit to connect the regulator to the calibration cup.
-
Set a voltmeter to measure in the millivolt (mV) range.
-
Remove the junction box cover, then plug the voltmeter leads into the test points on the amplifier.
Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point labeled TP-. -
Use the following formula to determine the correct test points output for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 2.5% CO2 and a fullscale setting of 5% CO2, the correct output is 300 mV. 300 (mV) = (2.5/5) X 400 +100 -
Place the controller into its calibration program or disable external alarms.
NOTE: Calibrating the CO2 transmitter may cause alarms. Be sure to put
the controller into its calibration program or disable external alarms before
continuing.
Setting the Zero Reading
Since there is a background of CO2 in air of typically 300 – 600 ppm (0.03 –
0.06 %volume), it is necessary to use a calibration kit with a 100% nitrogen
cylinder to set the zero signal of a CO2 transmitter with a low range
detector. Fresh air can be used to zero the transmitter if a 0-50 %volume or
0-100 %volume detector is being used.
- Screw the regulator into the 100% nitrogen cylinder.
- Turn the regulator knob counterclockwise to open the regulator.
- Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use the zero potentiometer on the amplifier to adjust the reading to 100 mV (±2 mV).
- Turn the regulator knob clockwise to close the regulator.
- Unscrew the regulator from the 100% nitrogen cylinder. Leave the sample tubing connected to the regulator and the calibration cup.
Setting the Response Reading
- Screw the regulator into the calibration cylinder. Verify that the calibration gas is representative of the transmitter’s target gas.
- Turn the regulator knob counterclockwise to open the regulator.
- Allow the calibration gas to flow for one minute and verify that the reading matches the response reading (±2 mV) you determined earlier. If necessary, use the span potentiometer on the amplifier to adjust the reading to match the correct response reading.
- Turn the regulator knob clockwise to close the regulator.
- Unscrew the regulator from the calibration cylinder.
Returning to Normal Operation
-
Remove the voltmeter leads from the amplifier test points.
-
Unscrew the calibration cup from the detector.
NOTE: For convenience, leave the regulator and calibration cup connected by the sample tubing. -
Secure the junction box cover to the junction box.
-
When the display reading falls below the alarm setpoints, return the controller to normal operation.
-
Verify that the controller display reading decreases and stabilizes at a typical background CO2 level. The 0-5,000 ppm and 0-5 %volume detectors will display a reading in a fresh air environment because they have low ranges. The 0-50 %volume and 0-100 %volume detectors will not display a reading in fresh air because their ranges are too large.
-
Store the components of the calibration kit in a safe and convenient place.
Parts List
Table 7 lists replacement parts and accessories for the CO2 transmitter.
Table 7:Parts List
Part Number | Description |
---|---|
06-1248RK | Sample tubing (3/16 in. x 5/16 in.; specify length when ordering) |
18-0405RK-01 | Junction box (without cover; pre-drilled for amplifier) |
18-0406RK | Junction box cover (cover only) |
57-1050RK | Amplifier (specify detector part number when ordering) |
61-0191RK-02 | Infrared CO2 detector, 0 – 5,000 ppm |
61-0191RK-03 | Infrared CO2 detector, 0 – 5% Volume |
61-0191RK-05 | Infrared CO2 detector, 0 – 50% Volume |
61-0191RK-10 | Infrared CO2 detector, 0 – 100% Volume |
65-2391RK-02 | CO2 transmitter, 0 – 5,000 ppm (includes detector, junction box, |
and amplifier)
65-2391RK-03| CO2 transmitter, 0 – 5% Volume (includes detector, junction box,
and amplifier)
65-2391RK-05| CO2 transmitter, 0 – 50% Volume (includes detector, junction
box, and amplifier)
65-2391RK-10| CO2 transmitter, 0 – 100% Volume (includes detector, junction
box, and amplifier)
71-0121RK| 65-2391RK CO2 Transmitter Operator’s Manual (this document)
81-0070RK-01| Steel calibration cylinder, 2,000 ppm CO2, 34-liter
81-0072RK-01| Steel calibration cylinder, 2.5% CO2, 34-liter
Part Number| Description
---|---
81-0073RK-01| Steel calibration cylinder, 15% CO2, 34-liter
81-0076RK-01| Zero air calibration cylinder (34-liter)
81-1050RK| Regulator, 0.5 liter/minute, with gauge and knob, for 17- and
34-liter calibration cylinders
81-1103RK| Calibration cup
82-0006RK| Pot adjust screwdriver, for calibration
1.888.610.7664
www.calcert.com
sales@calcert.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>