1370 Digital Oscilloscope
PeakTech® 1340 – 1375
Operating instructions 2 CH & 4 CH
Digital Memory Oscilloscopes
Table of contents
1. Safety instructions…………………………………………………………………………………………………………….. 3 2. Safety
symbols …………………………………………………………………………………………………………………. 5 3. Quick guide
………………………………………………………………………………………………………………………. 6 Structure of the
oscilloscope…………………………………………………………………………………………………….. 6 Front
…………………………………………………………………………………………………………………………………….. 6 Function
keys…………………………………………………………………………………………………………………………. 7 Back cover
…………………………………………………………………………………………………………………………….. 8 Controls 2
CH…………………………………………………………………………………………………………………………. 9 Controls 4
CH…………………………………………………………………………………………………………………………. 10 Introduction to User
Control……………………………………………………………………………………………………….11 Testing before
commissioning……………………………………………………………………………………………………12 Function
test…………………………………………………………………………………………………………………………… 12 Probe
compensation…………………………………………………………………………………………………………………13 Setting the probe
damping factor………………………………………………………………………………………………. 14 Safety instructions for
using the scanner……………………………………………………………………………………..14 Self-
calibration………………………………………………………………………………………………………………………… 15 Introduction
Vertical System…………………………………………………………………………………………………….. 15 Introduction
Horizontal System…………………………………………………………………………………………………..16 Introduction
Trigger System……………………………………………………………………………………………………… 17 Introduction
Touchscreen Operation………………………………………………………………………………………….. 17 Setting the menu
via touch screen…………………………………………………………………………………………….. 17 Gestures in Normal
Mode ……………………………………………………………………………………………………….. 19 Gestures in Wave Zoom
Mode…………………………………………………………………………………………………..22 Other Touchscreen
Settings………………………………………………………………………………………………………24 4. Advanced user
instructions……………………………………………………………………………………………….. 26 Adjusting the vertical
system…………………………………………………………………………………………………….. 27 Mathematical
functions……………………………………………………………………………………………………………..29 FFT
function…………………………………………………………………………………………………………………………… 30 Using “VERTICAL
POSITION” and “VOLTS/DIV” knobs………………………………………………………………. 32 Set horizontal
system………………………………………………………………………………………………………………. 33 Enlarge waveform
(zoom)………………………………………………………………………………………………………… 33 Set trigger
system…………………………………………………………………………………………………………………….34 Single
Trigger…………………………………………………………………………………………………………………………. 34 Trigger Brief
Description……………………………………………………………………………………………………………35 Detailed trigger
description………………………………………………………………………………………………………. 36 Operating the function
menu…………………………………………………………………………………………………….. 48 Sampling settings
…………………………………………………………………………………………………………………… 48 Display menu
Settings………………………………………………………………………………………………………………50 XY mode
………………………………………………………………………………………………………………………………..51 Frequency counter
…………………………………………………………………………………………………………………..51 Save and Recalling
Waveforms………………………………………………………………………………………………… 52 Screenshots and USB memory
………………………………………………………………………………………………… 53 Saving and playing back
recordings………………………………………………………………………………………….. 55 Set additional system
functions…………………………………………………………………………………………………. 58 Automatic measuring
functions…………………………………………………………………………………………………. 61 Cursor
measurements……………………………………………………………………………………………………………… 64 Autoscale
function…………………………………………………………………………………………………………………… 66 Help
Function…………………………………………………………………………………………………………………………. 68 Executing
keys……………………………………………………………………………………………………………………….. 68 Print
screenshot……………………………………………………………………………………………………………………….70 5. Arbitrary
function generator ……………………………………………………………………………………………….71
Connection…………………………………………………………………………………………………………………………….. 71 Channel
settings …………………………………………………………………………………………………………………….. 71 Signal
settings………………………………………………………………………………………………………………………… 71 Sine
signal……………………………………………………………………………………………………………………………… 72 Frequency settings
…………………………………………………………………………………………………………………. 72 Period
settings…………………………………………………………………………………………………………………………73
– 1 –
Amplitude settings ………………………………………………………………………………………………………………….. 73 Offset
settings …………………………………………………………………………………………………………………………73 High Level
Setting…………………………………………………………………………………………………………………… 73 Low Level
Setting……………………………………………………………………………………………………………………. 73 Rectangle
signal……………………………………………………………………………………………………………………… 73 Ramp signal
………………………………………………………………………………………………………………………….. 73 Symmetry
adjustment……………………………………………………………………………………………………………….73 Pulse signal
…………………………………………………………………………………………………………………………… 73 Pulse width
setting……………………………………………………………………………………………………………………73 Duty cycle
setting……………………………………………………………………………………………………………………. 74 Arbitrary
Signals……………………………………………………………………………………………………………………… 74 Integrated
waveforms………………………………………………………………………………………………………………. 74 User-generated
waveforms………………………………………………………………………………………………………. 76 6. Multimeter
……………………………………………………………………………………………………………………….. 78 Input
sockets………………………………………………………………………………………………………………………….. 78 DMM
menu…………………………………………………………………………………………………………………………….. 78 DMM information
window ………………………………………………………………………………………………………… 79 Multimeter Measuring
functions …………………………………………………………………………………………………79 AC/DC voltage measurement
…………………………………………………………………………………………………… 79 AC/DC current
measurements………………………………………………………………………………………………….. 80 Resistance
measurement………………………………………………………………………………………………………….80 Diode
measurement………………………………………………………………………………………………………………… 80 Capacity measurement
…………………………………………………………………………………………………………….81 Continuity test
…………………………………………………………………………………………………………………………81 Multimeter
Functions……………………………………………………………………………………………………………….. 81 Data Hold
Mode……………………………………………………………………………………………………………………….81 Information display
…………………………………………………………………………………………………………………. 81 Auto or Manual Range Selection
……………………………………………………………………………………………….81 Relative value measurements
………………………………………………………………………………………………….. 81 7. Communication with the
PC……………………………………………………………………………………………….. 82 USB
interface………………………………………………………………………………………………………………………… 82 LAN interface
………………………………………………………………………………………………………………………….83 Direct connection
…………………………………………………………………………………………………………………….83 Connection via
router………………………………………………………………………………………………………………. 85 WiFi connection to PC
…………………………………………………………………………………………………………….. 87 Using Wi-Fi as an access point
………………………………………………………………………………………………… 87 Connect to Wi-Fi Station
…………………………………………………………………………………………………………..88 8. Application examples
…………………………………………………………………………………………………………90 Example 1: Measuring a simple
signal……………………………………………………………………………………….. 90 Example 2: Measuring the gain of
an amplifier……………………………………………………………………………. 91 Example 3: Measuring a single
signal………………………………………………………………………………………… 92 Example 4: Detailed signal
analysis…………………………………………………………………………………………… 93 Example 5: Application of the
X-Y function…………………………………………………………………………………. 94 Example 6: Video signal
trigger………………………………………………………………………………….95 9.
Troubleshooting……………………………………………………………………………………………………………….. 95 10.Technical
specifications …………………………………………………………………………………………………… 96 General
specifications …………………………………………………………………………………………………………….. 101 11.Appendix
…………………………………………………………………………………………………………………………. 102 Appendix A: Scope of
delivery …………………………………………………………………………………………………..102 Appendix B: Maintenance
and care ……………………………………………………………………………………………102 Appendix C:Cloning a
Waveform………………………………………………………………………………………………. 102 Appendix D: Multimeter
Recorder Recording………………………………………………………………………………..106 Li-Po battery and safety
…………………………………………………………………………………………………………….109 NOTE :
This series of oscilloscopes offers various models with many different
additional functions and specifications.
In this manual, these functions are also described for models that may not
correspond to the functions of
your model. For detailed information on the built-in functions of each model,
refer to the technical
specifications starting on page 98.
– 2 –
1. safety instructions for operating the appliance
This product complies with the requirements of the following European Union
Directives for CE conformity: 2014/30/EU (Electromagnetic Compatibility),
2014/35/EU (Low Voltage), 2011/65/EU (RoHS).
Overvoltage category II; pollution degree 2.
To ensure the operational safety of the unit and to avoid serious injuries due
to current or voltage surges or short circuits, it is essential to observe the
following safety instructions when operating the unit.
Damage caused by non-compliance with these instructions is excluded from
claims of any kind.
- This unit must not be used in high energy circuits. Before connecting the unit to a mains socket, check that the voltage setting on the unit corresponds to the
existing mains voltage Connect the appliance only to sockets with earthed protective conductor. Do not place the appliance on a damp or wet surface. Do not operate the unit near strong magnetic fields (motors, transformers, etc.) Do not exceed the maximum permissible input values under any circumstances (serious risk of injury
and/or destruction of the unit). The specified maximum input voltages must not be exceeded. If it cannot be excluded beyond doubt that
these voltage peaks are exceeded due to the influence of transient disturbances or for other reasons, the measuring voltage must be pre-damped accordingly (10:1). Before switching to another measuring function, disconnect the test leads or probe from the measuring circuit. Check the unit, test leads and other accessories for possible damage or bare or bent cables and wires before commissioning. If in doubt, do not carry out any measurements. Only carry out measuring work in dry clothing and preferably in rubber shoes or on an insulating mat. Do not touch the measuring tips of the test leads. It is essential to observe the warnings on the unit. Unit must not be operated unattended Do not expose the unit to extreme temperatures, direct sunlight, extreme humidity or moisture. Avoid strong vibration. Keep hot soldering guns away from the immediate vicinity of the unit. Before starting measurement operation, the unit should be stabilised to the ambient temperature (important when transporting from cold to warm rooms and vice versa). Do not exceed the set measuring range during any measurement. This will prevent damage to the device. Clean the cabinet regularly with a damp cloth and a mild detergent. Do not use corrosive abrasive cleaners. This unit is suitable for indoor use only. Avoid any proximity to explosive and flammable substances. Opening of the unit and maintenance and repair work may only be carried out by qualified service technicians. Do not place the front of the unit on the workbench or work surface to avoid damage to the controls. Do not make any technical changes to the unit. -Measuring instruments do not belong in children’s hands-
– 3 –
Warning: If the oscilloscope is connected to an input signal greater than 42V
peak (30Vrms) or circuits greater than 4800VA, please follow the instructions
below to avoid fire or electric shock: -Use only insulated probes and test
leads. -Inspect all accessories before use and replace if damaged. If in
doubt, do not take measurements. -Remove the USB cable connecting the
oscilloscope to the computer. Never exceed the maximum specified input
voltages. Since the voltage is transmitted directly to the oscilloscope with
the aid of the probe, the unit may be damaged or there is a risk of injury
from electric shock. -Do not use exposed BNC or banana plugs. -Do not insert
any metal objects into the connections.
Cleaning the appliance: Before cleaning the appliance, disconnect the mains
plug from the socket. Clean the appliance only with a damp, lint-free cloth.
Use only commercially available detergents. When cleaning, make absolutely
sure that no liquid gets into the interior of the unit. This could lead to a
short circuit and destruction of the unit.
– 4 –
2. safety symbols and terms
You can find the following symbols in this operating manual or on the meter.
WARNING!
“Warning” indicates conditions and operating steps that pose a danger to the operator.
CAUTION!
“Caution” indicates conditions and operations that may cause damage to the product or other property.
Danger: High- See operating
Protective
Ground terminal
voltage
instructions conductor terminal Unit dimensions
(earth)
In order to avoid physical damage and damage to the measuring device and the
objects to be measured, you should read the following paragraph carefully and
also keep it in mind at all times during future use. This device may only be
used for the intended applications.
Warning:
The channels of the oscilloscope are not electrically isolated from each
other. Therefore, the measuring channels should be on a common ground during a
measurement. To avoid short circuits, the earth terminals must not be
connected to different, non-insulated DC levels.
Diagram of the internal ground wiring (GND):
Probe
Oscilloscope
Electrical Outlet
Signal Input
Power Cord
Ground Clip
Due to the internally connected earth terminals between the BNC socket, USB
port and IEC plug, no measurement of the mains voltage should be carried out
during mains voltage operation of the oscilloscope in conjunction with a mains
voltage operated PC. In the event of an error, a voltage flashover via the GND
of the PC could a
– 5 –
3. quick guide
This chapter covers the following topics:
Structure of the oscilloscope Introduction to the user guidance Testing before
commissioning Function test Probe compensation Probe attenuation Safety when
using the scanner Carrying out the self-calibration Introduction to the
vertical system Introduction to the horizontal system Introduction to the
trigger system Introduction to touch screen operation
Structure of the oscilloscope When you receive a new oscilloscope, the first
thing you should do is familiarise yourself with its control panel. This
chapter provides a simple description of the operation and functioning of the
oscilloscope’s control panel so that you can quickly become familiar with its
use.
Front
The oscilloscope has a simple control panel with rotary knobs and function
keys that can be used to set the various functions for performing basic
operations. The functions of the knobs are very similar to those of other
oscilloscopes. The 5 keys (F1 ~ F5) to the right of the screen or in the row
below the screen (H1 ~ H5) are menu selection keys that allow you to set the
different options for the current menu. The other keys are function keys that
allow you to enter different function menus or access a specific function
directly.
1. Display range 2. Control (push buttons and rotary switch) Range 3. Probe
compensation: Measuring signal (5 V / 1 kHz) Output 4. EXT trigger input 5.
Signal input sockets (2 to 4, depending on model) 6. Copy button: Saves
waveform directly 7. USB host port: For connecting an external storage medium
8. Power button ON/OFF: Red = unit switched off ; Green = unit switched on
Below the display are the H1 – H5 keys for operating the lower menu row. To
the right of the display are the F1 – F5 keys for controlling the
(superimposed) menu control on the right side of the screen.
– 6 –
F1 F2 F3 F4 F5
H1 H2 H3 H4 H5
Function keys
Switch off left/right menu Right display menu
Lower display menu
– 7 –
Back cover
1. Handle 2. Ventilation slots 3. Multimeter sockets 4. Mains voltage socket
5. Fuse 6. Feet 7. VGA connection: connection of an external monitor 8. LAN
connection: For connection to a network 9. USB device connection: For
connection to the PC 10. Anti-theft device: Opening for fastening 11. AV Port:
Signal output socket (optional) 12. Triger Out (P/F) Port: Trigger signal
output & Pass/Fail connection 13. Out 1 Port: Output socket for function
generator
– 8 –
Controls 2-channel
1. Function key area: Main control for all function menus 2. Additional
functions (depending on model): Menu keys e.g. for the function generator 3.
Trigger control: Rotary knob and menu buttons for trigger control 4.
Horizontal range: Rotary and push buttons for the horizontal range 5. Vertical
range: Rotary and push buttons for channel 1 and channel 2, as well as MATH
function key 6. Reset button: Resets measurement function to standard 7. Print
button: Prints screenshot 8. Counter button: Frequency counter function 9. DMM
button: digital multimeter function 10. Arrow keys: Align cursor to parameter
11. M-rotary knob: Multifunctional rotary knob for navigation in all main
menus
– 9 –
Controls 4-channel
1. Function key area: Main control for all function menus 2. Additional
functions (depending on model): Menu keys e.g. for the function generator 3.
Trigger control: Rotary knob and menu buttons for trigger control 4.
Horizontal range: Rotary and push buttons for the horizontal range 5. Vertical
range & channel selection keys: Keys for all channels, plus MATH function key
6. Quick selection buttons (model dependent): Reset, Print or Decode 7. Arrow
keys: Align cursor to parameter 8. M-rotary knob: Multifunctional rotary knob
for navigation in all main menus
– 10 –
Introduction to user control
Example: 2 channel model 1. Waveform display area 2. RUN/STOP: Can also be
used via touch screen 3. Trigger status 4. Activate touch screen function menu
5. Blue line shows cursor position A and B (only with cursor measurement) 6.
The [T] mark shows the horizontal trigger position 7. Displays the trigger
position in the memory length 8. Shows the current trigger value and the
location in the internal memory 9. Lock touch screen operation ( ) / enable (
) 10. Shows the current time (see configuration) 11. WiFi is activated (See
Connection to the Android device) 12. Indicates the connection of an external
USB device to the oscilloscope 13. Shows battery status (optional) 14.
Multimeter window 15. Arrow shows the trigger level position 16. Waveform of
CH1 17. The blue line shows the horizontal cursor position (only with cursor
measurement) 18. Waveform from CH2 19. Shows the activated measuring function
together with the corresponding measured value. 20. Shows the current trigger
type 21. Shows the currently active channel for function control in the lower
menu 22. Shows the memory depth 23. Shows the frequency of the trigger signal
24. Shows the current sampling rate 25. Shows voltage division, zero position
and bandwidth limit, as well as symbol for coupling type 26. Displays the Main
Time Base 27. Displays values of the cursor measurement 28. Blue pointer
indicates the zero point line of CH2, if not present the channel is not active
29. Yellow pointer indicates the zero point line of CH1, if not present the
channel is not active
– 11 –
Testing before commissioning It is recommended that after receiving a new
oscilloscope, a check of the unit is carried out as follows:
1. check if the unit has been damaged during transport. If you find that the
cardboard packaging or the protective foam pads are badly damaged, save them
until the entire unit and its accessories have passed the electrical and
mechanical test. 2. checking the accessories The accessories supplied are
described in Appendix B “Accessories” of this manual. Check the accessories
against this description to ensure they are complete. If any accessories are
missing or damaged, please contact your dealer. 3. checking the unit If you
notice any damage to the exterior of the unit or if the unit does not function
properly or does not pass the performance test, please contact your dealer. If
the unit has been damaged during transport, please keep the outer packaging
and also inform your dealer about the damage. Function test Check the proper
functioning of the meter as follows 1. connect the mains cable to a power
source. Switch on the unit with the mains switch ” “. The unit performs a
self-test and displays the PeakTech logo. First press the Utility button and
then the H1 button to access the Function menu. Select “Calibrate” using the
multi-rotary knob and press the H3 button to select “Factory Set”. The default
value for the probe attenuation in the menu is 10X. Set an attenuation of 10x
on the probe and connect the probe to the CH1 socket. Align the slot on the
probe with the BNC connector of channel 1 and turn the probe clockwise to
secure it. Connect the probe tip and earth terminal to the plug of the task
head compensator. 3. press the “Autoset” button. The square wave signal with a
frequency of 1 KHz and a 5V SS value is displayed after a few seconds
– 12 –
Probe compensation When you connect the probe to an input channel for the
first time, you must adapt the probe to the input channel. An uncompensated or
incorrectly compensated probe will result in measurement errors. Perform the
probe compensation as follows:
1. Set the damping factor of the probe to 10X in the menu, also set the
switch on the probe to 10X and connect the probe to channel 1. When using the
hook tip, make sure that it remains securely connected to the probe. Connect
the probe tip to the signal connector of the probe compensator and connect the
clamp of the reference cable to the earth terminal of the probe compensator;
then press the AUTOSET button.
2. Check the displayed waveforms and adjust the probe until correct
compensation is achieved
Overcompensated Correct compensationUndercompensated 3. Repeat the process if
necessary to obtain the most uniform image possible.
– 13 –
Setting the probe damping factor
The probe has several probe damping factors that affect the vertical scaling
factor of the oscilloscope.
If the set probe attenuation factor is to be changed or checked, press the
function menu key for the respective channel and then the selection key
corresponding to the probe until the correct value is displayed.
This setting remains valid until it is changed again.
Note: The damping factor of the probe in the menu is preset to 10X at the
factory.
Make sure that the value set on the attenuation switch of the probe
corresponds to the attenuation value set on the oscilloscope.
The values that can be set with the switch on the probe are 1 X and 10X (see
picture).
Note: When the
attenuator switch is set to 1X, the
probe limits the bandwidth of the oscilloscope to 5 MHz. You must set the switch to 10X if you
want to use the entire bandwidth of the oscilloscope.
Safety instructions for the use of the scanner head
The handle protection ring around the probe handle prevents it from
accidentally reaching over or slipping off and thus from touching the possibly
live metal parts (see Fehler! Verweisquelle konnte nicht gefunden werden.. )
Handle protection Warning: To avoid electric shock, always keep your fingers
behind the safety protection ring of the probe. To protect you from electric
shock, do not touch any conductive metal parts of the probe tip when it is
connected to a power source. Before taking any measurements, always connect
the probe to the oscilloscope first and then connect the earth terminal to the
housing of the DUT.
– 14 –
Self-calibration With the auto-calibration, the oscilloscope can be quickly
set to the optimum state for highly accurate measurements. You can run this
programme at any time, but you must do so if the ambient temperature varies by
more than 5°C. Remove all probes and cables from the input sockets before
performing the auto-calibration. Press the “UTILITY” button, then the “H1”
button to enter the FUNCTION menu; turn the “Multipurpose” knob to select
“Adjust”. Press the “H2” menu selection button to access the “Selfcal” option
and start the programme after confirming “OK” that all settings are correct.
Introduction to the vertical system
The following picture shows the buttons and keys for the VERTICAL CONTROL. The
following exercises will familiarise you with the vertical control step by
step.
1. With the rotary knob
“VERTICAL POSITION” you can
display the signal in the
centre of the waveform window. With
the
“VERTICAL
POSITION” adjustment knob, you set
the vertical display position of the signal. Turning the “VERTICAL POSITION” adjustment knob moves
the pointer of the channel’s zero position up and down, following the waveform.
Measurement capabilities
If DC coupling is set for the channel, you can quickly measure the DC
component of the signal by observing the difference between waveform and
signal ground.
If AC coupling is set for the channel, the DC component is filtered out. This
mode helps you to display the AC component of the signal with higher
sensitivity.
– 15 –
Set the vertical offset to 0 again: Turn the VERTICAL POSITION knob to change
the vertical position of the channel and press the VERTICAL POSITION knob to
reset the vertical position to 0. This is especially useful if the position
track goes far out of the display area and the signal should immediately
reappear in the centre of the screen. 2. Change the vertical setting and
observe the resulting change in status information.
With the status information displayed at the bottom of the wave window, you
can see changes in the vertical scaling factor for the channel.
Turn the vertical “VOLT / DIV” knob and change the “Vertical scaling factor”
(voltage range). Now you can see that the scaling factor of the channel has
been changed according to the status bar.
Press the CH1 ~ CH2 and Math keys. In the operating menu, symbols, waveforms
and scaling factor status information of the corresponding channel are
displayed on the screen.
You can also change the vertical system via touchscreen. Refer to the
paragraph “Adjust the vertical system via touchscreen”. Introduction to the
horizontal system The following picture shows a button and two adjustment
knobs for HORIZONTAL CONTROL. The following exercises will familiarise you
with the horizontal control step by step.
Horizontal control panel 1. Use the “Scale” adjustment knob to change the
settings for the horizontal time base; you can then
observe the resulting changes in the status information. Turn the “Scale” knob
to change the horizontal time base; you will then see the corresponding
changes in the “Horizontal Time Base” display in the status line. Use the
“HORIZONTAL POSITION” adjustment knob to adjust the horizontal position of the
signal in the waveform window. The “HORIZONTAL POSITION” adjustment knob is
used to control the trigger shift of the signal or for other applications.
When you use it to trigger the shift, you can observe that the waveform moves
horizontally and follows the rotation of the “Horizontal Position” adjustment
knob. 3. Trigger shift back to 0 Turn the HORIZONTAL POSITION knob to change
the horizontal position of the channel, press the HORIZONTAL POSITION knob to
set the shift back to 0.
4. press the “HOR” key to determine the window section.
– 16 –
Introduction to the Trigger System
Picture 0-12 shows a setting knob and three buttons for the TRIGGER CONTROL.
The following exercises will familiarise you step by step with the settings
for the trigger system.
Trigger control panel
Press the “Trigger MENU” trigger settings.
button to open the Trigger menu. Use the 5 menu items to change the
Use the “TRIG” adjustment knob to change the trigger level settings. Turn the TRIG LEVEL knob and watch the trigger indicator on the screen move up and down as you turn the knob. As the trigger indicator moves, the trigger level value displayed on the screen changes.
Note: Pressing the trigger knob resets the trigger to the zero position.
3. press the “FORCE” key to preset a trigger signal that is mainly applied to the trigger modes “Normal” and “Single”.
Introduction to touch screen operation (depending on model)
You can control various oscilloscope functions on touchscreen models by touching the display. Furthermore, you can always use the keys and knobs on the unit for the same settings.
The touchscreen lock at the top of the screen is for turning the touchscreen function on ( ) and off ( ). Tap the icon to make a change.
Set menu control via touch screen
Select menu item: Tap the menu items in the bottom menu, in the right menu or in the left menu directly on the display.
Select menu items: If there are selectable options that can be switched in the
menu, you can always change the area of the menu item by touching the display,
or press the corresponding key. See
illustration:
– 17 –
Press repeatedly to switch options Scroll through the list: If there is a
scroll bar in the left menu or in the file system window, you can
“swipe” up and down to scroll through the list. Touchable menu windows: Tap on
the symbol and a menu window is displayed in the upper left edge
of the display area. Tapping on the menu item with your finger has the same
function as pressing the corresponding key.
– 18 –
Gesture control in normal mode Select a channel (CH1 to CH4 button): Press the
pointer (yellow/blue) belonging to the channel on the
left side of the display.
Set the trigger level (trigger level rotary knob): “Swipe” at the pointer
(yellow/blue) belonging to the channel on the right side of the display.
Adjust the horizontal position (Horizontal Position knob): “Swipe” left/right
in the display.
– 19 –
Double and single zoom If Double-Zoom is selected in the touchscreen menu
window ( ), you can change the time base in the display area by a horizontal
swiping movement and the voltage division of the current channel by a vertical
swiping movement:
– 20 –
If you select Single Zoom, a control panel for these functions appears when
you touch any point in the main window:
Set Voltage/Division (Vertical Scale knob): Tap the upper left field of the
displayed Single Zoom menu to change the Volt/Div of CH1 (yellow) or the lower
left field to decrease it. For CH2, perform this operation in the right
fields.
Change the horizontal time base (Horizontal Scale knob): Use the left field of
the displayed Single Zoom menu (S) to increase the time base and the right
field (nS) to decrease this value.
– 21 –
Gesture control in Wave Zoom mode Press the HOR button to switch to zoom mode.
The upper half of the display shows the main window and the lower half shows
the zoom mode. Zoom mode is the enlarged display of the main window.
Swipe up/down in the centre to change vertical position
Swipe right up/down to change trigger level
– 22 –
Swipe right/left in the centre of the zoom window to change the horizontal
position. Changing Horizontal/Vertical Settings in Double Zoom Mode
In the zoom window, swipe left/right simultaneously to change horizontal time
base and swipe up/down simultaneously to change vertical volt/division.
– 23 –
Changing Horizontal/Vertical Settings in Single Zoom Mode
Tap in the zoom window to open the menu, then tap the corresponding Tap the
icon to change the associated value.
Other Touchscreen Settings Cursor measurements:
Select here to switch between horizontal and vertical lines
– 24 –
Run/Stop: Double-tap the display area or select the corresponding icons or to
start or stop the waveform display.
Touch Keyboard: To enter words, e.g. when naming files, you can use the
keyboard that appears: Set menu value: In some menus there are sliders for
different values. You can change these by
tapping the respective symbol:
– 25 –
4. User instructions (for advanced users) In the preceding paragraphs, the
user has already been familiarised with the basic functions of the function
areas, keys and buttons on the front panel of the oscilloscope. Based on the
introduction of the previous chapter, the user should already have gained
first insights into changing the oscilloscope settings, selecting and
evaluating the status bars and general operation. The following chapter covers
the following topics:
Set vertical system Set horizontal system Set trigger system Perform sampling
settings Set display system Save and recall Cut waveform and output again
Waveform recording and playback Carry out supporting system settings Perform
automatic measurements Perform cursor measurements Use Autoscale Function Use
executive buttons It is recommended that you read this chapter carefully in
order to be able to use the various measurement functions and other operation
methods of the touch screen oscilloscopes.
– 26 –
Adjust vertical system The VERTICAL functions include 5 menu buttons such as
CH1 CH2 (2CH models) and Math, and 8 rotary controls such as VERTICAL
POSITION, VOLTS/DIV for each measuring channel.
Settings CH1 CH2
Each channel has an independent vertical menu with functions based on that
channel.
Switch waveform display on/off (CH & Math function) Pressing the CH1 CH2 or
Math buttons has the following effect: – If the waveform is switched off, it
is switched on and the channel menu is displayed. – If the waveform is already
switched on, the channel menu is also displayed. – If the waveform is already
switched on and the channel menu is displayed, switch the waveform and the
channel menu off again with this action
Description of the functions / menu items of the channel menu:
Menu function Coupling Inverted
Adjustmentungen DC AC Ground
ON OFF
Description
AC and DC signals are detected DC components are blocked Input signal switched
off Waveform is displayed inverted. Waveform is displayed in the original.
Touch divider
X0,001 to X1000
Adjust this value to the selected (on the probe) probe attenuation value in 1-2-5 steps.
Current measurement
Bandwidthlimit
A/V (mA/V) V/A (mV/A)
Full 20M
Turn the M knob to set the amps / volts ratio. The range is 100 mA / V – 1 KA / V. Ampere / Volt ratio = 1 / resistance value The volt / amp ratio is calculated automatically. Full bandwidth. Bandwidth limited to 20MHz to reduce interference.
– 27 –
1. Set channel coupling For example, to represent channel 1 with a square
wave signal on a DC coupling basis, proceed as follows: (1) Press the CH1
button to display the channel menu. (2) Select Coupling in the lower menu. (3)
Select DC in the right-hand menu. DC and AC components are thus recorded. (4)
Alternatively, select AC in the right-hand menu. DC components are now
blocked.
2. Set probe damping For correct recording of the measured values, it is
absolutely necessary that the probe attenuation selected on the probe has also
been set correctly in the oscilloscope menu. Otherwise, measurement deviations
may occur. If the probe attenuation is set to 1:1, for example, 1:1 must also
be set in the channel menu. Example- Select settings as 10:1 for channel CH1:
(1) Press the CH1 button. (2) Select Touch Divider in the bottom menu and
press the ×10 menu function in the menu that appears on the right.
3. Measure current via voltage drop To measure a current via CH1 through a
voltage drop, e.g. at a 1 resistor, proceed as follows: (1) Press the CH1
button to open the channel menu (2) Select Sample and select MeasCurr as YES
in the right menu to open the “A/V Ratio” menu. Now set the desired value for
the current/voltage ratio using the multipurpose dial or the touch screen
operation. For our example with a 1 resistor, set the A/V ratio to 1.
4. Invert waveform With an inverted waveform, the displayed signal is rotated
180° against the phase of the earth potential. Example- Inverted
representation of channel 1: (1) Press the CH1 key for menu selection. (2)
Activate ON for the Inverted option by pressing the button. (3) Selecting OFF
returns the waveform to normal.
5. Set bandwidth limit For lower measuring frequencies than 20MHz, a
bandwidth limit can be set to filter high-frequency interfering signals.
Signals above 20MHz are now blocked. Example- Activate bandwidth limitation
for channel 1: (1) Press the CH1 button to open the Channel 1 menu. (2)
Activate the bandwidth option in the menu. (3) Select Full in the right-hand
menu to capture the full bandwidth. (4) Alternatively, select 20M to limit the
bandwidth to 20 MHz.
– 28 –
Mathematical function The mathematical functions are used to display measuring channels added, subtracted, multiplied or divided. Alternatively, the FFT function can be activated.
Functional scope of the mathematical functions:
Function menu Factor1
Wfm Math
Sign
Factor2
Vertical Source
Window FFT
Format
Setting CH1 CH2 CH3 CH4
+ – × /
CH1 CH2 CH3 CH4
Div Voltage
CH1 CH2 CH3 CH4 Rectangle Hanning Hamming Blackman Bartlett Emperor
Vrms dB
Description Select signal source factor 1 Selects the mathematical function to
calculate from source 1 to source 2 (e.g. CH1 + CH2) Select signal source
factor 2
Select voltage/division for displayed mathematical waveform with
multicontroller Select channel as FFT source
Select window type for FFT display.
Select Vrms format. dB Select format.
Hori
Hz
Select horizontal position for FFT
Hz/div waveform with multi-controller
Vertical
div
Select vertical position for FFT waveform
v or dB with multi-controller
Example- Add waveforms CH1 & CH2:
1. Press the Math button to display the maths menu. The additional
mathematical waveform (pink) is now shown in the display.
2. Select Dual Wfm Math from the bottom menu. 3. For Factor1, select channel
1 in the left selection menu. 4. Select the addition function + in the right-
hand menu as the mathematical operator. 5. For Factor2, select channel 2 in
the left selection menu. 6. Select Vertical in the right menu; select the menu
several times to change the symbol in the
display to the upper position or lower position. Depending on the position,
change the zero position of the mathematical curve (pink) or the
voltage/division for this curve by pressing the multi-controller.
– 29 –
User defined function Press the Math button to display the Math menu below.
Select User Function in the bottom menu, an expression input keyboard appears.
3. create a representation. When you are finished, select on the keyboard to
confirm. The division of the Math waveform is displayed at the bottom left of
the screen.
FFT function
The FFT (Fast Fourier Transformation) function converts a time-based waveform
into its individual frequency components. This function can be very useful in
evaluating the input signals. You can match these obtained frequencies with
known system frequencies from system clocks, oscillators or voltage supplies,
for example. In the audio range, the FFT function splits any waveform into its
individual components and thus shows the composition of a sound and the
distance ratios of the individual tones in the frequency band, as well as the
average levels. Example- Activate and use the FFT function: 1. Press the Math
button to display the maths function menu. The mathematical waveform (blue) is
displayed. 2. Select FFT from the bottom menu. 3. Select Source in the right-
hand menu; select for channel 1. 4. Select Window from the right-hand menu and
choose a useful window type (see table). 5. Select Format in Vrms or dB. 6.
Select Hori in the right-hand menu; press the menu function several times to
display the symbol above
or below this menu item. This allows you to change the time base of the FFT
wave with the multi-controller depending on the horizontal position. 7. Select
Vertical in the right menu; press the menu function several times to display
the symbol above or below this menu item. This allows you to change the
voltage/division or zero line of the FFT wave with the multi-controller
depending on the position.
– 30 –
Select FFT window
There are six FFT windows. Each window makes trade-offs between frequency resolution and amplitude accuracy. Choose the window based on what you want to measure and the characteristics of your source signal. The following table will help you choose the best window:
Art
Characteristics
Windows
Rectangle (Rectangle)
Hanning
This window is best for frequency resolutions, but is the worst for accurately
measuring the amplitude of these frequencies. It is the best window for
measuring the frequency spectrum of nonrepetitive signals and measuring
frequency components near DC.
Use the rectangular window for measuring transients or peaks where the signal
level before and after the event is almost the same.
Also usable for sine waves with the same amplitude and with fixed frequencies
Broadband noise with relatively slowly varying spectrum. This window is well
suited for measuring amplitude accuracy, but less so for frequency
resolutions.
Use the Hanning window to measure sine, periodic and narrowband noise.
Best suited for transients or peaks where the signal levels before and after the event differ significantly.
Hamming
This is a very good window for frequency resolution with slightly better
amplitude accuracy than the rectangular window. It has a slightly better
frequency resolution than the Hanning window.
Use the Hamming window to measure sine, periodic and narrowband noise.
Blackman
Best suited for transients or peaks where the signal levels before and after
the event differ significantly.
This is the best window for measuring the amplitude of frequencies, but offers
the poorest frequency resolution. Use the Blackman-Harris window for single
frequency signals
and finding higher order harmonics.
Bartlett
The Bartlett window is a slightly narrower version of the triangular windows, with “zero weight” at both ends.
Emperor
The frequency resolution when using the Kaiser window is adequate, the
spectral leakage and amplitude accuracy are both good.
The Kaiser window is best when the frequencies are very close but have very
different amplitudes (the sidelobe level and form factor are close to the
traditional Gaussian RBE). This window is also good for the random signals.
– 31 –
Notes for FFT use The waveform zoom function also works for FFT. Use the dBV
RMS scale for a detailed view of multiple frequencies, even if they have
different amplitudes.
Use the linear RMS scale to compare all frequencies in an overall view.
Signals that contain a DC component or offset can lead to incorrect FFT signal
amplitude values. To
minimise the DC component for the source signal, select AC coupling. To reduce
noise and aliasing in repetitive or single measurement waveforms, set the
oscilloscope’s
acquisition mode to average. Nyquist frequency: The highest frequency that an
oscilloscope that digitises in real time can measure is equal to half the
sampling rate and is called the Nyquist frequency. If not enough sampling
points are acquired and the frequency is higher than the Nyquist frequency,
the phenomenon of the “false waveform” occurs. Therefore, pay more attention
to the relationship between the sampled and measured frequency. Using VERTICAL
POSITION and VOLTS/DIV knobs 1. Use the VERTICAL POSITION knob to change the
vertical position of the waveforms of all channels
(including those created by mathematical calculation). The resolution of this
adjustment knob changes with the vertical division.
2. Use the VOLTS/DIV adjustment knob to set the vertical resolution of the
waveforms of all channels (including those created by mathematical
calculation), which determines the sensitivity of the vertical division in the
order 1-2-5. The vertical sensitivity increases when you turn the adjustment
knob clockwise and decreases when you turn it counterclockwise.
– 32 –
Set horizontal system The HORIZONTAL CONTROLS consist of the HORIZ-MENU button
and adjustment knobs such as HORIZONTAL POSITION and SEC/DIV. 1. adjustment
knob HORIZONTAL POSITION: with this adjustment knob you control the horizontal
positions of all channels (including those created by mathematical
calculation) whose resolution changes with the time base. 2. setting knob
SEC/DIV: with this you set the horizontal scale factor with which you
determine the main time base or window. 3. key HORIZ Menu: Press this key to
display the operating menu on the screen. Enlarge waveform (zoom) Press the
HOR button and the display splits in the middle. The upper half of the display
shows the main window and the lower half shows the zoom window. The zoom
window is a magnified area of the main window.
When Main Window is selected in the bottom menu, Horizontal Position and SEC /
DIV are used to adjust the horizontal position and time base of the main
window. The time base of the zoom window is also changed.
When Zoom Window is selected in the bottom menu, the Horizontal Position and
SEC / DIV are used to adjust the horizontal position and time base of the zoom
window. Hint: If the time base of the zoom window is the same as in the main
window, the interface is automatically switched back to normal mode and the
zoom mode is switched off. You can also zoom in on the waveform using the
touch screen, as described in the associated chapter.
– 33 –
Set trigger system
The trigger determines when the oscilloscope starts acquiring data and
displaying the waveform. Once set correctly, the trigger can convert a
fluctuating display into a meaningful waveform.
When the oscilloscope starts acquiring data, it records enough data to display
the waveform to the left of the trigger point. The oscilloscope continues to
record data while waiting for a trigger condition. When a trigger is detected,
the unit continuously records enough data to display the waveform to the right
of the trigger point.
The trigger control area consists of a 1 rotary knob and 2 menu buttons.
TRIG LEVEL: This knob sets the trigger level. When you press the knob, the level is reset to zero.
Force:
Press this button to create a trigger signal. This function is primarily used with the trigger modes “Normal” and “Single”.
Trigger Menu: The button calls up the trigger control menu.
Single trigger
Trigger control
The unit provides four types of triggers: single trigger, alt trigger, logic
trigger and bus trigger. Each trigger type has different submenus. There are
two ways to get into trigger mode: Key operation: Press Menu in the trigger
field to open the trigger menu. With H1 you can then select the extended
trigger menu, which you can scroll (turn) and select (press) with the
multipurpose knob. Touch screen operation: Press the house icon to open the
touch menu. Select Trig Menu and then the trigger (Single, Alt, Logic, Bus) in
the lower menu. The trigger type can then be selected in the right-hand screen
menu under Type.
Single: Uses a single trigger to display a stable waveform on both channels. Logic Trigger: Triggers a signal according to the conditions of the logic ratio Bus Trigger: Sets bus timing trigger
– 34 –
Trigger Brief description The single, logic and bus trigger menus are described below: Edge trigger: Occurs when the trigger input passes through a specific voltage level with the specified slope.
Video Trigger: Trigger on fields or lines for a standard video signal.
Slope Trigger: The oscilloscope starts triggering according to the rate of rise or fall of the signal.
Pulse Trigger: Finds pulses with specific widths. Runt Trigger: Trigger pulses that pass through one trigger level but not through the other trigger level.
Windows trigger: Gives a high trigger level and low trigger level. The oscilloscope triggers when the input signal passes through the high or low trigger level.
Timeout Trigger: The oscilloscope triggers when the time interval from the time of the
rising edge (or falling edge) by the trigger level when the adjacent falling edge (or rising edge) by the trigger
level
is
greater
than
the
set
timeout
time.
Nth Edge Trigger: The oscilloscope triggers on the Nth edge that appears on the specified idle time. Detailed trigger description
RS232 Trigger: RS232 is a serial communication mode used in data transfer
between PCs or between PC and terminal.
I2C trigger The I2C serial bus consists of SCL and SDA. The transmission rate
is determined by SCL and the transmission data is determined by SDA.
SPI trigger Trigger the specified data when the timeout is met. When using SPI
Trigger, you must specify the SCL and SDA data sources.
CAN bus trigger CAN (Controller Area Network) is a serial communication
protocol of the international ISO standardisation.
– 35 –
Detailed trigger description:
1. edge trigger (edge) An edge trigger occurs at the trigger threshold of the input signal. Select the edge trigger mode to trigger on the rising or falling edge of the signal.
Edge Trigger menu:
Menu Trigger Mode
Source
Coupling Gradient
Mode & Holdoff
Setting
Flank
CH1 CH2 EXT EXT/5 AC Line AC DC Rising Falling Car Normal Single Holdoff
Reset
Description
Set vertical trigger type as edge triggering
Channel 1 as trigger signal. Channel 2 as trigger signal. External trigger as
trigger signal. 1/5 of the external trigger as trigger signal. AC mains
voltage as trigger signal. Blocks the DC component. Allows all components to
pass through. Trigger on rising edge. Trigger on falling edge Capture waveform
even if no trigger occurs Capture waveform when trigger occurs Capture a
waveform when trigger occurs, then stop. 100ns~10s, use the knob to set the
time interval before another trigger occurs. Set the holdoff time as the
default value (100ns).
Trigger level:
The trigger level shows vertical trigger position of the channel. Turn the
Trigger Level knob or “slide” up or down on the touch screen to move the
trigger level. While adjusting, an orangered dashed line is displayed to show
the “trig” position, and the value of the trigger level changes is shown at
the right corner. After adjustment, the dotted line disappears.
– 36 –
2. video trigger
Select the video mode to trigger on video fields or video lines of NTSC, PAL
or SECAM standard video signals. In video trigger mode, the setting
information is displayed at the bottom right of the screen, e.g.:
indicates that video trigger on CH1 and sync type “even” has been selected.
Video trigger menu: Menu
Trigger Mode Source
Modu
Sync
Fashion Holdoff
Setting Video CH1 CH2 EXT EXT/5 NTSC PAL SECAM Line Field Odd Even Line NO.
Car
Description Set vertical trigger type as video triggering Channel 1 as trigger
signal. Channel 2 as trigger signal. External trigger as trigger signal. 1/5
of the external trigger as trigger signal.
Select video modulation
Synchronous trigger in video line Synchronous trigger in video field
Synchronous trigger in odd video field. Synchronous trigger in straight video
field. Synchronous trigger in created video line; set the line number using
the knob.
Capture waveform even if no trigger occurs.
3. slope trigger (slope)
Slope mode allows the oscilloscope to trigger on the rising/falling edge of a
signal within a specified time period. In slope trigger mode, the setting
information is displayed at the bottom right of the screen, e.g.:
indicates that slope trigger is selected on CH1, slope rising and the
difference between the up-level and low-level threshold is 0.00mV.
Slope Trigger Menu: Menu
Trigger mode Source
When
Threshold &Slew Rate
Fashion Holdoff
Setting Slope CH1 CH2 slope
High level Low level Slew Rate
Car Normal Single Holdoff
Reset
Description Set vertical trigger type as slope triggering Channel 1 as trigger
signal. Channel 2 as trigger signal.
Set flank
Set the slope condition; set the time using the Multi ( ) knob.
Adjust the high level using the G knob. Adjust the low level using the G knob.
Rise rate = (High level – Low level)/ settings Capture waveform even if no
trigger occurs Capture waveform when a trigger occurs. When trigger occurs
capture a wave and then stop it 100ns~10s, use the Multi ( ) knob to set the
time interval before another trigger occurs. Set the holdoff time as 100ns.
– 37 –
4. pulse width trigger
The pulse trigger lets the oscilloscope trigger according to the pulse width
of the signal. Unusual signals can be detected by adjusting the pulse width
conditions.
In pulse trigger mode, the setting information is displayed at the bottom
right of the screen, e.g.: indicates that pulse trigger on CH1 with DC
coupling has been selected and the polarity
is positive and the trigger level is 0.00mV.
Pulse Trigger menu: Menu
Trigger mode Source
Coupling
When
Setting
Pulse CH1 CH2 AC
DC
Description
Set vertical trigger type as pulse triggering
Channel 1 as trigger signal. Channel 2 as trigger signal. Blocks the DC component.
Allows all components to pass through.
Select polarity Select the pulse width conditions with the Multi ( ) knob
or touch +/- for the time setting and cursor.
to move the
Fashion Holdoff
Car Normal Single Holdoff
Reset
Capture waveform even if no trigger occurs Capture waveform when a trigger occurs. When trigger occurs capture a wave and then stop it 100ns~10s, use the Multi ( ) knob to set the time interval before another trigger occurs. Set the holdoff time as 100ns.
5. runt trigger
With the runt trigger, pulses are detected that run through one trigger level
but not through another, as shown in the graphic.
In run-trigger mode, the setting information is displayed at the bottom right
of the screen, e.g.: indicates that the run-trigger on CH1 has been selected
with positive polarity and the
difference between the up-level and low-level threshold is 0.00mV.
– 38 –
Runt Trigger Menu: Menu
Trigger mode Source
Threshold
Condition
Setting
Runt CH1 CH2 Up Level Low Level
Polarity
Description
Set Vertical Trigger Type as Runt Triggering
Channel 1 as trigger signal. Channel 2 as trigger signal.
Select the level setting with the Multi ( ) rotary knob or
tap +/- via touch screen for threshold setting
Positive polarity: The unit triggers on the positive runt pulse. Negative polarity: The unit triggers on the negative runt pulse. Select the pulse width conditions with the Multi ( ) knob
or touch +/- for pulse width and the desired digital digit.
to move the cursor to
Triggers when runt pulse is greater than set pulse width.
Triggers when the runt pulse is equal to the set pulse width.
Triggers when runt pulse is smaller than set pulse width.
Fashion Holdoff
Car Normal Single Holdoff
Reset
Capture waveform even if no trigger occurs Capture waveform when a trigger occurs. When trigger occurs capture a wave and then stop it 100ns~10s, use the Multi ( ) knob to set the time interval before another trigger occurs. Set the holdoff time as 100ns.
– 39 –
Windows Trigger
Provides a high and a low trigger level, whereby the oscilloscope triggers when a signal passes through the high or low trigger level.
In Windows trigger mode, the setting information is displayed at the bottom
right of the screen, e.g.:
indicates that the Windows trigger on CH1 has been selected with positive
polarity and the difference between the up-level and low-level threshold is
0.00mV.
Windows Trigger Menu:
Menu Trigger mode
Source Threshold
Condition
Setting
Windows CH1 CH2
Up Level Low Level
Polarity
Description Set Vertical Trigger Type as Windows Triggering Channel 1 as
trigger signal. Channel 2 as trigger signal. Select the level setting with the
Multi ( ) rotary knob or tap +/- via touch screen for threshold setting
Positive polarity: The unit triggers on the positive Windows pulse. Negative
polarity: The unit triggers on the negative Windows pulse. Enter: Triggers
when the signal enters the specified trigger level range.
Exit: Triggers when the trigger signal leaves the specified trigger level
range.
Fashion Holdoff
Car Normal Single Holdoff
Reset
Time: Triggers when the hold time is greater than the Windows time. Available
is 30ns to 10s. Default setting is 100ns
Capture waveform even if no trigger occurs Capture waveform when a trigger
occurs. When trigger occurs capture a wave and then stop it 100ns~10s, use the
Multi ( ) knob to set the time interval before another trigger occurs. Set the
holdoff time as 100ns.
– 40 –
7. timeout trigger
The unit triggers when the time interval (from when the rising edge (or falling edge) passes through the trigger level to when the adjacent rising or falling edge passes through the trigger level) is greater than the set timeout time.
In timeout trigger mode, the setting information is displayed at the bottom
right of the screen, e.g.: indicates that the timeout trigger on CH1 has been
selected with positive polarity
and the up-level and low-level threshold is 0.00mV.
Timeout trigger menu:
Menu Trigger mode
Source Threshold
Polarity
Configure
Fashion Holdoff
Setting Timeout
CH1 CH2 Up Level Low Level Polarity
Idle Time
Car Normal Single Holdoff
Reset
Description
Set vertical trigger type as timeout triggering Channel 1 as trigger signal.
Channel 2 as trigger signal. Select the level setting with the Multi ( )
rotary knob or
tap +/- via touch screen for threshold setting
Starts timing when the rising edge passes through the trigger level. Starts
timing when the falling edge passes through the trigger level. Sets the idle
time. This means the minimum time of an idle time before the trigger
conditions can be met. Selectable is 30ns-10s, default is 100ns.
Capture waveform even if no trigger occurs Capture waveform when a trigger
occurs. When trigger occurs capture a wave and then stop it 100ns~10s, use the
Multi ( ) knob to set the time interval before another trigger occurs. Set the
holdoff time as 100ns.
– 41 –
8 Nth Edge Trigger
The oscilloscope triggers on the Nth edge that appears after a specified idle
time. As shown in the diagram, the unit shall trigger on the second falling
edge after the specified idle time P1/P2/P3/P4
In Nth edge trigger mode, the setting information is displayed at the bottom
right of the screen, e.g.: indicates that the trigger on CH1 has been selected
as edge trigger and the up or low
level threshold is 0.00mV.
Menu Trigger mode
Source Edge
Configure
Fashion Holdoff
Setting Nth Edge
CH1 CH2 Edge
Idle Time
Edge Num
Car Normal Single Holdoff
Reset
Description
Set Vertical Trigger Type as Nth Edge Triggering Channel 1 as trigger signal.
Channel 2 as trigger signal. Triggers on the rising edge of the input signal
when the voltage level hits the specified trigger level. Triggers on the
rising edge of the input signal when the voltage level hits the specified
trigger level. Sets the idle time. This means the minimum time of an idle time
before the trigger conditions can be hit. Selectable is 30ns-10s, default is
100ns.
Sets the edge number value of “N” from Nth Edge for triggering.
Capture waveform even if no trigger occurs Capture waveform when a trigger
occurs. When trigger occurs capture a wave and then stop it 100ns~10s, use the
Multi ( ) knob to set the time interval before another trigger occurs. Set the
holdoff time as 100ns.
– 42 –
Logic Trigger
Trigger regarding the logic relation. In logic trigger mode, the setting
information is displayed at the bottom right of the screen, e.g.:
indicates that the trigger is in logic mode AND, CH1 is 2.00V as high trigger
level and CH2 is 0.00mV as low trigger level.
Menu
Setting
Description
Trigger mode Logic Mode Input Mode
Out Mod
Fashion Holdoff
Logic AND OR XNOR XOR CH1
CH2
Goes True Goes False
Is True> Is True< Is True=
Car Normal Single Holdoff
Reset
Set Vertical Trigger Type as Logic Triggering Set logic mode as AND Set logic mode as OR Set logic mode as XNOR Set logic mode as XOR Sets CH1 as high level, low level high or low level, as well as rising or falling Sets CH2 as high level, low level high or low level, as well as rising or falling Trigger when condition changes from False to True Trigger when condition changes from True to False Triggers when True time is higher than Setting Triggers when True time is less than Setting Triggers when True time is equal to the time of the setting Capture waveform even if no trigger occurs Capture waveform when a trigger occurs. When trigger occurs capture a wave and then stop it 100ns~10s, use the Multi ( ) knob to set the time interval before another trigger occurs. Set the holdoff time as 100ns.
Note: If one channel is set as “Rise” or “Fall”, the other channel cannot also be set as “Rise” or “Fall” at the
same time
– 43 –
Bus Trigger 1. SPI
Trigger on specified data when timeout conditions are met. When using the SPI
trigger, SCL and SDA data must be specified.
In SPI trigger mode, the setting information is displayed at the bottom right
of the screen, e.g.: indicates that the trigger is in SPI mode, CH1 as trigger
level is 0.00mV
and CH2 as trigger level is 0.00mV.
Menu Bus Type
Source Timeout
ClockEdge & Data
Fashion Holdoff
Setting SPI CH1 CH2
Time out
Clock Edge
Description
Set vertical bus type as SPI triggering Set CH1 as SCL or SDA Set CH2 as SCL
or SDA Set the minimum time that SCL must be inactive. A range of 100ns~10s is
available before the oscilloscope starts searching for the measurement data
(SDA) to be triggered. Use the Multi ( ) knob or the touch control to set. Set
Clock Edge as a rising or falling edge. Thus, the SDA data is sampled on the
rising or falling edge.
Data Bits
Current Bits Data
All Bits Car
Normal Single
Set the number of the serial data bit string. This value can be set between 4
and 32. Use the Multi ( ) rotary knob or the touch control. Set the number of
the data bits from 0-31.
Use the Multi ( ) knob or touch control to set the Current Data Bit value as
H, L or X(H or L). Set all data bits as specified values.
Capture waveform even if no trigger occurs Capture waveform when a trigger
occurs. When trigger occurs capture a wave and then stop it
– 44 –
2. I2C trigger
The I2C Serial Bus consists of SCL and SDA. The transmission rate is
determined by SCL and the transmission data by SDA. As shown in the picture,
the oscilloscope can be triggered to Start, Restart, Stop, Ack Lost, a
specific device address or a data value.
In I2C trigger mode, the setting information is displayed at the bottom right
of the screen, e.g.: indicates that the trigger is in I2C mode, CH1 as trigger
level is 0.00mV and
CH2 as trigger level is 0.00mV.
Menu Bus Type
Source
When
Fashion Holdoff
Setting I2C CH1 CH2
Start
Description
Set vertical bus type as I2C triggering Set CH1 as SCL or SDA Set CH2 as SCL
or SDA Triggers when SDA data passes from High to Low while SCL is High
Restart
If another start state occurs before a stop state
Stop
Ack Lost
Address Addr Bits Address
Adr Format Direction
Data
Data Format
Byte length Current
Bit Data All Bits
Addr / Data
Car Normal Single
Triggers when SDA data passes from Low to High while SCL is High Triggers when
SDA data is “High” during a confirmation of the SCL clock position.
Triggers a read or write bit when the set address is hit.
Set bit address aud 7, 8 or 10
Set address according to the set bit address. Address
range is 0-127, 0-255, 0-1023
Set data direction to Read or Write
Note: If address bit is set to 8, this is not available
Searches for the preset data value of SDA and triggers on the falling edge of
the SCL, on the last bit of the data range Set data byte length, available is
1-5. Use the multi ( ) knob or the touch control to set the byte length.
Select data bit, range from 0 to (byte length *8-1). Set data to H, L or X (H
or L). Set all data bits as specified values. Triggers when address and data
conditions are met at the same time Capture waveform even if no trigger occurs
Capture waveform when a trigger occurs. When trigger occurs capture a wave and
then stop it
– 45 –
3. RS232 trigger
RS232 is a serial communication type used for data transmission between PC and
terminal. A character is transmitted as a frame of data consisting of 1 start
bit, 5-8 data bits, 1 check bit and 1-2 stop bits.
In RS232 trigger mode, the setting information is displayed at the bottom
right of the screen, e.g.: indicates that the trigger is in RS232 mode and the
trigger level of CH1 is 0.00mV.
Menu Bus Type Input
When
Fashion Holdoff
Setting RS232 Source
Polarity
Start
CH1 CH2 Normal Inverted
Error
Chk Error
Data
Start
Error
Chk Error
Data Car Normal Single
Description
Set vertical bus type as RS232 triggering Set CH1 as SCL or SDA Set CH2 as SCL
or SDA
Set the polarity of the data transmission as Normal Set polarity of data
transmission as Inverted
Triggers on the start frame position. After selecting this condition, select
Configure for detailed setting options.
Triggers on the error frame position. After selecting this condition, select
Configure for detailed setting options. Trigger when a Chk Error was found.
After selecting this condition, select Configure for detailed setting options.
Triggers on the last bit of preset data. After selecting this condition,
select Configure for detailed setting options. Common Baud: Multi ( ) to
select common baud Custom Baud: Multi ( ) knob to select custom baud, range
from 0 to 10000000 possible. Stop bit: Select 1 or 2 Parity: “No”, “Even” or
“Odd Common Baud: Multi ( ) to select common baud Custom Baud: Multi ( ) knob
to select custom baud, range from 0 to 10000000 possible. Even-Odd: Select
Even or Odd Common Baud: Multi ( ) to select common baud Custom Baud: Multi (
) knob to select custom baud, range from 0 to 10000000 possible. Data Bits:
Set 5, 6, 7 or 8 bits Data: Set associated data bits from 0-31, 0-63, 0-127 or
0255. Capture waveform even if no trigger occurs Capture waveform when a
trigger occurs. When trigger occurs capture a wave and then stop it
– 46 –
4. CAN bus trigger
CAN (Controller Area Network) is a serial communication protocol of the international ISO standardisation. When using the CAN bus trigger, you can trigger on Frame Start, Frame Type, Identifier, Data, ID & Data, Frame End, Missing Acknowledgement or Bit Fill Error. You must specify the signal source, trigger signal type, sample point and signal rate of the CAN signal. In CAN bus trigger mode, the trigger setting information is displayed at the bottom right of the screen, e.g.
Menu
indicates that the trigger type is CAN, the CH1 trigger level is -126 mV
Setting
Description
Bus type
CAN
Sets vertical bus types as CAN triggers
CH1
CH1 as trigger source
CH2
CH2 as trigger source
Source
CH3
CH3 as trigger source
CH4
CH4 as trigger source
Type
CAN_H CAN_L TX RX
Actu. CAN_H Bus Signal Actu. CAN_L Bus Signal Transm. Signal on CAN signal line Receive signal on CAN signal line
Input / Input Condition
Sample Point Common baud
Turn the M knob (or tap on the touch screen) to set the sample point, which is
a point within a bit time. The oscilloscope samples the bit level at this
point. “Sample point” is represented by the percentage of “the time from the
beginning of the bit time to the sample point time” in the “bit time”. The
bandwidth is 5% to 95%.
Turn the M button to select the Baud number from the list on the left.
Custom Baud
Start Type
Type
(Lower menu)
ID DATA
ID&Data
Configure (Lower Menu)
Configure (Lower Menu)
Configure (Lower Menu)
Turn the M knob (or tap on the touch screen) to set the baud. The range is 10,000 to 1,000,000. Tip: You can select the closest value in Common Baud and then adjust it in this menu.
Trigger on the start frame of the data frame
Data
Remote
Error
Sets trigger on selected frame
Overload Format
Standard or Advanced
ID
M-knob and arrow keys for selection
Byte Length
Data Format ID Byte Length
M-regulator for setting between 1 and 8 bytes M-knob and arrow keys for
selection
Standard or Advanced
M-knob and arrow keys for selection
M-regulator for setting between 1 and 8 bytes
– 47 –
Fashion Holdoff
Data
M-knob and arrow keys for selection
End
Triggers on end frame of the data frame
Missing Ack Bit Stuffing
Triggers on missing ack (good message) Triggers on Bit-Stuffing Error
Car Normal Single
Capture waveform even without trigger Capture waveform with trigger Capture waveform with trigger, then stop
Operating the function menu
The operating area of the function menu comprises 8 function menu keys:
Measure, Acquire, Utility, Autoscale, Save, Display and Help, as well as 3
instant selection keys: Autoset, Run/Stop, Single.
Sampling settings (sampling function)
Press the Acquire button; select Acyu Mode, Length or PERF Mode to set the sampling function.
Description of the Acqu Mode menu:
Menu
Acqu Mode
Sample Peak Detect
Setting
Description
General scanning mode. Used to detect interference peaks and reduce
interference
Average
41664128
Used to reduce random interference of any kind with an optional number of averages.
Description of the Length menu:
Menu Length
Setting
1000 10K 100K 1M 10M 20M 40M (Single CH)
Description Selection of the recording length
Description of the PERF Mode menu:
Menu
Setting
Description
PERF Mode
8-bit 12-bit
Set vertical resolution (A/D)
If the sample rate is <=250MS/s, the ADC resolution is set to 12-bit by default.
If the sample rate is > 250MS/s, the ADC resolution is set to 8-bit by default.
If the sample rate is <250MS/s and only one channel is activated, the ADC resolution can be set to 8-bit or
12-bit.
Menu
Setting
Description
Interpl
Sinx/x x
Use sine(x)/x interpolation Use linear interpolation
– 48 –
The interpolation method is a processing method to connect the sampled points,
using some points to calculate the whole appearance of the waveform. Select
the appropriate interpolation method according to the actual signal. Sine (x)
/ x-interpolation: Connect the sampled points with curved lines. Linear
interpolation: Connect the sampled points with straight lines. This method is
suitable for reconstructing the straight-line signals, such as the square wave
or pulse wave.
Figure 4-3 Sine(x)/x interpolation
– 49 –
Display menu Settings (in Acquire)
Press the Acquire button; select Wave Display, XY Mode or Cymometer as the setting. Description of the Acquire menu:
Menu Type
Persist
XY Fashion Counter Clear
Setting
Dots Vect
OFF 1 sec. Time 2 sec. 5 sec. Infinity ON OFF ON OFF
Description Only the sample points of the waveform are displayed. The space
between the sampling points is connected with a line.
Persist adjusts the afterglow of the waveform. Use the multi-rotary control (
) to change the value settings. “Infinity” means maximum persistence.
Turns on the XY function. Switches the XY function off. Switches frequency
counter ON and OFF
Clears the afterglow waveforms from the display and restarts the afterglow.
Display: Press the Display menu selection button. In the lower selection menu, select Type and press the button to switch between Dot and Vect.
Afterglow (Persist ): With the Persist function, you can simulate the afterglow effect of a tube oscilloscope: the stored original data is faded, the new data is displayed in vivid colour.
(1) Press the display button (2) Select Persist from the bottom menu (3) Use
the right menu buttons or the touch screen operation to change the afterglow
of the display between
OFF, 1second, 2second, 5second and Infinity. If you select Infinity, the
waveform will continue to be overlaid until you change this setting again or
press the Clear button. Select OFF to turn off the afterglow and “clean” the
display of the superimposed waveforms. (4) Select Clear to clear the already
displayed overlaid waveforms from the display and start the afterglow again
Colour The colour temperature function uses colour correction to indicate
frequency of occurrence. The hot colours like red / yellow indicate frequently
occurring events and the colder colours like blue / green indicate rarely
occurring events.
(1) Press the display key. (2) Select Persist & Colour from the bottom menu. (3) Select Colour in the right menu, choose between ON / OFF.
– 50 –
XY Mode: XY mode is used to display the amplitude from one waveform against
the amplitude from another. The data point from the first waveform sets the
horizontal position, while the corresponding data point from the second
waveform shows the vertical position for each point. The oscilloscope is in
untriggered sampling mode: the data is displayed as bright dots. How to use
the knobs and buttons: The Vertical Scale and Vertical POSITION adjustment
knobs for Channel 1 are used to adjust the
horizontal scale and position of the yellow waveform. The Vertical Scale and
Vertical POSITION adjustment knobs for channel 2 are used to adjust the
vertical
scale and position of the blue waveform.
The following functions cannot be activated in XY mode Reference or digital
waveform Cursor Trigger control FFT
Application steps: 1. Press the display key 2. Select XY Mode as ON or OFF in
the lower menu Frequency counter: This is a 6-digit frequency counter. This
can measure frequencies from 2Hz to full bandwidth. However, it can only
measure the frequency accurately if the measured channel has a trigger signal
and is in edge mode of the single trigger mode. The counter is displayed at
the bottom right of the screen. 2 CH:
4CH:
Application steps: 1. Press the trigger menu button and set it to Single, the
trigger mode to Edge and select the source to be
measured. 2. Press the display key. 3. Select the counters as ON or OFF
– 51 –
Save and recall waveform
Press the Save button to open the Save menu at the bottom of the screen. Here you can save waveforms, configurations, screenshots or record waveforms as a movie.
Menu
Setting Wave
Description Select memory type
Configure
Configuration
Type
Image
Create screenshot
Record
Record waveform as a film
Clone
Clone waveform between cursors on generator
If the type is Wave, the menu shows the following:
Source
CH1 – CH4 Math
Select the waveform to be saved
Object
0-99
Select the address/memory number where the waveform is to be stored or retrieved
Object & Show
Show
ON OFF
Recall or close the waveform stored in the currently selected address. When “Show” is ON, the waveform stored at the address is displayed, the associated address number and relevant information is shown in the top left of the display. If the memory address is empty, a message “None is saved” is displayed.
Close All
Closes all waveforms that are stored under the object address.
Save
Saves the waveform at the selected address. Regardless of the type selected in the Save menu, you can always save the current waveform directly as a BIN file with the Copy button, without having to take a diversion via the Save menu.
Storage
Internal External
Saves to internal or external memory (USB). If the file is saved to an external USB memory, you can set the file name yourself and call up the saved file via the PC software supplied.
If the type is Configure, the menu shows the following:
Setting 1
The address of the settings to be saved
Configure
…..
Setting 8
Save
Saves the current oscilloscope configuration at the selected memory address.
Load
Recalls the saved configuration from the selected memory address
If the type is Image, the menu shows the following:
Save
Saves the current screen completely with all displays as a screenshot. This is only possible on an external memory, so please connect a USB memory first. Data format is BMP
– 52 –
Waveform recording The oscilloscope can store 100 waveforms, which can be
displayed again simultaneously with the current waveform. The recalled
waveform cannot be adjusted subsequently, but remains in the form in which it
was recorded. For example, to save a waveform from CH1 to memory address 1,
proceed as follows: 1. Press the Save button 2. Save: Select Type in the lower
menu and use the M-rotary knob to go to Wave in the left menu. 3. Select
Source in the lower menu and then CH1 in the right menu to select channel 1 as
the source. 4. Select Object & Show from the lower menu, then turn the M knob
to select 1 as the memory address
from the left menu. 5. Then select Storage from the bottom menu and Internal
from the right menu. 6. Then select Save in the lower menu to save the
waveform. 7. Recall: Select Object&Show in the lower menu, then select the
location 1 in the left menu. Then select
Show as ON so that the waveform saved under 1 is displayed. Address number and
all other relevant information are shown in the top left of the display. Tip:
Regardless of the type selected in the Save menu, you can always save the
current waveform directly as a BIN file by pressing the Copy key without
having to go via the Storage menu. If the storage has been set to external,
make sure that there is also an external storage medium attached to the
oscilloscope. Please carry out the setup of the USB storage as shown in the
following chapters.
Save screenshot A screenshot can only be stored on an external USB memory. 1.
Install a USB memory: Insert a USB stick into the “USB Host Port” of the unit.
If a USB symbol is
displayed at the top right, the USB memory has been correctly recognised and
connected. If the USB memory is not recognised correctly, please proceed as
described in the corresponding chapter. 2. After the USB memory has been
connected, press the Save button so that the Save menu is displayed at the
bottom of the screen. 3. Select Type in the lower menu and turn the M knob
until Image appears. 4. Press the Save button to save the screenshot. A
virtual keyboard opens, which you can operate with the M rotary knob or via
the touch screen function. Enter a file name with a maximum of 16 characters
and confirm with the Enter key .
USB memory requirements The supported file format of a USB memory is: FAT21
file system with a cluster size of no more than 4KB. A USB mass storage device
is also supported. If a connected USB memory does not work, format it as
described in the following two options: The system tool or a formatting tool.
(8Gbyte or larger USB memory sticks can be formatted with the second method).
– 53 –
System’s own function for formatting 1. Connect the USB stick to the PC. 2.
Right-click on Computer Manage (Win7) or right-click on the Windows icon
(Win10) and then on
Disk Management. 3. In the Disk Management menu you will find all the
information about the connected data carriers, Select
the USB storage device as marked in red in the following example 1 and 2:
4. Right-click on the area marked in red and select Format and a warning
message appears, which you confirm with YES.
5. Set the file format as FAT32 with a cluster size of 4096. Select quick
format “Perform a quick Format” and confirm with OK and confirm with YES.
6. Formatting process
– 54 –
7. Check that the formatting has been carried out and that FAT32 is now displayed with a cluster size of 4096.
Cloning a waveform
To clone a waveform that lies between the cursors, please refer to Appendix C.
Saving and playing back recordings
Waveform recording can record the current waveform and save it as a movie. You can set the interval between 1ms and 1000s, with a maximum of 1000 frames recorded. You can save the recording internally or externally.
For internal storage, you can choose between 4 options: OFF, Record, Playback and Storage.
For external storage, you can choose between 2 options: OFF and Record.
Record: To record a waveform until the set interval reaches the end frame. The Record menu is displayed as follows:
Menu Fashion
Record Mode Frame Set
Refresh Operate
Setting OFF
Record Playback Storage End Frame
Interval
ON OFF Play Stop
Description Close Wave Record Set Record Menu Set Playback Menu Set memory menu Turn the M-rotary knob to set the number of the recording frames (1-1000). Turn the M-rotary knob to set the interval of the recording frames (1ms-1000s). Renew waveform during recording No renewal Start recording Stop recording
Hint: Both waveforms – channel1 and channel2- are recorded. If a channel is
switched off during recording, this channel is invalid during playback.
Playback: Playback plays back the recorded waveforms The playback menu is
displayed as follows:
– 55 –
Menu Playback Mode
Frame Set
Play Mode Operate
Setting Start Frame
End Frame
Cur Frame
Interval
Loop Once Play Stop
Description Turn the M-rotary knob to set the number of the start recording frame (1-1000). Turn the M-rotary knob to set the number of the end recording frame (1-1000). Turn the M-rotary knob to set the number of the current recording frame (1-1000). Turn the M-rotary knob to set the interval of playback (1ms-1000s) Playback is repeated as an endless loop Playback only once Starts playback Stops playback
Storage: Saves the current waveform recording with respect to the defined start and stop frames.
The storage menu is displayed as follows:
Menu Storage Fashion Frame Set
Save Load
Setting Start Frame
End Frame
Description Turn the M-rotary knob to set the number of the start recording frame (1-1000). Turn the M-rotary knob to set the number of the end recording frame (1-1000). Saves selected waveform recording to internal memory Loads selected waveform recording from internal memory
Use the waveform pick-up as described below:
1. Press the Save button. 2. Select Type in the lower picture menu, select
Record with the M-rotary knob or touch screen. 3. Select Mode in the lower
picture menu and OFF in the right menu. 4. In the bottom menu, select Storage
as Internal. 5. Select Mode in the bottom menu and then Record in the right
menu. 6. Then select FrameSet in the bottom menu, set the EndFrame and the
Interval in the right menu. 7. Press Refresh in the lower menu. 8. Then select
Operate as Play. 9. Select Mode in the lower menu, then Playback in the right
menu. Set FrameSet and Playmode, then
Operate as Play. 10. To save a waveform image, select Mode in the lower image
menu, then Storage in the right menu. Select
FrameSet in the lower menu to select the range of frames, which are then saved
via Save. 11. To load the waveform from the internal memory, select Load in
the lower menu and then Playback of
the mode to analyse the waveform.
Note: Sampling, triggering and display functions are not available during waveform playback. If the storage medium is set to external, only two modes are available: OFF and Record.
– 56 –
The Record menu (external memory) is displayed as follows:
Menu Fashion Record mode FrameSet
Refresh Operate
Setting OFF
Record End Frame
Interval
Infinity ON OFF Play Stop
Description Closes shaft form receptacle Sets recording menu Turn the M-rotary knob to set the number of the end recording frame (1-1000). Turn the M-rotary knob to set the interval of playback (1ms1000s) Recording until the storage medium is full Renew waveform during recording Stop renewal Start recording Stop recording
Hint: Both waveforms (channel 1 & 2) are recorded. If you switch off a channel during recording, the channel is not available during playback.
Use the waveform pick-up as described below:
1. Press the Save button. 2. Select Type in the lower picture menu, select
Record with the M-rotary knob or touch screen. 3. Select Mode in the lower
picture menu and OFF in the right menu. 4. In the bottom menu, select Storage
as External. 5. Select Mode in the bottom menu and then Record in the right
menu. 6. Then select FrameSet in the lower menu, set the EndFrame and the
Interval in the right menu. If you
want to record the waveform without limit, select Infinity in the right menu.
The end frame display is shown as “-“. 7. Press Refresh in the lower menu. 8.
Then select Operate as Play. Connect the external storage medium to your
computer, open the PC software and load the
“wave_record_0.bin” file.
1. Select Communications Auto Player. 2. Transforming the waveforms from the
unit (Transform recording…). 3. Add the transformed files. 4. Set Play Mode
and Time Delay. 5. Click on the green play button to start playback:
– 57 –
Set additional system functions
Config
Press the Utility button and select Function in the lower screen menu, then select Configure in the left menu.
The configuration menu is displayed as follows:
Menu Language
Setting Chinese English
Others
Description Sets the system language. The selectable languages may vary with different firmware versions.
Display
ON Date display Switch on/off
OFF
Set Time
Hour Min
Set hour/minute
Day Month
Set day/month
Year
Set year
KeyLock
Key lock: To deactivate, press the Trigger Menu key and then the Force key. Repeat this three times quickly
About
Show version number and serial number
Display
Press the Utility button, then Function in the lower menu. Then select Display in the left menu.
Use the display menu as shown below:
Menu Backlight
Setting 0% – 100%
Description Turn the M-rotary knob to adjust the display illumination
Graticule
Select the four options to display and adjust the grid line
Battery Menu Time
ON OFF OFF, 5s – 30s
Switch on battery indicator or switch off if battery is installed in the unit.
Set the display duration of the pop-up menus
Adjust
Press the Utility button, then Function in the lower menu. Then select Adjust in the left menu.
Use the Adjust menu as shown below:
Menu
Description
Self Cal
Carries out a self-calibration of the unit
Default
Resets the unit to factory settings
Sample CH
Checks the probe compensation
– 58 –
Self-Calibration
Self-calibration can help to achieve a better measurement result in the event
of faulty measurement or a large influence of ambient temperatures. If the
ambient temperature increases significantly (over 5°C), selfcalibration should
always be carried out to achieve the best possible accuracy.
Before performing a self-calibration, remove all probes from the terminals of
the unit. Press the Utility button, then select Function from the lower menu
and then Adjust. Finally, press the SelfCal button in the lower screen menu to
perform the self-calibration.
Sample Checking Checks whether the probe compensation is OK. The result shows
three possibilities: Excessive compensation, Good probe compensation and
Insufficient probe compensation. Proceed as follows: 1. Connect the probe to
CH1 and set the Probe Attenuation to maximum. 2. Press the Utility button and
select Function from the bottom menu, then Adjust from the left picture
menu. 3. Select ProbeCh. in the lower menu and notes are shown in the display.
4. Press ProbeCh. again and the test starts. The result is then displayed.
Pass/Fail
The pass/fail check detects whether the input signal is within the limits of
the rule. If it exceeds the rule limits, it does not pass the check and is
classified as a “fail”; if it is within the rule limits, it is allowed as a
“pass”. It can also output fail or pass signals via an integrated and
configurable output port.
Description of the Pass/Fail menu:
Menu Operate
Output
Rule
SaveRule
Setting Enable Operate Passport
Fail Beep Stop Info Source Horizontal Vertical Create Number Save Load
Description Activate control Switch on/off Tested signal fulfils rule Checked
signal does not fulfil rule Beep when rule is fulfilled Stops when rule is
fulfilled Show Pass Fail information window Select source CH1, CH2 or Math
Change horizontal tolerance via M-rotary knob Change vertical tolerance via
M-rotary knob
Select the name of the rule between Rule 1 and Rule 8. Select Save to save the
rule Load a rule as a test rule
Pass/Fail Test:
The pass/fail check detects whether the input signal is within the limits of
the rule. If it exceeds the rule limits, it does not pass the check and is
classified as a “fail”; if it is within the rule limits, it is allowed as a
“pass”. It can also output fail or pass signals via an integrated and
configurable output port. To perform a pass/fail check:
1. Press the Utility button, then Function in the lower menu. Select
Pass/Fail in the left menu. 2. Enable switch on: Select Operate in the lower
menu and then Enable in the right menu as ON. 3. Create rule: Select Rule in
the bottom menu, then Source in the right menu and Source selection in the
left menu. Set the Horizontal and Vertical Tolerance in the right menu.
Finally, select Create in the right menu to create the rule.
– 59 –
4. Set output type: Select Output in the lower menu to make the output
settings. Use one/two of the following options: “Pass”, “Fail” or “Beep”.
Pass” and “Fail” are mutually exclusive options that cannot be activated at
the same time. “Stop” means the unit stops when the set conditions are
successfully met.
5. Begin to test: Select Operate in the lower menu, then Operate in the right
menu as Start so that the
test begins.
6. Save Rule: Select SaveRule in the lower menu. Then select the storage
location in the left menu, which you use with Save in the right menu to save
the defined test rule. You can then load and use it again if necessary. Select
Load to recall a saved rule.
Hint:
1. If Pass/Fail is on and you switch to XY or FFT, Pass/Fail is closed and is
no longer available as long as XY or FFT is used.
2. Factory setting closes Pass/Fail, AutoSet and AutoScale. 3. If no more
“save settings” are available, a message “NO RULE SAVED” is displayed. 4.
Under the status “Stop”, the data comparison is interrupted. If the status is
changed back to “Running”,
Pass/Fail will continue to run at the point where it was stopped and will not
restart completely. 5. When the waveform recording is running as playback,
Pass/Fail can be used to analyse the recorded
waveform.
Output
Press the Utility button, then Function in the lower menu. Then select Output in the left menu.
The Output menu sets the behaviour of the Trig Out (P/F) output on the back of the unit.
Use the output menu as shown below:
Menu Type
Setting Trig Level Pass/Fail
AG Output
Description Output synchronous trigger signal Outputs a high level signal on pass and a low level signal on fail. Outputs the signal of the arbitrary waveform generator (AG)
In the VGA menu you can activate the video signal outputs of the unit.
Use the video menu as shown below:
Menu Video
Setting OFF VGA AV
Description Switch off video output Enable VGA output Activate AV output (optional)
Device / Print Setup:
Device and Print Setup menus are described in the paragraph “Print screenshot”
LAN Set: To use the LAN network connection or Wi-Fi (wi-fi) with the computer
or data connection with an Android phone via WiFi, please refer to the related
articles “Connection with the PC”.
Update: Use the USB port on the front to update the unit firmware using a USB
memory device. DAQ: You can use the multimeter data recorder to record the
measurements when measuring current /
voltage with a multimeter.
– 60 –
Automatic measuring functions
Press the Measure button to switch to the menu for the automatic measurement functions.
The oscilloscope has 30 parameters for automatic measurement such as: Period,
Frequency, Mean, PK-PK, RMS, Max, Min, Top, Base, Amplitude,
Overshoot,Preshoot, Rise Time, Fall Time, +PulseWidth, –
PulseWidth, +Duty Cycle, -Duty Cycle, Delay AB , Delay AB , Cycle RMS, Cursor
RMS, Screen Duty, Phase, +PulseCount, -PulseCount, RiseEdgeCnt, FallEdgeCnt,
Area, and Cycle Area….
The “Automatic measurement functions” menu is described below:
Menu Meas Type (left menu)
Add
Source
Add
Remove
Show CH1 / CH2 Snapshot (4 CH models)
Meas Type (left menu)
Remove Remove All
ON OFF ON OFF
Setting
Description Selecting the measuring function
CH1
Selecting the source for the measurement function
CH2
to be applied (2 & 4 CH models)
CH3
Selecting the source for the measurement function
CH4
to be applied (4 CH models)
Adds the measurement function to the
measurement field (displayed at the bottom left),
with a maximum of 8 measurements possible at
the same time
Selects the measurement functions to remove
Removes the selected measuring function Removes all measuring functions Displays all measurement functions for CH1 or CH2 simultaneously Hides the measurement window for all CH1 or CH2 measurements Displays all measurement functions for CH1 to CH4 simultaneously Hides the measurement window for all CH1 to CH4 measurements
Trade fairs A measurement can only be carried out when the channel is switched on. The automatic measurement function cannot be activated in the following three situations:
- With a stored waveform 2) For the Dual Wfm Math waveform (mathematical
function) 3) In video trigger mode.
Period and frequency cannot be measured in scan format.
For example, to measure period and frequency for channel 1, proceed as follows:
1. Press Measure in the lower menu to open the automatic measurement functions. 2. Select Add from the bottom menu. 3. In the right-hand menu, select CH1 under the Source menu. 4. In the left menu, turn the M knob to select the period. 5. In the right menu, press Add to add the period to the measurement field. 6. In the left menu, turn the M-rotary knob to select the frequency. 7. In the right menu, press Add to add the frequency to the measurement field. The measured values are automatically displayed in the lower, left-hand measuring field (red marking).
– 61 –
Automatic measurement of the voltage parameters
The oscilloscope provides automatic voltage measurements including Vpp, Vmax,
Vmin, Vavg, Vamp, Vrms, Vtop, Vbase, Overshoot and Preshoot. The image below
shows a pulse with some voltage measurement points.
Overshoot Max
Vtop
PK-PK
Vamp
Vbase
Min
Preshoot .
Mean: Arithmetic mean value over the entire waveform. PK-PK: Peak-to-peak
tension. Max: Maximum amplitude. The highest positive peak voltage measured
over the entire curve. Min: Minimum amplitude. The highest negative peak
voltage measured over the entire curve. Vtop: Flat-top voltage of the curve,
useful for square/pulse signals. Vbase: Flat-base voltage of the curve, useful
for square/pulse signals. Vamp: Voltage between Vtop and Vbase of a curve.
Overshoot: (Overshoot) Defined as (Vmax-Vtop)/Vamp, useful for square wave and
pulse signals. Preshoot: Defined as (Vmin-Vbase)/Vamp, useful for square and
pulse signals. Cycle RMS: True Root Mean Square voltage over the entire length
of the period of the current waveform. Cursor RMS: True Root Mean Square
voltage over the range of two cursors.
– 62 –
Automatic measurement of the time parameters
The oscilloscope provides automatic measurements of time parameters including
Frequency, Period, Rise Time, Fall Time, +Width, -Width, Delay AB , Delay AB
and Duty Cycle.
The following picture shows a pulse with some timing points:
Rise Time
Fall Time
+Width
-Width
Rise Time: The time it takes for the leading edge of the first pulse in the
curve to rise from 10% to 90% of its amplitude. Fall Time: The time it takes
for the leading edge of the first pulse in the curve to fall from 90% to 10%
of its amplitude. +D Width: The width of the first positive pulse at the 50%
amplitude point. -D Width: The width of the first negative pulse at the 50%
amplitude point. + Duty: + duty cycle, defined as +width/period. – Duty: –
duty cycle, defined as -width/period. Delay AB : The delay between the two
channels on the rising edge. Delay AB : The delay between the two channels at
the falling edge. Duty cycle: Defined as (Width of positive pulse)/(Total
period) Phase: Compares the rising edge of CH1 and CH2, calculates the phase
difference of the two channels.
Phase difference= [(delay between channels of the rising edge) x PI]/period
Other measurement functions
+PulseCount
-PulseCount
+PulseCount reference value.
-PulseCount reference value.
: Displays the number of positive pulses that rise above the centre reference
transition. : Displays the number of negative pulses that are under the centre
reference transition. : Displays the number of positive passes from the low
reference value to the high
: Displays the number of negative passes from the high reference value to the
low
Area
: Calculates the total area of the waveform as volt-seconds. The area above the zero reference
(the vertical offset) is positive and the area below the zero reference is negative. The measurement area is
the algebraic sum of the waveform displayed on the screen.
Cycle Area
: Calculates the range of the first period of the waveform as a volt-second. The area
above the zero reference (the vertical offset) is positive and the area below the zero reference is negative.
The measurement area is the algebraic sum of the first period of the waveform.
Note: If less than a complete period is shown on the display, the measuring range is 0.
– 63 –
Cursor measurements
Press the cursor key to switch on a cursor and show it in the display. Press the key again to switch the cursor off.
Cursor measurements in normal mode
The cursor menu is described below:
Menu
Setting
Description
Type
Line Type (time and voltage type)
Window (Wave Zoom Mode)
Line
Source
Voltage Time
Time&Voltage Auto Cursor
Time Voltage
Main Extension
a b from
CH1 – CH4
Displays the cursor of the voltage measurement Displays the timing cursor Displays cursor of time and voltage measurement Horizontal cursors are displayed as the intersections of the vertical cursors with the waveform Activates vertical cursor Activates horizontal cursors Activates cursor in main window Activates cursor in zoom window (HOR button) Move M-rotary knob to move line a Move M-rotary knob to move line b Move M-rotary knob to move line a and b at the same time, as they are connected Select the channel as source which is used for the cursor measurement.
Follow the steps below to set a time and voltage cursor for channel 1:
1. Press the cursor key to open the menu.
2. Select Source as CH1.
3. Press the first menu field in the lower screen menu (H1) to change the
setting to Time&Voltage. Two blue dash lines for the horizontal voltage range
and two blue dash lines for the vertical time range are now displayed.
4. In the bottom menu, please select LineType as Time to activate the
vertical cursors. If the line is selected as “a” in the lower menu, turn the
M-rotary knob to move the line to the right or left in the display. If “b” was
selected, move this line to the right or left with the M-rotary knob.
5. In the lower menu, please select LineType as Voltage to activate the
horizontal cursors. If the line is selected as “a” in the lower menu, turn the
M-rotary knob to move the line up or down in the display. If “b” was selected,
move this line up or down with the M-rotary knob.
6. Press the horizontal control HOR to activate the wave zoom mode. In the
lower menu under Window, select Main for cursor control in the main window or
Extension for cursor control in the zoom window.
– 64 –
Auto Cursor With the Auto Cursor setting, the horizontal cursor is set at the
intersections of the vertical cursors with the waveform.
Move the cursors with gesture control Please read the article “Other
Touchscreen Settings”.
– 65 –
Cursor measurements in FFT mode:
In FFT mode, press the cursor key to open the cursor menu.
A description of the cursor menu in FFT mode is given below:
Menu
Setting
Description
Type
Line Type (time and voltage type)
Window (Wave Zoom Mode)
Line
Source
Vamp Freq Freq&Vamp Auto Cursor
Freq Vamp Main Extension
a b from
Math FFT
Displays the vamp cursor measurement Displays the frequency measurement cursor of the Displays cursor of frequency and vamp measurement Horizontal cursors are displayed as the intersections of the vertical cursors with the waveform Activates vertical cursor Activates horizontal cursors Activates cursor in main window Activates cursor in FFT zoom window Move M-rotary knob to move line a Move M-rotary knob to move line b Move M-rotary knob to move line a and b at the same time, as they are connected Select the channel as source which is used for the cursor measurement.
Perform the following steps for amplitude and frequency measurement via cursor in FFT mode as follows:
1. Press the Math button and select FFT.
2. Press the cursor key to open the menu.
3. Select Window as extension in the lower menu.
4. Press the first menu field in the lower screen menu (H1) to change the
setting to Freq&Vamp. Two blue dash lines for the horizontal voltage range and
two blue dash lines for the vertical time range are now displayed.
5. In the lower menu, please select LineType as Freq to activate the vertical
cursors. If the line is selected as “a” in the lower menu, turn the M-rotary
knob to move the line to the right or left in the display. If “b” was
selected, move this line to the right or left with the M-rotary knob.
6. In the lower menu, please select LineType as Vamp to activate the
horizontal cursors. If the line is selected as “a” in the lower menu, turn the
M-rotary knob to move the line up or down in the display. If “b” was selected,
move this line up or down with the M-rotary knob.
7. In the lower menu under Window, select the setting Main for the cursor
control in the main window.
Autoscale function
This is a very useful function for first-time users who want to perform a
simple and quick check of the input signal. This function is used to
automatically track signals even if the signals change at any time. Using the
Autoscale function, the instrument can automatically set up the trigger mode,
voltage division and time scale according to the type, amplitude and frequency
of the signals.
– 66 –
Press Autoscale to display the following menu:
Menu
Setting
Description
AutoScale
ON OFF
Switching on the autoscale function. Switch off the autoscale function.
Track and adjust both vertical and horizontal settings.
Fashion
Track and adjust horizontal scale only. Track and adjust vertical scale only.
Wave
Display waveforms with multiple periods. Show only one or two periods.
To measure the autoscale signal:
Press the Autoscale button. The function menu is displayed. Press the H1 button to select the ON option.
Press H2 and select
for Mode.
Press H3 and select
for Wave.
1. When you call up the autoscale function, a is displayed at the top left of
the screen. 2. In autoscale mode, the oscilloscope can estimate the “trigger
type” (single) as well as the “mode” (edge,
video) itself. At this point, the trigger menu is not available. 3. In XY mode
and STOP status, press the Autoset button to switch to Autoscale mode. The
oscilloscope
switches to YT mode and AUTO triggering. 4. In autoscale mode, the
oscilloscope is always set to DC coupling and AUTO triggering. In this case,
making trigger or coupling settings has no effect. 5. In autoscale mode, if
the vertical position, voltage division, trigger level or time scale is
adjusted from
CH1 to CH4, the oscilloscope turns off the autoscale function. Press the
Autoset button to return to the Autoscale function. 6. If you switch off the
submenu in the autoscale menu, the autoscale is off; if you switch on the
submenu, you switch on the function. 7. With video triggering, the horizontal
time scale is set to 50µs. If one channel is set to edge signal while the
other is set to video, the time scale is set to 50µs. 8. In autoscale mode,
the following settings are enforced: If zoom mode is activated, it is
deactivated.
– 67 –
Help function
1. Press the Help button and the overview appears. 2. Press Prev Page or Next
Page to scroll through the help topics or use the M-rotary knob and the touch
screen function to select. 3. Press OK if you want to view details of the
theme. 4. Select Quit to leave the help menu again.
Executing keys
The executing keys are AUTOSET, RUN/STOP, SINGLE and COPY.
Autoset
This key is used to automatically set all the control values of the unit needed to generate a viewable waveform. Press the AUTOSET key; the oscilloscope then performs a quick automatic measurement of the signal
The following table shows the parameter values of the AUTOSET function:
Menu
Setting
Vertical Coupling
Current
Channel Coupling
Current
Vertical Scale
Adjust to appropriate pitch.
Bandwidth
Full
Horizontal Level
Medium or +/- 2 div
Horizontal Scale
Adjust to appropriate pitch.
Trigger Type
Slope or video
Trigger Source
CH1 or CH2 (2 channel) or CH1 to CH4 (4 channel)
Trigger Coupling
DC
Trigger Slope
Current
Trigger Level
3/5 of the waveform
Trigger Mode
Car
Display format
YT
Force
Stop
Help
Leaving
Pass/Fail
From
Inverted
From
Zoom Mode
Leaving
– 68 –
Evaluate waveform via autoset 5 types: Sine, Square, Video Signal, DC Level,
Unknown Signal The corresponding menus are shown below: Sine (Multi-Period,
Single-Period, FFT, Cancel Autoset)
Rectangle (Multi-Period, Single-Period, Rising Edge, Falling Edge, Cancel
Autoset)
Video Signal:
DC Level, Unknown Signal:
Description of the icons:
Multi Period:
Displays several periods at the same time
Single Period:
Displays one period
FFT:
Switches to FFT mode
Rising Edge:
Shows rising edge of the waveform
Falling Edge:
Shows falling edge of the waveform
Cancel Autoset:
Back to previous menu and waveform information
Run/Stop: Activates or deactivates the sampling of the waveform display.
Note: If there is no sampling in STOP status, the vertical divisions or the
horizontal time base can still be adjusted in a certain range. In other words,
the signal can still be extended in the horizontal or vertical range. If the
horizontal time base is <50ms, the horizontal time base can still be reduced
by 4 divisions.
Single: Pressing this button allows the trigger mode to trigger a single
sample, with waveform capture stopping after one sample.
Copy: You can save the waveform using the Copy button in any menu. The source
of the waveform and the storage location are defined in the Save menu. Please
also read the corresponding chapter “Saving and recalling data”.
– 69 –
Print screenshot To print a screenshot directly, proceed as follows: 1.
Connect a printer to the USB device port on the back of the unit.
Note: The printer must have PictBridge compatible drivers. 2. Press the
Utility button and select Output under Function. 3. In the lower picture menu
under Device, select Pict. (If the PC option is selected, you can transfer a
screenshot to the PC). 4. If you have activated Pict in the lower menu, a new
menu symbol Print Set appears on the right in which
you can define various settings for printing. Activate Ink Saver with ON to
print the image on a white background and thus save printer ink. 5. Once you
have set up the printer in this way, you can initiate another print by simply
pressing the Print button.
– 70 –
5. arbitrary function generator The oscilloscope has an integrated single-
channel arbitrary function generator with 25 MHz. The generator provides 4
basic waveforms (sine, square, ramp and pulse), as well as 46 integrated
arbitrary waveforms (e.g. noise, exponential rise, exponential fall, sin(x)/x,
stair step). You can also create your own waveforms and save them internally
or externally on USB. Connection Press the Utility button, then Function in
the lower menu. Under the Output setting in the left menu under Type, select
AG Output so that the BNC socket on the back of the unit is used for the
generator. Connect the BNC cable to the BNC socket marked Out on the back of
the unit.
Channel settings Press the OUT key to switch the output on and off. The
corresponding symbol (values) is displayed in the upper left corner of the
screen when the output is switched on. Signal settings 1. Press the AFG key to
display the generator menu. 2. Select the desired waveform in the lower screen
menu . 3. The corresponding menu with the settings of the selected waveform
appears on the right edge of the
screen.
– 71 –
Sine signal The parameters of the sine signal in the right menu are:
Frequency/Period, Amplitude/High Level, Offset/Low Level. Frequency settings
Set the Frequency in the right menu (if Frequency is not displayed, select
Period and press this button again to switch back to Frequency). To set the
parameters in the right-hand menu, proceed as follows: There are three ways to
change the parameters: 1. M-rotary knob: Turn the M-rotary knob to change the
value of the current cursor position. Press the
arrow keys under the M knob to move the cursor and highlight the desired
frequency setting location in white. 2. Input field: Press the M rotary knob
and a numeric input field appears for direct input of the frequency via the M
rotary knob or touch screen:
3. Touchscreen: Use the touchscreen fields that appear next to the frequency
value:
– 72 –
Period setting Select Period from the right-hand menu (if Period is not
displayed, select Frequency and then press this button again to switch back to
Period).
Amplitude setting Select Amplitude in the right-hand menu (if Amplitude is not
displayed, select High Level and then press this button again to switch back
to Amplitude).
Offset settings Select Offset in the right-hand menu (if Offset is not
displayed, select Low Level and then press this button again to switch back to
Offset).
High Level Settings Select High Level in the right-hand menu (if High Level is
not displayed, select Low Amplitude and then press this button again to switch
back to High Level).
Low Level Settings Select Low Level in the right-hand menu (if Low Level is
not displayed, select Offset Amplitude and then press this button again to
switch back to Low Level).
Rectangle signal The parameters of the square wave signal in the right menu
are: Frequency/Period, Amplitude/High Level, Offset/Low Level. Note: Do the
possible settings for frequency, amplitude etc. as described on the previous
page for the sine signal.
Ramp signal The parameters of the ramp signal in the right menu are:
Frequency/Period, Amplitude/High Level, Offset/Low Level, Symmetry. Note: Do
the possible settings for frequency, amplitude etc. as described on the
previous page for the sine signal. Symmetry adjustment Select Symmetry in the
right menu of the Ramp waveform, then set the parameters in the right menu.
Pulse signal The parameters of the pulse signal in the right menu are:
Frequency/Period, Amplitude/High Level, Offset/Low Level, PulseWidth/Duty.
Note: Do the possible settings for frequency, amplitude etc. as described on
the previous page for the sine signal. Pulse width settings Select Width in
the right menu of the pulse waveform, then set the parameters in the right
menu. (If Width is not displayed, select Duty and then press this button again
to switch back to Width).
– 73 –
Duty cycle setting Select Duty Cycle in the right menu of the pulse waveform,
then set the parameters in the right menu. (If Duty Cycle is not displayed,
select Width and then press this button again to switch back to Duty Cycle).
Arbitrary signals The parameters of the arbitrary signal in the right menu
are: Frequency/Period, Amplitude/High Level, Offset/Low Level, Built-in
Waveform, Editable Waveform. Note: Do the possible settings for frequency,
amplitude etc. as described on the previous page for the sine signal. There
are two types of arbitrary waveforms: Built-in waveforms and user-generated
waveforms. Integrated waveforms: 1. Press the AFG button on the unit to
display the generator menu. 2. Select Arb in the bottom menu, then Built-in in
the right menu. 3. Select Common, Math, Windows or Others from the right-hand
menu. Use Others, for example, to
open the following interface:
4. Turn the M-rotary knob to select the desired waveform (or via touchscreen)
and confirm the selection with the Select function in the right menu.
Note: To output a DC voltage, press the DC button on the unit.
– 74 –
Integrated arbitrary waveforms
Name Common waveforms StairD Stair U Stair DU Trapezia RoundHalf AbsSine AbsSineHalf SineTra SineVer NegRamp AttALT AmpALT CPulse PPulse NPulse Mathematical waveforms ExpRise ExpFall Sinc Tan Cot Sqrt XX HaverSine Lorentz In Cubic Cauchy Besselj Bessely Erf Airy Windows Rectangle Gauss Hamming Hann Bartlett Blackmann Laylight Triang Other DC Heart Round LFMPulse Rhombus Cardiac Noise
Explanation
Step down Stair step upwards Stairs down and up Trapezoidal waveform
Semicircular waveform Absolute value of a sine Absolute value of a half sine
Transverse truncated sine wave Vertical truncated sine wave Negative ramp
Increasing oscillator curve Weakening oscillator curve Coded pulse Positive
pulse Negative pulse
Exponentially increasing Exponentially decreasing Splitting function Tangent
Kotangens Square root Square function Semiversus Lorentz function Natural
logarithm function Cubic function Cauchy distribution Bessel Genus-J Function
Bessel Gen Y Function Error function Airy function
Rectangle window Gaussian distribution Hamming windows Hanning windows
Bartlett Fernster Blackmann windows Laylight window Triangular window (Fejer)
DC Signal Heart signal Circle signal Linear FM pulse Rhombus signal
Cardiosignal Noise signal
– 75 –
User generated waveforms
1. Press the AFG key to open the generator menu. Select Arb in the lower
menu, then Others and New to enter waveform creation.
2. Set the number of points of the waveform: Select Points in the right-hand
menu, then turn the M knob to set the number or use the touch screen field.
You can create a waveform with 2 to 8192 points.
3. Set interpolation: For Intrpl, select On/Off in the right-hand menu to
connect the points of the waveform to be adjusted in an increasing/decreasing
manner or to insert them without connection. Without connection, the voltage
between the points is not automatically adjusted and only jumps to the new
value when the next point is reached.
4. Edit waveform points: Select Edit Points in the right-hand menu:
Under Point, select the point of the waveform you want to edit. Only as many
points are available as you have selected for the waveform.
Select Voltage to set the voltage value of the selected point.
Repeat these steps for all necessary points of your waveform.
Save the created waveform with Save. If you want to save the waveform in the
internal memory, press Memory in the right menu until Internal is displayed
(default if no USB is connected). Turn the M-turn control or use the touch
screen function to select a memory location USER0 to USER31. Then press Save
in the right-hand menu.
If a USB memory is connected and you want to save the waveform to it, select Memory and then USB. The unit lists the folders on the USB memory. Select a folder with the M knob to scroll up and down the list. To change to the folder, select Change Dir in the right-hand menu. Select this again to change to the parent directory.
Enter the file name via the soft keyboard and confirm it with the BIN file in
the selected folder.
– 76 –
key. The waveform is now saved as a
Note: You can enter file names with a maximum of 35 characters. Recalling a
saved waveform 1. Press AFG and then select Arb from the bottom menu. Then
select Others and File browse. 2. Select the desired waveform from one of the
memory locations in the internal or external memory. 3. Confirm Read in the
menu on the right.
– 77 –
6. multimeter Input sockets The input sockets of the digital multimeter are
located on the back of the unit and are marked accordingly with 10A, mA, COM
and V//C.
DMM menu Press the DMM button on the front of the unit to switch the
multimeter function on and off. When the multimeter function is on, this
button is illuminated.
The multimeter menu is described below:
Menu
Setting
Description
Current Voltage
R On-Off
Configure
ACA DCA ACV DCV
R
C
Hold ON OFF Show Info ON OFF Fashion Car Manu Relative
AC current measurement DC current measurement AC voltage measurement DC
voltage measurement Resistance measurement Diode measurement Capacity
measurement Switches to continuity test (On-Off measurement) Freezes the
measured value acquisition
Shows/hides the multimeter window
Selects automatic or manual mode
Sets the current measurement value to zero. The measurement is the difference
value of the reference value to the current measurement display.
Current
Selects the current measuring range
mA 10A
– 78 –
DMM information window The digital multimeter window is shown in the upper right corner of the display.
Description
1. Manual/Auto range indicator: Manual means that the measuring range must be
set manually, whereas Auto switches the measuring range automatically.
2. Measuring mode pointer:
V
Voltage measurement
A
Current measurement
R
Widertandsmessung
Diode measurement
C
Capacity measurement
Continuity check
3. Current measuring range 4. Measured value with unit 5. Data hold is
activated 6. Multimeter recorder 7. Reference value of the relative value
function 8. Measuring range for current measurement: mA or 10A 9. AC or DC
display for current or voltage measurement
AC or DC voltage measurement:
1. Press the DMM button on the front of the unit. Then press Voltage in the
lower menu several times to switch between ACV (alternating voltage) and DCV
(direct voltage).
2. Connect the black test lead to the COM socket and the red test lead to the
V//C socket. 3. Connect the test probes to the voltage to be measured and read
the measured value in the digital
display of the multimeter window.
– 79 –
AC/DC current measurement
Measure a current below 400mA as follows:
1. Press the DMM button on the front of the unit. Then press Current in the
lower menu several times to switch between ACA (alternating current) and DCA
(direct current).
2. Connect the black test lead to the COM socket and the red test lead to the
mA socket. 3. Select Configure in the lower screen menu, then switch to mA in
the right menu. 4. Switch off the circuit to be tested and discharge all
capacitors still present in the circuit. 5. Disconnect the circuit under test
and connect the black test lead to the open end of the disconnected
lead facing the load and the red test lead to the other open end of the
disconnected lead facing the voltage source. If this has been reversed, a
negative “-” sign is displayed. 6. Switch the circuit on again and read the
current value in the measuring display. 7. Disconnect the circuit under test
and discharge any remaining capacitors in the circuit before reconnecting the
disconnected line.
Measure a current above 400mA to 10A as follows: 1. Press the DMM button on
the front of the unit. Then press Current in the lower menu several times to
switch between ACA (alternating current) and DCA (direct current). 2. Connect
the black test lead to the COM socket and the red test lead to the 10A socket.
3. Select Configure in the lower screen menu, then switch to 10A in the right
menu. 4. Switch off the circuit to be tested and discharge all capacitors
still present in the circuit. 5. Disconnect the circuit under test and connect
the black test lead to the open end of the disconnected
lead facing the load and the red test lead to the other open end of the
disconnected lead facing the voltage source. If this has been reversed, a
negative “-” sign is displayed. 6. Switch the circuit back on and read the
current value in the measuring display. 7. Disconnect the circuit under test
and discharge any remaining capacitors in the circuit before reconnecting the
disconnected line.
Resistance measurement: 1. Press the DMM button on the front of the unit. Then
press R in the lower menu several times to switch
between resistance, diode and capacitance until R is highlighted. 2. Connect
the black test lead to the COM socket and the red test lead to the V//C
socket. 3. Connect the test probes to the (voltage-free) electrical conductor
to be measured and read the
measured value in the digital display of the multimeter window. Note: Never
perform a resistance measurement on a live conductor to avoid damaging the
unit.
Diode measurement: 1. Press the DMM button on the front of the unit. Then
press R in the lower menu several times to switch
between resistance, diode and capacitance until is highlighted. 2. Connect the
black test lead to the COM socket and the red test lead to the V//C socket. 3.
Connect the red test probe to the anode of the diode and the black test probe
to the cathode of the
diode. The cathode is marked here with a ring around the body. Read the
measured value in the digital display of the multimeter window.
– 80 –
Capacity measurement:
1. Press the DMM button on the front of the unit. Then press R in the lower
menu several times to switch between resistance, diode and capacitance until C
is highlighted.
2. Connect the black test lead to the COM socket and the red test lead to the
V//C socket. 3. Connect the test probes to the (discharged) capacitor to be
measured and read the measured value in
the digital display of the multimeter window. Note: Measurements on a charged
capacitor can damage the device. If you want to measure a capacitance of less
than 5nF, use the relative function before the measurement to improve the
measurement result.
Continuity test:
1. Press the DMM button on the front of the unit. Then press On-off in the
lower menu. 2. Connect the black test lead to the COM socket and the red test
lead to the V//C socket. 3. Connect the test probes to the (voltage-free)
electrical conductor to be measured and read the
measured value in the digital display of the multimeter window. If the
resistance value is below 50, the continuity signal sounds. Note: Never
perform a resistance measurement on a live conductor to avoid damaging the
unit.
Other multimeter functions
Data Hold Mode: You can freeze the measured value in the display. 1. Select
Configure in the lower menu and then press Hold in the right menu as ON. The
display shows
HOLD. 2. Press OFF to close the data storage again.
Information display: 1. Select Configure in the lower menu and then press Show
Info in the right menu as ON. The
measurement display is shown. 2. Press OFF to switch off the multimeter’s
measuring display.
Auto or Manual Range Selection: You can freeze the measured value in the
display. 1. Select Configure in the lower menu and then press Hold in the
right menu as ON. The display shows
HOLD. 2. Press OFF to close the data storage again.
Relative value measurements: Sets the current measurement value to zero. The
measurement is the difference value of the reference value to the current
measure