PHILIPS 867216 Smart Hopping 2.0 Access Point Installation Guide
- June 8, 2024
- Philips
Table of Contents
Smart-hopping 2.0 Access Point 1.4 GHz
Installation Guide
867216 Smart Hopping 2.0 Access Point
Proprietary Information
This document contains proprietary information that is protected by copyright.
Copyright
Copyright © 2022 Koninklijke Philips N.V. All Rights Reserved.
Manufacturer
Philips Medizin Systeme
Böblingen GmbH
Hewlett-Packard-Str 2
71034, Böblingen, Germany
United States Address
Philips Medical Systems North America Company
222 Jacobs Street
Cambridge, MA 02141
http://healthcare.philips.com
This document was published in the United States of America.
Trademark Acknowledgments
All trademarks, trade names and company names referenced herein are used for
identification purposes only and are the property of their respective owners.
Warranty
The information contained in this document is subject to change without
notice. Philips Medical Systems Nederland B.V. reserves the right to make
changes in specifications and/or to discontinue any product at any time
without notice or obligation and will not be liable for any consequences
resulting from the use of this publication.
Printing History
New editions of this document will incorporate all material updated since the
previous edition. The documentation printing date and part number indicate its
current edition. The printing date and edition number change when a new
edition is printed. The document part number changes when extensive technical
changes are incorporated.
First Edition ……………………………. June 2022
About This Guide
This Smart-hopping 2.0 Access Point 1.4 GHz Installation Guide provides complete instructions and procedures for installing the Philips Smart-hopping 2.0 Access Point 1.4GHz. This chapter describes the document and includes:
- Audience
- Document Organization
- Notational Conventions
- Related Documentation
- Terminology
Audience
Philips wrote the Smart-hopping 2.0 Access Point 1.4 GHz Installation Guide
for trained service personnel who install the Smart-hopping 1.4 GHz Access
Point as part of an overall Smart- hopping deployment.
Document Organization
The information in this guide is organized and presented as follows:
- Chapter 1, Overview, describes the Smart-hopping 2.0 1.4 GHz Access Point and how it provides a bidirectional data flow between the Philips Patient Information Center iX (PIC iX) server and Patient Monitors.
- Chapter 2, Mount and Install the Smart-hopping 2.0 Access Point 1.4 GHz, includes instructions for mounting the Access Point to a wall, flush with a ceiling tile, below a ceiling tile, and tether mounted.
- Chapter 3, Installing the Smart-hopping 2.0 Access Point 1.4 GHz, provides procedures for physically installing the Philips Smart- hopping 2.0 Access Point 1.4GHz.
- Chapter 3, Maintaining the Smart-hopping 2.0 Access Point 1.4 GHz, provides procedures to maintain and troubleshoot operation of the Philips Smart-hopping 2.0 Access Point 1.4GHz.
Notational Conventions
This guide uses the following notational conventions to convey information:
Note
Notes direct your attention to important information.
Caution
Cautionary statements direct your attention to a condition that could
result in loss of data or damage to equipment.
Warning
Warnings direct your attention to a condition that could result in physical
injury.
Related Documentation
Refer to these other documents for other installation service information
about the Smart-hopping infrastructure:
- Smart-hopping 2.0 Access Point Controller Installation Guide – provides procedures for physically installing and powering the Smart-hopping Access Point Controller at the clinical site.
- Smart-hopping 2.0 Infrastructure Installation and Service Guide – provides complete information and procedures to install, configure, inter-connect, and deploy the Smart-hopping infrastructure at the clinical site. This document includes site planning guidelines, procedures for use of the APC command line and graphical user interfaces, Access Point configuration procedures, and APC and Access Point firmware deployment procedures.
- Smart-hopping Synchronization Unit Installation Guide – lists procedures to install the Smart-hopping Sync Unit at the clinical site.
- Smart-hopping 2.0 Upgrade Guide – gives instructions on upgrading Philips Smart-hopping infrastructure (Access Points and Access Point Controllers).
Terminology
Note the following terms, acronyms, and abbreviations used throughout this
document and in related documentation:
- Access Point (AP) – A Smart-hopping component that provides bidirectional wireless access to the monitoring network for Patient Monitors.
- Access Point Controller (APC) – A Smart-hopping component used to manage the operation of the Access Points. One APC is elected as the Primary APC. The Primary APC supports the web- based interface to the system and manages the overall configuration.
- Access Point Group/AP Group – A logical grouping of Access Points. Access Point members of the same Access Point Group inherits common configuration settings (defaults). Access Point groups often map logically to the clinical units in which you install the Smart-hopping Infrastructure.
- IntelliVue Network – This term refers to the entire IntelliVue Network. In a routed topology, the IntelliVue Network includes the routers, switches, firewalls, and the Smart-hopping wireless infrastructure.
- Patient-Worn Monitor (PWM) – The Patient-Worn Monitor relays real time Physiological waveforms and trends to the Philips Patient Information Center iX (PIC iX).
- Smart-hopping infrastructure – Philips proprietary wireless network designed for continuous monitoring that provides two-way communications between Patient Monitors and the PIC iX server.
- Smart-hopping Infrastructure Service Tool – The software used to upgrade Smart-hopping APCs and Access Points, verify that APCs on your network are configured correctly, and display warning and error messages that you may use to troubleshoot any configuration errors that may exist on your Smart-hopping network. The Smart-hopping Infrastructure Service Tool is also referred to as the Upgrade Tool. This tool was previously referred to as the Upgrade Wizard.
- Partnered APC – Configurable element within an Access Point Group used to determine which APC manages the operation of the Access Point members of a particular Access Point Group.
- Power over Ethernet (PoE) Switch – The Power over Ethernet (PoE) Switch is a Power-over-Ethernet device that provides 48- VDC power to Access Points (and also remote Sync Units if connected) via 100Base-TX Ethernet LAN cabling.
- RF (Radio Frequency) Access Code – Configurable element in the Smart-hopping Access Point defaults shared among Access Points and wireless clients to control wireless access to the monitoring network. Wireless devices only connect to access points with which they share access codes. The RF Access Code allows a specific wireless client that is programmed with a matching Access Point RF Access Code to connect to that Access Point.
- Synchronization (Sync) Unit – The Smart-hopping Sync Unit provides a necessary common clock signal to synchronize all the Access Points in the system. When patients move around the hospital coverage area their transmitted data are handed over from one Access Point to another seamlessly without interruption or data loss.
- Smart-hopping Network- This term is used to describe the infrastructure used in a network topology to connect Smart- hopping devices (Access Points, Access Point Controllers, Synchronization Units).
- System ID – Configurable element in the APC Configuration that logically associates Access Points and Access Point Controllers operating within the same Smart-hopping Infrastructure.
- Uninterruptible Power Supply (UPS) – The UPS supplies backup power to protect against hospital generator change-over interruptions, and short power line transients.
Chapter 1: Overview
Introduction
The Smart-hopping infrastructure uses a cellular wireless architecture to
provide two-way communications between Patient-Worn Monitors and the PIC iX
server.
Using the Smart-hopping wireless protocol, the Smart-hopping infrastructure
provides monitoring capabilities for ambulatory patients within a wide
coverage area. The Patient- Worn Monitors and the Smart-hopping infrastructure
operate on the 1.4 GHz US Wireless Medical Telemetry Service
You can configure the Access Point Controller to communicate with 1.4 GHz
Smart-hopping Access Points (APs). 1.4-GHz Access Points can only communicate
with 1.4 GHz Patient-Worn Monitors.
Smart-hopping networks utilize a cognitive radio that senses the RF
environment and adapt to it. Dynamic wireless channel allocation ensures best
use of available wireless spectrum. The Smart-hopping infrastructure is
designed to co-exist with other 802.11 wireless deployments.
A General Description of the Smart-hopping Access Point
The Smart-hopping Access Point (AP) and remote antennas (RA), shown in Figure
1-2, provide an air-link to transmit and receive data between Patient-Worn
Monitors and the Information Center server via the Smart-hopping
infrastructure.
Note
Mounting Access Points must be performed by qualified personnel, using
certified conductors and following national electrical codes.
The Access Point is a modular antenna infrastructure consisting of an Access
Point (AP) with up to two connected Remote Antennas (RAs). A 74-foot (22.6 m)
coaxial and unshielded twisted pair (UTP) cable bundle is used to connect a
Remote Antenna to an Access Point.
Caution
The coaxial cable bundle cable length may vary from 69 – 79 feet. When
planning remote antenna placement, ensure the maximum length between the
Access Point and the
remote antennae is 69 feet.
The effective range of the Access Point and of each Remote Antenna is
typically 32 feet. The Access Point supports a maximum of 18 Patient-Worn
Monitors regardless of its component configuration.
When monitored patients are ambulatory, data roaming is handled seamlessly
between the other Access Points in the coverage area. Each RA is always used
with its two supplied antennas installed. The Access Point and its attached
Remote Antennas can be mounted out of the way on corridor walls, or above or
below ceiling tiles.
Access Point Mounting Options
Wall-mounting hardware is standard. There are also options available for
mounting Access Points to ceiling tiles (mounted on the tile or flush-mounted
with the tiles. The part numbers for these options include the following:
- 867216 – Basic Access Point Kit (includes hardware for wall mounting)
- Optional 453564979681 (IM4) Plastic Access Point cosmetic cover
- 867416 – IntelliVue Smart-hopping Access Point mounting hardware options:
- Basic Ceiling Mount – 453564979661 Metal rail and hardware (IM2) with the (optional) 453564979681 (IM4) Plastic Access Point cosmetic cover
- Flush Ceiling Mount – 453564979661 (IM2) Metal rail and hardware and 453564979671 (IM3) Plastic cosmetic ring
- If you plan to connect Remote Antennas to the Access Point, you must order the 453564979661 Metal rail and hardware (IM2) with the 453564979681 (IM4) Plastic Access Point cosmetic cover
Power Source
The Smart-hopping 2.0 Access Point 1.4 GHz receives its 48-V DC operating
power source via its Ethernet LAN cabling from Power over Ethernet via the
Smart-hopping Sync Unit. The Access Point is not equipped with a power socket.
The Access Point internally generates a variety of voltages used for its
internal components.
The CAT-5e (or higher) UTP cable within the 74 ft.-cable bundle carries a 5
VDC (maximum of 5.5 V DC power), Transmit and Receive control signals, and
Antenna Diversity signals from the Access Point to a connected Remote Antenna.
The 75-Ohm coaxial cable within the 74 ft.-cable bundle carries RF and DC
sense signals from the Access Point to a connected Remote Antenna.
Synchronization Signal
The Access Point receives a synchronization signal from a network of Sync
Units that enables a Patient-Worn Monitor to hand over data seamlessly between
Access Points within the coverage area when a patient is ambulatory and to
transfer data to the PIC iX server without interruption. Each Sync Unit
provides synchronization for up to 12 Access Points. The sync signal
distributes the common reference clock signal needed by the Smart-hopping
infrastructure.
Wireless Client Mobility
The Smart-hopping infrastructure supports seamless roaming of wireless
clients within the area of coverage. This roaming is accomplished via
communications between the wireless clients and the Access Points (AP) as
follows.
As a wireless client is moved around a building, it automatically monitors the
quality of the wireless link to its current Access Point (and it also detects
the presence of other Access Points). When the quality starts to deteriorate,
the wireless client automatically establishes a new connection to another
Access Point.
The Patient-Worn Monitor remains connected to two Access Points for a finite
time, and thus the same data is received by these Access Points. During this
time, information for header compression and other data for the connection is
routed to the new Access Point.
One of the two Access Points then releases the radio connection. If a packet
is in progress when a handover occurs, then the packet is reassembled co-
operatively between the two Access Points.
The Information Center server receives an unbroken flow of complete IP
packets.
Technical Alerts
Each Access Point is configured to signal alert conditions that are viewed
by the PerformanceBridge Focal Point management system.
The Access Point is configured to raise alerts on error conditions. The alerts
are sent over the Smart-hopping and Ethernet LAN infrastructure to a
monitoring station or Database Server. The Access Point raises an alert if:
- If data loss exceeds the configured threshold
- It loses its synchronization signal
For small systems having only one Access Point and no Sync Unit, you can
configure the Access Point to suppress the loss of synchronization signal
alert.
Note that the Smart-hopping APC provides System Alerts such as loss of
synchronization, high data loss, Access Point hardware failure, and over
capacity. Also, when redundant APCs are installed, the APC provides a System
Alert for APC hardware or software failure.
Firmware Updates
Firmware on the Smart-hopping 2.0 Access Point 1.4 GHz is upgraded from the
Upgrade Tool. Refer to the service document entitled Smart-hopping 2.0
Infrastructure Installation and Service Guide for details. The Access Point
firmware image is provided on the Smart-hopping Telemetry Service Tool CD-ROM.
Management Interfaces
The Access Point supports the full range of management interfaces via the
Smart-hopping Access Point Controller (APC). These interfaces also enable you
to view the status of the Access Point including the following information:
- System configuration such as firmware version
- Connection information such as numbers of packets received and transmitted, and number of errors
- Waveform data information such as total number of seconds of data sent and lost
The Access Point statistics can be read by remote devices (for example, an
IntelliVue Database Server), using the Scaleable Node Address Protocol (SNAP).
You can display the following status items using the APC web-based interface:
- Access Point Name
- Physical Address
- Partnered Access Point Controller
- IP Address
- Subnet Mask
- Default Gateway
- Access Point Type
Connectors and Status Indicators
Figure 1-3 shows the connectors on the Access Point:
Smart-hopping Access Point Connectors
Note the following connectors on the Smart-hopping Access Point:
- Ethernet Interface (LAN) – The Access Point provides a 100-Base-T Ethernet interface with an RJ45 connector to connect the Access Point to the Smart-hopping Sync Unit.
The Access Point Ethernet interface provides data communications to and from the IntelliVue Information Center over the Smart-hopping LAN infrastructure. It also presents the power and synchronization signals required by the Access Point. The synchronization signal is superimposed on the power supply voltage. Table 1-1 lists the pin signals for the Access Point Ethernet interface.
Table 1-1: Access Point Ethernet Interface Pin Signals Pin| Signal Description
---|---
1| Transmit Pair TX + Conductor
2| Transmit Pair TX – Conductor
3| Receive Pair RX + Conductor
4| + 48 V DC Power and Synchronization
5| + 48 V DC Power and Synchronization
6| Receive Pair RX – Conductor
7| 0 V Power Return
8| 0 V Power Return - UTP Cable Connectors to Remote Antennas (RA1 and RA2) – Two standard RJ45 connectors are provided for the UTP cables that connect the Access Point to its Remote Antennas. Each UTP cable carries power, Transmit, Receive, and Antenna Diversity Control signals to the Remote Antenna.
- Coaxial Cable Connectors to Remote Antennas – Two standard 75-Ohm connectors are provided for the coaxial cables that connect the Access Point to its Remote Antennas. Each coaxial cable carries RF and DC sense signals from the Remote Antenna.
Note
If an installed, powered Remote Antenna becomes disconnected from its Access
Point via its Coax/UTP cable bundle, you must reconnect the Coax/UTP cable
bundle to the RA and Access Point, and then cycle power to the connected
Access Point before the RA can re- establish communications with the Access
Point.
Front Panel
Smart-hopping Access Point Status LEDs
The Smart-hopping Access Point provides the following status LEDs on the front
panel of the Access Point. At initial power on the Access Point runs a Power
On Self-Test (POST). During the POST, the LED indicators flicker illuminate
continuously (AMBER) to indicate correct startup operation. Next, the Power ON
LED illuminates (GREEN) continuously to indicate that the 48-V DC power and
sync signal are being supplied, and the other two (AMBER) LEDs turn change
color (or turn off), depending on the status of the Access Point.
Table 1-2 summarizes the status the Access Point LEDs:
310236883
2022-06-07 19:50:31
Access Point Status
LED1 -Power/Sync| LED2 – Network| LED3 – Radio| LED4|
Remote Antenna Status
---|---|---|---|---
Green | Flashing Green | Flashing Green | Off | Power/Sync, Network, and Radio
are working properly.
FlashingGreen | Flashing Green | Flashing Green | Off | The Access Point has
lost remote synchronization. Check remote (upstream) synchronization.
Green | Flashing Green | Flashing Amber | Off | Power/Sync, Network, Radio,
and Remote Antennas are working properly.
Green | Flashing Green | Flashing Red | Off | Power/Sync and Network are
working properly. There is an issue with the radio.
Amber | Amber | Amber | Amber | LEDs that display during the Access Point
Initial POST (Power- On-Self-Test)
Amber | Amber | Amber | Off | The Access Point is not registered to any APC.
Flashing Red | Flashing Green | Flashing Green | Off | The Access Point has
lost local synchronization. Check local (upstream) synchronization.
Red | Off | Off | Off | If the Access Point is in this state for more than 5
seconds, this indicates that the Access Point has power but has a hardware
fault in another area. Replace the Access Point.
Flashing Red | Flashing Red | Flashing Red | Off | Hardware POST failure
(there is an issue with one or more of the following components: DRAM, Flash
Memory, CPU, Radio, LAN Interface)
Off | Off | Off | Off | The Access Point either does not have power or
configured to turn the LEDs off.
White | White | White | White |
Blue | Blue | Blue | Blue |
310236883
2022-06-07 20:00:10
Blue – not in service
See Specs in Windchill 867216-90002
Specifications
Table 1-3: Smart-hopping 2.0 Access Point 1.4 GHz Specifications
Specification | Value |
---|
Physical:
Chassis (only) Dimensions (H x W x L)| 30 mm x 163 mm x 163 mm (1.2 in x 6.38
in x 6.38 in)
Weight with Internal Antennas| 320 g (0.71 lb)
Mounting| Flush Mount with Ceiling tile (includes Cosmetic Ring, Below
Ceiling, Below Ceiling with Cosmetic Cover, Wall Mount, or Wall Mount with
Cosmetic Cover
Environmental:
Operating Temperature| 0 to +55o C (32 to 131oF)
Storage Temperature| -40 to +60o C (-40 to 140oF)
Humidity Range (Operating)| < 95% RH @ 40o C non-condensing
Humidity Range (Storage)| < 90% RH @ 60o C
Altitude| Operating and Storage up to 3,048 m (10,000 ft)
Electrical:
Power (Input)| 48 VDC nominal (37-57 VDC)
8 watts, 287 mA (Access Point)
Power Sensing| Auto-sensing PoE, compliant with 802.3af and 802.3at (Type 1)
Power (Output)| 5 VDC @ 50 mA at Remote Antenna RJ-45 connectors
LED Indicators| Four LEDs for:
• Power/Sync
• Network
• Radio Activity
• Future use
Smart-hopping Radio
Smart-hopping Antenna Frequency Range| 1395-1400 MHz + 1427-1432 MHz
WMTS – Wireless Medical Telemetry Service (USA)
Antenna Type| Smart-hopping dual internal antennas
RF Diversity| Uses two antennas and selects the antenna with the best signal.
Frequency Diversity| Dynamic, selects RF channels for best signals.
Compatibility Mode (6 channel mode):
Channel spacing| 1.6 MHz
Modulation| GFSK
Power output| 12.5 (±1.5) dBm (12.6 mW to 25 mW) into Antenna Load (2:1 VSWR
maximum)
Antenna gain| 2 dBi
Time slot types supported| Single or double/long
Total time slots available| 32
Specification| Value
---|---
Time slots allocated for wireless clients
• No Remote Antennas connected
• One Remote Antenna connected
• Two Remote Antennas connected
• Time slots allocated for roaming| 18
9 (and 9 for the Remote Antenna)
6 (and 6 for each Remote Antenna)
14
Advanced Mode (4 channel mode):
Channel spacing| 1.728 MHz
Modulation| PI/2-DBPSK, PI/4-DQPSK
Power output| 14 (±1.5) dBm (17.8 mW to 35.5 mW) into Antenna Load (2:1 VSWR
maximum)
Antenna gain| 2 dBi
Time slot types supported| Single
Total time slots available| 32
Time slots allocated for wireless clients
• No Remote Antennas connected| __
18
Time slots allocated for roaming| 14
Electrical Installation:
Fire Safety| Smart-hopping 2.0 Access Point 1.4 GHz are Listed for use within
“Other Spaces Used for Environmental Air (Plenum)” per NFPA70: 2011, Article
300.22.
Note: The term “plenum” as used in Article 300.22 Section C correlates
with the use of the term “plenum” in NFPA 90A-2009, Standard for the
Installation of Air-Conditioning and Ventilating Systems, and other mechanical
codes where the plenum is used for return air purposes, as well as some other
air- handling spaces. The area above dropped ceilings is an example of plenum
space.
Interface Connections:
LAN Input (Data): One-port Ethernet 10/ 100 Base-T (100-Mbps connections must
be Full Duplex)| RJ45 Female Socket
LAN Cable to Network Switch| CAT-5e (or higher) with a length of up to 100 m
(328 ft.).
Remote Antenna UTP Cable Connectors| Two RJ45 Female Socket Connectors are
provided to connect the UTP cables with 5 V DC power, Transmit and Receive
control signals, and Antenna Diversity signals to the remote Antennas are
carried. Connectors are protected against damage from unexpected connection to
LAN with PoE.
Remote Antenna 75 Ohm Coaxial Cable Connectors| Two standard 75-Ohm connectors
are provided for the coaxial cables on which RF and DC sense signals are
carried to the Remote Antenna.
Ordering Information
Table 1-4: Smart-hopping 2.0 Access Point 1.4 GHz Part Numbers and Contents
Item | Commercial Part Number | Contents |
---|---|---|
Smart-hopping Access Point | 867216 | Access Point, Wall anchors, #6 |
1.25-Inch self- tapping screws, Mounting Bracket
Philips Smart-hopping 2.0 Installation Material:| 867416| See
options (IM1, IM2, IM3, IM4 below:
• Smart-hopping Smart-hopping 2.0 Access Point
1.4 GHz Remote Antenna Cable
| Option IM1| Applicable for USA and Puerto Rico only. One cable required for
each Remote Antenna.
• Basic Ceiling Mount Kit| Option IM2| Mounting Rail, nuts (4), push nuts
(8), adapter plates (2), adhesive strips for adapter plates (2)
• Flush Mount Ceiling Ring Kit| Option IM3| Cosmetic Ring
• Ceiling/Wall Mount Cover Kit| Option IM4| Plastic Cover, machine screws
(2), and plastic grommets (2)
Table 1-5: Smart-hopping 2.0 Access Point 1.4 GHz Mounting Options by Mount Type
Mounting Option | Order these product options |
---|---|
Wall Mounting with Cosmetic Cover | 867416 |
Below Ceiling Mount | 867416 |
Below ceiling Mount with Cover | 867416 |
Flush Ceiling Mount with Cover | 867416 |
Regulatory Information
Table 1-6: -Smart-hopping 2.0 Access Point 1.4 GHz Regulatory Information
Description| Philips part#:| FCC ID:| Model#|
Software
---|---|---|---|---
Smart-hopping 2.0 Access Point 1.4 GHz with Remote Antenna| 453564883261
867216| PQC-867216A| ITS867216A| 5.0 or greater
FCC Compliance
Operation of this equipment requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry Service.
The transceiver and the Smart-hopping infrastructure are subject to radio
frequency interference. If suspected radio frequency interference with your
device, contact your service provider. This device complies with Parts 15 and
95H of the Federal Communications Commission (FCC) Rules. Operation is subject
to the condition that this device does not cause harmful interference.
This equipment complies with the FCC RF radiation exposure limits set forth
for an uncontrolled environment. This equipment should be installed and
operated with a minimum distance of 20 cm between the radiator and any part of
your body.
Symbol Definitions
Table 1-7 illustrates and defines the symbols appearing on the Smart-hopping
Access Point hardware:
Table 1-7: Smart-hopping 2.0 Access Point 1.4 GHz Symbol Definitions
Symbol | Description |
---|---|
__ | Philips shield logo |
**** | Legal Manufacturer Information |
__ | Country of manufacturer and date of manufacture |
__ | Tested to comply with FCC standards |
__ | Non-ionizing radiation |
__ | Contains parts that you must not put into normal waste disposal but must |
be recycled or dealt with as chemical waste. Dispose of in accordance with the
local country requirements.
| Tested to comply with CSA Group standards
****| Consult instructions for use
| See Warnings (Table 1-8)
****| Catalog Number (867216)
Table 1-7: Smart-hopping 2.0 Access Point 1.4 GHz Symbol Definitions
Symbol | Description |
---|---|
Serial Number (DKYWWNNNNN) | |
__ | Access Point MAC (Machine Access Code) address (MMMMMMMMMMMM) |
__ | BLE Antenna MAC (Machine Access Code) address (MMMMMMMMMMMM) |
__ | Service Number (453564883261) |
__ | Rated power input, DC (48 VDC) |
__ | 10/100 Base-T RJ-45 LAN connection |
__ | Local Area Network Connection |
__ | Rated power, output (5 VDC) |
__ | Data input/output connection |
RA1
| Access Point to Remote Antenna 1 Connection
RA2
| Access Point to Remote Antenna 2 Connection
__| Coaxial Remote Antenna connection
http://www.datenfreihafen.org/projects/iec16022.html
| 2D Bar code that includes the following:
• REF, Serial and Service numbers
• Access Point MAC
• BLE MAC address
Warnings
See Table 1-8 for product warnings:
Table 1-8: Product Warnings
_| Warnings
Consult the instructions for use.
This product can expose you to chemicals including Lead and Lead Compounds,
which are known to the State of California to cause cancer. For more
information, go to www.P65Warnings.ca.gov
The EU REACH Regulation 1907/2006 requires Philips to provide chemical
content information for Substances of Very High Concern (SVHC), if they are
present in the relevant article above a concentration of 0.1% weight by
weight. Information on substances, contained in the Philips products, can be
found on the Philips REACH website:
www.philips.com/REACH
_Disposables (product cannot be repaired):
• Wash hands after use, wear gloves, where possible, and avoid food intake
while handling the product.
Operation of this equipment requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry Service.
This device complies with part 15 of the FCC rules. Operation is subject to
the following two conditions:
• That this device may not cause harmful interference
• This device must accept any interference received, including interference
that may cause undesired operation.
---|---
Chapter 2: Mount and Install the Smart-hopping 2.0 Access Point 1.4 GHz
Access Point Placement Guidelines
Note the following important guidelines when locating Smart-hopping Access
Points:
- Consider building construction when placing Access Points to account for interference from ceramic wall tile, lead lined walls, elevator shafts, reinforced windows, and other obstacles which may cause signal degradation.
- Place Access Points in locations where there is no more than one wall between the Access Point and the coverage area.
- Ensure that the coverage area takes into account bathrooms, hallways, and windows.
- Access Point (internal) antennas must be more than four inches (10 cm) away from metal structures. If the antennas are too close to the structure, antenna performance can be degraded.
- Access Points or Remote Antennas shall not be placed closer than three feet (1 m) from other Access Points or RAs to prevent signal overload conditions.
- Each Smart-hopping Access Point requires a 100Mbps/Full Duplex or 1000 Mbps switch port connection.
- The coaxial cable bundle cable length may vary from 69 – 79 feet. When planning remote antenna placement, ensure the maximum length between the Access Point and the remote antenna is 69 feet.
- Try to avoid placing Access Points close to other electrical devices (exit lights, light fixtures, speakers, etc.). Devices like florescent light ballasts can create a significant amount of interference that can impact system performance.
- When installing a Smart-hopping network, Philips requires that the following LAN cables are Shielded Twisted Pair cables:
- Cables connecting the switch to the Access Point Controllers
- Cables connecting the switch to the Synchronization Units (that connects to the Access Points)
Warning
Qualified personnel must mount Access Points and use certified conductors and
follow national and local electrical codes.
All Smart-hopping devices are intended for indoor use only.
Fixed Mounting the Smart-hopping Access Point to a Wall
You can mount the Smart-hopping Access Point to a wall using the mounting
screws and screw anchors supplied with the Access Point. We recommend that you
mount the Access Point high on the wall as close to the ceiling as possible.
To mount the Smart-hopping Access Point to a wall:
Table 2-1: Required Wall Mounting Material
Required Parts
Illustration| Part Name| Quantity
867216 Access Point Kit -includes the following:| 1
| Access Point| 1
| Wall anchors| 2
| #6 1.25-Inch self- tapping screws| 2
| Mounting Bracket| 1
Table 2-2: Wall Mounting Tools
Required Tools
| Tape Measure
| Level
| Hammer (when using wall anchors)
| Drill
| 7/32 -inch drill bit
Required Tools
| Phillips head screwdriver
| Marker
Installation
-
Using a 7/32-inch drill bit, drill two pilot holes 0.219 inches (5.5 mm) in diameter, 3.394 inches (100 mm) apart (from the center points of the holes), and horizontally level with each other.
-
If the screws are going directly into the wall (not into building studs), tap a supplied plastic screw anchor into each pilot hole until it is flush with the wall surface.
-
Place the mounting bracket against the wall and line up the two holes.
-
Screw a supplied #6 x 1 1/4 inch self- tapping screw through each hole in the mounting bracket and into each screw anchor (or pilot hole).
-
Pull the cables out from the ceiling and connect them to the appropriate jacks on the Access Point.
Connect the Ethernet (network connection) cable to the LAN port on the Access Point. This provides power to the Access Point and lights up the LEDs on the front of the Access Point (if the other end of the Ethernet cable is connected to a Power over Ethernet Switch that is powered on).
If using two Remote Antennas, perform this step on both the RA1 and RA2 connectors on the Access Point:
• Connect the Coaxial cables (from the same side of the strand) to the RA coaxial connector.
• b) Connect the RJ-45 connector to the RA RJ-45 connector. -
While holding the Access Point chassis, fit the holes in the back of the Access Point chassis over the keyhole screws that protrude from the mounting bracket. Slide the Access Point chassis down so that the screw heads are securely in the grooves in the back of the Access Point chassis. You hear a click noise when the Access Point is securely placed in the mounting bracket.
Fixed Mounting the Smart-hopping Access Point to a Wall (With Cosmetic
Cover)
You can mount the Smart-hopping Access Point to a wall using the mounting
screws and screw anchors supplied with the Access Point. We recommend that you
mount the Access Point high on the wall as close to the ceiling as possible.
To mount the Smart-hopping 2.0 Access Point 1.4 GHz and the plastic cover to a
wall:
Table 2-3: Required Wall Mounting Material
Required Parts
Illustration| Part Name| Quantity
867216 Access Point Kit -includes the following:| 1
| Access Point| 1
| Wall anchors| 2
| #6 1.25-Inch self-tapping screws| 2
| Mounting Bracket| 1
Access Point Mounting Kit IM4 – includes the following:| 1
| Cosmetic Cover| 1
| M3 x 0.5 x8 machine screws| 2
| Plastic Grommets (included with kit)| 2
Table 2-4: Wall Mounting Tools
Required Tools
| Tape Measure
| Level
| Hammer (when using wall anchors)
| Drill
| 7/32 -inch drill bit
| One-inch hole saw bit
| Phillips head screwdriver
| T-8 Torx screwdriver
| Marker
| Silicone Adhesive
Installation
Caution
To mount the Access Point to a wall, connector cables must come through the
wall.
-
Using a 7/32-inch drill bit, drill two pilot holes 0.219 inches (5.5 mm) in diameter, 3.394 inches (100 mm) apart (from the center points of the holes), and horizontally level with each other.
-
Tap a supplied plastic screw anchor into each pilot hole until it is flush with the wall surface.
-
Place the mounting bracket against the wall and line up the two holes.
-
Screw a supplied #6 x 1 1/4 inch self- tapping screw through each hole in the mounting bracket and into each screw anchor (or pilot hole).
-
Mark two holes at the locations shown in the illustration below.
-
Using a one-inch hole saw, drill out a one-inch hole around the two marks. This is for connector cables to come out of the wall and attach to the Access Point.
On the finished side of the wall, secure the one-inch grommets to the one-inch holes you drilled. You need adhesive to secure the grommet -
Attach the access point to the Cosmetic Cover using the two M3 x 0.5 x8 screws.
Place the Access Point face down.
Place the plastic cover over the Access Point.
Make sure the two holes in the plastic cover align with the corresponding holes in the Access Point.
Secure the Access Point to the cosmetic ring by screwing in the two M3 x 0.5 x8 machine screws into the holes in the cosmetic ring and into the Access Point. -
Pull the cables out from the wall (through the grommets) and connect them to the appropriate jacks on the Access Point.
Connect the Ethernet (network connection) cable to the LAN port on the Access Point. This provides power to the Access Point and lights up the LEDs on the front of the Access Point (if the other end of the Ethernet cable is connected to a Power over Ethernet Switch that is powered on).
If using two Remote Antennas, perform this step on both the RA1 and RA2 connectors on the Access Point:
• Connect the Coaxial cables (from the same side of the strand) to the RA coaxial connector.
• b) Connect the RJ-45 connector to the RA RJ-45 connector. -
While holding the Access Point chassis and attached plastic cover, fit the holes in the back of the Access Point chassis over the keyhole screws that protrude from the mounting bracket. Slide the Access Point chassis down so that the screw heads are securely in the grooves in the back of the Access Point chassis. You hear a click noise when the Access Point is securely placed in the mounting bracket.
Flush Ceiling Mount (with Cosmetic Ring)
You can mount the Smart-hopping Access Point on a ceiling tile and make the
Access Point flush with the tile (the Access Point protrudes 7-10 mm from the
finished side of the ceiling tile).
To mount the Smart-hopping 2.0 Access Point 1.4 GHz and cosmetic ring to a
ceiling tile:
Note
Wall anchors and #6 1.25-inch self-tapping screws are included with the Access
Point.
These are not used in this mounting method.
Table 2-5: Required Flush Mounting with Cosmetic Ring Material
Required Parts
Illustration| Part Name| Quantity
867216 Access Point Kit -includes the following:| 1
| Access Point| 1
| Mounting Bracket| 1
Access Point Mounting Kit IM2 – includes the
following:| 1
| Mounting Rail| 1
| M6 Push Nut| 2
| M6 x 1 Nut| 2
Table 2-5: Required Flush Mounting with Cosmetic Ring Material
Required Parts
Illustration| Part Name| Quantity
| Adapter Plates| 2
Access Point Mounting Kit IM3 – includes the
following:| 1
| Cosmetic Ring| 1
Table 2-6: Ceiling Flush Mounting Tools
Required Tools
| Tape Measure
| Pliers (or wrench)
| Marker
| Utility Knife
Installation
Prepare the Tile
-
Remove (or get) the ceiling tile on which you wish to mount the Access Point.
-
Measure the tile and mark the middle points on the long side of the finished side of the tile.
-
Place the mounting rail against the finished side of the tile, aligning the center of the mounting rail with the marks you made on the tile.
Make sure the side of the rail labeled faces you.
Cable Access
Flush Mount -
Place four marks on the inner edge of the ceiling tile at the points where the arrows of the markings on the mounting rail end.
<–Flush Mount Cutout–> -
Mark an X on the edge of the tile where the Cable Access marking on the mounting rail points. This ensures that the mounting rail faces the correct direction when you flip the tile to mount the mounting rail on its unfinished side (in a later step).
-
(Optional) Remove the mounting rail from the tile and draw lines that connect the points together.
-
After taking the mounting rail off of the tile, cut out the area of the tile between the dots you marked.The open space is where you mount the Access Point.
-
Put the tile and mounting rail aside.
Mount the Hardware
-
Perform this step to connect both adapter plates to the mounting bracket:
• Remove the adhesive strip from the adapter plate.
• Orient the adapter plate so the bolts face the same direction as the shoulder screws that are already attached to the bracket.
• Place the bolts through one set of outer holes on the bracket.
• Press the adapter plate firmly against the bracket to ensure a strong adhesive hold. -
Connect the mounting bracket to the Access Point. Make sure the side of the mounting bracket with the rectangular cutout lines up with the connectors on the back of the Access Point.
Line up the shoulder screws on the bracket with the 4 keyholes in the Access Point and slide the bracket toward the connectors. You hear a click noise when the bracket is secure on the Access Point. -
Place the Access Point assembly, finished side down, on a flat surface and place the Cosmetic Ring over the Access Point assembly.
-
Place the ceiling tile, with the unfinished side facing you, over the Access Point and Cosmetic Cover.
Make sure the Access Point is aligned along the bottom edge of the rectangular cutout in the ceiling tile (the bottom edge of cutout refers to the edge furthest away from the mark you made on the tile when preparing the tile).
This leaves space for the minimum bending radius of the cables. -
Place the mounting rail over the Access Point assembly, on top of the unfinished side of the tile.
Make sure the Flush Mount label on the Mounting Rail faces you and the Cable Access label points toward the Access Point connectors. Feed the 4 bolts through the holes marked F.
The Flush Mount Cutout arrows must align with the rectangular cutout in the tile. -
Secure the Access Point assembly and Mounting Rail to the tile:
• Place an M6 x 1 nut on each of the bolts from the adapter plate that stick through the holes in the mounting rail.
• Tighten the nuts and make sure you do not over tighten the nuts, which may break the tile. -
To secure the nuts, place a push nut over each of the bolts and push down until it sits on top of each nut.
You can do this by hand or use a tool, such as pliers or an open-ended wrench. -
Connect the Ethernet (network connection) cable to the LAN port on the Access Point. This provides power to the Access Point and lights up the LEDs on the front of the Access Point (if other end of Ethernet cable is connected to a powered Synchronization Unit).
If using two Remote Antennas (perform this step on both the RA1 and RA2 connectors on the Access Point:
• Connect the Coaxial cables (from the same side of the strand) to the RA coaxial connector.
• Connect the RJ-45 connector to the RA RJ-45 connector. -
Place the tile into the ceiling.
Replace other adjacent ceiling tiles, if necessary.
Below Ceiling Mount
You can mount the Smart-hopping Access Point below a ceiling tile. To mount
the Smart-hopping 2.0 Access Point 1.4 GHz to a ceiling tile:
Note
Because this installation option does not include a Cosmetic Cover to hide
the Remote Antenna cabling, only use this option with installations of stand-
along Access Points.
Table 2-7: Required Below Mounting Material
Required Parts
Illustration| Part Name| Quantity
867216 Access Point Kit -includes the following:
| Access Point| 1
| Mounting Bracket| 1
Access Point Mounting Kit IM2 – includes the
following:
| Mounting Rail| 1
| M6 Push Nut| 2
| M6 x 1 Nut| 2
| Adapter Plates| 2
Table 2-8: Required Below Mounting Tools
Required Tools
| Tape Measure
| Drill
| 1/4 -inch drill bit
| One-inch hole saw bit
| Marker
| Pliers (or wrench)
Installation
Prepare the Tile
-
Remove (or get) the ceiling tile on which you wish to mount the Access Point.
-
Measure the tile and mark the middle point in the area between the finished an unfinished areas on the long side of the tile.
-
Place the mounting rail against the finished side of the tile, aligning the center of the mounting rail with the marks you made on the tile. Make sure the side of the rail labeled faces you.
Cable Access
Below Mount -
Place marks on the ceiling tile in the holes labeled B on the mounting rail.
-
Place a mark on the ceiling tile at the left Cable Access arrow, one inch above the edge of the Mounting Rail.
-
Remove the Mounting Rail.
With the finished side of the ceiling tile facing you, use the drill and 1/4-inch drill bit to drill four holes where you marked in the B holes in the previous steps.
Using the mark you made in Step 5 as the center point, use a one-inch hole saw to drill out a one-inch hole in the ceiling tile. This hole is to pass the Ethernet/LAN cable through the ceiling to the Access Point. -
Put the tile and mounting rail aside.
Mount the Hardware
-
Perform this step to connect both adapter plates to the mounting bracket:
• Remove the adhesive strip from the adapter plate.
• Orient the adapter plate so the bolts face the same direction as the shoulder screws that are already attached to the bracket.
• Place the bolts through one set of outer holes on the bracket.
• Press the adapter plate firmly against the bracket to ensure a strong adhesive hold. -
Place the Mounting Rail on the unfinished side of the tile. Make sure the Cable Access arrows on the Mounting Rail are pointed toward the one-inch hole you drilled for the cables.
The four 1/4-inch holes you drilled must align with the B holes in the Mounting Rail. -
Connect the Mounting Bracket and Rail to the ceiling tile:
• Pick up the Mounting Bracket and orient it so that the rectangular cutout on the Bracket is in the same direction as the Cable Access arrows on the Mounting Rail.
• With the adapter plates adhered to the Mounting Bracket, insert the 4 bolts from the finished side of the tile through the four 1/4-inch holes in the ceiling tile and through the B holes in the Mounting Rail. -
Secure the Mounting Bracket and Mounting Rail to the tile:
• Place an M6 x 1 nut on each of the bolts from the adapter plate that stick through the holes in the mounting rail.
• Tighten the nuts and make sure you do not over tighten the nuts, which may break the tile. -
To secure the nuts, place a push nut over each of the bolts and push down until it sits on top of each nut.
You can do this by hand or use a tool, such as pliers or an open-ended wrench -
Connect the Ethernet (network connection) cable to the LAN port on the Access Point. This provides power to the Access Point and lights up the LEDs on the front of the Access Point (if other end of Ethernet cable is connected to a powered Synchronization Unit).
-
With the LAN cable now connected, attach the Access Point to the Mounting Bracket and make sure that the rectangular cutout in the Mounting Bracket aligns with the shape of the Access Point.
Insert the four small shoulder screws on the Bracket into the keyholes on the AP and slide the AP toward the LEDs. You hear a click noise when the Access Point is secured to the Bracket. -
Place the tile into the ceiling.
Replace other adjacent ceiling tiles, if necessary.
Below Ceiling Mount (with Cosmetic Cover)
You can mount the Smart-hopping Access Point below a ceiling tile and place a
cosmetic cover around the Access Point to hide exposed cables. To mount the
Smart-hopping 2.0 Access Point 1.4 GHz and cosmetic ring to a ceiling tile:
Note
Wall anchors and #6 1.25-inch self-tapping screws are included with the
Access Point.
These are not used in this mounting method.
Table 2-9: Required Below Mounting with Plastic Cover Material
Required Parts
Illustration| Part Name| Quantity
867216 Access Point Kit -includes the following:| 1
| Access Point| 1
| Mounting Bracket| 1
Access Point Mounting Kit IM2 – includes the
following:| 1
| Mounting Rail| 1
| M6 Nut| 2
| M6 x 1 Nut| 2
Table 2-9: Required Below Mounting with Plastic Cover Material
Required Parts
Illustration| Part Name| Quantity
| Adapter Plates| 2
Access Point Mounting Kit IM4 – includes the
following:| 1
| Cosmetic Cover| 1
| M3 x 0.5 x8 machine screws| 2
| Plastic Grommets (included with kit)| 2
Table 2-10: Required Below Mounting Tools
Required Tools
| Tape Measure
| Drill
| 1/4 -inch drill bit
| One-inch hole saw bit
| Pliers (or wrench)
Required Tools
| T-8 Torx screwdriver
| Marker
| Silicone Adhesive
Installation
Prepare the Tile
-
Remove (or get) the ceiling tile on which you wish to mount the Access Point.
-
Measure the tile and mark the middle point in the area between the finished an unfinished areas on the long side of the tile.
-
Place the mounting rail against the finished side of the tile, aligning the center of the mounting rail with the marks you made on the tile. Make sure the side of the rail labeled faces you.
Cable Access
Below Mount -
Place marks on the ceiling tile in the holes labeled BC on the mounting rail.
-
Using a ruler (recommended) or visually, mark two RA Cable Hole marks on the ceiling tile where the
Cable Access arrows and the lines at the ends of the Center of RA Cable Holes text on the Mounting Rail intersect. -
Remove the Mounting Rail. With the finished side of the ceiling tile facing you, use the drill and 1/4-inch drill bit to drill four holes where you marked in the BC holes in the previous steps.
Using the marks you made in Step 5 as the center point, use a one-inch hole saw to drill out two one-inch holes in the ceiling tile. These holes are to pass cable connectors through the ceiling to the Access Point. -
Lightly coat both plastic grommets with silicone adhesive and insert them into both one-inch holes in the tile (Insert the grommets from the finished side of the tile).
-
Put the tile and mounting rail aside.
Mount the Hardware
-
Attach the access point to the Cosmetic Cover using the two M3 x 0.5 x8 screws.
• Place the Access Point face down.
• Place the plastic cover over the Access Point.
• Make sure the two holes in the plastic cover align with the corresponding holes in the Access Point.
• Secure the Access Point to the cosmetic ring by screwing in the two M3 x 0.5 x8 machine screws into the holes in the cosmetic ring and into the Access Point. -
Perform this step to connect both adapter plates to the mounting bracket:
• Remove the adhesive strip from the adapter plate.
• Orient the adapter plate so the bolts face the same direction as the shoulder screws that are already attached to the bracket.
• Place the bolts through one set of outer holes on the bracket.
• Press the adapter plate firmly against the bracket to ensure a strong adhesive hold. -
Place the Mounting Rail on the unfinished side of the tile. Make sure the Cable Access arrows on the Mounting Rail are pointed toward the one-inch holes you drilled in the tile for the cables.
The four 1/4-inch holes you drilled must align with the BC holes in the Mounting Rail. -
Connect the Mounting Bracket and Rail to the ceiling tile:
• Pick up the Mounting Bracket and orient it so that the rectangular cutout on the Bracket is in the same direction as the Cable Access arrows on the Mounting Rail.
• With the adapter plates adhered to the Mounting Bracket, insert the 4 bolts from the finished side of the tile through the four 1/4-inch holes in the ceiling tile and through the B holes in the Mounting Rail. -
Secure the Mounting Bracket and Mounting Rail to the tile:
• Place an M6 x 1 nut on each of the bolts from the adapter plate that stick through the holes in the mounting rail.
• Tighten the nuts and make sure you do not over tighten the nuts, which may break the tile. -
To secure the nuts, place a push nut over each of the bolts and push down until it sits on top of each nut.
You can do this by hand or use a tool, such as pliers or an open-ended wrench -
Pull the LAN cable and Remote Antenna cables (if applicable) through the 1-inch holes on the unfinished side of the ceiling tile so the cable connectors end up on the finished side of the tile.
Note: If connecting to two remote antennas, pull two cables through one of the holes and three cables through the other hole.
If using two Remote Antennas (perform this step on both the RA1 and RA2 connectors on the Access Point):
• Connect the Coaxial cables (from the same side of the strand) to the RA coaxial connector.
• Connect the RJ-45 connector to the RA RJ-45 connector. -
With all of the necessary cables now connected, attach the Access Point and the attached Cosmetic Cover to the Bracket, making sure the rectangular cutout in the Bracket aligns with the shape of the Access Point.
Insert the 4 small shoulder screws on the Bracket into the keyholes on the Access Point and slide the Access Point with Cosmetic Cover in the direction of the arrow molded into the Cover. When the Access Point is secured to the bracket, you hear a click noise.
Note: there are 4 “tick” marks on the outside surface of the Cosmetic that indicate the location of the keyholes to help with alignment to the shoulder screws. -
Place the tile into the ceiling.
Replace other adjacent ceiling tiles, if necessary.
Mounting the Smart-hopping Access Point with a Tether Mount (Optional)
In addition to (not as a substitute for) any of the other installation options
described in this chapter, you can add a tether from the tether point on the
Access Point to a permanent building structure for extra security (where local
building codes require this type of installation [such as earthquake-prone
areas]).
Note
You must connect the other end of the tether to a stable part of the
building frame.
Perform these steps before replacing the ceiling tile onto the ceiling frame
structure, or prior to attaching the Access Point to a wall:
-
There is a tether point on the Access Point housing to allow for tethering to other permanent structures. Use a 18- gauge galvanized steel wire. Make sure the tether wire complies with national and local building standards.
Thread the wire through the tether point in the Access Point housing. The tether wire must have three tight turns of the wire within 1.5 inches of the end of the wire. -
Connect the other end of the tetherwire to a permanent structure withinthe ceiling. The tether wire may be attached to any of the following:
• Wood or metal stud framing
• Blocking attached to wood or metal framing
• Plywood adequately attached to wood or metal stud framing, reinforced concrete, or reinforced masonry
• Reinforced concrete
• Reinforced masonry
Installation Procedure
To install the Smart-hopping 2.0 Access Point 1.4 GHz:
- Mount the Access Point where they can communicate with Patient Monitors. Note these guidelines when installing the Smart-hopping Access Point:
• You may mount each Smart-hopping Access Point as follows:
– To a wall (fixed mount) as described on page 2-3 and page 2-7
– Below a ceiling (flush mount) as described on page 2-12
– Below a ceiling (fixed mount) as described on page 2-20 and page 2-29
• Mounting instructions for Smart-hopping Remote Antenna are available in the Smart-hopping Remote Antenna Installation Guide document, which is available on the Philips InCenter web site.
• Use a category 5e (or better) Unshielded Twisted Pair cable to connect each Smart- hopping Access Point to the Smart-hopping infrastructure.
• The total length of UTP cable from the Smart-hopping access point to the Smart- hopping Synchronization Unit to the network switch cannot exceed 328 ft. (100 m).
• Use only the supplied, unmodified 74 ft. (22.6m) Coax and UTP cable bundle when connecting Remote Antennas to the Access Point.
Caution
The coaxial cable bundle cable length may vary from 69 – 79 feet. When
planning remote antenna placement, ensure the maximum length between the
Access Point and the remote antenna is 69 feet.
-
If you have installed Remote Antennas, connect each Smart-hopping Access Point to its installed Remote Antenna(s) using the supplied, unmodified 74 ft. (22.6m) Coaxial and UTP cable bundle(s).
Caution
Ensure you do not kink the RA Coax-and-UTP-cable-bundle during installation. You must maintain a minimum 2.5 inch (64-mm) bend radius for the RA Coaxial and UTP cable bundle throughout the installation.
a) Be sure to label the UTP cable bundles and the Remote Antennas themselves as RA 1 and RA 2 corresponding to the cable connections you made when mounting the Access Point.
b) Route a Category 5e (or higher) RJ-45 cable between each Smart-hopping Access Point and the equipment closet in which the Smart-hopping infrastructure devices are installed. -
Connect the Access Point to the Smart-hopping Sync Unit only when appropriate as part of the overall Smart-hopping installation.
a) After initially connecting and powering the Access Point, verify that the two status LEDs on each of its connected Remote Antennas are lit. You can verify this by viewing the Access Point status page presented in the APC web interface.
Note
If after connecting and powering the Access Point, its LEDs are lit red and the Remote Antenna LEDs do not light at all, open the APC web interface, click System in the View Device tree, click Configure and then select the Advanced tab. Verify that the Allow new APs to be added automatically option is set to True. This option must be set to True for the Access Point and Remote Antenna LEDs to light properly.
b) Reboot the Access Point.
The Access Point and its connected Remote Antennas ought to be listed in the APC web interface. Press F5 to refresh the web interface display if the Access Point and RAs are not listed.
c) For instructions on the Access Point startup sequence, see page 2-43.
Access Point Startup Sequence
Upon power on/start up, the Smart-hopping Access Point performs a Power-on-
Self-Test (POST) to ensure that its basic components are fully functional.
This test detects critical system failures. As the test progresses the state
of the system is displayed through a series of color combinations on the four
LEDs on the front of the Access Point. All four LEDs illuminate Amber color,
the Power/Sync LED illuminates green, the Network and Radio LEDs flash green,
and the LED turns off (not illuminated) indicating correct startup. If a test
fails, the LEDs are left in a state which indicates the point at which the
test failed. When the system is transmitting data the radio and network LEDs
on the Access Point flicker on and off as data is transmitted.
Access Point Configuration Information
Refer to the Smart-hopping 2.0 Infrastructure Installation and Service Guide
for complete details about configuring the Smart-hopping Access Point.
Chapter 3: Maintaining the Smart-hopping 2.0 Access Point 1.4 GHz
Troubleshooting the Access Point Using its LEDs
Complete test and inspection procedures for the Access Point are provided in
the Smart- hopping 2.0 Infrastructure Installation and Service Guide. Perform
these procedures
when the Smart-hopping Access Point is initially installed and after servicing
the unit. For information on the Status LEDS, see “Smart-hopping Access Point
Status LEDs” on page 1-8. If the Status LED indicators are not functioning
properly as indicated, perform the appropriate troubleshooting procedures. If
you cannot resolve the problem with the LEDs, then note the LED lighting and
sequence of colors, replace the unit, and send the defective unit back to
Philips with the explanation of LEDs sequence and colors displayed. Refer to
the Smart-hopping 2.0 Infrastructure Installation and Service Guide for
complete Access Point replacement procedures.
Replacing the Access Point
To repair a non-working Access Point, replace the unit. Individual Access
Point, PC assemblies, or sub-assemblies are not available for purchase.
If an Access Point is not working properly, call the Philips Customer Care
Solutions Center for information on ordering a replacement unit and returning
the defective unit.
Use the following toll-free number to contact the Philips Customer Care
Solutions Center:
• Telephone: (+1) 800-722-9377
When replacing defective equipment on your Smart-hopping infrastructure, be
sure to follow the procedures given in the Smart-hopping 2.0 Infrastructure
Installation and Service Guide.
Ordering Replacement Units
This section contains information for ordering replaceable parts and
assemblies for the Smart- hopping 2.0 Access Point 1.4 GHz and Remote Antenna.
The parts are listed in tables by major assemblies. Each table contains the
orderable part number (new or exchange) and a description of the part. You can
order only the part numbers listed in Table 3-1.
To order a replacement part:
- Identify the faulty component, part, subassembly or assembly.
- Locate the replacement part number for that major assembly in Table 3-1.
- Order replaceable parts from your nearest Philips Sales/Service office or from the Philips Support Materials Organization. A complete listing of the Philips Sales/Service Offices addresses with phone numbers is located at the end of this chapter.
Ordering Information.
Table 3-1: Ordering Access Point and Remote Antenna Replacement Parts
Device | Part Numbers |
---|---|
Smart-hopping 2.0 Access Point 1.4 GHz | 453564883261 |
Smart-hopping Mounting Options:
Basic Access Point Kit (includes hardware for wall mounting)| 453564979681
(IM4)
• Wall Mounting with Cover| 453564979681 (IM4) with 453564979681 (IM4) Plastic
Access Point cosmetic cover
• Below Ceiling Mount| 453564979661 (IM2) Metal rail and hardware
• Below ceiling Mount with Cover)| 453564979661 (IM2) Metal rail and hardware
with 453564979681 (IM4) Plastic Access Point cosmetic cover
• Flush Ceiling Mount with Cover| 453564979661 (IM2) Metal rail and hardware
with 453564979671 (IM3) Plastic cosmetic ring
• Remote Antenna connections| 453564979661 Metal rail and hardware (IM2)
Philips Sales and Support Offices Worldwide
Please call your local sales office listed in your telephone directory or a
regional office listed below for the location of your nearest sales office.
CORPORATE HEADQUARTERS:
Philips Medical Systems
Netherlands B.V.
Postbus 10.000
5680 DA Best
Netherlands| UNITED STATES:
Philips Medical Systems North America Company
222 Jacobs Street
Cambridge, MA 02141
800-291-6743
---|---
CANADA:
Philips Medical Systems
281 Hillmount Road
Markham, ON L6C 2S3
800-291-6743| EUROPE, MIDDLE EAST
AND AFRICA:
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard Str. 2
71034 Böblingen
Germany
Fax: (+49) 7031 463 1552
LATIN AMERICA HEADQUARTERS:
Philips Medical Systems
1550 Sawgrass Corporate Parkway #300
Sunrise, FL 33323
Tel: 954-835-2600
Fax: 954-835-2626| ASIA PACIFIC
HEADQUARTERS:
Philips Medical Systems
30/F Hopewell Centre
17 Kennedy Road
Wanchai
Hong Kong
Tel: (852) 2821 5888
Fax: (852) 2527 6727
Any Changes or modifications not expressly approved by the party responsible
for compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- onsult the dealer or an experienced radio/TV technician for help.
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator & your body.
Part Number: 453464900911
Printed in the U.S.A.
June 2022
First Edition
453564900911
DRAFT – 03 JUN 2022 13:32
Documents / Resources
|
PHILIPS 867216 Smart Hopping 2.0 Access
Point
[pdf] Installation Guide
867216A, PQC-867216A, PQC867216A, 867216 Smart Hopping 2.0 Access Point,
867216, Smart Hopping 2.0 Access Point, 2.0 Access Point, Access Point
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>