Diamond SP Mustang
High resolution strong cation exchange resin
Instruction for use
Introduction
Ion exchange chromatography(IEC)is a very effective method for the separation
and purification of biomolecule. The method mainly relies on the interaction
between positive and negative charges, and uses the charge properties and
differences of different biological molecules under specific conditions to
separate them. It has the characteristics of high load, good resolution,
controllable condition and easy scale-up. It has been widely used in medicine,
chemical industry, metallurgy, food and other fields.
IEC resin is composed of three parts: (1)Cross-linked agarose matrix, that has
the characteristics of porous, hydrophilic and good chemical stability;
through the modification to traditional agarose molecules, it has stronger
mechanical properties, which is called high rigid agarose, namely Diamond
matrix; (2)Functional groups fixed to the matrix, which are electrically
charged groups, determining the properties of the IEC resin; (3)An ion(called
an equilibrium ion)that has an opposite charge to the functional group and can
be reversibly bound to the functional group. Diamond SP Mustang is a strong
cation exchange resin formed by coupling sulfopropyl to a fine, highly rigid
agarose microsphere with a higher resolution than Diamond SP, which is
suitable for fine or medium purification in the chromatography step.
Technical characteristics
Appearance | White slurry, can be layered |
---|---|
Matrix | High rigidity agarose |
Functional group | Sulfopropyl |
Average particle size+ | 36-44μm |
Total ionic capacity | 130~160μmol H |
Dynamic loads | > 70 mg lysozyme/mL resin |
95mg BSA/mL resin
Max pressure| 0.5 MPa
Max flow rate| 450cm/h(0.5MPa, BXK100/500, H=20cm,20 ℃)
Chemical Stability| Stable in common aqueous buffers: 1M NaOH++,1M HAC++,70%
ethanol , 20% ethanol, 6M GuHCl,8M Urea, 100mM thioglycerol+25mM
Tris pH8.5, 0.1M citric acid +8M urea +1M NaCl pH2.5
Avoid contact with oxidizing agents, cationic detergents.
pH stability| 3~14(CIP), 4~12 (Working)
Temperature tolerance| 2~40 ℃, Can’t freeze, Can tolerate 121 ℃ high pressure sterilization
(50mM PB , pH7)
Storage| 2~30℃, 20% ethanol with 0.2M NaAc
Recommend flow rat| 100~300cm/h
+ Average particle size is the accumulated resin particle size of packing
volume distribution
.++1M NaOH and 1M HAc only be used for cleaning.
Method of chromatographic
3.1 Column packing
Note:It is best to equilibrate the resin slurry to room temperature
before column packing.
- According the column volume to calculate the amount of resin.
Resin volume=column volume×1.15(Compression factor=1.15) According to the volume of the settlement resin required, the suspended slurry of the resin required is calculated by the follow:
Required resin slurry 1 volume = Settlement resin volume ÷ Resin slurry 1 concentration. The original concentration of resin slurry 1 is shown in the follow table.
Pack size | Resin slurry1 concentration (%) |
---|---|
25mL、100mL、500mL、1L、5L、10L | 80 |
20L、40L | 75 |
1: It refers to the original packaging resin slurry sold by Bestchrom.
Note: For non-original packaging, customer can calculate the required
volume according to the actual concentration of resin slurry.
-
Washing the resin: Suspend the resin by shaking and pour into a funnel, remove the liquid, and wash 3 times with about 3mL distilled water/mL resin. Use a glass stick or stirrer to stir each time you add the distilled water, in order to better clean the storage solution.
-
Prepare the packing slurry: Move the washed resin from the funnel into a beaker or other appropriate container, and add distilled water to obtain a 50%~75% slurry concentration, stir well and set aside for use.
-
Take a cleaned BXK column (BXK series columns with diameters ranging from 1cm to 30cm can satisfy different scale chromatography applications). Take BXK16/20 for example, purge the bubbles trapped at the end-piece net by draining some distilled water through the column outlet. Leave about 1cm water at the bottom of the column and close the bottom outlet. Adjust the column so that it is perpendicular to the ground.
-
Slowly pour the slurry into the column at one time (use a packing reservoir if necessary). Do not bring any air bubbles into the column. Packing reservoir: Empty glasstube with same diameter as the BXK column.
-
Fill the remainder of the column with packing solution. Connect the packing reservoir to the chromatography system, open the flow rate, drain the bubbles in the hose, close the flow rate, and tighten the top cover of the packing reservoir.
-
After pouring, stir well again with stirrer, and then wash the resin particles on the inner wall of the column from top to bottom with the packing solution, and let the resin settle naturally until there is about 1cm of clarifying solution on the suspension.Mount the adapter and connect the adapter to the chromatography system or peristaltic pump.Lower the adapter to descend to contact with the clarifying solution and tighten the sealing ring after it is fully immersed in the clarifying solution.With the outlet of the top piece is opened, slowly move the adapter down until all bubbles are drained.
Note: This operation is only applicable to BXK 100 and above columns.Flushing the inner wall reduces the resin particles sticking between the seal ring and the column wall, avoiding the risk of leakage. -
When the bed height is 10cm, the flow rate can be set to 600cm/h. Open the bottom outlet, start the pump and run the setting flow rate until the bed is stability. Mark the bed height.
-
Remove the packing reservoir (if any), lower the adapter to about 0.5cm above the resin surface, and continue to press the column using the above flow rate until the bed is completely consolidated, marking the consolidated bed height.
-
Stop the pump, open the outlet of the top piece, close the outlet of the bottom piece, loosen the seal ring slightly, press the rubber surface according to the compression ratio of 1.15,tighten the seal ring, close the outlet, and complete the column packing.
3.2 Evaluation of Packing
-
Test the column efficiency to check the quality of packing.The tests are required after the column packing, during the column working life and when the separation and purification performance is deteriorated. The method of the expressing the efficiency of a packed column is in terms of the height equivalent to a theoretical plate(HETP) and the asymmetry factor(As).
-
Acetone or NaCl can be used as sample for the testing. Sample solution and eluent buffer can be prepared according to the following table.
| Acetone method| NaCl method
---|---|---
Sample| 1.0%(v/v)acetone in water| 0.8M NaCl in water
Loading| 1.0%CV| 1.0%CV
Buffer| Water| 0.4M NaCl in water
Flow rate| 30cm/h| 30cm/h
Monitor| UV280 nm| Conductivity -
Method for measuring HETP and As:
According the UV curve or the conductivity curve to calculate the column efficiency(HETP), and the asymmetry (As):
HETP=L/N
N=5.54(VR/Wh)²
Note: VR = retention volume
Wh = half-peak width
L = column height
N = the number of theoretical plates
(The units of VR and Wh should be the same) As=b/a Note:
a= 1st half peak width at 10% of peak height
b= 2nd half peak width at 10% of peak height -
Evaluation the column packing As a guideline, if the value of HETP is less than 3 times the average particle size (d50)of the resin and the As is between 0.8~1.8, it is very efficient. The unsatisfactory results should be analyzed and the column should be repacked .
3.3 Chromatographic method
- Buffer selection: Buffer salts whose buffer groups do not act on the resin should be selected. The buffer solution with low salt (less than 5mS/cm) and low pH (usually 1 pH unit lower than the isoelectric point of the target) should be adopted to facilitate the combination of substances. Meanwhile, the stability of samples in the buffer solution should be considered. Elution buffers are usually made by adding a high concentration of salt (e.g.1M NaCl) or high pH elution to balance buffer.
- Flow rate: According the column bed high to use the flow rate 100~300cm/h ,the higher column bed high and lower flow rate.
- Sample preparation: In order to prevent blocking of the column, the sample needs to be filtered by microporous membrane of 0.45μm before loading, the pH and conductivity of the sample are adjusted to be consistent with the equilibration buffer (the pH and conductivity of the sample can be adjusted by dilution, ultrafiltration, and desalination with Bestdex G-25).
- Equilibration: Washing the column with equilibration buffer until the pH and conductivity of the column outlet buffer are basically the same as the equilibration buffer, which usually needs 3-5C V.
- Sampling: The loading volume is determined according to the substance content in the sample and the binding load of Diamond SP Mustang.
- Rinse: Wash the column with equilibration buffer until the UV absorption value is reduced to an appropriate value.
- Elution: Linear gradient or step-gradient can be used to increase the elution strength in the elution buffer, eluting substances with different binding strength from the chromatography column, collecting different components and detecting the location of the target.
- Regeneration: Flush the column with a high concentration of salt (eg: 2M NaCl)
- Rebalancing: After rinsing with equilibration buffer, the second sample can be loaded and repeated.
Application
Wheat germ agglutinin and β-Lactoglobulin were separated by Diamond SP Mustang
Column: EzScreen 4.2mL
Equilibrium buffer:20Mm HCOOH pH 4.0
Elution buffer : 20mM HCOOH +0.75M LiCl pH 4.0
Sample: WGA and β -lactoglobulin mixture
Cleaning-in-place(CIP)
With the increasing use of chromatography resin, the accumulation of
contaminants on the chromatography column is also increasing.Cleaning-in-place
can prevent the accumulation of contaminants and maintain a stable working
state. Determine the frequency of CIP according to the degree of contamination
of the resin (if the contamination is serious, CIP should be carried out after
each use to ensure repeatability of the results).
The recommended CIP for different types of impurities and contaminants are as
follows:
- 2~3CV of 2M NaCl was used to wash out the proteins with relatively tight binding.
- Removal of strong hydrophobic proteins and precipitating proteins: Clean with 1M NaOH of 2~3CV first, then rinse immediately with 5~10CV pure water.
- Removal of lipoproteins and lipids: Clean with 70% ethanol or 30% isopropanol by volume of 5~10CV first, then rinse with pure water by volume of 5~10CV.
- The above two cleaning conditions can also be combined for cleaning, namely 30% isopropanol solution containing 1M NaOH.
Note: 70% ethanol or 30% isopropanol should be degassed before use. In the CIP process, the flow rate can be chosen as 30~60cm/h. Reverse flushing can be used hen the blockage is serious.
Sterilization
Since the 20% ethanol with 0.2M NaAc preservation solution does not have sterilization and depyrogenation, it is recommended that Diamond SP Mustang can be treated with 1M NaOH for more than 0.5-1h to reduce the risk of microbial contamination before and during use. It can also be autoclaved at 121 ℃ (50mM PB, pH7) for 30min.
Storage
Diamond SP Mustang is supplied in 20% ethanol with 0.2M NaAc or 2% benzyl alcohol with 0.2M NaAc. It should be stored in 20% ethanol with 0.2M NaAc and sealed at 2-30 ℃ after use, in order to prevent ethanol volatilization and microbial growth, it is recommended to replace 20% ethanol with 0.2M NaAc every 3 months.
Disposal and Recycling
Diamond SP Mustang is very difficult to degrade in nature, incineration is recommended to protect the environment.
Order information
Product | Code No. | Pack size |
---|---|---|
Diamond SP Mustang | AI0191 | 25mL |
AI0192 | 100mL | |
AI303211 | 500mL | |
AI0193 | 1L | |
AI0194 | 5L | |
AI0195 | 10L | |
AI303215 | 20L | |
AI303216 | 40L | |
Prepacked columns | Code No. | Pack size |
--- | --- | --- |
EzFast SP Mustang | EI303251 | 5×1mL |
EI303203 | 1×5mL | |
EI303253 | 5×5mL | |
EzScreen SP Mustang | EI303202 | 1×4.2mL |
EI303252 | 5×4.2mL | |
EzLoad 16/10 SP Mustang | EI303204 | 1 pcs |
EzLoad 26/10 SP Mustang | EI303206 | 1 pcs |
SMSAI019 V2.01
021-58953301
[ www.bestchrom.com](http://www.bestchrom.com)
Documents / Resources
|
BESTCHROM Diamond SP Mustang High Resolution Strong Cation Exchange
Resin
[pdf] Instruction Manual
Diamond SP Mustang High Resolution Strong Cation Exchange Resin, Mustang High
Resolution Strong Cation Exchange Resin, Resolution Strong Cation Exchange
Resin, Strong Cation Exchange Resin, Cation Exchange Resin, Exchange Resin,
Resin
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>