APG Sensors MPX Magnetostrictive Level Sensors Installation Guide
- June 2, 2024
- APG Sensors
Table of Contents
- MPX Magnetostrictive Level Sensors
- Description
- How To Read Your Label
- Warranty
- Dimensions
- Installation Guidelines & Instructions
- Sensor and System Wiring Diagrams
- General Care
- Repair Information
- Hazardous Location Wiring
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Sensors MPX Magnetostrictive Level Sensors
Installation Guide
For Custody Transfer API 18.2 Compliant Probes
MPX Magnetostrictive Level Sensors
Thank You
Thanks for purchasing a MPX series magnetostrictive level sensor from us! We
appreciate your business and your trust. Please take a moment to familiarize
yourself with the product and this manual before installation. If you have any
questions, at any time, don’t hesitate to call us at 888525-7300.
NOTE: Scan the QR code to the right to see the full user manual on your
tablet or smartphone. Or visit
www.apgsensors.com/support to find it on
our website.
https://www.apgsensors.com/sites/default/files/manuals/MPX- API-182-manual.pdf
Description
The MPX series magnetostrictive level sensor provides highly accurate and repeatable level and temperature readings in a wide variety of liquid level measurement applications, including API 18.2 Custody Transfer. It is certified for installation in hazardous areas in the US and Canada for Class I, Division 1 & 2 and Class I, Zones 1 and 2 environments.
How To Read Your Label
Each label comes with a full model number, a part number, and a serial number.
The model number for the MPX will look something like this:
SAMPLE: MPX-T5-JI-P2ST-180-AP
The model number correlates with all the configurable options and tells you
exactly what you have. Compare the model number to the options on the
datasheet to identify your exact configuration. You can also call us with the
model, part, or the serial number and we can help you.
You’ll also find all hazardous certification information on the label.
Warranty
This product is covered by APG’s warranty to be free from defects in material
and workmanship under normal use and service of the product for 24 months. For
a full explanation of our Warranty, please visit https://www.apgsensors.com
/about-us/terms-conditions. Contact Technical Support to receive a Return
Material Authorization before shipping your product back.
Scan the QR code below to read the full explanation of our Warranty on your
tablet or smartphone.
https://www.apgsensors.com /about-us/terms-conditions
Dimensions
IMPORTANT: WARNING — A SEAL SHALL BE INSTALLED WITHIN 18 inches OF THE
ENCLOSURE;
DANGER: OPEN CIRCUIT BEFORE REMOVING COVER or KEEP COVER TIGHT WHILE
CIRCUITS ARE ALIVE;
DANGER: WARNING — EXPLOSION HAZARD — SUBSTITUTION OF COMPONENTS MAY
IMPAIR SUITABILITY FOR CLASS I, DIVISION 2;
Installation Guidelines & Instructions
The MPX should be installed in an area–indoors or outdoors–which meets the following conditions:
- Ambient temperature between -40°C and 85°C (-40°F to +185°F)
- Relative humidity up to 100%
- Altitude up to 2000 meters (6560 feet)
- IEC-664-1 Conductive Pollution Degree 1 or 2
- IEC 61010-1 Measurement Category II
- No chemical corrosive to stainless steel (such as NH3, SO2, Cl2, etc.) for MPX-R
- No chemicals incompatible with Titanium Grade 2 for MPX-T
- Ample space for maintenance and inspection
Additional care must be taken to ensure:
- The probe is located away from strong magnetic fields, such as those produced by motors, transformers, solenoid valves, etc.
- The medium is free from metallic substances and other foreign matter.
- No ignition hazards exist due to impact or friction with the titanium stem.
- The probe is not exposed to excessive vibration.
- The float(s) fit through the mounting hole. If the float(s) does/do not fit, it/they must be mounted on the stem from inside the vessel being monitored.
- The float(s) is/are oriented properly on the stem (See Figure 5.1 below). MPX-R and MPX-T floats are typically installed by customer.
IMPORTANT: MPX level sensor MUST be installed according to drawing 9003468 (Hazardous drawing 9003468 (Hazardous Installation and Non-Inc endive Wiring Drawing) in section 9 to meet listed approvals. Faulty installation will invalidate all safety approvals and ratings.
Installation Instructions:
- If your sensor’s stem and floats fit through the mounting hole, carefully lower the assembly into the vessel, then secure the sensor to the vessel.
- If the floats do not fit, mount them on the stem from inside the vessel being monitored. Then secure the sensor to the vessel.
- For sensors with float stops, refer to the assembly drawing included with the sensor for float stop installation locations.
Electrical Installation Instructions:
- Remove the housing cover of your MPX.
- Feed system wires into MPX through 3/4″ NPT conduit openings. Fittings must be UUCSA Listed for CSA installation.
- Connect wires to MPX terminals. Use crimped ferrules on wires, if possible.
- Replace housing cover.
- For EMI protection, either connect the grounding screw (see Dimensions section 4) to an earth ground, or ensure that tank mounting of MPX is grounded.
See Modbus System Wiring Diagrams (section 6) for Modbus wiring examples.
IMPORTANT: Floats must be oriented properly on the stem, or sensor
readings will be inaccurate and unreliable. Untapered floats will have a
sticker or “N” indicating the top of the float. Remove sticker prior to use.
DANGER: WARNING — EXPLOSION HAZARD — DO NOT DISCONNECT EQUIPMENT UNLESS
POWER HAS BEEN SWITCHED OFF OR THE AREA IS KNOWN TO BE NON-HAZARDOUS;
Warning: — The model MPX-T contains titanium in excess of 7.5% for Group
II and care needs to be taken to avoid an ignition hazards due to impact or
friction;
Sensor and System Wiring Diagrams
Modbus System Wiring
Note: Earth Ground terminal is factory connected to internal housing
ground lug.
NOTE: For APG Modbus programming instructions, please see MPX user manual.
APG Modbus software can be downloaded from
www.apgsensors.com/support.
Modbus System Wiring with RST-6001
Note: An independent +12-24 Vdc power supply is required when using an
RST-6001 Modbus Controller. The RST-6001 can only supply ±5 Vdc, not the
+12-24 Vdc required by the MPX.
Note: Earth Ground terminal is factory connected to internal housing
ground lug.
IMPORTANT: For EMI protection, either connect the grounding screw (see
section 4) to an earth ground, or ensure that tank mounting of the MPX is
grounded.
IMPORTANT: Refer to section 9 for Hazardous Location Wiring.
General Care
Your level sensor is very low maintenance and will need little care as long as
it was installed correctly. However, in general, you should periodically
inspect your MPX unit to ensure the stem is free of any heavy buildup that
might impede the movement of the float(s). If sediment or other foreign matter
becomes trapped between the stem and float(s), detection errors can occur.
If you need to remove the floats(s) from the stem of your MPX, be sure to note
the orientation of the float(s) prior to removal. This will help ensure proper
re-installation of the float(s).
Also, ensure that the housing cover is snuggle secured. If the cover becomes
damaged or is misplaced, order a replacement immediately.
DANGER: WARNING — POTENTIAL ELECTROSTATIC CHARGING HAZARD – CLEAN ONLY
WITH A DAMP CLOTH;
Repair Information
If your MPX level sensor needs repair, contact us via email, phone, or online chat on our website. We will issue you an RMA number with instructions.
- Phone: 888-525-7300
- Email: sales@apgsensors.com
- Online chat at www.apgsensors.com
IMPORTANT: All repairs and adjustments of the MPX level sensor must be made by the factory. Modifying, disassembling, or altering the MPX is strictly prohibited.
Hazardous Location Wiring
INSTALLATION IN:
CLASS I, DIV 1 & 2, GROUPS C & D, MAX. TEMP. 85° C
Ex d IIB : Ex nA IIB
CLASS I, ZONE 1; AEx d IIB : C_ASS I, ZONE 2; AEx nA IIB
WARNING — OPEN CIRCUIT BEFORE REMOVING COVER OR KEEP COVER TIGHT WHILE
CIRCUITS ARE ALIVE.
WARNING EXPLOSION HAZARD — SUBSTITUTION OF COMPONENTS MAY IMPAIR
SUITABILITY FOR CLASS I, DIVISION 2.
WARNING — EXPLOSION HAZARD — DO NOT DISCONNECT EQUIPMENT UNLESS POWER HAS
BEEN SWITCHED OFF OR THE AREA IS KNOWN TO BE NON—HAZARDOUS.
SENSORS WITH COATING OR SLEEVE OPTION:
WARNING (WRN) — POTENTIAL ELECTROSTATIC (ES) CHARGING HAZARD —CLEAN WITH
DAMP CLOTH or AVOID STATIC DISCHARGE BY WIPING PLASTIC WITH. DAMP CLOTH. .
- INSTALL IN ACCORDANCE WITH SECTION 18 OF THE CEC OR ARTICLE 500 OF THE NEC.
- CSA LISTED OR NRTL/UL LISTED CONDUIT SEAL AT LOCATION A & B.
NON-INCENDIVE WIRING FOR INSTALLATION IN CLASS I, DIVISION 2, GROUPS C AND D, MAX. TEMP. 85°C
REVISIONS
ZONE| REV| DESCRIPTION| CHANGE ORDER| DATE|
APPROVED
–| D| INCLUDE MPX-G.| CO- 2406| 4/21/16| K. RED
–
–| E| ADD MPX-E & -R TO ES WRN.| CO-3436| 3/6/19| R. OBORN
–
Automation Products Group, Inc.
1025 W 1700 N Logan, UT 84321
www.apgsensors.com
Phone: 888-525-7300
E-mail: sales@apgsensors.com