TROTEC TA400 Dynamic Pressure Anemometer Instruction Manual
- June 3, 2024
- TROTEC
Table of Contents
TROTEC TA400 Dynamic Pressure Anemometer Instruction Manual
Notes regarding the operating manual
Symbols
Warning of electrical voltage
This symbol indicates dangers to the life and health of persons due to
electrical voltage.
Warning
This signal word indicates a hazard with an average risk level which, if not
avoided, can result in serious injury or death.
Caution
This signal word indicates a hazard with a low risk level which, if not
avoided, can result in minor or moderate injury.
Note
This signal word indicates important information (e.g. material damage), but
does not indicate hazards.
Info
Information marked with this symbol helps you to carry out your tasks quickly
and safely.
Follow the manual
Information marked with this symbol indicates that the operating manual must
be observed.
Safety
Read this manual carefully before starting or using the device. Always store the manual in the immediate vicinity of the device or its site of use.
Warning
Read all safety warnings and all instructions. Failure to follow the warnings
and instructions may result in electric shock, fire and/or serious injury.
Save all warnings and instructions for future reference.
- Do not use the device in potentially explosive rooms or areas and do not install it there.
- Do not use the device in aggressive atmosphere.
- Do not immerse the device in water. Do not allow liquids to penetrate into the device.
- The device may only be used in dry surroundings and must not be used in the rain or at a relative humidity exceeding the operating conditions.
- Protect the device from permanent direct sunlight.
- Do not expose the device to strong vibrations.
- Do not remove any safety signs, stickers or labels from the device. Keep all safety signs, stickers and labels in legible condition.
- Do not open the device.
- Never charge batteries that cannot be recharged.
- Different types of batteries and new and used batteries must not be used together.
- Insert the batteries into the battery compartment according to the correct polarity.
- Remove discharged batteries from the device. Batteries contain materials hazardous to the environment. Dispose of the batteries according to the national regulations.
- Remove the batteries from the device if you will not be using the device for a longer period of time.
- Never short-circuit the supply terminal in the battery compartment!
- Do not swallow batteries! If a battery is swallowed, it can cause severe internal burns within 2 hours! These burns can lead to death!
- If you think batteries might have been swallowed or otherwise entered the body, seek medical attention immediately!
- Keep new and used batteries and an open battery compartment away from children.
- Observe the storage and operating conditions (see Technical data).
Intended use
Only use the device for indoor measurements of air pressure, velocity, volume
flow and temperature within the measuring range specified in the technical
data. Observe and comply with the technical data.
To use the device for its intended use, only use accessories and spare parts which have been approved by Trotec.
Foreseeable misuse
Do not use the device in potentially explosive atmospheres, for measurements
in liquids or at live parts. Any unauthorised changes, modifications or
alterations to the device are forbidden.
Personnel qualifications
People who use this device must:
- have read and understood the operating manual, especially the Safety chapter.
Residual risks
Warning of electrical voltage
There is a risk of a short-circuit due to liquids penetrating the housing!
Do not immerse the device and the accessories in water. Make sure that no
water or other liquids can enter the housing.
Warning of electrical voltage
Work on the electrical components must only be carried out by an authorised
specialist company!
Warning
Risk of suffocation!
Do not leave the packaging lying around. Children may use it as a dangerous
toy.
Warning
The device is not a toy and does not belong in the hands of children.
Warning
Dangers can occur at the device when it is used by untrained people in an
unprofessional or improper way! Observe the personnel qualifications!
Caution
Keep a sufficient distance from heat sources.
Note
To prevent damages to the device, do not expose it to extreme temperatures,
extreme humidity or moisture.
Note
Do not use abrasive cleaners or solvents to clean the device.
Information about the device
Device description
The device can measure the following parameters:
Air pressure
- psi
- mbar
- inH2O
- mmH2O
- Pa
Air velocity
- metres per second (m/s)
- feet per minute (ft/min)
- kilometres per hour (km/h)
- miles per hour (mph)
- nautical miles per hour in knots (kn)
Air volume flow
- CFM (cubic feet per minute)
- CMM (cubic metres per minute)
Air temperature
- degrees Celsius
- degrees Fahrenheit
The device is equipped with a HOLD function as well as with a Max/Min value display.
Optionally, it also possible to read and save measurement data directly on a PC by means of the software included in the scope of delivery.
Device depiction
No. | Designation |
---|---|
1 | Temperature sensor |
2 | Tripod thread |
3 | USB port |
4 | Battery compartment cover |
5 | Display |
6 | AVG/REC button |
7 | Unit button |
8 | SAVE/CLEAR button |
9 | HOLD/ZERO button |
10 | Power button |
11 | P/V/F button |
12 | MAX/MIN button |
13 | Unit button |
14 | Setup /illumination button |
15 | – input |
16 | + input |
No. | Designation |
--- | --- |
17 | Temperature/time display |
18 | °C/°F indication for display 17 |
19 | Indication of the units CFM/CMM |
20 | Measurement value indication |
21 | °C/°F indication for display 20 |
22 | REC, MAX, MIN, AVG indication |
23 | MEM indication (memory) |
24 | Memory space indication |
25 | Clear memory indication |
26 | Setup active indication |
27 | Length unit indication |
28 | HOLD function active indication |
29 | Height (H) / width (W) / diameter (D) indication |
30 | Air duct shape indication (round / angular) |
31 | Air duct cross-section selection mode active indication |
32 | Battery status indication |
33 | Automatic switch-off active indication |
34 | Indication of the pressure unit |
35 | Indication of the velocity unit |
36 | Measuring mode indication |
37 | TEMP indication |
38 | Time format indication |
Technical data
Parameter | Value |
---|---|
Model | TA400 |
Dimensions of the device
(height x width x depth)
| 210 x 75 x 50 mm
Weight of the device incl. Pitot tube and battery| 540 g
Length of the Pitot tube| 335 mm
Diameter of the Pitot tube| 8 mm
Hose lengths| 850 mm each
Operating conditions| 0 °C to +50 °C, < 90 % RH
Storage conditions| 0 °C to +50 °C, < 90 % RH
Power supply| 1 x 9 V battery
Air pressure
Accuracy| ± 0.3 % at +25 °C
Pressure range| 0 to 5000 Pa
Pressure, max.| 5000 Pa
Measuring range| PSI: 0.7252
mbar: 50.00
inH2O: 20.07
mmH2O: 509.8
Pa: 5000
Resolution| PSI: 0.0001
mbar: 0.01
inH2O: 0.01
mmH2O: 0.1
Pa: 1
Air velocity
Measuring range| m/s: 1 to 80.00
ft/min: 200 to 15733
km/h: 3.6 to 288.0
MPH: 2.24 to 178.66
knots: 2.0 to 154.6
Resolution| m/s: 0.01
ft/min: 1
km/h: 0.1
MPH: 0.01
knots: 0.1
Accuracy| m/s: ±2.5 % at 10 m/s ft/min, km/h, MPH, knots: The accuracy depends
on the air velocity and the size of the air duct.
Parameter| Value
---|---
Air volume flow
Measuring range| CFM: 0 ft3/min to 99,999 ft3/min
CMM: 0 m3/min to 99,999 m3/min
Resolution| CFM: 0.0001 to 100
CMM: 0.001 to 100
Temperature
Measuring range| °C: 0 °C to 50 °C
°F: 32.0 °F to 122.0 °F
Resolution| °C: 0.1
°F: 0.1
Accuracy| °C: ± 1.0 °C
°F: ± 2.0 °F
Scope of delivery
- 1 x Device TA400
- 1 x 9 V battery
- 1 x Pitot tube
- 1 x Tube, white
- 1 x Tube, black
- 1 x Transport case
- 1 x Mini USB cable
- 1 x CD-ROM with software
- 1 x Quick guide
Transport and storage
Note
If you store or transport the device improperly, the device may be damaged.
Note the information regarding transport and storage of the device.
Transport
For transporting the device, use the transport case included in the scope of
delivery in order to protect the device from external influences.
Before transporting the device, please observe the following:
- Remove the hoses from the connections at device and Pitot tube.
Storage
When the device is not being used, observe the following storage conditions:
- dry and protected from frost and heat
- protected from dust and direct sunlight
- For storing the device, use the transport case included in the scope of delivery in order to protect the device from external influences.
- the storage temperature complies with the values specified in the Technical data
- Remove the battery from the device.
Operation
Inserting the battery
Note
Make sure that the surface of the device is dry and the device is switched
off.
- Open the battery compartment at the rear of the device by sliding the battery compartment cover (4) down at the arrow mark.
- Use the battery clip to connect the 9 V battery with correct polarity.
- Place the battery with the battery clip into the battery compartment.
- Slide the battery compartment cover (4) back on the battery compartment.
The cover should click into place.
Switching the device on
Info
Please note that moving from a cold area to a warm area can lead to
condensation forming on the device’s circuit board. This physical and
unavoidable effect can falsify the measurement. In this case, the display
shows either no measured values or they are incorrect. Wait a few minutes
until the device has become adjusted to the changed conditions before carrying
out a measurement.
- Press the Power button (10).
The device is switched on.
Measuring the differential pressure
In the measuring mode for differential pressure measurements the differential pressure of a zone 1 can be determined with regard to a reference environment (zone 2 / equipment location).
The differential pressure can be indicated in 5 different units:
-
PSI
-
mbar
-
inH2O
-
mmH2O
-
Pa
-
Connect the white tube to the + input (16).
The – input (15) will not be connected to a tube.
-
Press the P/V/F button (11) until PRESS appears in the measuring mode display (36).
-
Press the Unit __ button (13) to select the desired unit for the measurement.
The selected unit appears in the pressure unit display (34). -
Press the HOLD/ZERO button (9) for approx. 2 s to reset the saved measured values to zero.
-
Position the free end of the tube in the area (zone 1) the differential pressure of which is to be determined with regard to the measuring device (zone 2).
The measured differential pressure value will be indicated in the measurement value indication (20).
A positive measured value indicates that the pressure in zone 1 is higher than in zone 2.
A negative measured value indicates that the pressure in zone 1 is lower than in zone 2.
The measured value 0 indicates an identical pressure in both zones.
-
Note:
Additionally, you can connect the black tube to the – input (15). Bear in mind
that in that case the reference environment zone 2 is equivalent to the end of
the black tube, not the device location.
Measuring the air velocity
In the measuring mode for air velocity measurements the current air velocity
is measured with defined standard conditions (temperature 21.1 °C / 70 °F, air
pressure 14.7 psi / 1013 mbar).
The air velocity can be indicated in 5 different units:
-
metres per second (m/s)
-
feet per minute (ft/min)
-
kilometres per hour (km/h)
-
miles per hour (mph)
-
nautical miles per hour in knots (kn)
-
Connect the white tube to the + input (16) of both the device and the Pitot tube.
-
Connect the black tube to the – input (15) of both the device and the Pitot tube.
-
Press the HOLD/ZERO button (9) for approx. 2 s to reset the measured values to zero.
-
Press the P/V/F button (11) until VEL appears in the measuring mode display (36).
-
Press the Unit __ button (13) to select the desired unit for the measurement.
ð The selected unit appears in the velocity unit display (35). -
Point the upper end of the Pitot tube towards the air flow. In doing so, make sure that the Pitot tube is not inclined more than 10° with regard to the air current.
The measured value will be indicated in the measurement value indication (20).
If a negative measured value or the message Error is displayed, please check the connections at Pitot tube and device for proper fit and correct polarity.
-
Measuring the air volume flow
In the measuring mode for air volume flow measurements the air volume flow is measured with defined standard conditions (temperature 21.1 °C / 70 °F, air pressure 14.7 psi / 1013 mbar).
To achieve a measurement that is as accurate as possible, you can indicate round and angular current cross-sections incl. the precise cross-sectional areas.
The air volume flow can be indicated in 2 different units:
-
CFM (cubic feet per minute)
-
CMM (cubic metres per minute)
-
Connect the white tube to the + input (16) of both the device and the Pitot tube.
-
Connect the black tube to the – input (15) of both the device and the Pitot tube.
-
Press the P/V/F button (11) until FLOW appears in the measuring mode display (36).
-
Press the Unit __ button (13) to select the desired unit for the measurement.
The selected unit appears in the CFM/CMM units display (19). -
Point the upper end of the Pitot tube towards the air flow. In doing so, make sure that the Pitot tube is not inclined more than 10° with regard to the air current.
The measured value will be indicated in the measurement value indication (20).
-
Adjust air duct shape and dimensions
If you want to change details regarding the shape and dimensions of the air
duct, please proceed as follows:
-
Press the Setup/illumination button (14) for approx. 2 s.
The setup menu will be opened. -
Repeatedly press the Unit (7) or Unit __button (13) until the indication Duct Shape (31) appears.
-
Press the AVG/REC button (6) to confirm the selection.
The submenu for the air duct shape will be opened. -
Press the Unit (7) or Unit __ button (13) to choose the quadrangular or round air duct shape.
-
Press the AVG/REC button (6) to confirm the selection.
The air duct shape is selected.
If you selected a round air duct, you can now specify the diameter (D=) in cm:
D= is displayed on the indication (29).
-
Press the Unit (7) or Unit __(13) button to change the position of the decimal point.
-
Press the SAVE/CLEAR button (8) to select a digit.
The currently selected digit flashes. -
Press the Unit (7) or Unit __button (13) to adjust the value (ranging between 0 and 9).
-
Repeat these steps until the air duct diameter is indicated correctly.
-
Press the AVG/REC button (6) to save the setting.
The menu item Type is displayed. -
Press the Setup button (14) for approx. 2 s to exit the settings menu.
If you selected a quadrangular air duct, you can now specify its width (W=)
and height (H=) in cm:
W= is displayed on the indication (29)
-
Press the Unit (7) or Unit __(13) button to change the position of the decimal point.
-
Press the SAVE/CLEAR button (8) to select a digit.
The currently selected digit flashes. -
Press the Unit (7) or Unit __button (13) to adjust the value (ranging between 0 and 9).
-
Repeat these steps until the width of the air duct is indicated correctly.
-
Press the AVG/REC button (6) to save the setting.
H= is displayed on the indication (29). -
Repeat the steps for setting the width until the height of the air duct is also indicated correctly.
-
Press the AVG/REC button (6) to save the setting.
The menu item Type is displayed. -
Press the Setup button (14) for approx. 2 s to exit the settings menu.
Measuring the air temperature
The air temperature will be indicated in the temperature/time display (17).
In the measuring mode you can always switch between the units °C and °F:
- Press the Unit button (7).
The displayed unit will change from °C to °F or vice versa.
An acoustic confirmation signal will be emitted.
Displaying MIN / MAX / AVG values
The minimum (MIN), maximum (MAX) and average (AVG) values can be determined via a measurement interval.
-
Press the MAX/MIN button (12) until the desired function is indicated on the display (22).
The temperature/time display (17) will be switched from temperature to time (min:sec).
A new measurement interval will be started.
REC is displayed on the indication (22). -
Press the MAX/MIN button (12) to switch between the functions.
-
Press the MAX/MIN button (12) for approx. 2 s to return to the normal measuring mode.
Hold function
The currently measured value can be frozen in all measuring modes.
- Press the HOLD/ZERO button (9) to freeze the currently measured value.
- Press the HOLD/ZERO button (9) again to deactivate the hold function.
Saving and calling up measured values
In each measuring mode you can save up to 99 data sets.
To save a measured value, please proceed as follows:
- Press the SAVE/CLEAR button (8).
The current measured value will be saved.
An acoustic confirmation signal will be emitted.
The number of saved values below the MEM indication (23) is increased by one.
To call up a measured value, please proceed as follows:
-
Press the AVG/REC button (6) for approx. 2 s.
REC will be displayed in the temperature/time indication (17). -
Press the Unit (7) or Unit (13) button to choose the desired memory location.
The saved measured value is displayed in the measurement value indication (20). -
Press the AVG/REC button (6) for approx. 2 s to return to the measuring mode.
Settings
In the setting menu you can adjust some of the basic settings for the device.
The following table shows an overview of the available options.
Menu text | Designation | Setting options |
---|---|---|
Unit | Unit for the air duct dimensions | for determining the unit in which the |
air duct dimensions will be indicated
Duct Shape| Settings for the air duct| for specifying the shape and dimensions
of the air duct
Type| Display options for measuring modes| 1 = measuring modes for air
pressure and air velocity
2 = measuring modes for air velocity and air volume flow
3 = measuring modes for air pressure, air velocity and air volume flow
Sleep| Automatic switch- off| for de-/activating the automatic switch-off
function
ALL| Clear memory| for clearing the memory; yes or no
To access the setup menu, please proceed as follows:
The device is switched on.
-
Press the Setup button (14) for approx. 2 s.
The setup menu will be opened. -
Select the desired option using the Unit (7) or Unit __button (13).
-
Press the AVG/REC button (6) to confirm the selection.
The desired submenu will be opened.
Setting the measuring unit for the air duct
-
Press the Setup button (14) for approx. 2 s.
The setup menu will be opened. -
Repeatedly press the Unit (7) or Unit __button (13) until unit is displayed.
-
Press the AVG/REC button (6) to confirm the selection.
The length indication (27) displays the currently selected unit (in or cm). -
Repeatedly press the Unit (7) or Unit __button (13) until the desired unit is displayed.
-
Press the AVG/REC button (6) to save the setting.
-
Press the Setup button (14) for approx. 2 s to exit the settings menu.
Deleting saved measurements
There are two ways to delete measured values:
- Deleting all saved measurements
- Deleting a certain measured value
To delete all saved entries, please proceed as follows:
-
Press the Setup button (14) for approx. 2 s.
The setup menu will be opened. -
Repeatedly press the Unit (7) or Unit __button (13) until ALL is displayed.
-
At the same time the clear memory indication (25) appears in the bottom right corner.
-
Press the AVG/REC button (6) to confirm the selection.
The measuring modes are indicated in the measuring mode display (36).
YES will be indicated in the temperature/time display (17). -
Press the Unit (7) or Unit __button (13) to choose whether to delete the memory (yes) or not (no).
-
Press the AVG/REC button (6) to confirm the selection.
-
Press the Setup button (14) for approx. 2 s to exit the settings menu.
To delete one specific measured value, please proceed as follows:
-
Press the AVG/REC button (6) for approx. 2 s.
ð REC will be displayed in the temperature/time indication (17). -
Press the Unit (7) or Unit (13) button to choose the desired memory location.
The saved measured value is displayed in the measurement value indication (20). -
Press the SAVE/CLEAR button (8).
The selected measurement will be deleted.
An acoustic confirmation signal will be emitted.
The next saved measurement will be displayed. -
Press the AVG/REC button (6) for approx. 2 s to return to the measuring mode.
Setting the automatic switch-off
If automatic switch-off is activated, the device switches off automatically after approx. 5 minutes of non-use.
-
Press the Setup button (14) for approx. 2 s.
The setup menu will be opened. -
Repeatedly press the Unit (7) or Unit __button (13) until SLEEP is displayed.
-
Press the AVG/REC button (6) to confirm the selection.
On or off (activated or deactivated automatic switch-off) will be indicated in the temperature/time display (17). -
Press the Unit (7) or Unit __button (13) to choose the desired setting.
-
Press the AVG/REC button (6) to confirm the selection.
-
Press the Setup button (14) for approx. 2 s to exit the settings menu.
Setting the background illumination
The display comes with a background illumination that can be switched on as
needed.
- Press the Setup / illumination button (14) to switch the
background illumination on or off.
Switching the device off
- Press the Power button (10).
The device is switched off
Software
The supplied free software is designed for useful basic functionalities. Trotec assumes no liability with regard to this free software and also provides no support on that score. Trotec accepts no liability concerning the use of this free software and is under no obligation to make adjustments or to further develop updates or upgrades.
Installation requirements
Ensure that the following minimum requirements for installing the PC software
are fulfilled:
- Supported operating systems (32 or 64 bit version):
- Windows 10
- Windows 8
- Windows 7
- Windows Vista
- Windows XP
- Hardware requirements:
- processor speed: min. 90 MHz
- 32 MB RAM, minimum
- 7 MB hard disk space, minimum
- a minimum screen resolution of 1024 x 768 with a 16 bit colour depth
Installing the PC software
Administrator rights are required for the software installation.
- Place the supplied software CD into the CD-ROM drive.
- Open the CD-ROM contents in an Explorer window and double-click the setup.exe file.
- Follow the instructions of the installation wizard.
Starting the PC software
- Connect the measuring device to your PC via the mini USB cable provided in the scope of delivery.
- Switch on the measuring device if necessary.
- Start the Manometer&Flowmeter software.
Information about using the PC software is provided in the online help.
Errors and faults
The device has been checked for proper functioning several times during production. If malfunctions occur nonetheless, check the device according to the following list.
Indication | Cause | Remedy |
---|---|---|
OL | air pressure or air velocity above the measuring range |
- Check the battery voltage and insert a new high quality battery for testing purposes.
- Choose a different location for measuring.
If the message continues to be displayed, carry out a reference measurement at a known location:
- Choose a site within the measuring range for this measurement.
- Press the HOLD/ZERO button (9) for approx. 2 s to reset the saved measured values to zero.
- Read the measured value from the measurement value display (20).
If the error code is still displayed, the device may be defective.
Please contact the Trotec customer service.
-OL| air pressure below the measuring range
Error| air velocity or air volume flow below the measuring range
Maintenance and repair
Battery change
A battery change is required when the battery status indication (32) lights up
or the device can no longer be switched on (see chapter Inserting the
battery).
Cleaning
Clean the device with a soft, damp and lint-free cloth. Make sure that no
moisture enters the housing. Do not use any sprays, solvents, alcohol-based
cleaning agents or abrasive cleaners, but only clean water to moisten the
cloth.
Repair
Do not modify the device or install any spare parts. For repairs or device
testing, contact the manufacturer.
Disposal
Always dispose of packing materials in an environmentally friendly manner and in accordance with the applicable local disposal regulations.
The icon with the crossed-out waste bin on waste electrical or electronic equipment stipulates that this equipment must not be disposed of with the household waste at the end of its life. You will find collection points for free return of waste electrical and electronic equipment in your vicinity. The addresses can be obtained from your municipality or local administration. You can also find out about other return options that apply for many EU countries on the website https:// hub.trotec.com/?id=45090. Otherwise, please contact an official recycling centre for electronic and electrical equipment authorised for your country.
The separate collection of waste electrical and electronic equipment aims to enable the re-use, recycling and other forms of recovery of waste equipment as well as to prevent negative effects for the environment and human health caused by the disposal of hazardous substances potentially contained in the equipment.
In the European Union, batteries and accumulators must not be treated as domestic waste, but must be disposed of professionally in accordance with Directive 2006/66/EC of the European Parliament and of the Council of 6 September 2006 on batteries and accumulators. Please dispose of batteries and accumulators according to the relevant legal requirements.
Only for United Kingdom
According to Waste Electrical and Electronic Equipment Regulations 2013 (SI
2013/3113) (as amended) and the Waste Batteries and Accumulators Regulations
2009 (SI 2009/890) (as amended), devices that are no longer usable must be
collected separately and disposed of in an environmentally friendly manner.
Maintenance and repair
Battery change
A battery change is required when the battery status indication (32) lights up
or the device can no longer be switched on (see chapter Inserting the
battery).
You can download the current version of the operating manual and the EU declaration of conformity via the following link:
TA400
https://hub.trotec.com/?id=43622
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>