AKTiiA G1 Blood Pressure Monitoring Bracelet User Manual

June 3, 2024
AKTiiA

logo

AKTiiA G1 Blood Pressure Monitoring Bracelet

AKTiiA G1-Blood0Pressure-Monitoring-Bracelet-fig-
\(2\)

Introduction

Welcome to the Aktiia 24/7 Blood Pressure Monitor System! The Aktiia System is the smart way to track your Blood Pressure over days and nights without any effort. The Aktiia Bracelet is designed to be discreet, with no lights or alarms bothering you during your daily life. With its accuracy and ease of use, Aktiia is the perfect solution to track your blood pressure 24 / 7.
Aktiia offers a new way to monitor your heart health that easily integrates into your life. Our device tracks your Blood Pressure through unnoticeable measurements performed multiple times per day. All you need to do is to wear the Bracelet. You can access your data anytime by simply looking at the Aktiia mobile app. Speaking of the app, make sure you download the Aktiia App from the Apple app store. The app has all the information you need to get started with your Aktiia System.

Contact Information

Please read this User Manual carefully to gain a complete understanding of the device’s functions and safety-related information. In case you have any additional questions, you encounter any issue, or you would like to suggest some improvements, please contact Aktiia’s Customer Service at support@aktiia.com or visit our website at www.aktiia.com.
For an electronic version of this user manual and for common symbols, please visit www.aktiia.com/uk/manuals/electronicsymbols

Intended purpose

The Aktiia Bracelet is a non-invasive blood pressure (BP) monitor intended to measure optical Photoplethysmography (PPG) signals on the user’s wrist and to calculate blood pressure values using a Pulse Wave Analysis (PWA) technique, following an initialization process using an oscillometric blood pressure monitor. The Aktiia Bracelet can also calculate estimated pulse rate based on the same measurement and analysis technology.

Indications for use

The Aktiia 24/7 Blood Pressure Monitor System is a non-invasive monitor comprising of an oscillometric cuff and an optical bracelet for both on-demand & continual measurements of blood-pressure and estimated pulse rate.
The system is intended for OTC use for adult patients (ages 21-60) in an at- home environment.

Intended Patient Population

The Aktiia 24/7 Blood Pressure Monitor System is appropriate for users that are 21 to 60 years old, with an arm circumference of 22cm to 42cm (8 ¾ inches to 16 ½ inches) and a wrist circumference of 14cm to 21 (5 ½ inches to 8 ¼ inches).

Contraindications

Patients with the following conditions are NOT part of the intended patient population:

  • Pregnant women
  • Patients with the following heart conditions:
    • Patients with Tachycardia and resting pulse rate greater than 120 beats per minute
    • Congestive heart failure with reduced heart function
    • Atrial fibrillation
    • Severe valvular disease
    • Implanted devices such as a pacemaker or defibrillator
  • Patients with the following medical conditions:
    • Renal dysfunction
    • Raynaud’s disease
    • Actively undergoing treatment for the following endocrine disorders:
  • Diabetes
  • Pheochromocytoma
  • Hyperthyroidism
    • Arteriovenous fistula
    • Lymphoedema of the arm
    • Presence of an intravascular device
  • Other conditions:
    • Any condition that might interfere with readings, such as trembling and shivering
    • Exfoliative skin diseases involving the wrist, such as eczema or psoriasis
    • Arm amputation(s)

Important safety information

Please read the important safety information in this user manual before using the device. Any serious incident occurring in relation to the Aktiia Bracelet should be reported to Aktiia.

Warnings

  • The “WARNING” sign in this user manual indicates a potentially hazardous situation which, if not avoided, could result in serious injury or death.
  • Your Aktiia Bracelet Pod is small enough to be swallowed by a young child. To prevent risk of suffocation, never leave your Pod unattended!
  • The Aktiia Bracelet is not intended to diagnose any medical condition. The Akttia Bracelet is a monitoring device and will not alarm you based on your blood pressure readings or pulse rate physiological alarm conditions. Consult your physician for an appropriate analysis of your blood pressure data for diagnosis.
  • No modification of the equipment is allowed.
  • The Aktiia System does not display information that would indicate a cardiovascular event, such as a heart attack, stroke, or arrhythmia. If you believe you are having a medical emergency, call emergency services immediately

The Aktiia System is designed for assessment and monitoring of blood pressure, and it is not intended to diagnose any medical condition. The Akttia Bracelet is a monitoring device and will not alarm you based on your blood pressure readings or pulse rate. Consult your physician for an appropriate analysis of your blood pressure data for diagnosis and management, and do not change medications without discussing it first with your medical professional.

Cautions

  • The “Caution” sign in this user manual indicates a potentially hazardous situation which, if not avoided,could result in minor injury to the user or patient or damage to the equipment or other property.
  • This device may only be used for the purposes described in this User Manual. Aktiia cannot be held liable for damage or injury caused by incorrect use. Always follow the operating procedures described in this User Manual to measure your blood pressure accurately and safely.
  • Aktiia Bracelet is designed as a device for personal use (single user) only.
  • Do not share your device with others as it may result in inaccurate blood pressure readings.
  • The strap of your Aktiia Bracelet contains silicone material. Avoid wearing the Bracelet in case of known allergy to silicone. In case of skin irritation, stop wearing your
  • Aktiia Bracelet immediately and consult your doctor.
  • Use of this device adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this device and the other equipment should be observed to verify that they are operating normally.
  • Wireless communication devices (including home networking devices & peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of this device. Otherwise, degradation of the performance of this product could result.
  • Use of this device with active implantable medical devices, such as cardiac pacemakers and ICDs, has not been evaluated and is therefore not recommended
  • Avoid exposure of this device to known sources of EMI (electromagnetic interference) such as magnetic resonance imaging (MRI) systems, diathermy, lithotripsy, electrocautery, RFID (radio frequency identification), and electromagnetic security systems such as metal detectors. Note that the presence of RFID devices may not be obvious.
  • If such interference is suspected reorient equipment if possible to maximize distances.
  • The performance of the Bracelet can be affected by extremes of temperature, humidity and altitude.
  • Do not use the device with other medical equipment simultaneously
  • If, at any time, the device loses Bluetooth connectivity with the phone, the user must manually reestablish connectivity with the device through the Aktiia App
  • The Aktiia System is intended to be used by people who have general familiarity and experience with mobile device platforms

Technological characteristics

Aktiia Bracelet uses PPG technology (Photoplethysmography) to acquire optical data on your wrist. The PPG data are then transferred through the Aktiia App to a secured cloud server on which Aktiia’s algorithms estimate your blood pressure through advanced processing of the optical data. Aktiia Bracelet detects if the user is moving before initiating a PPG measurement and will provide blood pressure measurement only when the user is still.
This product is licensed under patents owned by CSEM SA, Switzerland.

Package contents

Your Aktiia Bracelet is supplied in a box containing the following items:AKTiiA G1-Blood0Pressure-Monitoring-Bracelet-fig-
\(3\)

Latest version of the Aktiia Bracelet user manual can also be found online at: https://aktiia.com/uk/manuals/

Operating requirements

To use the Aktiia System, you should have general familiarity and experience using mobile device platforms.
To use the Aktiia Bracelet you need to download the free Aktiia App from the Apple App store. Creation of a user account and login are required to operate the Aktiia Bracelet.

Access to the Internet is required to:

  • Download the Aktiia App.
  • Set up your Aktiia user account.
  • Update Aktiia Monitor’s firmware.
  • Visualize your new Blood Pressure data.
  • Receive insights on your Blood Pressure readings.

Authorized accessories and compatible devices

Your Aktiia Bracelet requires initialization and periodic re-initialization every 7 days. The Aktiia Bracelet must be used with the complementary Aktiia Cuff. The Aktiia Cuff is an oscillometric device provided by Aktiia in a separate package.
The Aktiia mobile application requires a mobile device with iOS 11 or later, Bluetooth 4.2 or higher and Wi-fi or 3G / 4G / 5G connection.

Quality of Service of Aktiia Bracelet G1: In regard to Quality of Service, the Aktiia Bracelet connects to the app using Bluetooth Low Energy (BLE). The BLE radio of the bracelet is always on and discoverable. Once the device has been paired to the app, no specific action is required for an authenticated user to connect to it. The Aktiia System is primarily intended for retrospective review of blood pressure and pulse rate data, there are no risks associated with data latency and throughput. The device is not being used as diagnostic tool and there are no alarms

Set up and initialization procedures

  1. Charge Bracelet before First Use.

    1. To charge the Bracelet:
  2. Connect the charging station to a USB port using the charging cable provided

  3. Remove the thin plastic protective film that is attached to the bottom of the pod of the Bracelet before charging

  4. Place the Aktiia Bracelet, with the sensor on the bottom, onto the charging station.

  5. A red light at the bottom of the charger will begin glowing to indicate that the bracelet has been placed correctly on the charging station.

  6. When charging is completed, the red light is steady.

  7. Download the Aktiia App.
    Scan the QR code on this page or go to the Apple App Store, then download and install the Aktiia App.

  8. Create a user account on the Aktiia App.
    Open the Aktiia App on your mobile device and follow the instructions to register and set up your personal account.

  9. Pair your Aktiia Bracelet with your mobile device.
    Place the Aktiia Bracelet on the charging station and enable Bluetooth connection on your mobile device. Wait until pairing is confirmed by the Aktiia App.

  10. Pair your Aktiia Cuff with your mobile device.
    Switch ON the Aktiia Cuff by turning on the toggle button. A blue light should start blinking. Wait until pairing is confirmed by the Aktiia App andthe light indicator is a steady blue. For further information on the Aktiia Cuff operation, please refer to the user manual

  11. Fit your Aktiia Bracelet on your wrist.
    For optimal measurement by the Aktiia Bracelet, please carefully read the instructions in the “Correct Aktiia Bracelet positioning” section of this user manual.

  12. Fit your Aktiia Cuff on the opposite arm.
    The Aktiia Cuff must be placed on the opposite arm than the one wearing the Bracelet (i.e if you wear your Aktiia Bracelet on your RIGHT wrist, place the Cuff over your LEFT arm and vice versa).
    The accuracy of blood pressure measurement done by the Aktiia Cuff depends on the right Cuff positioning. Please carefully read the instructions in the section “Correct Aktiia Cuff positioning” of this user manual.

  13.  Initialize your Aktiia Bracelet.
    On the Aktiia App, carefully follow the instructions on the screen to start the initialization procedure. The Cuff will start inflating.

  14. Switch OFF your Aktiia Cuff.
    The accuracy of the initialization process depends on your body posture during the initialization. Carefully read the section “Body Posture during Initialization Procedure” of this user manual.

Congratulations! Your Aktiia Bracelet is now initialized and ready to track your blood pressure!

The Aktiia Bracelet is not intended to diagnose any medical condition. The Akttia Bracelet is a monitoring device and will not alarm you based on your blood pressure readings or pulse rate physiological alarm conditions. Consult your physician for an appropriate analysis of your blood pressure data for diagnosis.
You need to complete an initialization procedure at least once every 7 days, when prompted by the mobile application. The notification to reinitialize will appear beginning on Day 5. If you do not reinitialize by Day 7, you will no longer see new data on your App. Data obtained outside the period may be inaccurate.

Correct Aktiia Bracelet positioning

Fasten your Aktiia Bracelet around your wrist with the sensor (Aktiia Pod) facing toward or placed against the skin on top of your wrist. The clasp will be on the underside of the wrist. The bracelet should be about 1 inch from your most prominent wrist bone. This ensures that the Bracelet does not move around as much whenyou are using your hands. The Bracelet should be snug but comfortable

Correct Aktiia Cuff positioning

The Aktiia Cuff must be placed on the opposite arm with respect to the one wearing the Bracelet (i.e if you wear your Aktiia Bracelet on your RIGHT wrist. Place the Cuff over your LEFT arm and vice versa).AKTiiA G1
-Blood0Pressure-Monitoring-Bracelet-fig- \(6\)

  1. Remove garments from your upper arm. If you roll up your sleeve, please ensure that the garment is not too tight so that it does not cause any blood flow constriction.
  2. Place your bare arm through the Cuff and position the Cuff 1″ (2-3 cm) above your elbow joint. The middle of the Cuff should be placed as the same level as your heart.
  3. Tighten the Cuff around your arm so that it fits closely but you can still insert two fingers between your arm and the Cuff. Secure the Cuff closed with the Velcro fastener.
  4. While seated, place your hand, palm side-up in front of you so that it is supported by a flat surface and the Aktiia Cuff is at the same height as your heart. Your Aktiia Cuff monitor is positioned on the inner side of your arm, over the artery and the logo is towards your elbow joint.
    Please check the instructions for use provided with the Aktiia Cuff initialization device

Body posture during the initialization procedure

Please sit down and relax for 5 minutes before starting the initialization procedure.

Note : Blood Pressure measurements can be affected by the position of the Cuff and your physiological and emotional condition.

  1. Sit upright with your back straight, keep your legs uncrossed, and your feet flat on the floor.
  2. Place both your hands palm-side down in front of you on a flat surface such as a desk or a table up to and including your elbow.
  3. Do not move or tense your arm muscles during measurement.
  4. Relax, and do not talk.

Accessing your data

Your Aktiia Bracelet automatically tracks your blood pressure, periodically measuring throughout the day and night. To ensure optimal data security, your data is stored on a secured cloud server, and not the Aktiia Bracelet or on your phone. Visualizing your data in the Aktiia App therefore requires internet connection to synchronize with Aktiia’s secure server.

Syncing your data

  • Make sure Bluetooth and Internet connection on your mobile device are enabled.
    Open the Aktiia App to synchronize data from your Aktiia Bracelet. The synchronization is done automatically and a sync view at the top of the screen shows you the sync status. This might take few seconds to several minutes depending on how often you open the app and synchronize your data.

  • In case you do not see a sync view when opening the app, swipe down to manually activate the synchronization.

  • Make sure you open the app regularly so that the synchronization is fast, and you avoid losing data. For this reason, synching once a day is recommended.

Visualizing your Blood Pressure data

  • Press the Home button on the Aktiia App to see your blood pressure data in a chart format.
  • The top is the systolic (SYS) value and the bottom the diastolic (DIA) value. Tap on the line to read the numeric values, hold, and move your finger on the screen to navigate through the values.
  • In the Aktiia App you will be shown metrics such as the average Blood Pressure values for the day.
  • You are also able to swipe left and right to navigate through your historical data, in day, week, and month view.

How to evaluate your blood pressure

  • The Aktiia Bracelet is not intended to diagnose any medical condition. The Akttia Bracelet is a monitoring device and will not alarm you based on your blood pressure readings or pulse rate physiological alarm conditions. Consult your physician for an appropriate analysis of your blood pressure data for diagnosis.
  • The following classifications are based on measurements taken on a seated person after few minutes of rest. It is important to note that Blood Pressure readings in normal life conditions might be higher.
  • These charts are not intended to provide a basis for any type of diagnosis or emergency assessment; these charts only depict different classifications of blood pressure.
  • Consult your physician for an interpretation and diagnosis based on your personal blood pressure results. If you feel that your readings are abnormal, contact your physician.

The American Heart Association (AHA) has created the following guide for classifying blood pressure values.

BLOOD PRESSURE CATEGORY| SYSTOLIC BP

mm Hg (upper number)

|

AND/OR

| DIASTOLIC BP

mm Hg (lower number)

| COLOR INDICATOR
---|---|---|---|---
NORMAL| LESS THAN 120| AND| LESS THAN 80| Green
ELEVATED| 120-129| AND| LESS THAN 80| Yellow
HIGH BLOOD PRESSURE (HYPERTENSION) STAGE I| 130-139| OR| 80-89| Orange
HIGH BLOOD PRESSURE STAGE (HYPERTENSION) 2| 140 OR HIGHER| OR| 90 OR HIGHER| Light Red
HYPERTENSIVE CRISIS (consult your doctor immediately)| HIGHER THAN 180| AND/OR| HIGHER THAN 120| Dark Red

Note : Various factors such as age, obesity and medical condition should be considered for a correct evaluation. Consult with your physicians for an accurate assessment and diagnosis of your health condition.

Battery handling and power supply

The battery of Aktiia Bracelet is a built-in rechargeable li-polymer battery.
When the monitor is connected to a mobile device, the battery volume will be displayed in the mobile app, in the Devices section. If the battery is lower than 10%, please charge your device.

  1. Connect the USB microB connector (the smaller side) of the USB cable to the charger’s USB connector port.

  2. Connect the USB A connector (the larger side) of the USB cable to a power source.

  3. Place Aktiia Bracelet on the charger as depicted below.
    A red light on the bottom part of the charger starts glowing when the bracelet is charging.

  4. A steady red light indicates that the device is fully charged.

It is recommended that you charge the battery when the battery is less than 10%. Aktiia Bracelet charger is designed to work when plugged to a constant 5V DC source.
The battery in this device is a fixed battery and can only be changed by an authorized Aktiia Service agent.
Should your battery be defective, please contact Aktiia Customer Service for a replacement.
Lithium battery replacement by inadequately trained personnel could result in a hazard such as a fire or explosion.
Note : Battery life and charge cycles vary by use and settings.

Unpair your device

Aktiia Bracelet is designed for personal use (single user) only.
During the pairing procedure, the devices are linked to your account. In case you wish to reset the pairing and allow another person to use the device you must reset the pairing first.

To complete the unpairing procedure you should:

  1. Login to your account
  2. Tap on the device tab
  3. Tap on the button with the three dots
  4. Press unpair

Note : Unpairing is needed if Aktiia Bracelet and / or Aktiia cuff need to be linked with a new user account
Note : Unpairing is not needed if a new mobile device is used to download Aktiia data with the same user account

Troubleshooting

Aktiia Bracelet is designed to be discreet, with no lights or alarms bothering you during your daily life. To know if your Aktiia Bracelet is ON:

  1. Take off your Aktiia Bracelet.
  2. Look at the sensor on the bottom side of the pod.
  3. Gently tap the Bracelet’s pod twice – a green light should blink within 5 seconds.
    If the green light does not blink, your Aktiia Bracelet may be out of power. Please put your device on the charging station.
Problem Solution

I don’t see a red light when charging my Aktiia Bracelet

| Verify that the charging station is well connected to a power supply through the USB cable provided.

Try replacing the Pod by turning it by 180°. The charging pins on the Pod should be aligned with the ones on the charging station. The Pod should be magnetically locked on the charging station.

My Aktiia Bracelet does not connect to my mobile device

| Check that your mobile device Bluetooth is activated and that you enabled permissions for Bluetooth connection and location.

If your mobile device Bluetooth is activated but your Aktiia Bracelet still does not connect,

place your Aktiia Bracelet on the charging station to force a reset (you will not lose your personal data or account information). If the problem persists, contact Aktiia customer service.

I cannot Sync my Aktiia Bracelet with the app

| Check that your Aktiia Bracelet is connected in the Devices section. On the Home  screen scroll down to activate Sync.

I cannot Pair my Aktiia Bracelet

| Verify that your Aktiia Bracelet is correctly placed on the charger and that a red light is visible at the bottom of the charging station.

Initialization recording failed

| Verify that Bluetooth and Location services are enabled, and devices are connected. Verify that you have a stable internet connection.

Verify that the Aktiia Bracelet and Aktiia Cuff are positioned correctly.

Do not move during recording and take 5 deep breaths before starting the initialization procedure again.

Aktiia was not found

| Place your Aktiia Bracelet on the charging station and connect it to the power supply. Verify that Bluetooth is on and permissions are granted and try again.
---|---

There is no reliable measurement data for this period

| Aktiia Bracelet only measures your Blood Pressure while you are at rest. You might not get any measurements during periods of high activity. If you don’t get any measurements over more than 24 hours, and other troubleshooting steps do not resolve this, please contact our customer support.

Care and maintenance

  • The Aktiia Bracelet does not contain any part or component that requires maintenance operation by the user.
  • To avoid skin irritation, it is recommended to regularly clean your Aktiia Bracelet. Simply wipe your Aktiia Bracelet with soft cloth lightly moistened with water. Then dry your Aktiia Bracelet with a dry cloth.
  • The Aktiia Bracelet battery can maintain the performance characteristics for a maximum of 300 charge cycles, which corresponds to an expected service life of 3 years.
  • Do not attempt to disassemble The Aktiia Bracelet as this will result in permanent damages and will void your warranty. If you encounter troubles with your Aktiia Bracelet battery, please contact Aktiia customer support.
  • Do not immerse the Aktiia Bracelet in water and do not expose the Aktiia Bracelet to water moving with force, such as water running from a tap, shower, ocean waves, or waterfalls.

Warranty

Your Aktiia Bracelet is warranted to be free from defects in materials and workmanship within two years from the date of purchase when used in accordance with the provided instructions. Issues related by strap’s wearing are not covered by the warranty.
The warranty extends only to the end user. We will, at our option, repair or replace without charge Aktiia Bracelet covered by the warranty. Repair or replacement is our only responsibility and your only remedy under the warranty.

Specifications

Product description Blood pressure monitor
Model Aktiia Bracelet G1
Measurement method Photoplethysmography (PPG)
Sensor to measure PPG Green light (wavelength 526nm); silicon receiver

diode
Measurement accuracy (Blood Pressure)| ± 5mmHg
Measurement accuracy (Pulse Rate)| ± 5 pulse/min
Measurement range| 40-250 mmHg for blood pressure, 40-180 beats/min for heart rate


Classifications

| Internally powered

Applied parts: Pod (type BF).

IP33 (protected from spraying water and tools >2.5mm)



Communication module

| ****

BLE 5.0

Frequency range 2.4 to 2.4835 GHz Modulation GFSK


Operating conditions

| +5°C to 35°C

700hPa to 1060hPa15 % to 90 % RH

---|---


Storage and transportation conditions

| -20°C to +60°C

700hPa to

1060hPa10% to 95% RH

Aktiia Bracelet size| 14 cm to 21 cm / 5 ½ to 8 ¼ inches (circumference)
Aktiia Bracelet weight| approximatively 18g (0.6 ounces).
Bracelet essential performance: Safety| The bracelet connects with the app
Bracelet essential performance: EMC| Normal Mode : If the EUT does not sync measured data after performing a test, or if the measurements are unsatisfactory (in accordance to the customer’s standards), or if the Bluetooth connectivity with the phone is lost, then it is considered a failure.

Charging Mode : If the EUT’s Bluetooth connectivity with the phone is lost or if the EUT stops charging, then it is considered a failure.

Power Source| Li-Ion rechargeable battery 3.7 Vdc 55mAh
Charging Cable| USB Micro B to USB A cable
Note: The user needs to supply an external USB power supply that meets the following requirements:| ·          Voltage: 5VDC

·          Current: 100mA (minimum)

Electromagnetic Compatibility (EMC) and Radio Frequency (RF) statements

  • The Aktiia 24/7 Blood Pressure Monitor System (hereafter, the Aktiia System) needs special precautions regarding EMC and needs to be installed and put into serviceaccording to the EMC information provided in the following section.

  • The Aktiia System is suitable for use in all establishments, including domestic establishments and those directlyconnected to the public low voltage power supply network that supplies buildings used for domestic purposes.

  • The Aktiia System uses RF energy only for its internal function. Therefore, its RF emissions are very lowand are not likely to cause any interference in nearby electronic equipment.

  • The Aktiia System is intended for use in an electromagnetic environment in which radiated RF disturbances arecontrolled. Interference may occur in the vicinity of equipment marked with the following symbol .

  • Portable and mobile RF communication equipment (e.g. cell phones) can affect the Aktiia System. CAUTIONS :

  • The Aktiia Bracelet should not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary, it should be observed to verify normal operation in the configuration in which it will be used.

  • Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.

  • Wireless communication devices (including home networking devices & peripherals such as antenna cables and external antennas)
    should be used no closer than 30 cm (12 inches) to any part of the Aktiia Bracelet, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.”

  • Avoid exposure of this device to known sources of EMI (electromagnetic interference) such as magnetic resonance imaging (MRI) systems, diathermy, lithotripsy, electrocautery, RFID (radio frequency identification), and electromagnetic security systems such as metal detectors. Note that the presence of RFID devices may not be obvious.

  • If such interference is suspected reorient equipment if possible, to maximize distances.

  • If, at any time, the device loses Bluetooth connectivity with the phone, the user must manually reestablish connectivity with the device through the Aktiia App.

Electromagnetic compatibility and coexistence information

Guidance and manufacture’s declaration – electromagnetic emissions

The Aktiia System is intended for use in the electromagnetic environment specified below. The user of Aktiia Bracelet should ensure that it is used in such an environment. The Aktiia System is suitable for use in “Home Healthcare Environment”, i.e. all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Emission test| Compliance| Electromagnetic environment – guidance
Conducted emissions CISPR11| Groupe 1| The Aktiia System uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Radiated emissions CISPR11| Class B
Harmonic emissions IEC 61000-3-2| ****

Not Applicable

Voltage fluctuations/ Flicker emissions

IEC 61000-3-3

| ****

Not Applicable

Guidance and manufacturer’s declaration – electromagnetic immunity

The Aktiia System is intended for use in the electromagnetic environment specified below. The user of the Aktiia System should ensure that it is used in

such an environment. The Aktiia System is suitable for use in “Home Healthcare Environment”, i.e. all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Immunity test| 60601-1-2 test levels| Compliance
Electrostatic discharge

IEC 61000-4-2

| ± 8 kV contact

± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV in air

| ± 8 kV contact

± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV in air

Radiated RF electromagnetic field IEC 61000-4-3| 10 V/ m

80 MHz à 2.4 GHz

80 % MA à 1 kHz

| 10 V/ m

80 MHz à 2.4 GHz

80 % MA à 1 kHz





Proximity fields from RF wireless communications equipment

IEC 61000-4-3 (interim method)

| 9 V/ m

710 MHz, 745 MHz,

780 MHZ, 5240 MHz, 5550 MHz, 5785 MHz


27 V/ m

385 MHz


28 V/ m

450 MHz, 810 MHz,

870 MHz, 930 MHz,

1720 MHz, 1845 MHz, 1970 MHz, 2450 MHz

| 9 V/ m

710 MHz, 745 MHz,

780 MHZ, 5240 MHz, 5550 MHz, 5785 MHz


27 V/ m

385 MHz


28 V/ m

450 MHz, 810 MHz,

870 MHz, 930 MHz,

1720 MHz, 1845 MHz, 1970 MHz, 2450 MHz

Power frequency magnetic field IEC 61000-4-8| 30 A/ m

50 Hz

| 30 A/ m

50 Hz

---|---|---
Olivier|  |
 |  |

Essential Performance:

Performance as required by IEC 60601-1-2 and IEC 80601-2-30 clause 202.8.1.101 for basic safety, the bracelet connects with the app. For the EMC the following were monitored during testing:

Normal Mode : If the EUT does not sync measured data after performing a test, and/or if the measurements are unsatisfactory in accordance to IEC 80601-2-30 clause 202.8.1.101 as below then it is considered a failure.
Charging Mode: If the EUT stops charging, then it is considered a failure.

Performance criterion per IEC 80601-2-30 clause 202.8.1.101 were monitored as below:

  • component failures; (Normal Mode & Charging Mode)
  • changes in programmable parameters or settings; (Normal Mode & Charging Mode)
  • reset to default settings; (Normal Mode & Charging Mode)
  • Per manufacturer, to meet to appropriate technology, CUFF pressure at any point of the
    NOMINAL measurement range greater than 2 mmHg (0,3 kPa) per IEC 80601-2-30 clause 202.8.1.101 clause d) was monitored with below parameters. SBP EMC bias, DBP EMC bias, RF_Num_Rec, Spread_DueToNoise_SBP, NRF2_Num_Precise_Rec/NRF2_Num_meas_Rec, and NRF2_acceptance were calculated internally by Aktiia SA.

During the tests, the following rules need to be met:

  • The percentage of removed measurements to be <= Floor(0.04 x RF_Num_Rec)

  • SBP_EMC_bias < 2 mmHg

  • DBP_EMC_bias < 2 mmHg

  • SBP_EMC_bias + (2.576 x Spread_DueToNoise_SBP) < 2

  • SBP_ EMC_bias – (2.576 x Spread_DueToNoise_SBP) > -2

  • DBP_EMC_bias + (2.576 x Spread_DueToNoise_DBP) < 2

  • DBP_EMC_bias – (2.576 x Spread_DueToNoise_DBP) > -2
    After the tests, the following rules need to be met:

  • NRF2_Num_Precise_Rec/ NRF2_Num_meas_Rec > 0.96.

  • NRF2_acceptance is calculated.

Aktiia SA monitored the accuracy of parameters from the raw data provided by Intertek. These parameters were monitored,

  • during and after exposure to non-transient phenomena; and
  • after exposure to transient phenomena.

If, at any time, the EUT loses Bluetooth connectivity with the phone, the user must manually reestablish connectivity with the EUT through the Aktiia’s proprietary app. Disconnection with the phone will not create any risk of harm to the user and will only cause a slight procedural delay in uploading collected raw data.

EMC Test Plan & Summary

The Aktiia System is intended for use in the electromagnetic environment specified below from the EMC testing.report The user of Aktiia Bracelet should ensure that it is used in such an environment. The Aktiia System is suitable for use in “Home Healthcare Environment”, i.e. all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

IEC 60601-1-2 (Edition 4.0): 2014+AMD1:2020, FCC Part 15 Subpart B, ISED ICES-003

Proposed Tests for Home Healthcare Environment


Basic Standard| Test Specifications| Applicable Ports| Test Mode and Configuration
CISPR 11| Radiated Emission| –| Charging Mode at 120 Vac 60 Hz Normal

Mode on Battery Mode

CISPR 11| Conducted Emission| –| Charging Mode at

100V 50Hz, 120V 60Hz, 240V 50Hz

ANSI C63.4| Radiated Emission| –| Charging Mode at 120 Vac 60 Hz Normal Mode on Battery Mode
---|---|---|---
ANSI C63.4| Conducted Emission| AC Power| Charging Mode at 120V 60Hz
EN/IEC 61000-4-2| Electrostatic Discharge,

± 8 kV contact discharge

| Enclosure|

Charging Mode at 120 Vac 60 Hz Normal Mode on Battery Mode

Electrostatic Discharge,

± 15 kV air discharge

| Enclosure
Electrostatic Discharge,

± 8 kV contact discharge

| PATIENT

coupling PORT

| Not Applicable3
Electrostatic Discharge,

± 15 kV air discharge

| PATIENT

coupling PORT

EN/IEC 61000-4-3| Radiated Immunity, 10 V/m, 80 –2700 MHz,

80% AM at 1 kHz

& Proximity fields from RF Wireless communications equipment

| Enclosure|

Charging Mode at 120 Vac 60 Hz Normal Mode on Battery Mode

EN/IEC 61000-4-4| Fast Transient/Burst

± 2 kV, 5/50 nsec pulse, 100 kHz repetition freq.

| AC Power|

Charging Mode at 120 Vac 60 Hz

---|---|---|---
Fast Transient/Burst

± 2 kV, 5/50 nsec pulse, 100 kHz repetition freq.

| DC Power| Not Applicable4
Fast Transient/Burst (Capacitive clamp)

± 1 kV, 5/50 nsec pulse,

100 kHz repetition freq.

|

Input/ Output

Lines

| Not Applicable1
EN/IEC 61000-4-5| Surges, 1.2/50 (8/20) S

± 2 kV (line to earth)

± 1 kV (line to line)

| AC Power| Not Applicable5
Surges, 1.2/50 (8/20) S

± 2 kV (line to earth)

± 1 kV (line to line)

| DC Power| Not Applicable4
Surges, 1.2/50 (8/20) S

± 2 kV (line to earth)

| Input/ Output

Lines

| Not Applicable2
IEC 60601-1-2 (Edition 4.0): 2014+AMD1:2020, FCC Part 15 Subpart B, ISED ICES-003

Proposed Tests for Home Healthcare Environment


Basic Standard| Test Specifications| Applicable Ports| Pass/Fail

Comments

EN/IEC 61000-4-6| Continuous Conducted RF, 80% AM (1 kHz)

3 Vrms, 0.15-80 MHz

6 Vrms in ISM and amateur radio Bands within

150kHz – 80MHz

| AC Power|

Charging Mode at 120 Vac 60 Hz

Continuous Conducted RF, 80% AM (1 kHz)

3 Vrms, 0.15-80 MHz

6 Vrms in ISM and amateur radio Bands within 150kHz – 80MHz

| DC Power|

Not Applicable4

 | Continuous Conducted RF,|  |
---|---|---|---
80% AM (1 kHz)| Input/ Output Lines| Not Applicable1
3 Vrms, 0.15-80 MHz|  |
6 Vrms in ISM and amateur radio Bands

within

|  |
150kHz – 80MHz Cables longer than 3 m|  |
Continuous Conducted RF, 80% AM (1 kHz)

3 Vrms, 0.15-80 MHz

6 Vrms in ISM and amateur

radio Bands within 150kHz – 80MHz

| PATIENT|
coupling PORT| Not Applicable3
EN/IEC 61000-4-8| Power Frequency Magnetic Field

30A(rms)/m at 50Hz /60Hz

|

Enclosure

| Charging Mode at 120 Vac 60 Hz Normal Mode on Battery Mode
 | Continuous Conducted RF,|  |
---|---|---|---
80% AM (1 kHz)| Input/ Output Lines| Not Applicable1
3 Vrms, 0.15-80 MHz|  |
6 Vrms in ISM and amateur radio Bands

within

|  |
150kHz – 80MHz Cables longer than 3 m|  |
Continuous Conducted RF, 80% AM (1 kHz)

3 Vrms, 0.15-80 MHz

6 Vrms in ISM and amateur

radio Bands within 150kHz – 80MHz

| PATIENT|
coupling PORT| Not Applicable3
EN/IEC 61000-4-8| Power Frequency Magnetic Field

30A(rms)/m at 50Hz /60Hz

|

Enclosure

| Charging Mode at 120 Vac 60 Hz Normal Mode on Battery Mode
 |  |  | b l

e 5

---|---|---|---
Voltage Dips > 95% reduction, 1 period At 0°| AC Power| N

o t A

p p l i c a b l

e 5

Voltage Interruptions > 95% reduction, 250/300 periods| AC Power| Not Applicable5
EN 61000-3-2| Emissions: Harmonics

Disturbances

| AC Power| Not Applicable5
---|---|---|---
EN 61000-3-3| Emissions: Voltage

Fluctuation

| AC Power| Not Applicable5

The EUT does not contain signal input/output lines longer than 3 m in length.

  • The EUT does not contain signal ports that connect directly to the outdoors.
  • The EUT does not contain any patient coupling ports.
  • The EUT does not contain DC power ports.
  • Per the client, the power adapters are not provided with the EUT; only power cable and docking station will be provided. See Section 5.4 for details

IEC 60601-1-2 (Edition 4.0): 2014+AMD1:2020, FCC Part 15 Subpart B, ISED ICES-003

Test Summary for Home Healthcare Environment


Basic Standard| Test Specifications| Applicable Ports| Pass/Fail Comments
ANSI C63.4| Radiated Emission| –| Complies
ANSI C63.4| Conducted Emission| AC Power| Complies
CISPR 11| Radiated Emission| –| Complies
---|---|---|---
CISPR 11| Conducted Emission| AC Power| Complies
EN/IEC 61000-4-2| Electrostatic Discharge,

± 8 kV contact discharge

| Enclosure| Complies6
Electrostatic Discharge,

± 15 kV air discharge

| Enclosure
Electrostatic Discharge,

± 8 kV contact discharge

| PATIENT

coupling PORT

| Not Applicable3
Electrostatic Discharge,

± 15 kV air discharge

| PATIENT

coupling PORT

EN/IEC 61000-4-3| Radiated Immunity, 10 V/m, 80 –

2700 MHz,

80% AM at 1 kHz

& Proximity fields from RF Wireless communications equipment

| Enclosure| ****

Complies6

EN/IEC 61000-4-4| Fast Transient/Burst

± 2 kV, 5/50 nsec pulse, 100 kHz repetition freq.

| AC Power| Complies6
---|---|---|---
Fast Transient/Burst

± 2 kV, 5/50 nsec pulse, 100 kHz repetition freq.

| DC Power| Not Applicable4
Fast Transient/Burst (Capacitive clamp)

± 1 kV, 5/50 nsec pulse, 100 kHz

repetition freq.

| ****

Input/ Output Lines

| Not Applicable1
EN/IEC 61000-4-5| Surges, 1.2/50 (8/20) S

± 2 kV (line to earth)

± 1 kV (line to line)

| AC Power| Not Applicable5
Surges, 1.2/50 (8/20) S

± 2 kV (line to earth)

± 1 kV (line to line)

| DC Power| Not Applicable4
Surges, 1.2/50 (8/20) S

± 2 kV (line to earth)

| Input/ Output Lines| Not Applicable2
IEC 60601-1-2 (Edition 4.0): 2014+AMD1:2020

Test Summary for Home Healthcare Environment

|
---|---
Basic Standard| Test Specifications| Applicable Ports| Pass/F ail Comm

ents

EN/IEC 61000-4-6| Continuous Conducted RF, 80% AM (1 kHz)

3 Vrms, 0.15-80 MHz

6 Vrms in ISM and amateur radio Bands within 150kHz – 80MHz

| AC Power| Compli es6
Continuous Conducted RF, 80% AM (1 kHz)

3 Vrms, 0.15-80 MHz

6 Vrms in ISM and amateur radio Bands within 150kHz – 80MHz

| DC Power| ****

Not Applica ble4

 | Continuous Conducted RF, 80% AM (1 kHz)

3 Vrms, 0.15-80 MHz

6 Vrms in ISM and amateur radio Bands within 150kHz – 80MHz Cables longer than 3 m

| ****

Input/ Output Lines

| Not Applica ble1|
---|---|---|---|---
Continuous Conducted RF, 80% AM (1 kHz)

3 Vrms, 0.15-80 MHz

6 Vrms in ISM and amateur radio Bands within 150kHz – 80MHz

| PATIENT coupling PORT| ****

Not Applica ble3

EN/IEC 61000-4-8| Power Frequency Magnetic Field 30A(rms)/m at 50Hz

/60Hz

| ****

Enclosure

| Compli es6



EN/IEC 61000-4-39

| Radiated Fields in close proximity 8A/m, CW, 30kHz; 65A/m (rms), PM at 2.1 kHz, 50% duty

cycle, 134.2kHz;

7.5A/m (rms), PM at 50 kHz, 50% duty cycle, 13.56MHz

| ****


Enclosure

| ****

Compli es6

|
---|---|---|---|---
EN/IEC 61000-4-11| Voltage Dips 30% reduction, 25/30 periods At 0°| AC Power| Not Applica

ble5

Voltage Dips > 95% reduction,

0.5 period

At 0°, 45°, 90°, 135°, 180°, 225°,

270° and 315°

| AC Power| N

o t A

p p l i c a

b

 |  |  | l

e 5

|
---|---|---|---|---
Voltage Dips > 95% reduction, 1 period At 0°| AC Power| N

o t A

p p l i c a b l

e 5

Voltage Interruptions > 95% reduction, 250/300 periods| AC Power| Not Applica ble5
EN 61000-3-2| Emissions: Harmonics

Disturbances

| AC Power| Not Applica

ble5

|
---|---|---|---|---
EN 61000-3-3| Emissions: Voltage Fluctuation| AC Power| Not

Applica ble5

The EUT does not contain signal input/output lines longer than 3 m in length.

  • The EUT does not contain signal ports that connect directly to the outdoors.
  • The EUT does not contain any patient coupling ports.
  • The EUT does not contain DC power ports.
  • Per the client, the power adapters are not provided with the EUT; only power cable and docking station will be provided. See Section 5.4 for details.
  • The Performance Criteria per IEC 80601-2-30:2018 clause 202.8.1.101 were applied, refer to report section 5.7 for details.

Radio Characteristics

AKTiiA G1-Blood0Pressure-Monitoring-Bracelet-fig-
13

FCC Statement

The present product has the FCC ID: 2AUVE-AKTIIAG1
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this
device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
The Device wireless operation is safe and complies to RF Exposure requirements.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

The FCC-ID is located in the settings menu of the Aktiia App.

Disposal

This symbol applied on the device or its packaging means that at the end of its useful life the product must not be disposed of with domestic waste.
At the end of the device’s useful life, the user must deliver it to the able collecting centers for electric and electronic garbage or give back to the retailer when purchasing a new device. Disposing of the product separately prevents possible negative consequences for the environment and for health, deriving from inadequate disposal. It also allows the recovery of materials of which it’s made up in order to obtain an important saving of energy and resources and to avoid negative effects to the environment and health. In case of abusive disposal of device by the user, will be applied administrative endorsements in compliance with current standard. The device and its parts are made with regard to disposal, as appropriate, in accordance with national or regional regulations

Network security recommendations

The following recommendations detail security measures that Aktiia users should follow to ensure appropriate protection of their personal data. Failure to comply with these recommendations may lead to user personal data leakage or destruction.
Only use mobile application authorized by Aktiia. Aktiia only makes its mobile application and subsequent updates available on official app stores (e.g. Apple App store).
Use unique credentials (username and password) for login to your Aktiia account. Safely store your password so that no other person can access it. It is recommended to regularly update your password, at least once every 3 months.
Do not let other people login to your Aktiia account on your behalf.

Clinical Safety Information

The Aktiia Bracelet was validated in a clinical study compared to double- auscultation measurements at the upper arm. The data set used for comparison comprised of 4’517 data pairs generated from 85 adults ages 22-60 years old, from a multitude of body and arm positions, under conditions of natural changes in blood pressure (BP), under conditions of induced changes in BP by physical maneuvers, and at different times throughout the 7-day calibration interval.

The primary outcome results including all body positions and all time points (n=4’517 data pairs) are shown in the following table:

Primary outcome| Mean value of differences| Standard deviation of all differences
---|---|---
Systolic Blood Pressure Performance| -1.0 mmHg| 10.8 mmHg
Diastolic Blood Pressure

Performance

| -4.9 mmHg| 9.5 mmHg

  • Because the Aktiia system is not intended to be used only for a single measurement, but to repeatedly measure the BP of a user over the long-term, it is also important that the device provides a reliable representation of BP over time. The results of a subject-level analysis that included all the data collected over the one-week study period found that the device error for systolic BP is within 10 mm Hg for 85% of users and for diastolic BP within 10 mm Hg for 75% of users.
  • A subgroup analysis found that there is no systematic bias of results across the following subgroups: age, sex, skin tone, or hypertension status, apart from an underestimation of the diastolic BP (but not systolic BP) for subject with highly pigmented skin.
  • Device performance was assessed in the following body positions: sitting, lying, and standing and at different elevations of the arm. For most body and arm positions, device performance is similar to the overall performance presented above. However, the device did show a systematic under-estimation of diastolic blood pressure when measurements are made in the standing position.
  • Data from a motion study that evaluated 31 subjects found that an average of 27 measurements are performed over 24 hours, and that those measurements are only obtained when the subject is quiet in any of the sitting, lying or standing positions.

Aktiia Bracelet G1 Blood pressure monitor
Aktiia SA, Bassin 8a, 2000 Neuchâtel, Switzerland Website : www.aktiia.com
Customer Service: support@aktiia.com
User Manual & Symbols Glossary: www.aktiia.com/uk/manuals/electronicsymbols
Version 1 – 11/10/2022