APERA PC400 Portable pH or Conductivity Meter Instruction Manual
- June 3, 2024
- APERA
Table of Contents
- BRIEF INTRODUCTION
- TECHNICAL SPECIFICATIONS
- INSTRUMENT DESCRIPTION
- pH MEASUREMENT
- mV MEASUREMENT
- CONDUCTIVITY MEASUREMENT
- PARAMETER SETTING
- COMPLETE KIT
- APPENDIX 1: TABLE OF PARAMETER SETTING AND FACTORY DEFAULT SETTING
- APPENDIX 2: TABLE OF SELF-DIAGNOSIS
- APPENDIX 3: ICONS AND ABBREVIATION
- WARRANTY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
PC400 Portable pH/Conductivity Meter
PH400 Portable pH Meter
EC400 Portable Conductivity Meter
PC400 Portable pH/Conductivity Meter
Instruction Manual APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de
BRIEF INTRODUCTION
Thank you for purchasing Apera Instruments PC400 Portable pH/Conductivity
Meters. Before using the product, please read this manual carefully to help
you properly use and maintain the product. For technical support, please
contact us at info@aperainst.de
Apera Instruments reserves the right to update the content of this manual
without giving prior notices.
1.1 Measuring Parameters
Measuring Parameters | PH400 | EC400 | PC400 |
---|---|---|---|
pH/mV | √ | √ | |
Conductivity/TDS | √ | √ | |
Temperature | √ | √ | √ |
1.2 Features and Functions
- The built-in microprocessor chip enables advanced functions such as auto calibration, auto temperature compensation, auto electrode recognition, parameter setting, self-diagnosis, calibration reminder, calibration time check, auto power-off, low-battery reminder, etc.
- The meter adopts advanced digital processing technology, intelligently improves the response time and accuracy of the measurements. Stable reading and auto lock display mode are available for choice.
- Comes with a carrying case, which includes the meter, the electrode(s), calibration solutions, and other accessories, convenient for in-field use.
- Meets IP57 Waterproof and dustproof rating, ideal for use under harsh environment.
1.3 Features in pH Measurement (for PH400 and PC400)
- 1 to 3 points auto calibration with calibration guide and self-diagnosis function.
- Automatic recognition of calibration solutions. Two series of standard solutions for choice: USA and NIST
1.4 Features in Conductivity Measurement (for EC400 and PC400)
- 1 to 4 points auto calibration with calibration guide and auto-check function.
- Single-tap switch between conductivity and TDS.
- Automatic recognition of conductivity calibration solutions.
TECHNICAL SPECIFICATIONS
2.1 Parameter Specifications
| Technical Parameters| Applicable Models
---|---|---
pH| Measuring
Range| (0 to 14.00) pH| PH400
PC400
Resolution| 0.01 pH
Accuracy| ±0.01 pH ±1 digit
Temperature
Compensation
Range| (0 to 100) °C, Automatic or Manual
mV| Measuring
Range| -1000 mV to 0 to 1000 mV
Resolution| 1 mV
Accuracy| ±0.2% ES ±1 digit
Conductivity| Measuring Range| Conductivity: 0 to 200 mS/cm, including 5
ranges: (0.00 to 19.99) pS/cm, (20.0 to 199.9) pS/cm, (200 to 1999) pS/cm,
(2.00 to 19.99) mS/cm, (20.0 to 199.9) mS/cm
TDS: (0 to 100) g/L, including 5 ranges:
(0.00 to 9.99) mg/L, (10.0 to 99.9) mg/L,
(100 to 999) mg/L, (1.00 to 9.99) g/L,
(10.0 to 100.0) g/L| EC400
PC400
Resolution| 0.01/0.1/1 pS/cm, 0.01/0.1 mS/cm
Accuracy| ±1.0% ES ±1 digit
Temperature
Compensation
Range| (0 to 50)°C, Automatic or Manual
Electrode
Constant| 0.1 / 1 / 10 cm-‘|
Temperature| Measuring
Range| 0 to 100 °C| PH400
Resolution| 0.1 °C| EC400
PC400
Accuracy| ±0.5 °C±1 digit|
2.2 Other Specifications
Power Supply | AA Batteries 4 (1.5V4) |
---|---|
IP Rating | IP57 Waterproof and Dustproof |
Dimensions and Weight | Meter: 86×196×33 mm/ 0.335 kg |
With case: 360×275×80 mm/ 1.6 kg
INSTRUMENT DESCRIPTION
3.1 LCD Display
- Measuring parameters
- Measuring value
- Reminder icons
- Measurement unit
- Temperature unit
- Measuring unit in calibration
- Calibration value, numberings of data storage, and reminder icons
- Temperature and reminder icons
- ATC—Auto Temperature Compensation; MTC— Manual Temperature Compensation
- Stable reading icon
- Completed calibration icons
- Auto-Lock reading mode
- Low battery reminder. Please replace batteries when this icon is displayed.
3.2 Keypad
3.2.1 Short Press — <1.5 s ;Long Press — >1.5 s 。
3.2.2 Power On: Short press to power on: LCD displays the measuring mode used
last time (Backlight turned on for 1 minute).
3.2.3 Power Off: The meter can only be turned off in measuring mode by short
pressing
Special Notes: Pressing in calibration mode or parameter setting mode will NOT
turn the meter off.
Users need to press first to go back to measuring mode, and then press to
power off.
Table- 1 Keypad Operation and Functions
Keypad | Operation | Functions |
---|---|---|
Short Press | • Power on/off | |
Short Press | Choose measuring mode: |
• PH400 pH Meter: →
• EC400 Conductivity Meter: →
• PC400 pH/Cond. Meter: → → →
Long Press| • Enter parameter setting
| Short Press| • Cancel any operation and enter measurement mode
Long Press| • Enter calibration mode
| Short Press| • In measurement mode: press to turn on/off the backlight
• In calibration mode: press to calibrate
• In parameter setting mode: press to confirm choice
| Short Pressor Long Press| • In manual temperature compensation (MTC) mode:
Short press to adjust temperature, long press to adjust swiftly.
• In parameter setting mode: press to change the numbering of parameters in
main menu and sub-menu.
• In sub-menu, press to change parameters and settings.
3.3 Connectors
The meter adopts 8-pin connector, into which pH, ORP, and conductivity
electrode can be connected. When connected, the meter will automatically
switch to the correspondent measurement mode.
3.4 Display Mode
3.4.1 Stable Reading Display Mode
When the measuring value is stable, the screen displays as shown in Figure-2.
If does not appear or is flashing, that means the measuring value has not been
stable. Users should wait for the smiley face and not record the readings or
conduct calibrations at that moment.
3.4.2 Auto-Lock Display Mode
In parameter setting P4.6, select “On” to turn on the auto-lock display mode,
In which the reading will be automatically locked after the measuring value
has been stable for more than 10 seconds, and the HOLD icon will come up as
shown in Figure-3. Short press to cancel the hold.
pH MEASUREMENT
4.1 Information regarding the pH Electrode
The instrument is equipped with a 201T-S 3-in-1 Combination pH Electrode. Its
built-in temperature sensor allows for auto. Temperature compensation. There
is a vial stored with 3M KCl storage solution on top of the electrode. Before
using, loosen the cap of the vial, take out the electrode and rinse it in
distilled water and then do measurement. When not in use, place the electrode
back to the vial and tighten the cap so that the pH sensor can be stored in
the best condition. When the pH electrode is dipped into the test sample
solution, stir it for a few seconds to remove potential air bubbles inside the
probe to help the measurement get stabilized quickly.
When connecting the electrode into the instrument, please slowly rotate the
connector, identify the location of the mount to insert it, and then screw it
on. Please note that do not pull on the cables in case of poor contact.
Please keep the connector clean and dry. For detailed information of the
electrode’s maintenance, please refer to section 4.5.
4.2 Information regarding pH Calibration
4.2.1 Standard Buffer Solutions
The instrument adopts two series of standard buffer solutions: USA and NIST as
shown in Table-2. Users can select which one to use in P1.1 (refer to section
7.3).
Table-2
Calibration Icons | pH Standard Calibration Solution Series |
---|---|
USA Series | NIST Series |
3-point
calibration| L
| 4.00 pH| 4.01 pH
M| 7.00 pH| 6.86 pH
H
| 10.01pH| 9.18 pH
4.2.2 pH Calibration Modes
The instrument has 1 to 3 points auto calibration mode. The 1st point must be
7.00 pH (or 6.86 if using NIST). Then choose other calibration solutions to
conduct 2nd and 3rd points (see Table-3 for details). In the process of
calibration, the meter will display the electrode’s slope in acid and alkaline
ranges.
Table-3 3-point Calibration Mode
| USA| NIS| Calibration icon| When to adopt
---|---|---|---|---
1-Point Calibration| 7.00 pH| 6.86 pH| M
| accuracy ±0.1 pH
2-Point Calibration| 7.00 pH and 4.00/1.68 pH| 6.86 pH and
4.01/1.68 pH| LM
| Measuring range: 0 to 7.00 pH
7.00 pH and 10.01 pH| 6.86 pH and 9.18
pH| MH
| Measuring range: >7.00 pH
3-Point Calibration| 7.00pH, 4.00/1.68 pH and 10.01
pH| 6.86pH, 4.01/
1.68pH, 9.18 pH| LMH
| Wide measuring range
4.2.3 How often to calibrate
The frequency that you need to calibrate your meter depends on the tested
samples, performace of electrodes, and the requirement of the accuracy. For
High-Accuracy measurements (≤ ±0.02pH), the meter should be calibrated before
test every time; For ordinary-accuracy Measurements (≥±0.1pH), once
calibrated, the meter can be used for about a week or longer. In the following
cases, the meter must be re-calibrated:
- The electrode hasn’t been used for a long time or the electrode is brand new.
- After measuring strong acid (pH<2) or strong base (pH>12) solutions.
- After measuring fluoride-containing solution and strong organic solution
- There is a big difference between the temperature of the test sample and the temperature of the buffer solution that is used in the last calibration.
4.3 pH Calibration (Take 3-point calibration as an example)
4.3.1 Press to enter calibration mode. CAL1 icon will flash in the upper right
corner of the LCD. 7.00 pH will flash in the lower right corner of the LCD,
reminding you to use pH 7.00 buffer to conduct 1st point of calibration.
4.3.2 Use distilled water to rinse off electrode and then dry it. Dip it into
pH 7.00 buffer solution, stir gently and let it stand still and wait for the
reading to become stable. In the lower right corner of LCD, the process of
auto recognizing the buffer solution will be displayed. Pressing before the
buffer is recognized will generate Er2 (please refer to table 6).
4.3.3 When the meter locks 7.00 pH, displays on LCD. Press key to calibrate
the meter. End icon appears after calibration is done. The 1st point
calibration is finished. In the meanwhile, CAL2 will flash at the upper right
corner, and 4.00 pH & 10.01 pH will flash alternately at the bottom right,
indicating using pH 4.00 or pH 10.01 buffer solution to make the 2nd point
calibration.
4.3.4 Take out pH electrode, rinse it in distilled water, dry it, and dip
iinto pH 4.00 buffer solution. Stir the solution gently and let stand still in
the buffer solution until a stable reading is reached. The meter’s displaywill
show the recognition process of calibration buffer solution at the lower right
of LCD. When the meter recognizes 4.00 pH,
displays on LCD. Press key to calibrate the meter. End icon and electrode
slope of acidity range display after calibration is done. In the meanwhile,
CL3 will flash at the upper right corner of the LCD, and 10.01 pH will flash
at the lower right, indicating using pH10.01 buffer solution to make the 3rd
point calibration.
4.3.5 Take out pH electrode, rinse it in distilled water, dry it, and dip it
into pH 10.01 buffer solution. Stir the solution gently and let it stand still
in the buffer solution until a stable reading is reached. The meter’s display
will show recognition process of calibration buffer solution at the bottom
right of LCD. When the meter recognizes 10.01 pH, displays on LCD. Press key
to calibrate the meter. End icon and electrode slope of alkalinity range
display after calibration is done. The meterreturns to the measurement mode,
displays stable measuring value andcalibration guide icons. Please see
Figure-4 for the above calibrationprocess.
During the calibration process, press key to exit from the calibration mode.
Correspondent calibration guide icons will appear on the LCD.
4.4 Sample Measurement
4.4.1 Rinse the pH electrode in distilled water, dry it, and dip it into
sample solution. Stir the solution gently and let it stand still in the sample
solution until icon appears and stays on LCD, get the pH reading, which is pH
value of sample solution, please refer to Figure-5 for calibration and
measurement process of the pH meter.
Figure-5 Calibration and measurement process of pH meter
4.4.2 Self-Diagnosis Information
Table – 4 Self-diagnosis information
Icons | Contents | How to fix |
---|---|---|
Wrong pH buffer solution or the buffer solution out of range. | 1.Check |
whether pH buffer solution is correct. 2.Check whether the meter connects the
electrode properly.
3.Check whether the electrode is damaged.
| Press when measuring value is not stable during calibration.| Press key when
icon appears and stays.
| During calibration, the measuring value being unstable for over 3 minutes|
1.Check whether there are air bubbles in the glass bulb. To remove the air
bubble, hold the electrode firmly and flick the electrode in a downward motion
for several times.
2.Replace with a new pH electrode.
| pH electrode zero electric potential out of range
<-60mV or >60mV )| 1.Check whether there are air bubbles in the glass bulb. To
remove the air bubble, hold the electrode firmly and flick the electrode in a
downward motion for several times.
2.Check whether pH buffer solution is correct. 3.Replace with new pH
electrode.
| pH electrode slope out of range ( <85% or >110% )
4.4.3 pH isothermal measurement principle
According to the pH isothermal measurement principle, the closer the test
sample’s temperature is to the calibration solution’s, the higher the accuracy
of the measurement. So this principle is recommended to follow when conducting
tests.
4.4.4 Restore to factory default
The instrument has a function to return to factory default setting, which can
be set up in P1.2 (refer to section 7.3). Returning to factory default setting
is to restore the meter to the theoretical value (zero potential pH is 7.00,
slope is 100%), and set all the parameters to default settings (see appendix
1). When the meter’s calibration or measurement is performing abnormally,
users can use this function to let the meter return to factory default mode,
and conduct calibration and then test again. Please note that this function is
irreversible once used.
4.5 Maintenance of the pH Electrode
4.5.1 Daily maintenance
The soaking solution contained in the supplied protective vial is used to
maintain the sensitivity of the glass bulb and junction. Loosen the cap,
remove the electrode and rinse in distilled water before taking a measurement.
Insert the electrode and tighten the cap after measurements to prevent the
reference solution from leaking. If the soaking solution is turbid or moldy,
replace it with 3M KCL solution. (SKU:
AI1107). Soaking the probe in other brands’ storage solution could potentially
cause damage to the probe.
The electrode should NOT be soaked in pure or distilled water, protein
solution or acid fluoride solution.
In addition, do not soak the electrode in organic lipids.
4.5.2 Calibration buffer solution
For calibration accuracy, the pH of the standard buffer solution must be
reliable. The buffer solution should be refreshed often, especially after
heavy use. We recommend 10-15 times of use before replacing the pH buffers.
4.5.3 Protect glass bulb
The sensitive glass bulb at the front of the combination electrode should not
come in contact with hard surfaces. Scratches or cracks on the electrode will
cause inaccurate readings. Before and after each measurement, the electrode
should be rinsed with distilled water. If a sample sticks to the electrode or
it’s contaminated, the electrode should be thoroughly cleaned using a soft
brush and then rinsed with distilled water. After that, soak it in the 3M KCL
(SKU: AI1107) solution again for 6 hours.
4.5.4 Renew glass bulb
Electrodes that have been used over a long period will become aged. Soak the
electrode in 0.1mol/L hydrochloric acid for 24 hours, then wash the electrode
in distilled water, then soak it in 3M KCL (SKU: AI1107) storage solution for
24 hours.
The method to prepare 0.1mol/L hydrochloric acid: dilute 9mL hydrochloric acid
in distilled water to 1000mL.
For serious passivation, dip the bulb in 4% HF (hydrofluoric acid) for 3-5
seconds, and wash it in distilled water, then soak it in the 3M KCL storage
solution (SKU: AI1107) for 24 hours to renew it.
4.5.5 Clean contaminated glass bulb and junction (please refer to
Table-5)
Table-5 Clean contaminated glass bulb and junction
Contamination | Cleaning Solutions |
---|---|
Inorganic metal oxide | Dilute acid less than 1mol/L |
Organic lipid | Dilute detergent (weak alkaline) |
Resin macromolecule | Dilute alcohol, acetone, ether |
Proteinic haematocyte sediment | Acidic enzymatic solution (saccharated yeast |
tablets)
Paints| Dilute bleacher, peroxide
Notes:
The instruments will NOT give accurate and stable pH readings when measuring
distilled or deionized water. This because distilled and deionized water do
not have enough ions present for the electrode to function properly. To
measure distilled or deionized water’s pH, users need to use a specialized
electrode.
Contact us for more details.
The electrode housing is polycarbonate. When using cleaning solutions, take
cautions on carbon tetrachloride, trichlorethylene, tetrahydrofuran, acetone,
etc., which will dissolve the housing and invalidate the electrode.
mV MEASUREMENT
5.1 ORP measurement
Press key, and switch the meter to mV measurement mode. Connect ORP electrode
(the 301Pt-S combination ORP electrode is sold separately) and dip it in
sample solution, stir the solution briefly and allow it to stay in the
solution until icon appears and get the reading.
ORP means Oxidation Reduction Potential. The unit is mV.
5.2 Notes on ORP measurement
5.2.1 ORP measurement does not require calibration. When the user is not sure
about ORP electrode quality or measuring value, use ORP standard solution to
test mV value and see whether ORP electrode or meter works properly. Table-6
is the data of standard ORP solution for 222 mV.
Table-6
°C | 10 | 15 | 20 | 25 | 30 | 35 | 38 | 40 |
---|---|---|---|---|---|---|---|---|
mV | 242 | 235 | 227 | 222 | 215 | 209 | 205 | 201 |
5.2.2 Clean and activate ORP electrode
After the electrode has been used over a long period of time, the platinum
surface will get polluted which causes inaccurate measurement and slow
response. Please refer to the following methods to clean and activate ORP
electrode:
(a) For inorganic pollutant, soak the electrode in 0.1mol/L dilute
hydrochloric acid for 30 minutes, then wash it in distilled water, then soak
it in the 3M KCL storage solution for 6 hours.
(b) For organic or lipid pollutant, clean the platinum surface with detergent,
then wash it in distilled water, then soak it in the 3M KCL storage solution
for 6 hours.
(c) For heavily polluted platinum surface on which there is oxidation film,
polish the platinum surface with toothpaste, then wash it in distilled water,
then soak it in the 3M KCL storage solution for 6 hours.
CONDUCTIVITY MEASUREMENT
6.1 Information regarding the Conductivity Electrode
6.1.1 Conductivity electrode
Model 2301T-S plastic conductivity electrode with constant K=1.0 and built-in
temperature sensor, can realize automatic temperature compensation. The
electrode housing is POM plastic which is corrosion resistant and impact
resistant. When dipping the conductivity electrode in solution, stir the
solution briefly to eliminate the air bubbles and improve response and
stability.
6.1.2 Conductivity electrode constant
The meter matches conductivity electrodes of three constants: K=0.1, K=1.0 and
K=10.0. Please refer to table-7 for measuring range. Set constant in parameter
setting P2.1 and refer to section 7.4
Table– 7 Electrode constant and measuring range
Range | <20 pS/cm | 0.5 pS/cm to 100 mS/cm | >100mS/cm |
---|---|---|---|
Conductivity electrode constant | K=0.1 cm-1 | K=1.0 cm-1 | K=10 cm-1 |
Standard solution | 84pS/cm | 84 pS/cm | 1413 pS/cm |
Electrode’s model | DJS-0.1T-S | 2301T-S | 2310T-S |
Note: When testing ultra-pure water with conductivity less than 1.0
μS/cm, a flow test should be conducted in a flow cell.
6.1.3 When connecting the electrode, please rotate slowly to identify the
location of the mount before plugging in. The nut on the connector should be
screwed on tightly. Once the conductivity electrode is connected, the meter
will automatically switch to conductivity mode (no need to switch manually).
Please do not pull the cable in case of poor contact. Make sure the connector
is clean and dry. See more information regarding the conductivity electrode’s
maintenance in section 6.6.
6.2 Information regarding Conductivity Calibration
6.2.1 Conductivity Standard Calibration Solutions
The meter uses conductivity standard solution of 84 μS/cm, 1413 μS/cm, 12.88
mS/cm and 111.8 mS/cm.
The meter can recognize the standard solution automatically, and can perform 1
to 4 points of calibration.
The calibration indication icons correspond to the four standard values as
shown in Table-8.
Table-8 Conductivity Calibration Solutions
Calibration Icons | Calibration Solutions |
---|---|
L |
84 μS/cm |
M |
1413 μS/cm |
H |
12.88 mS/cm |
H+ |
111.8 mS/cm |
6.2.2 How often to calibrate
(a) The meter has been calibrated before leaving the factory and can generally
be used right out of box.
(b) Normally perform calibration once per month.
(c) For high accuracy measurements or large temperature deviation from the
reference temperature
(25°C), perform calibration once per week.
(d) Use conductivity standard solution to check whether there is error.
Perform calibration if error is large.
(e) For new electrode or the meter has been set to factory default, perform
3-point or 4-point calibration. For general use, choose standard solutions
that are closer to the sample solution to perform 1-point or 2-point
calibration.
6.2.3 Reference Temperature
The factory default setting for reference temperature is 25℃. The reference
temperature can be set from 15℃ to 30℃. Users can set it up in parameter
setting P2.2 (see Section 7.4 for details).
6.2.4 Temperature compensation coefficient
The temperature compensation coefficient of the meter setting is 2.0%/°C.
However, the conductivity temperature coefficient is different from solutions
and concentration. Please refer to Table – 9 and the data collected during
testing. Do the setting in P2.3. (see section 7.4 for more).
Note: When the coefficient for the temperature compensation is set to
0.00 (no compensation), the measurement value will be based on the current
temperature.
Table – 9 Temperature compensation coefficient of special solutions
Solution | Temperature compensation coefficient |
---|---|
NaCl solution | 2.12%/°C |
5% NaOH solution | 1.72%/°C |
Dilute ammonia solution | 1.88%/°C |
10% hydrochloric acid solution | 1.32%/°C |
5% sulfuric acid solution | 0.96%/°C |
6.2.5 Precaution for calibration solution’s contamination
Conductivity standard solution has no buffer. Please avoid being contaminated
during usage. Before submerging the electrode in standard solution, please
rinse the electrode in distilled water and dry it with clean cloth or tissue
paper. Please do not use the same cup of conductivity standard solution
repeatedly, especially for standard solution of low concentration (84μS/cm).
The contaminated standard solution will affect the accuracy.
6.3 Conductivity Calibration
6.3.1. Rinse the electrode in distilled water, allow it to dry, wash with a
little of standard solution and dip it in standard solution. Stir the solution
briefly and allow it to stay in the solution until a stable reading is
reached.
6.3.2. Press key to enter the calibration mode.
The meter’s display will show blinking “CAL” at the top right, and
scanning and locking process of calibration solution at the bottom right. When
the meter locks1413 mS, stable icon will display on LCD. Press key to complete
calibration.
The meter will return to measuring mode and M
is displayed on bottom
left of the LCD screen. See Figure-11.
6.3.3. Notes:
(a) Pressing key before stable icon appeared on LCD screen, Error indication
icon Er 2 will be shown. See Table -11.
(b) Press key before confirmation to exit calibration mode (calibration will
not be completed).
6.3.4. For multi-point calibrations, please repeat the steps in 6.3.1 to 6.3.2
until all calibrations are finished.
The meter can be calibrated in the same calibration solution repeatedly until
the reading is stable and repeatable.
6.4 TDS & Conductivity
6.4.1. TDS and conductivity is linear related. The conversion factor is
0.40-1.00. Adjust the factor from parameter P2.6. The factory default setting
is 0.71 and please refer to section 7.4. The meter only needs to be calibrated
in Conductivity mode, then after calibration of conductivity, the meter can
switch from conductivity to TDS.
6.4.2. Adjust TDS conversion factor in parameter setting P2.7 according to the
data collected during testing and experience. Table – 10 lists some commonly
used Conductivity and TDS conversion factors. This is for your reference only.
Table – 10 Conductivity and TDS conversion factors
Conductivity of solution | TDS conversion factor |
---|---|
0-100 μS/cm | 0.6 |
100-1,000 μS/cm | 0.71 |
1-10 mS/cm | 0.81 |
10-100 mS/cm | 0.94 |
6.5 Sample test
6.5.1. Rinse conductivity electrode in distilled water, dry it, and dip it in
the sample solution. Stir the solution briefly and allow it to stay in the
sample solution until a stable reading is reached and icon appears on LCD,
then get the reading value, which is the conductivity value of the solution.
6.5.2. Press key to switch to TDS.
6.5.3. During the process of calibration and measurement, the meter has self-
diagnosis functions, indicating the relative information as below: Table – 11.
Table – 11 Self-diagnosis information of conductivity measurement mode
Display Icons | Contents | Checking |
---|---|---|
Wrong conductivity calibration solution or the meter recognition of | ||
calibration solution out of range. | 1.Check whether conductivity calibration |
solution is correct.
2.Check whether the meter connects the electrode well.
3.Check whether the electrode is damaged.
| Press key when measuring value is not stable during calibration.| Press key
after icon appears.
| During calibration, the measuring value being unstable for over 3 minutes.|
1.Shake the electrode to eliminate bubbles in electrode head.
2.Replace with new conductivity electrode.
6.5.4. Factory default setting
For factory default setting, please refer to parameter setting P2.7 (Section
7.4). With this function, all calibration data is deleted and the meter
restores to the theory value. Some functions restore to the original value
(refer to appendix -1). When calibration or measurement fails, please restore
the meter to factory default setting and then perform re-calibration or
measurement. Please note once set the factory default, all the data deleted
will be irretrievable.
6.6 Maintenance of the Conductivity Electrode
6.6.1. Always keep the conductivity electrode clean. Before taking a
measurement, rinse the electrode in distilled water. It is recommended to
rinse it again in the sample solution. When dipping the electrode in solution,
stir the solution briefly to eliminate air bubbles and allow it to stay until
a stable reading is reached. Conductivity electrodes are usually stored dry.
For conductivity electrodes that haven’t been used for a long time, users can
soak the electrode in 12.88 mS calibration solution for 5-10 minutes, or to
soak it in tap water for 1 to 2 hours. Rinse the electrode in distilled water
after measurement.
6.6.2. The sensing rod of Model 2301T-S conductivity electrode is coated with
platinum black to minimize electrode polarization and expand measuring range.
The platinum black coating adopted our special processing technology, which
improves the electrode performance and the firmness of the coating. If the
platinum black electrode is stained, gently clean the electrode with soft
brush in warm water containing detergent or alcohol.
PARAMETER SETTING
7.1 Main Menu
In the measurement mode, press key to enter in P1.0, then press or to switch
main menu:
P1.0→P2.0→P4.0. Please refer to Figure – 7.
P1.0: pH parameter setting menu,
P2.0: Conductivity parameter setting menu,
P4.0: Basic parameter setting menu.
7.2 Sub-Menu
7.2.1 In P1.0, press to enter the submenu P1.1 for pH setting, press and to
change submenu: P1.1→P1.2→ … →P1.5. See Figure-7 for details.
7.2.2 In P2.0, press to enter the submenu P2.1 for conductivity setting, press
and to change submenu: P2.1→P2.2→ … →P2.7. See Figure-7 for details.
7.2.3 In P4.0, press to enter the submenu P4.1 for basic parameter setting,
press and to change submenu: P4.1→P4.2→…→P4.8. See Figure-7 for details.
7.3 pH Setting Sub-Menu
| P1.1 — Select pH buffer series (USA—NIST)
- In measurement mode, long press to enter P1.0, press to enter P1.1
- Press , “USA” flashes; Press to choose USA→NIST; Press to confirm. USA—USA NIS—NIST
- Press to enter P1.2, or press to return to measurement mode.
---|---
| P1.2 — Return to factory default mode (No—Yes)
Press , “No” flashes, press to choose No→Yes; Press to confirm, the meter returns to measurement mode.
No— not return to factory default mode; Yes—return to factory default mode If not choosing Yes, Press to return to measurement mode.
7.4 Conductivity Setting Sub-Menu
| P2.1 — Select electrode’s constant(1.0—10.0—0.1 )
1. In P2.0, press to enter P2.1
2. Press , “1.0” flashes, press to choose 1.0→10.0→0.1; press to confirm
3. Press to enter P2.2, or press to return to measurement mode.
---|---
| P2.2—Select reference temperature (25.0℃—18.0℃—20.0℃)
1. Press , “25.0℃” flashes, press or to adjust temp. from 15℃ to 30℃;
press to confirm.
2. Press to enter P2.3, or press to return to measurement mode.
| P2.3—Adjust temp. compensation coefficient (0.00-9.99%)
1. Press , “2.00” flashes, press or to adjust from 0.00-9.99; press to
confirm.
2. Press to enter P2.4, or press to return to measurement mode.
| P2.4—Adjust TDS Factor (0.40-1.00)
1. Press , “0.71” flashes, press or to adjust from 0.40-1.00; press to
confirm.
2. Press to enter P2.5, or press to return to measurement mode.
| P2.5 — Return to factory default mode (No—Yes)
Press , “No” flashes, press to choose No→Yes; Press to confirm, the meter
returns to measurement mode.
No— not return to factory default mode; Yes—return to factory default mode If
not choosing Yes, press to return to measurement mode.
7.5 Basic Parameter Setting Sub-Menu
| P4.1 —(°C—°F)
1. Press , “°C” flashes, press , “°F” flashes; when parameter flashes, press
to confirm.
2. Press to enter P4.2, or press to return to measurement mode.
---|---
| P4.2—Select lasting time for backlight (1—2—3—On)
1. Press , “1” flashes, press to choose from 1→2→3→On; press to confirm
Select On to have the backlight always turned on; time unit is minute.
2. Press to enter P4.3, or press to return to measurement mode.
| P4.3 — Select auto power-off time (10—20—30—On)
1. Press , “20” flashes, press to choose from 20→30→On→10, press to confirm.
Select On to shut off the auto power-off function; time unit is minute.
2. Press to enter P4.4, or press to return to measurement mode.
| P4.4 — Set up Auto-Lock reading mode (Off—On)
1. Press , “Off” flashes, press to choose Off→On; press to confirm.
2. Press to return to measurement mode.
COMPLETE KIT
| Content| Quantity| PH400| EC400| PC400
---|---|---|---|---|---
1| PH400 Portable pH Meter| 1| √| |
2| EC400 Portable Conductivity Meter| 1| | √|
3| PC400 Portable pH/Conductivity Meter| 1| | | √
4| 201T-S Plastic 3-in-1 Combination pH Electrode| 1| √| | √
5| 2301T-S Plastic Combination Conductivity Electrode| 1| | | √
6| pH Standard Buffer (4.00 pH,7.00 pH,10.01pH/50m1)| 1 for each| √| | √
7| Conductivity Standard Solutions (84pS,1413pS,12.88mS/50m1)| 1 for each| |
√| √
8| Carrying Case| 1| √| √|
9| Instruction Manual| 1| √| √| √
APPENDIX 1: TABLE OF PARAMETER SETTING AND FACTORY DEFAULT SETTING
Mode| Symbo| Parameter| Abbreviation| Content|
Factory Default
---|---|---|---|---|---
P1.0
pH| P1.1| Select Buffer Series| | USA—NIST| —
P1.2| Restore to factory default settings| | No—Yes| No
P2.0
Conductivity| P2.1| Select electrode’s constant| | 1.0-10.0-0.1| 1.0
P2.2| Select reference temperature| | 15°C — 30°C| 25°C
P2.3| Adjust temperature compensation coefficient| | 0.00 9.99| 2.00
P2.4| Adjust TDS factor| | 0.40 – 1.00| 0.71
P2.5| Restore to factory default setting| | No—Yes| No
P4.1| Select temperature unit| /| °C – °F| —
P4.2| Select lasting time for backlight| | 1-2-3—On| —
P4.3| Select time for auto power-off| | 10-20-30—On| —
P4.4| Set up auto-lock reading mode| /| Off—On| —
APPENDIX 2: TABLE OF SELF-DIAGNOSIS
Symbol | Self-Diagnosis Information | pH | Conductivity |
---|---|---|---|
Wrong conductivity calibration solution or the meter recognition of | |||
calibration solution out of range. | √ | √ | |
Press key when measuring value is not stable duringcalibration. | √ | √ | |
During calibration, the measuring value is not stable for ?.. 3 min. | √ | √ | |
pH electrode zero electric potential out of range (<-60mV or >60mV) | √ | ||
pH electrode slope out of range (<85% or >110%) | √ |
APPENDIX 3: ICONS AND ABBREVIATION
Mode | Symbol | Abbreviation | Content |
---|
P1.0
pH| P1.1| | Standard buffers
P1.2| /| Factory default setting
P2.0
Conductivity| P2.1| | Electrode constant
P2.2| | Reference temperature
P2 3| | Temperature compensation coefficient
P2.4| | TDS Factor
P2.5| | Factory default setting
P4.0
Basic Parameters| P4.1| /|
P4.2| | backlight
P4.3| | Auto power-off
P4.4| /|
Others| | | USA Series
| NIST Series
| Off
| On
| No
| Yes
WARRANTY
We warrant this instrument to be free from defects in material and workmanship
and agree to repair or replace free of charge, at option of APERA INSTRUMENTS,
any malfunctioned or damaged product attributable to responsibility of APERA
INSTRUMENTS for a period of TWO YEARS (SIX MONTHS for the probe) from the
delivery.
This limited warranty does not cover any damages due to: Transportation,
storage, improper use, failure to follow the product instructions or to
perform any preventive maintenance, modifications, combination or use with any
products, materials, processes, systems or other matter not provided or
authorized in writing by us, unauthorized repair, normal wear and tear, or
external causes such as accidents, abuse, or other actions or events beyond
our reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Address: Wilhelm-Muthmann-Str.18, 42329 Wuppertal, Germany
Email: info@aperainst.de |
www.aperainst.de |Tel.: +49 202 51988998