CooperVision Biofinity Multifocal User Guide

June 4, 2024
CooperVision

Biofinity® multifocal lens fitting guidelines
User Guide

A simplified fitting philosophy

Our new fitting process is based on eye care professionals’ real-world experiences.

  • Fitting lower ADD powers are now simpler than ever, by using the same D lens design for both eyes
  • Fitting higher ADD powers continues to be flexible, giving you more options for exceptional vision performance

Initial visit
Step 1 Start with new refraction and verification of eye dominance (fogging technique).
Step 2 Select the distance prescription based on spherical equivalent corrected for the vertex distance. Choose D or N lens design based on needed ADD power:

ADD Dominant eye Non-Dominant eye
+1.00 D D
+1.50 D D
+2.00 D D
+2.50 D D

Visual acuity expectations when using the D and N lens combination

Lens Distance Near
Binocularly 20/20 20/20
D Lens 20/20 20/40 better
N Lens 20/40 better 20/20

Step 3 Although the lens will equilibrate quickly, allow patients to adapt to lenses for a minimum of 15 minutes before assessing vision. If binocular vision is unacceptable, perform a monocular over-refraction, using hand-held trial lenses, to determine which eye needs improvement.
To improve distance vision add +/-0.25D (up to +/-0.50D) to the eye that needs improvement.
To improve near vision add +/-0.25D (up to +/-0.50D) to the eye that needs improvement.
Clinical Tips Prescribe maximum plus power for distance vision (Do not over minus)
Choose the lower ADD power when possible; not necessary to overprescribe the ADD power
Test patient’s near function vision with their cell phone
Check visual acuity with room lights on

A unique multifocal lens for unique eyes

Balanced Progressive” Technology

  • Optimized for exceptional vision at all distances: near, intermediate, and far
  • Allows for personalized fitting for each wearer and each eye
  • The streamlined fitting process helps ensure success for presbyopic patients

Follow-up visits one week later
If the patient requires further enhancement to distance or near visual acuity.
Step 1 Evaluate binocular visual acuity.
Step 2 Check monocular visual acuity.
Step 3 Perform over-refraction using hand-held trial lenses (avoid using a phoropter).
FIRST OPTION: To improve either distance or near vision, modify distance vision by +/- 0.25D in the eye that needs improvement.
SECOND OPTION: To improve near vision add +0.50D to the ADD power of the eye that needs improvement.

Product specifications

Biofinity® multifocal

Base Curve| 8.6 mm
Diameter| 14.0mm
Sphere Power| +6.00D to -10.00D (0.50D after -6.00D)
ADD Power| +1.00, +1.50, +2.00, +2.50
Lens Design| D Lens. N Lens
Material| con’ ‘Icon A
Water content| 48%
Dk| 128
Wearing schedule| | Daily Wear or Extended Wear up  6 nights/7 days

800 341 2020
For additional fitting tips, tutorials, and more information on Biofinity multifocal, visit
www.coopervision.com/biofinity-multifocal
The eye care professional retains the independent clinical judgment on how to fit and prescribe lenses.
432015 CooperVision. Inc. 1131 04/15 XLITERAKIT210

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

CooperVision User Manuals

Related Manuals