AMETEK Echotel 940 Ultrasonic Level Switches User Manual

June 3, 2024
AMETEK

AMETEK Echotel 940 Ultrasonic Level Switches

UNPACKING

Unpack the instrument carefully. Make sure all components have been removed from the foam protection. Inspect all components for damage. Report any concealed damage to the carrier within 24 hours. Check the contents of the carton/ crates against the packing slip and report any discrepancies to Magnetrol. Check the nameplate model number to be sure it agrees with the packing slip and purchase order.
Check and record the serial number for future reference when ordering parts.

These units are in compliance with:

  1. The EMC directive 2014/30/EU.

MOUNTING

AMETEK Echotel 940 Ultrasonic Level Switches-2

WIRING – RELAY VERSION: ECHOTEL 940

OUTPUT

Level Relay Wiring
**** De-energized
**** Energized

IMPORTANT: When power is lost or gap is dry; the relay de-energizes.

WIRING – CURRENT SHIFT VERSION: ECHOTEL 941

OUTPUT

Level Current
**** 8 mA (± 1 mA)
**** 16 mA (± 1 mA)

SPECIFICATIONS

PERFORMANCE SPECIFICATIONS

Description Specification
Response time 0,5 s typical
Repeatability ± 2 mm (0.078″)
Ambient Temperature -40 °C up to +85 °C (-40 °F up to +185 °F)
Humidity 0-99 %, non condensing
Electromagnetic Compatibility Meets CE requirements (EN 61326: 1997 + A1 +

A2)

PHYSICAL/FUNCTIONAL SPECIFICATIONS

Description Specification
Input voltage 12 – 35 V DC
Power consumption < 1 W
Signal output Current Shift

gap
Relay| SPDT:    1 A @ 30 V DC
Potted cable| 45 cm (18″) flying leads of 18 AWG wires
Net Weight| 510 g (unit with 3/4″ NPT transducer and shortest length)

TRANSDUCER SPECIFICATIONS

Description Specification
Material 316/316L SST (1.4401/1.4404)
Mounting Threaded (NPT/BSP)
Actuation length From 3 cm up to 250 cm (1.2″ up to 98.4″)
Operating process temp. -40 °C up to +85 °C (-40 °F up to +185 °F)
Max pressure 103 bar @ +85 °C (1500 psi @ +185 °F) – transducer length > min.

length 138 bar @ +85 °C (2000 psi @ +185 °F) – transducer length = min. length Flanged models are downrated to the design pressure of the selected flange

TEMPERATURE/PRESSURE RATING

MODEL IDENTIFICATION

A complete measuring system consists of:

  1. One code for electronics and transducer
  2. OPTION: Standard cable length is 45 cm (18″). Specify cable length separately when a longer cable is required.
  3. OPTION: 052-7012-002: code for Echotel 941 cast aluminium housing with terminal block, 3/4″ NPT cable entry, IP 65 052-7012-003: code for Echotel 940 cast aluminium housing with terminal block, 3/4″ NPT cable entry, IP 65

BASIC MODEL NUMBER

AMETEK Echotel 940 Ultrasonic Level Switches-11

DIMENSIONS IN mm (inches)

AMETEK Echotel 940 Ultrasonic Level Switches-12

IMPORTANT

SERVICE POLICY

Owners of Magnetrol products may request the return of a control; or, any part of a control for complete rebuilding or replacement. They will be rebuilt or replaced promptly. Magnetrol International will repair or replace the control, at no cost to the purchaser, (or owner) other than transportation cost if:
a. Returned within the warranty period; and,
b. The factory inspection finds the cause of the malfunction to be defective material or workmanship.
If the trouble is the result of conditions beyond our control; or, is NOT covered by the warranty, there will be charges for labour and the parts required to rebuild or replace the equipment.

In some cases, it may be expedient to ship replacement parts; or, in extreme cases a complete new control, to replace the original equipment before it is returned. If this is desired, notify the factory of both the model and serial numbers of the control to be replaced. In such cases, credit for the materials returned, will be determined on the basis of the applicability of our warranty.
No claims for misapplication, labour, direct or consequential damage will be allowed.

RETURNED MATERIAL PROCEDURE
So that we may efficiently process any materials that are returned, it is essential that a “Return Material Authorisation” (RMA) form will be obtained from the factory. It is mandatory that this form will be attached to each material returned. This form is available through Magnetrol’s local representative or by contacting the factory. Please supply the following information:

  1. Purchaser Name
  2. Description of Material
  3. Serial Number and Ref Number
  4. Desired Action
  5. Reason for Return
  6. Process details

Any unit that was used in a process must be properly cleaned in accordance with the proper health and safety standards applicable by the owner, before it is returned to the factory.
A material Safety Data Sheet (MSDS) must be attached at the outside of the transport crate or box.
All shipments returned to the factory must be by prepaid transportation. Magnetrol will not accept collect shipments.
All replacements will be shipped Ex Works.

European Headquarters & Manufacturing Facility
Heikensstraat 6
9240 Zele, Belgium
Tel: +32-(0)52-45.11.11
e-mail: info@magnetrol.be
www.magnetrol.com

BULLETIN N°: BE 51-647.7
EFFECTIVE: AUGUST 2021
UNDER RESERVE OF MODIFICATIONS SUPERSEDES: August 2017

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals