APERA PH700 Benchtop pH Meter Instruction Manual
- June 3, 2024
- APERA
Table of Contents
APERA PH700 Benchtop pH Meter
Brief Introduction
Thank you for purchasing Apera Instruments PH700 Benchtop pH meter.
This instrument is an outstanding combination of advanced electronics design
and sensor technology. It’s the most economical choice of a reliable lab-grade
benchtop pH meter. Please read this manual carefully to properly use and
maintain the meter. Aprea Instruments reserves the right to update the
contents of this manual without giving prior notices.
Main Features
- The meter has a built-in microprocessor chip, which enables intelligent functions such as automatic calibration, automatic temperature compensation, data storage, parameter setup, max/min reading display, etc.
- The advanced automatic calibration mode has calibration guide and self-diagnosis functions, automatically recognizing standard buffer solutions (USA or NIST series).
- The advanced digital processing technology intelligently improves the meter’s response time and measuring accuracy with stable reading display mode available.
- The kit comes with a 3-in-1 combination pH Electrode, measuring pH and temperature simultaneously.
- IP54 dustproof and spill proof. Connectors are protected by silicone seal caps, effectively ensuring the reliability and service life of the meter under harsh environment.
- Detachable electrode holder and 3 bottles of pH buffer solutions come with the kit for convenient use.
Technical Specifications
pH | Range | 0 to 14.00 pH |
---|---|---|
Resolution | 0.01/0.1 pH | |
Accuracy | ±0.01 pH ±1 digit | |
Temperature Compensation | 0 to 100 ˚C (Auto or Manual) | |
Calibration | 1 to 3 points automatic calibration |
mV
| Range| ±1999 mV
Resolution| 1 mV
Accuracy| ±0.1% F.S ±1 digit
Temp.| Range| 0 to 100˚C; 32 to 212°F
Resolution| 0.1˚C; 0.1/1°F
Accuracy| ±0.5˚C ±1 digit
Others| Data Storage| 50 groups
Storage Content| Numberings, Measurement, Unit, Temperature, Temperature
Compensation Status
Power| DC9V/300mA
IP Ratings| IP54 dustproof and spill proof
Size and Dimension| (240235103) mm/1kg
What’s in the Box?
| Content| Quantity
---|---|---
3.1| PH700 Benchtop pH Meter| 1
3.2| 201T-F Plastic 3-in-1 Combination pH Electrode| 1
3.3| pH standard buffer solutions (4.00/7.00/10.01 pH)| 1 for each 50ml
3.4| 9V Power Adapter| 1
3.5| Electrode Holder| 1 set
3.6| Instruction Manual| 1
Instrument Description
LCD Display
-
Measurement value
-
Calibration icons and numberings
-
Measurement unit
-
Temperature and reminder icons
-
The pH unit displayed during calibration
-
The pH value displayed during calibration, numberings for storage and recall, and reminder icons
-
Icons for data storage and recall — M+: store measuring data; RM: Recall measuring data
-
Self-Diagnosis icons and numberings
-
Temperature compensation mode icons:
ATC: Auto Temperature Compensation;
MTC: Manual Temperature Compensation -
Completed calibration icons
-
Stable reading icon
-
Icons for maximum and minimum values
Keypad
Keypad operations
Short press —– <1.5 seconds; Long press —– >1.5 seconds.
For keypad operation, please see Table-1.
Table – 1 Keypad operations and descriptions
Keypad | Operation | Functions |
---|---|---|
Short Press |
- Power on/off
| Short Press|
- Select measuring mode: pH→ mV.
Long Press|
- Enter parameter setting:P1→P2→…P4.
| Short Press|
- Enter Calibration mode
- Cancel any operation and go back to measurement mode
| Short Press|
- In calibration mode, press to conduct calibration
- In parameter setting, press to confirm selection
| Short Press|
- Store measuring data
Long Press|
- Recall stored measuring data
| Short Press or Long Press|
- In manual temperature compensation mode, short press to change temperature, long press to change quickly
- In parameter setting mode, short press to change parameter and settings.
- In recall mode, short press to change numberings, long press to change quickly.
| Short Press|
- Buttons for max/min values (referee to section 3.6)
Connectors
Symbol| Connector Name| Connector Type
REF| Socket for reference electrode| Φ2 banana type
pH/mV| Socket for pH and ORP electrode| BNC
TEMP| Socket for temperature sensor| RCA
DC9V| DC9V power socket| Φ2.5 direct type
Stable reading display mode
When the measuring value is stable, smiley face icon appears on LCD, see
Figure-3. Please do not get the reading value or make calibration until the
icon appears and stays.
Data storage, recall, and deletion
Storage
When the measurement is stable, press key, the meter displays M+ icon and
storage numberings on LCD, and stores the measuring data. In Figure-4, the
meter stores the first group of the measuring data. Press again to keep
storing.
Recall stored data
In the measurement mode, long press key to recall the last stored Measuring
data. The meter displays RM icon and storage numberings.
In Figure-5, the meter is recalling the 28th stored measuring data.
Keep pressing and to recall the stored measuring data successively. Press and
hold and to quickly recall the stored measuring data. Short press to go back
to measurement mode.
Clear stored data
The meter can store 50 groups of data. Once the storage is full, if user
presses again, the FUL icon will be displayed on LCD (see Figure-6).
The storage needs to be cleared in parameter setting P4 so as to keep storing
data. The numberings in M+ indicates the total number of stored data; The
numberings in RM indicates the number of stored data in current mode (ph/mV).
For example, if 20 groups of data is stored in pH mode, and 10 groups Figure –
6 in mV mode, then the numberings for M+ will be “M+30” in both pH and mV
mode; the numberings for RM in pH mode would be “RM20”, in mV mode “RM10”.
Max/Min Function
Use this function to display the maximum or minimum records out of groups of measuring data. In measurement mode, press , LCD will display flashing icons of “MAX/MIN”, indicating the meter has entered the mode of recording maximum and minimum data. After testing, press again, LCD will alternately display the maximum and minimum value in the past measurements. Press to go back to regular measurement mode.
pH Measurement
Preparation
Install the electrode holder
The electrode holder is composed of a metal stand base, a supporting pole, and an electrode clip. First, install the metal stand base to the meter (screw it on). Please note that the metal base can be installed on either side of the meter. Then install the supporting bar on the metal stand base, and then install the electrode clip on the supporting bar.
Connect the power adapter
Plug the power adapter tightly into the DC9V socket. The power’s voltage
should meet the requirements listed on the power adapter.
Things needed in addition to what’s included in the box
A clean cup, distilled water (8-16oz), and tissue papers for rinsing and
drying the probe.
pH Electrode Information
pH Electrode
The meter comes with the 201T-F 3-in-1 combination electrode with a built-in
temperature sensor, which enables the automatic temperature compensation. This
electrode is only suitable for regular water solutions’ pH testing. Please
refer to Section 7 for ideal pH electrodes to use for other specific
applications. The electrode’s housing adopts polycarbonate materials, which is
resistant to shocks and corrosion. A 3M KCL storage bottle comes with the
electrode, which is for storing the electrode when not in use to keep the
sensitivity of the probe.
Technical Specifications of the 201T-F pH Electrode
- Measurement Range: 0 – 14 pH, 0 – 80˚C (32 – 176˚F)
- Junction: Single Ceramic
- Reference Electrode: Ag/AgCl
- Connector: BNC/RCA
- Size: ø12*160 mm
- Temperature unit: 30K Thermistor
Electrode Connectors
The pH electrode has two connectors: the BNC connector connects the pH
probe; the RCA connector connects the temperature sensor. Plug these two
connectors into ‘pH/mV” and “TEMP” sockets. Please note not to pull the cables
in case of poor contact. Please keep the connectors clean and dry. Refer to
section 4.7 regarding how to properly maintain the pH electrode.
Use the Electrode
Screw off the KCL storage bottle, and put it aside (do not dump or spill the
KCL solution). Prepare a cup of distilled water, and rinse the electrode in it
for a few seconds. Gently shake the probe to remove excess water, and dry it
with clean tissue paper (DO NOT rub or wipe the probe, just use paper to dap
off excess water). Gently stir the electrode for a few seconds after it’s
dipped into the test solution and then let it stand still. Wait for the stable
measurement (a smiley face appears and stays) and then record the readings.
When test is finished, place the electrode in the storage bottle and tighten
the cap in order to keep the sensitivity of the pH sensor.
Information regarding Calibration
pH buffer solutions
The meter has 2 series of standard buffer solutions: USA and NIST (see
Table-2). They can be selected in parameter P1 (see 4.6.1).
Table-2
Calibration icon | pH standard buffer series |
---|---|
USA | NIST |
3-Point calibration | |
7.00 pH | 6.86 pH |
10.01 pH | 9.18 pH |
3-Point Calibration
The meter can adopt 1 to 3 points of calibration. The 1st point must be using
7.00 pH (or 6.86 pH) buffer solution, and then choose other buffers to do 2nd
point and 3rd point. Please refer to Table-3 for details. In the process of
calibration, the slope of acidity range and alkalinity range will be
displayed, which indicates the health of your probe. The closer the slope is
to 100%, the better the condition of the probe is.
Table-3 Calibration Mode
| USA| NIS| Calibration icon| When to adopt
---|---|---|---|---
1-Point Calibration| 7.00 pH| 6.86 pH| | accuracy≤ ±0.1 pH
2-Point Calibration| 7.00 pH and 4.00/1.68 pH| 6.86 pH and 4.01/1.68 pH| |
Measuring range: 0 to 7.00 pH
7.00 pH and 10.01 pH| 6.86 pH and 9.18 pH| | Measuring range: >7.00 pH
3-Point Calibration| 7.00pH, 4.00/ 1.68 pH and 10.01 pH| 6.86pH, 4.01/ 1.68pH,
9.18 pH| | Wide measuring range
How often to calibrate
The frequency that you need to calibrate your meter depends on the tested
samples, condition of electrodes, and the requirement of the accuracy. For
High-Accuracy measurements (≤ ±0.02pH), the meter should be calibrated before
test every time; For ordinary-accuracy measurements (≥±0.1pH), once
calibrated, the meter can be used for about a week or longer. In the following
cases, the meter must be re-calibrated:
- The electrode hasn’t been used for a long time or a new electrode is connected.
- After measuring strong acid (pH<2) or strong base (pH>12) solutions.
- After measuring fluoride-containing solution and strong organic solution
- There is a significant difference between the temperature of the test sample and the temperature of the buffer solution that is used in the last calibration.
pH Meter Calibration
-
Press to enter calibration mode. CAL1 icon will flash in the upper
-
right corner of the LCD. 7.00 pH will flash in the lower right corner of the LCD, reminding you to use pH 7.00 buffer to conduct 1st point of calibration.
-
Use distilled water to rinse off electrode and then dry it (with a tissue). Dip it into pH 7.00 buffer solution, stir gently and let it stand still and wait for the reading to become stable. In the lower right corner of LCD, the process of auto recognizing the buffer solution will be displayed. Pressing
before the buffer is recognized will generate Er2 (please refer to table 5). -
When the meter locks 7.00 pH in the lower right corner, stable icon displays on LCD. Then press key to calibrate the meter.
End icon appears after calibration is done.
The 1st point calibration is finished. In the meanwhile, CAL2 will flash at the upper right corner, and 4.00 pH & 10.01 pH will flash alternately at the bottom right, indicating using pH4.00 or pH10.01 buffer solution to make the 2nd point calibration. -
Take out the pH electrode, rinse it in distilled water, dry it, and dip it into pH 4.00 buffer solution. Stir the solution gently and let stand still in
the buffer solution until a stable reading is reached. The meter’s display will show the recognition process of calibration buffer solution at the lower right corner. When the meter recognizes 4.00 pH, stable icon displays on LCD. Press key to calibrate the meter. End icon and electrode slope of acidity range will be displayed after calibration is done.
In the meanwhile, CL3 will flash at the upper right corner of the LCD, and 10.01 pH will flash at the lower right, indicating using pH10.01 buffer
solution to make the 3rd point calibration. -
perform one-point, two-point and three-point calibration. Calibration guide icons appear on LCD. During the calibration process, press key to exit from the calibration mode. The meter can perform one-point, two-point and three-point calibration. Calibration guide icons appear on LCD.
Measuring pH
Screw off the KCL storage bottle, and put it aside (do not dump or spill the
KCL solution). Prepare a cup of distilled water, and rinse the electrode in it
for a few seconds. Gently shake the probe to remove excess water, and dry it
with clean tissue paper or Kimwipes (DO NOT rub or wipe the probe, just use
paper to dap off excess water). Gently stir the electrode for a few seconds
after it’s dipped into the test solution and then let it stand still. Wait for
the stable measurement (a smiley face icon appears and stays) and then record
the readings. Please refer to Figure-8 for calibration and measurement process
of the pH meter.
Figure-8 Calibration and measurement process of pH meter
Notes:
- This default 201T-F electrode will NOT give accurate and stable pH readings when testing purified water such as distilled or deionized water. This because those water do not have enough ions present for the electrode to function properly. To measure distilled or deionized water’s pH, users need to use a specialized pH electrode with the meter such as LabSenTM 803 Pure Water pH Electrode. Contact us at 1-614-285-3080 for more details. When testing purified water like spring water or drinking water, it will take longer for the readings to get stabilized (typically 3-5 minutes) because there is very few ions left to be detected by the sensor in those purified water.
- Purified water such as distilled water and deionized water are recommended for only rinsing the probe for the best result.
- The 201T-F pH Electrode is NOT suitable for testing high-temperature samples (>176˚F). Testing high-temperature samples could cause permanent damage to the electrode. A specialized electrode such as Lab Sen TM 213 (up to 225˚F) should be used in this situation.
- The electrode housing is polycarbonate. When using cleaning solutions, take cautions on carbon tetrachloride, trichlorethylene, tetrahydrofuran, acetone, etc., which will dissolve the housing and invalidate the electrode.
- According to the pH isothermal measurement principle, the closer the test sample’s temperature is to the calibration solution’s, the higher the accuracy of the measurement. This principle is recommended to follow when conducting tests for the best result. Example: If users were to test samples at 150˚F, we recommend warming up the calibration solutions to the same temperature before performing calibrations in order to get the most accurate readings.
- The instrument has a function to return to factory default setting, which can be set up in P5 (refer to section 5.6). Returning to factory default setting is to restore the meter to theoretical value (zero potential pH is 7.00, slope is 100%), and set all the parameters to default settings. When the meter’s calibration or measurement is performing abnormally, users can use this function to let the meter return to factory default mode, and then conduct calibration and test again. Please note that this function is irreversible once used.
- Self-Diagnosis
In the process of calibration and measurement, the meter has a self-diagnosis function. Please see table 5.
Table – 5 Self-diagnosis InformationDisplay Icons| Contents| How to fix
---|---|---
| Wrong pH buffer solution or the buffer solution is out of range.|
1. Check whether pH buffer solution is correct.
2. Check whether the meter connects the electrode properly.
3. Check whether the electrode is damaged.
4. Make sure there is no bubble in the electrode’s bulb.
| pH electrode zero electric potential out of range (<-60mV or >60mV)|
1. Make sure there is no bubble in the electrode’s bulb.
2. Check whether the electrode is damaged.
3. Check if pH buffer solutions are correct.
4. The pH Electrode is aged. Replace a new one.
| pH electrode slope is out of range (<85% or >110%)
| Press key when reading is not stable during calibration.| Press key when
icon appears and stays on the screen.
| During calibration, the measuring value is not stable for ≥3min.|
1. Make sure there is no bubble in the electrode’s bulb.
2. Check whether the electrode is damaged.
3. The pH Electrode is aged. Replace a new one.
- If you find any air bubble in the glass bulb of the pH sensor, simply shake the probe for a few times to remove it. The existence of an air bubble in the glass bulb will significantly decrease the stableness of measurement.
Parameter Setting
Symbol | Parameter setting content | Parameter | Factory default setting |
---|---|---|---|
P1 | Select pH buffer series | USA – NIST | USA |
P2 | Select resolution | 0.1/0.01 pH | 0.01 pH |
P3 | Select temperature unit | ℃ – ℉ | ℃ |
P4 | Clear stored data | No-Yes | No |
P5 | Return to factory default | No-Yes | No |
How to do the parameter setting:
In measurement mode, long press to enter parameter setting P1. Then press to
change menu from P1 to P2…P5. For details, please see Table-7.
Table-7 pH Parameter Setting
pH Electrode’s Maintenance
Daily maintenance
The soaking solution contained in the supplied protective bottle is used to
maintain activation in the glass bulb and junction. Loosen the cap, take out
the electrode and rinse in distilled or water before taking a measurement.
Insert the electrode and tighten the cap after measurements to prevent the
solution from leaking. If the soaking solution is cloudy or moldy, replace the
3M KCL solution.
- We DO NOT recommend using other brand’s soaking solution, since different chemicals can be used and potential damage can be caused to the electrode)
- The electrode should NOT be soaked and stored in purified or distilled water, protein solution, acid fluoride solution, or organic lipids. Distilled water or purified water is only recommended for rinsing electrodes.
Calibration buffer solution
For calibration accuracy, the pH of the standard buffer solution must be
reliable. The buffer solution should be refreshed often, especially after
heavy use. We recommend 10-15 times of use before replacing the pH buffers.
Protect glass bulb
The sensitive glass bulb at the front of the combination electrode should not
come in contact with hard surfaces. Scratches or cracks on the electrode will
cause inaccurate readings. Before and after each measurement, the electrode
should be rinsed with distilled water or purified water. If a sample sticks to
the electrode or it’s contaminated, the electrode should be thoroughly cleaned
using a soft brush with soap water and then rinsed with distilled water. After
that, soak it in the KCL solution again for at least 6 hours.
Renew glass bulb
Electrodes that have been used over a long period of time will become aged.
Submerge the electrode in 0.1mol/L hydrochloric acid 24 hours, then rinse the
electrode in distilled water, then submerge it in KCL soaking solution for 24
hours.
The method to prepare 0.1mol/L hydrochloric acid: dilute 9mL hydrochloric acid
in distilled water to 1000mL. For serious passivation, submerge the bulb in 4%
HF (hydrofluoric acid) for 3-5 seconds, and wash it in distilled water, then
submerge it in the soaking solution for 24 hours to renew it.
Clean contaminated glass bulb and junction (please refer to Table-8)
Table-8 Clean contaminated glass bulb and junction
Contamination | Cleaning Solutions |
---|---|
Inorganic metal oxide | Dilute acid less than 1mol/L |
Organic lipid | Dilute detergent (weak alkaline) |
Resin macromolecule | Dilute alcohol, acetone, ether |
Proteinic haematocyte sediment | Acidic enzymatic solution (saccharate yeast |
tablets)
Paints| Dilute bleacher, peroxide
mV measurement
ORP measurement
Press key, and switch the meter to mV measurement mode. Connect ORP electrode
(the 301Pt-C combination ORP electrode is sold separately) and dip it in
sample solution, stir the solution briefly and allow it to stay in the
solution until icon appears and get the reading.
ORP means Oxidation Reduction Potential. The unit is mV.
Notes on ORP measurement
ORP measurement does not require calibration.
When the user is not sure about ORP electrode quality or measuring value, use
ORP standard solution to test mV value and see whether ORP electrode or meter
works properly. Table-9 is the data of standard ORP solution for 222 mV.
Table-9
°C | 10 | 15 | 20 | 25 | 30 | 35 | 38 | 40 |
---|---|---|---|---|---|---|---|---|
mV | 242 | 235 | 227 | 222 | 215 | 209 | 205 | 201 |
Clean and activate ORP electrode
After the electrode has been used over a long period of time, the platinum
surface will get polluted which causes inaccurate measurement and slow
response. Please refer to the following methods to clean and activate ORP
electrode:
- For inorganic pollutant, submerge the electrode in 0.1mol/L dilute hydrochloric acid for 30 minutes, then wash it in distilled water, then submerge it in the soaking solution for 6 hours.
- For organic or lipid pollutant, clean the platinum surface with detergent, then wash it in distilled water, then submerge it in the soaking solution for 6 hours.
- For heavily polluted platinum surface on which there is oxidation film, polish the platinum surface with toothpaste, then wash it in distilled water, then submerge it in the soaking solution for 6 hours.
Recommended pH Electrodes for Specific Applications
Application | Ideal Apera pH Electrodes to Use with PH700 Meter |
---|---|
Regular water solutions | 201T-F, LabSen 213 |
Beverage, beer, or wine analysis | LabSen 213 |
Cosmetics | LabSen 851-1, (MP500 temp. probe required) |
Dairy products (milk, cream, yogurt, mayo, etc.) | LabSen 823 |
High-Temperature liquid | LabSen 213 |
Low-temperature liquid | LabSen881 (MP500 temp. probe required) |
Meat | LabSen 763 |
Micro sample testing | LabSen 241-6, LabSen 241-3 (MP500 temp. probe required) |
Purified Water (Low ion concentration samples) | LabSen 803, LabSen 813 |
Soil | LabSen 553 |
Solid or semi-solid samples (cheese, rice, fruit, etc.) | LabSen 753 |
Strong acid samples | LabSen 831 (MP500 temp. probe required) |
Strong alkalined samples | LabSen 841 (MP500 temp. probe required) |
Surface test (skin, paper, carpet, etc.) | LabSen 371 (MP500 temp. probe |
required)
Titration| LabSen 223
TRIS buffer solutions| LabSen 213, LabSen 223
Viscous liquid samples| LabSen 223, LabSen851-1
Wastewater or emulsion| LabSen 333
Warranty
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS(Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS(Europe) GmbH for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery.
This limited warranty does not cover any damages due to:
Transportation, storage, improper use, failure to follow the product
instructions or to perform any preventive maintenance, modifications,
combination or use with any products, materials, processes, systems or other
matter not provided or authorized in writing by us, unauthorized repair,
normal wear and tear, or external causes such as accidents, abuse, or other
actions or events beyond our reasonable control.