Monmouth T800 Circulaire Filtration Fume Cabinet Instruction Manual
- September 20, 2024
- Monmouth
Table of Contents
T800 Circulaire Filtration Fume Cabinet
“`html
Specifications
Description: Filtration Fume Cabinet
Compliance: Complies with international
standards including BS7989:2001 for filtration fume cupboards
Product Usage Instructions
Installation
-
The cabinet should be sited in a draught-free position.
-
The cabinet is recirculating and requires no connection to
ductwork. -
Lift up the front cover and remove the inner filter access
cover using the key.
General Operation
Control System
The cabinet is controlled and monitored by a microprocessor with
an LCD touch screen. The touch screen provides the operator with
general information about the cabinet and displays the current face
velocity and filter condition. Control of light level, fan speed,
and other operating settings are all accessed via a menu
system.
The control system is factory set to maintain the following
parameters:
- Normal face velocity: 0.50m/sec.
- Low fan speed: 0.35m/sec.
- Low airflow alarm: 0.30m/sec.
These settings can be changed by a Monmouth engineer. Please
contact Monmouth Scientific for further advice.
Start-Up
When first turned on, the screen provides information about the
filters fitted to the cabinet. Ensure correct filters are fitted to
suit the work being carried out. Basic filter information can be
displayed by pressing the types key. Contact Monmouth Scientific
for further advice on filter types.
If procedures inside the cabinet are always the same, this
screen can be bypassed in the Supervisor Settings.
Control Buttons
-
Light On / Off
-
Reduce Brightness
-
Increase Brightness
-
Fan On / Off
-
Low Fan Speed – Reduces fan speed to 0.35m/sec. for use when
weighing -
Settings – Access to settings menu
-
Information – Cabinet and filter information
-
Standby – Puts cabinet into standby
FAQ
Q: What should I do if the cabinet is not maintaining the
specified parameters?
A: If the cabinet is not maintaining the specified parameters,
contact Monmouth Scientific for further assistance and possible
adjustments.
Q: Can I change the control system settings on my own?
A: The control system settings should be changed by a qualified
Monmouth engineer. Contact Monmouth Scientific for guidance on
adjusting the settings.
“`
FILTRATION FUME CABINET
MODELS T800 T1800 OPERATING AND MAINTENANCE MANUAL
ISSUE A 6TH December 2013
2
CONTENTS SECTION 1 ……………………………………………………………………. 5
DESCRIPTION OF THE CABINET ………………………………………………………………. 5
SECTION 2 ……………………………………………………………………. 6
INSTALLATION ………………………………………………………………………………………… 6 TESTING / COMMISSIONING
…………………………………………………………………………… 6
SECTION 3 ……………………………………………………………………. 7
GENERAL OPERATION ……………………………………………………………………………. 7 CONTROL SYSTEM
………………………………………………………………………………………… 8 START-UP
……………………………………………………………………………………………………… 8 FACE VELOCITY
…………………………………………………………………………………………….. 9 CARBON FILTER CONDITION
…………………………………………………………………………. 9 CONTROL
BUTTONS…………………………………………………………………………………….. 10 SETTINGS
……………………………………………………………………………………………………. 11 SUPERVISOR SETTINGS
………………………………………………………………………………. 12
SECTION 4 ………………………………………………………………….. 15
FILTERS ………………………………………………………………………………………………… 15 PRE-FILTERS –
CHANGING……………………………………………………………………………. 15 MAIN CARBON FILTERS –
CHANGING……………………………………………………………. 15 MAIN HEPA FILTERS – CHANGING
………………………………………………………………… 16 EXHAUST FILTER CHANGING (IF FITTED)
…………………………………………………… 16 FILTER SELECTION………………………………………………………………………………………. 17
MAXIMISING FILTER LIFE……………………………………………………………………………… 18 CARBON FILTER
EFFICIENCIES ……………………………………………………………………. 18 ABSORBTION CAPACITIES
…………………………………………………………………………… 18
SECTION 5 ………………………………………………………………….. 19
MAINTENANCE………………………………………………………………………………………. 19 FUSES
…………………………………………………………………………………………………………. 19 LIGHTING
…………………………………………………………………………………………………….. 19 CHECKING THE LOW AIRFLOW ALARM
………………………………………………………… 20 CHECKING THE CARBON FILTER CONDITION ALARM
…………………………………… 20
3
Warning This system must be used in compliance with these instructions and any
repairs or maintenance carried out by qualified personnel. For parts or
service information please contact Monmouth Scientific on: +44 (0) 1278 458090
4
SECTION 1
DESCRIPTION OF THE CABINET
The Circulaire range of filtration fume cabinets are designed to provide
operator and environmental protection. The cabinet provides an inflow of air
0.5m/sec through the working aperture to provide operator protection. The contaminated air is then passed through electrostatically charged pre-filters to remove particulate and then through Activated Carbon main filters to remove chemical contaminates before exhausting the air back to the laboratory. An additional carbon or HEPA exhaust safety filter may be fitted if required.
When installed correctly the cabinet complies fully with international standards including BS7989:2001 for filtration fume cupboards.
External Dimensions
Internal Dimensions
Circulaire T800
800mmWide 700mmDeep 1210mm High 785mmWide 680mmDeep 840mm High
Circulaire T1100
1100mmWide 700mmDeep 1210mm High 1085mmWide 680mmDeep 840mm High
Circulaire T1400
1400mmWide 700mmDeep 1210mm High 1385mmWide 680mmDeep 840mm High
Circulaire T1800
1800mmWide 700mmDeep 1210mm High 1785mmWide 680mmDeep 840mm High
5
SECTION 2
INSTALLATION
The cabinet should be sited in a draught free position The cabinet is
recirculating and requires no connection to ductwork Lift up the front cover
and remove the inner filter access cover using the key
provided and check that the main Carbon Filter/s are in place and the
identification tags are connected. If filters are not fitted see section 4 for
instructions. Check the pre-filters are in place by rotating the small plastic
catch located inside the enclosure, which will allow the pre-filter retaining
frame to be lowered. Connect the cabinet to a 13A outlet socket TESTING /
COMMISSIONING A test certificate will be supplied for conformity to CE
marking, and electrical test. The airflow should be checked using a vane
anemometer and the results recorded. The operation of the filter condition
alarm should also be checked, see section 5. THE CABINET MUST BE TESTED EVERY
14 MONTHS TO COMPLY WITH C.O.S.H.H REGULATIONS.
6
SECTION 3
GENERAL OPERATION
The main on/off switch is located on the right hand side near the top of the
cabinet. When first turned on the LCD control panel will be displayed
providing operator instructions for cabinet use. The cabinet can be put into
standby mode by pressing the key. When in standby mode the fan, lights and
control system are turned off. Normal operation can be resumed by touching the
screen. An ECO mode can be activated (see page 11) which turns off the lights
and reduces the fan speed if no activity is seen inside the cabinet for a time
period set in the supervisor settings (see page 11). Normal operation is
resumed by touching the screen. If the cabinet is not to be used for several
days the cabinet can be turned off at the main on/off switch if preferred.
7
CONTROL SYSTEM The cabinet is controlled and monitored by a microprocessor with an LCD touch
screen. The touch screen provides the operator with general information about the
cabinet and displays the current face velocity and filter condition. Control of light
level, fan speed and other operating settings are all accessed via a menu system.
The control system is factory set to maintain the following parameters:
Normal face velocity:
0.50m/sec.
Low fan speed:
0.35m/sec.
Low airflow alarm:
0.30m/sec.
These settings can be changed by a Monmouth engineer. Please contact Monmouth
Scientific for further advice.
START-UP When first turned on the screen below is displayed providing
information about the
filters fitted to the cabinet. It is vitally important that the correct
filters are fitted
to suit the work being carried out. Basic filter information can be displayed
by
pressing the types.
key. Please contact Monmouth Scientific for further advice on filter
If the procedures carried out inside the cabinet are always the same this
screen can be bypassed in the Supervisor Settings (see page11)
8
When the filter type has been accepted the normal running screen below is
displayed.
FACE VELOCITY The microprocessor controls the fan speed to compensate for
filter blockage and other changes in airflow. The face velocity is
continuously monitored and displayed. The value will fluctuate slightly during
normal operation, this is normal and is an indication that the fan is under
microprocessor control. If preferred the units displayed can be changed to
ft/min. in the Supervisor Settings (see page 11) CARBON FILTER CONDITION The
cabinet is fitted with a hydrocarbon sensor which monitors the airflow after
it has passed through the main filter/s. The filter condition is displayed on
the bar graph indicator in the middle of the display. If filter breakthrough
is detected the green line will progress to the right and warning messages
will be displayed. If the main filter fitted to the cabinet is a HEPA filter
the bar graph will not be displayed.
9
CONTROL BUTTONS Light On / Off Reduce Brightness Increase Brightness Fan On
Fan Off Low Fan Speed – Reduces fan speed to 0.35m/sec. for use when weighing
Settings – access to settings menu Information – Cabinet and filter
information Standby – Puts cabinet into standby
10
SETTINGS Pressing the key on the main screen enters the Set-Up menu access
screen.
The supervisor access code is supplied with the cabinet and allows access to
change operating preferences. The engineer access code is reserved for service
engineers to carry out maintenance procedures. The key returns to the normal
display screen.
11
SUPERVISOR SETTINGS When the correct Supervisor code is entered the screen
below is displayed.
Face Velocity Display – select preferred velocity units. Exhaust Sensor
Sensitivity – Changes the sensitivity of the carbon filter sensor to enable
detection of less volatile chemicals. Audible Alarm – Turns off the audible
low airflow alarm. Keypad Sounds – Turns off the audible key beeps. ECO Mode –
turns off the lights and reduces the fan speed if no activity is seen inside
the cabinet for a time period set by using the arrow keys at the bottom of the
screen. Normal operation will be resumed when operator touches the screen.
Display Acceptance Screen – Turns on / off the Filter type acceptance screen
at start up. Set Time – displays a time / date setup screen
Pressing the key returns to the previous screen.
12
INFORMATION SCREEN Pressing the key on the main screen will display the
information screen below.
Service information is reset by a Monmouth engineer during a service visit.
When a service becomes due a warning screen will be displayed momentarily on
start-up during the proceeding month. Pressing the key will display the screen
below providing information on the type of filters fitted. The key will return
to the previous screen.
13
Pressing the key again will display information on the types of filter
available.
14
SECTION 4
FILTERS
Genuine Monmouth Filters are fitted with an electronic tag which transmits
filter type, part number etc. to the control system automatically on start-up.
Filters from alternative suppliers may be used but the efficiency and capacity
cannot be guaranteed. The control system will detect non genuine filters and
display a warning on start-up. IMPORTANT: Filters concentrate dust, pollutants
etc. and care must be taken when changing them. Personal Protective Equipment
must be worn including gloves and particulate face mask.
PRE-FILTERS – CHANGING This may be carried out with the cabinet running to
provide additional protection to the operator. Rotate the small plastic catch
inside the enclosure which will allow the pre-filter
retaining frame to be lowered and the filter replaced.
MAIN CARBON FILTERS – CHANGING Check filters to be fitted are the correct
grade for intended use. Contact Monmouth Scientific for information if
required. The cabinet should be turned off whilst changing the main Carbon
Filters. Lift up the front panel and remove the inner filter access cover with
the key
provided. Disconnect the wire from the electronic filter tag(s). Rotate each
filter clamp handle 180° to raise filters. Slide filters out and seal in a
marked bag for disposal. Fit new filters ensuring they are pushed fully in
before rotating filter clamp
handles to clamp the filter in place. Re-connect the wire to the electronic
filter tag(s). Re fit the inner filter access cover. Close the front panel.
Cabinet is now ready for use and the updated filter information will be
displayed
on the general information screen.
15
MAIN HEPA FILTERS – CHANGING Carry out safety hazard assessment for safe
changing Follow the procedure for changing Carbon Filters taking extra care
with operator
protection. (A dropped filter can release particulate). Dispose of filter as
hazardous waste. EXHAUST FILTER CHANGING (IF FITTED) The exhaust filter/s
are mounted on top of the cabinet. Remove the top fixing frame by unscrewing
the fixing bolts. Disconnect the wire from the electronic filter tag(s) and
remove the filter(s) Place used filters in a polythene bag and seal. Label bag
with disposal
instructions. Apply silicone grease to new filter seals and place in position.
Re-connect the wire to the electronic filter tag(s). Replace frame and bolt
down evenly. If the cabinet is used with toxic particulates and a HEPA exhaust
filter fitted the
filter should be checked using DOP equipment.
16
FILTER SELECTION It is most important that filters fitted are correct for the
particular application. A guide to filter selection is as follows:
Gaseous fumes Activated Carbon filters. Different grades are available to
improve efficiency and extend filter life. Particulates HEPA filters.
Circulaire HEPA filters are 99.997% efficient for particulates greater than
0.3 microns. For maximum protection against penetration the safety exhaust
HEPA filter should be selected. The exhaust filter has seals under negative
pressure to eliminate possible filter bypass. The main filter can either be
HEPA or activated carbon.
Activated Carbon Filters Standard activated carbon is suitable for a wide
range of pollutants including hydrocarbons. Activated carbon can be
impregnated with chemicals to neutralise types of chemicals and there is a
list on the following page to indicate the types available. The given weight
is approximate to standard activated carbon. Impregnated carbons have higher
densities and will increase filter weight.
Filter Type HEPA ACTIVATED CARBON A/C ACID FORM SUL AMM ETHER SCHOOLS
Application Particulates Hydrocarbons
Acid gasses Aldehydes Sulphur compounds Ammonia Ethers Educational, Animal
odours
Typical Chemicals Asbestos / powders Alcohols, Hydrocarbons, General use So2,
HCL, H2So4 Formalin Glutaraldehyde H2S, mercaptans NH3, NH4
SO2, H2SO4, BR2, H2S, NH3, CCL4, hydrocarbons
All grades of activated carbon have general use capability for hydrocarbons. Other grades are available for applications not listed above. Filters can be manufactured in layers suitable for more than one application.
To determine correct filter type please contact Monmouth Scientific with details of application, volumes, concentrations, temperatures etc.
Exhaust Filters The cabinet may be fitted with a safety exhaust filter if required to provide an additional level of safety / capacity if required. Cabinets used for toxic particulates must be fitted with a HEPA exhaust filter. (The main filter may be Carbon or HEPA.)
17
MAXIMISING FILTER LIFE Handle minimum volumes of chemicals Minimise surface area of exposed chemicals to reduce evaporation rates Cover containers as far as practical Do not boil off large volumes of chemicals Minimise use of heat Acids should be at room temperature and covered as far as practical
CARBON FILTER EFFICIENCIES Typical filter efficiencies are >99% and this efficiency is maintained for most of the filter life. Filters should be changed when efficiency has reduced to below 90%.
ABSORBTION CAPACITIES Circulaire cabinets have very large filter capacities, with a typical value of >30% for hydrocarbons. The cabinets have the following nominal absorption capacities:
Model
Circulaire 800 Circulaire 1100 Circulaire 1400 Circulaire 1800
Carbon Weight
1 X 16Kg 2 X 14Kg 2 X 16Kg 2 X 16Kg + 1 X 14Kg
Hydrocarbon capacity at 30% absorption 4.8Kg 8.4Kg 9.6Kg 18Kg
Impregnated filters have different densities and filter capacities. Contact Monmouth Scientific for absorption capacities for different applications.
18
SECTION 5
MAINTENANCE
The cabinet should be isolated from the electricity supply before carrying out
any maintenance procedures. FUSES The main fuses are located in the mains
inlet socket on the top of the cabinet. Remove the mains lead and withdraw the
fuses using a small screwdriver. Always replace fuses with the correct type
and rating. LIGHTING The high efficiency, low voltage LED light tubes are
fitted to the inside of the front cover. They should provide many years of
service without requiring replacement. Spare tubes are available from Monmouth
Scientific and have the following part numbers: Circulaire CT800 – GS-01514
Circulaire CT1100 – GS-01539 Circulaire CT1400 – GS-01540 Circulaire CT1800 –
GS-01540 Care should be taken to ensure the tubes are fitted in the correct
orientation with the red+ marking on the tube aligning with red mark on the
fitting.
19
CHECKING THE LOW AIRFLOW ALARM The low airflow alarm will be activated if the face velocity drops below 0.35m/sec. To check the function of the alarm follow the procedure below:
- With the cabinet running measure the face velocity using a rotating vane anemometer. The average reading should be 0.55m/s ± 0.02
- Using paper or similar material progressively block the air inlet grille/s as evenly as possible inside the cabinet to simulate a blocked filer.
- The fan will automatically speed up to try and compensate for the loss in airflow, however as the filters are increasingly blocked a point will be reached where the fan has no more capacity and the airflow will begin to drop.
- Monitor the face velocity to check that the alarm is activated at around
0.3m/sec. 5) Remove paper to allow the cabinet to return to a normal running
condition.
CHECKING THE CARBON FILTER CONDITION ALARM 1) Turn the cabinet off, open the front cover and release the filter clamp to provide a
filter bypass. 2) Close the cover and turn the cabinet on. 3) The filter condition alarm is inhibited for 5 minutes during start up to allow the
sensor to stabilise. After 5 minutes proceed to step 4. 4) Pour a small amount of Isopropyl Alcohol onto a piece of tissue paper inside the
cabinet. 5) The hydrocarbon sensor should sense the presence of the fumes, the bar graph
indicator will indicate an alarm condition and activate the audible alarm. 6) Turn the cabinet off, re-clamp the filter and run the cabinet to contain any fumes
still being released. 7) The bar graph indicator will drop to indicate a safe condition and the cabinet is
safe to use. 8) If no response is observed from the challenge test, or the cabinet does not
recover to a safe condition after the test the main carbon filters may need to be replaced or the system may need to be re-calibrated. Contact Monmouth Scientific for further advice.
20
21
Monmouth Scientific Ltd
Units 5 & 6 Kilnside East Quay Bridgwater Somerset TA6 4DB Email:
info@monmouthscientific.co.uk
22
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>