ZOLL CL-2295A Cool Line Intravascular Heat Exchange Catheter Instruction Manual
- September 16, 2024
- ZOLL
Table of Contents
- ZOLL CL-2295A Cool Line Intravascular Heat Exchange Catheter
- Device description
- Contraindications
- Warnings, precautions, adverse reactions
- Check for a catheter leak
- Materials required
- Use sterile technique.
- Preparing the catheter
- Inserting the catheter
- Disconnecting the catheter from the console
- Reconnecting the catheter to the console
- Removing the catheter
- MRI safety information
- Guidewire Instructions for Use
- Cool Line catheter
- Precautions For Use
- Transparent Film Dressing
- Specifications:
- FAQ:
- Q: Can the Cool Line Catheter be reused?
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ZOLL CL-2295A Cool Line Intravascular Heat Exchange Catheter
INSTRUCTION MANUAL
Federal law restricts this device to sale by or on the order of a physician.
Cool Line® Intravascular Heat Exchange Catheter Kit includes
Quantity | Description |
---|---|
1 | Cool Lineintravascularheatexchange catheter |
9.3 French x 22 cm Triple infusion Luer Extension line clamps Radiopaque shaft
Applause heparin coated
1| Guidewires 0.032″ (0.81 mm) x 65 cm
1| Vessel dilator 10 F x 0.038″
(3.4 mm x 1.0 mm)
1| Detachable suture tab & clip
1| 18 ga x 2½” (1.3 mm x 63 mm) Radiopaque
PTFE OTN catheter
1| 000 Silk suture
1| Chloraprep® triple swabstick prep pack
6| 4″ x 4″ (10 cm x 10 cm) Gauze sponges
---|---
1| No. 11 Surgical blade with long handle
---|---
1| 3 cc Syringe & 25 ga x 1½” (0.5 mm x 25 mm) Needle
1| Fenestrated drape
1| 18 ga x 2¼” (1.3 mm x 70 mm) Needle
1| Needle disposal cup
Device description
The Cool Line intravascular heat exchange catheter (“Cool Line catheter” or “catheter”) is a sterile, single use flexible 9.3 F catheter designed for placement in the central venous circulation from an insertion site in the jugular, subclavian, or femoral veins. The catheter is to be connected to a ZOLL® single use disposable Start-Up Kit and the Coolgard®/Thermogard® family of consoles (“console”) (all supplied separately). A dilator and guidewire are required for the percutaneous insertion of the catheter. Three Luers are available for infusion, measuring central venous pressure or sampling.
Infusion port | Flow rate | Priming volume |
---|---|---|
Proximal (blue) | 1400 ml/hr | 0.3 cc |
Medial (white) | 1200 ml/hr | 0.3 cc |
Distal (brown) | 2100 ml/hr | 0.4 cc |
The catheter blood contact surfaces (tip, balloon, and shaft) are treated with an ant?thrombotic Applause heparin coating.
Sterility
Ethylene oxide sterilized. The catheter is supplied sterile for single use only and should not be resterilized. The package should be inspected prior to use to ensure that the sterility barrier has not been compromised.
Storage
Store in a cool, dry place. Avoid freezing and excessive heat
Indications for use
The Cool Line Catheter Model CL-2295A, when used with the ZOLL Thermal Regulation System, is indicated for use in fever reduction, as an adjunct to antipyretic therapy, in adult patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.
WARNING – Fever Reduction
The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. The safety and effectiveness of this device was examined in a randomized controlled trial of 296 patients. The mortality results reported in this trial, for the four patient cohorts enrolled, are presented in the table below (Cl – cerebral infarction, ICH – intracerebral hemorrhage, PTBI – prt mary traumatic brain injury, SAH – subarachnoid hemorrhage)
| Cool Line| Control|
---|---|---|---
| n| N| %| n| N| %| p*
Cl| 3| 16| 18.8| 3| 14| 21.4| 0.74
ICH| 8| 33| 24.2| 7| 27| 25.9| 1.00
PTBI| 10| 44| 22.7| 4| 38| 10.5| 0.24
SAH| 13| 61| 21.3| 7| 63| 11.1| 0.15
Fischer’s exact test
For more details on the clinical trial results, refer to the Physi- cian’s Manual – “Normothermia for the Neuro-critically Ill stroke patient”.
Contraindications
1. The risks of the catheter are essentially those of a central line. The
catheter should not be used in patients for whom central line placement is not
indicated.
2. Bleeding diathesis.
3. Active sepsis.
4. Infection or active bleeding at the site of catheter insertion.
5. Patients with no vascular access, or a vascular system that Vvill not
accommodate a catheter.
6. Patients for whom the required temperature monitoring cannot be
established.
7. Hypothermia is contraindicated in patients who have hematological diseases
that will be made worse with hypothermia (e.g., any disease that produces
cryoglobulinemia, any hemoglobinopathy in which hemolytic anemia can
be precipitated by cold including Sickle Cell Disease or Thalassemia).
8. Not intended for pediatric or neonatal use.
Warnings, precautions, adverse reactions
Central venous catheterization should only be performed by welltrained personnel well-versed in anatomical landmarks and safe technique. Personnel should also have knowledge of potential complications.
WARNING
The catheter should be placed via a jugular, subclavian, or femoral vein approach only. Do not allow the catheter to be placed into the right atrium or right ventricle. Placement in the right atrium or right ventricle can result in severe patient injury or death.
Single use only. The product is designed for single use only. Do not
resterilize or reuse. Do not reinsert, once removed from the patient. Do not
alter the catheter in any way. Maximum use period: 7 days.
Potential risks with reuse of a single use device include but are not limited
to
- Potentially life threatening infection
- Toxic shock due to degradation of materials Increased risk of thrombosis
- Reduced heat exchange power
- Device failures
WARNING. The catheter and Start-Up Kit could potentially misconnect with other devices with small bore connectors. Such misconnection errors could result in patient injury or death.
Caution. The custom Luers on the catheter and Start-Up Kit may reduce the risk of misconnections but still have the potential for misconnections with these specific medical device applications: Breathing Systems & Driving Gases applications, Enteral & Gastric applications, Urethral & Urinary applications, Limb Cuff Inflation applications, Neuraxial applications, and lntravascular or Hypodermic applications. Always use caution when connecting ZOLL catheters and Start-Up Kits to these and other medical device applications.
Caution. Ensure that the catheter and/or Start-Up Kit are not connected to another medical device.
1. The use of intravascular cooling devices controls fever, including fever
due to sepsis. Care must be taken to assess patients for sepsis.
2. Caution If blood is observed within the sterile saline circuit, stop the
procedure and perform the check for a catheter leak
3. Do not allow the catheter to be placed into the right atrium or right
ventricle. If placed via the jugular or subclavian veins, the catheter should
be positioned so that the distal tip of the catheter is in the superior vena
cava above its junction with the right atrium and parallel to the vessel wall.
X-ray examination should be used to ensure that the catheter is not in the
right atrium or ventricle. The distal tip of the catheter should be positioned
at a level above either the azygos vein or the carina of the trachea,
whichever is better visualized If placed via the femoral vein, the catheter
should be positioned so that its distal tip is in the inferior vena cava,
below its junction with the right
atrium and parallel to the vessel wall.
4 Possible complications with central venous catheters include: atrial or
ventricular perforation, cardiac tamponade, air embolism, catheter embolism,
thoracic duct laceration, bacteremia, septicemia, thrombosis, inadvertent
arterial puncture, hematoma formation, hemorrhage, nerve damage and
dysrhythmias.
5 All Luer-Lock connections and covers must be securely tightened to prevent
air embolism or fluid or blood loss.
6 Never use excessive force in moving the catheter or guideVvire If
resistance is encountered, an x-ray should be pe rformed to identify the
reason for the resistance.
7 Passage of the guideVvire into the right heart can cause dysrhythmias,
right bundle branch block, vessel wall, atrial or ventricular perforation.
8. Use only sterile normal saline for catheter priming. It is the circulating
fluid in the catheter.
9. The catheter should be routinely inspected for flow rate, security of
dressing, correct catheter position, and for secure Luer-Lock connections. Use
the centimeter markings to identify if the catheter position has changed A
potential risk associated with multi-lumen balloon catheters is a lack of
therapy, including failure to infuse dru g therapy through the infusion
lumens
10. Only x-ray examination can ensure that the catheter tip has not entered
the heart or no longer lies parallel to the vessel wall If the catheter
position has changed, perform an x-ray examination to confirm the catheter tip
position
11. For blood sampling, temporarily shut off the remaining infusion ports
through which solutions are being infused
12. Use only a 30 cc or smaller syringe for blood sampling 13. Alcohol and
acetone can weaken the structure of the shaft material. Care should therefore
be taken when infusing drugs containing alcohol or when using alcohol or
acetone when performing routine catheter care and maintenance. Alcohol should
not be used to declot the catheter.
14 Use only the ZOLL suture tab and clip provided in the kit to prevent
catheter damage
15. Use of a syringe smaller than 10 ml to irrigate or declot an occluded
catheter may cause intraluminal leakage or catheter rupture.
16 Do not infuse into the orange IN and OUT Luer-Lock connections, as this
will result in lack of therapy
17 Fever may have infectious and/ or non-infectious causes in patients.
Mitigation of fever as a sign of possible infection necessitates daily,
meticulous assessment for other signs of infection.
18 Use care when infusing drugs that may be affected by cool temperatures (as
low as 4°(). Solutions containing mannitol are temperature-sensitive and must
not be delivered through the catheter except for a rapid push of a solution of
up to 20% mannitol, followed by a saline flush. Higher than a 20%
concentration of mannitol drip or infusion pump delivery of mannitol must be
done via a separate line.
19. Cardiac Tamponade: Placement of indwelling catheters in the right atrium
is a practice that may lead to cardiac perforation and tamponade.
Practitioners placing central venous catheters must be aware of this
potentially fatal complication before advancing the catheter too far relative
to patient size The actual position of the tip of the indwelling catheter
should be confirmed by x-ray after insertion. Central venous catheters should
not be placed in the right atrium unless specifically required for special
relatively short term procedures, such as aspiration of air emboli during
neurosurgery. Such procedures are nevertheless risk-prone and should be
closely monitored and controlled.
20. The catheter is coated with heparin. This may induce or aggravate pre-
existing heparin induced thrombocytopenia (HIT)
21. WARNING. When connecting infusion sets/injection systems to ZOLL
catheters, do not exceed 100 psi/689 kPa.
22. For patients being made hypothermic, the hypothermia itself may
exacerbate some disease states. Care should be taken to properly monitor
patient homeostasis during hypothermia.
- Cardiac rhythm disturbances-both bradycardia and ventricular tachyarrhythmia.
- Clotting and coagulations function. Patients at risk for disturbances of their clotting or coagulation function should be closely monitored during hypothermia.
- Blood gas and pH analysis. Hypothermia modifies resting pH and Pa CO2. Physicians should be aware of the effect of temperature upon the result.
- Prolonged hypothermia depresses the immune response and lung function.
WARNING. lntraluminal or balloon leakage. lntraluminal leakage (between the saline lumen and infusion lumens) or balloon leakage is a potential catheter failure mode. In the event of such a failure, sterile saline from the cooling circuit is introduced into the patient. lntraluminal leakage or balloon leakage is typically associated with a fluid loss alarm once the saline bag has been depleted and stops the system. Always investigate fluid level alarms. The cooling circuit is a closed loop system – usually fluid loss alarms indicate a breach somewhere in this closed loop. With any fluid loss alarm, check both the integrity of the catheter and the Start-Up Kit (see below).
WARNING. If you notice a depleted saline bag or an air trap alarm, do not
replace the saline bag prior to identifying the location of the leak and
taking the appropriate mitigation. Check for system leaks according to the
instructions in the Check for a Start-Up Kit leak and Check for a catheter
leak sections below. (Note that a leak could be external or internal.)
Replacing the saline bag repeatedly without investigating the leak or loss of
saline may result in unintended infusion of saline into the patient. Saline
infusion may lead to the following adverse effects: local swelling that can
cause subsequent local tissue damage; systemic fluid overload that can lead
to dependent edema and subsequent skin breakdown; internal organ fluid
overload, with subsequent overloading of the brain, lungs or heart. In some
cases, this fluid overload may lead to life threatening events.
Caution. The console emits an alarm when the saline bag is empty. The bag must be completely empty and additional saline must have drained between the saline spike and the air trap for the saline level in the air trap to drop sufficiently to trigger the alarm.
Check for a Start-Up Kit leak
1. Check the air trap for condensation. If the air trap shows signs of
condensation, wipe the air trap and reinstall it in the console. In the case
of an air trap alarm, verify that the air trap alarm is cleared after this
step.
2. Carefully check the saline path from the saline bag to the console for any
leaks. Check if there is saline on the floor, console, or the patient’s bed.
3. If there is any saline on the floor, console, or the patient’s bed, check
that the Luers on the catheter and Start-Up Kit are not cracked or damaged and
that the connections are tight enough to prevent leaks.
4. If you find a leak in the Start-Up Kit, replace the Start-Up Kit and see
if there is also a leak in the catheter.
5. If you do not find a leak in the Start-Up Kit, there is likely a leak in
the catheter. Investigate further.
Check for a catheter leak
1. Disconnect the Start-Up Kit from the catheter. Properly cap both the catheter and Start-Up Kit using an aseptic technique.
2. Fill a sterile 10 ml slip tip syringe with sterile saline.
3. Connect the syringe to the IN Luer of the catheter and disconnect the OUT
cap. Infuse 10 ml of saline – the saline should flow out the OUT Luer. If the
saline does not flow out of the OUT Luer, a catheter leak is indicated.
4. Cap the OUT Luer and pull 5 cc of vacuum. Sustain for at least 10 seconds.
Up to 4 ml of saline (not blood) should enter the syringe and you should be
able to maintain the vacuum. If traces of blood are seen in the syringe or
vacuum cannot be maintained, it indicates a catheter leak.
5. If you find a leak in the catheter, replace the catheter.
6. Replace the saline bag and re-prime the Start-Up Kit.
7. Ensure leak-tight Luer connections to the Start-Up Kit and continue the
therapy_
Materials required
Quantity Description
Quantity | Description |
---|---|
1 | Cool Line Catheter Kit for percutaneous introduction |
1 | 500 cc bag of sterile normal saline (not provided) |
1 | Start-Up Kit (provided separately) |
6 ft ( 183 cm) standard tubing or
9 ft (274 cm) extended tubing
1| Coo lg a rd or Thermogard console (provided separately)
Catheter preparation and insertion
Note. The catheter has a radiopaque marker band to assist in identification of the catheter during and after insertion when viewed using x-ray equipment. The proximal end of the proximal balloon has one marker band. The tip of the catheter contains barium sulfate to make it radiopaque.
Use sterile technique.
Caution. The IN and OUT Luer-Locks on the catheter are custommanufactured and are intended to connect only with the ZOLL Start Up Kit listed in Materials required.
Preparing the catheter
1. Place the patient in a slight Trendelenburg position as tolerated to
reduce the risk of air embolism. If the femoral approach is used, place the
patient in a supine position.
2. Prep and drape the puncture site as required.
3. Carefully remove the catheter from the package, leaving on the catheter
membrane cover.
4. Remove the caps from the IN and OUT Luers. With the catheter cover in
place, fill a syringe (5 cc or larger) with sterile saline and attach the
syringe to the female IN Luer.
WARNING. Never inject positive pressure into the IN Luer with the OUT Luer cap
in place.
5. Gently inject saline through the catheter until it begins to exit from the
OUT Luer.
6. Using a 5 cc or larger syringe, flush the distal, proximal, and medial
infusion Luers with sterile saline. Clamp or attach the injection caps to the
proximal and medial infusion Luers. Leave the distal Luer uncapped for
guidewire passage.
Caution. Always prime the catheter infusion Luers before inserting the catheter into the patient.
7. Remove the catheter membrane cover. If there is resistance in removing the membrane cover from the catheter, flush the membrane cover with sterile saline. Inspect the catheter to ensure that air has been purged from the heat exchange membrane. Inspect the catheter for leaks.
Caution. Avoid excessive wiping of the coated catheter. Avoid wiping the catheter with dry gauze, as this may damage the catheter coating. Avoid using alcohol, antiseptic solutions, or other solvents to pre-treat the catheter, because this may cause unpredictable changes in the coating, which could affect the device safety and performance.
WARNING. Do not cut the catheter to alter length.
Inserting the catheter
Caution. Use jugular, subclavian, or femoral vein approach only.
1. Obtain jugular, subclavian, or femoral venous access using standard
percutaneous techniques. Access should be maintained with a 0.032″ (0.81 mm)
guidewire. See Guidewire Instructions for Use.
WARNING. Do not attempt to reinsert a partially or completely withdrawn OTN
(over the needle) introducer needle from its catheter.
Caution. Do not use a guidewire larger than 0.032″ (0.81 mm) with the
catheter.
2. Holding the guidewire in place, remove the introducer catheter. Caution.
Maintain a firm grip on the guidewire at all times.
3. Enlarge the cutaneous puncture site with the cutting edge of the scalpel
positioned away from the guidewire.
WARNING. Do not cut the guidewire. Use a vessel dilator to enlarge the site as
required. Do not leave the vessel dilator in place as an indwelling catheter,
to minimize the risk of a possible vessel wall perforation.
4. Thread the tip of the catheter over the guidewire. Maintain a sufficiently
firm grip on the guidewire during catheter insertion. Grasping the catheter
tip near the skin, advance the catheter into the vein. Continue to advance the
catheter over the guidewire, placing your fingers just proximal to the
balloon.
5. Using centimeter marks on the catheter as positioning reference points,
advance the catheter to at least the minimum marker number to ensure the
proximal infusion port is in the vessel.
6. Hold the catheter at the desired depth and remove the guidewire. If
resistance is encountered when attempting to remove the guidewire after
catheter placement, the guidewire may be kinked at the tip of the catheter. If
resistance is encountered, withdraw the catheter relative to the guidewire
about 2-3 cm and attempt to remove the guidewire. If resistance is encountered
again, remove the guidewire and catheter simultaneously.
Caution. Do not apply undue force to the guidewire.
7. Verify that the guidewire is intact upon removal.
8. Check catheter placement by attaching a syringe to the distal infusion
Luer and aspirate until a free flow of venous blood is observed. Connect the
infusion Luers to the appropriate LuerLock line(s) as required. Unused
infusion port(s) may be
“locked” through the injection cap(s) using standard hospital protocol. Slide
clamps are provided on the tubing to occlude flow through the infusion Luers
during line and injection cap changes.
Caution. To minimize risk of damage to the tubing from excessive pressure,
the clamp must be opened prior to infusing through the Luer.
Caution. Do not clamp or occlude the IN or OUT lines. This can cause line
blockage and possible failure.
9. Secure and dress the insertion site and catheter temporarily.
10. If subclavian or jugular access is used, verify the catheter tip position
by chest x-ray immediately after placement. The x-ray examination must show
the catheter located in the right side of the mediastinum in the SVC with the
distal end of the catheter parallel to the vena cava wall. The catheter distal
tip must be positioned at a level above either the azygos vein or the carina
of the trachea, whichever is better visualized. If the catheter tip is
malpositioned, reposition and reverify. If femoral access is used, x-ray
examination must show the catheter located in the IVC with the distal end of
the catheter parallel to the vena cava wall. If the catheter tip is ma I
positioned, reposition and reverify.
11. The proximal radiopaque marker indicates the proximal end of the balloon.
Ensure that the balloon and proximal port reside completely in the vessel. If
the catheter is malpositioned, reposition and reverify.
12. Secure the catheter to the patient. Use the suture tab side \Mngs as the
primary suture site to minimize the risk of catheter migration.
13. The ZOLL suture tab and clip can also be used as an additional attachment
point. Ensure that the catheter body is secure and does not slide.
Caution. Use only the ZOLL suture tab and clip provided in the kit. Catheter
damage may result if other tabs or clips are used. Caution. Do not suture
directly to the outside diameter of the catheter, to minimize the risk of
cutting or damaging the catheter or impeding catheter flow.
14. Dress the puncture site per hospital protocol. Maintain the insertion
site with regular meticulous redressing using aseptic technique.
15. Record on the patient’s chart the indwelling catheter length using the
centimeter marks on the catheter shaft as reference. Frequent visual
reassessment should be made to ensure that the catheter has not moved.
16. Attach a primed Start-Up Kit to the catheter: Connect the male Luer of
the Start-Up Kit to the female IN Luer of the catheter (labeled “IN”). Connect
the female Luer of the Start-Up Kit to the male OUT Luer of the catheter
(labeled “OUT’). White
“ZOLL” tags are fitted loosely to the IN and OUT extension tubes to help
identify them.
Note. The Start-Up Kit IN and OUT Luers are only intended to connect to the
catheter IN and OUT Luers and are not intended to connect to standard Luer
Lock syringes. They have ZOLL custom fittings and are orange in color for
easy identification.
17. Ensure that a sufficient amount of sterile saline is present at the ends
of the Luers to make an air-free connection. Refer to the operation manual.
WARNING. Failure to connect the Start-Up Kit correctly to the catheter could
result in catheter failure. Do not attach the StartUp Kit (orange) Luers to
the blue, white, or brown infusion Luers.
Caution. Do not attach the Start-Up Kit to the distal port. Caution. Do not
place any stopcocks in line that may be inadvertently shut off. This can
cause line blockage and possible failure.
18. Pump saline through the Start-Up Kit and catheter to ensure that all
connections are secure and that there is no leaking. Allow any remaining air
in the system to be purged out as described in the operation manual.
Disconnecting the catheter from the console
1. Stop circulation of saline through the catheter.
2. Disconnect the Start-Up Kit from the catheter.
3. To maintain sterile connections, immediately cap off the Luer connectors
of both the catheter and Start-Up Kit using sterile Luer caps, or connect the
IN and OUT Luers together.
Reconnecting the catheter to the console
-
Remove the Luer caps from the Luer connectors of the catheter and Start-Up Kit. Discard the Luer caps or disconnect the IN and OUT Luers from each other.
-
Attach the Start-Up Kit to the catheter: Connect the male Luer of the Start-Up Kit to the female IN Luer of the catheter. Connect the female Luer of the Start-Up Kit to the male OUT Luer of the catheter. The Start-Up Kit and catheter IN and OUT Luers are orange in color. Ensure that a sufficient amount of sterile saline is present at the ends of the Luers to make an airfree connection.
WARNING. Failure to connect the Start-Up Kit correctly to the catheter could result in catheter failure.
WARNING. Do not use the IN and OUT Luer fittings for starr dard central line infusion ports. They are for connection to the Coolgard or Thermogard console only. -
The Start-Up Kit IN and OUT Luers are only intended to connect to the catheter IN and OUT Luers. They have ZOLL custom fittings and are orange in color for easy identification.
Caution. Do not place any extra stopcocks in line that may be inadvertently shut off. This can cause line blockage and possible failure.
Removing the catheter
1. Stop pumping saline through the catheter.
2. Disconnect the Start-Up Kit from the catheter. Uncap or leave uncapped the
IN and OUT Luers of the catheter. This allows residual saline within the
circuit to be expressed. As the catheter is withdrawn, the balloons are
compressed. Saline within the balloons must be free to pass out of the balloon
or the balloon will not deflate, making the catheter difficult to remove.
3. Optionally, attach a 20 or 25 cc syringe to the catheter IN Luer. Pull and
hold a vacuum for 15 seconds to allow residual saline to be removed from the
catheter balloon section prior to removing the catheter.
4. Place the patient in a supine position. Remove the dressing. Remove the
sutures from the suture site.
WARNING. Do not place a Luer cap on the OUT Luer.
5. Slowly remove the catheter from the patient. As the catheter exits the
site, apply pressure with a dressing impermeable to air (e.g. Vaseline gauze).
WARNING. Do not move the catheter if resistance is felt. Check to ensure
that the IN and OUT Luers of the cooling circuit are NOT capped. If they are
capped, uncap them, deflate the balloon, and try removing the catheter again.
If resistance is still encountered, an x-ray should be performed to identify
the reason for the resistance.
MRI safety information
The Cool Line lntravascular Heat Exchange Catheter is MR Conditional. A patient with this catheter may be safely scanned under the followingconditions. Failure to follow these conditions may result in injury to the patient.
Parameter | Condition |
---|
Nominal Values of Static
Magnetic Field (T)| 1.5 Tand 3.0 T
Maximum Spatial Field Gradient
(T/m and gauss/cm)| 40 T/m (4,000 gauss/cm)
Type of RF Excitation| CircularlyPolarized(CP)(i.e., quadrature
driven)
Transmit RF Coil Information| There are no transmit RF coil restrictions.
Accordingly, the following may be used: body transmit RF coil and all other RF
coil combinations (i.e., body RF coil com- bined with any receive-only RF
coil, transmit/receive head RF coil, transmit/
receive knee RF coil, etc.)
Operating Mode of MR System| Normal Operating Mode
Maximum Whole Body
Averaged SAR| 2-W/kg (Normal Operating Mode)
Limits on Scan Duration| Whole body averaged SAR of 2-W/kgfor 60 minutes of
continuous RF exposure (i.e., per pulse sequence or back to back
sequences/series without breaks)
MR Image Artifact| The presence of this implant produces an imaging artifact.
Therefore, carefully select pulse sequence parameters if the implant is
located in the area of interest.
WARNING| The ZOLL Coolgard and Thermogard consoles are MR Unsafe and, thus,
these consoles are not allowed in the MR sys- tern room. Therefore, the
catheter must be disconnected from the console prior to moving the patient
into the MR sys-
tern room.
Guidewire Instructions for Use
Caution. Federal law (USA) restricts this device to sale and use by or on the order of a physician.
Note. This information applies only to the use of guidewires in the Seldinger technique of catheter placement in the vasculature.
Warnings
The supplied guidewire is designed for single use only. Do not resterilize or
reuse. Do not reinsert, once removed from the patient.
Should resistance occur during insertion or withdrawal, do not continue to
move the guidewire. Determine the cause under fluoroscopy and take action as
needed.
Use extreme caution when moving a guidewire through a stent. Use of a
guidewire in stented vessels creates additional patient risk.
Cautions
Avoid withdrawing the guidewire through metal needles; the guidewire may
shear.
Because of the delicate and fragile nature of guidewires, extra care in
handling must be taken. Avoid bending or kinking. Do not use damaged
guidewires.
During storage or procedures, avoid coiling the guidewire in less than 8 inch
(20.32 cm) diameters, as smaller diameters place unnecessary stress on the
guidewire and may cause it to kink. The provided dispenser is the best means
of storage and handling of the guidewire.
Sufficient guidewire length must remain exposed to maintain a firm grip on the guidewire at all times.
Dispenser
Every guidewire is provided in a dispenser packaged within a Tyvek pouch.
Remove the guidewire assembly from the Tyvek pouch. Remove the guidewire
protective cap immediately prior to guidewire use. Prepare the guidewire prior
to insertion. It is recommended that the dispenser be filled with heparinized
solutions (e.g. saline or dextrose) to bathe the guidewire during insertion.
The preformed “J” guidewire will resume its shape when removed from the
product dispenser.
1. Guidewire protec tive cap
Inspection
Inspect the guidewire prior to use and discard if any deformities are present in the guidewire. Guidewire placement should be routinely monitored by x-ray or fluoroscopic procedure.
Technique
1. Puncture the vessel.
2. Insert the guidewire into the needle hub and gently advance 5-10 cm of the
guidewire into the punctured vessel. Navigate the guidewire to the desired
position.
Caution. Avoid rough or overly vigorous manipulation of the guidewire to
prevent damage to the guide or the vessel.
3. Remove the needle from the guidewire.
4. Dilate the tissue and vessel with the dilator using a slight rotary
motion.
5. Remove the dilator (the vessel dilator is intended for vascular dilation
only).
6. Introduce the catheter by sliding it over the guidewire.
7. Remove the guidewire.
Cool Line catheter
1. IN Luer
2. Out Luer
3. Infusion ports
4. Distal Luer
5. Proximal port
©2024 ZOLL Medical Corporation. All rights reserved. ZOLL and Cool Line are
trademarks or registered trademarks of ZOLL Medical Corporation and/or ZOLL
Circulation Inc. in the United States and/or other countries. All trademarks
are the property of their respective owners.
Patent: www.zoll.com/patents
Precautions For Use
Suresite Window may be used on clinlcallv infected wounds if the following precautions are followed:
• The patient should bill! 1.1nd1.n medic.-l(clinical SllJ>ftrll i:sion.
• The dressing should be changed daily.
• The p;1tient should be receiving suitable systemic treatmen1. lmmuoo-
comprcmised patients and diabeticpatients may require extra superviaion. Care
!.hould
be taken lo avoid skin damage from repeated applicatiorts on patien1s with
thin or fr.agile skin.
Sterile. Single usa. Do not use contents if package is opened or damaged.
Store at room tempe..ature, 59-86’F.
www.medine.com ©2004 Medline Industries, Inc.
Medline Ia A registered traciemark of Mediine Industries, Inc.
Made [n USA tor: Medine Industries, Inc.,
Mundelein, IL 600! USA 1-800-MEDLINE RI4PCI
Transparent Film Dressing
Description
Suresite* Window Transparent Film Dressing consists of a polyurethane film
with acrylic adhesive.
The dressing is moisture vapar penmeable, thus.
atlowing oXygen and moisture vapor to pass through the dressing. When praperly
applied, Suresite is impermeabie to liguids and bacteria.
Indications:
Suresite dressings are intended for minor abrasions, skin tears and to help
prevent skin breakdown. May also be used on pressure ulcers {stagas | & 1)
with minirnal drainage, partial-thickness wounds, clean, closed surgical
incisions, first and secand degree burns and for autolytic debridement. Aiso
indicated for the management of peripharal and central LV, catheter sites.
Contraindication:
Suresite is contraindicated for use as a primary dressing on maderately ta
heavily draining wounds.
Application:
1. Prepartehe site according to facility guidelines.
Clip any excess hair at the site. Shaving is not recommended, Allow any skin
preparation to dry completely.
2. Peet the I’ner from the dressing, exposing the adhasive surface.
3. Position the dressing over the site. If securing over an LV., center the
slit portian of the frame over the catheter hub
4. Gently remove the paper frame, smoothing the dressing down as you pull the
frame away.
5. For L. catheter sites, seal the dressing around and under the catheter
hub.
6. Firmly smooth dressing from the center toward the edges.
7. Date and initial the label and apply to edge of diessing if desired.
Removal:
1. Gently grasp the edge and slowly peel the dressing fram the skin in the
direction of hair growth or grasp one edge of the dressing and gently pull it
straight out (o stretch and rafease adhesion.
2. An adhesive solvent can be used to facilitate dressing removal,
Specifications:
-
Catheter Type: Intravascular heat exchange
-
French Size: 9.3
-
Length: 22 cm
-
Features: Triple infusion, Luer Extension line clamps,
Radiopaque shaft, Applause heparin coated -
Guidewires: 0.032 (0.81 mm) x 65 cm
-
Vessel Dilator Size: 10 F x 0.038 (3.4 mm x 1.0 mm)
-
Detachable Suture Tab & Clip
-
Silk Suture: 000
-
Gauze Sponges: 4 x 4 (10 cm x 10 cm)
-
Surgical Blade: No. 11 with long handle
-
Fenestrated Drape
-
Needle Disposal Cup
-
Infusion Ports: Proximal (blue), Medial (white), Distal
(brown) -
Flow Rates: 1400 ml/hr, 1200 ml/hr, 2100 ml/hr
-
Priming Volumes: 0.3 cc, 0.3 cc, 0.4 cc
-
-
FAQ:
Q: Can the Cool Line Catheter be reused?
A: No, the Cool Line Catheter is supplied sterile for single-use only and should not be resterilized or reused to prevent potential risks associated with reuse of a single-use device.
Q: What are the indications for using the Cool Line Catheter?
A: The Cool Line Catheter is indicated for use in fever reduction as an adjunct to antipyretic therapy in adult patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and are intubated and sedated.
Q: What are the storage recommendations for the catheter?
A: Store the catheter in a cool, dry place. Avoid freezing and excessive heat to maintain product integrity.
Q: How should catheter placement be performed?
A: Catheterization should only be performed by well-trained personnel familiar
with anatomical landmarks and safe techniques.
Ensure placement via jugular, subclavian, or femoral vein approach only to
prevent severe patient injury or death.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>