Laviband Scuba III Smart Pooltester
Product Information
Specifications
- Product Name: Smart Pooltester Scuba III
- Parameters List: Free Chlorine, Total Chlorine, Combined Chlorine, Bromine, pH, Cyanuric Acid, Total Alkalinity, Calcium Hardness, Active Oxygen, Copper, Phosphate
- Measuring Range:
- Free Chlorine: 0.1 – 6 mg/L
- Total Chlorine: 0.1 – 6 mg/L
- Combined Chlorine: 0.1 – 6 mg/L
- Bromine: 0.2 – 13.5 mg/L
- pH: 6.5 – 8.4 mg/L
- Cyanuric Acid: 20 – 120 mg/L
- Total Alkalinity: 5 – 300 mg/L
- Calcium Hardness: 50 – 500 mg/L
- Active Oxygen: 0.2 – 10 mg/L
- Copper: 0.1 – 5 mg/L
- Phosphate: 0.06 – 3 mg/L
- Range for safe swimming conditions (Green light): Free Chlorine: 0.3 – 4 mg/L
Product Usage Instructions
-
Battery Replacement:
- Use non-rechargeable batteries as specified in technical data.
- Ensure the instrument is dry before opening the battery compartment.
- Pay attention to the correct polarity while inserting batteries.
- Tighten the cover securely after battery replacement.
- Transfer measurement data before replacing batteries.
-
Disposal Instructions for Batteries and
Accumulators:- Return used batteries and accumulators to designated facilities or shops selling batteries.
- Electrical equipment must not be disposed of with normal household waste; return to authorized disposal centers.
-
Important Safety Instructions:
- Read the manual carefully before use.
- The intended use is for chemical water testing only in non-commercial applications.
- Suitable for use in wet environments, both indoors and outdoors.
FAQs
Q: Can I use rechargeable batteries with the Smart Pooltester Scuba III?
A: No, only non-rechargeable batteries as specified in the technical data should be used to power the device.
Q: What is the measuring range for Free Chlorine in safe swimming conditions?
A: The recommended range for safe swimming conditions for Free Chlorine is between 0.3 - 4 mg/L, indicated by a green light on the device.
Q: Can the Smart Pooltester Scuba III be used for commercial applications?
A: No, this instrument is designed for non-commercial applications to determine dissolved matter in water through chemical testing and analysis only.
Parameters List
Parameter | Test name on display | Measuring Range |
---|---|---|
Free Chlorine | Cl F | 0.1 – 6 mg/L |
Total Chlorine | Cl T | 0.1 – 6 mg/L |
Combined Chlorine | Cl Comb. | 0.1 – 6 mg/L |
Bromine | Br | 0.2 – 13.5 mg/L |
pH | pH | 6.5 – 8.4 mg/L |
Cyanuric Acid | CyA | 20 – 120 mg/L |
Total Alkalinity | TA | 5 -300 mg/L |
Calcium Hardness | CaH | 50 – 500 mg/L |
Active Oxygen | O2 Act. | 0.2 – 10 mg/L |
Copper | Cu | 0.1 – 5 mg/L |
Phosphate | PO4 | 0.06 – 3 mg/L |
*Range for safe swimming condition (Green light)** | Required |
Tablet Reagents
---|---
0.3 – 4 mg/L| DPD No.1
0.3 – 4.5 mg/L| DPD No.1 + DPD No.3 Evo
0 – 0.5 mg/L| DPD No.1 + DPD No.3 Evo
1.5 – 8.5 mg/L| DPD No.1
6.8 – 7.6 mg/L| Phenol Red Photometer
30 – 50 mg/L| CyA-Test
50 -150 mg/L| Alka-M-Photometer
175 – 400 mg/L| Calcio-H No.1 + Calcio-H No.2
3 – 8 mg/L| DPD No.4 Evo
0.05 – 0.3 mg/L| Copper No.1
0.06 – 0.25 mg/L| Phosphate No.1 LR + Phosphate No.2 LR
- These are globally considered measuring ranges for safe swimming.
- For individual pool control, users should always follow local regulations or recommendations by their Chemical suppliers.
Congratulations on your purchase of the Scuba3s. You are the owner of a state- of-the-art smart polyester for the analysis of poolwater quality with the latest technology.
Battery replacement
Notes/Caution
- Only use non-rechargeable batteries as specified in technical data.
- The instrument must be dry before opening the battery compartment.
- While inserting batteries, pay attention to the correct polarity of the batteries.
- Make sure that the cover is tightly screwed after batteries replacement.
- Please make sure to transfer measurement data before replacing batteries.
Important information for consumers in the EU
Disposal instructions for batteries and accumulators EC Guideline 2006/66/CE requires users to return all used and worn-out batteries and accumulators. They must not be disposed of in normal domestic waste. Your local authority may have a disposal facility; alternatively, you can hand them in at any shop selling batteries and accumulators. You can also return them to the company which supplied them to you; the company is obliged to accept them.
Disposal of electrical equipment
Due to the European Directive 2012/19/EU your electrical instrument must not
be disposed of with normal household waste! Tintometer GmbH will dispose of
your electrical instrument in a professional and environmentally responsible
manner. This service, excluding the cost of transportation, is free of charge.
Send your electrical Tintometer instruments for disposal freight prepaid to
your supplier.
Important safety instructions
Read instructions before use
This manual provides important information on the safe operation of the
product. Please read it carefully and familiarise yourself with the product
before use.
Intended use:
This instrument is a battery-operated measuring device to determine dissolved
matter in water for non-commercial applications. The instrument and reagents
are intended to be used for chemical water testing and analysis only and may
not be used for other purposes. For use inside and outside buildings. Suitable
for use in wet environments.
Caution
The reagents are potentially harmful if they are swallowed. Do not swallow!
Keep them out of the reach of children. Contact for Poison Center:
www.lovibond.com
Requirements for safe use
Note the following points for safe use:
- If the product is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
- The product may only be used under the ambient conditions mentioned in this operating manual.
- The housing of the product must not be opened or modified.
The product must not be used if:
- It is visibly damaged (e.g. after being transported)
- It was stored under adverse conditions for a lengthy period (ambient conditions, see “Technical Data”)
Guidelines for photometric measurements
- Use only blister tablets with black printing.
- Always add the reagent tablets to the water sample straight from the foil without touching them with your fingers.
- Close the blue lid on top of the sample chamber tightly before zeroing/reading.
- Always ensure that the sample chamber, the lid, and the stirring rod are thoroughly rinsed when changing from one test to the other.
Maintenance guidelines
- Keep the instrument in a clean condition.
- No solvent, abrasive materials, or brushes should be used to clean the instrument. Clean the sample chamber using a soft tissue.
- Dry the instrument when not in use, and store it under the advised ambient conditions.
- Remove batteries and don´t close the battery lid if the instrument is being stored for an extended period.
Instrument Overview
General Measurement procedure
Switch Scuba3s on:
- When not in use, Scuba3s turns off automatically after 5 minutes.
- To turn it off manually, press the Down key for 3 seconds.
Select Parameters
- Select the parameter on the screen with the Test key or scroll up and down to select any other parameter.
- After parameter selection, the Scuba3s shows a “Zero measurement” screen.
Collect Sample:
- Collect samples by dipping Scuba3s into pool water multiple times.
- Make sure the sample chamber is filled.
Cover measuring chamber
- Cover the measuring chamber with the rubber cap.
- Make sure that the chamber is tightly closed.
Perform Zero
- Zero measurement sets a reference point by eliminating the influence of the measuring chamber and sample. It is required to perform zero before each Test measurement.
- Press the Test key to perform zero. When finished, the icon on the left side of the display will change from empty vial to full vial.
Add tablet
- Add tablet reagent directly from the blister by pressing from the top of the blister as shown in the picture.
- Refer to the parameters list on the first page to choose the appropriate reagents for each test.
Dissolve tablet
- Use a stirring rod delivered in Scuba3s Box to crush and dissolve the tablet.
- Make sure that the tablet is completely dissolved.
Perform Test
- Close the chamber with a rubber cap and swirl the device back and forth for approx. 15 seconds, do not shake it. Press the Test key to perform the test.
- Some parameters such as Cl.Comb requires two tests, in this case, 2nd test will follow the first test immediately.
Reaction timers
- Cl F, Br, TA, CyA, pH: No timers
- Cl T, O2 Act., CaH, Cu: 2 minutes
- PO4: 10 minutes
- Scuba3s has integrated timers which ensure that the reaction time between reagent and water sample is kept and colour is fully developed.
- Countdowns start automatically, and when finished, measurement takes place.
- You can skip timer by pressing Test key again, but it is not recommended.
Results
- The results are stored automatically.
- The results of some parameters can be switched into different forms (PO4 to P) by pressing the Up key. In the case of Cl Comb, press the Up / Down key to see the results of Cl F or Cl T.
- The instrument indicates if the water is safe for swimming or not.
Attention:
Please follow local regulations and recommendations provided by your chemical
providers to know the ideal concentrations for your pool.
Settings
Scuba3s settings menu
- When in the home menu press the Up key for 3 seconds to enter the settings menu.
- Scroll through the list to see different setting options.
- Press the Test key to enter the sub-menu.
Results
- The latest results will be shown on screen.
- Scroll up or down to see the next results.
- Press the Back key to go back to the main settings menu.
Transfer Results
- The instrument shows NFC next to the battery icon.
- Follow instructions from Lovibond® PoolAssistant App to receive data.
- Note: All data sets are transferred from the memory.
Delete results
- Press the Test key to select the data delete option.
- Press the Test key again to confirm the action.
- All results will be deleted.
- Note: Make sure to transfer all your results before deleting.
Date Setting
- Press the Test key to select the date setting.
- Scroll up or down to see other date formats or press the Test key to confirm.
- Press the Up or Down key to increase/ decrease the value.
- Press the Test key to confirm the value and move to the next entry.
- When the arrow is at the last entry, press the Test key to save the date.
Time Setting
- Press the Test key to select the time setting.
- Press the Up or Down key to increase/ decrease the value.
- Press the Test key to confirm and move to the next entry.
- When the arrow is at the last entry, press the Test key to save time.
Error Codes
- E1 – Indicates an insufficient light on the detector.
- E2 – Indicates calculation error.
- E4 – Indicates invalid calibration.
- E7 – Indicates method not found.
Technical Data
- Optics: Spectrometric sensor, multi-wavelength
- Wavelength: Multi
- Display: LCD Dot Matrix with Backlight
- Interface: NFC (Near Field Communication)
- App Connectivity: Lovibond®-PoolAssistant
- Memory: 100 data sets
- Power Supply: 2 x AA batteries (Mignon AA/LR6) 3V DC / 55mA
- Battery Life: 5000 measurements
- Auto – Off: 5 minutes
- Protection class: IPX7 according to DIN EN 60529
- Conformity: CE, UKCA
- Warranty: 2 years
- Ambient conditions:
- Temperature: 10 – 50 °C
- Humidity: 30 – 95 %
- Pollution degree: 2
- Maximum altitude above sea level: 3500 meters
Delivery Content
- 1 x Scuba 3s in a robust plastic case with a hanger
- 2 x AA batteries
- 1 Stirring rod
- 1 Instruction Manual
- 1 Blue lid
Reagents
- 20 x DPD No.1 tablets
- 10 x DPD No.3 Evo tablets
- 20 x Phenol Red Photometer tablets
- 10 x Alka-M-Photometer tablets
- 10 x CyA-Test tablets
- 10 x Calcio-H No.1 tablets
- 10 x Calcio-H No.2 tablets
List of Accessories & Reagents
- 525620 – Refill Pack Cl2-pH
- 525621 – Refill Pack Cl2-pH-Alka-m-CyA
- 525652 – Refill Pack Calcium Hardness
- 525651 – Refill Pack Active Oxygen
- 525653 – Refill Pack Copper
- 525654 – Refill Pack Phosphate
- 364107 – Stirring rod
- 364140 – Stirring rod 10er pack
- 19802218 – Blue lid