BIOWAVE BioWraps Thigh and Calf Biowrap User Guide

September 9, 2024
BioWave

BioWraps Thigh and Calf Biowrap

“`html

Specifications:

  • Product: Thigh & Calf Biowrap Silver Compression Electrode
    Garment

  • Technology: Advanced Silver Fabric Electrode

  • Compatibility: BioWaveHOME & BioWavePRO

  • Sizes Available:

    • S/M: 12 – 24 inches circumference around thigh or calf
    • L/XL: 24 – 36 inches circumference around thigh or calf

Product Usage Instructions:

Quick Set Up and Use:

  1. Ensure the 4 Snap Cables are snapped onto each of the 2 snaps
    on the outside of the BioWrap.

  2. Connect the blue connectors of each BioWrap cable to your
    BioWave Leadwire Cable.

  3. Clean the skin with soap and water where the silver fabric
    electrodes will touch.

  4. Apply a pea-sized amount of conductive Theracream on your skin
    where the silver fabric electrodes will touch.

  5. Put on the BioWrap, ensuring it provides strong but comfortable
    compression over the silver fabric locations.

  6. Connect the leadwire cable to the BioWave device and turn it
    on.

  7. Adjust treatment intensity using the PLUS (+) button for
    comfort.

  8. After the 30-minute treatment, disconnect the cables, remove
    the BioWrap, and store it in a cool, dry place.

Directions for Use:

  1. Apply BioWrap Conductive Cream on the skin where the silver
    fabric electrodes will touch.

  2. Attach the 4 Snap Cables to the snaps on the back of the
    BioWrap and connect the Leadwire cable to the blue connectors.

  3. Thread and secure the elastic straps through the D-rings for
    proper fit and compression.

  4. Ensure the silver fabric electrodes are centered over the pain
    areas on your thigh.

FAQ:

Q: Can I use the Thigh & Calf Biowrap without BioWaveHOME

& BioWavePRO devices?

A: No, this Biowrap is specifically designed for use with
BioWaveHOME & BioWavePRO devices for optimal functionality and
safety.

Q: How should I clean the Thigh & Calf Biowrap?

A: It is recommended to spot clean the Biowrap with a damp cloth
and mild soap. Do not immerse it in water or machine wash.

“`

Quick Reference Guide

THIGH & CALF BIOWRAP
SILVER
Compression Electrode Garment
Utilizes Advanced Silver Fabric Electrode Technology
See video at
https://biowave.com/support-biowraps/

NOTE: ONLY for use with BioWaveHOME & BioWavePRO.

Questions? Call us!
1-877-BIOWAVE · (1-877-246-9283)
support@biowave.com

Size

Circumference Around Thigh or Calf

S / M*

12″ – 24″

L / XL

24″ – 36″
*Calf requires S/M size

This Package Contains
· One Extremity BioWrap Compression Electrode Garment ­ for use on RIGHT or LEFT thigh or calf. · One 4.2 oz. Bottle of Theracream Conductive Cream
Inside View
Top

Elastic Straps

Built-in Silver Fabric Electrodes treat any side of the upper or lower leg

Bottom

Elastic Strap Provides Compression Around the Upper Thigh or Calf Elastic Strap Provides Compression around the Lower Thigh or Calf

Outside View
Top

D-Rings for the Elastic Straps

Velcro Closures

Bottom Snap Locations for the 4″ Snap Cables

Channels for the 4″ Snap Cables

Quick Set Up and Use
See video at biowave.com/support/biowrap
1. The 4″ Snap Cables should already be snapped onto each of the 2 snaps on the outside of the BioWrap.
2. Plug the blue connectors at the other end of each BioWrap cable into your BioWave Leadwire Cable.
3. Clean your skin with soap and water using a washcloth in the locations that the silver fabric will touch your skin.
4. Squeeze a separate pea size amount of conductive Theracream onto your fingertip and rub into your skin thoroughly over the entire area that the silver fabric electrodes will touch your skin. Use a pea size amount of cream for each silver fabric electrode.
5. This BioWrap can be attached while sitting on the edge of a chair. See diagrams on opposite side to put on the BioWrap and velcro it closed firmly so there is strong but comfortable compression over the silver fabric locations.
6. Plug the single end of the leadwire cable into the BioWave device.
7. Press the power button to turn on the BioWave device.
8. Start the treatment by pressing the PLUS (+) button to increase intensity so the sensation felt is strong but still comfortable. As your body adapts to the electrical field and the sensation fades, keep pressing the PLUS (+) button to increase the intensity to maintain a strong but comfortable sensation at the treatment location.
9. When the 30-minute treatment is over, disconnect the BioWave Leadwire cable from the 4″ Snap Cables, remove the BioWrap from your body and store in a cool dry place.

Directions for Use

· These instructions are for attaching the Thigh & Calf BioWrap to the RIGHT or LEFT hamstring.
· This BioWrap can be used to treat any location around the thigh including hamstring, quadriceps, IT Band, abductor, adductor. The S/M Thigh & Calf BioWrap can also be used to treat any location around the calf including the calf muscle, pain from shin splints or pain on either side of the lower leg.
· The silver fabric electrodes must be oriented so they are directy over the location(s) of pain to be treated.

Squeeze a pea size amount of BioWrap Conductive Cream onto your finger tips and rub into your skin thoroughly in the two locations that the silver fabric
will touch the skin.

Both 4″ Snap Cables should already be snapped onto the 2 snaps on the back side of the BioWrap. Attach your BioWave
Leadwire cable to the two blue connectors on the
back of the BioWrap.

Insert top elastic strap through top D-ring.
Pull top elastic strap so BioWrap is tight.
Make sure silver fabric electrodes are centered over locations of pain on your thigh.

Hamstring Example

Pull top strap across front of thigh so BioWrap is tight. Secure top Velcro strap across back of thigh to close.
Insert bottom elastic strap through bottom D-ring.
Pull bottom elastic strap so BioWrap is tight.

Pull bottom strap across front of thigh so BioWrap is tight.
Secure bottom Velcro strap across of back of thigh to close.

The silver fabric electrodes should be centered over locations of pain on your thigh. Both the top and
bottom strap should provide a similar feeling of compression
beneath each strap.

Questions?
biowPualklvnloeewe.ecsroostB+m riao1p-W/8asracu7pr7opis- spsBtifogrIohOrntt.tW/obfAioVwEr(a+p1-8·77s-u2p4p6o-r9t@ 28b3i)owave.com

Directions for Use

· These instructions are for attaching the Thigh & Calf BioWrap to the RIGHT or LEFT quadriceps.
· This BioWrap can be used to treat any location around the thigh including hamstring, quadriceps, IT Band, abductor, adductor. The S/M Thigh & Calf BioWrap can also be used to treat any location around the calf including the calf muscle, pain from shin splints or pain on either side of the lower leg.
· The silver fabric electrodes must be oriented so they are directy over the location(s) of pain to be treated.

Squeeze a pea size amount of BioWrap Conductive Cream onto
your finger tips and rub into your skin thoroughly in the two locations that the silver fabric
will touch the skin.

Both 4″ Snap Cables should already be snapped onto the 2 snaps on the back side of the BioWrap. Attach your BioWave
Leadwire cable to the two blue connectors on the
back of the BioWrap.

Insert top elastic strap through top D-ring.
Pull top elastic strap so BioWrap is tight.
Make sure silver fabric electrodes are centered over locations of pain on your thigh.

Quadriceps Example

Pull top strap across back of thigh so BioWrap is tight. Secure top Velcro strap across front of thigh to close.
Insert bottom elastic strap through bottom D-ring.
Pull bottom elastic strap so BioWrap is tight.

Pull bottom strap across back of thigh so BioWrap is tight.
Secure bottom Velcro strap across front of thigh to close.

The silver fabric electrodes should be centered over
locations of pain on your thigh. Both the top and bottom strap should provide a similar feeling of compression beneath each
strap.

Questions?
bioPwulal lvoewe.cr ostmrap/sacurposps ofrorntt/obf iowrap · support@biowave.com knee so B+io1W-8ra7p7is-BtigIhOt.WAVE (+1-877-246-9283)

Directions for Use

· These instructions are for attaching the Thigh & Calf BioWrap to the RIGHT or LEFT calf.
· This BioWrap can be used to treat any location around the thigh including hamstring, quadriceps, IT Band, abductor, adductor. The S/M Thigh & Calf BioWrap can also be used to treat any location around the calf including the calf muscle, pain from shin splints or pain on either side of the lower leg.
· The silver fabric electrodes must be oriented so they are directy over the location(s) of pain to be treated.

Squeeze a pea size amount of BioWrap Conductive Cream onto your finger tips and rub into your skin thoroughly in the two locations that the silver fabric
will touch the skin.

Both 4″ Snap Cables should already be snapped onto the 2 snaps on the back side of the BioWrap. Attach your BioWave
Leadwire cable to the two blue connectors on the
back of the BioWrap.

Insert top elastic strap through top D-ring.
Pull top elastic strap so BioWrap is tight.
Make sure silver fabric electrodes are centered over locations of pain on your calf.

Calf Example

Pull top strap across front of lower leg so BioWrap is tight. Secure top Velcro
strap across back of calf to close.
Insert bottom elastic strap through bottom D-ring.
Pull bottom elastic strap so BioWrap is tight.

Pull bottom strap across front of lower leg so BioWrap is tight.
Secure bottom Velcro strap across back of calf to close.

The silver fabric electrodes should be centered over locations of pain on your
calf. Both the top and bottom strap should provide a similar
feeling of compression beneath each strap.

Questions?
biowPualklvnloeewe.ecsroostB+m riao1p-W/8asracu7pr7opis- spsBtifogrIohOrntt.tW/obfAioVwEr(a+p1-8·77s-u2p4p6o-r9t@ 28b3i)owave.com

TROUBLESHOOTING
My BioWave shows a connection error but my BioWrap is plugged in
If your BioWave device shows a connection error when you are fully connected to the BioWrap, remove the BioWrap and follow these steps:
If you forgot to apply Conductive Cream:
· CLEAN YOUR SKIN. Remove BioWrap and clean your skin with soap and water in the region that the silver fabric electrodes touch your skin. Shave any excessive hair in the locations that the silver fabric electrodes touch your skin.
· APPLY A PEA-SIZE AMOUNT OF CONDUCTIVE CREAM directly to skin areas that will be against each of the silver fabric electrodes inside the BioWrap. Gently massage Conductive Cream into the skin allowing it to absorb, before putting on the BioWrap.
If you applied too much Conductive Cream (more than a pea-size amount):
· CLEAN YOUR SKIN. Remove BioWrap and use a dry towel to remove all excess Conductive Cream from your skin and any excess cream from the silver fabric in the BioWrap. Next, reattach the BioWrap and begin the treatment.
The treatment feels stronger under one electrode versus the other
· If the treatment feels stronger under the upper electrode as compared to under the lower electrode, it may be because the upper elastic strap around the thigh or calf is too tight and/or the lower elastic strap around the thigh or calf is too loose. Adjust the compression of each strap to increase or decrease the relative sensation during the treatment.
· Tighter Compression and/or More Conductive Cream = Stronger Treatment Sensation. · Looser Compression and/or Less Conductive Cream = Less Treatment Sensation.
Questions? Call us!
Call technical support: 1-877-BIOWAVE (1-877-246-9283)
biowave.com/support/biowrap support@biowave.com

IMPORTANT
Contraindications When Not to Use BioWraps
· DO NOT use if you have an implanted cardiac pacemaker. · DO NOT place the silver fabric electrodes in the BioWrap over open
wounds, broken skin, rashy skin or sensitive skin areas like sunburned skin.
NOTE: · BioWrap Wearables MAY BE PLACED over or in the proximity of implanted
orthopedic hardware, including total joint replacements, anchors, plates, rods, screws, pins and clips. · Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.
Cleaning
· DISCONNECT and REMOVE the two 4″ SNAP CABLES before hand washing the BioWrap.
· HAND WASH using mild soap and RINSE in cold water. · PAT DRY with a TOWEL, then let AIR DRY. · DO NOT wash in a washing machine. · DO NOT place in a dryer.

Need help?
Contact us at:
1-877-BIOWAVE · (1-877-246-9283) support@biowave.com

The information, including but not limited to text, graphics, images
and other material contained herein are for informational purposes only. Learn more at BioWave.com/terms-of-use

BioWave Corporation 8 Knight St., Suite 201 Norwalk, CT 06851
© 2024 BioWave Corporation Rev 4 – 240627

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals