QCORE ADDENDUM 3 Sapphire Multi-Therapy and Dedicated Infusion Pumps User Manual
- June 3, 2024
- QCORE
Table of Contents
- QCORE ADDENDUM 3 Sapphire Multi-Therapy and Dedicated Infusion Pumps
- Clause ‘Warning’
- Introduction; Clause ‘Product Overview and Indications’
- Components, Accessories, and Administration Sets
- Fundamental Concepts and Operations;
- Getting Started
- Using the Infusion Modes
- Basic Infusion Operations
- Options Menu: Configuring, Viewing and Testing
- Using Advanced Features
- Alarms and Troubleshooting
- Maintenance and Storage
- Technical specification
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
QCORE ADDENDUM 3 Sapphire Multi-Therapy and Dedicated Infusion Pumps
Clause ‘Warning’
Update: The following sentences were added to the end of the first paragraph:
Alternatively, administration sets from different manufacturers that are
regulatory cleared and labeled for use with the Sapphire infusion system, can
be used as well.
The list of regulatory cleared sets needs to be obtained from the official
manufacturer’s publication such as manufacturer’s website, catalog or any
other formally published document.
Using anything other than administration sets regulatory cleared and labeled
for use with Sapphire infusion system, may result in operation that is not
within the constraints and parameters of the device.
Note: The use of Administration sets manufactured by ICU Medical is
approved for sale through March, 2023.
Introduction; Clause ‘Product Overview and Indications’
Update: The third paragraph was updated with an added intended use – air
transportation
The pump is intended to be used by both licensed health care professionals in
a clinical environment, and home users in an ambulatory environment and in
pre-hospital medical air and ground transportation.
Safety and compliance information
Update: An additional standard was added to the Compliance and
Classification section
IEC 60601-1-12 – Requirements for medical electrical equipment and medical
electrical systems intended for use in the emergency medical services
environment.
Proper Use of the Pump
Administration Sets
Update: The warning was replaced with the following:
Use Q Core standard administration sets listed here or in Q Core’s approved
list of products:
http://www.qcore.com/. Alternatively, administration sets from different
manufacturers that are regulatory cleared and labeled for use with the
Sapphire infusion system, can be used as well. The list of regulatory cleared
sets needs to be obtained from the official manufacturer’s publication such as
manufacturer’s website, catalog or any other formally published document.
Using anything other than administration sets regulatory cleared and labeled
for use with Sapphire infusion system, may result in operation that is not
within the constraints and parameters of the device. Severe injury or death
may result from using sets other than those indicated in Q Core’s approved
list of products. For more information refer to Administration Sets
Note: The use of Administration sets manufactured by ICU Medical is
approved for sale through March, 2023.
Basic Infusion Safety Information
Update: The following sentences were added to the first paragraph:
When clamping the administration set, ensure the clamp is at least 20 cm (8
in) away from the pump, when possible.
Note that if the dose rate is beyond the pump resolution of 0.1mL/h
increments, the pump will increase or decrease the rate by up to 0.05mL/h.
This flow rate (mL/h) is presented on the running screen during infusion.
Administering Infusions: General Safety Precautions
Update: Three bullets were added to the warnings list:
- Air detection
- Air detection is an important safety feature of the Sapphire pump. If the air detection is disabled (OFF), use a set with an air-eliminating filter to prevent injury to the patient due to an air embolus.
- Air detection serves as a safety component. Disabling the air detection hinders the pump’s ability to alert on hazardous situations.
- Always ensure that the administration set is primed before starting an infusion
- The air detector working range when delivering fatty acids is 2%-20% lipids.
Warnings and Safety Precautions
Electromagnetic Compatibility
Update: the Electromagnetic Compatibility sub-clause has been rephrased:
The Sapphire pump is designed to conform with electromagnetic compatibility
(EMC) standard IEC 60601-1-2 and to operate accurately in conjunction with
other medical equipment which also meets the requirements of this standard. To
avoid electromagnetic interference that may affect the operation of the pump,
do not use the pump near sources of strong electric and magnetic interference
(EMI), such as MRI, CT, diathermy (deep heat treatment), electromagnetic
security systems (e.g metal detectors) and large electric motors.
Portable and mobile RF communications equipment, such as RF emitters, cellular
telephones, 2-way radios, Bluetooth™ devices, microwave ovens in close
proximity to this device may affect wireless communications with the Infusion
pump and/or the operation of the Infusion pump.
Some of these EMI sources (mostly RF emitters) may not be visible and the
device can potentially be exposed to fields from these EMI sources without the
user’s awareness.
Special precautions need to be exercised regarding EMC. These include:
-
Maintaining a minimum separation distance of 2 ½ ft (¾ m) between the Infusion pump system and portable/mobile RF communications equipment
-
Manage the electromagnetic environment to permit the device to perform as intended without disturbing other equipment.
-
Separate the device from all other electronic equipment. If the device must be used near other electrical equipment, monitor the equipment to ensure there is no electromagnetic interference.
-
Devices should not be used adjacent to or stacked with other equipment. If the device must be used adjacent to or stacked with other equipment, monitor the device to verify normal operation.
-
If you identify or suspect external RF sources or other equipment are influencing device operation (from known or unknown source), try to (as applicable) increase the pump’s distance from the EMI source, re-orient the device, relocate the device, connecting device to different outlet, contact the biomedical engineering department for additional guidelines concerning electromagnetic immunity or decrease emitting device output power (to 30 dBm).
-
Contact the biomedical engineering department for additional information in the service manual concerning operating devices near RF sources.
The EMC limits for the Medical Device Directive 93/42/EEC (EN301489-1/-17 IEC/EN 60601-1-2:2007) are designed to provide reasonable protection against harmful interference in a typical medical installation. The equipment generates uses and can radiate radio frequency energy, and if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices, which can be determined by turning the equipment Off and On, the user is encouraged to try to correct the interference by one or more of the following measures:- Reorient or relocate the receiving device
- Increase the distance separating between the equipment parts
- Connect the equipment to an outlet on a circuit different from that to which the other device(s) is connected
- Consult the manufacturer or field service technician for help
Components, Accessories, and Administration Sets
Using Pump Accessories
Update: the following information was added to the first paragraph, after
the first sentence:
Alternatively, administration sets from different manufacturers that are
regulatory cleared and labeled for use with the Sapphire infusion system, can
be used as well. The list of regulatory cleared sets needs to be obtained from
the official manufacturer’s publication such as manufacturer’s website,
catalog or any other formally published document.
Note: The use of Administration sets manufactured by ICU Medical is
approved for sale through March, 2023.
Q Core Approved Administration sets
Update: The warning was replaced with the following:
Use Q Core standard administration sets listed here or in Q Core’s approved
list of products:
http://www.qcore.com/. Alternatively, administration sets from different
manufacturers that are regulatory cleared and labeled for use with the
Sapphire infusion system, can be used as well. The list of regulatory cleared
sets needs to be obtained from the official manufacturer’s publication such as
manufacturer’s website, catalog or any other formally published document.
Using anything other than administration sets regulatory cleared and labeled
for use with Sapphire infusion system, may result in operation that is not
within the constraints and parameters of the device.
Note: The use of Administration sets manufactured by ICU Medical is
approved for sale through March, 2023.
Update: the following information was added below the approved Q Core
administration sets table:
All filters used in Q Core approved administration sets are air eliminating
filters.
Fundamental Concepts and Operations;
Working with the Main Display
Alphanumeric keypad’
Update: The first paragraph was updated:
In some instances, for example, defining a new Preset program or entering a
drug name, the keypad displays letters and symbols in addition to numbers.
Overview of Icons
Update: An icon was added to the icons table:
Icon | Meaning | Notes |
---|---|---|
Air detection is disabled (OFF) | No Air in Line alarm is triggered when the |
pump air detection is disabled (OFF). A technician authorization code is
required to enable or disable the air detection (this can only be set manually
on the pump and not by the DLE).
If air detection is disabled (OFF), use a set with an air-eliminating filter
to prevent injury. Always ensure the administration set is primed before
starting an infusion.
---|---|---
Enabling Special Features
A technician authorization code is required to enable/disable the following
additional features
Update: The following replaces the New Patient feature in the table:
Feature | Delivery Mode(s) | Description/Notes |
---|---|---|
New Patient | All | Allow users to associate an infusion with a patient and |
reset the Accumulated VI (accumulated volume infused)
Update: The following was added to the features table:
Feature | Delivery Mode(s) | Description/Notes |
---|---|---|
Air Detection | All, except Epidural | Determines whether the pump air detection |
is disabled (OFF) or enabled (ON) during infusion. This feature should be used when meeting the clinical practice and guidelines and coupled with an alternative method of eliminating air. When the air detection is disabled (OFF) the user is prompted to use a set with an air eliminating filter.
Getting Started
Connecting the Infusion Container to the Administration Set
Update: The ‘> To connect the container to the administration set:’ workflow was updated:
- Open the sterilized administration set package.
- Close the clamps and the AFFV to block the administration set. Ensure the clamp is located at least 20 cm (8 in) from the pump, when possible.
- Spike the administration set into the container.
Using the Infusion Modes
Continuous Mode
To begin a continuous infusion without drug library’
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
To begin a continuous infusion with a drug library
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
To begin a Multi-step infusion without drug library
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
To begin a Multi-step infusion with drug library
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
TPN Mode
To begin a TPN infusion
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
Intermittent Mode
To begin an Intermittent infusion without drug library
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
To begin an Intermittent infusion with drug library
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
PCA Mode
Infusion
Parameters: PCA Mode
Update: the following replaces the Boluses per 1 h or Total dose per 1h
in the table:
Parameter | Description |
---|---|
Bolus per 1h (or 4hrs) OR Total dose per 1hr (or 4hrs) | The maximum number of |
boluses OR the maximum dose that can be delivered during a 1 hour (or 4 hours)
period. (User with high authorization codes can set the 1 hour or 4 hours
parameters). The Total dose limit takes into account medication delivered via:
Continuous rate Yes
Demand Bolus Yes
Loading Dose Yes
Clinician bolus Yes
All doses, including boluses given by clinician, are taken into account.
When the Total dose limit is reached, the patient is locked out from
activating additional boluses.
To begin a PCA infusion without drug library
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
To begin a PCA infusion with drug library
Update: The following step was added after step #2:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
PCA Mode: Mid-infusion actions
Updates:
- The term Viewing Bolus History is replaced with the term Viewing Delivery History
- The following replaces the text in ‘To update parameters using the View/edit function key’ section, third bullet in step #5:
Delivery History : Displays summary of medication delivery events. For more information, refer to Viewing Delivery History section.
Epidural Mode
Infusion Parameters: PCEA Mode
Update: the following replaces the Boluses per 1 h or Total dose per 1h
in the table:
Parameter | Description |
---|---|
Bolus per 1h (or 4hrs) OR Total Dose per 1hr (or 4hrs) | The maximum number of |
boluses OR the maximum dose that can be delivered during a 1 hour (or 4 hours)
period. (User with high authorization codes can set the 1 hour or 4 hours
parameters). The Total dose limit takes into account medication delivered via:
Continuous rate Yes
Demand Bolus Yes
Boluses given by a clinician, are not taken into account for the Total dose
limit: Loading Dose No
Clinician bolus No
When the Total dose limit is reached, the patient is locked out from
activating additional boluses.
Infusion Parameters: Epidural Intermittent Mode
Update: the following replaces the Boluses per 1 h or Total dose per 1h
in the table:
Parameter | Description |
---|---|
Bolus per 1h (or 4hrs) OR Total dose per 1hr (or 4hrs) | The maximum number of |
boluses OR the maximum dose that can be delivered during a 1 hour (or 4 hours)
period. (User with high authorization codes can set the 1 hour or 4 hours
parameters). The Total dose limit takes into account medication delivered via:
Continuous rate Yes
Intermittent dose Yes
Demand Bolus Yes
Boluses given by a clinician, are not taken into account for the Total dose
limit: Clinician bolus No
When the Total dose limit is reached, the patient is locked out from
activating additional boluses.
Epidural Intermittent Mode: Mid-infusion Actions
Updates:
-
The term Viewing Bolus History is replaced with the term Viewing Delivery History
-
The following replaces the text in ‘To update parameters using the View/edit function key’ section, third bullet in step #5:
Delivery History: Displays summary of medication delivery events. For more information, refer to Viewing Delivery History section. Applies only when Epidural Intermittent with PCEA infusion was programmed. -
The following replaces the first paragraph in ‘Administering a Clinician Bolus’ section
A bolus of any amount (within the predefined range) can be delivered by clinicians who have a High authorization level code. A clinician bolus can be given only while the infusion is running. The lockout time is reset after delivering a clinician bolus. Clinician bolus is applicable only when programming Epidural Intermittent with PCEA.
Epidural Intermittent Mode: Mid-infusion Action
Administering a Clinician Bolus
Updates: the note at the end of the Sub-clause was replaced with the
following note
The RATE of the Clinician Bolus, as all other boluses in PCEA mode is defined
prior to programming the infusion. It can be set to 125 mL/h or 200 mL/h (for
more information, refer to Epidural Mode Options Menu on page 205).
Basic Infusion Operations
Starting New Infusions: Shortcuts
Repeating last infusion
Updates:
-
The first note was replaced with the following note:
When using Repeat Last Infusion option, the Delivery History, Accumulated VI parameter and the remaining Lockout Time are not cleared; instead, they continue counting from the previous infusion. For more information about Accumulated VI, and Delivery History, refer to Using the New Patient feature chapter. -
To repeat the last infusion’ section, the following step was added after step #1:
If a warning that the air detection is disabled (OFF) appears, ensure that a set with an air-eliminating filter is used and press OK.
Using a Preset Program
Update: ‘To start an infusion using Preset programs function’ section,
the following step was added after step #1:
If a warning that the air detection is disabled (OFF) appears, ensure that a
set with an air-eliminating filter is used and press OK.
Resuming Infusions After Pump Shutdown
Update: ‘To resume the infusion’ section was updated:
To resume the infusion:
- On the Attention screen, press OK.
- If a warning that the air detection is disabled (OFF) appears, ensure that a set with an air-eliminating filter is used and press OK.
- The Paused infusion screen appears, press Request continue to resume the infusion.
- On the attention screen press OK to confirm.
- The running screen appears.
Options Menu: Configuring, Viewing and Testing
Main Options: Overview
Update: the last sentence in the overview was replaced with the following
sentence:
The screen also provides access to testing components and viewing pump (as
opposed to infusion) parameters (e.g. System parameters, Event logs, and
Delivery History)
Managing Configuration Settings
View Menu
-
Update: View system parameters
-
The description of Single Air detector, Accumulated Air detector and Accumulated Threshold has been updated
These settings are associated with the amount of air that triggers an Air in Line alarm, when the air detection is enabled (ON). These settings can be modified by Technicians only. For more information, refer to the Service Manual. Note: While a non-epidural infusion is running at a rate of 4 mL/h or lower, the Single air detector switches automatically to ON. -
A setting was added below the accumulated threshold, “Air detection”.
This setting indicates the air detection is disabled (OFF), and it is displayed instead of the single and accumulated air settings mentioned above.
-
-
Update: The bolus history category in the table was replaced with the following:
Category| Description/Notes
---|---
Delivery History (PCA, PCEA and Epidural intermittent delivery modes only)| Provides a view of the boluses and the total amount of medication delivered during PCA, PCEA or PIEB infusion. The Delivery History is associated with a patient. For more information, refer to Viewing Delivery History section.
Managing Alarm Settings
Update: In the description of ‘Infusion near end’ option the note has
been removed
Viewing Bolus History
Update: Title and first paragraph were replaced with the following:
Viewing Delivery History
This screen, which appears only in PCA, PCEA and PIEB delivery modes, provides
a summary of all bolus-related events that occurred during a specified time
frame and the total amount of medication delivered throughout the treatment.
Note: To access the Delivery History during PCA, PIEB or PCEA infusions:
From the toolbar select View/edit, then, select Edit PCEA –> Delivery History.
When the pump is locked, the Delivery History can be accessed from the toolbar
without unlocking the pump. Delivery History information includes:
Update: The Bolus History information table was updated:
- Total Dose given parameter was added
- History period parameter was replaced with the text below
Name of Value| Description/Notes
---|---
Total Dose given| The total amount of drug delivered to the patient during a treatment through Boluses, Loading dose, Continuous rate, KVO if applied and Intermittent doses. When using Repeat Last Infusion, this value accumulates from the previous infusion/s.
Bolus History Period| The number of hours over which the displayed boluses occurred. The default history period is 1 hour, and it can be set from 1 hour up to the number of hours that the infusion has been running.
The setting can be modified by pressing >, entering a value using the keypad, and then pressing OK. - The note below the table was replaced with the following note
When using the Repeat Last Infusion option (for the same patient), the Delivery History, accumulated VI and lockout time are not cleared, they continue counting from the previous infusion.
Using Advanced Features
Using the New Patient Feature
Update: The entire clause was replaced with the following:
When the New Patient feature is enabled, and either a New Infusion or a PreSet
program is selected, the pump will prompt you to indicate whether the infusion
to be programmed is for a new patient or not. When selecting Repeat Last
Infusion the New Patient screen will not appear, and the pump will indicate
that the infusion to be repeated will be used for the last patient selected.
Note: The New Patient feature can be enabled/ disabled by technicians
only.
When a New Patient is selected, entries associated with the patient can be
tracked in the Event Log (Viewing the Event Log section). In addition, when
Repeat Last Infusion is used the pump calculates the accumulated volume
infused (Accumulated VI) for all infusions associated with the patient, and
the Delivery History. When a new patient is selected, the Accumulated VI and
Delivery History are automatically cleared (for more information, refer to
Monitoring the Accumulated Volume Infused
(Shift’s Total) section). The current accumulated Delivery history can be
viewed via the Options menu. When an infusion is running, the Delivery History
can be accessed via the View/Edit soft key in the toolbar (for more
information, refer to the Viewing Delivery History section).
To select a new patient:
- From the startup screen, select New Infusion or Preset Programs
- On the New Patient screen, select Yes.
Alarms and Troubleshooting
Level 3 Alarms
Update: Additional Alarms were added to the table.
Maintenance and Storage
Preventive Maintenance
Cleaning and Disinfecting the Pump
Update: Cleaning and Disinfection: Safety Precautions
Alarm Title | Displayed Text |
---|---|
Potential Air in Line | Press OK to test for air |
Pump Stopped | Please quit and then restart the infusion |
-
A safety precaution was added to the list:
Do not clean the pump with Bleach (8.25% concentration, mixed at 1 part bleach in 10 parts water), as deterioration may occur. -
The precaution note was replaced with the following precaution note:
Before using materials other than the products listed above for cleaning and disinfecting the Sapphire Infusion pump, make sure they are listed in Q Core’s official approved list of materials (published at www.qcore.com))
Update: Cleaning and Disinfection Procedure – one cleaning agent was added to the table:
Cleaning/Disinfecting Solution | Manufacturer |
---|---|
Virox® AHP 5 RTU | Diversey |
Alarm Testing
Updates:
-
The first paragraph was replaced with the following:
It is recommended to perform manual testing of the following alarms at least once a year. Alarm testing can be conducted as part of the yearly certification.
For the Sapphire Epidural pump manual alarm testing, refer to the testing protocols available for authorized technicians (for more information refer to the service manual). -
A note was added to the Air in Line test:
Note – To test the Air in Line alarm, ensure the air detection is enabled (ON) in the technician options. If the air detection disabled (OFF) icon is displayed and a warning message stating the air detection is disabled (OFF) appears when programing an infusion, the Air in Line alarm will not be triggered.
Battery Care Information
Update: The first paragraph (preceding the table) was replaced with the
following:
The Sapphire pump can operate on battery power, enabling operation of the pump
during an electrical power failure, during patient transport or during
ambulatory care.
When working on battery power (disconnected from main power supply) the
battery charge level icon, on the upper right corner of the indicators bar,
indicates remaining battery capacity. Check the status of the battery charge
level icon regularly:
Technical specification
Environmental specifications
Operating Conditions
Update: Transient conditions were added to the Temperature and Humidity
conditions:
Humidity : (15% to 90% at transient state)
Temperature: (-20°C to +50°C at transient state)
References