REFINE HP-5L Ultrasonic Scaler Handpiece Instruction Manual

August 22, 2024
Refine

REFINE HP-5L Ultrasonic Scaler Handpiece Instruction Manual

INSTRUCTION MANUAL For HANDPIECE OF ULTRASONIC SCAL ER

Description of device

Description
Handpiece Of Ultrasonic Scaler is the accessory of scalers, It is the handheld energy transformation part of ultrasonic piezo scaler, by using piezoceramics to transfer electric energy to mechanical energy to eliminate the dental calculus.

Models
Note: The exact model see package labels.

Product structure
The HANDPIECE is composed of cover and ultrasonic transducer.

Scope of applicaton
The HANDPIECE is used for the dental calculus elimination and root canal cleaning.

Contraindication
The hemophilia disease patient is not allowed to use this equipment.
The patients with heart pacemaker are forbidden to use this equipment.
The doctors with heart pacemaker are forbidden to use this equipment.
The heart disease patient, pregnant woman and children should be cautious to use the equipment.

The classification of the device
Operating mode: Continuous operation
Protection degree against water(used on the pedal): IPX1
Degree of safety of application in the presence of a Flammable Anaesthetic Mixture with air or with Oxygen or Nitrous Oxide :
Equipment not suitable for being used in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide

The main technical specifications
Power source Input: 190VAC-230VAC
Output primary tip Vibration excursion: < 200μm
Output tip Vibration frequency. 28kHz士5kHz
Output half-excursion force: 0.5N-5N
Tip’s output power: 3W-20W
Water pressure: 0.01MPa-0.5MPa

Working condition
Environmental temperature +5°C-+40°C
Relative humidity: 30%-75%
Atmospheric pressure: 70kPa-106kPa

Installation of the product

Installation of handpiece
Open the packing box, make sure that all the parts and accessories are complete according to the packing list. Contact with dealer if necessary.
Keep dry the handpiece connector and cable plug socket before installing the handpiece.
Make sure that the bottom jack point is at the plug, make the handpiece fixed to the cable corr e ctly. (Picture 1)

Installation of tips
Fix the tip to handpiece by torque wrench. (Picture 2)

Please use “REFINE” brand scaling lips Notes: The detachable handpiece 1s a high-tech device which should be used carefully. Don’t use too much torque force to, screw the tips, by torque wrench. In order to avoid the vibration by stepping on the foot switch by accident, unscrew the scaling tips after each operation.

Product function and operation

Scaling function
Connect the handpiece and the connector of cable correctly, and adjust the water and vibration intensity to a suitable level.
The handpiece can be handled in the same gesture as a pen in hand.
Select a suitable scaling tip as you need, screw it on the handpiece lightly by the torque wrench.
During the clinical treatment, don’t make the end of tip touch the teeth vertically.
Don’t make the tip overexert on the surface of the teeth in case of hurting the teeth and damaging the lip.
After finishing operation, keep the machine working for 30 seconds on the water supply condition in order to clean the handpiece and the scaling tip.

Endo funtion

Usage process
a) Fix endochuck to handpiece by endo wrench.
b) Unscrew the screw cap on the endochuck.
c) Put the ultrasonic file into the hole in the front of endochuck.
d) Screw down the screw cap with endo wrench to tight up the ultrasonic file.
e) Press option key, turn to endo root canal cleaning function.
f) Press option key, turn the water and vibration intensity to a suitable level.
g) Put the ultrasonic file into the patient’s root canal slowly, step on the foot pedal, then make endo root canal cleaning. During the treatment, turn up the power gradually according to the needs.

Notes
a) When fixing endochuck, it must be screwed down.
b) The screw on the endochuck must be screwed down.
c) Don’t press it too much when the ultrasonic file in root canal.
d) Don’t start up the foot pedal until the ultrasonic file is in root canal.
e)The power range is supposed from 1st to 5th grades.
f) Please sterilize the endo chuck and handpiece after operation.

Instruction for main components of detachable handpiece

Nipple: The nipple can be removed, for easy cleaning.
Light pipe: light guide, can be autoclaved under high temperature and pressure.
LED lamp: illuminate the oral area, detachable and sterilized.
Handpiece: The main part of the whole handpiece, can be autoclaved under the high temperature and pressure.

Notes

Don’t knock or rub the handpiece.
Don’t screw or unscrew the scaling lip and endochuck when stepping on the foot pedal.
The handpiece, scaling tip, torque wrench, endo wrench and endochuck must be sterilized before treatment.
The internal screw thread of the scaling tips produced by some manufactures is coarse, rusty and collapsed. This will damage the external screw thread of the handpiece irretrievably. Please use “REFINE” brand scaling tips.
We are only responsible for the safety on the following cond仆ions: The maintenance, repair and modification are made by the manufacturer or the authorized dealer.

Trouble shooting

If the problem still can’t be solved, please contact with local dealer or manufacturer.

Cleaning and sterilization

Clean and sterilize the handpiece after using it.
Remove the nipple, light pipe, and LED lamp before autoclaved, clean by pure water and dry ii by three-way syringe (Picture 3)
Autoclaved in high temperature/ pressure :
121°C/1bar(0.1MPa),
134°C/2 2bar(O 22MPa),
Using sterile towel or sterilized bags to pack the handpiea, nipple, light pipe and LED lamp before autoclaved, install them as showninpicture3 after cooling down.

Notice
Clear the cleaning liquid in the handpiece with three-way syringe before sterilicalion.
Be sure that the scaling lip has been unscrewed from the handpiece and it cannot be sterilized with others.
Please confirm whether the outer of the handpiece is damaged during the treatment or sterilication.
Please don’t smear any protective oil on the surface of handpiece.
Please take out the handpiece, nipple, light pipe, LED lamp as Picture 3 and sterilize separately.
There are two waterproof “O” rings at the end of handpiece Please lubricate them with dental lube frequently, as sterilication and repeated pulling and inserting will reduce their using life Change a new one once it is damaged or worn excessively.

The following sterilizing methods are forbidden:
Put handpiece into any liquid for boiling.
Dip handpiece in disinfectors such as iodine, alcohol and glutaraldehyde.
Put handpiece into oven or microwave oven for baking.
Put handpiece into ultrasonic cleaner for cleaning.

Transportation, Storage And Maintenance

Transportation
Excessive impact and shake should RF-USH-M015 Edition V1 6 Modify 20230808 be prevented during transportation Lay it carefully and lightly and don’t invert it.
Don’t put it together with dangerous goods during transportation.
Avoid solarization and getting wet in rain or snow during transportation.

Storage
Don’t store the machine together with the articles which are combustible, poisonous, caustic, or explosive.
This equipment should be stored in a room where the relative humidity is 10%- 93%, atmospheric pressure is 70kPa-106kPa, and the temperature is -20°C-+55°C.

Maintenance
The equipment should be handled carefully and lightly. Be sure that it is far from the vibration, and installed or kept in a cool, dry and ventilated place.

Environmental Protection

There is not any harm factor in our products You can deal with it based on the local law.

After Service

We offer one year’s free repair to the equipment according to the warranty card. The repair of the equipment should be carried out by our professional technician We are not responsible for any irretrievable damage caused by the non-professional person.

Symbol instruction

Statement

All rights of modifying the product are reserved to the manufacturer without further notice The pictures are only for reference. The final interpretation rights belong to GUILIN REFINE MEDICAL INSTRUMENT CO , LTD The industrial design, inner structure, etc, have been claimed for several patents by REFINE, any copy or fake product responsibilities.

Before installing the handpiece, make sure that the handpiece connector and the cable plug socket are completely dry.

website: htts://www.alandental.com

RF-USH-M015 Edition V1 6 Modify 20230808

Documents / Resources

| REFINE HP-5L Ultrasonic Scaler Handpiece [pdf] Instruction Manual
HP-5L Ultrasonic Scaler Handpiece, HP-5L, Ultrasonic Scaler Handpiece, Scaler Handpiece, Handpiece
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals