Refine RF-USH-M015 Handpiece Of Ultrasonic Sclaer Instruction Manual
- June 21, 2024
- Refine
Table of Contents
INSTRUCTION MANUAL
For HANDPIECE OF
ULTRASONIC SCALER
Description of device
1.1 Description
Handpiece Of Ultrasonic Scaler s the accessory of scalers, It s the handheld
energy transformation part of ultrasonic piezo scaler, by using piezoceramics
to transfer electric energy to mechanical energy to eliminate the dental
calculus.
1.2 Models
Note: The exact model see package labels.
1.3 Product structure
The HANDPIECE s composed of cover and ultrasonic transducer.
1.4 Scope of applicaton
The HANDPIECE s used for the dental calculus elimination and root canal
cleaning.
1.5 Contraindication
1.5.1 The hemophilia disease patient is not allowed to use this equipment.
1.5.2 The patients with heart pacemaker are forbidden to use this equipment.
1.5.3 The doctors with heart pacemaker are forbidden to use this equipment.
1.5.4 The heart disease patient, pregnant woman and children should be
cautious to use the equipment.
1.6 The classification of the device
1.6.1 Operating mode: Continuous operation
1.6.2 Protection degree against water(used on the pedal): IPX1
1.6.3 Degree of safety of application in the presence of a Flammable
Anaesthetic Mixture with air or with Oxygen or Nitrous Oxide : Equipment not
suitable for being used in the presence of a flammable anaesthetic mixture
with air or with oxygen or nitrous oxide
1.7 The main technical specifications
1.7.1 Power source Input: 190VAC-230VAC
1.7.2 Output primary tip Vibration excursion: < 200µm
1.7.3 Output tip Vibration frequency: 28kHz±5kHz
1.7.4 Output half-excursion force: 0.5N-5N
1.7.5 Tip’s output power: 3W-20W
1.7.6 Water pressure: 0.01 MPa-0.5MPa
1.8 Working condition
1.8.1 Environmental temperature +5°C-+40°C
1.8.2 Relative humidity: 30%-75%
1.8.3 Atmospheric pressure: 70kPa-106kPa
Installation of the product
2.1 Installation of handpiece
2.1.1 Open the packing box, make sure that all the parts and accessories are
complete according to the packing ist. Contact with dealer if necessary.
2.1.2 Keep dry the handpiece connector and cable plug socket before installing
the handpiece.
2.1.3 Make sure that the bottom jack point is at the plug, make the handpiece
fixed to the 2.2 Installation of tips
2.2.1 Fix the tip to handpiece by torque wrench. (Picture 2) 2.2.2 Please use
“REFINE” brand scaling tips.
Notes: The detachable handpiece is a high-tech device which should be
used carefully. Don’t use too much torque force to, screw the tips, by torque
wrench. In order to avoid the vibration by stepping on the foot switch by
accident, unscrew the scaling tips after each operation.
Product function and operation
3.1 Scaling function
3.1.1 Connect the handpiece and the connector of cable correctly, and adjust
the water and vibration ntensity to a suitable level.
3.1.2 The handpiece can be handled n the same gesture as a pen in hand.
3.1.3 Select a suitable scaling tip as you need, screw t on the handpiece t
ghtly by the torque wrench.
3.1.4 During the clinical treatment, don’t make the end of tip touch the teeth
vertically.
3.1.5 Don’t make the tip overexert on the surface of the teeth in case of
hurting the teeth and damaging the tip.
3.1.6 After finishing operation, keep the machine working for 30 seconds on
the water supply condition n order to clean the handpiece and the scaling tip.
3.2 Endo funtion
3.2.1 Usage process
a) Fix endochuck to handpiece by endo wrench.
b) Unscrew the screw cap on the endochuck.
c) Put the ultrasonic file nto the hole in the front of endochuck.
d} Screw down the screw cap with endo wrench to tight up the ultrasonic file.
e) Press option key, turn to endo root canal cleaning function.
f) Press option key, turn the water and vibration ntensity to a suitable
level.
g) Put the ultrasonic file into the patient’s root canal slowly, step on the
foot pedal, then make endo root canal cleaning. During the treatment, turn up
the power gradually according to the needs.
3.2.2 Notes
a) When fixing endochuck, it must be screwed down.
b) The screw on the endochuck must be screwed down.
c) Don’t press it too much when the ultrasonic file n root canal.
d} Don’t start up the foot pedal until the ultrasonic file s in root canal.
e)The power range is supposed from 1st to 5th grades.
f) Please sterilize the endo chuck and handpiece after operation.
3.3 Instruction for main components of detachable handpiece
3.3.1 Nipple: The nipple can be removed, for easy cleaning.
3.3.2 Light pipe: ight guide, can be autoclaved under high temperature and
pressure.
3.3.3 LED amp: illuminate the oral area, detachable and ster lized.
3.3.4 Handpiece: The main part of the whole handpiece, can be autoclaved under
the high temperature and pressure.
Notes
4.1 Don’t knock or rub the handpiece.
4.2 Don’t screw or unscrew the scaling tip and endochuck when stepping on the
foot pedal.
4.3 The handpiece, scaling tip, torque wrench, endo wrench and endochuck must
be sterilized before treatment.
4.4 The nternal screw thread of the scaling tips produced by some manufactures
is coarse, rusty and collapsed. This will damage the external screw thread of
the handpiece irretrievably. Please use “REFINE” brand scaling tips.
4.5 We are only responsible for the safety on the following conditions: The
maintenance, repair and modificat on are made by the manufacturer or the
authorized dealer.
Trouble shooting
Fault | Possible causes | Solutions |
---|---|---|
Handpiece doesn’t work | The tip is loosened | Tighten it |
Something wrong with the Handpiece | Take out the handpiece and send it back |
for repair
Something wrong with the Cable| Contact our dealers or us
The connect plug between the cable and circuit board is loosened| Contact our
dealers or us
There is no spray when getting through the electricity| The water control
switch is not on| Turn on the water control switch
No pressure or low pressure| Check the pressure
There is no water when getting through the electricity| The water control
switch is not on| Turn on the water control switch
There is impurity in the sdenoid valve| Contact our dealers or us
Water way clogs| Clean the water line by three-way syringe
Low vibration| The tip hasn’t been screwed on the handpiece tightly| Screw the
tips tightly by torque wrench
The tip is shaken loose by vibration| Screw on the tip tightly
The coupling between the handpiece and the cable isn’t dry| Dry it, especially
the water between handpiece and connector
The tip is worn out or deformed| Change another tip
The handpiece generates heat| The water control switch is in a low setting|
Turn the water control switch to a higher grade
LED light doesn’t work| Poor contact| Contact tightly
Something wrong with LED light| Change a new one
LED lamp installed backwards| Please install the “+” of the LED lamp to the
“+” of the handpiece
There is water seeping from the coupling between the handpiece and the cable|
The waterproof „0 ring is damaged| Change a new “0 ” ring
If the problem st ll can’t be solved, please contact with ocal dealer or manufacturer.
Cleaning and sterilization
6.1 Clean and sterilize the handpiece after using t.
6.2 Remove the nipple, ight pipe, and LED amp before autoclaved, clean by pure
water and dry it by three-way syr nge. (Picture 3)
6.3 Autoclaved in high temperature/ pressure:
6.3.1121°C/1bar(0.1MPa);
6.3.2 134 °C/2.2bar(0.22MPa);
6.3.3 Using sterile towel or sterilized bags to pack the handpiea, nipple,
ight pipe and LED lamp before autoclaved, nstall them as showninpicture3 after
cooling down.
6.4 Notice
6.4.1 Clear the cleaning iquid n the handpiece with three-way syringe before
sterilication.
6.4.2 Be sure that the scaling tip has been unscrewed from the handpiece and
it cannot be ster lized with others.
6.4.3 Please confirm whether the outer of the handpiece s damaged during the
treatment or sterilication.
6.4.4 Please don’t smear any protective oil on the surface of handpiece.
6.4.5 Please take out the handpiece, nipple, ight pipe, LED lamp as Picture 3
and sterilize separately.
6.4.6 There are two waterproof “O” rings at the end of handpiece. Please
lubricate them with dental lube frequently, as sterilication and repeated
pulling and nserting will reduce their using life. Change a new one once it is
damaged or worn excessively.
6.5 The following sterilizing methods are forbidden:
6.5.1 Put handpiece nto any iquid for boiling.
6.5.2 Dip handpiece n disinfectors such as odine, alcohol and glutaraldehyde.
6.5.3 Put handpiece into oven or microwave oven for baking.
6.5.4 Put handpiece into ultrasonic cleaner for cleaning.
Transportation, Storage And Maintenance
7 .1 Transportation
7.1.1 Excessive impact and shake should be prevented dur ng transportation.
Layt carefully and lightly and don’t nvert it.
7 .1.2 Don’t put t together with dangerous goods during transportation.
7 .1.3 Avoid solarization and getting wet rain or snow during transportation.
7.2 Storage
7.2.1 Don’t store the machine together with the articles which are
combustible, poisonous, caustic, or explosive.
7.2.2 This equipment should be stored n a room where the relative humidity is
10%- 93%, atmospheric pressure is 70kPa-106kPa, and the temperature s
-20°C-+55°C.
7.3 Maintenance
The equipment should be handled carefully and ightly. Be sure that it s far
from the ibration, and nstalled or kept in a cool, dry and ventilated place.
Environmental Protection
There s not any harm factor in our products. You can deal with it based on the
local law.
After Service
We offer one year’s free repair to the equipment according to the warranty
card. The repair of the equipment should be carried out by our professional
technician. We are not responsible for any irretrievable damage caused by the
non-professional person.
Symbol instruction
| Caution
---|---
| Refer to instruction manual/ booklet
| Manufacturer
|
According to the type of protection against electric shock: CLASS I EQUIPMENT
| Handle with care
| Keep dry
| Used indoor only
| Can be autoclaved
| Recovery
| Atmospheric pressure for storage 70kPa-106kPa
| Temperature imitation -20°C-+55°C
| Humidity limitation 10%-93%
Statement
All rights of modifying the product are reserved to the manufacturer without
further notice. The pictures are only for reference.
The final nterpretation rights belong to GUILIN REFINE MEDICAL INSTRUMENT CO.,
LTD. The ndustrial design, nner structure, etc, have been claimed for several
patents by REFINE, any copy or fake product must take egal responsibilities .
Before nstalling the handpiece, make sure that the handpiece connector and the
cable pluq socket dry.
Guilin Refine Medical Instrument Co., Ltd.
No.8-3, Information Industrial Park, High-Tech Zone,
Qixing District, Guilin, Guangxi, 541004, P.R. China
Tel: +86-773-7796686 Email:refine@refine-med.com
Website: http://Avww.refine-med.com
MedNet EC-REP GmbH
Borkstrasse 10 – 48163 Muenster – Germany
RF-USH-M015 Edition: V1 6 Modify: 20230808
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>