REFINE Swan Dental LED Curing Light Instruction Manual

June 1, 2024
Refine

REFINE Swan Dental LED Curing Light

Specifications

  • Dimensions: 86mm*223mm
  • Net weight: 265g
  • Classified by power supply: Powered by rechargeable lithium battery
  • Input: Rechargeable lithium battery Power supply (Adapter)
  • Performance of LED lamp: Light Intensity

Product Usage Instructions

Safety Precautions

  • Ensure you read and understand the safety precautions before using the equipment. Avoid use if you are a heart disease patient, pregnant woman, child, or allergic to blue light.

Turn On/Off

  • The device can be turned on/off using the specified buttons.
  • Refer to the manual for detailed instructions.

Modes and Time Setting

  • The device has different modes like TURBO, NORMAL, ORTHO, and CHECK modes with varying light intensities and durations. Select the appropriate mode for your requirements.

Installation and Demounting

  • Insert the large end of the light guide into the mounting hole of the curing light ensuring a proper fit without any oblique positioning.

FAQ

Q: What should I do if an accessory is damaged?

A: Depending on the damaged accessory (power adapter, charging base, battery, or light hood), follow the replacement instructions provided in the manual.

Q: How do I know if the LED lamp is operating correctly?

A: When operating correctly, the LED lamp will light up, indicating its good condition. Ensure it matches the wavelength of dental resin materials used in clinics like 3M and Dentsply.

Safety Precautions

  1. This product is intended for use in hospitals and dental clinics only. The use of the product must comply with the requirements of the relevant operating specifications and relevant regulations of the medical department. Users must be professionally trained and qualified as dentists or technicians. Do not look directly into the light output. Patients, clinicians, and assistants are always suggested to wear protectors (safety goggles, face shield…) to avoid crossing infection.
  2. The products are reusable: Before each use, place a new barrier sleeve over the curing light Swan. The hygienic barrier sleeve is custom-fitted to the curing light and keeps the surface of the curing light clean. The barrier sleeve helps prevent cross-contamination, helps keep dental composite material from adhering to the surface of the lens and curing light, and prevents discoloration and corrosion from cleaning solutions. The curing lights must be cleaned and sanitized with appropriate cleaning and/or sanitizing agents after each patient. The sleeve is not equipped with a curing light. Please note that the sleeve must meet the requirements of relevant medical device regulations and it’s disposable.
  3. During the operation, the light should be aimed straight at the composite resin to ensure the effect of solidification.
  4. Please use the original light hood and follow the right installation ways to avoid the damage of blue light to the eyes.
  5. Please do not use other than the specified voltage to recharge the battery. Only the original pedestal charger, adapter, and Lithium battery could be used, because other brand pedestal chargers, adapters, and Lithium batteries are likely to damage the circuit.
  6. Please recharge the battery at least 4 hours before first time usage. It’s forbidden to connect the charger to the base for a long time while not charging and it’s not allowed to use the curing light while charging.
  7. It is forbidden to touch the charging connector with metal or other conductor, to avoid damage to the circuit of charge or the battery.
  8. Please recharge the battery in a cool and ventilated room. Please press the main unit to the buckle of the charging base, to avoid a poor connection.
  9. It is forbidden to self-taking-apart the battery, in order not to result in short-circuit or leakage.
  10. It is forbidden to extrude, shake, or rock the battery. The Lithium battery is forbidden to be in short-circuiting situations and it is forbidden to put the battery with metal or other conductors.
  11. The instrument has electromagnetic interference. Do not use it around electronic surgery, and use the instrument with caution in an environment with strong electromagnetic interference.
  12. This product does not contain toxic or hazardous substances and it is discarded under the relevant laws and regulations on waste medical devices. Please deal with the replaced battery according to national regulations
  13. As a professional manufacturer of medical instruments, we are only responsible for the safety on the following conditions:
  • The maintenance, repair, and modification are made by the manufacturer or the authorized dealer.
  • The changed components are original of our company and operated correctly according to the instruction manual.

Product contraindications:

The heart disease patient, pregnant women, children, and the person who are allergic to blue light should be cautious about using this equipment.

Symbol instruction

REFINE-Swan-Dental-LED-Curing-Light-FIG-1 REFINE-Swan-Dental-LED-Curing-Light-FIG-2

  • 1 Product introduction
  • The curing light manufactured by Guilin Refine Medical Instrument Co., Ltd. uses the principle of light radiation to irradiate photosensitive resin to quickly cure it. This product is intended for use in hospitals and dental clinics and it’s used for prevention and treatment of dental complications of adults or children.
  • Scope of application: suitable for dentistry, used to cure light-curing materials.
  • Product composition: main unit, LED light, light hood, charging base, battery and power adapter, etc.

The features of this product are:

  • A variety of different working modes and working hours can be set.
  • Constant light power output and the decrease in battery power will not affect the curing effect.
  • Large-capacity battery. While fully charged, under 10 seconds of light per time, it can be used continuously for more than 500 times.
  • Product schematicREFINE-Swan-Dental-LED-Curing-Light-FIG-3

Replacement instructions for accessories

Accessories Period of replacement Method of replacement
Power adapter Damaged Replace directly according to the requirements of the

manual.
Charging base| Damaged| Replace directly according to the requirements of the manual.
Battery| Damaged| Take out the damaged battery and replace it with a new one.
Light hood| Damaged| Replace directly according to the requirements of the manual.

Technical parameters

Technical Parameters of each model (table 1)

Dimensions *Φ86mm223mm**
Net weight 265g
Classified by power supply Powered by rechargeable lithium battery
Input Charging: DC5V 1A  Working: 3.6V lithium battery(ICR 18650)
Rechargeable lithium battery Battery model: ICR 18650

Battery voltage and capacity: 3.6V/2600mAh

The battery has overvoltage, overcurrent, and short circuit protection.

Power supply (Adapter)| Input : 100-240V AC 50Hz/60Hz 0.4A Max Output: DC 5V/1A

Built-in fuse: T1A 250V

Performance of LED lamp| a) 10W high-power blue and purple LED lamp.

b) Wavelength: 385nm-515nm

c) Category: class I.d)Shoot limit(AEL): 3.9*10―³J

Light Intensity| 1000mW/cm²-2500mW/cm²
Inspection method of LED lamp| When operating correctly, the LED lamp lights up, indicating that the lamp is in good condition. Dental resin materials commonly used in the clinic can match the wavelength of this curing light, such as 3M, Dentsply, and other resins.
Modes and time setting| TURBO mode: Blue and purple light, 2500mW/cm2 for 1s, 3s NORMAL mode: Blue and purple light, 1200mW/cm2 for 5s, 10s, 15s, 20s ORTHO mode: Blue and purple light, 2500mW/cm2 for 3s to work 10 times every cycle, or 5s to work 10 times every cycle, interval 1s CHECK mode: purple light, 900mW/cm2 for 30s, 60s
---|---
The effective area of irradiation light| 66.4mm²

Environmental factors

  • Operation: +5°C – +40°C
  • Relative humidity: 10% – 93%
  • Atmospheric pressure: 70kPa-106kPa

Equipment safety classification

  • Classified by type of protection against electric shock: Class Ⅱ
  • Classified by degree of protection against electric shock: Type B application part
  • Classified according to the degree of protection against incoming liquid: Ordinary equipment (IPXO), non-waterproof
  • Classified by operation mode: Short-term operation equipment
  • Classification of safety degree when using flammable anesthetic gas mixed with air or oxygen or nitrous oxide: Equipment that cannot be used with flammable anesthetic gas mixed with air or oxygen or nitrous oxide.

Installation and demounting

  • Insert the large end of the light guide (front connector) into the mounting hole at the upper end of the curing light (please insert it into the bottom in a rotating way, do not be oblique).
  • Clamp and install the light hood at the root of the light guide (front joint).
  • When removing or replacing the light guide, proceed in the opposite direction above.

Operation

  • When the machine is under standby mode, press the time/mode key to light up the screen.
  • Mode setting: press the time/mode key for more than 2 seconds to select the mode; during mode switching, the display screen has a transition picture (the picture shows the full English name of the mode to be switched to), and every time the key is pressed, there’s a beep. Without operating within 1 minute, the screen will turn off and the main unit will enter the sleep state.
  • Time setting: press the time/mode key to select time, see Table 1 for settable curing time.
  • Point the light guide rod (front connector) light source at the right position, press the switch button and there’s a beep. The curing light will start to work as the mode setting. The display screen starts counting down from the selected time, and after the timing ends, the display screen displays the setting time.
  • Before the end of setting time, at any time pressing the on/off key can finish the work.
  • After the end of an irradiation cycle, you can immediately press the switch to start the next irradiation cycle. If the handle starts to become hot, please stop working until the handle is completely cooled down. It is recommended that the number of consecutive irradiations should not exceed 10 times.
  • If the transparent cover is stained with resin after use, please wipe it with a cotton cloth to avoid affecting the light intensity.
  • The effective light intensity of this machine is many times higher than that of the halogen lamp curing light, and the curing depth of the light-curing composite resin is not less than 4mm for 10 seconds.
  • The front connector at the front end can be removed, but high temperature and high-pressure sterilization are not allowed, otherwise, it will cause damage.

Low battery prompt and charging

  • Low voltage   indication: The battery symbol on the display starts to flicker, accompanied by a sound prompt;
  • When charging, insert the charging cable into the TYPE-C socket at the back of the base, the blue light at the bottom of the base will light up, and the brightness will change;
  • Correctly put the main unit into the charging base, and the base display will automatically light up and display the power;
  • The charging base has a light meter function. When the base is in standby mode, the display is off. Only when the main unit is charged or when testing light intensity that the display automatically light up. If it is not charged or measured, the base display will automatically turn off.

Transportation, storage, and maintenance

Transportation

  • During transportation, avoid shock and vibration, handle with care, and avoid inversion.
  • Do not mix with dangerous goods during transportation.
  • Avoid sunlight, rain, or snow during transportation

Storage

  • Do not mix with toxic, corrosive, flammable, or explosive materials during storage.
  • The product should be stored in an environment where the relative humidity does not exceed 10% – 93%, the atmospheric pressure is 70kPa – 106kPa, and the temperature is -20°C – +40°C.

Maintenance

  • This equipment should be handled with care, far away from the source of the earthquake, and should be installed or stored in a cool, dry and ventilated place.
  • When the device is not in use, turn off the power switch and unplug the power plug; when not in use for a long time, charge the main unit every 3 months.
  • To clean or disinfect, wipe the surface with clean water or disinfectant, do not soak.
  • After use, check if there is any resin remaining on the surface of the lamp, so as not to affect the life of the front joint or affect the curing effect.

Troubleshooting

Faulty Possible cause solutions
No reaction 1. Out of battery 2.Faulty battery 3.Battery protection 1.

Charge/Change a new battery 2. Change a new battery 3. Activate it by putting it on the base
Insufficient Light intensity| There is resin remaining on the surface of the transparent cover of the light guide.| 1.   Remove the resin

2.   Change the transparent cover

Not charging when connecting with the adapter| 1.   The adapter is not connected well

2.   Faulty adapter or encompass- ible

3.   The charging point is impurity.

| 1. Reconnect. 2. Change the adapter

3. Clean the inner connector of the charging base with the alcohol

Less working life after full charge| The capacity of the battery decreased| Change a new battery
The mode indicator twin- ankles when charging| Low voltage| Back to normal after 15 minutes of charging.

If the fault still cannot be resolved, please contact your local dealer or our company.

Electromagnetic compatibility

Attention:

  • This Curing Light meets the relevant requirements of YY 0505 standard electromagnetic compatibility. The basic performance of the light curing machine is: that the front LED light can emit blue (UV) when working.
  • The user should install and use it according to the electromagnetic compatibility information provided by the accompanying documents.
  • Portable and mobile radio frequency communication equipment may affect the performance of the curing light, avoid strong electromagnetic interference when using it, such as close to mobile phones, microwave ovens, etc.;
  • The guidelines and manufacturer’s declaration are detailed in the attachment.
  • Without the express consent of Guilin Refine Medical Instrument Co., Ltd., unauthorized changes or modifications to the equipment may cause damage to the equipment or other
  • The electromagnetic compatibility of the equipment.

Alert:

  • This light-curing machine should not be used close to or stacked with other equipment. If it must be used close or stacked, it should be observed and verified that it can be used in the configuration used.

Normal operation

  • Except for the cables sold by the manufacturer of the curing light as spare parts of internal components, the use of accessories and cables other than those specified may cause the curing light to Increase in emission or decrease in immunity.
  • The following cables must be used to comply with electromagnetic emission and anti-interference requirements:
Cable Name Cable Type Cable Length
Power Line (Output end) Unshielded Parallel Lines 1.2 m

Key components of electromagnetic compatibility

  • The key components of electromagnetic compatibility of this product are the IC chip, main PCB, lithium battery, power adapter, and LED lamp.
  • Using or replacing the accessories and cables that are not designed to match, will cause the electromagnetic compatibility emission and anti-interference performance to be significantly reduced.
  • Do not change machine parts without authorization.

Enclosure:

Guidance and manufacturer’s declaration – electromagnetic emissions

The models Curing Light are intended for use in the electromagnetic environment specified below. The customer or the user of the model Curing Light should ensure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment – guidance
RF emissionsGB 4824| Group 1| The models Curing Light use RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissionsGB 4824| Class B| The models Curing Light are suitable for use in domestic establishments and in establishments directly connected to a low voltage power supply network that supplies buildings used for domestic purposes- es.
Harmonic emissions 17625.1| Unapplicable
Voltage fluctuations/flicker emissions 17625.1| Complies
Guidance & Declaration — electromagnetic immunity

The models Curing Light are intended for use in the electromagnetic environment specified below. The customer or the user of the model Curing Light should ensure that It is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment – guidance
Electrostatic discharge (ESD)

GB/T 17626.2

| ±8 kV contact

±15 kV air

| ±8 kV contact

±15 kV air

| Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrical fast transient/ burst

GB/T 17626.4

| ±2kV for power supply lines ±1 kV for Input/output lines| ±2kV for power supply lines| Mains power quality should be that of a typical commercial or hospital environment.
Surge GB/T 17626.5| ±1 kV line to line ±2 kV line to earth| ±1 kV line to line| Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions, and voltage variations on power supply input lines GB/T 17626.11| <5 % UT (>95% dip in UT.) for 0.5 cycles 40 % UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95 % dip in UT) for 5 sec| <5 % UT (>95% dip in UT.) for 0.5 cycles 40 % UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95 % dip in UT) for 5 sec| Mains power quality should be that of a typical commercial or hospital environment. If the user of the model Curing Light requires continued operation during power mains interruptions, it is recommended that the model Curing Light be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field GB/T 17626.8| 3A/m| 3A/m| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage before application of the test level.
Guidance & Declaration – Electromagnetic immunity

The models Curing Light are intended for use in the electromagnetic environment specified below. The customer or the user of the model Curing Light should ensure that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment – guidance
Conducted RF GB/T 17626.6 Radiated RF GB/T 17626.3| 3 Vrms 150 kHz to 80 MHz 3 V/m 80 MHz to 2.5 GHz| 3 Vrms ** 3 V/m| Portable and mobile RF communications equipment should be used no closer to any part of the model Curing Light, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d=[3,5/V1]×P1/2 d=[1.2/E1]×P1/2  80 MHz to 800 MHz d=[2.3/E1]×P1/2  800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d Is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range Interference may occur In the vicinity of equipment marked with the following symbol:
---|---|---|---
 |  |  |
NOTE:** 1 At 80 MHz end 800 MHz the higher frequency range applies.

NOTE: 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcasts, and TV broadcasts cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the model Curing lights are used exceeds the applicable RF compliance level above, the model Curing Light should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model Curing Light. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
The Rated maximum the output power of the transmitter W| The distance between different frequencies of the transmitter
---|---
150kHz to 80MHz d=1.2×P1/2| 80MHz to 800MHz d=1.2×P1/2| 800MHz to 2,5GHz d=2.3×P1/2
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.22.3| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For the maximum rated output power of the transmitter not listed in the above table, the recommended separation distance (d) which is in meters (m) that can be determined by the formula in the frequency column of the corresponding transmitter. The P is the maximum rated output power of the transmitter provided by the transmitter manufacturer which is in watts (W).
Note 1: At 80MHz and 800MHz frequency points, the higher frequency band formula is used.
Note 2: These guidelines may not be suitable for all situations. Electromagnetic propagation is affected by the absorption and reflection of buildings, objects, and human bodies.

Environmental protection

Components hazardous substance or element
(Pb) (Hg)
Main units
Charging base
Adaptor
Mechanical elements (Including screws, nuts, joint rings, etc.)○:

Indicates that the content of this toxic and hazardous substance in all homogeneous materials of this component is below the limit requirement in SJ/T-11363-2006 “Limit Requirements for Toxic and Hazardous Substances in Electronic Information Products”.
×: It means that the content of this toxic and hazardous substance in at least one of the homogeneous materials of the part exceeds the limit requirement of SJ/T11363-2006. (This product complies with EU RoHS environmental protection requirements; currently, there is no mature technology in the world that can replace or reduce the lead content in electronic ceramics, optical glass, steel, and copper alloys) According to the “Administrative Measures for Restricting the Use of Hazardous Substances in Electrical and Electronic Products”, “Regulations for Recycling and Disposal of Waste Electrical and Electronic Products” and relevant standards, please comply with the safety and use precautions of the products, and recycle or dispose of the products in appropriate ways according to local laws and regulations after the use of the products.

After sales service

  • If this equipment cannot work normally due to quality problems from the date of sale, please show the warranty card and our company will be responsible for the maintenance.
  • Please refer to the product warranty card for the warranty period and warranty scope.
  • This product does not contain replaceable parts. All maintenance, adjustment, calibration, and modification of technical parameters of the product can only be carried out by our company’s technicians or special repair shops. If customers need to repair themselves, our company can provide them on request Circuit diagrams, component lists, legends, calibration specifications, or other materials needed to help qualified technical personnel repair the equipment parts designated by our company, but we will not bear any consequences.
  • The user must use the original parts, and purchase please contact the local dealer or the manufacturer. Do not use the related accessories of other brands, so as not to cause damage to the light curing machine or other dangers.
  • After the power adapter, charging base, and other accessories are damaged, users should not repair it by themselves. Please buy new parts and replace them by yourself. If you need relevant information, please contact the manufacturer.

Special instructions

The company reserves the right to modify the machine design, product technology or accessories, operation manual, and machine packaging content at any time without prior notice.
The product is subject to the real object, and the picture is for reference only. The final interpretation right belongs to Guilin Refine Medical Instrument CO., LTD.

After service and Warranty Instruction

  1. Period validity: two-year free repair for the main unit, charging base, and power adapter from the date of purchase. One year’s free repair for a lithium battery. A warranty card is necessary while submitting the repair request.
  2. Range of warranty: Within the warranty period of validity, we are responsible for any troubles caused by quality problems or product technique and structure.
  3. The following are beyond our warranty:
  4. The damage caused by disobeying the operation instructions or lack of the needed condition.
  5. The damage is caused by unsuitable operation or disassembly without authorization.
  6. The damage caused by unadvisable transportation or preservation.
  7. There isn’t the seal of the distributor or the warranty card isn’t filled in completed.
  8. Light tip and a protective light shield.

After service cardREFINE-Swan-Dental-LED-Curing-Light-
FIG-5

  • MedNet EC-REP GmbH
  • Borkstrasse 10 · 48163 Muenster · Germany
  • Product life: 5 years
  • Production date: please refer to the packaging label.
  • https://www.alandental.com

Documents / Resources

| REFINE Swan Dental LED Curing Light [pdf] Instruction Manual
Swan Dental LED Curing Light, Dental LED Curing Light, LED Curing Light, Curing Light, Light
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals