Microlife E-V10 BP A2 Classic Blood Pressure Monitor Instruction Manual
- August 12, 2024
- microlife
Table of Contents
Microlife E-V10 BP A2 Classic Blood Pressure Monitor
Product Specifications :
-
Brand: Microlife
-
Model: BP A2 Classic
-
Intended Use: Measuring non-invasive blood pressure in
individuals aged 12 years or older -
Manufacturer: Microlife Corporation
-
Country of Manufacture: Taiwan, China
-
Power Source: Battery operated or mains adapter
-
Display: LCD Display
-
Memory Capacity: Stores multiple readings
Product Usage Instructions
- Important Facts about Blood Pressure:
- Blood pressure is the pressure of the blood flowing in the arteries generated by the pumping of the heart. It is crucial to monitor systolic and diastolic values as well as pulse rate.
- Using the Device for the First Time:
- Inserting the Batteries:
- Open the battery compartment
- Insert the required batteries
- Close the battery compartment securely
- Selecting the Correct Cuff:
- Choose a cuff size that fits comfortably around your upper arm
- Taking a Blood Pressure Measurement:
- Checklist for Taking a Reliable Measurement:
- Sit in a comfortable position
- Rest for a few minutes before taking the measurement
- Place the cuff on your upper arm at heart level
- Follow the device instructions for accurate readings
- Appearance of the Irregular Heartbeat (IHB) Symbol:
- The device may display an IHB symbol if it detects an irregular heartbeat. Consult your healthcare provider if this symbol appears frequently.
- Data Memory:
- Viewing Stored Values:
- Access the memory function on the device to view previous readings
- Clearing All Values:
- Follow the device instructions to clear all stored readings
- Battery Indicator and Battery Change:
- Low Battery:
- Replace the batteries when the low battery indicator appears
- Flat Battery – Replacement:
- Follow the device manual for battery replacement instructions
FAQ:
-
Q: What should I do if I consistently get high blood pressure readings?
A: Consult your healthcare provider for further evaluation and guidance on managing high blood pressure. -
Q: Can I use rechargeable batteries with the device?
A: Yes, you can use rechargeable batteries following the specified procedure outlined in the user manual.
Microlife Corporation
9F, 431, RuiGuang Road, NeiHu Taipei 11492, Taiwan, China
www.microlife.com
Microlife UAB P. Lukšio g. 32 08222 Vilnius Lithuania
Microlife BP A2 Classic
Guarantee Card
- Name of Purchaser
- Serial Number
- Date of Purchase
- Specialist Dealer
Microlife BP A2 Classic
-
ON/OFF button
-
Display
-
Cuff socket
-
Mains Adapter Socket
-
Battery compartment
-
Cuff
-
Cuff connector
-
M-button (memory)
Display -
Systolic value
-
Diastolic value
-
Pulse rate
-
Irregular heartbeat (IHB) symbol AM Pulse
-
Battery display
-
Stored value
-
Traffic light indicator
Intended use
- This oscillometric blood pressure monitor is intended for measuring non-invasive blood pressure in people aged 12 years or older.
- It is clinically validated in patients with hypertension, hypotension, diabetes, pregnancy, pre-eclampsia, atherosclerosis, end-stage renal disease, obesity and the elderly.
- Dear Customer,
- This device was developed in collaboration with physicians and clinical tests carried out prove its measurement accuracy to be of a very high standard.*
- If you have any questions, problems or want to order spare parts please contact your local Microlife-Customer Service. Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your country. Alternatively, visit the internet at
- www.microlife.com where you will find a wealth of invaluable information on our products.
- Stay healthy – Microlife Corporation!
- This device uses the same measuring technology as the award winning «BP 3BTO-A» model tested according to the British and Irish Hypertension Society (BIHS) protocol.
Important facts about blood pressure
- Blood pressure is the pressure of the blood flowing in the arteries generated by the pumping of the heart. Two values, the systolic (upper) value and the diastolic (lower) value, are always measured.
- The device indicates the pulse rate (the number of times the heart beats in a minute).
- Permanently high blood pressure values can damage your health and must be treated by your doctor!
- Always discuss your values with your doctor and tell them if you have noticed anything unusual or feel unsure. Never rely on single blood pressure readings.
- There are several causes of excessively high blood pressure values. Your doctor will explain them in more detail and offer treatment where appropriate.
- Under no circumstances should you alter the dosages of drugs or initiate a treatment without consulting your doctor.
- Depending on physical exertion and condition, blood pressure is subject to wide fluctuations as the day progresses. You should therefore take your measurements in the same quiet conditions and when you feel relaxed! Take at least two readings every time (in the morning: before taking medications and eating / in the evening: before going to bed, bathing or taking medication) and average the measurements.
- It is quite normal for two measurements taken in quick succession to produce significantly different results.
- Deviations between measurements taken by your doctor or in the pharmacy and those taken at home are quite normal, as these situations are completely different.
- Several measurements provide much more reliable informa-tion about your blood pressure than just one single measure-ment.
- Leave a small break of >5 minutes between two measure-ments.
- If you suffer from a cardiac arrhythmia consult with your doctor before using the device. See also chapter «Appearance of the irregular heartbeat (IHB) symbol» of this user manual.
- The pulse display is not suitable for checking the frequency of heart pacemakers!
- If you are pregnant, you should monitor your blood pressure regularly as it can change drastically during this time.
- A number of factors can affect the accuracy of measurements taken on the wrist. In some cases, the result may differ from the measurement taken on the upper arm. We therefore advise you to compare these values with those produced by the upper arm measurement and discuss them with your doctor.
How do I evaluate my blood pressure
Table for classifying blood pressure values in adults in accordance with the
World Health Organisation (WHO) in 2003. Data in mmHg.
Range | Systolic | Diastolic | Recommendation |
---|---|---|---|
1. | blood pressure normal | < 120 | < 80 |
Range | Systolic | Diastolic | Recommendation |
--- | --- | --- | --- |
2. | blood pressure normal | 120 – 129 | 80 – 84 |
3. | blood pressure slightly high | 130 – 139 | 85 – 89 |
doctor
4.| blood pressure too high| 140 – 159| 90 – 99| Seek medical
advice
5.| blood pressure far too high| 160 – 179| 100 – 109| Seek medical
advice
6.| blood pressure dangerously high| ≥180| ≥110| Urgently seek
medical advice!
The higher value is the one that determines the evaluation. Example: a readout
value between 150/85 or 120/98 mmHg indicates «blood pressure too high».
An index of 140/90 is shown adjacent to the traffic light bars AP. Your blood
pressure is too high if the bars reach or exceed this index.
Using the device for the first time
-
Inserting the batteries
After you have unpacked your device, first insert the batteries. The battery compartment 5 is on the bottom of the device. Insert the batteries (4 x 1.5 V, size AA), thereby observing the indicated polarity. -
Selecting the correct cuff
Microlife offers different cuff sizes. Select the cuff size to match the circumference of your upper arms (measured by close fitting in the centre of the upper arm).
Cuff size | for circumference of upper arm |
---|---|
S | 17 – 22 cm |
M | 22 – 32 cm |
M – L | 22 – 42 cm |
L | 32 – 42 cm |
L – XL | 32 – 52 cm |
- Pre-shaped cuffs are optionally available.
- Only use Microlife cuffs.
- Contact your local Microlife Service if the enclosed cuff 6 does not fit.
- Connect the cuff to the device by inserting the cuff connector 7 into the cuff socket 3 as far as it will go.
Taking a blood pressure measurement
Checklist for taking a reliable measurement
- Avoid activity, eating or smoking immediately before the measurement.
- Sit down on a back-supported chair and relax for 5 minutes. Keep your feet flat on the floor and do not cross your legs.
- Always measure on the same arm (normally left).
- Remove close-fitting garments from the upper arm. To avoid constriction, shirt sleeves should not be rolled up – they do not interfere with the cuff if they are laid flat.
- Always ensure that the correct cuff size is used (marking on the cuff).
- Fit the cuff closely, but not too tight.
- Make sure that the cuff is positioned 1-2 cm above the elbow.
- The artery mark on the cuff (ca.3 cm long bar) must lie over the artery which runs down the inner side of the arm.
- Support your arm so it is relaxed.
- Ensure that the cuff is at the same height as your heart.
- Press the ON/OFF button 1 to start the measurement.
- The cuff will now pump up automatically. Relax, do not move and do not tense your arm muscles until the measurement result is displayed. Breathe normally and do not talk.
- When the correct pressure is reached, the pumping stops and the pressure falls gradually. If the required pressure was not reached, the device will automatically pump some more air into the cuff.
- During the measurement, the pulse indicator AP flashes in the display.
- he result, comprising the systolic 9 and the diastolic AT blood pressure and the pulse rate AK are displayed. Note also the explanations on further display symbols in this booklet.
- When the device has finished measuring, remove the cuff.
- Switch off the device. (The monitor does switch off automatically after approx. 1 min.).
Appearance of the irregular heartbeat (IHB) symbol
- This symbol AL indicates that an irregular heartbeat was detected. In this case, the measured blood pressure may deviate from your actual blood pressure values. It is recommended to repeat the measurement.
- Information for the doctor in case of repeated appearance of the IHB symbol:
This device is an oscillometric blood pressure monitor that also measures the pulse during blood pressure measurement and indicates when the heart rate is irregular.
Data memory
-
This device automatically stores the last 30 measurement values.
-
Viewing the stored values
-
Press the M-button 8 briefly, when the device is switched off. The display first shows «M» AO, and «28A»which stands for the average of all stored values.
-
Pressing the M-button again displays the previous value. Pressing the M-button repeatedly enables you to move from one stored value to another.
-
Memory full
Pay attention that the maximum memory capacity of 30 memories per user is not exceeded. When the 30 memory is full, the oldest value is automatically overwritten with the 31th value. Values should be evaluated by a doctor before the memory capacity is reached – otherwise data will be lost. -
Clearing all values
-
If you are sure that you want to permanently remove all stored values, hold down the M-button (the device must have been switched off beforehand) until «CL» appears and then release the button. To permanently clear the memory, press the M-button while «CL» is flashing. Individual values cannot be cleared.
-
How not to store a reading
-
As soon as the reading is displayed press and hold the ON/OFF button 1 until «M» AO is flashing. Confirm to delete the reading by pressing the M-button 8.
Battery indicator and battery change
-
Low battery
When the batteries are approximately ¾ empty the battery symbol AN will flash as soon as the device is switched on (partly filled battery displayed). Although the device will continue to measure reliably, you should obtain replacement batteries. -
Flat battery – replacement
When the batteries are flat, the battery symbol AN will flash as soon as the device is switched on (flat battery displayed). You cannot take any further measurements and must replace the batteries.- Open the battery compartment 5 on the bottom of the device.
- Replace the batteries – ensure correct polarity as shown by the symbols in the compartment.
-
Which batteries and which procedure?
- Use 4 new, long-life 1.5 V, size AA alkaline batteries.
- Do not use batteries beyond their date of expiry.
- Remove batteries if the device is not going to be used for a prolonged period.
-
Using rechargeable batteries
-
You can also operate this device using rechargeable batteries.
- Only use «NiMH» type reusable batteries.
- Batteries must be removed and recharged when the flat battery symbol appears. They should not remain inside the device as they may become damaged (total discharge as a result of low use of the device, even when switched off).
- Always remove the rechargeable batteries if you do not intend to use the device for a week or more.
- Batteries cannot be charged in the blood pressure monitor. Recharge batteries in an external charger and observe the information regarding charging, care and durability.
Using a mains adapter
- You can operate this device using the Microlife mains adapter (DC 6V, 600 mA).
- Only use the Microlife mains adapter available as an original accessory appropriate for your supply voltage.
- Ensure that neither the mains adapter nor the cable are damaged.
- Plug the adapter cable into the mains adapter socket 4 in the blood pressure monitor.
- Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.
Error Messages
If an error occurs during the measurement, the measurement is interrupted
and an error message, e.g. «Err 3», is displayed.
Error | Descrip tion | Potential cause and remedy |
---|---|---|
«Err 1» | Signal too weak | The pulse signals on the cuff are too weak. |
Re-position the cuff and repeat the measurement.
«Err 2» AR| Error signal| During the measurement, error signals were
detected by the cuff, caused for instance by movement or muscle tension.
Repeat the measurement, keeping your arm still.
«Err 3» AQ| No pressure in the cuff| An adequate pressure cannot be
generated in the cuff. A leak may have occurred. Check that the cuff is
correctly connected and is not too loose. Replace the batteries if necessary.
Repeat the measurement.
«Err 5»| Abnormal result| The measuring signals are inaccurate and no
result can therefore be displayed. Read through the checklist for taking a
reliable measurement and then repeat the measurement.
«HI»| Pulse or cuff pressure too high| The pressure in the cuff is too
high (over 299 mmHg) OR the pulse is too high (over 200 beats per minute).
Relax for 5 minutes and repeat the measurement.
«LO»| Pulse too low| The pulse is too low (less than 40 beats per
minute). Repeat the measurement.
Please immediately consult your doctor, if this or any other problem occurs repeatedly.
- If you think the results are unusual, please read through the information in «Section 1.» carefully.
Safety, care, accuracy test and disposal
Safety and protection
-
Follow instructions for use. This document provides important product operation and safety information regarding this device. Please read this document thoroughly before using the device and keep for future reference.
This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application. -
This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the «Technical Specifications» section.
-
The cuffs are sensitive and must be handled with care.
-
Only pump up the cuff once fitted.
-
Do not use this device if you think it is damaged or notice anything unusual.
-
Never open this device.
-
Read the additional safety information provided within the individual sections of this instruction manual.
-
The measurement results given by this device is not a diag-nosis. It is not replacing the need for the consultation of a physician, especially if not matching the patient’s symptoms. Do not rely on the measurement result only, always consider other potentially occurring symptoms and the patient’s feedback. Calling a doctor or an ambulance is advised if needed.
-
Ensure that children do not use this device unsupervised; some parts are small enough to be swallowed. Be aware of the risk of strangulation in case this device is supplied with cables or tubes.
Contraindications
Do not use this device if the patient’s condition meets the following
contraindications, to avoid inaccurate measurements or injuries.
- The device is not intended for measuring blood pressure in pediatric patients of age younger than 12 years old (children, infant, or neonates).
- Presence of significant cardiac arrhythmia during measurement may interfere with blood pressure measurement and affect the reliability of blood pressure readings. Consult with your doctor about whether the device is suitable for use in this case.
- The device measures blood pressure using a pressured cuff. If the measuring limb suffers from injuries (for example open wounds) or under conditions or treatments (for example intrave-nous drip) making it unsuitable for surface contact or pressurization, do not use the device, to avoid worsening of the injuries or conditions.
- Patient motions during measurement may interfere with the measurement process and influence results.
- Avoid taking measurements of patients with conditions, diseases, and susceptible to environment conditions that lead to incontrollable motions (e.g. trembling or shivering) and inability to communicate clearly (for example children and unconscious patients).
- The device uses oscillometric method to determine blood pres-sure. The arm being measure should have normal perfusion. The device is not intended to be used on a limb with restricted or impaired blood circulation. If you suffer with perfusion or blood disorders, consult your doctor before using the device.
- Avoid taking measurement on the arm on the side of a mastectomy or lymph node clearance.
- Do not use this device in a moving vehicle (for example in a car or on an aircraft).
WARNING
Indicates a potentially hazardous situation, which if not avoided, could
result in death or serious injury.
- This device may only be used for the intended uses described in this Instructions for Use. The manufacturer cannot be held liable for damage caused by incorrect application.
- Do not change the patient medication and treatment based the result of one or multiple measurements. Treatment and medica-tion changes should be prescribed only by a medical profes-sional.
- Inspect the device, cuff, and other parts for damage. DO NOT USE the device, cuff or parts if they appear damaged or oper-ating abnormally.
- Blood flow of the arm is temporarily interrupted during measure-ment. Extended interruption of blood flow reduces peripheral circulation and may cause tissue injury. Beware of signs (for example tissue discoloration) of impeded peripheral circulation if taking measurements continuously or for an extended period of time.
- Prolonged exposure to cuff pressure will reduce peripheral perfusion and may lead to injury. Avoid situations of extended cuff pressurization beyond normal measurements. In the case of abnormally long pressurization, abort the measurement or loose the cuff to depressurize the cuff.
- Do not use this device in oxygen rich environment or near flam-mable gas.
- The device is not water resistant or water proof. Do not spill or immerse the device in water or other liquids.
- Do not dissemble or attempt to service the device, accessory and parts, during use or in storage. Access to the device internal hardware and software is prohibited. Unauthorized access and servicing of the device, during use or in storage, may compromise the safety and performance of the device.
- Keep the device away from children and people incapable of operating the device. Beware of the risks of accidental ingestion of small parts and of strangulation with the cables and tubes of this device and accessories.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result
in minor or moderate injury to the user or patient, or cause damage to the
device or other property.
- The device is intended only for measuring blood pressure at upper arm. Do not measure other sites because the reading does not reflect your blood pressure accurately.
- After a measurement is completed, loosen the cuff and rest for > 5 minutes to restore limb perfusion, before taking another measurement.
- Do not use this device with other medical electrical (ME) equip-ment simultaneously. This may cause device malfunction or measurement inaccuracies.
- Do not use this device in proximity of high frequency (HF) surgical equipment, magnetic resonance imaging (MRI) equip-ment, and computerized tomography (CT) scanners. This may cause device malfunction and measurement inaccuracies.
- Use and store the device, cuff and parts in temperature and humidity conditions specified in the «Technical Specifications».
- Usage and storage of the device, cuff and parts in conditions outside ranges given in the «Technical Specifications» may results in device malfunction and the safety of usage.
- Protect the device and accessories from the following to avoid damaging the device:
- water, other liquids, and moisture
- extreme temperatures
- impacts and vibrations
- direct sunlight
- contamination and dust
- This device has 2 user settings. Please ensure it is cleaned and disinfected between users to prevent any cross contamination.
- Stop using this device and cuff and consult with your doctor if you experience skin irritation or discomfort.
Electromagnetic Compatibility Information
- This device is compliant with EN60601-1-2: 2015 Electromagnetic Disturbances standard.
- This device is not certified to be used in vicinity of High Frequency (HF) medical equipment.
- Do not use this device close to strong electromagnetic fields and portable radio frequency communication devices (for example microwave oven and mobile devices). Keep a minimum distance of 0.3 m from such devices when using this device.
- Device care
- Clean the device only with a soft, dry cloth.
- Cleaning the cuff
- Carefully remove spots on the cuff with a damp cloth and soapsuds.
- WARNING: Do not wash the cuff in a washing machine or dishwasher!
- WARNING: Do not dry the cuff cover in a tumble dryer!
- WARNING: Under no circumstances may you wash the inner bladder!
Accuracy test
We recommend this device is tested for accuracy every 2 years or after
mechanical impact (e.g. being dropped). Please contact your local Microlife-
Service to arrange the test (see foreword).
Disposal
Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
Guarantee
This device is covered by a 5 year guarantee from the date of purchase. During
this guarantee period, at our discretion, Microlife will repair or replace the
defective product free of charge. Opening or altering the device invalidates
the guarantee.
The following items are excluded from the guarantee:
- Transport costs and risks of transport.
- Damage caused by incorrect application or non-compliance with the instructions for use.
- Damage caused by leaking batteries.
- Damage caused by accident or misuse.
- Packaging/storage material and instructions for use.
- Regular checks and maintenance (calibration).
- Accessories and wearing parts: Batteries, power adapter (optional).
- The cuff is covered by a functional guarantee (bladder tightness) for 2 years.
- Should guarantee service be required, please contact the dealer from where the product was purchased, or your local Microlife service. You may contact your local Microlife service through our website: www.microlife.com/support
- Compensation is limited to the value of the product. The guarantee will be granted if the complete product is returned with the original invoice. Repair or replacement within guarantee does not prolong or renew the guarantee period. The legal claims and rights of consumers are not limited by this guarantee.
Technical Specifications
Operating conditions:
- 10 – 40 °C / 50 – 104 °F
- 15 – 90 % relative maximum humidity
- Storage conditions: -20 – +55 °C / -4 – +131 °F
- 15 – 90 % relative maximum humidity
- Weight: 375 g (including batteries)
- Dimensions: 120 x 87 x 59 mm
- Cuff size: from 17-52 cm according to the cuff sizes (see «Selecting the correct cuff»)
- Measuring procedure: oscillometric, corresponding to Korotkoff method: Phase I systolic, Phase V diastolic
- Measurement range: SYS: 60 – 255 mmHg
- DIA: 40 – 200 mmHg
- Pulse: 40 – 199 beats per minute
- Cuff pressure display range: 0 – 299 mmHg
- Resolution: 1 mmHg
- Static accuracy: within ± 3 mmHg
- Pulse accuracy: ± 5 % of the readout value
- Voltage source: 4 x 1.5 V alkaline batteries; size AA
- Mains adapter DC 6V, 600 mA (optional)
- Battery lifetime: approx. 920 measurements (using new batteries)
- IP Class: IP 20
Reference to standards:
- IEC 80601-2-30; IEC 60601-1;
- IEC 60601-1-2 (EMC); IEC 60601-1-11
- Expected service life: Device: 5 years or 10000 measurements, whichever comes first
- Accessories: 2 years or 5000 measurements, whichever comes first
This device complies with the requirements of the Medical Device Directive
93/42/EEC.
Technical alterations reserved.
References
- life.cz - výroba je pro nás potěšením
- Microlife ®
- Hypertension and Fever Management - Microlife AG
- Support - Microlife AG
- Microlife. Partner pre ľudí. Pre život. - Microlife AG
- Softvér a návody - Microlife AG
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>