Specialized Care Mouth Opener and HVE Adapter Instructions

August 10, 2024
Specialized Care

Specialized Care Mouth Opener and HVE Adapter

Product Information

Specifications:

  • Manufacturer: Specialized Care Co, Inc.
  • Product: Mouth Props, Mouth Opener, and HVE Adapter
  • Maximum Temperature: 285 degrees Fahrenheit
  • Website: www.specializedcare.com
  • Address: 24 Stickney Terrace Ste. 2, Hampton, NH 03842-4902
  • Contact: 603-926-0071, Fax: 603-926-5905, Toll-Free: 800-722-7375

Product Usage Instructions

Cleaning and Sterilization Procedure:

  1. Clean the Mouth Props, Mouth Opener, and HVE Adapter after each use to prevent cross-contamination.
  2. Use an appropriate cleaning solution recommended for dental instruments.
  3. Follow the manufacturer’s guidelines for cleaning and sterilization.

Important Note

Alcohol is not recommended for cleaning these products in the presence of bioburden as it may cause staining and make bioburden more adherent. Follow OSHA and CDC guidelines for infection control procedures.

FAQ

  • Q: Can I use alcohol to clean the Mouth Props, Mouth Opener, and HVE Adapter?
    • A: No, alcohol is not recommended as it may cause staining and make bioburden more adherent. It’s best to use cleaning solutions recommended for dental instruments.

Cleaning and Sterilization

Procedure : Mouth Props, Mouth Opener and HVE Adapter

  • Applies to the Open Wide® Wraparound (OWR), Tab (OWT) mouth props, the Molini™ Mouth Opener (MMO) and the Corbin™ HVE Adapter (CA-HVE 4).
  • The Molini™ Mouth Opener, Open Wide® Reusable Mouth Props and Corbin™ HVE Adapter are not sterile when packaged. These devices are autoclavable by either steam or chemical methods. They should be cleaned using an ultrasonic cleaner or manually with a brush, a mild detergent and water, removing all organic or inorganic matter, prior to placing in the autoclave. If a saliva ejector tip is included as a demonstration with your product, it is not autoclavable. The Corbin Adapters can be autoclaved up to 10 times*. Replace after that, or when they look worn.
  • Note that the maximum temperature is 285 degrees Fahrenheit.
  • Please Note: Alcohol is a poor cleaning agent in the presence of bioburden (blood and saliva, bacterial spores.) After exposure to alcohol, denatured bioburden becomes more insoluble and tenaciously adherent to most surfaces and may cause staining.
    *Always follow OSHA and CDC guidelines for infection control procedures.