Testex RTMMIL Digital Micrometer Instruction Manual
- August 7, 2024
- Testex
Table of Contents
Testex RTMMIL Digital Micrometer
Introduction
Testex Press-O-Film Replica Tape consists of a layer of compressible foam
affixed to an incompressible 50 µm (2 mils) polyester layer. When pressed or
‘burnished’ against a roughened steel surface, the foam collapses and forms an
impression of the surface. Placing the burnished Testex tape between the
anvils of a micrometer thickness gage, such as the Testex Digital Micrometer,
and subtracting the thickness of the incompressible layer gives a measure of
the surface profile height.
How to Measure
- Locate a representative site for measurement. Clean the surface to be tested using a cleaning putty or other suitable method (such as compressed air or pressure-sensitive tape) to remove dust, debris, or residual blast media from the surface.
- Select the appropriate grade of replica tape based on the expected height of the profile being measured, using the following chart:
Replica Tape Grade | Range | Order Code |
---|---|---|
Coarse Minus | 13 – 25 µm | 0.5 – 1.0 mil |
Coarse | 20 – 50 µm | 0.8 – 2.0 mils |
X-Coarse | 40 – 115 µm | 1.5 – 4.5 mils |
X-Coarse Plus | 100 – 150 µm | 4.0 – 6.0 mils |
NOTE : If a measurement falls outside of the specified range for the tape
grade used, disregard the reading and re-measure the surface with the
appropriate tape grade.
Pull a single adhesive backed replica tape free from its release paper, being
careful not to touch either side of the replica foam during application.
3. Apply the replica tape to the surface.Press down on the adhesive backing ofthe tape around the replica foam to ensure it remains in place during the burnishing process.
4a. For flat or convex surfaces: Using a Testex High Accuracy Burnishing
Tool, press the ball of the tool onto the replica foam with just enough
pressure so that the flat base rests on the tape. Move the tool’s ball around
the circular replica foam opening.
For concave surfaces : Flip over the High Accuracy Burnishing Tool and
hold it at an angle. Use the curved area near the Testex logo at the top of
the tool to burnish.
— or —
4b. If using a Classic Burnishing Tool, hold the tool like a pencil, and apply
consistent pressure with the ball end, being careful not to apply too much
force. Excessive force can lead to inaccurate measurements.
NOTE : The stick-type legacy burnishing tool remains a valid measurement practice and is recognized by international standards, but does not have an associated accuracy statement.
5. Alternate between linear (x-y) and circular motions with the burnishing tool to completely burnish the replica foam until a uniform stippled gray appearance is achieved across the entire measurement area.
The time required to properly burnish a replica will vary based on the surface profile. Profiles at the high end of a replica tape grade’s range may only require burnishing for 10-15 seconds. Profiles at the low end of the range may require up to 60 seconds. Refer to the below examples of properly and improperly burnished replicas.
6. Carefully remove the burnished replica tape from the surface. A small piece of release paper remains on the top-left corner of some grades of replica tape to facilitate easy removal.
7. Inspect the replica to ensure that no contamination from the surface is
present (e.g. dust, debris, residual blast media).
When held up to a light source, ‘points’ of light should be visible
consistently across the replica where the foam has been compressed by the
peaks of the surface profile.
8. Adhesive residue or other debris can build up on the faces of the anvils during use. To clean the anvils, place a cleaning card or piece of paper inside the closed anvils, and gently move back and forth.
NOTE : Clean the anvils prior to turning on the gage, zeroing (pg. 6), or
whenever the gage does not display -50 μm (-2.0 mils) when replica tape is
removed after taking a measurement.
After cleaning, power on the Testex Digital Micrometer by pressing the button.
9. Ensure the correct grade of replica tape is selected (pg. 1) and the HL
icon (pg. 7) is displayed at the bottom of the screen.
NOTE : If using Coarse Minus grade replica tape, HL mode should be
disabled. Any grade of replica tape can be selected from the Menu.
10. Press down on the lever to open the micrometer anvils, and insert the piece of replica tape, centering the circular opening of the replica foam between the anvils. Slowly release the lever, allowing the anvils to gently close on the replica.
Once the measured value stabilizes, the displayed value will flash, and be
recorded if Memory Mode is enabled (pg. 6).
NOTE : The gage should return to -50 μm (-2.0 mils) after the replica
tape is removed. If a different number is displayed after a measurement is
taken and the replica tape is removed, clean the anvils as detailed above.
Number of Readings
At least two measurements should be taken at each location. If the two
replicas differ by 5 μm (0.2 mils) or less, their average should be recorded
as the profile.
If the two initial replicas differ by more than 5 μm (0.2 mils), a third
replica is recommended. The average of the two closest replicas should be
recorded.
Accuracy
In 2024, a comprehensive study1 was performed to determine an accuracy
statement for replica tape. That study determined the following accuracy and
precision statements:
Replica Tape Grade | Precision | Accuracy |
---|---|---|
Coarse | ± 2 µm (0.1 mils) | ± 8 µm (0.3 mils) |
X-Coarse | ± 6 µm (0.2 mils) | ± 8 µm (0.3 mils) |
X-Coarse Plus | ± 8 µm (0.3 mils) | ± 10 µm (0.4 mils) |
The above accuracy statement is applicable under the following conditions:
- The procedure detailed in these instructions is followed
- The High Accuracy Burnishing Tool is used
- At least two readings are taken per measurement, in accordance with the procedure in Number of Readings (above)
- All readings are converted, using a Testex Digital Micrometer (pg. 7) or conversion table
NOTE : The Classic Burnishing Tool remains a valid measurement practice and is recognized by standards, but does not have an associated accuracy statement.
1 Beamish, M. (2024, March 3-7). Improvements to the Replica Tape Test Method. AMPP 2024, New Orleans, LA, United States.
Digital Micrometer Features & Operation
Menu Navigation
Press the button to enter the gage menu. Navigate between menu items using the
UP and DOWN buttons. Use the button to select the highlighted menu item. The
following menu items are displayed:
Zero
The Testex Digital Micrometer will automatically zero when powered on, or
can be zeroed by selecting from the gage menu. Ensure the micrometer anvils
are clean prior to zeroing or powering on the gage.
NOTE : If the gage does not display -50 µm (-2.0 mils) after a
measurement is taken and the replica tape is removed, clean the anvils as
detailed in Section 8 of How to Measure (pg. 4) prior to preforming a zero.
Memory Mode
When Memory Mode is ON the gage will store up to 250 readings. Stored
measurements can be downloaded to PosiSoft Desktop Software (pg. 9).
Statistics are displayed on screen showing the number of readings, average,
lowest, and highest measured readings.
What is stored, and whether the micrometer automatically stores a reading,
depends on the mode:
- Replica Tape Grade Selected – When a grade of replica tape is selected (pg. 7), the gage automatically stores each reading into memory.
- Averaging Mode – When Averaging Mode is ON (pg. 8), the gage automatically stores the average of paired readings into memory. Individual readings are not stored.
- No Replica Tape Grade – When No Replica Tape Grade is selected, readings are not automatically added to memory. Press the UP button to save a displayed reading.
Press the DOWN button to remove the last reading taken from memory.
WARNING: Toggling the display units, toggling Conversion mode or turning
ON/OFF Averaging Mode will automatically reset Memory Mode, and clear all
stored readings.
Select from the gage menu to clear all stored readings from memory.
Alternatively, powering on the gage by pressing and holding the and DOWN
button will clear all stored readings from memory. All other settings will be
retained.
Select from the gage menu to turn OFF Memory Mode. Stored readings will be
retained in memory until cleared. Additional readings can be added by turning
Memory Mode ON from the gage menu.
Display Brightness
Adjusts the display brightness using the UP and DOWN buttons.
Units
Toggles the display units between thou/mils and microns.
Replica Tape Grade
When a replica tape grade is selected, the micrometer will automati-cally
subtract the 50 µm (2.0 mils) thickness of the incompressible polyester layer
from the measurement. With Memory Mode ON, the se-lected grade will also be
recorded.
NOTE: With Conversion Mode enabled, the correct grade of replica tape must be
selected to ensure accurate readings.
Toggle the selected grade of tape from the gage menu. The menu icon will show
the currently selected grade: C for Coarse, XC for X-Coarse, and XC+ for
X-Coarse Plus. The selected grade of tape is displayed in the bottom left of
the measurement screen.
Select to select No Replica Tape Grade and use the micrometer to measure the distance between the anvils, like a standard analog micrometer. Conversion Mode and Averaging Mode are unavailable when No Replica Tape Grade is selected.
Conversion Mode H HL
When Conversion Mode is selected (default), an HL icon HL is displayed at the bottom center of the measurement screen. The gage displays a more accurate measurement that has been adjusted for the non-linearity of replica tape.
When Conversion Mode is disabled, an H icon is displayed at the bottom center
of the measurement screen. The gage displays the measured thickness without
adjustment, comparable to an analog spring micrometer. A correction table must
be used to accurately measure surface profile using replica tape.
NOTE: Conversion Mode is unavailable when No Replica Tape Grade is selected
from the gage menu.
Averaging Mode X2
At least two measurements should be taken at each location. If the two
replicas differ by 5 µm (0.2 mils) or less, their average should be recorded
as the profile height.
When A X2 veraging Mode is enabled, the gage will display the average of two
measurements. The individual readings will be displayed as R1 and R2, and the
average of the two readings will be displayed on the main measuring screen.
NOTE: When both Averaging Mode and Memory Mode are ON, only the average of the
two readings is recorded. The number of readings and statistics will be based
on the recorded averages.
NOTE: Averaging Mode is unavailable when No Replica Tape Grade is selected
from the gage menu.
Gage Reset
Restores the gage to default settings, clears stored memory and zero
adjustments.
Power Off
Powers down the gage. Alternatively, the gage will automatically power down
after 5 minutes of no activity. All settings are retained.
Exit Menu
Exits the gage menu and returns to the main measurement screen.
NOTE: While in the menu, pressing down on the lever to open the anvils will
also exit the menu.
Verifying the Gage
Verification shims are available as an optional accessory.
Connect to Computer
Readings stored in memory can be downloaded to a computer for viewing,
storing, and creating reports. Use the supplied cable and connect to the gage
using the USB-C port on the side of the micrometer.
USB Drive Mode —Retrieves data in a manner similar to a flash drive in
.csv format. No software to install or internet connection required.
PosiSoft Desktop —Powerful desktop software (PC/Mac) for downloading,
viewing, printing and storing measurement data. Includes a customizable,
templated PDF Report Generator.
Learn more at www.defelsko.com/posisoft
Troubleshooting
Before returning the instrument for service:
- Install new or newly recharged battery in the proper alignment as shown within battery compartment.
- Clean the micrometer anvils.
- Perform a Reset and Zero the gage.
IMPORTANT:
If these steps do not resolve the issue and the Gage must be returned for
service, please follow the instructions provided at
www.defelsko.com/service
Limited Warranty, Sole Remedy and Limited Liability
DeFelsko’s sole warranty, remedy, and liability are the express limited warranty, remedy, and limited liability that are set forth on its website: www.defelsko.com/terms
www.defelsko.com
© 2024 DeFelsko Corporation USA All Rights Reserved
This manual is copyrighted with all rights reserved and may not be reproduced
or transmitted, in whole or part, by any means, without written permission
from DeFelsko Corporation.
DeFelsko, Testex, Press-O-Film, Coarse, X-Coarse, and X-Coarse Plus are
trademarks of DeFelsko Corporation registered in the U.S. and in other
countries. Other brand or product names are trademarks or registered
trademarks of their respective holders.
Every effort has been made to ensure that the information in this manual is
accurate. DeFelsko is not responsible for printing or clerical errors.
Analog Micrometer Conversion Table
X-Coarse Grade
Reading 1 **Mils| Result 2 ****Mils| Reading
1 ****µm| Result 2 ****µm
1.9| 1.5| 50| 40
2.0| 1.6| 53| 44
2.1| 1.8| 55| 47
2.2| 1.9| 58| 51
2.3| 2.0| 60| 54
2.4| 2.2| 63| 57
2.5| 2.3| 65| 60
2.6| 2.4| 68| 64
2.7| 2.6| 70| 67
2.8| 2.7| 73| 71
2.9| 2.8| 75| 74
3.0| 3.0| 78| 78
3.1| 3.1| 80| 81
3.2| 3.2| 83| 85
3.3| 3.4| 85| 87
3.4| 3.5| 88| 89
3.5| 3.6| 90| 94
3.6| 3.8| 93| 98
3.7| 3.9| 95| 101
3.8| 4.1| 98| 105
3.9| 4.2| 100| 108
4.0| 4.3| 103| 112
4.1| 4.5| 105| 115
| | |
| | |
| | |
| | |
Coarse Grade**
Reading 1 **Mils| Result 2 ****Mils| Reading
1 ****µm| Result 2 ****µm
0.9| 0.8| 20| 19
1.0| 1.0| 22| 21
1.1| 1.1| 24| 24
1.2| 1.3| 26| 27
1.3| 1.4| 28| 29
1.4| 1.6| 30| 32
1.5| 1.7| 32| 35
1.6| 1.8| 34| 37
1.7| 2.0| 36| 40
X-Coarse Plus Grade
Reading 1 ****Mils| Result 2 ****Mils| Reading
1 ****µm| Result 2 ****µm**
4.2| 4.0| 105| 100
4.3| 4.2| 108| 105
4.4| 4.3| 110| 107
4.5| 4.5| 113| 112
4.6| 4.6| 115| 114
4.7| 4.7| 118| 118
4.8| 4.9| 120| 121
4.9| 5.0| 123| 125
5.0| 5.1| 125| 128
5.1| 5.3| 128| 132
5.2| 5.4| 130| 134
5.3| 5.5| 133| 139
5.4| 5.7| 135| 141
5.5| 5.8| 138| 145
5.6| 6.0| 140| 150
This manual is copyrighted with all rights reserved and may not be reproduced
or transmitted, in whole or part, by any means, without written permission
from DeFelsko Corporation.
Testex and Press-O-Film are trademarks of DeFelsko Corporation registered in
the U.S. and in other countries. Other brand or product names are trademarks
or registered trademarks of their respective holders.
Every effort has been made to ensure that the information in this manual is
accurate. DeFelsko is not responsible for printing or clerical errors.