What do I do if I get a change sensor alert?

Firstly, when you receive this alert, take time to read the pump screen to find out why the alert has occurred.
The information below will give you the instructions.
This information is on page 255 of the Mini Med™ 780G System User Guide

Title and text Type Explanation Next steps

Change sensor
Insert new sensor and Start New Sensor.| Alert| No was selected in the Check sensor insertion message, indicating that the sensor is not fully inserted.|

  • Select OK to clear the alert.
  • Change the sensor. For details, see the sensor user guide.
  • After the sensor is changed, refer to Starting the sensor, on page 117.

Change sensor
Second calibration not accepted. Insert new sensor.| Alert| This alert occurs when two Calibration not accepted alerts are received in a row.|

  • Select OK to clear the alert.
  • Change the sensor. For details, see the sensor user guide.

Change sensor
Sensor not working properly. Insert new sensor.| Alert| This alert occurs when the transmitter diagnoses a problem with the sensor that cannot be resolved.|

  • Select OK to clear the alert.
  • Change the sensor. For details, see the sensor user guide.

Check connection
Ensure transmitter and sensor connection is secure, then select OK.| Alert| The pump fails to detect the transmitter and is unable to receive sensor signal.|

  • Select OK to clear the alert.
  • If the sensor is fully inserted, select Yes. If the sensor is not fully inserted, select No.
  • If the sensor was not fully inserted, insert a new sensor.
  • See Pump issues, on page 215 for additional assistance, if needed.

Always follow the directions for use

Information contained herein is not medical advice and should not be used as an alternative to speaking with your doctor. Discuss indications, contraindications, warnings, precautions, potential adverse events, and any further information with your healthcare professional.
©Medtronic. Medtronic, Medtronic logo, and Engineering the extraordinary are trademarks of Medtronic. All other brands are trademarks of a Medtronic company.

Customer Support

2 Alma Road Macquarie Park NSW 2113
Tel: +61 2 9857 9000
Fax: +61 2 9889 5167
Toll Free : 1800 777 808
medtronic-diabetes.com.au

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals