Product Usage Instructions

Before using the forehead thermometer, please ensure you adhere to the following guidelines:

  • Do not disassemble, repair, or modify the thermometer.
  • Clean the thermometer lens after each use.
  • Avoid direct finger contact with the lens.
  • Do not expose the thermometer to extreme conditions.
  • Allow the thermometer and patient to acclimate to the room conditions for 30 minutes before use.
  • Avoid measuring temperature immediately after exercise, bathing, or being outdoors.

To install or replace batteries:

  • Slide the battery cover back gently.
  • Remove old batteries and dispose of them properly.
  • Insert two 1.5V alkaline AAA(LR03) Size batteries with the correct polarity.
  • Slide the battery cover back into place securely.

FAQ

  • What should I do if the low battery warning appears?
  • If the low battery symbol appears, replace the batteries with two new 1.5V alkaline AAA(LR03) Size batteries as soon as possible to ensure proper functionality.
  • How should I clean the thermometer lens?
  • Use a clean, soft cloth with a mild detergent solution to wipe the thermometer lens after each use. Avoid using harsh chemicals or abrasive materials.
  • Can I use rechargeable batteries with the thermometer?
  • Avoid using rechargeable or different specification batteries as it may damage the device. Only use two 1.5V alkaline AAA(LR03) Size batteries for optimal performance.

Introduction

Intended Use
Touch Free Infrared Thermometer is intended for home use and the intermittent measurement of human body forehead temperature in people of all ages.
Features

  1. 2 in 1 thermometer: Human Body/Object

  2. Fever Indication (Human Body Mode only)

  3. °C/°F Switch-able Function

  4. Multi-memory Recall

  5. One-second Reading

  6. Power Saving-Auto Power Off
    Auto power-off after idling to ensure battery life.

  7. Warning indication
    “low battery” symbol, and “Lo”/ “Hi” symbol indicating out of measuring range.

  8. Large LCD Display

  9. Large Button Design

  10. Economic design and convenience
    This is a “Non-contact” medical thermometer that enables temperature readings, designed for sanitary conditions, cleanliness, and convenience. Simply move the thermometer close to the subject’s forehead or object at the distance indicated by the device.

  11. Accurate and Reliable

  12. User Friendly
    Due to our unique design of the measuring software, users don’t need to have special skills to operate this device.

  13. Instant Measurement
    By using our unique technology, users can get their precise body temperature instantly and accurately.

  14. Handicapped persons and children need assistance from another person to use the device.

Important Information Before Use
When using this product, please be sure to follow all the notes listed below. Any action against these notices may cause injury or affect the accuracy.

  1. Do not disassemble, repair, or remodel the thermometer.
  2. Be sure to clean the thermometer lens each time after usage.
  3. Avoid direct finger contact with the lens.
  4. No modification of this equipment is allowed.
  5. It is recommended that the user take 3 temperatures. If they are different, use the highest reading.
  6. Do not expose the thermometer to extreme temperatures, very high humidity, or direct sunlight.
  7. Avoid extreme shock or dropping the device.
  8. Before the measurement, patients and thermometers should stay in steady-state room conditions for at least 30 minutes.
  9. Avoid measuring temperature in 30 minutes after exercise, bathing, or returning from outdoors.
  10. To protect the environment, dispose of empty batteries at appropriate collection sites according to national or local regulations.
  11. Please use the thermometer solely for its intended purpose.
  12. There are no absolute body temperature standards. Keep reliable records of your temperature to serve as a reference for judging a fever.
  13. Under any circumstances, the temperature-taking result is ONLY for reference. Before taking any medical action, please consult your physician.
  14. The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy of temperature measurements, please contact an authorized distributor for sending calibration services, additional cost may apply for return shipping fee.
  15. CAUTION: Please keep this device away from pets, pests, and children.
  16. CAUTION: Please keep away from children and pets, because small parts may be inhaled or swallowed.
  17. If any serious incident has occurred concerning the device should be reported to the manufacturer and the competent authority of the Member State in which the user is established.
  18. If you require any assistance during setting up, using, or maintaining, please contact the manufacturer.
  19. Avoid using the device near an operating cell phone or microwave oven.

Contraindications

  1. Scarred or compromised tissues on measurement site.
  2. The patient’s skin is on a certain medication of measurement site.

Product Identification

  1. Memory Button
  2. Mode Button
  3. Battery Cover
  4. Display LCD
  5. Measure Button
  6. Measurement Sensor
  7. Distance Detection Sensor

AVITA-NT66B-Touch-Free-Infrared-Thermometer-FIG-2

Description of LCD Display

  • Measuring sequence in the process
  • Night Mode (Mute Mode)
  • Human Mode
  • Object Mode
  • Degree of measurement
  • Celsius Scale
  • Fahrenheit Scale
  • Low battery indicator
  • Frowning Face (temperature is equal to or higher than 38°C)
  • Smiling Face (temperature is lower than 38°C)
  • Memory Symbol and sets of records

Battery Installation

Low battery warning
When the battery power becomes low, the low battery symbol “ will appear on the display. The thermometer can still be used during this time, but the batteries should be replaced as soon as possible. If the batteries run out completely, “Lo” will be displayed along with the low battery symbol “ ”. In this case, the batteries will need to be replaced before using the thermometer again.

CAUTION

  1. Risk of explosion if the battery is replaced by an incorrect type.
  2. It is recommended not to use rechargeable, unqualified, or different spec of batteries as it may damage the device or cause a circuit shortcut.

Replacing the Battery

  1. Gently slide the battery cover back.
  2. Carefully remove the old batteries and properly discard them.
  3. Insert new batteries (Two 1.5V alkaline AAA(LR03) Size) according to the proper polarity.
  4. Slide the battery cover back on.

NOTE: Battery-operated

  1. Please properly dispose of the batteries away from small children and heat.
  2. It is recommended to remove the batteries if the unit will not be used for an extended time.
  3. Batteries must be disposed of under local environmental and institutional policies.
  4. Dispose of used batteries under the applicable legal regulations. Never dispose of batteries in the normal household waste.

Switching Between Fahrenheit and Celsius
Your thermometer can display results in either degrees Celsius (°C) or degrees Fahrenheit (°F). With a combination of Memory and Mode buttons can change the mode to either °C or °F with a beep sound.
Switching between 4 Kinds of Measuring Mode

  1. Under power-on status, you can press the Mode Button to switch different measuring modes. There are 4 kinds of modes which include Human, Object, Human/Night, and Object/Night Mode (in order).
  2. The beep sounds will be off (mute) when you choose in Human/Night Mode, and Object/Night Mode, and the Moon symbol “ ” will appear on the LCD in both Night Modes.AVITA-NT66B-Touch-Free-Infrared-Thermometer-FIG-14

NOTE
Each press will come with a beep sound to ensure the setting is activated. (Except both Night Modes)

Tips for Measuring Human Temperature
Bear in mind that the thermometer needs to have been in the room in which the measurement is taken for at least 30 minutes before use.

NOTE

  • Attempting to take temperature readings from sites on the body other than the forehead may produce inaccurate results.
  • The patient should remain still while the reading is being taken.
  • Readings taken while asleep should not be compared directly to readings taken while awake, as body temperature while asleep is typically lower.
  • Do not take body temperature readings within 30 minutes of being outdoors, exercising, or bathing.

Measuring Human Temperature
Taking a Measurement

  1. Press the Power Button to power on the thermometer. The unit will run a self-test and the LCD will briefly display all of its symbols during this time. When the device is ready, “–“appears on the screen, and the thermometer is ready to take a measurement.

  2. Select the desired mode by pressing and releasing the Human/Object Button.
    Note

    • When taking a patient’s temperature, ensure that the thermometer is in Human Mode; the Human symbol “ ” will appear on the display.
    • If the patient’s skin is covered with hair, sweat, or dirt, clean the area and wait 10 minutes before taking a measurement.
    • Ensure that the thermometer is held firmly during measurement and that the patient does not move until the measurement is complete. Movement can impact the measurement.
  3. Position the thermometer under 3cm (around 1 inch) from the center of the patient’s forehead with the sensor aimed between the eyebrows.

  4. Press and release the Measure Button.

  5. Slowly move the device toward or farther from the forehead until you have reached the correct distance. (If the distance is beyond the correct distance, the dash icon on the display flashes in this sequence until the correct distance is achieved.)AVITA-NT66B-Touch-Free-Infrared-Thermometer-FIG-15
    Note
    When the user presses the Measure Button, this workflow begins and they require time for the device to capture the temperature. After capturing the temperature, the display turns to standby mode.

  6. As you hear a short beep means this temperature reading has been completed and accompanied with a back-light.

  7. If the temperature measurement is below 38°C, a “Smiling Face” will appear next to the reading. If the reading is 38°C or above, a “Frowning Face” will display and the LED light up to alert.

  8. After about 60 seconds after use, the thermometer will automatically beep and shut off.

Fever Alarm
Only in Human Mode and Human/Night Mode.

Measuring Object/Liquid Temperature

  1. Press the “Measure Button” to turn the thermometer on. All symbols on the display will momentarily appear.
  2. Ensure that the thermometer is in Object Mode; the Object symbol “ ” will be on the display. To alternate between modes press and release the Human/Object Mode Button until you see the desired measurement symbol on the display.
  3. Position the thermometer under 3cm (around 1 inch) from the object.
  4. Release the “Measure Button” and the temperature reading will be displayed.
  5. After use, the thermometer will automatically beep and shut off after idling.

Memory Function

Memory Recall
You can recall up to 10 measurements currently stored in memory to share with your physician or trained healthcare professional.

  1. When the device is on, press once briefly on the Memory Button, then pass it again to show the last measurement accompanied by “symbol.
  2. The “symbol or “symbol will appear with each measurement stored in memory to indicate whether a person or object’s temperature was taken.
  3. Each press of the same button recalls a previous measurement.

Memory Deletion

  1. Under power-on status, you may hold press the “Memory button” to delete all the readings. The LCD shows “- -” with one beep sound to indicate that all memories are cleared.
  2. Remove batteries, memory will be cleared.

NOTE
All the readings will be cleared matter record in Human Mode or Object Mode.

Cleaning and Disinfecting

For home-use device disinfection, 75% alcohol (available in the pharmacy) can be used.

  • Measurement Sensor
    Clean the measurement sensor with an alcohol swab before and after each measurement.

  • Thermometer:
    Use a soft, dry cloth to clean the thermometer body. Never use abrasive cleaning agents, thinners, or benzene for cleaning. Do not scratch the surface of the probe lens or the display. Do not expose the thermometer to extreme temperatures, humidity, direct sunlight, or shock.

Error Codes
When a malfunction or incorrect temperature measurement occurs, an error message will appear as described below.

Display Cause Solution
The temperature measured is higher than the measurement range. Human mode:

43°C (109.4°F)

Object mode:        100°C (212.0°F)

| Operate the thermometer only between the specified temperature ranges. If necessary, clean the sensor tip. In the event of a repeated error message, contact your retailer or Customer Services.
| The temperature measured is lower than the measurement range. Human mode: 34°C (93.2°F)

Object mode:           0°C (32.0°F)

| The operating temperature is not in the range:

15°C~40°C (59°F~104°F)

| Operate the thermometer only between the specified temperature ranges.

Troubleshooting

Trouble Probable Case Recommended Action
Failed to power on. Batteries are drained. Replace with a new battery

Batteries are not correctly aligned with

terminals.

| Reinsert batteries correctly.
The thermometer is

damaged.

| Contact your retailer or

Customer Services.

A low battery symbol appears.| Low battery.| Replace with a new battery as soon as

practical

In colder temperatures, batteries have weaker electrical charges.| Warm up the batteries or use the device in a warmer setting.
The measurement is abnormal or if there is any doubt about the measured result| Incorrect measure position.| Reposition the probe point center of the

forehead.

The measuring distance is too far.| Maintain a distance of 3 cm.
The measuring is hampered by hair, perspiration, oil, sweat,

makeup, etc.

| Remove obstructions from the forehead.
The lens of the probe is dirty.| Clean the lens according to the cleaning

and disinfecting.

You have just come| Stay in a steady state
| from extreme temperatures

environment.

| room at least for 30 minutes.
---|---|---
The thermometer has been stored in a cold room or exposed to

direct sunlight.

| Put the thermometer in a steady-state room at least for 30 minutes.
Measurement was taken in surface

temperature mode.

| Switch mode to human mode.
The power switches off automatically.| System design.| Push the power button

again.

Batteries are drained.| Replace with a new

battery

The thermometer is

damaged.

| Contact your retailer or

Customer Services.

Operating the Bluetooth function

This product is a TOUCH-FREE Infrared Thermometer with Bluetooth function. design without entering personal information. If the device has a transmission function, the transmission measurement data is designed to be encrypted and transmitted, and will not be tampered with or retrieve user-related information during the transmission process. The firmware and software of the product have been programmed in the production process, and the programming interface is different from the data transmission interface. When programming to the microcontroller, use an encrypted programmer, so there is no need to worry about the software being tampered with during transmission.

Bluetooth function requirement

  • An Android device with Android version 4.3 or above and hardware support for Bluetooth 4.2.
  • An iOS device with iOS version 5 or above and hardware support for Bluetooth 4.2.

How to activate the Bluetooth function:
Please refer to the instruction manual of your mobile phone or computer for how to activate the Bluetooth function.
Set-Up Process

  1. Please determine if your mobile phone or computer has BLE4.2.
  2. Turn on the NT66B device, when the Bluetooth icon shows on the device, it means the NT66B device is under broadcast condition.
  3. Please check the connecting condition from your mobile phone or computer. The device name should be “NT66B”.
  4. Every measure reading will be transferred to your mobile phone or computer automatically.

Technical Specification

  • Measuring range :
    • Human Body : 34°C~43°C (93.2°F~109.4°F)
    • Object : 0°C~100°C (32.0°F~212.0°F)
  • Calibration Accuracy:
    • Human Body : ±0.2°C(±0.4°F): from 34°C~43°C (93.2°F~109.4°F)
    • Object : < 40°C ± 2°C; >= 40°C ± 5%
  • Display resolution: 0.1°C
  • Operating environment: 15°C~40°C (59°F ~ 104°F) with a relative humidity of 15% to 95% (noncondensing)
    • Atmospheric pressure: 700~1060 hPa
  • Storage/ Transportation environment: -25 to 55 °C(-13 to 131 °F) with a relative humidity of 15% to 95% (noncondensing)
  • Atmospheric pressure: 700~1060 hPa
  • Power supply: 2 x 1.5V AAA size alkaline batteries
  • Weight: approx. 120g (with batteries)
  • Memory: 10 sets
  • Dimensions: approx. 142mm×42mm×43.9mm (L×W×H)
  • Operation Distance: 3 cm
  • Frequency: 2402~2480GHz
  • Output power range: <=4dBm
  • Battery Operation Time: about 90 minutes (Depending on different brands of batteries.)
  • Shelf Life :3 years
  • Reference body site: Ear temperature
  • For Customer Service

To obtain further service please contact AViTA Corp. for the address of the repair location. Enclosed is the Proof of Purchase. Include USD 10.00 for the return shipping and handling. Include a letter, with your name, address, phone number, and description of the specific problem. Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend ensuring the product with the return receipt requested.

EMC Tables

NT66B is intended for use in the electromagnetic environment specified below. The customer or the user of NT66B must make sure that it is

used in such an environment.


Guidance and manufacturer’s declaration – Electromagnetic emissions
Phenomenon| Professional healthcare facility environment a)| HOME HEALTHCARE

ENVIRONMENT a)

Conducted and radiated RF MISSIONS| CISPR 11

Group 1 Class A (Not BLE Function)

Group 2 Class A (With

BLE Function)

| CISPR 11

Group 1 Class B (Not BLE Function)

Group 2 Class B (With

BLE Function)

Harmonic distortion| Not applicable

(Note: Power by Battery or DC Input)

Only the AC input needs to be tested

Voltage fluctuations and flickering| Not applicable

(Note: Power by Battery or DC Input)

Only the AC input needs to be tested

a) The equipment is suitable for use in Home Health Environments and Professional Health Care Environments limited to patient rooms and respiratory treatment facilities in hospitals or clinics. The more restrictive acceptance limits of Group 1 Class B (CISPR 11) have been considered and applied. The equipment is suitable for use in the mentioned environments when directly connected to the Public Mains Network.

b) The test is not applicable in this environment unless the ME EQUIPMENT and ME SYSTEM used will be connected to the PUBLIC MAINS NETWORK and the power input is otherwise within the scope of

the Basic EMC standard.

Guidance and manufacturer’s declaration – Electromagnetic immunity

– Enclosure port

Phenomenon| Basic EMC standard or test method| Immunity test levels
Professional healthcare facility

environment

| HOME HEALTHCARE ENVIRONMENT
ELECTROSTATIC

DISCHARGE

| IEC 61000-4-2| ± 8kV contact

± 2 kV, ±4kV ±, ±8 kV, ±15 kV air

Radiated RF EM

fields

| IEC 61000-4-3| a)| 10 V/m b) 80MHz – 2.7

GHz 80% AM at 1kHz

Proximity fields from RF wireless communications equipment| IEC 61000-4-3| COMPLIANT

NOTE: Further information about distances to be maintained between portable and mobile RF communications equipment (transmitters) and the NT66B can be requested from the supplier using the contact information provided in this manual. However, it is advisable to keep the equipment at an adequate distance of, at least, 0.5 m from mobile phones or other RF communications transmitters to minimize possible interference.

---|---|---
RATED power frequency

magnetic fields.

| IEC 61000-4-8| 30 A/m c)

50 Hz or 60 Hz

a) The equipment is suitable for use in Home Health Environments and Professional Health Care Environments limited to patient rooms and respiratory treatment facilities in hospitals or clinics. The more restrictive IMMUNITY acceptance limits have been considered and applied.

b) Before modulation is applied.

c) This test level assumes a minimum distance of at least 15 cm between the ME EQUIPMENT or ME SYSTEM and sources of power frequency

magnetic fields.

FCC STATEMENT

Federal Communications Commission interference statement

This equipment has been tested and found to comply with the limits for a Class B digital device, under part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used by the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and the receiver.
  • Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/ TV technician for help.

CAUTION
Any changes or modifications not expressly approved by the grantee of this device could void the user’s authority to operate the equipment.

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:

  1. this device may not cause harmful interference, and
  2. this device must accept any interference received, including interference that may cause undesired operation.

RF exposure warning
The equipment complies with FCC RF exposure limits set forth for an uncontrolled environment.
The equipment must not be co-located or operating in conjunction with any other antenna or transmitter.

Explanation of Symbols

AVITA-NT66B-Touch-Free-Infrared-Thermometer-FIG-22 AVITA-NT66B-Touch-Free-Infrared-Thermometer-
FIG-23 AVITA-NT66B-Touch-Free-Infrared-
Thermometer-FIG-24 AVITA-NT66B-Touch-Free-
Infrared-Thermometer-FIG-25

CONTACT

  • MDSS GmbH
  • Schiffgraben 41
  • 30175 Hannover Germany
  • AViTA Corporation
  • 9F., No. 78, Sec. 1, Kwang Fu Rd.,
  • San Chung Dist., New Taipei City 24158, Taiwan, China
  • China, AViTA(WUJIANG)

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

AVITA User Manuals

Related Manuals