VICTOR 4090C Digital LCR Meter User Manual
- June 26, 2024
- VICTOR
Table of Contents
VICTOR 4090C Digital LCR Meter
Install
This chapter describes some of the checks you must make after you receive the instrument, and the conditions you must know and have before you install it.
Unpacking
Thank you for buying and using our products. After unpacking, you should first
check whether the instrument is damaged due to transportation. We do not
recommend you to power the instrument under the condition of appearance
damage.
Confirm that the bridge is attached with the following accessories. Optional
accessories need to be ordered separately. If any items are missing, please
contact the nearest sales office.
The standard accessories:
- A kelvin clamp.
- A power cord.
- A user manual.
Optional accessories:
- RS232 cable.
- USB cable.
- Handler cable
Power supply
- Voltage: AC220V±10% or AC110V ±10%, optional.
- Frequency: 45Hz-65Hz.
Warning : in order to prevent leakage of electricity from harming the instrument or people, the user must ensure that the earth wire of the power supply is connected to the earth reliably.
Fuse
The instrument is factory fused, users should use the company’s fuse.
Surrounding environment
- Please do not use in the environment with many dust, vibration, direct sunlight and corrosive gas.
- The normal operating temperature of the instrument is 0-40 ℃, and the working humidity is 15%-85% RH.
- The instrument has been carefully designed to reduce the clutter caused by AC power input, however, it should still be used in a low-noise environment as far as possible. If it cannot be avoided, please install a power filter.
- Keep away from strong electromagnetic field during instrument measurement, so as to avoid interference to the measurement.
Use Kelvin clamp
Please use the Kelvin clamp provided by our company. The Kelvin clamp and the
pin of the tested part should be kept clean to ensure the good contact between
the tested part and the Kelvin clamp. The Kelvin clamp is connected to the
four test terminal of Hcur, Hpot, Lpot and Lcur in the front panel of the
instrument. For a test part with a shielded case, the shield may be connected
to the earthing terminal of the front panel of the instrument. Two terminals
of Hcur and Hpot are connected to the red clamp, while two terminals of Lpot
and Lcur are connected to the black clamp.
Warm-up
- In order to ensure the accurate measurement of the instrument, the startup warm-up time should be no less than 30 minutes.
- Do not turn on and off the power of the instrument frequently, which may cause internal data confusion.
Other characteristics of the instrument
- Power : Less than 20W.
- Dimensions : 265mm×105mm×305mm(width × height ×length).
Brief
This series digital bridge adopts 3.5-inch LCD, has friendly user interface and various measurement functions, supports multiple test frequencies and test levels, with basic accuracy up to 0.1%. USB and RS232 can facilitate remote operation. The Handler interface facilitates the automatic sorting of components. This manual takes “C Type”(100kHz) as an example.
Basic features
- Measuring frequency up to 100 kHz.
- Level range (100-2000) mVrms.
- Internal bias voltage (0-1500) mV.
- Automatic identification of component type.
- Range can be automatically or locked.
- Support DC resistance, electrolytic capacitance measurement.
- Basic accuracy 0.1%.
- Support open, short calibration.
- Data recording function.
- 3.5-inch LCD, supports English and Chinese interface.
- USB, RS232 communication interface, support SCPI communication protocol.
- Handler interface, support comparator sorting, list scan.
Table 2- 1
Model | A Type | B Type | C Type |
---|
Excitation source frequency (Hz)
|
10 points(100、120、200、400、800、1k、2k、4k、8k、10k)
| 12 points(100、120、200、400、800、1k、
2k、4k、8k、10k、15k、
20k)
| 16 point(s 100、120、200、
400、800、1k、2k、4k、
8k、10k、15k、20k、40k、
50k、80k、100k)
Digits| Main parameter: 5 digits; secondary parameter: 5 digits
Parameter| Main parameter: L/C/R/Z; secondary parameter: X/D/Q/θ/ESR
Range(L)| 0.001μH – 9999H
Range(C)| 0.001pF – 9999μF
Range(R)| 0.0001Ω – 99.99MΩ
Basic Accuracy| 0.1%
---|---
Measurement Display Speed| 2 times/second (slow), 5 times/second (medium
speed), 10 times/second (fast)
Internal Bias| 0-1500mV adjustable, 1mV stepping
Test Level(Vrms)| 100mV、300mV、600mV、1000mV、1500mV、2000mV
Correction Function|
Open circuit correction, short circuit correction
Screening function| Tolerance range: -100% ~ +100%, with fixed points of 10%, 25% and
50%
Selection of comparator|
Five gear sorting, third gear qualified, one gear failed, one ancillary gear
Communication Interface| Standard configuration: USB, RS232 (or 485), Handle interface; Optional: GPIB, USB Host
Others
| Support DCR, electrolytic capacitor measurement mode, adjustable backlight brightness, optional in English and Chinese
Quick reference
The front panel
The rear panel
Keys
Power switch
Power switch: Press down to power on and pop up to power off.
The direction keys
Left and right direction keys to control cursor movement, up and down
direction keys to select parameters.
Basic operations
- Slide the voltage selector to select the voltage consistent with the power supply to be used. Example: the power supply to be used is 220V AC power, select the voltage selector to display 220V.
- Connect the power supply.
- Connect Kelvin clamp. Two terminals of Hcur and Hpot are connected to the red clamp, while two terminals of Lpot and Lcur are connected to the black clamp.
- Press down the power switch, LCD of front panel light up.
- Power on the default into the LCR measurement page. See figure 3-3. Where “–OL–” means the measurement data over the range.
- Connect part to be measured to the kelvin clamp and start the measurement. Example: connect the resistance of about 1KΩ to the kelvin clamp, and the page is shown in figure 3-4.
LCR page
-
Page title
Used to identify the page to display. LCR, DCR, ECA, COM SET, SYSTEM SET, LIST SCAN 6 pages. The measurement page in the document refers to the pages of LCR, DCR, ECA. -
Measurement conditions and parameter
-
Measurement result display Main parameter is Rp, said parallel equivalent models to measure resistance, the measured results is 999.4 Ω; Secondary parameter is X , said the reactance, measuring result is 0.0567 Ω.
-
The message column The LOCAL/REMOTE identification is displayed in the far right of the message column. The switch between local and remote needs to be carried out by SCPI command. In remote mode, the keyboard cannot be used. The instrument starts in local mode by default.
Basic function operation
Measurement conditions and parameters
Function
Cp-D is displayed at the FUNC, as shown in figure 4-1. The first character “C” is the main parameter, represents capacitance measurement; the second character “p” represents parallel equivalent model; the last character “D” is the secondary parameter, represents loss factor.
-
Main parameter
The main parameters include AUTO, L (inductance), C (capacitance), R (resistance) and Z (impedance). When selecting AUTO, the instrument will automatically judge whether the measurement is L, C or R according to the measured data. In this case, the secondary parameter cannot be set and the fixed combination is R-X, C-D and L-Q. The equivalent model is automatically switched according to the measurement data. In the measurement page, Press key to switch to the next main parameter(AUTO、R、C、L、Z), Loop switching. -
Equivalent model
Equivalent models include s (series) and p (parallel). The actual inductance, capacitance and resistance are not ideal pure reactance or resistance parts. They are usually presented as a composite impedance element. Herein, the part to be measured is equivalent to a simple series or parallel model of two elements, then performs measuring and calculating.
Note : generally, the series equivalent model is used for low impedance parts and the parallel equivalent model is used for high impedance parts.
In the measurement page, Press key to switch the equivalent model(s、p).
Secondary parameter
The secondary parameters include X (reactance), D (loss factor), Q (quality
factor), θ (impedance Angle), and ESR (equivalent series resistance). In the
measurement page, Press key to switch to the next secondary parameter(X、D、
Q、θ、ESR), Loop switching.
Level
There are two ways to set the level.
- Method one: In the LCR page, move the cursor to the LEVEL item with the left and right keys, and then switch to the previous or next level point by pressing the up and down keys (100mV, 300mV, 600mV, 1V, 1.5V, 2V) , Loop
- switching.
Method two: In the LCR page, press LEVEL key to quickly switch to the next level point, Loop switching.
Bias
In the LCR page, move the cursor to the BIAS item with the left and right
keys, press ENTER the input state. The page is shown in figure 4-2. key
to
Display in the BIAS a current bias set (For example, if the front bias is 1V, the bias in the field is 1000mV), Select the numeric digit to be changed by left and right keys, and the selected digit will be displayed in reverse. Change the value of the digit by up and down keys (up keys +1, down keys -1). Press the key ENTER to exit after setting.
Frequency
There are two ways to set the frequency.
- Method one: In the measurement page, move the cursor to the FREQ item with the left and right keys, and then switch to the previous or next frequency point by pressing the up and down keys(The fixed frequency point are:100Hz、120Hz、200Hz、400Hz、800Hz、1k、2k、4k、8k、10k、15k、20k、40K、50K、80K、 100K), Loop switching. The page is shown in figure 4-3.
- Method two: In the measurement page, press FREG key to quickly switch to the next frequency point, Loop switching.
Range
The instrument range can be set to automatic(AUTO) or locked(LOCK). When the
measuring range is set to automatic, the instrument will automatically adjust
the appropriate measuring range according to the impedance of the part to be
measured. When locked, the range is locked at the current range. In general,
it is recommended to use automatic range. Range locked is used for batch
testing in the factory to save the stable time of range switching. The usage
is as follows: measure one of the part in the range automatically, then lock
it in the current range and start batch measurement. There are two ways to set
the range.
- Method one: In the measurement page, move the cursor to the RANGE item with the left and right keys, and then switch by pressing the up and down keys(LOCK、AUTO).
- Method two: In the measurement page, press Range key to quickly switch.
Output impedance
The output impedance is the output impedance of the excitation source, the
instrument provides two choices: 30 Ω and 100 Ω. By default, the output
impedance is 100 Ω. When output impedance select 30 Ω and the measured
impedance is small, overcurrent may occur. As shown in figure 4-4, the main
parameter location “– OC –“. In the measurement page, move the cursor to the
ROUT item with the left and right key, then by up and down key switch output
impedance (30 Ω、100 Ω)
Speed
There are two ways to set the measurement speed.
- Method one:
- In the measurement page, move the cursor to the SPEED item with the left and right keys, and then switch to the next speed point by pressing the up and down keys(FAST、MED、SLOW), Loop switching.
Method two: In the measurement page, press Speed key to switch to the next speed point, Loop switching.
DCR mode
Press DCR key enter DC resistance measurement mode and press again to
exit. The page is shown in figure 4-5. The level, bias and frequency cannot be
changed in DC resistance measurement mode.
DC resistance mode measures the DC impedance of coils, transformers, etc.
Electrolytic capacitor mode
Press key enter electrolytic capacitor mode and press again to exit. The level
and bias cannot be changed in electrolytic capacitor mode. The page is shown
in figure 4-6. It is necessary to pay attention to the connection polarity of
electrolytic capacitor to be measured. The red clamp is connected to the
positive electrode of electrolytic capacitor; the black clamp is connected to
the negative electrode of electrolytic capacitor.
Δ function
In the measurement page, Press NULL key to turn on/off the Δ function.
When the Δ function is turned on, the instrument records the main parameter of
the measured value at the time of turning on as the reference value and
displays it at the position of the secondary parameter. The position of the
main parameter displays the difference between the current measured value and
the reference value. The page is shown in figure 4-7.
Data retention function
Press HOLD key to turn on the data retention function, as shown in figure
4-8, the red “ HOLD ” is displayed in the upper right corner of the page
and flashes, and the page data remains unchanged. Press key again to turn off
the data retention function.
Data recording function
Press MAX key to show maximum, Press MIN again cancel display. Press
key to show minimum, Press AVG again cancel display. Press key to show
average. Press again cancel display. The page is shown in figure 4-9. The
third column of the message column displays MAX (maximum value), MIN (minimum
value), AVG (average value) label and data units, and the fourth column
displays the corresponding data. The message column data in figure 4-9 shows
that the maximum value is 1.5818pF, which is the maximum value of all measured
data (main parameter) from the moment when the key MAX is pressed down to
the current.
Calibration
Open and short circuit calibration is used to eliminate the measurement error
introduced by the distributed parameters of the instrument and fixture. Open-
circuit calibration is usually performed before large impedance measurements
and short-circuit calibration before small impedance measurements. If accurate
measurements will be made, it is recommended to perform open and short circuit
calibration prior to measurement.
Open circuit calibration
Open the test clamp (red and black clamps are not connected to any element).
Stay away from human body and interference source more than 10 cm. Press
CAL key to enter calibration, the page is shown in figure 4-10. As the
calibration progresses, the right side number is automatically reduced by 1
until it is reduced to 0. When calibration is completed, the beep will give a
alert and then automatically return to the measurement page
Short circuit calibration
Short circuit the test clamp (red and black clamps together), Press CAL
key to enter calibration, The system automatically identifies open circuit,
short circuit, both are not: display blank, two beeps of notes ) ,The page is
shown in figure 4-11, As the calibration progresses, the right side number is
automatically reduced by 1 until it is reduced to 0. When calibration is
completed, the beep will give a alert and then automatically return to the
measurement page.
System settings
SYSTEM SET page
In the measurement page. Press the key SET to enter the SYSTEM SET page.
The SYSTEM SET page has two pages, see figure 5-1 and figure 5-2. Left and
right keys can move the cursor and turn pages. Press SET again to exit the
SYSTEM SET page.
System parameters
The options of system parameters is shown in the table below.
Table 5- 1
Items | Options |
---|---|
Language | 中文/English |
Brightness | 30%,50%,70%,100% |
Power on | Default set, Last set |
Beep | OFF / ON |
TRIG source | INT, MAN, EXT |
PASS tone | Long, Short, Off |
FAIL tone | Long, Short, Off |
-
Language
Move the cursor to “Language” item by left and right keys, and select language (English, 中文) by up and down keys. -
Brightness
Move the cursor to “Brightness” item by left and right keys, and select backlight brightness (30%, 50%, 70%, 100%) by up and down keys. Loop switching. -
Power on
Move the cursor to “Power on” item by left and right keys, and select power on (Default set, Last set) by up and down keys. When the power on is set to the Last value, the function, level, frequency, bias, range, output impedance and measurement speed of the measurement page are the setting parameters when the
power off. If “Power on” is set to the Default set, these parameters revert to the default value. -
Beep
Move the cursor to “Beep” item by left and right keys, and switch beep (OFF, ON) by up and down keys. When switch beep on, a beep with a key (a short sound), switch beep off, no beep with a key. This parameter mainly controls the beep of the key, and has no effect on beep of notes and alerts. -
TRIG source
- Move the cursor to the “TRIG source” item by left and right keys., then switch the trigger source (INT, MAN, EXT) by the up and down keys. Loop switching. When set as INT, trigger signal is generated inside the instrument. When the trigger source is set to MAN, press the “TOL%” key of the front panel to generate a trigger signal manually. When the trigger source is set to EXT, the Handler interface receives the trigger signal. Every time it is triggered, the data is updated.
- When the trigger source selects MAN or EXT, exit SYSTEM SET and return to the measurement page, as shown in figure 5-3. The trigger mode is displayed in the first column of the message column, which represents MAN trigger. Press the “TOL%” key of the front panel each time to generate a trigger signal and then update the result data. A blank or last measured value is displayed at the page before the data is updated.
-
PASS tone
Move the cursor to the “PASS tone” item with the left and right keys, then switch the PASS tone style (Long, Short, Off) by the up and down keys. Loop switching. Each trigger signal with one beep, long style for 80 ms, short for 20 ms. -
FAIL tone Move the cursor to the “FAIL tone” item with the left and right keys, then switch the FAIL tone style (Long, Short, Off) by the up and down keys. Loop switching. Each trigger signal with one beep, long style for 80 ms, short for 20 ms.
Comparator
The comparator function is used for inspection of warehousing, and for the measurement, discrimination of components on production line. Together with the use of Handler interface, make it suitable for automatic sorting measurement system.
Comparator parameter setting
Nominal value
- Step 1: In the measurement page, place the reference part on the instrument test clamp for measurement.
- Step 2: Press COMPARE key to enter the COM SET page, the cursor position defaults to the “Nom” item, Nominal value obtained by taking 4 significant digits of reference part’s value. If the nominal value needs to be adjusted, move the cursor to the “Nom” item with the left and right key and press the key ENTER to enter input state, Select the numeric digit to be changed by left and right keys, and the selected digit will be displayed in reverse. Change the value of the digit by up and down keys (up keys +1, down keys -1). Press the key ENTER to exit after setting.
Note : every time you enter the COM SET page, the nominal value will be updated according to the current measurement value.
Comparator switch
- Method one: In the COM SET page, move the cursor to the “Comp” item with the left and right keys, and then ON/OFF the comparator by the up and down keys.
- Method two: In the measurement page, move the cursor to the “Comp” item with the left and right keys, and then ON/OFF the comparator by the up and down keys.
Tolerance
-
Method one: Use the left and right keys to move the cursor to the Low and High limits of the set classification row(1, 2, 3, 2nd), press up and down keys to select the next tolerance point..(-50%、-25%、-10%、0%、
10%、25%、50%). Loop switching. -
Method two: Use the left and right keys to move the cursor to the Low and High limits of the set classification row(1, 2, 3, 2nd), press ENTER to enter input state.(set range : -99.99%~+99.99%, resolution :0.01%). The setting method can refer to bias input state setting.
Note : Press the key SAVE to store currently setting. and the setting after power down is not lost.
Sorting mechanism
The machine provides five levels (BIN:1, BIN:2, BIN:3, BIN:AUX, BIN:OUT) for
sorting. The sorting process is shown in figure 6-2. P is the main parameter
of the measurement result, Pn_L is the low limitof the main parameter of n
level (n=1,2,3), Pn_H is the high limit of the main parameter of n level
(n=1,2,3), 2nd is the measurement result of the secondary parameter, 2nd_L is
the low limit of the secondary parameter, and 2nd_H is the high limit of the
secondary parameter. When the main parameter of the measurement results meets
the first level and the secondary parameter passes, the sorting result is
BIN:1. When the measurement results show that the main parameter does not meet
the first level, but the second level, and the secondary parameter passes, the
sorting results are BIN:2. When the measured results show that the main
parameters do not meet the first and second levels, but meet the third level,
and the secondary parameters passes, the sorting results are BIN:3. When the
main parameters of measurement results do not meet the first, second and third
levels, the sorting results are BIN:OUT. When the main parameters of the
measured results meet the first, second or third levels, and the secondary
parameters does not pass, the sorting result is BIN:AUX. When the comparator
is opened, the sorting results are displayed on the measurement page, as shown
in figure 6-3. At the same time, the sorting results are output through the
Handler interface.
If the result of sorting is BIN:1, BIN:2, BIN:3, BIN:AUX file, the beep alert will prompt according to the PASS tone setting in the system setting. If the sorting result is BIN:OUT, a beep alert will prompt according to the FAIL tone setting in the system Settings.
List scan
The list scan function is mainly used to test the same part at different frequencies.
Parameter setting and operation
Enter the LIST SCAN page. In the LCR page, move the cursor to LIST item with
the left and right keys, as shown in figure 7-1. Press the up and down keys to
enter the LIST SCAN page, as shown in figure 7-2.
Note : When the FUNC item is AUTO, the LIST SCAN page cannot be entered.
Set the scan mode.
On the LIST SCAN page, move the cursor to “Mode” item with the left and right
keys, and switch modes by the up and down keys.
There are two modes, Seq and Step. With the setting Seq, each start signal will automatically complete all measurement and comparison of sequence number 1-7. With the setting Step, each start signal completes the measurement and comparison of the current sequence number. The current sequence number increased by 1 (if the current sequence number is 7, the sequence serial number becomes 1).
Set parameters.
The parameters that can be set for each sequence number include frequency,
high limit and low limit. Setting of Low and High limits: As shown in figure
7-3, move the cursor to the high or low limit field to be set and press the
key ENTER to enter input state. Select the numeric digit to be changed by
left and right keys, and the selected digit will be displayed in reverse.
Change the value of the digit by up and down keys (Up key value increases by
1, down key value decreases by Press the key ENTER to exit after setting.
In the non-input state, press the up and down key to disable the field and
display as “——–“.
The frequency setting is consistent with the frequency setting of the measurement page. In the non-input state,a disable option is added to the frequency options and displayed as “——-”.
Start scan.
It can be started by the external signal line (trigger signal of Handler) or
by pressing the key TOL. In this case, the internal trigger source(INT)
is used, regardless of the TRIG source setting in SYSTEM SET page. After each
measurement and comparison, the measurement value and comparison result are
displayed on the right side of the current sequence number row, as shown in
Figure 7-4. The comparison result is I, the measured value is within the low
and high limits. The comparison result is H, the measured value is higher than
the high limit. The comparison result is L, the measured value is lower than
the low limit. Note: when the frequency field is set to disable option, the
current sequence number will be skipped directly. If both the high and low
limits are in the disable state, there is no comparison result, which is
displayed as “–“.
Output the scanning result in the Handler interface.
The list scan results can be output through the Handler. The comparison
results of sequence number 1, 2, 3 and 7 can be directly output through the
Handler. At the same time, the overall scan results of sequence number 1-7 can
also be output through the Handler. At this time, as long as comparison
results of H and L appear in sequence number 1-7, it will be judged as the
overall unqualified. Clear all scan results each time the measurement of
sequence number 1 is started. See 9.3 Handler interface for specific output
signals and pins. The list scan result can also give the beep prompt, the
comparison result is I, and the beep prompt is based on the PASS tone setting;
the comparison result is H, L, the beep prompt is based on the FAIL tone
setting. The overall scan result are given in Seq mode, and the comparison
results of current sequence numbers are given in Step mode.
Exit the LIST SCAN page.
In the LIST SCAN page, move the cursor to the “LIST” item with left and right
keys, and exit the LIST SCAN page by the up and down keys.
Characteristics
Basic accuracy
The following data were measured under the following conditions:
- Temperature condition: 23℃±5℃
- Humidity conditions: ≤65% R.H.
Calibration: open circuit and short circuit calibration before measurement Warm-up time: >30 minutes
- C: 0.10%*(1+ Cx/Cmax+ Cmin/Cx )(1+Dx)(1+ks+kv+kf);
- L: 0.10%*(1+ Lx/Lmax+ Lmin/Lx )(1+1/Qx)(1+ks+kv+kf);
- Z: 0.10%*(1+ Zx/Zmax+ Zmin/Zx )(1+ks+kv+kf);
- R: 0.10%*(1+ Rx/Rmax+ Rmin/Rx )(1+Qx)(1+ks+kv+kf);
- D: ±0.0010(1+ Zx/Zmax+ Zmin/Zx )(1+Dx+DxDx)(1+ks+kv+kf );
- Q: ±0.0010*(1+ Zx/Zmax+ Zmin/Zx )(Qx+1/Qx )(1+ks+kv+kf );
Among them:
- L, C, R and Z are relative errors. D, Q are absolute errors.
- Where the subscript is x, is the measured value of this parameter; where the subscript is Max, is the maximum value; and min is the minimum value.
- Ks is the speed factor, kv is the voltage factor and kf is the frequency factor. The maximum and minimum values of measurement parameters that affect accuracy, as shown in the following table.
Table 8- 1
frequenc
y(Hz)
| 100| 120| 200| 400| 800| 1K| 2K| 4K| 8K
---|---|---|---|---|---|---|---|---|---
Cmax| 15.9| 13.25| 7.95| 3.98| 1.99| 1.59| 0.795| 0.398| 0.199
Cmin| 15.9| 13.25| 7.95| 3.98| 1.99| 1.59| 0.795| 0.398| 0.199
Lmax| 159| 132.5| 79.5| 39.8| 19.9| 15.9| 7.95| 3.98| 1.99
Lmin| 159| 132.5| 79.5| 39.8| 19.9| 15.9| 7.95| 3.98| 1.99
Z/Rmax| 100000
Z/Rmin| 100
frequenc
y(Hz)
| 10K| 15K| 20K| 40K| 50K| 80K| 100K
Cmax| 0.159| 0.106| 0.0795| 0.0398| 0.0318| 0.0199| 0.0159
Cmin| 0.159| 0.106| 0.0795| 0.0398| 0.0318| 0.0199| 0.0159
Lmax| 1.59| 1.06| 0.795| 0.398| 0.318| 0.199| 0.159
Lmin| 1.59| 1.06| 0.795| 0.398| 0.318| 0.199| 0.159
Z/Rmax| 100000
Z/Rmin| 100
Where, the unit of Cmax is uF; The unit of Cmin is nF. Lmax is in H; The unit
of Lmin is mH.
Zmax/Rmax ,Zmin/Rmin unit is Ω;
Measurement speed factor ks:
speed | slow | middle | fast |
---|---|---|---|
ks | 0 | 1 | 3 |
Measurement Frequency factor kf:
freq(Hz) | 100-1K | 2K-10K | 15K | 20K | 40K-100K | |
---|---|---|---|---|---|---|
kf | 0 | 0.5 | 1 | 4 |
Measurement level error factor kv:
level(mV) | 100 | 300 | 600 | 1000 | 1500 | 2000 |
---|---|---|---|---|---|---|
kv | 9 | 3 | 1 | 0 | 1 | 2 |
Accuracy diagram
DCR measurement accuracy
Range | Display range | Accuracy Re |
---|---|---|
100MΩ | 20.00MΩ-99.99MΩ | 10.0%+20 count |
20MΩ | 10.00MΩ-20.00MΩ | 5.0%+10 count |
10MΩ | 4.000MΩ-9.999MΩ | 2.0%+5 count |
4MΩ | 400.0kΩ-3.9999MΩ | 1.2%+3 count |
400kΩ | 40.00kΩ-399.99kΩ | 0.3%+3 count |
40kΩ | 4.000kΩ-39.999kΩ | 0.2%+2 count |
4kΩ | 400.0Ω-3.9999kΩ | 0.1%+2 count |
400Ω | 40.00Ω-399.99Ω | 0.2%+2 count |
40Ω | 4.000Ω-39.999Ω | 0.3%+3 count |
4Ω | 0.400Ω-3.999Ω | 1.0%+5 count |
0.4Ω | 0.000Ω-0.399Ω | 3.0%+10 count |
Excitation source
-
frequency
10Hz-100kHz, ±0.02% of reading -
level
100mVrms -2000mVrms ,±10% of reading -
Bias
0mV-1500mV, ±(5% of reading+20mV) -
Output impedance
100Ω/30Ω ±1% of reading
External interface instructions
The USB interface
USB communication interface to achieve online communication with the computer.
(reference SCPI part of communication protocol).
RS232 interface
RS232C interface introduction
Using asynchronous serial communication, the serial interface of the
instrument is not strictly based on RS-232 standard, but only provides a
minimal simplified subset. As shown in table 9-1, sending and receiving data
is relative to the machine.
Table 9- 1
Single | Abbreviations | Pin |
---|---|---|
Send | TXD | 3 |
Receive | RXD | 2 |
Ground | GND | 5 |
RS232C connector use black DB9 hole standard socket, pin sequence is shown in figure 9-1:
Note :
- To avoid electrical shock, switch off the power when you plug and unplug the connector;
- Do not arbitrarily short connect the output terminal, or short connect with the housing, so as to avoid damage to the instrument.
Handler interface
The desktop bridge provides the user with a Handler interface that supports
the output of the comparator sorting results and the output of the list scan
results.
Technical description
- Output signal: low effective, open collector output, Opto-couplers isolation.
- Input signal: Opto-couplers isolation
- DC power supply voltage supported by power pin: 3V-25V
Pin definition
The Handler interface uses blue DB9 hole standard socket, specific pin definitions and locations, as shown in figure 9-2 and table 9-2. The following table “/” separates the output signals of the comparator and the list scan. If the comparator and the list scan have the same signal name, there is only one signal name.
Table 9- 2
pin | name | attribute | Describe |
---|---|---|---|
1 | nBIN1 | Output | Comparator function, result output, BIN:1; |
List scan function, result output of sequence number 1. if the comparison result is H or L, then output effective signal.
2| nBIN3| Output| Comparator function, result output, BIN:3;
List scan function, result output of sequence number 3. if the comparison result is H or L, then output effective signal.
3| nAUX_BIN/ nBIN7| Output| Comparator function, result output, BIN :AUX;
List scan function, result output of sequence number 7. if the comparison result is H or L, then output effective signal.
4| nEXT_TRIG| Input| External trigger signal, rising edge trigger. The high level voltage
should be consistent with the external DC power source.
5| EXTCOM1| Ground| Reference ground for external DC power supply and output
signal.
6| nBIN2| Output| Comparator function, result output, BIN:2;
List scan function, result output of sequence number 2. if the comparison result is H or L, then output effective signal.
7| nOutOfBINs| Output| Comparator function, result output, BIN:OUT;
List scan function, Overall comparison result, After startup, as long as one result of all sequence numbers is H or L, output effective signal.
8| nEOM| Output| End of Measurement signal.
9| EXTDC1| Power| External DC power source. Reference ground is EXTCOM1
Sequence diagram
Table 9- 3
Time | The minimum value | The largest number |
---|---|---|
T1 | 1ms | —- |
T2 | —- | Time of measurement |
T3 | Time of measurement | Time of measurement +1ms |
T4 | 0 | —- |
The value of one measurement time in the table is judged by the speed, for example, non-DCR mode, slow speed, and one measurement time is 0.5s.
SCPI Command reference
A little. (If necessary, please contact the manufacturer to ask for it.)
Precautions and warranty
-
Packaging
The measuring instrument should be packed with plastic bags together with accessories, spare parts, instruction manual and product qualification certificate in the strong packing case which is dustproof, shockproof and moisture-proof. -
Transportation
The measuring instrument should be handled carefully in the transportation process, and be protected against moisture and rain. -
The storage
The measuring instrument should be stored in the ventilation room with ambient temperature of 5℃-40℃ and relative humidity of 15% to 85%RH. The air should not contain harmful impurities that corrode the measuring instrument. -
The warranty
This instrument maintenance must be professional and technical personnel, otherwise the instrument damage or precision loss does not belong to the scope of warranty. Please do not replace the internal components of the instrument during maintenance. The instrument must be re-measured and calibrated after maintenance.