stryker EasyFuse Dynamic Compression System Instruction Manual
- June 8, 2024
- stryker
Table of Contents
stryker EasyFuse Dynamic Compression System Instruction Manual
Introduction
Easy Fuse dynamic compression system is an internal fixation system intended for fractures, osteotomies, and jointarthrodesis of the mid foot and hind foot. The system is provided as a single-use sterile pack comprising of a bone staple implant and select instruments for implantation. Multiple implants sizes are available consisting of two-leg and four-leg variants with features such as low-profile and wide bridges, as well as multiple leg lengths. Additional instruments are provided in separate, single-use sterile packs. The staple is composed of nickel-titanium (nitinol) alloy per ASTM F2063. EasyFuse implants are provided pre-loaded on a single-use disposable cartridge. Thiscartridge is then attached to an inserter creating the assembly used to implant the staple. The implant is designed to provide sustained compression to facilitate bony fusion.
Proper surgical procedures and techniques are the responsibility of the medical professional. The following guidelines are furnished for information purposes only. Each surgeon must evaluate the appropriateness of the procedures based on his or her personal medical training and experience. Prior to use of the system, the surgeon should refer to the product package insert for complete warnings, precautions, indications, contraindications and adverse effects. Package inserts are also available by contacting the manufacturer. Contact information can be found on the back of this surgical technique and the package insert is available on the website listed.
Acknowledgments:
Surgeon design team – The EasyFuse dynamic compression system was developed in
conjunction with: John R. Clements, DPM (Roanoke, VA), Kent Ellington, MD
(Charlotte, NC), Carroll Jones, MD (Charlotte, NC), John S. Lewis, Jr., MD
(Louisville, KY)
Indications and contraindications
Indications
The Easy Fuse dynamic compression system is intended to be used for fracture fixation, osteotomy fixation, and joint arthrodesis of the foot and ankle
Contraindications
There are no product specific contraindications.
Operative technique
Mid/Hind foot surgical technique
Prepare fusion site
Create the osteotomy and/or prepare the fusion site needed to implant Easy Fuse.
Size
Place the Universal Drill Guide perpendicular across the fixation site to determine the appropriate implant size. Rotate the knob clockwise and select the preferred staple size. Notice the distance between drill holes will change as the knob is rotated to each size.
Universal Drill Guide
Drill
Use the Drill to create a pilot hole in the bone. Use the laser markings on the Drill to measure the drill depth. Prior to drilling any additional holes, place a corresponding
Locator Pin
Drill
Prepare inserter
Place the Universal Implant Inserter into its unlocked position by lifting the lever up. Assemble the selected implant Cartridge onto the Universal Inserter by aligning the tabs on the implant Cartridge with the grooves of the Universal Inserter and rotating clockwise until locked. Proceed to press the lever of the Universal Inserter down to its locked position to outwardly displace the legs of the EasyFuse implant.
Universal Inserter
Implant Cartidge
Insert implant
Remove the Locating Pins and Drill Guide before inserting the implant. Position the legs of the EasyFuse over the pilot holes and advance the implant into the holes by hand until fully seated.
Remove inserter
Unlock the Universal Implant Inserter from the implant by moving the inserter lever to its unlocked position. Slide outward or twist counter-clockwise the Universal Inserter to disengage cartridge from implant.
Final seat and fluoro check
If necessary, place the implant Cartridge on the EasyFuse bridge and lightly tap with a mallet on the back of the Inserter until the implant is flush to the bone. Check the final position of the EasyFuse implant under fluoroscopy.
Additional implants
Repeat steps 2 through 7 for each additional Easy Fuse implant used. Tip: If placing 2 EasyFuse implants in any orientation other than parallel to one another, stagger the implant placement so the legs do not obstruct one another inside the bone.
4-Leg surgical technique
Prepare fusion site
Create the osteotomy and/or prepare the fusion site needed to implant Easy Fuse.
Size
Place the 4-leg Sizer perpendicular across the fixation site to determine the appropriate implant size. Adjust the distance across the Universal Drill Guide to the selected size by rotating the knob clockwise. Attach the 4-leg Clip to the Universal Drill Guide.
4-Leg Sizer
4-Leg Clip
Drill
Place the Universal Drill Guide across the fixation site. Use the Drill to
create a pilot hole in the bone. Use the laser markings on the Drill to
measure the drill depth.
Prior to drilling any additional holes, place a corresponding Locator Pin in
the first hole through the drill guide. Prepare the outer most holes first
prior to preparing the inner holes.
Prepare inserter
Place the Universal Implant Inserter into its unlocked position by lifting the lever up. Assemble the selected implant Cartridge onto the Universal Inserter by aligning the tabs on the implant Cartridge with the grooves of the Universal Inserter and rotating clockwise until locked. Proceed to press the lever of the Universal Inserter down to its locked position to outwardly displace the legs of the Easy Fuse implant
Insert implant
Remove the Locator Pins and Drill Guide before inserting the implant. Position the legs of the EasyFuse implant over the pilot holes and advance the implant into the holes by hand until fully seated.
Remove inserter
Unlock the Universal Implant Inserter from the implant by moving the inserter lever to its unlocked position. Slide outward or twist counter-clockwise the Universal Inserter to disengage cartridge from implant.
Final seat and fluoro check
If necessary, place the implant Cartridge on the Easy Fuse bridge and lightly tap with a mallet on the back of the Inserter until the implant is flush to the bone. Check the final position of Easy Fuse implant under fluoros copy
Additional implants
Repeat steps 2 through 7 for each additional Easy Fuse implant used. Tip: If placing 2 EasyFuse implants in any orientation other than parallel to one another, stagger the implant placement so the legs do not obstruct one another inside the bone
Removal and reinsertion
The Easy Fuse implant can be removed using the Universal Implant Inserter and
the appropriate implant Cartridge. Assemble the implant Cartridge onto the
Universal Inserter. Ensure that the Universal Inserter lever is in its
unlocked position.
To remove an Easy Fuse implant, use a flat-sided instrument, like an
osteotome, to wedge the bridge of the implant slightly off the bone. Place the
Cartridge Tip underneath the implant bridge and lock onto the implant by
moving the Universal Inserter lever to its locked position. Pull up on the
inserter to remove the implant from the bone. If required, the Easy Fuse
implant can be repositioned and inserted again following step 5 in the
surgical technique.
Explant information
If the removal of the implant is required due to revision or failure of the device, the surgeon should contact the manufacturer using the contact information located on the back cover of this surgical technique to receive instructions for returning the explanted device to the manufacturer for investigation.
Postoperative management
Postoperative care is the responsibility of the treating physician.
Procedural sizing chart
The diagram and chart below highlight some of the suggested procedures and recommended sizing.
Index | Procedure | Implant sizing |
---|---|---|
1 | Hallux IP Fusion | 15×12 |
2 | MTPJ Fusion | 18×15, 20×15, MTP |
3 | Lapidus Fusion | 15×15, 18×20, 18×25, 20×25 4 Leg |
4 | Naviculocuneiform Fusion | 18×15, 18×20, 20×15, 20×20 |
5 | Talonavicular Fusion | 18×20, 18×25, 20×20, 20×25 |
6 | Calcaneocuboid Fusion | 18×25, 20×20, 20×25, 4 Leg |
7 | TMT Fusion | 15×15, 15×20, 18×15, 18×20,20×15, 20×20 |
8 | Chevron Osteotomy | 15×15, 15×20,18×15, 18×20 |
9 | Metatarsal Osteotomy | 15×15, 15×20, 18×15, 18×20,20×15 |
10 | Proximal Base Osteotomy | 15×15, 15×20, 18×15, 20×15 |
11 | Cotton Osteotomy | 18×15, 18×20, 20×15, 20×20 |
12 | Evans Osteotomy | 20×20, 20×25, 25×20, 25×25 |
13 | Calcaneal Osteotomy | 20×20, 20×25, 25×20, 25×25 |
14 | Subtalar Fusion | 20×20, 20×25, 25×20, 25×25 |
15 | Jones Fracture | 15×12, 18×15 |
Ordering information
2-Leg Implant Part Numbers
Part number | Description |
---|---|
FFS21512 | EasyFuse Implant Procedure Pack, 15×12 |
FFS21515 | EasyFuse Implant Procedure Pack, 15×15 |
FFS21520 | EasyFuse Implant Procedure Pack, 15×20 |
FFS21815 | EasyFuse Implant Procedure Pack, 18×15 |
FFS21820 | EasyFuse Implant Procedure Pack, 18×20 |
FFS21825 | EasyFuse Implant Procedure Pack, 18×25 |
FFS22015 | EasyFuse Implant Procedure Pack, 20×15 |
FFS22020 | EasyFuse Implant Procedure Pack, 20×20 |
FFS22025 | EasyFuse Implant Procedure Pack, 20×25 |
FFS22520 | EasyFuse Implant Procedure Pack, 25×20 |
FFS22525 | EasyFuse Implant Procedure Pack, 25×25 |
FFSP1530 | EasyFuse Instrument Procedure Pack |
4-Leg Implant Part Numbers
Part number | Description |
---|---|
FFS4MTPS | EasyFuse Implant Procedure Pack, MTP, Small |
FFS4MTPL | EasyFuse Implant Procedure Pack, MTP, Large |
FFS42520 | EasyFuse Implant Procedure Pack, 4-Leg, 25×20 |
FFS43020 | EasyFuse Implant Procedure Pack, 4-Leg, 30×20 |
FFSP1530 | EasyFuse Instrument Procedure Pack |
Coustamer Support
Foot & Ankle
This document is intended solely for the use of healthcare professionals. A
surgeon must always rely on his or her own professional clinical judgment when
deciding whether to use a particular product when treating a particular
patient. Stryker does not dispense medical advice and recommends that surgeons
be trained in the use of any particular product before using it in surgery.
The information presented is intended to demonstrate a Stryker product. A
surgeon must always refer to the package insert, product label and/or
instructions for use, including the instructions for cleaning and
sterilization (if applicable), before using any Stryker product. Products may
not be available in all markets because product availability is subject to the
regulatory and/or medical practices in individual markets. Please contact your
Stryker representative if you have questions about the availability of Stryker
products in your area.
Stryker Corporation or its divisions or other corporate affiliated entities
own, use or have applied for the following trademarks or service marks:
EasyFuse, Stryker. All other trademarks are trademarks of their respective
owners or holders. AP-015450, 09-2021 Copyright © 2021 Stryker
Manufacturer
Stryker Corporation 1023 Cherry Road Memphis, TN 38117 800 238 7117 901 867 9971 www.wright.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>