UNIPad FT-810 THERMOTENS for WAIST Instruction Manual
- June 3, 2024
- UNIPad
Table of Contents
UNIPad FT-810 THERMOTENS for WAIST
Revision record
Date| Doc. No.| Version| Contents of
revision
| Reason
2021/11/12| IFUT-FT-810-002| 0.1| | New release
2021/11/30| IFUT-FT-810-002| 0.2| Remote controller Function| Revised the
function
2021/11/30| IFUT-FT-810-002| 0.21| Shelf Life and manufacturer
website
| Revised information
2021/12/8
| ****
IFUT-FT-810-002
| ****
0.3
| Modify model Name, NCC warring| Revised model name
and warring for NCC requirement
2021/01/03
| ****
IFUT-FT-810-002
| ****
0.4
| ****
-Remove Remote controller
-Modify Warring for wireless
-Rename Model name,
-Modify manufacture address
| -Update the latest
configuration
-Revised warring for wireless requirement
-Rename Model name
– Modify manufacture address
2021/01/03| IFUT-FT-810-002| 0.41| -Modify manufacture address
R.O.C
| – Modify manufacture
Address R.O.C
2021/01/04| IFUT-FT-810-002| 0.42| -Update manufacture address
(ZIP Code)
| Same as left side
2022/01/06| IFUT-FT-810-002| 0.43| -Update bottom view for the
latest manufacture address
| Same of left side
2022/03/01| IFUT-FT-810-002| 0.44| -Add FCC warning (Annex 4)| Same of left
side
Read this instruction manual carefully and keep it for later use; be sure to
make it accessible to other users and observe the information it contain.
Signs and symbols
The following symbols appear in this instruction manual and on the device
Device description
The FT-810 is a self-adhesive TENS device with 15 adjustable intensity level
for pain relief. Moreover, it also provides a heat function which can be used
in conjunction with the TENS function. TENS, Transcutaneous Electrical Nerve
Stimulation, refers to the electrical stimulation of nerves through the skin
which is an effective non pharmacological method of pain relief. It has no
side-effects if administered correctly, and can be used for self-treatment.
Any symptoms that could be relieved using TENS must be checked by your general
practitioner who will also give you instruction on how to carry out a TENS
self-treatment regime. TENS device works by passing electrical currents over
the skin via a set of gel pads. As a transfer medium, the gel pads are subject
to natural wear and tear and must be replaced when they stop providing
sufficient contact or the main device no longer sticks to the skin completely.
Failure to replace the gel pad may lead to skin irritation as a result of
heightened current density in particular areas.
Intended use
This device is designed to be used for temporary relief of pain associated
with sore and aching muscles in the shoulder, waist, back, upper extremities
(arm) and lower extremities (leg) due to strain from exercise or normal
household work activities. In addition, it also provides a heat function
intended to temporarily relief of minor aches and pains.
Included in delivery
- 1 x FT-810 main unit
- 2 x self-adhesive gel pads (61 mm x 89 mm)
- 1 x USB-C charging cable (for charging only)
- 1 x Adaptor
- 1 x Storage board
- This instruction manual
WARNINGS
- To avoid damage to health, do not use this device in the following circumstances:
- If you have implanted electrical devices (e.g. pacemaker).
- If you have metal implants such as copper coils.
- If you use an insulin pump.
- If you have a high fever (e.g. >39℃)
- If you have a known or acute cardiac arrhythmia or disorders of the heart’s impulse and conduction system.
- If you suffer from a seizure disorder (e.g. epilepsy).
- If you are pregnant.
- If you have cancer.
- After an operation, if strong muscle contractions could affect the healing process.
- Near the heart. This device must not be placed on any part of the front ribcage (where the ribs and breastbone are located), and especially not on the two large pectorals, as this can increase the risk of ventricular fibrillation and include cardiac arrest.
- On the skeletal skull structure, or around the mouth, throat or larynx.
- In the neck/carotid artery area.
- In the genital area.
- On acutely or chronically diseased (injured or irritated) skin (e.g. inflamed skin – whether painful or not, reddened skin, rashes, e.g. allergies, burns, bruises, swellings, both open and healing wounds, and post-operative scars where the healing process could be affected).
- In humid environments (e.g. in the bathroom) or when bathing or showering.
- After consuming alcohol.
- If you are connected to a high-frequency surgical device.
- In cases of acute or chronic disease of the gastrointestinal tract.
- Before using this device, consult your doctor if any of the following apply to you:
- Serious illness, in particular if you suspect or have been diagnosed with high blood pressure, a blood coagulation disorder, propensity to thrombo-embolic conditions or in the case of malignant neoplasms.
- Skin diseases or open wounds.
- Unexplained chronic pain in any part of the body.
- Diabetes.
- Any sensory impairment that reduces the feeling of pain (e.g. metabolic disorders).
- If you are receiving medical treatment.
- In the event of complaints linked to the stimulation treatment.
- If you suffer from persistently irritated skin due to long-term stimulation at the same application site.
- Keep this device away from children to prevent potential hazards.
- If the device does not work properly, or you feel unwell or experience pain, stop using it immediately.
- Switch off the device before attaching or removing it.
- Do not modify this device in any way (e.g. by cutting it), as this will produce a higher current density and may be dangerous.
- During the initial few minutes, use the device while sitting or lying down, to minimize the risk of accidental injuries arising from vagal responses (feeling of faintness) that have been infrequently reported as a side effect. It you feel faint, switch off the device immediately, lie down and support the legs in an elevated position (approx. 5-10 minutes).
- Do not use this device whilst asleep, driving a vehicle or operating machinery.
- Do not use this device whilst undertaking any activity where an unexpected reaction (e.g. strong muscle contractions even at low intensity) could be dangerous.
- Do not use this device whilst using other devices that transmit electrical impulses to your body.
- The actual temperature of the device may vary depending on the condition of your skin, your age, the location of the pain, etc.
- If the heat function feels too hot, stop treatment immediately. You can continue the TENS treatment without the heat function.
- Wait at least one hour after eating before starting treatment with this device.
- Do not use this device on wet skin as this impairs the adhesion of the gel pads.
- Ensure that no metallic objects (e.g. belt buckles or necklaces) come into contact with the electrodes during stimulation. if you are wearing jewelry or have piercings in the area to be treated (e.g. a navel piercing), these must be removed before using the device; failure to do this could result in spot burns.
- Pre-treating the skin with greasy creams or lotions is not recommended.
- Do not use adhesive tapes, bandages or any medium other than the gel pads provided to adhere the device to your skin, as the uneven adhesion of the gel pads could lead to lesions.
- Not on handling battery:
- If a battery has leaked, put on protective glove to handle the device.
- If your skin or eyes come in to contact with fluid from a battery cell, flush out the affected area with water and seek medical assistance.
- Protect the device from excessive heat.
- Risk of explosion! Never throw this device into a fire.
- Do not disassemble, split or crush this device.
- Only use chargers specified in this Instruction Manual.
- Battery must be charged corrected prior to use. The instructions from the manufacturer and the specifications in this Instruction Manual regarding correct charging must be observed at all times.
- Fully charge the device prior to initial use.
- In order to achieve as long a battery service life as possible, fully charge the battery at least every 6 months.
ADVERSE REACTIONS
- Patients may experience skin irritation and burns beneath the stimulation electrodes applied to the skin.
- Patients may experience headache and other painful sensations during or following use.
- Patients should stop using the device and should consult with their physicians if they experience adverse reactions from the device.
IMPORTANT
- This device is not a substitute for medical consultation and treatment. Consult your doctor first if you are experiencing any pain or suffering from an illness.
- Only use this device:
- On people
- For the intended purpose and as specified in this instruction manual. Improper use can be dangerous.
- For external use
- With the original accessories supplied which can be re-ordered.
- Never direct contact the electrode area or apply the electrode to any part of the body. This product must be used in conjunction with a set of self-adhesive electrode gel pad. Please follow the steps in Chapter 8 to install the gel pads to the electrode area before use.
- Always check the sight of stimulation when using TENS function with the heat function simultaneously to avoid being insensitive to pain result from overheating.
- This product is intended for single user only.
- Protect the device from strong impact.
- Do not use this device if it shows signs of damage or does not function properly. In such cases, contact customer services.
- Do not expose this device to direct sunlight or high temperatures.
- Protect this device from dust, dirt and humidity. Never immerse the device in water or other liquids.
- Do not use the device by or in the vicinity of other heating devices, heating elements such as dryers or ovens, and do not use it in the vicinity (~1 m) of shortwave or microwave devices (e.g. mobile phones), as doing so can result in unpleasant current peaks.
- If the device is damaged or you have doubts, do not use it and contact your retailer or customer services at the address provided.
- Switch the device off immediately if it is faulty or not working properly.
- Never attempt to open and/or repair the device yourself.
- Repairs may only be carried out by customer services or authorized retailers.
Device description
FT-810 Top View
FT-810 Bottom View
A) Increase intensity button
B) Low heat function indicator
C) High heat function indicator
D) Heat function button
E) TENS function button
F) TENS function indicator /Battery level indicator
G) Decrease intensity button
H) USB Type C Port
I) Heating surface
J) Left/Right electrode
Initial use
Charging the device
Note
Before using this device for the first time, let it charge for a minimum of 2
hours. How to do this:
-
Connect the USB Type C charging cable provided to the adapter and the device.
-
Then insert the mains adapter into a suitable socket.
-
Alternatively, you can charge this device through your computer/laptop. To do so, connect the device to a USB port on your computer/laptop using the USB charging cable.
-
Whilst charging, the battery level indicator flashes green. As soon as the device has been completely charged, the battery level indicator will be continuously illuminated in green.
Note- You cannot use this device while it is being charged.
- Use this product only with the genuine charging cable and adaptor which is come with the product.
Attaching the gel pads
- Remove one of the protective films with care.
- Attach the gel pads to the electrode area and ensure that the edge of the gel pad does not protrude outside the electrode profile. Only attach one gel pad to each electrode.
- Remove the second protective film slowly and carefully to make sure the gel pads are not damaged.
Note: Use this product only with the genuine gel pads (61 mm x 89 mm, single size) which are come with the product and can be purchased separately from the manufacturer afterward.
Ini tial use
Starting use
Note
In total, this device has two levels `of heat function, and 15 intensity
levels for TENS application. We recommended you start off using a low
intensity level until you have got used to the device. This device has an
automatic switch-off function. After 20 minutes of use, it automatically
switches itself off.
- Stick the device onto the painful area.
- Long press the “TENS function button” to switch “on the device”, the “TENS function indicator/Battery level indicator” flashes green. To start stimulation, short press the “increase intensity button” on the device once. The lowest intensity is now activated.
- To switch on superficial heat, long press the “Heat function button” on the device/remote controller. The “Heat function indicator” and “TENS function indicator/Battery level indicator” flash green. To start stimulation, short press the “Increase intensity button” on the device once. The lowest intensity is now activated. To adjust the heat function level, short press the “Heat function button” on the device or remote controller.
- If you want to switch off superficial heat, long press the “Heat function button” on the device. Now only the “TENS function indicator/Battery level indicator” is flashing and you can start stimulation by short pressing the “Increase intensity button” on the device/remote controller.
- Select your desired intensity level using the “Increase intensity button” or “decrease intensity button” on the device. We recommend you set the intensity to a non-painful, maximum tolerable level. The intensity should be increased continually throughout the treatment session to maintain an intense sensation and avoid habituation.
- After the device output for 20 minutes, it will automatically switch off. However, you can switch off the device, long press the “TENS function button” and “Heat function button” on the device. The “TENS function indicator” and “Heat function indicator” go out.
- Take care when removing it from your body. Reapply the storage board to the gel pads to protect them.
- If the “TENS function indicator/Battery level indicator” flashes red, please charge the device as described in the previous section.
Note on use
- Switch off the device before removing it.
- Wait at least 30 minutes between applications.
- Only use the device as often as you find comfortable. It treatment with this device starts to feel unpleasant, stop using it.
- This device has a function which detects whether the device is in contact with the body properly, if there is not sufficient contact with the body, the device beeps and switches itself off. If the device is attached to your body but still beeps, clean the gel pads and try starting the device again. If the continuous beeping persists, replace the gel pads.
- If the adhesive capability of the gel pads decreases, replace them immediately. Do not use the device again until the gel pads have been replaced, as the uneven adhesion of the gel pads could lead to skin lesions.
- In the event of unpleasant current peaks, switch off the device and move it slightly
- If you find that using the device cause you to have a much heavier flow, stop using it.
Superficial heat (SH)
In addition to the TENS function, this device also offers two levels of the
superficial heat (SH), delivered through the gel pads which can be activated
as required for all programs. You can activate the first level of the SH by
short pressing the Heat function button on the device whenever the electrical
stimulation is outputting. Then wait a moment until the temperature stops
increasing. If the temperature is too low for you, you can activate the high
second level of the function by short pressing the” HEAT function button” on
the device. If you would like to deactivate the heat function, you can do so
by long pressing the “HEAT function button” on the device.
If you want to use the SH function independently, without additional electrical stimulation, proceed as follows:
- Stick the device onto the painful area.
- Long press the “Heat function button” to switch “on the device, the “Heat function indicator” flashes green.
- If you want to adjust the heat function level, press the “Heat function button” on the device.
- After the device output for 10 minutes, it will automatically switch off. However, you can switch off the heat function at any time, long press the “Heat function button” on the device. The “Heat function indicator” go out.
- Take care when removing it from your body. Reapply the storage board to the gel pads to protect them.
Cleaning and maintenance
- Clean the device after use with a soft, slightly damp cloth. If it is very dirty, you can also moisten the cloth with a mild soapy solution.
- Ensure that no water enters the device. If this happens, only use the device again once it has fully dried out.
- To ensure that the gel pads retain their adhesion for as long as possible, clean them carefullyth a damp, lint-free cloth.
- Do not use any chemical or abrasive cleaning agents for cleaning.
Troubleshooting
Problem | Possible cause | Solution |
---|
The device does not switch on when the “TENS function button” or “Heat
function button” is pressed.
| The battery may be flat.| Fully charge the battery. If it is still not possible to switch the device on, please contact customer service.
The device does not stick to the body.
| ****
The gel pad may be dirty or worn. The treatment area of the skin may not be clean enough.
| Clean the adhesive surface of the gel pads using a damp, lint-free cloth. The gel pads must be replaced if they still
do not stick securely.
Clean the skin prior to any applications; do not use skincare lotions or oils prior to treatment. Shaving may increase the life of gel
pads.
There is no noticeable simulation.
| ****
The gel pads may be dirty or worn, the battery may need charging.
| Ensure that the gel pads are in firm
contact with the treatment area.
The battery is almost flat. Charge it
fully.
An unpleasant sensation is experienced where the gel pads are applied.| The gel pads are worn. This may cause irritated skin, as even distribution of the current across the entire area is no longer
guaranteed.
| Replace the gel pads.
Skin in the treatment area turns red.| If redness under the gel pads
disappears quickly this is not dangerous, and is caused by the locally
stimulated, increased blood flow caused by TENS and heat.| Immediate stop
treatment and wait until your skin has returned to its normal condition.
However, consult your doctor before you continue treatment if the skin
irritation persists and if it is accompanied by an itch sensation or
inflammation, as this may
be caused by an allergic reaction to
The skin under the heat surface turns red.| Reddening under the heat surface
of the device can be explained by the heavily stimulated blood flow in the
heated tissue.| The skin should return to its normal condition a few minutes
after treatment. However, consult your doctor before continuing to use the
heat function if the redness persists and if you experience any itching or
inflammation.
---|---|---
The device becomes too hot.| —| Press the “Heat function button” to stop the
heat function and just continue using the TENS function.
Disposal
For environment reasons, do not dispose of the device in household waste at
the end of its service life. Dispose of the device at appropriate collection
points in your country. If you have any question, please contact the local
authorities responsible for waste disposal.
Technical specifications
Name | HEATING TENS/EMS |
---|---|
Model | FT-810 |
Output waveform | Biphasic rectangular pulses |
Pulse length | 400 μs |
Pulse frequency | 4~45 Hz |
Maximum output voltage | 63 Vp-p (at 1kΩ) |
Output current | 63 mAp-p (at 1kΩ) |
Heat settings | HI: Up to 43˚C |
LOW: Up to 41˚C
Voltage supply| Lithium polymer battery 1000 mAh
Treatment time| TENS-only mode: 20 minutes, and then switch off automatically
Heat-only mode: 10 minutes, and then switch off automatically TENS + Heat mode: 20 minutes, and then switch off automatically
---|---
Intensity| Adjustable from 0 to 15
Operating conditions| 5˚C to 40°C (41°F to 104°F)
at a relative humidity of 15% to 90%
Storage conditions| 0°C to 40°C (32°F to 104°F)
at a relative humidity of 0% to 90%
Dimension| Approx. 284 mm × 95 mm × 20 mm
Weight| 102 g
Shelf Life / Product Life| 3 years
Annex 1 – Electromagnetic Compatibility
Important information regarding Electro Magnetic Compatibility (EMC) With the
increased number of electronic devices such as computers and mobile (cellular)
telephones, this device in use may be susceptible to electromagnetic
interference from other devices, even if it complies with CISPR emission
requirements. Electromagnetic interference may result in incorrect operation
of this device and create a potentially unsafe situation. Besides, this device
should also not interfere with other devices. In order to regulate the
requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent
unsafe product situations, the IEC
60601-1-2 standard has been implemented. This standard defines the levels of
immunity to electromagnetic interferences as well as maximum levels of
electromagnetic emissions for medical devices. This device manufactured by
HIVOX conforms to this IEC 60601-1-2 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
- The use of the device adjacent to or stacked with other device should be avoided because dot could result in improper operation. In case such use is necessary, the device and other device should be observed to verify that they dare operating normally?
- The use of any parts other than those specified or provided by the HIVOX could result in increased electromagnetic emissions or decreased electromagnetic immunity of this device and result in improper operation.
- Portable RF communications device (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of this device. Otherwise, degradation of the performance of the device could result.
- Refer to further guidance below regarding the EMC environment in which the device should be used.
Guidance and manufacturer’s declaration – Electromagnetic emission
This device is intended for use in the electromagnetic environment specified
below. The user of this device should make sure it is used in such an
environment.
Test Standard| Test Item| Test Level| Electromagnetic
environment – Guidance
---|---|---|---
EN 55011 /
CISPR 11
|
Conducted Emissions of Powerline
|
Group 1
| The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
EN 55011 /
CISPR 11
|
Radiated Emissions
|
Class B
| This device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
EN/IEC 61000-3-2| Harmonic Current Emission| Not applicable (Note 1)|
—
---|---|---|---
EN/IEC 61000-3-3| Voltage Fluctuations and Flicker| Not applicable (Note 1)|
—
Note 1: The power supply of this device is from battery (DC voltage), so Harmonic current emission, Voltage fluctuations & flicker, EFT, Surge, CS and DIP tests are not applicable for this device.
Note 2: Standards applicable to other modes or EM ENVIRONMENTS of transportation for which use is intended shall apply.
Guidance and manufacturer’s declaration – Electromagnetic immunity
This device is intended for use in the electromagnetic environment specified
below. The user of this device should make sure it is used in such an
environment.
Test Standard| Test Item| Test Level| Electromagnetic
environment – Guidance
---|---|---|---
IEC 61000-4-2
|
Electrostatic discharge
|
Contact:
± 8 kV;
Air:±2 kV, ±4 kV,
±8 kV, and ±15 kV
| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Portable and mobile RF communications equipment should
be used no closer than 30 cm (12 inch) to any part of this device.
IEC 61000-4-3
|
Radiated RF EM fields
| 10 V/m;
80 MHz to 2.7 GHz;
80% AM at 1
kHz
| Portable and mobile RF communications equipment should be used no closer than 30 cm (12 inch) to any part of this device.
IEC 61000-4-3
| Proximity fields from RF wireless
communications equipment
|
Not applicable (Note 1)
|
—
IEC 61000-4-4
| Electrical fast
transients / bursts
| Not applicable (Note 1)|
—
IEC 61000-4-5| Surges line-to-line| Not applicable (Note 1)| —
IEC 61000-4-6
| Conducted disturbances induced by RF fields|
Not applicable (Note 1)
|
—
IEC 61000-4-8
|
Rated power frequency magnetic fields
|
30 A/m
50 Hz and 60 Hz
| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital
environment.
IEC 61000-4-11| Voltage dips and
interruptions
| Not applicable (Note 1)| —
ISO 7637-2
| Electrical Transient Conduction along Supply Lines|
Not applicable (Note 2)
|
—
---|---|---|---
Note 3: The power supply of this device is from battery (DC voltage), so
Harmonic current emission, Voltage fluctuations & flicker, EFT, Surge, CS and
DIP tests are not applicable for this device.
Note 4: Standards applicable to other modes or EM ENVIRONMENTS of transportation for which use is intended shall apply.
Recommended separation distance between portable and mobile RF
communications equipment and this device
This device is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of this
device can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and this device as recommended below, according to the maximum
output power of the communications equipment.
Test freque ncy
(MHz)
|
Band (MHz)
|
Service
|
Modulation
| Maximum power (W)| Immunity test level (V/m)|
Separation distance (m)|
---|---|---|---|---|---|---|---
385
|
380 – 390
|
TETRA 400
| Pulse modulation
b)
18 Hz
|
1.8
|
27
|
0.3
|
450
|
430 – 470
|
GSMR 460;
FRS 460
| FM c)
±5 kHz
deviation 1 kHz sine
|
2
|
28
|
0.3
|
710|
704 – 787
|
LTE Band 13,
17
| Pulse|
0.2
|
9
|
0.3
|
modulation
b)
|
745|
780|
217 Hz
|
810|
800 – 960
| GSM 800/900;|
Pulse
|
2
|
28
|
0.3
|
870|
TETRA 800;
iDEN 820;
| modulation
b)
|
930
|
CDMA 850;
LTE Band 5
| 18 Hz|
1720|
1700 – 1990
| GSM 1800;
CDMA 1900;
|
Pulse
|
2
|
28
|
0.3
|
1845|
GSM 1900; DECT; LTE
|
modulation
b)
|
1970|
Band 1, 3, 4,
25; UMTS
|
217 Hz
|
| | RFID 2450;
LTE Band 7
| 217 Hz| | | |
---|---|---|---|---|---|---|---
5240|
5100 – 5800
|
WLAN
802.11 a/n
| Pulse|
0.2
|
9
|
0.3
|
modulation
b)
|
5500|
5785|
217 Hz
|
a) For some service the uplink frequency are included.
b) The carrier shall be modulated using a 50% duty cycle square signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
|
NOTICE :
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:
- This device may not cause harmful interference, and
- This device must accept any interference received, including interference that may cause undesired operation.
FCC Warning
Any changes or modifications not expressly approved by the party responsible
for compliance could void your authority to operate the equipment.
FCC Warning
Note : This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
FCC Warning
RF exposure statements
This Transmitter must not be co-located or operating in conjunction with any
other antenna or transmitter.
HIVOX BIOTEK INC.
5F., No. 123, Xhingde Rd.,Sanchong Dist.,New Taipei City 241, Taiwan, R.O.C
https://www.hivox-biotek.com/
Documents / Resources
|
UNIPad FT-810 THERMOTENS for
WAIST
[pdf] Instruction Manual
FT-810, FT810, 2A32IFT-810, 2A32IFT810, FT-810 THERMOTENS for WAIST, FT-810,
THERMOTENS for WAIST, THERMOTENS, FT-810 THERMOTENS
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>