Ensure that you read and understand this instruction manual before using
this product. Failure to follow the instructions and safety precautions in
this manual can result in damage and/or serious injury.
Please keep these instructions in a safe location for future reference
WARNINGS (Risk of possible injury or other serious adverse reactions)
- Do not operate the Airflo 12 compressor in the presence of flammable anaesthetic mixtures with air, oxygen or nitrous oxide, as these mixtures can create a combustible environment leading to the potential for fire or even an explosion in the presence of a spark.
- Do not use the Raiser Lifting Cushion if under the influence of alcohol or if taking medication which may affect your ability to exercise judgment, coordination or motor control.
- NEVER use the Raiser Lifting Cushion as a step, stand or as a device for lifting items.
- The Raiser Lifting Cushion should only be used for its stated purpose and should not be modified in any way.
- Ensure that when in use, the Airflo 12 compressor, the hand control lead and air hose are laid out in such a way as to avoid any risk of tripping or strangulation.
- Do not allow children to play with the Raiser Lifting Cushion or Airflo 12 as operation of the equipment in an uncontrolled or unsupervised situation may lead to personal injury and/or damage.
- To avoid personal injury during use and/or damage to the equipment, keep animals away from the Raiser Lifting Cushion and Airflo compressor.
- Regularly check that there are no obvious signs of damage. If in doubt, do not use and seek advice from your supplier.
- If The Raiser Lifting Cushion is to be used in a chair which does not have armrests, the user MUST assess their ability to maintain stability whilst the cushion is being raised and lowered.
- To avoid the risk of electric shock, NEVER recharge the Airflo battery in the bathroom, outdoors or in a damp environment.
- Do not use any other air supply to inflate the Raiser Lifting Cushion.
- Ensure that the Raiser Lifting Cushion is protected from sharp objects.
CAUTIONS (Risk of possible problems including device malfunctions, device failure, damage to the device or damage to other property)
- Only recharge the battery with the Mangar Health battery charger supplied with the Airflo 12 compressor.
- Never use hot air to dry the Raiser Lifting Cushion following cleaning as this may affect the integrity of the welded seams and fabric and lead to premature failure of the cushion.
- Disconnect the Airflo 12 compressor from the Raiser Lifting Cushion before transporting so as to avoid accidental inflation of the cushion.
Product Description
Intended Use
The Raiser Lifting Cushion is intended solely for use as a lifting aid to
assist a person to raise themselves from and lower themselves into an
armchair. Users should have sufficient upper body strength to be able to hold
the armrests for added support if required. WARNING: If The Raiser Lifting
Cushion is to be used in a chair which does not have armrests, the user MUST
assess their ability to maintain stability whilst the cushion is being raised
and lowered. The Raiser Lifting Cushion is not intended for any other use and
should never be used as a step or for lifting items. When properly maintained,
the Raiser Lifting Cushion will give many years of trouble free service.
Construction
The Raiser Lifting Cushion is constructed from a very strong polyurethane
coated nylon material which has been precision cut into panels and then joined
together by high frequency welding. This produces a very strong light weight
and air tight construction. The Raiser Lifting Cushion incorporates a quick
exhaust valve which allows the cushion to fully deflate without the need to
keep pressing the down button on the hand control. The Raiser Lifting Cushion
is powered by the Mangar Health Airflo 12 battery powered compressor.
PLEASE READ the separate Airflo 12 compressor instructions before using your Raiser Lifting Cushion.
Standard Product Contents
1 x Raiser Lifting Cushion
1 x User Instructions
Installation and Use
Getting Started
The Airflo 12 does not have an ‘on/off’ switch.
To operate the compressor, press the ‘Up’ button on the Airflo or on the hand
control.
Connecting
- Position the Raiser Lifting Cushion on the chair so that the air hose is at the rear.
- Connect the air hose (A Fig. 1) and the hand control lead (B) to the Airflo 12.
- Place the hand control within easy reach.
Care and Maintenance
Operating – (Figs. 2 & 3)
WARNING: If The Raiser Lifting Cushion is to be used in a chair which
does not have armrests, the user MUST assess their ability to maintain
stability whilst the cushion is being raised and lowered.
- Keep your feet firmly on the floor, slightly apart and as close to the front of the armchair as possible.
- With your hands on the armrests to steady yourself, lean forward and transfer your weight to the front of the seat.
- To rise, press and hold the ‘Up’ button on the hand control. Release the button to stop at any height.
- To lower, press ‘Down’ on the hand control. To stop lowering at any point, momentarily press the ‘Up’ button.
- After leaving the seat, press and hold the ‘Down’ button until the initial release of air has occurred. The Raiser Lifting Cushion will carry on deflating after the ‘Down’ button is released, ready for when the cushion is next used.
NOTE: If the air hose is disconnected, the Raiser Lifting Cushion will deflate.
Cleaning
Routine cleaning
-
Fully inflate the Raiser Lifting Cushion for cleaning.
-
All components can be cleaned with a proprietary non-abrasive mild liquid cleaner or disinfectant. Always follow packaging instructions for using mild detergent products.
NOTE: Before using proprietary cleaners, always test on an inconspicuous area. -
Use a soft cloth when cleaning.
-
Do not wash the product in boiling water.
IMPORTANT: Following cleaning, rinse thoroughly with clean water to remove all soap deposits.
Wipe dry or leave to air dry. Never store the Raiser Lifting Cushion when it is wet or damp. Do not leave in bright sun light. Do not dry using hot air.
Disinfection -
The Raiser Lifting Cushion should be fully inflated and pre-cleaned by washing with detergent and water, thoroughly rinsed and then disinfected using a 1:10 solution of household liquid bleach and water (1 part bleach to 10 parts cold water). Leave the solution on the Raiser Lifting Cushion for approximately 10 minutes before thoroughly rinsing off with clean cold water.
WARNING: Never mix bleach with any other cleaning product as doing so can result in the formation of harmful acids and poisonous gases which can cause serious breathing problems.
Maintenance
Other than routine cleaning and disinfecting, the Raiser Lifting Cushion is
maintenance free. Mangar Health recommends that your Lifting Cushion is
inspected annually by either your supplier or other suitably qualified person.
If required, Mangar Health can provide the above inspection within an annual
service contract (UK only). Please contact our Customer Service Department for
further details on Freephone 0800 2800 485.
Transportation / Storage
Disconnect the Airflo 12 compressor from the Raiser Lifting Cushion before
transporting so as to avoid accidental inflation of the cushion. Ensure that
the Raiser Lifting Cushion is clean and thoroughly dry before storing. Failure
to thoroughly dry the cushion before storage may result in a build-up of mould
on the surface of the product leading to a weakening of the fabric and
consequential invalidation of the product warranty. Store in a dry
environment. To maintain the Airflo battery when being stored for long
periods, put the Airflo on charge for 16 hours every 3 months.
Disposal / Recycling
When the Raiser Lifting Cushion has reached the end of its working life,
dispose of it at your local household waste recycling centre (civic amenity
site).
When the Airflo has reached the end of its working life, remove the battery
for safe disposal in accordance with your local council guidelines, and
dispose of the Airflo at your local household waste recycling centre (civic
amenity site).
Trouble shooting guide
Problem| Remedy
• Compressor does not run when ‘up’ is selected.| • Ensure that the air hose
from the product being inflated is connected to the Airflo.
• Check that there are no kinks in the air hose.
• Compressor keeps stopping when trying to inflate the lifting aid.| • Check
that there are no kinks in the air hose.
• Bottom Battery Status Indicator light solid ‘Red’ and product will not
deflate.| • Low battery capacity, recharge for a minimum of 2 hours.
• Bottom Battery Status Indicator flashing red every 2 seconds.| • Internal
pump temperature too high – allow a cool down period of 60-90 minutes.
• Bottom Battery status indicator flashing Red every ½ second.| • Current
overload. Check there are no kinks in in the air hose.
• Compressor causes electrical interference with another piece of electrical
equipment.| • Move either the compressor or other item further away from each
other until the interference stops.
Specifications
Max. user weight: | 190kg (30st) |
---|---|
Inflated height: | 23cm |
Inflated width: | 43cm |
Inflated length: | 43cm |
Product weight: | 0.60kg |
Max. inlet pressure: | 0,35 bar (5psi) |
Max/Min temperature: | +50ºC / -15ºC |
Appendices
Spare Parts
Warranty
The warranty is given against defects in material and workmanship and
comprises piece, parts and labour. The warranty is not valid in the case of
fair wear and tear, wilful damage, common neglect, misuse or unauthorised
alteration or repair by unqualified persons.
The following statements do not affect the consumer’s statutory rights. This
product is sold on the understanding that in the event of any defect of
manufacture or material appearing within the warranty period, the defect will
be rectified free of charge providing that:
(a) Reasonable evidence is provided (e.g. purchase invoice, delivery note)
that the claim is being made within the warranty period.
(b) The defect is not attributable to accidental damage (either in transit or
otherwise), misuse or unauthorised repair.
Raiser Lifting Cushion serial number: …………………………
Date of purchase: …………………………
Invoice/Delivery note number: ………………………….
In the event of a warranty claim, please contact your supplier. A full warranty statement can be obtained by contacting Mangar Health on +44(0)1544 267674 or email sales@mangarhealth.com
Declaration of Conformity
Mangar International as manufacturer hereby declares that the following products conform to the:
- Applicable provisions of Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017
- UK Medical Devices Regulations 2002
and following standards:
BS EN 60601-1-6:2010 and A1:2015.
BS EN 60601-1:2006 and A12:2014.
BS EN 60601-1-11:2015. BS EN 62366-1:2015.
Trade name: Airflo 12 with Surfer Bather
Basic UDI-DI: 5020833MPCA040500PB
Product Classification: ISO 9999:2016, 12 36 18 Assistive products for lifting
persons
Classification: MDR Class 1
21.09.2020
Place and Date
Simon Claridge, CEO
Healthcare solutions designed for life
Mangar International Ltd
Presteigne, Powys, Wales, LD8 2UF
United Kingdom
Tel: +44 (0)1544 267674
Fax: +44 (0)1544 260287
E-mail: sales@mangarhealth.com
www.mangarhealth.com
WINNCARE France
4, Le Pas du Château
– 85670 Saint-Paul-MontPenit –
France
Tél : +33 (0)2 51 98 55 64
www.winncare.fr
MI0364 issue 7
(06/21)
Documents / Resources
|
Mangar HEALTH MPCA200500 Raiser Lifting
Cushion
[pdf] User Manual
MPCA200500, Raiser Lifting Cushion, MPCA200500 Raiser Lifting Cushion, Lifting
Cushion, Cushion
---|---
References
- Home Page - Mangar UK | Safe Lifting Products
- Groupe Winncare: Fabricant de lits médicalisés, matelas et coussins thérapeutiques, lèves-personnes et mobilier pour collectivités - Winncare France
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>