ANKOVO Non Contact Infrared Body Thermometer User Manual
- June 7, 2024
- ANKOVO
Table of Contents
- **ANKOVO Non Contact Infrared Body Thermometer User Manual
- Foreword
- Notice
- Safety Information
- Warning
- 1- OVERVIEW
- Thermometer Applications
- Equipment Symbols
- Technical Specifications
- Power Supply Requirements
- Calculated values of the indicators according to ISO 80601-2-56
- Safety classification of ME EQUIPMENT
- 2- OPERATION
- General Setup and Use
- 3- TROUBLESHOOTING
- 4-REPLACING THE BATTERY
- Warning
- 5-CLEANING. CARE AND STORAGE
- 6-DISPOSAL
- 7-WARRANTY
- 8-Calibration
- 9-EMC Declaration
- Read More About This Manual & Download PDF:
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
**ANKOVO Non Contact Infrared Body Thermometer User Manual
**
Foreword
The Non Contact Infrared Body Thermometer Operating Instructions intend to
provide the necessary information for proper operation of HTD8813C
Thermometer model.
General knowledge of Infrared Thermometer and an understanding of the features
and functions of the HTD8813C Thermometer model are prerequisites for proper
use.
The Non Contact Infrared Body Thermometer can be used repeatedly, which using
life is 3 years.
Please read the manual first before using it, if not fully understand the
usages, please stop using the thermometer.
Do not operate any of the model HTD8813C Thermometer without completely
reading and understanding these instructions.
Notice
Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would, alone or in combination with this device, fall within the scope of one of the relating patents.
Safety Information
This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application. The Non Contact Infrared Body Thermometer is designed to minimize the possibility of hazards from errors in the software program by following Sound and light Engineering Design Processes, Risk Analysis and Software Validation.
Warning
Warnings are identified by the WARNING symbol shown above.
- The Non Contact Infrared Body Thermometer is to be operated by consumers in the home setting and primary care setting as screening tool. This manual, accessories, direction for Use, all precautionary information, and specifications should be read before use.
- This product is designed to measure human body temperature on the forehead. Do not use it for any other purpose.
- This product is intended in the home setting and primary care setting as screening tool.
- Do not use the thermometer ifit malfunctions or has been damaged in any matter.
- When the ambient temperature of the thermometer changes too much, such as moving the Thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between15 Cto 35 C (59 “F – 95 ‘F).
- Remove primary batteries if equipment is not likely to be used for long time.
- This product is not waterproof, do not be immersed in water or other liquid; If cleaning and disinfection, please follow the “Care and Storage” section requirements.
- Do not touch the sensor of infrared detection with your fingers.
- If a cold compress on the forehead fever patients, or take other measures to cool down, the temperature data will be low; should be avoided in this case to measure body temperature.
- If measure human forehead temperature , please select “body” mode; for measure other objects, liquids, food and other temperature please select “surface” mode.
- This product must be operated in a stable environment, if the ambient environment was mutations, please should be note whether there is fog on the sensor, if any; before using accordance with the •care and Storage” section to removing the fog.
- Do not near strong electrostatic field or strong magnetic fields, thus avoiding the impact on the accuracy of the measurement data.
- Do not mix the old and new batteries to avoid damage to the product.
- It may affect the accuracy of measurements when the forehead is covered by hair, perspiration, cap or scart.
- The measuring result of this product is only for your reference. If you have any doubt, please measure the temperature in other methods.
The device should be kept out of the reach of children/pets. When not in use,
store the device in a dry room and protect it against extreme moisture, heat,
lint, dust and direct sunlight. Never place any heavy objects on the storage
case.
Do not throw batteries into fire.
Only use recommended batteries. Do not use rechargeable batteries.
This thermometer irreplaceable the diagnostic in hospitals.
Do not fall, disassemble or modify the device.
Do not use this device if you think it is damaged or notice anything unusual.
This device comprises sensitive components and must be treated with caution.
Observe the storage and operating conditions described in the ‘Technical
Specifications’ section.
Not servicing/maintenance while the Thermometer is in use.
When using, shall not touch battery and the patient simultaneously.
Do not use the device if it is damaged /degraded /lessened in any way. The
continuous use of a damaged unit may cause injury, improper results, or
serious danger.
Based on the current science and technology, other potential allergic
reactions are unknown.
This equipment needs to be installed and put into service in accordance with
the information provided in the ACCOMPANYING DOCUMENTS.
1- OVERVIEW
Intended Use
The Non Contact Infrared Body Thermometer are designed to be used for
intermittent measurement and monitoring of human body temperature by consumers
in the home setting and primary care setting as screening tool.
Description of Non Contact Infrared Body Thermometer
- Device principle and introduction The Non contact Thermometer are hand-held, reusable, battery operated devices, which can measure human body temperature on forehead, the skin temperature on one’s forehead. The operation principle is based on Infrared Sensor technology. The IR sensor can out put different signal when measuring different object temperature or in different ambient temperature, and the ASIC can turn the signal from IR Sensor to a digital value and display it on the LCD.
- Description on Controls, Indicators, and Symbols
Figure 1: HTD8813C infrared body thermometer.
- LCD Display
- Battery Cover
- ON/Scan Button
- Set Button
- MEMO Button
- MODE Button
- IR Sensor
- Surface Mode
- BodyMode
- Digital Display
- Smile or Cry Indicator
- Battery Indicator
- Buzzer on/ off Indicator
- Memory number
- °F
- ‘C
- .Memory
Thermometer Applications
Thermometer Model Number HTD8813C
Thermometer Style Non Contact Infrared Body Thermometers
Adult Ear Forehead
Pediatric Ear Forehead
Equipment Symbols
Warning
Non sterile Packaging
Refer to Operating instructions
Compliance with WEEE Standard
DO NOT THROW AWAY Intended for multiple use
Operating atmospheric pressure
Operating Temperature
Operating Humidity
recyclable
Manufacturer
serial number
This device complies with parts 15 of FCC ( Federal communications commission
) Rules
IP22: The first number 2: Protected against solid foreign objects of 12.5 mm l[I and greater. The second number: Protected against vertically falling water drops when enclosure titled up to 152,
Technical Specifications
Measurement Unit ˚C/ ˚F
Operating mode Adjusted mode(Body mode)
Direct mode(surface mode)
Measuring site Forehead
Reference Body Site Axillary
Rated output range Body mode: 35.0″C-
42.0″C/95.0″F-107.6″F Surface mode: o.o·c- 100.o·c; 32.0″F-212.0″F
Ext.ended output range Body mode: 34.0″C- 34.9″C/
93.2″F-94.8″F, 42.1″C-42.9″C/107.8-109.2″F
Range Body mode:34.0″C –
42.9″C / 93.2″F -109.Z”F Surface mode: o.o·c-100.o·c132-212″F
Accuracy Body mode: 34.0″C-
34.9″C :±0.3″C/ 93.2°F-94.8°F:±0.S°F; 35.0″C- 42.0″C :±o.z·c
/95.0″F-107.6″F:±0.4 “F; 42.l”C- 42.9″C :±0.3″C/107.8″F-109.2′ F: ±0.S’ F;
Surface mode:±2.0″q’±3.6″F
Display Resolution O.l”C/0.l”F Three-color
Bacldight (Color Alarm) 35.5″C-37.3″C / 95.9″F-99.l”F: Green (Normal
Temperature); 37.4″C-38.0″C(Alarm point)/ 99.3” F-100.4″F; Yellow (Slight
Fever) 38.l”C -42.9″C / 100.6″F-109.2″F; Rcd(High Fever) Note: 1.Surfacc mode
i5 always with Green backlight. 2. In body mode 34.0″C-35.4 •c is with green
backlight.
Auto Power Off Time les then 18 s
Measuring Time les then 2 s
Measuring Distance 1CM-SCM(0.4-2in)
Memories 50
Power Supply Requirements
Batteries 1.5 v {AAA)
alkaline batteryX2 (!EC Type LR03)
Adaptable Range 2.6V-3.6V
Environmental
Operating Condition I Operating
Temperature:15.0″C-35.0″C(59.0″F-95.0″F) , Relative Humiditys85% atmospheric
pressure;70Kpa -106Kpa
Transport and storage condition storage temperature. 20 c + 55 c /
-4 – 131 F relatively humidity 93% atmosphere pressure 70 – 108 Kpa
Calculated values of the indicators according to ISO 80601-2-56
Note: the above value is calculated from clinical data of HTD8818A
Safety classification of ME EQUIPMENT
Note; Not intended to he sterilized. Not for U5e in an OXYGEN RICH
ENVIRONMENT
2- OPERATION
2.1 Battery installation
Caution: The Non Contact Infrared Body Thermometer does not operate with
dead batteries and does not input outer power. Install new batteries.
- Pull the battery downward, toward the bottom of the Non Contact Infrared Body Thermometer, and remove the battery access door;
- Inserttwo pieces AAA size batteries according to the”+” and”-“;
- Close the battery cover.
2.2How to Operate
Before Applying the Thermometer Be sure to read and understand all Warnings listed of the instructions before use.
-
The thermometer is aligned with the middle of the forehead to measure body temperature (between the eyebrows above) and keep the vertical distance, press the On/Scan button, temperature display immediately, see figure 2.
Figure 2-Measuringposition and distance -
When the ambient temperature of the thermometer changes too much, such as moving the Thermometer from one place of lower temperature to another place of higher temperature, Allow the thermometer to remain in a room for 30 minutes where the temperature is between15°C to 35″C.
-
The ambient temperature around the test person should be stable, should keep away from the larger flow fan, air-conditioning vents and so on.
-
When people moving from one place of lower temperature to another place of higher temperature, should at least remain in the test environment more than 5 minutes, to be consistent with the ambient temperature after the re-measurement.
-
Wait at least 1 second for the next measurement. If the continuous measurement of five times, it is recommended to wait at least 30 seconds and then continue measurement.
-
You cannot use the thermometer in place where the sun is strong.
-
If for some reason the low forehead temperature measurement can try to align behind the ears. See figure 3.
Figure 3-Align behind the ears to measurement
General Setup and Use
■ Start measuring
- Turn on the thermometer by pressing the On/Scan button. The thermometer will perform self-test with all segments displayed for 2 seconds.
- Align staff forehead to keep the distance, and then press the On/Scan button to start the measurement, read the data.
Note: 1) After full display over, you will hear a rattle or “beep beep beep” three times, which means that the measurements have been completed, while the target value of the measured temperature is displayed on the LCD, while backlit display according to the appropriate setting among the three colors red, green, yellow one of. And the Green means ready for next measurement. When 37.4″C-38.0°C, it’s yellow ,means slight fever warning. Please pay attention to body temperature. When the body temperature is above 38.1 •c, it’s red , means high fever. Please take action to cool down or go for a doctor.
Note: 2) To ensure the accuracy of the measurement, wait at least 30 seconds after 5 consecutive measurements.
-
Mode conversion
When the device is running. press the MODE button to cycle conversion between “body” mode and “surface” mode. “body” mode is used for measuring human body temperature, the “surface ” mode is used to measure the surface temperature. (The factory default is “body” mode). -
Recalling and Erasing Memory Data The last temperature taken before the thermometer powers off is stored in memory, upto 50.
HTD8813C as below step:
- In the boot or shutdown state, short press the MEMO button to view the history of measured values.
- An empty memory cell shows •—°C” or “—°F”.
- Temperature readings can be stored in memory. Up to 50 temperature readings can be stored into the memory cells and automatically overwrite historical data.
- In boot mode, press the MEMO button until the LCD display “CLr”, which means that all stored data is cleared completely.
■ Parameter settings
This product can be set according to the subjects of different colors and different environments data to meet the different characteristics of populations or individuals. Long press the MODE button to modify the measurement parameters.
-
) Unit Set-Fl
Under the boot mode .Long press SET button to enter Fl, press the “MODE” or “MEMO” button to switch Celsius and Fahrenheit temperature units, press the SET button to confirm the unit settings (factory default is Fahrenheit). -
) Fever alert set-F2
Under Fl state, press SET button to enter the F2, press the “MODE” button to decrease 0.1 ‘C, press the “MEMO” button plus 0.1 ‘C, long press to accelerate th speed of temperature regulation, and finally press the SET button to save. (The factory default is 38.1 ‘C) -
) Prompt sound settings-F3
Under F2 state , short press SET button to enter F3, press MODE button or MEMO button to set voice switch, and press the SET button to confirm the settings. (The factory default is the voice Prompt to open). -
) Overall temperature offset value-F4
To meet the different colors, different characteristics of the population, or the environment caused by seasonal temperature change is large that need for temperature detection and debugging. Under F3 state , short press the SET button to enter F4, press the MEMO button to plus 0.1 ‘C, press the MODE button decrease 0.1 ‘C, long press fast subtraction temperature, and then SET button to confirm the parameter setting. Parameter adjustment range: -5-C and +S”C (factory default is O). -
) Exit setting mode
In the F4 mode, press the SET button will automatically turn off the screen, exit setting.-
3- TROUBLESHOOTING
4-REPLACING THE BATTERY
- Open and release battery cover following indicator on the surface of battery cover. Before changing the battery be sure the system is already power off.
- Remove the battery and replace with 2 new one, type AAA, Make sure align properly as indicated inside the battery cover .
- Slide the battery cover back in until it snaps in place.
- Do not dispose of used batteries in household waste. Take them to special local collection sites.
- In case, if system is latched up after changing battery. You may not follow up the process of rule one. Just take off battery, waiting 30 sec, then load battery again.
Warning
Do not recharge, disassemble or dispose of in fire.
- The typical service life of the new and unused batteries is 2000 measurements for the operation time is 18s.
- Only use the recommended batteries, do not recharge non-rechargeable batteries, and do not burn them.
- Remove the batteries if the thermometer is not to be used for a long period.
5-CLEANING. CARE AND STORAGE
The lens is very delicate. It is very important to protect the lens from dirt and damage. Use a clean, soft cloth to clean the surface of the device and LCD. Do not use solvents or immerse the device into water or other liquids. Always keep the thermometer a within the storage temperature and humidity range as specified. It is recommended to store the thermometer in a dry location free from dust. Always keep the thermometer within the storage temperature range (- 20 “Cto 55″C or – 4″F to 131’F) and humidity range (:!>93% non-condensing). It is recommended to store the thermometer in a dry location free from dust. Do not expose the thermometer to direct sunlight, high temperature/ humidity or any extreme environment, otherwise the function will be reduced. When the ambient temperature of the thermometer changes too much, such as moving the thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between15″C to 35°C.
6-DISPOSAL
- Used batteries should not be disposed of in the household rubbish. Used Batteries should be deposited at a collection point.
- At the end of its life, the appliance should not be disposed of in household rubbish. Enquire about the options for environment-friendly and appropriate disposal. Take local regulations into account
7-WARRANTY
Our company warrants Non Contact Infrared Body Thermometer at the time of its original purchase and for the subsequence time period of one year.
The warranty does not cover the followings:
- The device series number label is tom off or cannot be recognized.
- Damage to the device resulting from misconnection with other devices.
- Damage to the device resulting from accidents.
- Changes performed by users without the prior written authorization of the company.
- Batteries and packaging are not covered under warranty
When asked to provide warranty service, you must have a purchase date and purchase stamp dealers (including dealers name and address) of the warranty card. Be sure to ask the dealer to purchase this product signature on the warranty card. When asked to provide warranty service, please put the product to get our distribution points for repair. Products outside the warranty expires, will be charged accordingly. Note:
- If you have any problems with this device, such as setting up, maintaining or using, please contact with SERVICE PERSONNEL of us. Don’t open or repair the device by yourself.
- Please report to us if any unexpected operation or events occur.
- The patient is an intended operator. The patient can measure and change battery. Under normal circumstances and maintain the device and its accessories according to the user manual.
8-Calibration
The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to the use instructions ,periodic re-adjustment is not required .If at any time you question the accuracy of temperature measurements ,please contact us timely.
9-EMC Declaration
- This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS; This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
- Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
- Caution: this unit has been thoroughly tested and inspected to assure proper performance and operation!
- Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.