APERA PH20 Value pH Tester Kit User Manual
- June 7, 2024
- APERA
Table of Contents
PH20 Value pH Tester Kit
User Manual
APERA INSTRUMENTS, LLC
aperainst.com
Thank you for purchasing the Apera Instruments PH20 Pocket pH Tester Kit.
Please read this manual carefully before use in order to properly use and
maintain the product.
For video tutorials, please go to
support.aperainst.com
■ Latest Upgrade
The latest PH20 Tester comes with an upgraded probe structure, which is
equipped with a sensor shield that prevents the glass sensor breakage from
accidental collisions (see picture below). Users can remove the shield when
rinsing and cleaning the sensor and put it back on afterward.
Battery Installation
Please install batteries according to the following steps. *Please note the
correct direction of the battery installation:
The Positive Side (“+”) OF EVERY SINGLE Battery MUST FACE UP.
(THE WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER AND
POTENTIAL HAZARDS!)
Keypad Functions
■Short press——- < 2 seconds
■Long press——– > 2 seconds
| 1. Short press to turn on, long press to turn off;
2: When turned off, long press to enter setup;
3: In settings, short press to change parameter;
---|---
| 1.When turned on, long press to enter calibration mode.
2.In calibration mode, short press to finish calibration;
3.In settings, short press to confirm changes.
What’s in the Kit?
Preparation before Use
Things needed in addition to what’s in the box
A cup of distilled water (8-16oz) for rinsing the probe, and clean tissue
paper for removing excess water.
■ Note
Before shipment, the factory adds some water droplets to the tester’s cap to
maintain a hydrated atmosphere for the probe. This ensures that the pH sensor
keeps its sensitivity.
Therefore, users can begin using their tester immediately upon receiving it.
This practice of adding several drops of water into the cap is highly
recommended for frequent users when storing the tester. However, if the tester
has not been used for a long time (e.g. > 1 month) and doesn’t perform as well
as it used to be, users can soak it in a 3M KCL (SKU: AI1107) solution
overnight to recover its sensitivity.
- DO NOT use any other brand’s storage solutions because different chemicals may be used and potential damage could be caused to the probe.
Calibration
- Short press to turn on. Rinse in distilled water; use a clean tissue to dap off excess water.
- Long press to enter calibration mode (short press to go back to measurement mode if you want to cancel the calibration).
- Insert the probe in the 7.00 pH calibration solution to start 1 st point calibration; Stir gently; leave it to stand; Wait for to appear and stay on the screen (see Diagram 3); short press to finish 1st point calibration; tester returns to measurement mode; calibration M icon displays on the button left side of the screen.
- Rinse the probe in distilled water. Long press to enter calibration mode; insert the probe in the pH 4.00 calibration solution to start the 2 nd point calibration, stir gently; leave it to stand; wait for to appear and stay on the screen; then short press to finish the 2nd point calibration, tester returns to measuring mode, calibration icons L M display on the bottom left.
- Repeat step 4) and use 10.01 pH standard buffer (sold separately) to perform 3 rd point calibration L M H. will show up on the bottom left.
■ Notes
a) Tester will automatically recognize pH buffer solution, users can choose
calibration points: 1 point, 2 points, or 3 points. But the 1st point
calibration must be in calibration solution of 7.00 pH, then followed by 2nd
or 3rd point calibration. For details, please refer to the following table:
| Calibration Solution| Calibration icon| When to do
---|---|---|---
1-point
Calibration| 7.00 pH|
M
| Required Accuracy.? 0.3 pH
2-point
Calibration| 7.00 pH and 4.00 pH|
L M
| Target pH level< 8.5 pH
7.00 pH and 10.01 pH|
L M
| Target pH level> 8.5 pH
3-point
Calibration| 7.00 pH, 4.00 pH
and 10.01 pH|
L M H
| Wide measuring range
b) Automatic self-diagnostic information: In calibration mode, when the
measured value hasn`t been stable yet ( is not staying on LCD), pressing will
cause LCD to display
“Er2”). For more details, please refer to Section 8.
c) Only pH 4.00 and pH 7.00 buffer solutions are included in the test kit, pH
10.01 is not.
Users can purchase it separately if needed (when your target pH level is >8.5
pH).
d) Keeping the standard calibration buffers in good condition is the
foundation of taking reliable pH measurements. The bottled buffers (2 oz.) can
be used for about 10 times as long as they are not contaminated (make sure
they are tightly closed when not in use).
After that, we recommend replacing them with fresh ones to keep the accuracy.
Measurement
- Short press to turn on the tester. Rinse the probe in distilled water. Then shake off excess water and dap with tissue paper.
- Stir the probe in your sample solution to accelerate stabilization and leave it to stand.
Record the readings after comes up and stays on the screen.
■ Notes
a) After each test, users should rinse the pH probe thoroughly with distilled
water or purified water and remove excess water before the next test. Warm
soap water and a
soft brush can be used to remove tough contaminants on the glass sensor.
b) When testing purified water like RO water or distilled water, it will take
longer for the readings to get fully stabilized (typically 2-5 minutes).
c) Do NOT store the probe in purified water because that will cause permanent
damage to the pH electrode. Purified water is only recommended for rinsing the
probe.
Parameter Setting
Settings table
Prompt Mark| Parameter Setting Content| Code| Factory
Default Setting
---|---|---|---
P1| Select pH buffer series| USA — NIST| USA
P2| Select temperature unit| °F — °C| °F
P3| Back to the factory default setting
(erase calibration record)| No — Yes| No
Parameter Setting
When turned off, long press to enter setup → short press to switch P1-P2-P3→ Short press , parameter flashing →short press to choose, short press to confirm parameter change →Long press to go back to measurement mode.
Parameter Setting Instruction
Select standard pH buffer solution (P1): There are two options of standard buffer solutions: USA series and NIST series (refer details to the following table).
Icons | pH standard buffer series |
---|---|
USA series | NIST series |
Three-Point calibration | L |
M | 7.00 pH |
H | 10.01 pH |
Common Problems
The tester has a self-diagnosis function to help you fix common problems:
Symbol | Self-Diagnosis information | How to fix |
---|
| Wrong
pH calibration solution, which exceeds the recognizable range of the meter.|
1. Check if the correct buffer is used (1st point calibration must be 7.00
pH).
2.Check if the buffer is expired or contaminated.
3. Check if the probe is damaged.
4. Check if there is an air bubble in the glass bulb sensor
| Is pressed before the measurement is stable ( e) appears and stays)| Wait
for the smiley face to appear and stay, then press .
-
If any air bubbles are found inside the sensor shield, shake the probe in sample solution with force to remove them. The existence of air bubbles will cause unstable measurements.
-
For more FAQs, go to support.aperainst.com
Technical Specifications
pH | Range | 0 — 14.0 pH |
---|---|---|
Resolution | 0.1 pH | |
Accuracy | ±0.1 pH | |
Calibration Points | 1 — 3 points | |
Automatic Temperature Compensation | 0 — 50°C (32 — 122°F) | |
Temperature | Range | 0 — 50°C (32 — 122°F) |
Resolution | 0.1 °C | |
Accuracy | ±0.5°C |
Icons and Functions
8.1 Calibration points indication: L M H
8.2 Stable Measurement: appears and stays on the screen
8.3 Self-Diagnostic information: Er1, Er2
8.4 Low-Voltage warning: flashes, a reminder of battery replacement
8.5 Auto Power-Off in 8 minutes if no operation.
Warranty
We warrant this instrument to be free from defects in material and workmanship
and agree to repair or replace free of charge, at the option of APERA
INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to the
responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS (SIX MONTHS
for the probe) from the delivery. This limited warranty does not cover any
damages due to: Transportation, storage, improper use, failure to follow the
product instructions or performing any
preventive maintenance, modifications, combination or use with any products,
materials, processes, systems or other matter not provided or authorized in
writing by us, unauthorized repair, normal wear, and tear, or external causes
such as accidents, abuse, or other actions or events beyond our reasonable
control.
APERA INSTRUMENTS, LLC
Address: 6656 Busch Blvd, Columbus Ohio 43229
Tel: 1-614-285-3080
Email: info@aperainst.com
Website: aperainst.com