COVID-19 Test Kit
(Colloidal Gold Method )
INTENDED USE
COVID-19 Test Kit(Colloidal Gold Method)is a solid phase immunochromatographic assay for the rapid, qualitative detection of antigen to 2019 Novel Coronavirus. This test provides only a preliminary test result. Therefore, any reactive specimen with the COVID-19 Test Kit(Colloidal Gold Method)must be confirmed with alternative testing method(s) and clinical findings.
PACK FORMATS
1 test/box
20 tests/box
INTRODUCTION
The novel coronaviruses belong to the β genus. COVID-19 is an acute respiratory infectious disease. People are generally susceptible. Currently, the patients infected by the novel coronavirus are the main source of infection; asymptomatically infected people can also be an infectious source. Based on the current epidemiological investigation, the incubation period is 1 to 14 days, mostly3 to 7 days. The main manifestations include fever, fatigue, and dry cough. Nasal congestion, runny nose, sore throat, myalgia, and diarrhea are found in a few cases.
PRINCIPLE
The COVID-19 Test Kit(Colloidal Gold Method) is a colloidal gold
immunochromatographic assay. The test uses COVID-19 (SARS-CoV-2) antibody
(test line T) and goat anti-mouse IgG (control line C) immobilized on a
nitrocellulose strip. The burgundy-colored conjugate pad contains colloidal
gold conjugated to another COVID-19 (SARS-CoV-2) antibody conjugated with
colloid gold and mouse IgG-gold conjugates. When the processed buffer
containing the sample is added to the sample well, COVID-19 (SARS-CoV-2) will
combine with the COVID-19 antibody conjugate to form an antigen-antibody
complex. This complex migrates through the nitrocellulose membrane by
capillary action. When the complex meets the line of the COVID-19 antibody of
test line T, the complex is trapped forming a burgundy colored band which
confirms a reactive test result. The absence of a colored band in the test
region indicates a non-reactive test result.
The test contains an internal control (C band) which should exhibit a burgundy
colored band of the immunocomplex goat anti-mouse IgG/mouse IgG-gold conjugate
regardless of the color development on any of the test bands. Otherwise, the
test result is invalid and the specimen must be retested with another device.
MATERIALS SUPPLIED
Sealed pouches each containing a test cassette, a desiccant
Sampling cotton swabs
Antigen extract buffer
Antigen extraction tube
Paper workbench (The small one-test-box can be used as a workbench)
Instruction for use
MATERIAL REQUIRED BUT NOT PROVIDED
- Specimen collection containers
- Timer
STORAGE AND STABILITY
The kit can be stored at room temperature or refrigerated (4-30°C). The test
device is stable through the expiration date printed on the sealed pouch. The
test device must remain in the sealed pouch until use.
Do not freeze.
Do not use it beyond the expiration date.
WARNINGS AND PRECAUTIONS
- For professional In Vitro diagnostic use only. Do not use it after the expiration date.
- This package insert must be read completely before performing the test. Failure to follow the insert gives inaccurate test results.
- Do not use it if the tube/pouch is damaged or broken.
- The test is for single use only. Do not re-use under any circumstances.
- Handle all specimens as if they contain infectious agents. Observe established precautions against microbiological hazards throughout testing and follow the standard procedures for proper disposal of specimens.
- Wear protective clothing such as laboratory coats, disposable gloves, and eye protection when specimens are assayed.
- Humidity and temperature can adversely affect results.
- Do not perform the test in a room with strong airflow, ie. an electric fan or strong air-conditioning.
SPECIMEN COLLECTION
-
COVID-19 Test kit(Colloidal Gold Method )can be performed using the following sampling:
I. ) Saliva swab -
Testing should be performed immediately after specimen collection.
-
Bring specimens to room temperature prior to testing.
-
If specimens are to be shipped, they should be packed in compliance with local regulations covering the transportation of etiologic agents.
SPECIMEN COLLECTION
Swab collection (Recommended for self-test usage)
Remove the swab package, keep the sponge tip on the tongue until the tip is
fully soaking by saliva (at least two minutes).
TEST PROCEDURE
Allow test cassette, specimen, and Antigen extract buffer control to equilibrate to room temperature (15-30°C) prior to testing.
- Remove the test cassette from the sealed foil pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour.
- Place the test device on a clean and level surface.
Test procedure:
- Place the antigen extraction tube on the workbench. Place the antigen extraction buffer bottle vertically downward, squeeze the bottle to make the buffer drip freely into the extraction tube without touching the edge of the tube, and add 6 drops (about 200ul) to the extraction tube.
- Put the swab specimen into the extraction tube pre-added with the antigen extraction buffer, and rotate the swab about 10 times while pressing the swab head against the tube wall to release the antigen in the swab, then let it stand for about 1 minute.
- Remove the swab while squeezing the tip of the swab so that as much liquid in the swab can be discharged as possible. Dispose of used swabs in accordance with biohazard waste disposal methods.
- Install the dripper on the extraction tube and cap it tightly, and let it stand for about 1 minute.
- Open the aluminum foil bag and take out the test card, add 3 drops (about 100ul) into the sample hole of the test card (or use a pipette to add 100ul), and start the timer.
- Wait for the colored line to appear. The result should be read in 15 minutes. Do not interpret the result after 20 minutes.
INTERPRETATION OF RESULTS
NEGATIVE: If only the C band is present, the absence of any burgundy color in
the T band indicates that no COVID-19 (SARS-CoV-2) antigens are detected in
the specimen. The result is negative.
COVID-19 positive:
In addition to the presence of the C band, if the T band is developed, the
test indicates the presence of the COVID-19 (SARS-CoV-2) antigen in the
specimen. The result is COVID-19 positive.
INVALID:
The Control line fails to appear. Insufficient specimen volume or incorrect
procedural techniques are the most likely reasons for control line failure.
Review the procedure and repeat the test with a new test cassette. If the
problem persists, discontinue using the test kit immediately and contact your
local distributor.
LIMITATIONS
- Use fresh samples whenever possible.
- Optimal assay performance requires strict adherence to the assay procedure described in this insert sheet. Deviations may lead to aberrant results.
- A negative result for an individual subject indicates the absence of detectable COVID-19 (SARS-CoV-2) antigen. However, a negative test result does not preclude the possibility of exposure to or infection with COVID-19.
- A negative result can occur if the quantity of the COVID-19 (SARS-CoV-2) antigen present in the specimen is below the detection limits of the assay, or failed to collect the COVID-19 (SARS-COV-2) antigen
- As with all diagnostic tests, a definitive clinical diagnosis should not be based on the result of a single test, but should only be made by the physician after all clinical and laboratory findings have been evaluated.
PERFORMANCE CHARACTERISTICS
- Accuracy
A total of 518 samples were tested in this study. The results of the test reagent and control reagent both were 309 negative specimens and 209 positive specimensMethod| RT-PCR| Total
---|---|---
Positive| Negative
COVID-19 Test Kit| Positive| 206| 3| 209
Negative| 3| 306| 309
Total| 209| 309| 518
Sensitivity| 98.56 %| confidence Interval| 95.87 %~99.51 %
Specificity| 99.03 %| confidence interval| 97.18 %~99.67 %
Accuracy| 98.84 %| confidence interval| 97.50 %~99.47 % - Analytical Specificity
Cross-Reactivity| Human coronavirus 229E, human coronavirus OC43, human coronavirus HKU1, human Coronavirus NL63, adenovirus (type 5), adenovirus(type 7), adenovirus (type 18), human metapneumovirus (hMPV), parainfluenza virus(type 1), influenza A virus, influenza B virus, Haemophilus influenza, rhinovirus(type 2), rhinovirus (type 14), rhinovirus (type 16), respiratory syncytial virus(type A-2), Streptococcus pneumonia, and Streptococcus their mo, there is no crossover with this product.
---|---
Interfering Substances| 2mg/mL hemoglobin, 2mg/mL mucin, 5mg/L human anti- mouse antibody (HAMA), 10mg/mL biotin, 500μg/mL mucus, 3μg/mL gentamicin, 120μg/mL cromolyn sodium, 60μ g/mL oxymetazoline hydrochloride, 200μg/mL henylephrine hydrochloride, 200μg/mL N-Acetaminophen, 3μg/mL aspirin, 3μg/mL ibuprofen, 3μg/mL morpholino hydrochloride, 3μg/mL cephalexin kanamycin, 3μg/mL tetracycline, 3μg/mL chloramphenicol, 3μg/mL erythromycin, 3μg/mL vancomycin, 3μg/mL nalidixic acid, 3μg/mL hydrocortisone Pine and 3μg/mL human insulin will not affect the test results. - Limit of Detection
Limit of detection| 2ng/ml
---|---
REFERENCE
- Weiss SR, Leibowitz JL. Coronavirus pathogenesis. Adv Virus Res 2011; 81: 85-164.
- Masters PS, Perlman S. Coronaviridae. In: Knipe DM, Howley PM, eds. Fields virology. 6th ed. Lippincott Williams & Wilkins, 2013: 825-58.
- Su S, Wong G, Shi W, et al. Epidemiology, genetic recombination, and pathogenesis of coronaviruses. Trends Microbiol 2016; 24: 490-502.
- Cui J, Li F, Shi ZL. Origin and evolution of pathogenic coronaviruses. Nat Rev Microbiol 2019; 17: 181-192.
SYMBOLS USED ON PACKAGING NOTICE
Hangzhou Singclean Medical Products Co., Ltd.
No. 125(E), 10th street, Hangzhou Economic and Technological
Development Zone, Zhejiang, China. P.C.: 310018
SUNGO Europe B.V.
Add: Olympisch Stadion 24, 1076DE Amsterdam, Netherlands
www.goodtest.eu
Rev.: 8.129.05.020-A1
Date of last revision: 2. 3. 2021