C165 QUICK START GUIDE

GETTING STARTED

Press “Power” in the fresh air and allow to countdown for Fresh Air purge.

Rotate selector dial to Flue 1. In fresh air, the O2 reading should be 20.9% ±0.3%.
Rotate selector dial to Flue 2. In fresh air, the CO reading should be zero (0).
Test Equipment C165 - Test Equipment C165

Connect the flue probe thermocouple connector to T1.
Connect flue probe to the water trap.
Connect the optional probe to T2.

Test Equipment C165 - Test Equipment C16

BASIC CO/COMBUSTION ANALYSIS

Insert Flue probe into the stack. Adjust cone so the end of the probe is approximately at the center of the stack.
NOTE: You will have to drill a hole at least 3/8″. Use appropriate sealing method after testing.
Test Equipment C165 - Test EquipmentRotate selector dial to Flue 1 for CO2, O2, Gross efficiency, Flue temperature, Inlet temperature. Fuel type can be changed via MENU or STATUS setting
Test Equipment C165 - Test EquipmeRotate selector dial to Flue 2 for CO (ppm), Losses calculated, Excess air %, pressure reading Fuel type can be changed via
MENU or STATUS setting. Make any adjustments as needed for proper combustion and wait for the analyzer to display changes in readings. (Repeat as necessary).
Test Equipment C165 - Test

Once complete, remove the probe from the stack, allow the analyzer to purge in fresh air until CO sensor readings return to ZERO (0) and O2 readings reading return to 0% to 21%. Continue to the next test or power analyzer off by pressing On/Off, if finished.
NOTE: Print and store functions may be used at any time during testing.
Test Equipment C165 - Test Equipme

DIFFERENTIAL TEMPERATURE TEST

Rotate the selector dial to
Pressure/Temperature.
Test Equipment C165 - coPress “Pump” to ZERO (0) pressure sensor. Connect the probe to T1. Connect the probe to T2. Compatible with any K-Type thermocouple probe or clamp.Test Equipment C165 - connect
probr toConnect thermocouples in test location to start testing
Observe T1, T2, and differential (Delta T).
Test Equipment C165 - T1 T2

Press “Print” to print results. Press and Hold results.
Test Equipment C165 - pintar

DRAFT & STATIC PRESSURE TEST

Rotate the selector dial to Pressure/Temperature.Test Equipment C165 -
p Press “Pump” to ZERO (0) pressure sensor.

Connect a true draft hose to PI. Connect to P2 for differential. Place probe tip in the stack to measure draft.
Connect static pressure hose to measure differential pressure.
Test Equipment C165 - T2a

Place true draft probe tip in the flue to measure draft.
You can also use the combustion probe for measuring pressure.

Press “Print” to print results.
Press and Hold results.

HEAT EXCHANGER TEST

There are many methods to test heat exchanger integrity. One of these is to observe Excess Air, O2, AND CO readings both before and after the blower turn on. If the heat exchanger is sealed your O2 and CO readings should remain fairly stable. A breach in the heat exchanger may allow fresh air to be forced into the flue after the blower turns on due to a pressure increase in the plenum. The result may be a rise in the measured O2 in the stack gas and an increase in the Excess Air. In some sealed systems the fresh air drawn in the breach may reduce the combustion available leading to an increase in the CO reading. If either of these situations is present it is probable that there is a problem with the Heat Exchanger which may require additional testing and inspection.
NOTE: Many cracks are invisible to borescopes or the naked eye and only open or separate from pressure or temperature during operation.
Test Equipment C165 - ExchConfirm the Fuel Type. Scroll or switch a different fuel type.

Press  to select fuel type.
Test Equipment C165 - confirm the fuel

NOTE: The analyzer will perform a fresh air purge if it did not perform one on startup. Otherwise, it will skip past the zeroing step to the Blower Offscreen. See manual for details.  Press to continue.
Test Equipment C165 - NoteEnsure the analyzer is in fresh air during the countdown. The screen will automatically advance when complete.
Test Equipment C165 - picaa

Call for heat.
Observe and wait for O2 readings to stabilize. after the blower turns on, press the button to start the Post-Blower test.
Test Equipment C165 - callThe analyzer will wait for 60 seconds and then record the
Post-Blower values for CO,02, and Excess Air.
Test Equipment C165 - 60

Test results will automatic-cally be stored to exchange reports. The report includes both Pre and post Blower test segments.

Press to start printing. See the REPORTS section in this guide.

CO ROOM TEST
Great for checking ambient CO and back drafting situations
No probes or hose connections are required for this test.
Rotate selector dial to ROOM CO. Press “Print” to start the GENERAL test.
Test Equipment C165 - 1

Press Flue 2 to start the test.

CO readings will be logged every 2 minutes for a 30-minute time span.
Readings will be stored once the 30-minute test is completed.
Test Equipment C165 - 2

Press to print results.
Test results will automatically be stored in ROOM CO reports. See the REPORTS section in this guide.
Test Equipment C165 - 3

AUX SETTING
Programmable Auxiliary screen allows for Tech selectable test parameters to be chosen.
Rotate selector dial to Menu. Scroll or to Screen. Press “Print” to select.
Test Equipment C165 - AuxScroll or customize any Line. Press to select. [Line 5 shown]Test Equipment C165 - scroll Scroll or select the parameter for this line. Press to select. [Pressure shown]Test Equipment
C165 - ora

Rotate selector dial to AUX to view.
Custom selections will now display in A U X mode.
Test Equipment C165 - auxaa REPORTS: PRINTING & VIEWING
Print and Log easily from the following test screens: Aux, Flue 1, Flue 2, Commission Test, Pressure/Temperature, Exchange Test, and Room CO
Press **“Print” to start printouts of results from any test screen. Press during printing to cancel.
Test Equipment C165 - Rep1 Press and hold to log current readings. Align printer above Analyzer.
Test Equipment C165 - to logTo view test results, rotate the dial to
Menu. Scroll **or  to Report Press to select.
Test Equipment C165 - menu

Scroll  or  type of test wanted. Press to select. [Exchange shown] Press to select View.
Test Equipment C165 - to scroll
Press to select report log number. When indicator lights blink, use or scroll through logged reports. Press to view a selected report log.

**HIGH CO ALERT

**

At 1000 ppm CO the screen will display HIGH CO and an indicator light will flash and the analyzer will beep several times. Above 2000 ppm the purge pump starts.

MAIN MENU

| SUBMENU|

OPTIONS / COMMENTS

---|---|---
SETUP| LANGUAGE| English
SET TIME| HH:MM: SS format

e.g. 7 am = 07:00:00, 7 pm =19:00:00

SET DATE| MM/DD/YY format
PRINTER| KM IRP KANE IRP•2

WIRELESS lif installed)

SERIAL

PASSKEY| 1111(wait 5 secs after entering the last digit)
BACK|
UNITS| FUEL TYPE| NAT GAS, TOWN GAS, COKE GAS, PROPANE, BUTANE, LPG, LIGHT OIL, BIO-OIL. WOOD PELLETS, BIOGAS, USER 1 to 5
FUEL ORIGIN| UK, FRANCE, SPAIN, N AMERICA, BELGIUM, NETHERLAND
EFFICIENCY| GROSS. NET. GROSS COND. NET COND
PRESSURE| FILTER: OFF = normal response. ON = slower (damped) response
RESOLUTION: LOW = e.g. 0.00i inH20 resolution. HIGH = displays roan extra decimal place
UNITS: mbar,     Pa, PSI, mmHg, hPa, inH20, mmH20, kPa, psi
TIME: Test units not available in North American market
BACK:
GAS| PPm. PPal(n), mg/m3, mg/m3(n), mg/kWh, mg/kWh(n)
| TEMP| C. F
02 REF| Up/down to set value (3% default)
NOx CALC| Up/down to set value 15% default)
BACK|
SCREEN| CONTRAST| The factory setting is 14
BACKLIGHT| 0 to 300 secs
AUX| Enables users to customize the parameters on the AUX display: LINE 1, LINE 2, LINE 3. LINE 4, LINE 5, LINE 6, BACK
BACK|
REPORT| AUX| Stored AUX tests VIEW, DEL ALL, BACK
COMBUSTION| Stored combustion tests: VIEW, DEL ALL, BACK
COMMISSION| Stored commission tests: VIEW, DEL ALL, BACK
PRS/TEMP| Stored pressure tests: VIEW, DEL ALL, BACK
EXCH| Stored exchange tests: VIEW, DEL ALL, BACK
ROOM CO| Stored room CO tests:
VIEW, DEL ALL, BACK
HEADER| LINE 1
LINE 2
BACK
BACK|
SERVICE| CODE| Password protected for authorized service agents only. Leave set to 000000.