EXTECH CO200 Desktop Indoor Air Quality CO2 User Manual

June 6, 2024
EXTECH

EXTECH logo s123

USER MANUAL

CO2 Monitor

Model CO200

EXTECH CO200 Desktop Indoor Air Quality CO2

Introduction

Congratulations on your purchase of the Model CO200 Carbon Dioxide Meter. The CO200 measures and displays CO2 level, air temperature, and relative humidity. Visible and audible alarm (with alarm relay output), and real-time calendar clock, make this an ideal instrument for indoor air quality (IAQ) monitoring. The CO200 is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.

Description

EXTECH CO200 - METER Description

METER
  1. LCD Multifunction display
  2. Power LED indicator
  3. Control buttons

Power adaptor jack, alarm relay jack, and sensors on rear

EXTECH CO200 - LCD DISPLAY Description

LCD DISPLAY
  1. CO2 concentration in ppm
  2. Calendar clock
  3. Air Temperature
  4. % Relative Humidity
FUNCTION INDICATORS
ppm CO2 measurement value
GOOD CO2 (good air quality level)
NORMAL CO2 (normal air quality level)
POOR CO2 (poor air quality level)
Air Temp Air Temperature
Humidity % Relative Humidity
°C or °F Celsius or Fahrenheit units
MIN/MAX Minimum or Maximum readings
**** Flashes when CO 2 alarm is triggered (relay activates)
CONTROL BUTTONS
SET Long press to enter setup mode. Saves settings.
ESC Exits setup page/mode. Terminates CO2 calibration.
RESET Long press to clear the MIN/MAX memories.
Select a mode or increase a value in the setup menu.
MIN/MAX Access MIN/MAX function or use as down-arrow in setup

mode.
Long press SET, MIN/MAX, and ▲to enter the CO2 calibration mode.

Operation

Powering the CO200

Plug the supplied power adaptor (5V/0.5A) into the power jack on the back of the meter (ensure that you connect to the power jack on the left, not the alarm relay jack on the right; see diagram in the Alarm Relay section below).

Now connect the power adaptor to an AC power source, the meter will switch ON, the front panel LED power indicator will light, and the beeper will sound. If the power source voltage is incorrect, “ bAt “ will display and the LED indicator will flash. Do not use the CO200 with an incorrect power source.

In normal operation the LCD will display CO2, temperature, humidity, date, and time. The air quality level is also displayed (GOOD, NORMAL, POOR).

 EXTECH CO200 - Operation - Powering the CO200

Measurements

The meter starts measuring as soon as it is powered and updates readings once per second. If the operating environment changes (from high to low temperature, for example), it will require 2 minutes for the CO2 sensor to respond and 10 minutes for the Relative Humidity sensor to respond.

Note: Do not hold the meter close to your mouth or any other source of CO2.

MIN/MAX Mode

In the normal mode, press the MIN/MAX button to see the minimum or the maximum CO2, Temperature, and Humidity values. Each press of the MIN/MAX button steps through the MIN, the MAX, and then the normal display mode.

Long press the RESET button to clear the minimum and maximum values from memory.

Alarm Function

The CO200 features an audible (beeper) and visual (flashing fan icon) alarm that warn when the CO2 concentration exceeds the programmed threshold. There are two alarm thresholds, the ‘POOR’ air quality threshold (that triggers the alarm), and the ‘NORMAL’ air quality threshold (that cancels the alarm). See the Setup mode to set these threshold settings.

The alarm output relay switches on and off as the CO2 alarm switches on and off (see next section).

The audible alarm can be manually silenced by pressing any key, or it will automatically stop when the CO2 reading is no longer in the alarm region (the user can subsequently press and hold the RESET button to re-activate the alarm, if desired).

EXTECH CO200 - Operation - Alarm Function

If the beeper is manually silenced, it will sound again when readings move out of, and then back into, the alarm region.

The fan icon continues to flash even when the beeper is manually silenced. It disappears only when readings are no longer in the alarm region.

Alarm Relay Output

The CO200 includes an alarm relay output that can be connected to an external device such as an alarm light, warning sign, alarm buzzer, or controlling device (ON/OFF switch to a fan, for example).

The relay switches ON and OFF as the CO200’s CO2 alarm switches ON and OFF (see previous section for more on the alarm feature). The fan display icon flashes each time the relay closes.

The rear relay output jack requires a 2.5mm stereo phone plug. The relay is rated for 1A at 30VDC or 0.5A or 125VAC.

EXTECH CO200 - Operation - Alarm Relay Output 1 EXTECH CO200 - Operation - Alarm Relay Output
2

  1. AC adaptor input (5VDC)
  2. Alarm relay output
  3. Phone plug wiring
Setup Mode

Long press the SET button to enter the setup mode.

P1.1: Threshold setting for the ‘GOOD’ air quality display icon

When entering the setup mode, P1.0 and ‘CO2’ are displayed. Press the SET button again to access P1.1 for setting the CO2 upper limit (threshold) for the ‘GOOD’ air quality display icon. The current set value will appear blinking on the display.

EXTECH CO200 - Setup Mode - P1.0 EXTECH
CO200 - Setup Mode - P1.1 EXTECH CO200 - Setup Mode -
P1.2

Press the or button to increase or decrease the value. Each press adjusts in 100 ppm increments. The range is 0 to 700 ppm (default is 700 ppm).

When the value has been set, press the SET button to confirm the ‘GOOD’ threshold and move to P1.2 for setting the ‘NORMAL’ display threshold. Press the ESC button to exit without saving.

P1.2: Threshold setting for the ‘NORMAL’ air quality icon and the CO 2 Alarm

P1.2 is used for setting the CO2 high threshold for the ‘NORMAL’ air quality display icon. It is also the low threshold for the alarm beeper, the flashing fan alarm icon, and the output relay. The current set value will appear blinking on the display.

Press the or button to increase or decrease the value. Each press adjusts in 100 ppm increments. The range is 700 to 1000 ppm (default is 1000 ppm).

When the value has been set, press the SET button to confirm the ‘NORMAL’ limit and move to P1.3 for setting the ‘POOR’ display threshold. Press the ESC button to exit without saving.

P1.3: Threshold setting for the ‘POOR’ air quality icon and CO 2 Alarm

P1.3 is used for setting the high CO2 threshold for the ‘POOR’ air quality display icon, alarm beeper, flashing fan icon, and output relay. The current set value will appear blinking on the display.

Press the or button to increase or decrease the threshold. Each press adjusts in 100 ppm increments. The range is 1000 to 5000 ppm (default is 1000 ppm).

When the value has been set, press the SET button to confirm and return to the P1.0 screen. Press ESC to exit without saving.

EXTECH CO200 - Setup Mode - P1.3

Caution: Set the alarm value within the specified range of the CO200 where accuracy is assured. Out of specification readings are for reference only.

P2.0 Temperature Units

Press the button in P1.0 mode to access P2.0 for setting the temperature display units.

Press the SET button to access the P2.1 setting mode. The °C or °F will blink (default is °C).

Press the button to change the units. Press the SET button to confirm the setting or press the ESC button to exit without saving and return to P1.0.

EXTECH CO200 - Setup Mode - P2.0 EXTECH
CO200 - Setup Mode - P2.1

P3.0 Real-Time Clock

Press the button in P1.0 twice to access P3.0 for setting the real-time calendar clock. Press the SET button and the meter will access P3.1, showing the blinking YEAR on the lower left. To change the year, press the or button. Press the SET button to save the setting and to enter P3.2 (or press the ESC button to return to P3.0 without saving the setting).

Press the button from P3.1 to access P3.2. The current MONTH setting will blink. To change the month, press the or button. Press the SET button to save the setting and to enter P3.3 (or press the ESC button to return to P3.0 without saving the setting).

Repeat as above to finish setting the P3.3 (DATE), P3.4 (HOUR) and P3.5 (MINUTE)

EXTECH CO200 - Setup Mode - P3.0 EXTECH
CO200 - Setup Mode - P3.1

P4.0: Factory Default Reset

Press the button in P1.0 three times to access P4.0 where you can reset the meter to its default settings. Press SET and the meter will access P4.1 with a blinking “No”. Press the to switch the status to ‘Yes’ and then press either the SET button to confirm and continue or the ESC button to abort the process.

EXTECH CO200 - Setup Mode - P4.0 EXTECH
CO200 - Setup Mode - P4.1

Calibration

The meter is calibrated to a 400ppm CO2 concentration at the factory. If the accuracy becomes a concern, or after one year of use, return the meter to Extech for calibration or continue below.

ABC (Automatic Baseline CO2 Calibration)

ABC (Automatic Baseline Calibration) establishes a baseline calibration to eliminate the zero drift of the infrared sensor. The ABC function is always “ON” when the meter is powered.

ABC is designed to calibrate the meter at the minimum CO2 reading detected over 7.5 days of continuous monitoring (power on). It assumes that the area being tested receives fresh air with a CO2 level of approximately 400ppm at some point during the monitoring period. Note that It is not suitable to use the CO200 in closed areas with consistently high CO2 levels 24 hours a day.

Manual CO2 Calibration

Manual calibration is suggested to be done outdoors on a clear day with good ventilation and fresh air where the CO2 level is approximately 400 ppm. Do not calibrate on a rainy day because high humidity will affect the CO2 level in air.

CAUTION: Do not calibrate the meter in an atmosphere of unknown CO2 concentration. Do not calibrate in places crowded with people, pets, plants, or where high CO2 concentrations may exist such as near ventilation ducts or fireplaces.

EXTECH CO200 - Manual CO2 Calibration

Place the meter in the calibration area and switch it on. Press and hold the SET , , and buttons simultaneously for > 1 second to enter CO2 calibration mode. “400ppm” and “CO2” will blink while the calibration is in process.

The calibration will require approximately five (5) minutes. When the calibration is complete, the blinking stops and the meter returns to the normal operating mode. Press ESC to abort a calibration session (not recommended).

Specifications

Function Range Resolution Accuracy
CO2 0 to 9999ppm 1ppm ± (5%rdg + 50ppm)
Temperature -10 to 60°C (14 to 140°F) 0.1° ± 0.6°C (0.9°F)
Humidity 0.1 to 99.9% 0.1% ± 3% (10 to 90%)

± 5% (< 10% or > 90%)

Display Multi-function LCD
Sensor Type CO2: NDIR (non-dispersive infrared) technology

Humidity: Capacitance sensor
Temperature (air): Thermistor
Response| CO2: < 2min for 90% step change
Air temperature: < 2min for 90% step change
RH: < 10min for 90% step change
Alarm relay output| Rated: 1A 30VDC or 0.5A 125VAC
Relay opens/closes as the CO2 alarm switches ON/OFF
Operating Conditions| 14 to 140°F (-10 to 60°C); < 90% RH non-condensing
Storage Conditions| -4 to 140°F (-20 to 60°C); < 99% RH non-condensing
Power Supply| 5VDC (±10%), ≥ 500mA (supplied adaptor)
Dimensions| 4.6 x 4 x 4” (117 x 102 x 102 mm)
Weight| 7.2 oz (204 g)

Maintenance

Cleaning and storage
  1. The meter should be cleaned with a damp cloth and mild detergent when necessary. Do not use solvents or abrasives.
  2. Store the meter in an area with moderate temperature and humidity.

Troubleshooting

No power Check that the AC adaptor is properly connected.
Slow response Check that rear air flow vents are not blocked.
“BAt” and green LED flash The adaptor output voltage is too high or too low.
Error Display Codes

CO2 Display Code| EXTECH CO200 - Error - CO2 Display
Code
---|---
E01| CO2 sensor malfunction.| Return for repair
E02| CO2 reading out of range (low).| Re-calibrate the meter. If problems persist, return for repair.
E03| CO2 reading out of range (high).| Place the meter in fresh air and allow 5 minutes.

If problem persists, re-calibrate the meter.

Return for service if necessary.

E17| CO2 senor failure.| Return for repair.

Temperature Display Code| EXTECH CO200 - Error - Temperature Display
Code
---|---
E02| Air temperature out of range (low).| Place the meter at room temperature for 30 minutes. If problem persists, return for repair.
E03| Air temperature out of range (high).| Same as EO2.
E31| Sensor malfunction.| Return for repair.

Humidity Display Code| EXTECH CO200 - Error - Humidity Display
Code
---|---
E04| Temperature display shows error.| Refer to temperature error codes.
E11| RH calibration failure.| Please return for repair.
E34| RH sensor circuit failure.| Please return for repair.

CO2 Level Guidelines

Important: The public information provided below is offered for reference only. FLIR Systems, Inc. is not liable for injury to persons or property through the use or misuse of this device. It is the responsibility of the user to ensure proper air quality in the residential or commercial spaces where this device is to be used.
Non-enforced reference levels:

  • 250 – 350 ppm: Normal background outdoor levels.
  • 350- 1,000 ppm: Typical level found in occupied spaces with good air exchange.
  • 1,000 – 2,000 ppm: Level associated with complaints of drowsiness and poor air.
  • 2,000 – 5,000 ppm: Level associated with headaches, sleepiness, and stagnant, stale, or stuffy air. Poor concentration, loss of attention, increased heart rate and slight nausea may also be present.
  • 5,000 ppm: Exposure may lead to serious oxygen deprivation resulting in permanent brain damage, coma, and death.

Regulatory exposure limits:

  • ASHRAE Standard 62-1989 (1000ppm): CO2 concentration in occupied building should not exceed 1000ppm.
  • OSHA (5000ppm): Time weighted average over five 8-hour workdays should not exceed 5000ppm.
  • Building bulletin 101 (Bb101) 1500ppm: UK standards for schools indicate that CO2, averaged over an entire day (i.e. 9am to 3:30 pm), should not exceed 1500ppm.
  • Germany, Japan, Australia, UK (5000ppm): 8-hour weighted average occupational exposure limit is 5000ppm.
Two-year Warranty

FLIR Systems, Inc. warrants this Extech brand instrument to be free of defects in parts and workmanship for two years from date of shipment (a six-month limited warranty applies to sensors and cables). To view the full warranty text please visit: http://www.extech.com/support/warranties.

Calibration and Repair Services

FLIR Systems, Inc. offers calibration and repair services for the Extech brand products we sell. We offer NIST traceable calibration for most of our products. Contact us for information on calibration and repair availability, refer to the contact information below. Annual calibrations should be performed to verify meter performance and accuracy. Product specifications are subject to change without notice. Please visit our website for the most up-to- date product information: www.extech.com.

Contact Customer Support

Customer Support Telephone List: https://support.flir.com/contact
Calibration, Repair, and Returns: repair@extech.com
Technical Support: https://support.flir.com

Copyright © 2021 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in any form
www.extech.com

CO200-en-US v1.3 1/21

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals