METER ProCheck Installation Guide

June 6, 2024
METER

METER ProCheck -

INTRODUCTION

Thank you for choosing the PROCHECK Handheld Reader from METER Group.

The PROCHECK is a handheld readout device for use with all soil moisture sensors and
environmental monitoring sensors made or sold by METER. Visit the Datalogger compatibility tables(metergroup.com/environment/articles/data-logger- compatibility-tables for specific details on compatible sensors. The PROCHECK interfaces with both analog and digital soil moisture sensors to obtain instantaneous results. This manual covers all PROCHECK functions and capabilities.

Verify all PROCHECK components are included and appear in good condition:

  • PROCHECK control unit
  • USB-to-serial cable
  • Four AA batteries
  • Carrying case (if purchased)
  • Pigtail-to-stereo adapter (if purchased)

OPERATION

Please read all instructions before operating the PROCHECK to ensure it performs to its full potential.

PRECAUTIONS
METER sensors are built to the highest standards, but misuse, improper protection, or improper installation may damage the sensor and possibly void the manufacturer’s warranty. Before using the PROCHECK, follow the recommended installation instructions and arrange proper protections to safeguard sensors from damage.

CONFIGURING THE PRO CHECK

Before taking measurements, configure the PROCHECK system date and time to ensure accurate timestamps. ECH2O Utility automatically synchronizes the date and time of the PROCHECK to the computer time when they are connected. Select the Set Date/Time option on the Action menu (Section 4).

Use the following steps to configure the PROCHECK controller prior to readings.

  1. Turn on the PROCHECK.

  2. Press MENU to navigate to the Configuration tab (Figure 1).
    METER ProCheck - PROCHECK

  3. Use the UP and DOWN buttons to scroll down to Date.

  4. Press ENTER.
    The system date appears in the center of the screen in a mon day, year format (Figure 2). Arrows appear above and below the first value.
    METER ProCheck - PROCHECK2

  5. Use UP and DOWN to change the first value. Holding down the arrows will scroll quickly
    between values.

  6. Press ENTER to move to the next value.

  7. Repeat step 5 and step 6 until the correct date is selected.

  8. After the last value is changed, press ENTER to return to the Configuration tab.

  9. Use DOWN to highlight Time.

  10. Press ENTER.
    The system time appears in the center of the screen, in 24-h format ( Figure 3 ). Arrows appear above and below the first value.
    METER ProCheck - PROCHECK3

  11. Use UP and DOWN to change the value.

  12. Press ENTER to move to the next value or ESC to return to the previous value.

  13. Repeat steps 11 and steps 12 to select the accurate time.

  14. After the last number is changed, press ENTER to return to the Confi guration tab.

  15. Press MENU to return to the Measurement tab.

The PROCHECK is ready to take readings (Section 2.2)

TAKING A READING

METER sensors have a serial connector that plugs directly into the PROCHECK. If the sensor has bare wire ends, consult Section 2.6 prior to taking a reading.

  1.  Press the MENU button.

  2. Use UP and DOWN to scroll through the sensor list until the correct sensor type appears in the Measurement tab ( Figure 4 ).
    METER ProCheck - PROCHECK4 NOTE: The sensor version is only shown if the PROCHECK was able to retrieve sensor information directly from the sensor.

  3.  Plug the sensor into the stereo connector at the top of the PROCHECK.
    NOTE: Sensor units and calibrations can be customized in the Configuration tab (Section 3.3.2). A soil calibration type displays only if it is available for that sensor type.

  4. Press ENTER to engage the live update mode ( Figure 5 ).
    In live update mode, the PROCHECK will take a reading every 30 s and update the screen. A pair of arrows rotate in the lower right corner during updates
    METER ProCheck - PROCHECK5

  5. Press ESC to exit the live update mode.

  6. Press SAVE to save the current reading on the display (Section 2.3).

ANNOTATING A READING

Readings can be annotated with a Sample ID, which will be saved and downloaded with that record. The PROCHECK auto-fills the last annotation stored into memory.

  1. Press SAVE to show the Annotate screen ( Figure 6).
    The right side of the screen shows a reading summary and the left side of the screen shows the annotation field as well as the date and time of the reading.
    METER ProCheck - PROCHECK6

  2. Use UP and DOWN to change the current character (A-Z, 0-9, or special characters ., -,‘, #, or space). Holding down the arrow buttons will scroll quickly between values.

  3. Press ENTER to move to the next value.
    Press ESC to return to a value.

  4.  Repeat until the last character.

  5. Press SAVE to save the annotation or MENU to return to the Measurement tab without saving the record.

 DOWNLOADING DATA

METER recommends using the ECH2O Utility to download data. This function transfers all saved measurement data on the PROCHECK to a computer.

The USB driver should automatically install with ECH2O Utility. If it does not, install the USB driver from the included USB or from downloads.metergroup.com.

Use the following steps to connect to download data.

  1. Launch ECH2O Utility.

  2. Use the supplied USB-to-serial cable to connect PROCHECK to an available USB port on the computer.

  3. Choose the correct communication port from the Connect Via dropdown list.

  4. Click Connect.

  5. Click the Download button in the toolbar to download all new data (since the last download).
    Alternatively, go to the Data menu and select Download All Data.

  6. Edit the file name in the File Save dialog.

  7. Select the location on the computer to save the data.

  8. Choose a file format.

  9. Click Save to start the download.
    ECH2O Utility will show the progress of the download.
    NOTE: Cancel a download in progress using the Cancel button. ECH2O Utility does not create a data file if the
    download is canceled.

By default, ECH2O Utility saves measurement data as an Excel® workbook file. Sheet 1 contains processed data (including calibration coefficients used to calculate data), and Sheet 2 shows raw data. Data can also be saved as .txt or .csv file format.

ECH2O Utility will download data in the units chosen in the Preferences dialog. These options apply only to the Excel workbook and Processed Data Text file formats.

Data can also be downloaded through the PROCHECK if ECH2O Utility is not functioning properly. Contact Customer Support for more information.

 DELETING DATA

CAUTION: Erasing data permanently deletes all data from the PROCHECK. Data is not recoverable.

Using the PROCHECK, follow steps 1 through 3 to erase data.

  1. Navigate to the Data tab.
  2. Select Erase.
  3. Press ENTER in the confirmation screen to erase data.
    Press ESC to cancel.
    The screen will return to the Data tab.

ADDITIONAL CONNECTING CONSIDERATIONS

In order for METER sensors to be used with non-METER loggers, the connection and SDI-12 address may have been changed. Adjustments will need to be made so the PROCHECK can read these sensors again.

METER sensor cables can be stripped and tinned to use with non-METER loggers. To connect these sensors back to the PROCHECK or other METER loggers, a pigtail-to-stereo adapter cable will be needed. Contact Customer Support.

The SDI-12 address on METER digital sensors may also need to be changed to use with a non-METER logger. The PROCHECK can be used to change the default METER SDI-12 address (0). (For more information on SDI-12, see the sensor integrator guide)

To assign the addresses, follow steps 1 through 6.

  1.  Connect the SDI-12 compatible sensor to the PROCHECK stereo port.
    NOTE: If the sensor uses bare wires, use the METER pigtail adapter to connect. Contact Customer Support.

  2. Navigate to the Configuration tab ( Figure 7 )
    METER ProCheck - PROCHECK7

  3. Select SDI-12 Address to bring up the currently connected sensor.

  4.  Use the UP and DOWN arrows to select the new address from “0” (default) to 1–9, a–z, or A–Z ( Figure 8).
    METER ProCheck - PROCHECK8

  5. Press ENTER or SAVE to accept the new address.
    The screen refreshes to show the new address setting.
    NOTE: If setting addresses on multiple sensors, repeat steps 1 through 5 before proceeding to step 6.

  6. Press ESC or MENU to exit the SDI-12 Address menu.

APPLYING A CUSTOM CALIBRATION

The PROCHECK can use custom calibration for certain sensor types. Custom calibrations apply directly to the raw analog-to-digital count (ADC) in the form of a fifth-order polynomial.
For more information on creating a custom calibration, please refer to Soil- specific calibrations for METER soil moisture sensors (metergroup.com/environment/articles/method-a-soil-specific-calibrations-for- meter-soil-moisture- sensors)

NOTE: Custom calibrations are in terms of m3/m3. PROCHECK applies unit conversions after the custom calibration

To enter a custom calibration into the PROCHECK:

  1. Navigate to the Configuration tab ( Figure 9 ).
    METER ProCheck - PROCHECK9

  2. Select Calibration Type.

  3. Select the sensor type ( Figure 10 ).
    METER ProCheck - PROCHECK10

  4. Highlight Medium ( Figure 11 ).
    METER ProCheck - PROCHECK11

  5. Press ENTER until the medium type changes to Custom.

  6.  Select Coefficients.
    The Custom Calibration screen will appear ( Figure 12 )
    METER ProCheck - PROCHECK12Each coefficient corresponds to a letter in the polynomial at the top of the screen. Below the polynomial resides the value of each coefficient in scientific notation.

  7. Highlight the desired coefficient and press ENTER.
    The Edit screen appears ( Figure 13 ).
    METER ProCheck - PROCHECK13

  8. Use UP and DOWN to change the current value, and use the left and right arrow buttons (ESC and ENTER) to change the cursor position.

  9. Press ENTER to save the coefficient.
    Press MENU to return to the coefficient list without changing the coefficient.

  10. After all the coefficients have been updated, press SAVE to save the custom calibration.
    Press ESC to return to the Calibration Type screen without saving the calibration

SYSTEM

This section describes the specifications and components of the PROCHECK.

SPECIFICATIONS
Sensor Input Ports 1
Sensor Port Type 3.5-mm stereo plug connector
Sensor Types Firmware versions up to 1.5C: All environmental, soil moisture,

and water potential sensors sold by METER before November 2012.

Firmware versions after 1.5C: All environmental, soil moisture, and water potential sensors sold by METER

NOTE: See more specifics in the Datalogger compatibility tables (metergroup.com/environment/articles/data-logger-compatibility- tables)

Display| 128 x 64 graphical
Data Storage| 1 MB (5,000 readings)
Battery Capacity| Four AA alkaline batteries
Battery Life| 500 to 1,000 h
Computer Communication| Serial RS-232 to USB
Software Interface| ECH2O Utility
Enclosure| Splash-proof polycarbonate
Enclosure Rating| IP20, NEMA 1
Enclosure Size| 15.5 × 9.5 × 3.3 (6.1 × 3.7 × 1.3 in)
Operating Environment| 5 to 50 °C (0%–100% relative humidity)
COMPLIANCE
Manufactured under ISO 9001:2015
EM ISO/IEC 17050:2010 (CE Mark)

BUTTONS

The PROCHECK has six buttons to navigate through its functions (Figure 14).

  •   MENU: Turns on and off the PROCHECK when held down. Use to cycle through the menus at the top of the screen.

  • UP and DOWN arrows: Scrolls through sensors, menu options, and characters or numbers in the editing menus.

  • ESC: Cancels setting changes and sensor readings, and scrolls to the left in the editing menus.

  •   ENTER: Starts readings, chooses options, and scrolls to the right in the editing menu.
    Press ENTER to override the 30-s auto-update and receive the most recent reading.

  • SAVE: Saves instrument and annotation settings.

METER ProCheck - PROCHECK14

SCREENS

The PROCHECK features three main menus designed for ease of use. Press MENU to navigate between the tabs and the arrow buttons and ENTER to manage selections.

 MEASUREMENT TAB

The Measurement tab is the PROCHECK home screen and appears when the PROCHECK is turned on. Use this tab to take readings ( Section 2.2 ).

 CONFIGURATION TAB

The Configuration tab lists several different options for customizing the instrument ( Figure 15 ).

METER ProCheck - PROCHECK15

UNITS SCREEN
From this menu configure the unit preferences for various unit types ( Figure 16 ).

METER ProCheck - PROCHECK16

Turning on the Show Unprocessed Data option displays any formatted data in its raw analog-to-digital converter (ADC) state. All data readings on the Measurement and Data tabs are displayed in an unprocessed format as seen on the following screen. This is helpful when troubleshooting sensors or when creating a custom calibration.

Use the arrows to navigate to the measurement type, and press ENTER to cycle through the unit options. Some unit options are only available for specific sensors.

  • Water Content: m3/m3, % VWC, IPF, cm/m, or εa
  • Water Potential: pF, kPa, or bar
  • Temperature: °C or °F
  • Humidity: aw, kPa, RH, or %RH
  • Precipitation: mm or in
  • Drainage/Water Level: mm, in, or ft
  • Volume: liters, gal (US), ml, m3, or ft3
  • EC: bulk dS/m, mS/cm, microS/cm, or pore water dS/m
  • Wind Speed: m/s, km/h, or mph
  • Gauge Pressure: psig or kPag
  • Distance: km or mi

CALIBRATION TYPE SCREEN
Some sensors support a calibration adjustment for different media types (soil, Rockwool, etc.). The first screen shows a summary of the selected soil calibrations for each configurable sensor ( Figure 17 ).

METER ProCheck - PROCHECK17Select the sensor to select its calibration, using ENTER to toggle between options ( Figure 18 ).

METER ProCheck - PROCHECK18

Both preprogrammed and custom calibration options are available ( Section 2.7 ). View the calibration coefficients by selecting Coefficients… ( Figure 19 ).

NOTE: With the GS3 sensors (only), custom calibration equations must be constructed with εa (dielectric permittivity) as the x variable.

METER ProCheck - PROCHECK19

Specific instructions on how to make a custom calibration can be found on the METER
website (metergroup.com/environment/articles/method-a-soil-specific- calibrationsfor-meter-soil-moisture- sensors).

SENSOR LIST SCREEN
The Sensor List screen customizes which sensors are available for selection on the Measurement tab ( Figure 20 ). Each checked sensor appears on the Measurement tab. Use the arrow buttons to select sensors individually or select Check all and Uncheck all to change all the sensors at once.

METER ProCheck - PROCHECK20

SDI-12 ADDRESS SCREEN
SDI-12 is a communication protocol for serial data interface at 1,200 baud. Please see sdi-12.org for more information on using SDI-12.

METER digital soil moisture sensors use both the SDI-12 standard and a METER proprietary protocol. This SDI-12 feature is not necessary when using METER sensors with METER data loggers. The SDI-12 may be useful using the sensors with third-party data acquisition systems such as Campbell Scientific data loggers. For more information, refer to the respective sensor integrator guide.

The SDI-12 Address screen is to assign the use of a single-character sensor address. This character may consist of 0 (default) to 1-9, a-z, or A-Z. This address is necessary when multiple sensors are installed at a single site. Each sensor will need a different address assigned ( Section 2.6).

BATTERY
This line displays the percentage of remaining battery power in the device. It cannot be selected.

DATE SCREEN
The Date screen displays the current set date on the handheld ( Figure 21 ). The date must be updated prior to using PROCHECK ( Section 2.1 ).

METER ProCheck - PROCHECK21

TIME SCREEN
The Time screen displays the current set date on the handheld (Figure 22). The time must be updated prior to using PROCHECK ( Section 2.1 ).

METER ProCheck - PROCHECK22

SRS α-VALUES
SRS α-values may need to be changed (see the SRS User Manual for more information). To modify the SRS α-values for both the SRS-NDVI and SRS-PRI sensors, follow the directions below.

  1. Navigate to SRS α-Values in the Configuration tab.

  2. Press ENTER.
    The SRS-PRI α-value and the SRS-NDVI α-value are displayed on the screen ( Figure 23 ).
    METER ProCheck - PROCHECK23

  3. Use UP and DOWN to select the desired SRS variant.

  4. Press ENTER.

  5. Use UP and DOWN to change the value.

  6. Press SAVE to save this value and return it to the menu.
    Press ESC to exit without saving the value.

ATMOS 41 CONFIG

The ATMOS 41 Config option allows the Lightning Strike Reject Level to be changed (see the ATMOS 41 User Manual for more information). To modify the strike reject level in the ATMOS 41, follow the directions below.

  1. Connect the ATMOS 41 to PROCHECK.

  2. Navigate to the ATMOS 41 Config in the Configuration tab.

  3. Press ENTER.
    The PROCHECK will connect to the sensor.

  4. Press ENTER to select Strike Reject Level.
    A pair of arrows will appear next to the selected level ( Figure 24 ).
    METER ProCheck - PROCHECK24

  5. Use UP and DOWN to change the current value.

  6. Press SAVE to save this value and return it to the menu.
    Press ESC to exit without saving the value.

CONTRAST SCREEN
The Contrast screen controls the screen contrast (Figure 25). Use the arrow buttons to alter the contrast level. Press ENTER to save or ESC to cancel adjustments.

METER ProCheck - PROCHECK25

ABOUT SCREEN
This menu displays the PROCHECK serial number, device name, and firmware version.

DATA TAB

This tab allows data to be viewed, downloaded, or erased ( Figure 26 ). The PROCHECK can store up to 5,000 individual readings.

METER ProCheck - PROCHECK26

The View option is a summary of the saved data records. The first screen is a summary screen ( Figure 27 ).

METER ProCheck - PROCHECK27

This page shows a data record list that includes the date and the time that the reading was taken and the annotation (if any). To view more information about an individual record, press ENTER while highlighting the record ( Figure 28 ).

METER ProCheck - PROCHECK28

The left side of the screen shows a reading summary and the right side shows the annotation, date, and time of the reading. Press ESC to return to the previous view screen or use UP and DOWN to view records in detail.

The Download option allows data to be downloaded to a connected computer ( Section 2.4 ) and the Erase option deletes all data ( Section 2.5 ).

STATISTICS TAB

The Statistics tab collects summary statistics, including average and standard deviation, from measurements taken with the PROCHECK ( Figure 29 ).

NOTE: Statistics Saved cannot be downloaded into an exported .xls file.

METER ProCheck - PROCHECK29

  1. Use UP and DOWN to select the correct sensor.

  2. Press ENTER.
    The connected sensor will show the current reading on the left side of the screen.

  3. Press ENTER to add the current measurement to the summary statistics.
    The n value represents the number of measurements taken as part of the summary statistics.

  4. Press ESC to stop live updating. The measurements will continue to be added until ESC is pressed.

  5.  Press ESC a second time to reset the statistical information.

  6.  Press SAVE to save the statistics summary.
    This does not reset the n value, so measurements can continue to be added. The SAVE button only saves the summary statistics and not the data used to calculate the values.

ECH2O UTILITY

The PROCHECK is designed to work with ECH2O Utility to collect and manage data from the device. The ECH2O Utility installer software is located on the USB that shipped with the PROCHECK.

Download the latest version of ECH2O Utility from downloads.metergroup.com. The USB driver should automatically install with ECH2O Utility. If it does not, install the USB driver from the included USB or from downloads.metergroup.com.

NOTE: The USB driver must be installed before connecting the USB cable to the computer.

Save data as an Excel workbook file, tab-delimited text file, and raw, unprocessed data file. ECH2O Utility can also be used to delete measurement data stored in the PROCHECK and set the date and time in the PROCHECK.

NOTE: The ECH2O Utility is designed to work with both the PROCHECK and METER loggers. Some controls, features, and settings of ECH2O Utility are disabled when connected to a PROCHECK.

PROCHECK works with ECH2O Utility software for Microsoft Windows®. Minimum computer requirements include:

  • Windows 7operating system
  • Available serial port (or USB-to-serial adapter)
  • Microsoft Excel 97 or newer (if saving data as .xls files)

ECH2O Utility Help file within the software provides more information.

SERVICE

This section contains calibration and recalibration information, calibration frequencies, cleaning and maintenance guidelines, troubleshooting guidelines, customer support contact information, and terms and conditions.

MAINTENANCE

The PROCHECK may be returned to METER for maintenance in the following areas: system inspection, parts replacement, and instrument cleaning. Replacement parts can also be ordered from METER. Contact Customer Support for more information.

METER repairs products with manufacturer’s defects and those under warranty at no cost to the customer. For non-warranty repairs, the customer is responsible for the cost of parts, labor, and shipping.

UPDATE FIRMWARE

ECH2O Utility ( Section 3.4 ) can be used to check for updates to the PROCHECK firmware.

  1. Connect the PROCHECK to the computer.

  2. Open ECH2O Utility.

  3. Click Help.

  4. Choose Check for Device Firmware Updates from the Help menu to compare the firmware version in the PROCHECK with the latest version available from METER.
    NOTE: The computer must be connected to the internet for this feature to work.
    A dialog will state that the firmware is current or needs to be updated.

  5.  Click OK.

  6. A downloader will be downloaded to the computer.

  7.  Close ECH2O Utility.

  8. Open Firmware Downloader.

  9. Press Start.

  10. When the program is finished, close the updater and disconnect the PROCHECK

TROUBLESHOOTING

Table 1 lists common problems and their solutions. If the problem is not listed or these solutions do not solve the issue, contact Customer Support.

Table 1 Troubleshooting the PROCHECK

Problem

| Possible Solutions
---|---
PROCHECK does not have power| Replace the batteries. Insert the batteries in the correct orientation when replacing them.
PROCHECK is not returning accurate readings| Check that the sensor stereo connector is fully plugged in to the stereo port on the PROCHECK.

Select the correct sensor on the Measurement tab. Taking measurements with an incorrect sensor type selected will create incorrect measurements.

The display reads Above Range or Below Range| Check that the sensor stereo connector is fully plugged into the stereo port on the PROCHECK

Select the correct sensor on the Measurement tab. Taking measurements with an incorrect sensor type selected will create incorrect measurements.

This message could also indicate a broken sensor. Contact Customer Support for assistance.

The display reads Invalid Response or No Response| PROCHECK cannot communicate with a digital sensor. Check that the sensor is plugged in and the correct sensor type is selected.

Press the ENTER button several times to restart the sensor measurement.

The data seems incorrect| If using a custom calibration, check the coefficients in the Configuration tab.

Select the correct sensor on the Measurement tab.

If these steps do not resolve the problem, please consult the sensor user manual.

The serial-to-USB adapter does not show in the
communication port picker| Use a different computer port.
Use a different cable.
Download the USB driver (downloads.metergroup.com)..)
Contact Customer Support.
ECH2O Utility cannot connect to PROCHECK| Try one or more of the following to make sure there is a serial
connection to PROCHECK.
• Check the serial port choice. Make sure the Connect Via dropdown menu shows the name of the PROCHECK.
• Download the USB driver (downloads.metergroup.com).
• Check that the serial cable is securely plugged into the PROCHECK and computer.
• Check the PROCHECK battery level.
ECH2O Utility loses connection with
PROCHECK| Check or replace the serial cable.

Increase the number of times ECH2O Utility tries sending commands to the PROCHECK.

Increase the Direct Connect Retries on the Communication tab of the Preferences form.

ECH2O Utility tells me there is no data available for download| There is no measurement data stored in PROCHECK (previously saved data may have been erased).

After each measurement, press the SAVE button to save data for later use.

Some of the measurement data show *| Data stored by the PROCHECK is outside the expected range for the sensor type. This could indicate a broken sensor or inaccurate sensor type selection.
Sensor data does not seem correct after download| There are many issues that impact the quality of sensor measurements. Please see the sensor user manual for help troubleshooting the sensor.

If the issue is still not corrected, contact Customer Support.

CUSTOMER SUPPORT

NORTH AMERICA
Customer service representatives are available for questions, problems, or feedback Monday through Friday, 7:00 am to 5:00 pm Pacific time.

Email: support.environment@metergroup.com
sales.environment@metergroup.com
Phone: +1.509.332.5600
Fax: +1.509.332.5158
Website: metergroup.com

EUROPE
Customer service representatives are available for questions, problems, or feedback Monday
through Friday, 8:00 to 17:00 Central European time.

Email: support.europe@metergroup.com
sales.europe@metergroup.com
Phone: +49 89 12 66 52 0
Fax: +49 89 12 66 52 20
Website: metergroup.de
If contacting METER by email, please include the following information:

Name Email address
Address Instrument serial number
Phone Description of the problem

NOTE: For products purchased through a distributor, please contact the distributor directly for assistance.

TERMS AND CONDITIONS

By using METER instruments and documentation, you agree to abide by the METER Group, Inc. USA Terms and Conditions. Please refer to metergroup.com/terms- conditions for details.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

METER User Manuals

Related Manuals